Electrical Impedance Tomography for lung recruitment monitoring Technical aspects of clinical usage and interpretation Karel Roubík, Vladimír Sobota Faculty of Biomedical Engineering Czech Technical University in Prague, Czech Republic [email protected], www.ventilation.cz

Transcript of Electrical Impedance Tomography for lung recruitment monitoring ...

Page 1: Electrical Impedance Tomography for lung recruitment monitoring ...

Electrical Impedance Tomography for lungrecruitment monitoring— Technical aspects of clinicalusage and interpretation

Karel Roubík, Vladimír SobotaFaculty of Biomedical EngineeringCzech Technical University in Prague, Czech Republic

[email protected],  www.ventilation.cz

Page 2: Electrical Impedance Tomography for lung recruitment monitoring ...

Electrical impedance tomographyMonitoring of ventilation:+ noninvasive, + bedside, + continuous, + no using ionizing radiation,+ relatively inexpensive,+ ... 

+ “modern”

Page 3: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT systems currently used in lung ventilation monitoring

EIT System Number of Electrodes Electrode Belt

Maltron Sheffield Mk 3.5 8 NO

Goe MF II 16 NO

Dräger PulmoVista 500 16 YES, rubber

Timpel Enlight 32 YES, rubber

Swisstom‐based devices 32 YES, textile

Page 4: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT systems currently used in lungventilation monitoring

Maltron Sheffield Mk 3.5 Goe MF II

Page 5: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT systems currently used in lung ventilation monitoring

Dräger PulmoVista 500 Timpel Enlight (Brasil)

Page 6: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT systems currently used in lungventilation monitoring

Swisstom BB2 Salvia Elisa 800 VIT 

Page 7: Electrical Impedance Tomography for lung recruitment monitoring ...

Electrical impedance tomography

Monitoring of ventilation:+ noninvasive, + bedside, + continuous, + no using ionizing radiation,+ relatively inexpensive,+ ... 

‐ limitations, problems with evaluation and problems with EIT interpretation

Page 8: Electrical Impedance Tomography for lung recruitment monitoring ...

Electrical impedance tomographyMonitoring of ventilation:+ noninvasive, + bedside, + continuous, + no using ionizing radiation,+ relatively inexpensive,+ ... 

The aim of the presentation is to show how EIT can be used for lung recruitment evaluation and to show possible sources of error during EIT data processing.

Page 9: Electrical Impedance Tomography for lung recruitment monitoring ...

CT scan EIT image

Contrast in the lung EIT image is directly proportional to the changes of the lung volume.These changes in EIT image can be directly recalculated to changes in the lung volume. 

COULOMBE, N. et al. Physiol. Meas. 2005, 26, pp. 401‐411 

Comparison of CT and EITLow spatial resolution of EIT—does it matter?

Page 10: Electrical Impedance Tomography for lung recruitment monitoring ...

Electrical Impedance Tomographyprinciples

Tomographic plane

Page 11: Electrical Impedance Tomography for lung recruitment monitoring ...

00

Electrical Impedance Tomographyprinciples

Principle: Conductivity of lungs is strongly affected by presence of gas. The more gas in the lungs, the higher the lung resistance.

Principle:Measurement of impedances (resistances) of the lungs at different angles with a consequent image reconstruction similar to CT.

current source

V V

drop ofvoltage

00

current source

V V

drop ofvoltage

((      ))

high resistance

U=R.I

Page 12: Electrical Impedance Tomography for lung recruitment monitoring ...

Principles of EITLow spatial resolution

• El. current preferably flows through regions with lowest resistance• Resolution towards the center of the image decreases

Further increase in number of electrodes does not improve the image resolution significantly

X‐ray tube

X‐ray detector

electrode #1

electrode #2

Page 13: Electrical Impedance Tomography for lung recruitment monitoring ...

Principles of EITWide “U.F.O.‐disc‐like” tomographic plane

• The tomographic plane is more than 10 cm wide, the width increases with depth

• Contribution of impedance changes from neighboring organs (heart, stomach)

Courtesy of 365PSD.com

Stolen from Drëger

Page 14: Electrical Impedance Tomography for lung recruitment monitoring ...

Data AcquisitionPlacement of electrodes 

(when not integrated in a belt)• Time consuming procedure

• The reconstruction algorithm assumes a certain (equidistant) position of electrodes—when electrodes are placed improperly, reconstruction errors occur

http://en.wikipedia.org/wiki/Electrical_impedance_tomography

Page 15: Electrical Impedance Tomography for lung recruitment monitoring ...

Data AcquisitionRadial Rotation of the electrode belt

The image of electrode belt may be disturbing during X‐ray imaging—temporary repositioning of the electrode belt may cause its rotation 

Animal experiment (pigs)with PulmoVista 500.

Page 16: Electrical Impedance Tomography for lung recruitment monitoring ...

Data AcquisitionRadial Rotation of the electrode belt

Comparison of regional ventilation (in %) at layer‐defined regions of interest (ROIs) at a standard image and the same image rotated by 45°

9

39

47

5

16

35

37

12

Page 17: Electrical Impedance Tomography for lung recruitment monitoring ...

Data AcquisitionInsufficient contact of one or more electrodes

Comparison of two images within the same breath cycle when the electrode placed at sternum lost its contact at maximum inspirium

Page 18: Electrical Impedance Tomography for lung recruitment monitoring ...

Data ProcessingIn time domain

0 1 2 3 4 5 6 7 8 9

‐8000

‐6000

‐4000

‐2000

0

2000

4000

Time (min)

Relativ

e Im

pedance (AU)

In spatial domainContrary to CT, evaluation of EIT image by looking at it does not work!Image analysis is required. 

