LT5570 - Fast Responding, 40MHz to 2.7GHz Mean … · en dnc dnc out vcc in+ dec in ...
University of Wisconsin-Madison School of Veterinary Medicine€¦ · 2018-06-06 · 40MHz HRUS...
Transcript of University of Wisconsin-Madison School of Veterinary Medicine€¦ · 2018-06-06 · 40MHz HRUS...
Advanced Ultrasound Techniques
Ellison Bentley, DVM, Diplomate ACVOUniversity of Wisconsin-Madison
School of Veterinary Medicine
High resolution ultrasound• Ultrasound biomicroscopy
(UBM) and high resolution ultrasound (HRUS)– Low power histology-like
view• Resolution-20-80µm
– 10MHz-300-400µm• Tissue penetration limited
to 5-10mm40MHz HRUS image canine
High resolution ultrasound• Clinical Uses: anterior
segment evaluation, uveal tumors, cysts, scleral and corneal disease, drainage angle, changes in these structures in disease
• Research: effect of drugs on anterior segment structures, pathophysiology of anterior segment diseases
HRUS(20MHz) image of canine eye
Cornea
IrisCB
Lens
Ciliary cleft
• HRUS correlates with histo
40MHZ
UBM vs HRUS
UBM• 50MHz-higher
resolution• Smaller field of view• Less penetration• Positioning critical• Sedation/anesthesia• Higher cost
HRUS• 20MHz - 40 MHz-lower
resolution• Larger field of view• Deeper penetration• Easier positioning• Topical anesthesia• Lower cost
UBM vs HRUS
50 MHz
10 MHz20
MHz
Courtesy C. Kendall, Innovative Imaging, Inc.
UBM vs HRUS
UBM image High resolution ultrasound image40MHz probe
• Not all probes created equal
50MHz
50MHz
20MHz
40MHz
Technique• Topical anesthetic• Genteal Gel/Optixcare• Superior and temporal quadrants easiest
– Ventral and nasal may require sedation
Probe position-longitudinal
Can label by clock hour
40MHz 12AX
Probe position-transverse
Note ciliary processes20MHz
40MHz 12T
Cornea
• Corneal thickness assessment
• Sequestrums-deeply pigmented lesions hyperechoic
• Deeper, faintly stained area not visible
• Contralateral eye: control
Sequestrum
Hyperechoic sequestrum‘Normal’ cornea=0.27
Normal contralateral eyeCorneal thickness=0.58mm
Corneal tumors
• Corneal tumors in horses– Sometimes difficult to assess depth
• Limbal melanomas with corneal extension in dogs
Hemangioma in a horse
Uvea
• Useful to differentiate tumors versus cysts• Useful to evaluate tumor extension • Useful to determine which patients are good
candidates for laser therapy for tumors; progression
• Can use to determine location of ciliary processes for cyclophotocoagulation
Uveal cysts
2 yr FS Golden retriever Note bilobed uveal cysts
Uveal cysts
10 yr old DSH referred for ciliary body tumor
17 year cat referred for ciliary body tumor
20MHz
20MHz
Feline cysts-COPLOW
• 15 eyes removed for suspected neoplasia
• 14/15=uveal cysts• 1/15-PIFMFeline cysts masquerading as neoplasia in catsFragola, Dubielzig, Bentley, Teixeira, VO 2018
Uveal tumors
4 yr MN mixed CNPre laser=0.80mm
Immediatepost laser=1.17mm
Uveal tumors
3 months post laser= 0.58 mm
15 months post laser=0.82 mm; re-lasered
Location of ciliary processes
• 8 yr Shiba Inu-primary glaucoma; candidate for gonioimplant, laser
• Location of ciliary body:– 5.31 mm superiorly– 4.77-5.02
inferiotemporal
Lens
• Useful to detect peripheral lens abnormalities
9 yr FS mixed breed, diabeticLIU more severe OS; HRUSperipheral lens capsule rupture,closed cleft
Iridocorneal angle
• Can examine ciliary cleft in greater detail• Can be useful in glaucoma cases, pre-op
cataract cases (?)• Maybe predict onset in ’good’ eye?
