Sensor for pH DeGrandpre.pdf

download Sensor for pH DeGrandpre.pdf

of 11

Transcript of Sensor for pH DeGrandpre.pdf

  • 7/27/2019 Sensor for pH DeGrandpre.pdf

    1/11

    A sensor for in situ indicator-based measurements of seawater pH

    Matthew P. Seidel a,1, Michael D. DeGrandpre a,, Andrew G. Dickson b

    aDepartment of Chemistry, The University of Montana, 32 Campus Drive, Missoula, Montana 59812, United Statesb Scripps Institution of Oceanography, University of California, San Diego, 9500 Gilman Drive, La Jolla, California 92093, United States

    Received 12 June 2007; received in revised form 6 November 2007; accepted 28 November 2007

    Available online 15 December 2007

    Abstract

    Indicator-based spectrophotometric pH methods are now proven and commonly used for analysis of ocean samples; however,

    no autonomous system for long-term in situ applications has been developed based on this method. We describe herein an

    autonomous indicator-based pH sensor for seawater applications adapted from a design originally developed for freshwater pH

    measurements (SAMI-pH). The new SAMI-pH uses a different pH indicator, flow cell design, detection system, and mixing

    configuration to improve upon the freshwater performance. A new method was also tested that utilizes an indicator concentration

    gradient in the sample to correct for the pH perturbation caused by the indicator. With these design changes, laboratory tests found

    the precision improved from 0.004 to 0.0007 and the accuracy improved from 0.0030 to +0.0017 based on comparisons with

    benchtop UV/Vis measurements. In situ testing of two SAMI-pH instruments was completed off the pier at Scripps Institution of

    Oceanography. The average pH offset between the two instruments over the 22 d deployment period was 0.00420.0126 ( n =883),

    with the precision primarily regulated by large spatial and temporal variability at the site. The results demonstrate that the SAMI-

    pH can provide drift-free and precise pH measurements in adverse measurement conditions (extensive fouling and large tidal

    variability). With the current battery power (18 alkaline D-cells), the system can be deployed for periods up to 2 months with a

    0.5 h measurement frequency.

    2008 Elsevier B.V. All rights reserved.

    Keywords: pH sensor; in situ; autonomous; spectrophotometry; inorganic carbon; marine chemistry

    1. Introduction

    Quantifying and understanding the impacts of oceanic

    uptake of anthropogenic CO2 is an important contempor-

    ary problem that has brought intense focus on the ocean

    inorganic carbon system. The rapid rise in atmospheric

    CO2 has increased the concentration of CO2 in the oceans

    (Peng et al., 2003). Surface water pH has already

    decreased by

    0.1 units and the continued uptake ofCO2 is expected to decrease surface ocean pH by 0.20.3

    units over the next century (Orr et al., 2005). A decrease in

    pH of this magnitude could have large-scale impacts on

    marine ecosystems (Orr et al., 2005), will result in

    decreased uptake of CO2 by the ocean (Sabine et al.,

    2004), and will affect burial rates of organic and inorganic

    carbon (Feely et al., 2004). Understanding the inorganic

    carbon cycle is also key to understanding and predicting

    net community metabolism, airsea gas exchange, and

    export of carbon from the surface to depth. New and

    Available online at www.sciencedirect.com

    Marine Chemistry 109 (2008) 1828www.elsevier.com/locate/marchem

    Corresponding author. Tel.: +1 406 243 4118; fax: +1 406 243 4227.

    E-mail address: [email protected]

    (M.D. DeGrandpre).1 Tel.: +1 406 243 4118; fax: +1 406 243 4227.

    0304-4203/$ - see front matter 2008 Elsevier B.V. All rights reserved.doi:10.1016/j.marchem.2007.11.013

    mailto:[email protected]://dx.doi.org/10.1016/j.marchem.2007.11.013http://dx.doi.org/10.1016/j.marchem.2007.11.013mailto:[email protected]
  • 7/27/2019 Sensor for pH DeGrandpre.pdf

    2/11

    improved autonomous techniques are needed that can

    collect high quality, high temporal resolution CO2 data for

    characterizing both long and short term changes in the

    marine inorganic carbon system.

