Presentation Title Arial 28pt Bold Agilent Blue · 1 2009 - Stimuli article „Transfer of...
Transcript of Presentation Title Arial 28pt Bold Agilent Blue · 1 2009 - Stimuli article „Transfer of...
July 22, 2011Confidentiality Label
1
Infinitely betterinstrument to instrument method transfer
Agilent 1290 Infinity withIntelligent System Emulation Technology (ISET)
Christian GotenfelsProduct Manager
Agenda
• Why is instrument to instrument method transfer soimportant?
• Parameter affecting instrument to instrument method transfer
• What is ISET?
• Performance examples
• Comparison with current solutions
• Summary
Page 3
Why is instrument to instrument method transfer so important? - Historical view
July 22, 2011Page 4
2004 - 483 incidence observed by the FDA 1
2009 - Stimuli article „Transfer of Analytical Procedure:A Proposal for a New General Information Chapter 2
2011 - Proposal for a New General USP chapter <1224>
1 Johnson Y, „An FDA perspective on method transfer on chromatographic analytical methods in the pharmaceutical industry,“ Chrom. Forum Delaware, October 2004
2 Oscar Quattrocchi, Greg Martin Dennis Runser, Robert Iser, Fred Xi, Horacio Pappac, “Transfer of Analytical Procedures: A Proposal for a New General Information Chapter,” Pharmacopeia Forum, Vol. 35(5) [Sept.- Oct.2009]
Method Transfer Workflow
Page 5
Location A Regulated environment
Discovery Quality Control
Location B Regulated environment
Discovery Development Quality Control
Development
Method Transfer Workflow
Page 6
Location A Regulated environment
Discovery Quality Control
CRO
Development
Development
Method Transfer Workflow
Page 7
Location A Regulated environment
Discovery Quality ControlDevelopment
CMO Regulated environment
Quality Control
Method Transfer Workflow
Page 8
Methods from Literature: Scientific journals, applicationnotes, etc
Instrumentation:1100 Series, 1200 Series, Alliance, Acquity, etc
?
Design Differences (U)HPLC Systems
• Delay volumes(gradient formation)
• Power range(flow x pressure)
• Extra column volume
• Temperature
• Data rates
• Sensitivity
Page 9
Problem:Instrument to Instrument Method Transferability• Not possible• Requires Re-development• Requires Re- validation
Resulting in huge additional:cost factor:
Instrument to Instrument Method Transferability- Important Parameters
Page 10
Pump
Sampler
ColumnThermostat
Detector
Del
ay v
olum
eE
xtra
col
umn
volu
me
Retention TimeResolution
Retention TimeResolutionSensitivity
Retention TimeResolution
ResolutionRetention TimeSensitivity
Delay volumeGradient mixing behavior Pressure x flow rate
Delay volumeExtra column volumeInjection volume
Temperature profileExtra column volume
Data rateExtra column volumePath-length
Method Transferability: 1260 Quat / 1220 GradFull HPLC Methods Compatibility – Riskless Replacement
1100 / 1200 Series Quaternary system
Column: 4.6 x 150 mm, 5 µmGradient: 0 min 20 % B
10 min 95 % Flow rate: 1 mL/min
min0 2 4 6 8 10
Run existing conventional HPLC methods and get same result !
Page 11
1260 / 1220 Infinity LC Quat / Gradient system
1100/1200 Series HPLC
1260 InfinityQuaternary LC
1200 Series RRLC – Std.
1260 InfinityBinary LC
Max Flow Rate 5 mL/min 5 mL/min 5mL/min 5 mL/minDelay Volume 900-1200 µL 900-1200 µL 900-1200 µL* 900-1200 µL*Capillary ID 0.17mm 0.17mm 0.17mm 0.17mmDisp. Vol. w/o cell 15µl 15µl 15µl 15µlInjection Principle Variable Loop Variable Loop Variable Loop Variable LoopInj.Volume – Std/Ext. 100 / 1500 µL 100 / 1500 µL 100 / 1500 µL 100 / 1500 µLArea RSD <0.25 % <0.25 % <0.25 % <0.25 %Oven Design A A A AColumn Length 300 mm 300 mm 300 mm 300 mm