Typical glaucoma patient…
Iridocorneal angle
• Predictive value for onset of glaucoma?• Closed cleft at dx-6.5 mos• Open cleft at dx-22 mos
OD OS
• Ultrasound weakness=hyphemavs tumor
• Retinal detachment versus vitrealmembrane
Contrast-visualizing blood vessels• Doppler-detects
movement of blood and contrast agents
• Ocular movement=artifact
• Orbital cases• Labruyerre, Hartley,
et al JSAP 2011-contrast better than Doppler
Contrast ultrasound-seeing vessels
• Contrast agents-exogenous air or gas
• Specific agents: high molecular weight/low solubility gas microbubbles
Anesthesiology. 2004;101(3):614-619.
Contrast ultrasound-instruments
• Bubbles=scattered echoes
• Contrast agents-echogenic only at high doses; severe attenuation fundamental frequencies
Harmonic frequencies
• Fundamental-frequency-transmitted by probe
• Harmonic-multiples of transmitted
• 4 MHz transmitted– 2, 8, 12, 16 MHz
Pulse inversion imaging
• 2 waves-one inverted– Cancelled out=no signal
• Waves from contrast=signal
Contrast agent safety
• FDA-cardiac only• Anaphylactoid-2 dogs• 488 dogs/cats contrast
ultrasound vs 262 ultrasound alone– 0.2% (4 dogs)-vomiting
and syncope
Contrast disadvantage• Specialized machine required• Cost of contrast agent-$5/bottle
– 1 bottle/patientAgent Dosage Reference
Levovist 80 mg/kg Scharz et al. (2005) Rademacher et al. (2005)
Definity Dogs: <20kg, 0.1 mL, >20kg, 0.2 mL
O’Brien et al. (2004)
Optison Dogs: 0.5 mL Yamaya et al. (2002)Sonovue Dogs: 0.04-0.06 mL/kg
Dogs: <20kg, 0.5 mL, >20kg, 1.0mLDogs: 0.03 mL/kg
Nyman et al. (2005)O’Brien et al. (2004)Ohlerth et al. (2005)
J Ultrasound Med 21:299–307, 2002
90% divided by the number of frames, and thechange in video intensity was the increase frombaseline. The baseline was defined as the aver-age of 10 frames just before the rising phase ofthe curve. The calculated values from the curvesclosely correlate with tissue perfusion.
We checked the fundus after the examinationand compared the findings with those beforeexamination to confirm the safety of thismethod.
For statistical analyses, data were expressed asmeans. Correlation was performed by a Wilcoxonnonparametric rank sum test with SPSS version10 software (SPSS Inc, Chicago, IL). For differ-ences, P < .05 was considered significant.
ResultsWith the ultrasonographic contrast mediumand low-MI harmonic imaging, the ultrasono-graphic intensity of all orbital tissues increasedexcept in the tissues known to have no bloodflow in intact eyes. Contrast-enhanced tissueswere identified by comparison with the 10-MHzfundamental images obtained before contrast-enhanced imaging. Contrast enhancement firstappeared at the retrobulbar arteries, followedby the uveal tissue, extraocular muscles, andfinally the optic nerve. There was no change inthe vitreous humor, because we know that thenormal vitreous humor has no blood vessel.This enhancement pattern was reproducible(Fig. 5), and the brightness of the avascularregion showed little change, as expected (Fig. 6).
When the harmonic mode was switched to thefundamental mode, video intensities of theuvea, vessels, and muscles were diminished,and intensities in the remaining tissuesremained unchanged (Fig. 7).
Figure 8A illustrates the relationship of timeversus signal intensity in the uvea of a healthyeye, and Figure 8B shows the relationships inan eye with impaired blood flow. Table 1depicts the values of the rise-time slope and thechange in video intensity in both groups. Thesevalues for the uvea are significantly greater in
J Ultrasound Med 21:299–307, 2002 303
Hirokawa et al
Figure 4. Fundal ICG angiographic image of the same eye as inFigure 3 showing a defect in the avascular area of the fundus. Inthis method, near-infrared light is used, which is much more effi-cient in penetrating pigmented tissues of the fundus comparedwith the wavelength of visible rays. Therefore, this techniqueseems to be more proper for showing choroidal blood flow.