    Seawater pH has become a more appealing inorganic

    carbon parameter because improvements in the spectro-photometric method have made it consistent with the

    other CO2 system parameters (e.g., Byrne and Breland,

    1989) and it is readily measurable on ships with commer-

    cially available instrumentation. From our standpoint,

    another advantage is that spectrophotometric pH methods

    are, ostensibly, more easily adapted for in situ measure-

    ments. If pH is used in inorganic carbon calculations, it is

    highly desirable that the calculated dissolved inorganic

    carbon (DIC), total alkalinity (AT) orpCO2be comparable

    in precision and accuracy to measured values, i.e.,

    1.5 mol kg

    1

    (Byrne et al., 1999), 2.0 mol kg

    1

    (Dickson et al., 2003),and1atm (Murphy et al., 1998),

    respectively. To illustrate the sensitivity of calculated DIC

    andpCO2 to pH uncertainty, we used the CO2 equilibrium

    model CO2SYS (Lewis and Wallace, 1998). The carbonic

    acid constants in Mehrbach et al. (1973) refit by Dickson

    and Millero (1987) were used because these constants

    give the most accurate calculations of inorganic carbon

    parameters from spectrophotometric pH measurements

    (Clayton et al., 1995; Byrne et al., 1999). Calculations

    were made using a constant alkalinity of 2250 mol kg1

    over the pH range 7.78.3. For a pH uncertainty of

    0.005, the error in DIC ranges from 2.03.5 molkg1, lowest at the lower pHs typical of deep water. If the

    AT precision stated above is included in the calculation,

    the pH accuracy needs to be 0.002 to estimate DIC

    within 1.5 mol kg1. This level of uncertainty is

    adequate for most applications. Similar calculations for

    pCO2 reveal that the pH error is more critical at lower pH

    values wherepCO2 is higher. A pH uncertainty of 0.002

    gives a pCO2 error of 5.0 atm at pH 7.7 (pCO2=980

    atm) and an error of 1.1 atm at pH 8.3 (pCO2=182-

    atm). Previous shipboard studies have obtained pH

    accuracy (which typically limits the pH uncertainty) from0.0010.005 and precision from 0.0010.0004 (Byrne

    and Breland, 1989; Clayton et al., 1995; Tapp et al., 2000;

    Bellerby et al., 2002).

    Some work on in situ spectrophotometric pH sensors

    has been undertaken. An instrument for obtaining vertical

    depth profiles of pH and pCO2 from a ship has been

    developed (Nakano et al., 2006). The instrument collected

    data to 1000 m with accuracy of 0.0020.005 pH units

    when compared to discrete samples. The SEAS pH

    instrument, designed for in situ spectrophotometric pH

    and other inorganic carbon measurements, was tested in

    seawater and freshwater (Liu et al., 2006). Shipboard

    depth profiles in the Equatorial Pacific and Gulf of

    Mexico were collected. The in situ precision was 0.0014

    while accuracy was not reported. Both the Nakano et al.

    (2006) and Liu et al. (2006) designs require high power

    (67 W) light sources, which limit the duration they can

    be deployed in the field.We have also previously presented the design and

    performance of an in situ spectrophotometric pH system,

    optimized for freshwater measurements (Martz et al.,

    2003). The Submersible Autonomous Moored Instrument

    for pH, or SAMI-pH, accuracy and precision was 0.003

    and 0.004 pH units (n = 16), respectively, when deployed

    in a river over an 8 d period. No in situ indicator-based pH

    sensors, however, have been deployed for extended

    periods (weeks to months) in a marine environment. In

    this paper, we describe the optimization of the Martz et al.