* Smaller delay volumes possible
Seamless Method Transfer 1260/1220 InfinityBy unchanged critical specification
Page 12
Optimized for 3 – 4.6mm ID
Columns
1100/1200 Series HPLC
1260 InfinityQuaternary LC
1200 Series RRLC – Std.
1260 InfinityBinary LC
1260 InfinityBinary LC
1290 InfinityBinary LC
Max Flow Rate 5 mL/min 5 mL/min 5mL/min 5 mL/min 5 mL/min 5 mL/minDelay Volume 900-1200 µL 900-1200 µL 900-1200 µL* 900-1200 µL* 340 µL** 10 – 110 µL*Capillary ID 0.17mm 0.17mm 0.17mm 0.17mm 0.12mm** 0.12mm*Disp. Vol. w/o cell 15µl 15µl 15µl 15µl 7.5µL** 7.5µL*Injection Principle Variable Loop Variable Loop Variable Loop Variable Loop Variable Loop Variable LoopInj.Vol. – Std/Ext. 100 / 1500 µL 100 / 1500 µL 100 / 1500 µL 100 / 1500 µL 100 / 1500 µL 100 µLArea RSD <0.25 % <0.25 % <0.25 % <0.25 % <0.25 % <0.25 %Oven Design A A A A A AColumn Length 300 mm 300 mm 300 mm 300 mm 300 mm 300 mm
* Smaller delay volumes possible
Seamless Method Transfer What if critical specification change ...? * Optimized for 2.1 - 4.6mm ID
** Optimized for 2.1mm ID
July 22, 2011Page 13
Gradient Analyses with Different LC Systems- Impact of delay volume and mixing behavior
Page 14
2 4 6 8 10 12
mAU
0
50
100
150
200
2
.74
7
3.0
49 8
.12
1 8
.32
3
8.7
47
8.9
30
9.5
45 9
.97
5
12
.00
5 1
2.2
15
12
.39
3
2 4 6 8 10 12
mAU
0
50
100
150
200
DAD1 C, DAD1B, DAD: Signal B, 275 nm/Bw:4 nm Ref 400 nm/Bw:60 nm (B:\RIC\AGI...A\GV 1PHARMA\HPLC000001.D)
2.7
44 2
.88
4 6.0
88
6.3
28
6.6
73
6.8
72
7.4
80
7.9
23
9.9
37
10
.24
9
1200 Series LC Quatenary Pump
1290 Infinity LC
Sample: 0.5% impurities in formulation (metoclopramide)Xbridge C18, 150x3 mm, 3.5 µm dp, 0.45 ml/min, Eluent: A =0 .25 % AmAc, B = ACN, Gradient: 0-15 min; 5-57 % B
The result:• Difference in RT and Resolution• One peak is missing!
Instrument to Instrument Method Transferability- Impact of delay volume and mixing behavior
What happens with a programmed gradient?
Page 15
time
90%
10%
response
Programmed gradient step
1200 Series Binary Pump SL
1290 Infinity Binary Pump
1200 Series
1290 Infinity
Total delay volume of the system(sum of capillaries, mixer, cells, valves..)
Delayvolume
Gradient slope/Mixing behavior
July 22, 2011Page 16
Intelligent System Emulation Technology
(ISET)The Solution
NEW
What is ISET?- Two components
Page 17
A. Best in class performance of the 1290 Infinity LCBroad power range, ultra-low delay volume, superior sensitivity, unmatched flow and composition accuracy
What is ISET?- Two components
Page 18
A. Best in class performance of the 1290 Infinity LCBroad power range, ultra-low delay volume, superior sensitivity, unmatched flow and composition accuracy
B. The revolutionary ISET emulationalgorithmFull characterization of target LC systemsCreation of emulation function
FA
Peak 3Overlay of 10 runs
RSD = 0.005 %
Method Transferability: 1290 Infinity LC- System Emulation Technology
Page 19
Concept
EmulatedSystem A
FA
Select System to be emulated by a simple mouse click:
Agilent 1100 Series
Agilent 1200 Series LC
Agilent 1260Infinity LC
Agilent 1220Infinity LC Other
Intelligent System emulation technology- How it works (1290 to 1200)
Page 20
Programmed gradient
1290 gradient
1200 gradient
0 0.5 1 1.5 2 2.5 3 3.50
50
100
150
200
250
300
350
400
min
mAU
Injection
Same gradient conditions by moving the gradient and by emulating the gradient mixing behavior.
July 22, 201121
Intelligent System Emulation Technology ISET- How to?
Select pump
Select autosampler
Chromatographic Evaluation: Conditions
Instruments:(A) 1100 quaternary pump (G1311A) + G1367A autosampler,(B) 1290 Infinity LC system (C) 1290 Infinity system with ISET
Sample: 1 Atrazindesethyl, 2 Hexanzinon, 3 Metoxuron, 4 Cyanazin, 5 Methabenzthiazuron, 6 Chlortoluron, 7 Atrazin, 8 Diuron, 9 Metobromuron, 10 Metazachlor, 11 Sebuthylazin, 12 Nifedipin, 13 Terbuthylazin, 14 Linuron, 15 Prometryn, 16 Nimodipin
Conditions:Gradient: Water-Acetonitrile , 20 – 60% in 8 min,Flow rate: 0.8 ml/min, Column: 2.1x100 mm Zorbax Eclipse Plus, 1.8 µm column.
