Team West Pharmaceutical
Sponsor: Tim MillerAdviser: Dr. Harris
Sponsor and Business• Leadership in Drug
Administration Systems and Components
• Project Focus: rubber manufacturing for various rubber device components
The Problem• Rubber is being tested
far too late in the manufacturing process and takes too long
• Results in excessive material waste, as production continues with wrong formulations
Project Scope
• Create a rapid, at-line testing method to determine whether a rubber compound was properly mixed and will cure correctly.
Technical Basis
• Focus on Rubber Production
• Scorch time, cure rate, performance, viscosity
• Factors relate to quality of batch
Questions?
Wants/Needs/Constraints
• Wants– Simple Readout– Cost Effective– Automated
• Needs– Accuracy of
Prediction– Fast Testing– Minimize amount of
sample destroyed • Constraints– OSHA– FDA
Key Metrics
Metric Target
Uncertainty ± 2σ
Time < 10 min
Weight Destroyed < 1/2 lb
Selected ConceptAt-Line Viscosity Compression Test
Supporting Theory
Preliminary Testing• Tests were performed in the CCM to verify testing method using the Instron 5567
DesignDetermine Viscosity via compression test
Fabrication•The device was built from a combination of purchased motor components and machined framework
Calibrations and Programming• Labview & GalilTools
– Automatically converts and sends appropriate commands to the controller
– Allows user to set number of data points
– Safeguard to power off motors before critical loads are reached.
Video
Results and Statistical Analysis
Sample Viscosity (lb-s/in2)
A1 0.0055
A2 0.0052
A3 0.0055
A4 0.0056
A5 0.0058
A6 0.0059
Standard Deviation = 2.48×10-4 (lb-s/in^2)
Viscosity factor= (w^4)*8*(1/(h^3))*dh/dt
Meeting Key Metrics
Deliverables
• Prototype with ability to perform the desired tests
• Labview program for controlling the tests
• Method for performing tests and gaining useful data
Path Forward
• More Calibration Testing• Develop a more polished program• Build Database of viscosity data• The device needs to be a bit more refined, with proper electrical enclosures
Cost Summary• Sponsor given range: $5,000• Actual Costs: $4,851.67
– Zaber: Stepper Motors-> $1222.00– McMaster Carr and Materials-> $733.67– Galil: Controller-> $1817.00– Omega: Load Cell->$1079.00
Thank You