    (2003) design for seawater applications. The new SAMI-pH uses a different pH indicator, flow cell design,

    detection system, and mixing configuration to improve

    upon the freshwater performance. A 22 d in situ seawater

    pH time-series is presented and demonstrates that the

    SAMI-pH sensor is capable of collecting accurate and

    precise long-term in situ data on ocean moorings and other

    unmanned platforms.

    2. Principle of operation

    A theoretical basis for the spectrophotometric deter-

    mination of pH is given in Clayton and Byrne (1993)and readers should refer to that article and other articles

    by Byrne et al. for a more in depth discussion. The

    method is based upon the equilibrium reaction of a pH

    indicator, usually a diprotic sulfonephthalein indicator,

    HLpKa

    H L2 1

    where HL and L2 are the protonated and deprotonated

    forms of the indicator and pKa is the second apparent

    dissociation constant. The acidic H2L form is not present

    at seawater pHs. The sulfonephthalein indicator m-cresolpurple (mCP) was selected because it has been exten-

    sively used in past studies (Clayton and Byrne, 1993;

    Clayton et al., 1995; Byrne et al., 1999; DOE, 1994).

    Combining the log form of the indicator equilibrium

    expression and Beer's Law yields:

    pH pKa logR e1

    e2 Re3

    2

    whereR is the absorbance ratioA2/A1 corresponding to the

    peak absorbance wavelengths 2

    and 1

    of the L2 and

    HL forms, respectively, and the eis are the temperature-

    19M.P. Seidel et al. / Marine Chemistry 109 (2008) 1828

  • 7/27/2019 Sensor for pH DeGrandpre.pdf

    3/11

    dependent ratios of the molar absorptivities () of the two

    species at both 1 and 2,

    e1 HLe578

    HLe434

    e2 Le578

    HLe434

    e3 Le434

    HLe434

    : 3

    For mCP, the s are measured at the peak absorbance

    wavelengths for L2 (578 nm) and HL (434 nm).

    The adaptation of this methodology for autonomous in

    situ applications requires special considerations to retain

    the accuracy and precision of the benchtop method. Most

    lab measurements are performed at a fixed temperature,

    typically 25 C (DOE, 1994). For in situ applications,

    temperature is not controlled and the temperature de-

    pendence of pKa and eis must be accurately known. The

    eis also depend upon the detection system optical band-

    pass and must be determined for that bandpass andliterature values cannot be used. In situ instruments, in

    particular, may use a larger bandpass to increase light

    throughput to reduce the light source power requirements.

    Moreover, when using a short pathlength optical cell

    (for reasons described below), the pH indicator solution

    can significantly alter the sample pH thus requiring some

    method to correct for the pH perturbation. Most benchtop

    methods use 10 cm pathlength cuvets with5106 mol

    kg1 indicator (e.g., Clayton and Byrne, 1993). At this

    concentration, the magnitude of the perturbation is

    typically b0.001 pH units (Chierici et al., 1999). Even

    so, various methods have been used to account for the pHperturbation (Zhang and Byrne, 1996; Bellerby et al.,

    2002). In the design presented here, we use a custom-

    made 1.1 cm pathlength flow cell to maximize light

    throughput and minimize the light source power require-

    ments and flush volume. Light throughput with a 0.5 W

    tungsten lamp gives sufficient signal-to-noise to achieve

    the desired level of absorbance precision. From our ob-

    servations, throughput drops off roughly proportionally topathlength, so a 1 W lamp is required to achieve the

    same signal-to-noise for a 2 cm pathlength cell. The short

    pathlength also reduces the pumping required to flush the

    system completely for blank measurements. These power

    savings allow for longer deployments, critical for an au-

    tonomous system. However, because the 1.1 cm cell

    requires indicator concentrations 10 times greater than

    10 cm cells, the pH perturbation becomes significant. To

    address this problem, a perturbationcorrection metho-

    dology suggested by Martz et al. (2003) has been

    thoroughly evaluated here, as described below.

    3. Methods

    3.1. Instrument modifications

    Two instruments, SAMI-47 and SAMI-51, with 47 and 51

    corresponding to their serial numbers, were manufactured by

    Sunburst Sensors, LLC with modifications to the Martz et al.