Method transfer, pesticide example2.1x100 mm Zorbax Eclipse Plus, 1.8 µm column, flow: 0,8 mL/min
July 22, 2011Page 23
(C)1290 Infinity
1200 barwith ISET
(A)1100
Quaternary400 bar
(B)1290 Infinity
1200 bar
1 2,3 4 5 6,7 8 9 10,11,12 13 14 15 16
Method transfer, pesticide example Retention times and ΔRT
April 201024
A B1290 Infinity LC
1100 Quaternary
1290 Infinity LC with ISET
PesticidesColumn: 3 x 50 mm, 2.7 µm Poroshell Run time: 3.5 min
25
min1 1.5 2 2.5 3 3.5
mAU
0
100
200
300
400
500
600
700
800
1290 System
Rs= 1.90
1100 Binary System
Rs= 1.15
Sim
azin
e
Chl
orid
azon
e
Chl
orto
luro
n
Diu
ron
Pro
pazi
neTe
rbut
ylaz
ine
Pro
met
ryn
1290 System with iSETM
etam
itron
Cya
nizi
ne
Rs= 1.87
PesticidesColumn: 3 x 50 mm, 2.7 µm Poroshell Run time: 3.5 min
26
PesticidesColumn: 3 x 50 mm, 2.7 µm Poroshell Run time: 3.5 min
27
Intelligent System Emulation Technology ISET- Benefits
Achieve higher productivity for method developmentSpeed up your method development with UHPLC performance and then fine-tune your method by emulating the target system – and be confident that the method will run as intended
Simplify your instrument to instrument method transferNo more method transfer problems! Simply emulate the LC system on which the original method was developed –with a single mouse click.
Minimize your instrument-related costsRun your legacy methods with ISET and at the same time take full advantage of the UHPLC speed, resolution and sensitivity of the1290 Infinity LC. No need to maintain your old legacy LC systems.
Page 28
Instrument to Instrument Method Transferability- Comparison to current solution?
• From a lower to higher dwell volume LC systemA) Adding physical volume (plumbing solution)B) Modifying your gradient table with isocratic hold
• From a higher to lower dwell volume LC systemC) Delay the injection
Page 29
Method Transfer Possibilities Today- A) Adding physical volume (plumbing solution)
Page 30
1290 + 1000 μL dwell vol.
1100/1200 Quat Pump
Example: 1290 Infinity LC
Method Transfer Possibilities Today- A) Adding physical volume (plumbing solution)
Page 31
1290 + 1000 μL dwell vol.
Example: 1290 Infinity LC
Advantages• Works well for not so fast analysis• No Method change
Disadvantages• Requires hardware change, no flexibility• Hardware change difficult in validated environment• Requires manual determination of the dwell
volume (Pumps with damper the dwell volume is not constant and depends on pressure)
1100/1200 Quat Pump
Method Transfer Possibilities Today- B) Isocratic Hold
Example: 1290 to 1200
Page 32
Injection
0 0.5 1 1.5 2 2.5 3 3.50
50
100
150
200
250
300
350
min
Isocratic hold: move the gradient
Programmed gradient
1290 gradient
1200 gradient
mAU
Isocratic hold: move the gradient
Method Transfer Possibilities Today- B) Isocratic Hold
Example: 1290 to 1200
Page 33
Programmed gradient
1290 gradient
1200 gradient
Injection
0 0.5 1 1.5 2 2.5 3 3.50
50
100
150
200
250
300
350
min
mAU
Isocratic hold: move the gradient
Does not consider the gradient mixing behavior (different curves).
Method Transfer Possibilities Today- B) Isocratic Hold
Example: 1290 to 1200
Page 34
1200 Series LCBinary Pump
1290 Infinity LC+ 860 µl “Delay Volume”
1290 Infinity LC+ 900 µl “Delay Volume”
1290 Infinity LC+ 940 µl “Delay Volume”
Method Transfer Possibilities Today- B) Isocratic Hold
Example: 1290 to 1200
Page 35
1260 Infinity LC
1290 Infinity LC+ 900 µl hold
Advantages• Flexibile, Isocratic Hold is part of the method
Disadvantages• Works not in all cases• Requires manual determination of the dwell volume/
isocratic hold (Pumps with damper the dwell volume is not constant and depends on pressure)
• Requires modification of the methods (should be avoided in validated environment,but doesn’t require revalidation USP Chapter <621>)
LC Method Transfer Possibilities Tomorrow- Intelligent System Emulation Technology (ISET)
Page 36
Advantages• Simple and flexible
(select the emulated LC system simply by a mouse click)• No determination of the dwell volume etc. required
(perfectly covers pumps with damper and not constant dwell volume)• Works perfectly in all cases• Requires no modification of the methods
(no conflicts in regulated environment)
Agilent 1290 Infinity LC with ISET inside - The Art of Emulation
Emulates other (U)HPLC instruments – by a simple mouse click
Runs existing (U)HPLC methods– without modifying method or system
Delivers same retention times and peak resolution – for infinitely better method transfer
37
Thanks for your attention!!!
Questions?
Page 38