A.Adler, Mar 2008 Electrical Impedance Tomography 1

BIOM5200 – Medical Imaging

Electrical Impedance

Tomography:

Image Algorithms and

Applications

Andy AdlerAssistant Professor, SCE, Carleton U

Page 2: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 2

Outline

• Electrical Impedance Tomography

• Applications

• Physics

• Image Reconstruction

• Future Work

Page 3: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 3

Electrical Impedance Tomography

• Relatively new medical imaging

technique (early 1990’s)

• Body Surface Electrodes apply current

patterns and measure the resulting

voltages

• Distribution of conductivity is calculated

Page 4: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 4

EIT: Block Diagram

+

+

CurrentSource

Amplifiers

Controler

Dataacquisition

Medium

+

Imaging SystemΩ

Page 5: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 5

Electrode placement to monitor

the lungs and heart

Adult Preterm infant

Page 6: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 6

EIT: Applications

EIT can image/monitor processes involving movement of conductive fluids and gasses

• Lungs

• Heart / perfusion (blood flow)

• GI tract

• Brain

• Breast

Page 7: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 7

Application: Breathing

Chest images of tidal breathing in normal

Page 8: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 8

Application: Heart Beat

EIT signal in ROI around heart and ECG

Page 9: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 9

Why image lungs?

Respiratory FailureInadequate gas exchange by the respiratory system.

Hypoxemia PaO2 < 60 mmHg or Hyercapnia PaCO2 > 45 mmHg

Causes

• Pulmonary dysfunction – Asthma ,Emphysema , Chronic obstructive airway disease,

Pneumonia , Pneumothorax, Hemothorax, Acute Respiratory Distress Syndrome (ARDS), Cystic Fibrosis

• Cardiac dysfunction – Pulmonary edema, Arrhythmia, Congestive heart failure, Valve

pathology

Treatment– Emergency treatment: cardiopulmonary resuscitation.

– Treatment of the underlying cause is required.

– Mechanical ventilation may be required.

Page 10: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 10

Mechanical Ventilationused in acute settings (ICU). Often a life-saving technique, but has

many complications – pneumothorax,

– airway injury,

– alveolar damage,

Accordingly it is generally weaned off or to minimal settings assoon as possible.

Positive pressure in contrast to the more historically common negative pressure ventilators sucking air into the lungs.

Iron Lung

Page 11: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 11

Modes of Ventilation

classifications based on how to control the ventilator breath.

• Breath termination

• Breath initiation

• High Frequency Ventilation (HFV)

As microprocessors are incorporated into ventilator design, ventilators use combinations of all modes and flow-sensing

Page 12: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 12

Why image lungs?PneumoniaA: Normal chest x-ray.

B: Abnormal chest x-ray

shadowing from

pneumonia in the right lung

Page 13: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 13

FRC

Static Mechanics: ventilation

Stiffer lung

Stiffer lungs have decreased

resting volume (FRC)

Page 14: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 14

Regional ventilation

Data from

Frerichs et al (2003)

Intensive Care Med..

Page 15: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 15

No change in

lower lungs

until pressures

get really big

Page 16: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 16

Applications: Brain

Applications

• Hemorrhage

• Localization of epileptic foci

Newborn with EIT electrode cap on head

Page 17: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 17

Industrial Applications

Process Tomography

• Fluid/gas flow in pipes

• Metal Castings

Geophysics

• Undersurface geology

• Mine detection

Page 18: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 18

EIT: Advantages

EIT is a relatively low resolution imaging

modality, with several advantages

• Non-invasive

• Non-cumbersome

• Suitable for monitoring

• Underlying technology is low cost

Page 19: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 19

Non-invasive

0.1

1

10

100

30 100 300 1K 3K 10K 30K

Frequency (Hz)

Current

(mA)

Percentile 99

Percentile 50

Percentile 1

EIT system

Thresholds for cutaneous perception of electric current vs. frequency and EIT system

Page 20: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 20

Hardware: Electrodes

• Current stimulation is better than voltage, because it accounts for electrode contact impedance

• Traditionally EIT uses adjacent current drive.

• Some systems separate drive and measurement electrodes, using adaptive current patterns

Page 21: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 21

EIT: Physics

• Within medium Ω there is E and J.

dt

dod

c

EJ

EJ

εε

σ

=

=

( )EJ ojωεεσ −=

Page 22: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 22

EIT: Physics

In the absence of magnetic fields

No charge build up in conductive medium

We have

V−∇=E

0J =−=•∇t∂

∂ρ

( ) Ω=∇−•∇ in 0Vj oωεεσ

Page 23: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 23

EIT: Physics

Current is applied at electrodes

Body need to be grounded, somewhere

eIt

=−=•∇∂

∂ρJ

point someat 0=V

Page 24: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 24

EIT: Numerical Models

In order to calculate measurements from

conductivities, we can use:

• Analytic Techniques

– Analytic models exist for elliptic 3D media; however, numerical approximations of sums required