Page 19: Electrical Impedance Tomography for lung recruitment monitoring ...

Data ProcessingSelection of baseline

Change of color scale during PEEP steps—baseline in the beginning 

0 1 2 3 4 5 6 7 8 9

‐2000

0

2000

4000

6000

8000

10000

12000

Time (min)

Relativ

e Im

pedance (AU)

[B]

Expirium InspiriumInspirium Expirium

Page 20: Electrical Impedance Tomography for lung recruitment monitoring ...

Data ProcessingSelection of baseline

Change of color scale during PEEP steps—baseline in the end

0 1 2 3 4 5 6 7 8 9

‐8000

‐6000

‐4000

‐2000

0

2000

4000

Time (min)

Relativ

e Im

pedance (AU)

[B]

Expirium InspiriumInspirium Expirium

Page 21: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT in spatial domainRegions Of Interest (ROIs)

Functional ROIs• The edge criteria depend on the regional ventilatory function of the lungs

PULLETZ, S. et al. Physiol Meas 2006, 27, S115‐127

Standard deviation of relative impedance change of each pixel. 

The higher the SD, the brighter the grey color.

Page 22: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT in spatial domainRegions Of Interest (ROIs)

Functional ROIs• The edge criteria depend on the regional ventilatory function of the lungs

PULLETZ, S. et al. Physiol Meas 2006, 27, S115‐127

Values of linear regression coefficient calculated in each image pixel with pixel data as dependent variable and global (average) data as independent variable. 

The higher the value of the regression coefficient, the brighter the grey color.

Page 23: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT in spatial domainRegions Of Interest (ROIs)

Arbitrary ROIs• Defined as simple geometric objects

• Serve for calculation of regional ventilation (% of total VT)

Images by authorsPULLETZ, S. et al. Physiol Meas 2006, 27, S115‐127

Layers Quadrants 1 Quadrants 2

Page 24: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT in spatial domainEIT in morbidly obese patients

Distorted EIT images from a morbidly obese patient (top) at different PEEP levels and corresponding images from a normal weight patient (bottom).

ERLANDSSON, K. et al. Acta Anaesthesiol. Scand. 2006, 50, pp. 833‐839

Page 25: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT in spatial domainCenter of Ventilation (CoV)

„The Center of Ventilation (CoV) reflects the distribution of ventilation in ventral‐to‐dorsal direction“

BLANKMAN, P. et al. Critical Care 2014, 18:R95VAN HEERDE, M. Acta Anaesthesiol Scand 2010; 54: 1248‐1256

TIV = Tidal Impedance Variation

Page 26: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT in spatial domainCenter of Gravity (CoG)

The Center of Gravity (CoG) is calculated as a weighted average of row / column sums.

RADKE, O.C. et al. Anesthesiology 2012 Jun; 116(6): 1227‐1234

Note: Authors frequently interchange the terms CoVand CoG. Thus, in some articles CoV refers to CoG.

Page 27: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT in spatial domainRegional Ventilation Delay

MUDERS, T. et al. Crit Care Med 2012, Vol. 40, No. 3

Page 28: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT in spatial domainRegional Opening / Closing Pressure

PULLETZ, S. et al. J Crit Care. 2012 Jun, 27(3):323.e11‐8

Page 29: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT in spatial domainRegional Compliance

COSTA, E. L. V. et al. Intensive Care Med. 2009, 35, pp. 1132‐1137

Estimation of regional compliance helps to determine regional hyperdistension or collapse of lungs.

Page 30: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT in time domainEvaluation of lung recruitment

0 20 40 60 80 100 120 140 160 180

0

2000

4000

6000

8000

10000

12000

Time (s)

Relativ

e Im

pedance (AU)

ΔEELI

Example of recruitment maneuver. The difference in End‐Expiratory Lung Impedance (EELI) corresponds to the recruited parts of the lungs.

Page 31: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT in time domainEvaluation of lung recruitment

0 20 40 60 80 100 120 140 160 180

0

2000

4000

6000

8000

10000

12000

Time (s)

Relativ

e Im

pedance (AU)

ΔEELI

Example of recruitment maneuver failure. 

Probably insufficient PEEP unable to keep the lungs open.

Page 32: Electrical Impedance Tomography for lung recruitment monitoring ...

EIT ambiguitiesContinuous drift not caused by atelectasis

0

20

40

60

ROI (%)

0 5 10 15 20 25 30 35 40 45 50

0

2000

4000

6000

8000

10000

Time

Relativ

e Im

pedance (AU)

ROI 2

ROI 3

ROI 1ROI 2

Page 33: Electrical Impedance Tomography for lung recruitment monitoring ...

Conclusion• EIT is an interesting imaging monitoring¹modality offering real time information about lung ventilation. Data processing is necessary.

• EIT data always have to be interpreted depending on the actual situation and procedure (since EIT is a functional imaging).

• Due to the complex processing and recent fast  development, a ‘Gold Standard’ of data processing and visualization is still missing.

¹ based on imaging

Page 34: Electrical Impedance Tomography for lung recruitment monitoring ...

Conclusion• EIT is an interesting imaging monitoring modality offering real time information about lung ventilation. Data processing is necessary.

• EIT data always have to be interpreted depending on the actual situation and procedure (since EIT is a functional imaging).

• Due to the complex processing and recent fast  development, a ‘Gold Standard’ of data processing and visualization is still missing.

Thank you for your attention!