    (2003) design. The new design uses mCP instead of cresol red

    indicator because the mCP response better matches the full pH

    range of seawater. A commercially available static mixer (de-

    scribed below) replaced a coiled tube mixer to more effectively

    mix the indicator and sample. A custom made z-configuration

    flow cell (Fig. 1) replaced the 90o flow path optical cell used in

    the original SAMI design (DeGrandpre et al., 1995; Martz et al.,

    Fig. 1. Instrument design for the seawater SAMI-pH modified from the SAMI-pH for freshwater ( Martz et al., 2003). NO and NC refer to thenormally open and closed valve settings. Dotted lines represent fiber optics.

    20 M.P. Seidel et al. / Marine Chemistry 109 (2008) 1828

  • 7/27/2019 Sensor for pH DeGrandpre.pdf

    4/11

    2003) to reduce bubble trapping that occasionally occurred in the

    90 flow cell.Because of the orientation of the inlet relative to the

    optical channel, bubbles are more easily swept through the z-

    configuration. To improve light throughput, the spectrograph

    detection system was replaced with a fiber optic splitter con-

    nected to three silicon photodiode detectors with discrete inter-ference filters (Yuan and DeGrandpre, 2006).

    3.2. Flow design

    The sample and indicator solutions are connected to the

    normally open and normally closed ports, respectively, of a

    three-way solenoid valve (161K031, Neptune Research Inc.)

    (Fig. 1). A low power 50 l per pulse solenoid pump

    (LPLA1210050L, The Lee Co.) draws sample or indicator

    through the valve and dispenses it into the downstream

    plumbing. The pump and valve are enclosed in a silicone oil-

    filled housing with a diaphragm for pressure equilibration. The

    indicator solution, described below, is contained in a 150 ml gas-impermeable bag (Pollution Measurement Corp.). The static

    mixer, developed for HPLC, is a 350 l internal volume

    convoluted flow path encased in a stainless steel housing (421-

    0350B, Analytical Scientific Instruments). The connecting and

    exit tubing consists of 1.6 mm o.d., 0.76 mm i.d. PEEK.

    3.3. Optical design

    Light from a 2.3 mm diameter 0.5 W tungsten lamp (4115-B,

    Gilway Technical Lamp) is collected by butt coupling to a 0.37

    numerical aperture 1000 m fused silica fiber (ThorLabs Inc.).

    The fiber is fitted into the custom made PEEK z-cell (Fig. 1)using a nut (P-331, Upchurch Scientific Inc.) andferrules (P-359,

    Upchurch Scientific Inc.). The cell pathlength is 1.1 cm with a

    17.7 l internal volume. The fiber input and output ferrule ends

    create a watertightseal by compressing Teflon gaskets onto fused

    silica windows. The transmitted light is collected by another

    1000 m fiber optic coupled to a 1 3 fiber optic bundle splitter

    (FiberTech Optica Inc.) (Yuan and DeGrandpre, 2006). The light

    from each bundle channel passes through a series of optical

    filters, which isolate the three wavelengths of interest (434, 578,

    and 780 nm). Measurements at 780 nm, where mCP has no

    absorbance, serve as a correction for light throughput changes

    between blank measurements. Additional filters are used tominimize out of band (stray) light. For detection at the 780 nm

    referencechannel, two 780 nm bandpass filters (780-10-50; Intor,

    Inc.) are used. At the two analytical wavelengths, optical

    interference filters (435.8-10-45 and 577.7-10-50; Intor, Inc.) are

    combined with heat absorbing filter glass (040FG11 at 434 nm

    and 005FG13 at 578 nm, Andover Corp.). The combined filters

    reduced stray light to b0.15% (Yuan and DeGrandpre, 2006)

    which is critical for good absorbance and pH accuracy. All

    interference filters have a 10 nm bandpass at full width half

    maximum; however, the center wavelength can vary slightly

    between filters with the same part number, requiring thateis (Eq.