• Numerical Models

– Finite Element Techniques, main method

Page 25: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 25

Finite Element Models

Simple Model

with 64 elements

Used for inverse solution

-1 -0.5 0 0.5 1

-1

-0.8

-0.6

-0.4

-0.2

0

0.2

0.4

0.6

0.8

1

Page 26: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 26

Model of Borsic, Physiol Meas, 22:77-83, 2002

Page 27: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 27

-1-0.5

00.5

1

-1-0.5

00.5

1

-0.25

-0.2

-0.15

-0.1

-0.05

0

0.05

0.1

0.15

0.2

0.25

Finite Element Models

“Simple” 3D Model with

768 elements

Used for inverse solution

Page 28: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 28

Image Reconstruction:Static Imaging

Static imaging reconstructs the absolute

conductivity from measurements.

Algorithms:

• Iterative (Newton-Raphson)

• Layer Stripping

Page 29: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 29

CurrentInjection

Patient

and

EIT datameasurement

Real Data

Finite Element Model

EIT data

simulation Simulation Data

Does simulationapproximate

real data

Finishedyes

Update

conductivity

distribution

no

Iterative (Absolute) Image

Reconstruction

Page 30: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 30

Absolute Imaging Difficulties

• Extremely sensitive to uncertainties in

electrode position

– Need to know where electrodes are to and electrode shape to 1mm

– “Absolutely” must do 3D

• Numerical instability

• Slow reconstructions

• Is muscle in chest isotropic?

Page 31: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 31

Difference

Imaging: Example

Page 32: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 32

Difference Imaging

• Calculate ∆ conductivity

from ∆ measurements

• Inverse problem linearized

• reduced sensitivity to electrode and

hardware errors.

• Suitable for physiological imaging: lung,

heart, GI

Page 33: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 33

Inverse Techniques

• We can pose dynamic imaging as linear

inverse, using a sensitivity matrix

j

jhhj

δ

δσσ )()( +−=

zzz

σ∆= Hz

Page 34: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 34

Inverse Techniques

• Classic least-squares inverse

( ) zHHHx

Hxz

tt 1ˆ

−=

=

Page 35: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 35

Model based matrix inverses

Page 36: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 36

Matrix Techniques

However, problem is:

• ill-conditioned: measurements depend much more on data near electrodes

than in centre

• ill-formed: more unknowns than

measurements

Page 37: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 37

Regularized Imaging

Handwaving argument for regularization:

used for ill-posed and ill-formed problems to find a solution with:

• Low error: small ( z – Hx )

• Stable: small change in x for small ∆z

• Good looking:

– Somewhat hard to define, but includes

smoothness, clean edges, etc.

Page 38: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 38

MAP estimates

• MAP approach says choose x such that

f(x|z) is maximized

– In other words, choose the image that is most likely, considering the measured data

• Bayes Rule

( ) ( ) ( )( )z

xxzzx

f

fff =

Page 39: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 39

MAP estimates

f(z|x) the distribution of measurements given an image– Based on forward model and noise

properties

f(z) distribution of measurements– Not a parameter of MAP estimate

f(x) distribution of image– Based on a priori knowledge of physically

possible and likely images distributions

Page 40: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 40

Regularized Imaging

Given Linear Model:

Maximum A Posteriori (MAP) estimate is:

( ) ( )∞−−−−− ++=

+=

xRRHRHRHx

nHxz

11111ˆ xnt

xnt

z

Page 41: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 41

Image Reconstruction

• Forward Model (linearized)

= ×

Measurements(difference)

Image(difference)

Jacobian

System is underdetermined

+ noise

Page 42: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 42

Image Reconstruction

Regularized linear Inverse Model

– ×

Norm weighted by

measurementaccuracy

+ PenaltyFunction

2

Way to introduce

“prior knowledge”into solution

Page 43: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 43

Regularized Imaging

• Parameters Rx, Rn, x∞ , represent a priori

statistical knowledge of problem

[ ]

( ) ( )[ ] [ ]

[ ]

==

−==

=

∞∞∞∞

OM

L

22

21

0

0

σ

σ

nnR

xxxxx-xx-xR

x x

n

x

t

ttt

E

EE

E

Page 44: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 44

Choice of parameter Rx

• Parameter is a “penalty function”

• Many regularization approaches use a diagonal matrix– Tikhonov regularization uses the scaled

identity matrix

– This will penalize large amplitude pixels in image

• We choose a dense matrix– Penalize image frequency content above

maximum possible with measurements

Page 45: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 45

Regularization: Hyperparameters

Regularizations techniques must finally

introduce a “hyperparameter” (µ)

( ) ( )

nscorrelatio image relative the ie. ,

amplitudes noise relative the ie. ,

1

1

=

=

++=

x

n

RQ

RW

QxWzHQWHHx

2

2

1

1

1

ˆ

x

n

tt

where

σ

σ

µµ

Page 46: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 46

Regularization: Hyperparameters

µ is thus the ratio of image and noise

amplitudes,

it can be interpreted as a filter noise figure

2

2

y

x

σ

σµ =

Page 47: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 47

Regularized Inverse

Parameters:

