AZPES/APP Permit Field Method Training
Fall 2016
Not Approved Methods
Color Wheels
Test Strips
Pool Kits
Approved Methods
Standard Methods – 22nd Edition -4500-Cl B (2011) - Iodometric Method I
-4500-Cl C (2011) - Iodometric Method II
-4500-Cl D (2011) - Amperometric Titration
-4500-Cl E (2011) - LL Amperometric Titr.
-4500-Cl F (2011) - DPD Ferrous Titration
-4500-Cl G (2011) - DPD Colorimetric*
Most other meters by non-Hach companies are also acceptable
Calibration
Most meters (including Hach) are
factory calibrated and do not need
routine calibration.
General Procedure
General Procedure
General Procedure
*
General Procedure
General Procedure
Method Required QC Method Blank (or RB): 1 daily. If at or above Reporting Level, immediate corrective action. [DI water + DPD] LFB: At reporting level (control charts 1020B.13 or ADHS will allow recovery ± 10%). At least monthly (Goal – each run). [ADHS will allow 2° Stds, e.g. Spec√s APP Permits only] Duplicates: 1 for each matrix type daily or with each batch of 20 or fewer samples. Control Charts 1020B. Need Record for Each
NSI Solutions, Inc.
HACH
0.5 ml Pipetman
(0.5 ml of standard added) X (25-30 mg/L Cl2)
10.5 ml (sample cell volume + standard added)
= 1.19-1.43 mg/L
(12 ml sample cell volume + 0.5 ml standard added = 1.00-1.2 mg/L)
Formula for Hach Standard