    (3)) be measured for each SAMI. The filtered light is detected bysilicon photodiodes (S2386-45K, Hamamatsu Corp). The

    detector photocurrent is converted to voltage in an amplifying

    circuit, digitized into 12 bits and stored using a data logger (TFX-

    11, Onset Computer Corp.).

    3.4. Indicator solution

    The SAMI-pH and benchtop UV/Vis indicator solutions

    are comprised of1.0103 mol kg-soln1 mCP (211761-

    10G, 90% dye content, Aldrich) and 0.70 mol kg-soln1 NaCl

    (S671-3, Fisher Chemicals) adjusted to a pH of 7.5 with

    1.0 M HCl. The indicator concentration was tested over the

    range 0.020.0005 mol kg-soln1 and it was found that

    1.0103 mol kg-soln1 mCP gave the best balance between

    pH precision and flush times (results not shown). The

    influence of indicator pH is discussed below.

    3.5. Measurement sequence

    Each measurement begins by flushing 3.75 ml of sample(75 pump pulses) through the normally open (NO) port on the

    valve (Fig. 1). This flushing is done to completely clear the

    system of indicator from the previous measurement so a blank

    measurement can be taken. Next, a 50 l pulse of indicator is

    injected into the seawater sample through the normally closed

    (NC) port on the valve (Fig. 1) followed by an additional 25

    pump pulses (1.25 ml). Light intensity is recorded for each of the

    25 sample pump pulses as the indicator peak flushes through the

    optical cell. The first four of these measurements are prior to

    arrival of the indicator and are averaged to obtain a blank value,

    i.e., 100% transmittance. Optical blanks are therefore recorded

    for every pH analysis rather than every 12 h as done in Martzet al. (2003). The standard deviation of the four blank mea-

    surements is typically b0.0005 absorbance units. The full

    measurement sequence requires four min to complete. The

    SAMI 18 D-cell battery pack provides sufficient power for 2448

    measurements or50 d for a 0.5 h measurement frequency, 100

    d for a 1 h measurement frequency, etc.

    3.6. Molar absorptivity ratios (eis)

    To calculate the eis (Eq. (3)), molar absorptivities for the

    acidic and basic forms of mCP were determined at both

    434 nm and 578 nm for the SAMI-pHs and the reference

    spectrophotometer (Varian Cary 300 Bio UV/Vis). A 0.68 molkg1 NaCl solution with 0.02 mol kg1 NaOH was used for

    the determination of the basic L2 form of mCP. In this

    solution, greater than 99.995% is in the L2 form. A solution

    with pH of 5.5 was prepared for the HL form consisting of

    0.018 mol kg1 sodium acetate, 1.0 mol kg1 HCl, and

    sufficient NaCl to set the ionic strength at 0.70 mol kg1

    (Clayton and Byrne, 1993). At pH 5.5, the presence of H2L,

    which has a poorly documented pKa and cannot be accurately

    quantified, is minimized. L2, which is present at pH 5.5, was

    accounted for by calculating its concentration and correcting

    for its molar absorptivity contribution to the final absorbance

    value (Martz, 2005). The correction was approximately 53%for HL578 due to the much larger L578. However, because

    21M.P. Seidel et al. / Marine Chemistry 109 (2008) 1828

  • 7/27/2019 Sensor for pH DeGrandpre.pdf

    5/11

    HL578 is small, pH is relatively insensitive to errors in this

    parameter (Table 1). Corrections for the other three molar

    absorptivities were all less than 0.3% of the measured value.

    It is critical to account for the ei temperature dependence

    (Zhang and Byrne, 1996; Gabriel et al., 2005). For example,

    using constanteis for a 10 C temperature change creates a pH

    error of0.006 pH units. The molar absorptivity temperature

    dependence from 525 C was determined for each SAMI and

    the Cary UV/Vis. Replicate measurements were used to

    determine the precision and resulting uncertainty in

    calculated pH (Table 1). As can be seen in Table 1, uncertainty

    of the limits the pH accuracy to 0.0010.002 pH units. The

    measurement uncertainty is primarily controlled by indicator

    solution preparation.