Figure 5. Harmonic ultrasonography of the normal transverse eye. A, Just before injection of the contrast agent. B, Just after the injection. Immediatelythe video intensity of vessels and uvea was instantly enhanced.
A B
Pre contrast Post contrast
with this finding. Detection of the avascular areain the uvea especially showed excellent correla-tion with fundal angiography. A combination ofhigh-frequency harmonic imaging with a con-trast agent insonating at a low MI has achievedsuch excellent resolution and reproducibility.7,15
An additional factor for the favorable results isthat the target tissues were near the probe, andsignals from bubbles were efficiently received.However, variations in quantitative signal inten-sity were observed.
Nevertheless, our study is still limited by sever-al factors. First, there is no reference standard forperfusion examination in orbital organs as forthe myocardial scintigram in the heart. The fun-dal ICG angiography used as a control for perfu-sion in our study is like fluoroscopy with aniodine agent and provides little information ontissue viability.4 Also, the microbubble harmonicphenomenon at 5 MHz and higher has not yetbeen examined adequately. In our study, a simplein vitro experiment was performed only to con-firm resonance and insonation. Taking intoaccount the properties of the shell wall and bub-ble size, further studies are necessary and shouldfocus on biophysical problems.16 Another issue isthat information on blood flow direction andvelocity is not provided by this method. In thisaspect, it is very different from color Dopplerimaging. Some ophthalmologic facilities usecolor Doppler imaging for assessment of retro-bulbar hemodynamics by using the resistanceindex. However, this index is not highly repro-ducible, because it tends to be affected by thepassage of blood vessels. Furthermore, an MI
value greater than the FDA guideline is used forclear measurement. The major objective of ourexamination was to visualize perfusion, andblood flow direction and velocity are expressedmore indirectly by color Doppler imaging thanby our method. Furthermore, our method com-plies with the FDA guideline, and the qualitativedata have high intrasubject and intersubjectreproducibility.
Contrast-enhanced echography has propertiesthat can be exploited to provide informationrelated to tissue perfusion. Our method has thepotential for expansion to clinical ophthalmicpractice for investigation of complex etiologiesof diseases such as glaucoma and age-relatedmacular degeneration, especially glaucoma, theprevalence of which is 3% of all Americans 40years and older,1 some patients also havingischemic optic neuropathy, which is curable byneurosurgery.17 Nowadays, this disease isdiagnosed by traditional ophthalmologicexaminations that provide little retrobulbarhemodynamics assessment. Our method mayprovide some useful information for deter-mining indications for neurosurgery.
In this study, we have determined somerequirements for an ideal ultrasonographic con-trast agent for visualizing orbital perfusion. Anagent that is active at a low MI is preferable, andbubble destruction should occur at an MI oflower than 0.24 for quantitative measurement.There is obviously a need to develop a highly sen-sitive bubble detection method, such as higher-frequency harmonic imaging at a low MI,without compromising penetration.
306 J Ultrasound Med 21:299–307, 2002
Contrast-Enhanced Harmonic Ultrasonography of Uveal Perfusion
Figure 9. Fundus before (A) and after (B) ultrasonographic examination. No change is evident after the examination.
A B
Research• Would cavitation of contrast bubbles
cause tissue damage in the eye?
Pre contrast Post contrast
Constrast enhanced ultrasound• Good for:
– Retinal detachment• Differentiating retinal detachment vs vitreal membrane
– Differentiation of solid tumor vs hemorrhage or exudate
• Less helpful with diffuse tumors in tissues that light up anyway (ie no tissue expansion)
• Bad for:– Masses with lots of necrosis– Small tumors– Retinal detachment with devitalized retina
12AX
20MHz at 6AX
Thank you Gill McLellan!!
Contrast
Iridociliary adenoma
Contrast
3D ultrasound
• Pre-natal imaging• 2D computerized
reconstruction• Limited ophtho reports• Movement artifact
3D ultrasound
3D
3D
Thanks to:COPLOW
UW Ophthalmology:
Erin ScottAlly Gosling
Andrew LewinPaul Miller
Gillian McLellanKim ShermanBecky Telle
UW Radiology:Ken Waller
Fern Delaney