• W: models measurement noise

• Q: penalizes image features which are greater than data supports

• : represents the background conductivity distribution (heart,lungs,etc)

• µ: “hyper-parameter” amount of regularization

∞x

Page 48: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 48

Advantages of Regularization

• Stabilizes ill-conditioned inverse

• Introduction of a priori information

• Control of resolution-noise performance

trade-off

• MAP inverse justifies the formulation in

terms of Bayesian statistics

Page 49: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 49

Noise – Resolution Tradeoff

LittleRegularization

(small penalty)

No Noise

-3dB

SNR

Lots of

Regularization(large penalty)

Page 50: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 50

Applications …

• Electrode Errors

• Electrode Movement

• 3D Imaging / Electrode Placement

• Temporal Filtering

Page 51: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 51

Electrode Measurement Errors

Experimental measurements with EIT quite often show large errors from

one electrode

Causes aren’t always clear

– Electrode Detaching

– Skin movement

– Sweat changes contact impedance

– Electronics Drift?

Page 52: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 52

Example of electrode errors

Images measured in anaesthetised, ventilated dog

A. Image of 700 ml ventilation

B. Image of 100 ml saline instillation in right lung

C. Image of 700 ml ventilation and 100 ml saline

A B C

“Bad”

Electrode

Page 53: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 53

Measurements with “bad” electrode

“Bad”Electrode

2

3

4

5

6

7

0

1

7067564534231201

XX∗∗∗∗∗∗∗∗X70

XXX∗∗∗∗∗∗∗∗67

XXX∗∗∗∗56

∗∗∗∗∗∗∗∗XXX∗∗∗∗∗∗∗∗∗∗∗∗45

∗∗∗∗∗∗∗∗∗∗∗∗XXX∗∗∗∗∗∗∗∗34

∗∗∗∗XXX23

∗∗∗∗∗∗∗∗XXX12

X∗∗∗∗∗∗∗∗XX01

∗∗∗∗ “bad” measurement

X measurement at current injection

Page 54: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 54

“Zero bad data” solution

“Traditional solution” (in the sense that

I’ve done this)

– ×

2

σ1

σ2

σ3

σ4

-2

-2

-2

-2

Error Here

Replace

With zero

Page 55: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 55

Regularized imaging solution

Electrode errors are large measurement

noise on affected electrode

– ×

2

σ1

σ2

σ3

σ4

-2

-2

-2

-2

Error Here

Low

SNR

here

Replace

With zero

Page 56: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 56

Simulation

Data simulated with 2D FEM with 1024 elements

– not same as inverse model

“Bad”Electrode

Position in % of medium diameter

Small

targets simulated at

different

radial positions

Page 57: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 57

Simulation results for opposite driveNo Electrode Errors

Zero Affected Measurements

Regularized Image

Page 58: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 58

How does this work with real data?

A. Image of 700 ml ventilation

B. Image of 100 ml saline instillation in right lung

C. Image of 700 ml ventilation and 100 ml saline

A B C

“Bad”

Electrode

Page 59: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 59

Electrode Movement

Electrodes move

• with breathing

• with posture change

Simulations show broad central artefact in images

A B

Page 60: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 60

Imaging Electrode Movement

• Forward model image includes

movement

= ×

Jacobiannow includes

measurement changedue to movement

+ noise

“image” nowincludes

x,y sensormovement

Page 61: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 61

Images of electrode movement

Simulation: tank twisted in 3D

y x

Page 62: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 62

Top

slice

Middle

slice

Bottom

slice

Simulation Standard

Algorithm

with electrode

movement

Page 63: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 63

EIT makes fast measurements.

Can we use this fact?

……

0-1 +2 +n+1-2-n

past now future

=

Jacobian

Image sequenceMeasurement

sequence

0-1 +2 +n+1-2

past now future

Page 64: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 64

Temporal Reconstruction

Temporal Penalty Functions

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1 1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

likely quite likely unlikely

Standard EIT approaches to not take this into account

Page 65: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 65

GN vs. Temporal Inverse

Gauss-Newton solver

Solve time = 5.33 s

(with caching) = 0.22 s

Temporal solver

(4 time steps)Solve time = 34.81 s

(with caching) = 0.60 s

1. Noise free data (IIRC tank)

2. Data with added 6dB SNR noise

Page 66: Electrical Impedance Tomography: Image Algorithms …€¦ · A.Adler, Mar 2008 Electrical Impedance Tomography 1 BIOM5200 – Medical Imaging Electrical Impedance Tomography: Image

A.Adler, Mar 2008 Electrical Impedance Tomography 66

Gauss Newton vs. Temporal

Inverse (6db SNR)

Gauss-Newton solver

Solve time = 5.33 s(with caching) = 0.22 s

Temporal solver(4 time steps)

Solve time = 34.81 s(with caching) = 0.60 s