    3.7. Indicator pKa

    The pKa for mCP, determined on the total hydrogen ion

    concentration scale (Clayton and Byrne 1993; DOE, 1994) is:

    pKa 1245:69

    T 3:8275 0:00211 35 S ; 4

    where Tis temperature in Kelvin and S is salinity. The pKa in

    Eq. (4) was determined for temperatures ranging from

    293 KT303 K and salinities from 30S37. It should

    be noted that in situ deployments may have temperatures

    outside of this range. We used the van't Hoff equation to

    extrapolate the pKa temperature dependence. The change in

    entropy and enthalpy were calculated using the constants A, B,

    and C reported by Millero (1995).

    3.8. pH perturbation correction

    A gradient of indicator passing through the flow cell is used

    to correct for the indicator pH perturbation. The indicator and

    sample mix as they move toward the flow cell, generating a

    dilution curve (Fig. 2). The gradient of indicator absorbances is

    recorded and the pH is calculated at each point in the dilution

    curve using Eq. (2). The total indicator concentration [HL]T,

    which is the sum of the two indicator species (Eq. (5)), is

    calculated from absorbances and molar absorptivities using

    Eqs. (6) and (7),

    HL T HL L2

    5

    HL A434 Le578b A578 Le434b

    HLe434 Le578b2 HLe578 Le434b

    26

    L2

    A578 HLe434b A434 HLe578b

    HLe434 Le578b2 HLe578 Le434b2:

    7

    Table 1

    The temperature dependence of the mCP molar absorptivities

    Species SAMI-47 SAMI-51 UV/Vis SAMI

    precision

    SAMI-pH

    error

    HL434 17104

    26(t25)

    16911

    26(t25)

    17340

    26(t25)

    52 0.0012

    HL578 85 +

    (t25)

    110+

    (t25)

    79 +

    (t25)

    11 0.0010

    L434 2301+

    12(t25)

    2238+

    11.7(t25)

    2151+

    12(t25)

    27 0.0009

    L578 38040

    71(t25)

    37480

    71(t25)

    37973

    71(t25)

    170 0.0020

    Molar absorptivity values are given in kg mol1 cm1 and temperature (t)

    is in C. These values are only applicable to mCP dye lot 11517KC from

    Sigma Aldrich. The average SAMI precision is calculated as the standard

    deviation of replicate measurements (n =9) from both instruments. The

    SAMI-pH error is estimated by adding or subtracting the SAMI precisionfrom the molar absorptivity value and recalculating the pH value.

    Fig. 2. (Top) Absorbances calculated from intensity measurements taken

    for 25 pump pulses as the indicator slug passes through the flow cell

    (1 absorbance measurement obtained per pump pulse). Absorbances

    between 1.5 and 0.2 absorbance units are used to calculate the pH

    (symbols). Filled circles are A434 and open circles are A578. (Bottom) pH

    versus total indicator concentration [mCP]T for the SAMI data from the

    top figure (filled circles) and theoretical equilibrium model calculations

    (open circles). A linear extrapolation through each data set gives an R2

    value of 0.9998 for the SAMI data and N0.9999 for the pH perturbation

    model. The y-intercept (7.9169) represents the perturbation-free pH.

    22 M.P. Seidel et al. / Marine Chemistry 109 (2008) 1828

  • 7/27/2019 Sensor for pH DeGrandpre.pdf

    6/11

    The are in kg-soln mol1 cm1 and pathlength (b) in cm. A

    plot of pH versus [HL]T is extrapolated to zero indicator

    concentration to determine the perturbation-free pH (Fig. 2).

    Mixing occurs at both the leading edge and tail end of the

    indicator slug as shown in Fig. 2. The leading edge of the

    indicator pulse has only 23 absorbance values in the optimal

    absorbance range (0.21.5 a.u.) (Ingle and Crouch, 1988);

    however, the tail end of the dilution curve is more dispersed.

    Approximately 9 points in this portion of the dilution curve are

    used to create the perturbation correction plot (Fig. 2). Because

    in this case the indicator pH is 7.5, the pH decreases with

    increasing indicator concentration.

    It was proposed in Martz et al. (2003) that a linear ex-

    trapolation could be used to determine the pH with no indicator

    present. We examined the linearity of the pH perturbation using

    an equilibrium model. The model includes the CO2 and indicator

    equilibrium expressions, equilibrium constants, the original DIC

    andATof the sample and indicator, and dilution factors (calculated

    from the total indicator concentration). The program calculates apH value for a range of points with varying indicator con-

    centration and dilution factors. An example of a theoretical pH

    perturbation curve is shown in Fig. 2. TheR2 value for a linear fit

    to the theoretical data is N0.9999 over the [HL]T range used.

    Linear relationships are also found over the entire pH range tested

    (7.58.5) (not shown). Therefore, a linear fit is used with the

    dilution curve data from each sample measurement (Fig. 2) to

    determine the pH perturbation.

    3.9. Laboratory tests

    Laboratory pH measurements were made on certifiedreference materials (CRMs) (Dickson, 2001) or seawater

    collected in Hood Canal, Washington (USA). To prevent CO2exchange during replicate analyses, samples were stored in

    gas-impermeable bags (Pollution Measurement Corp.) and the

    bags were connected directly to the input line of the sensor

    (Fig. 1). All laboratory measurements were made at20 C.

    Temperature was controlled to 0.05 C using a circulating

    water bath (Thermo Neslab RTE 7) and a water chamber

    placed over the top of the SAMI-pH. The sample bags were

    also placed inside the chamber so that the sample and optical

    cell of the SAMI-pH were at the same temperature. All tem-

    perature probes and water baths were calibrated to a certifieddigital thermometer (4400, Eutechnics).

    Comparison measurements were obtained using the Cary

    double beam UV/Vis spectrophotometer. Dual 10-cm jacketed

    cell holders were used for temperature control. The 10 cm

    sample cuvet was mixed using a magnetic stirrer (Spectrocell,

    Inc.). Each pH measurement took sample directly from the same

    bag used for the SAMI-pH measurements. The outlet tube from

    the bag was placed inside the cuvet and sample was allowed to

    overflow the cuvet 10 volumes to insure the cuvet was

    thoroughly rinsed. The cuvet was then placed inside the UV/Vis

    and allowed to reach the desired temperature (5 min). Each

    measurement used three 40 l additions of 1103 mol kg1

    indicator with absorbance measurements taken for each addition.

    As in the SAMI-pH, the resulting plot of pH versus total

    indicator concentration was used to linearly extrapolate to a zero

    indicator concentration.

    3.10. Field tests

    In situ testing of SAMI-47 andSAMI-51 wasperformed in a

    seawater tank and off the SIO pier. Comparison measurements

    were made on a Cary 1E UV/Vis spectrophotometer. Samples

    were taken from the tank or pier in glass bottles with minimal

    head space. Samples were first placed in a water bath for 20 min

    set to the in situ water temperature and then a 10 cm cuvet at the

    same temperature was filled with sample, minimizing the head

    space. The cuvet was inserted into the Cary 1E for a blank

    measurement and indicator was added as described above but

    the cell containing sample and indicator was shaken by hand.

    All discrete pH samples were measured immediately after

    collection using the procedure described above.

    Preliminary tests were made in a 100-gallon tank filled withseawater collected from a pipeline that extends 400 m into the

    Pacific Ocean. A re-circulating pump was used to keep the water

    uniformly mixed. The SAMI-pH instruments were programmed

    to take measurements every 15 min. Discrete samples were

    collected at 1 h intervals.

    Immediately after these tests, the SAMI-pH instruments

    were deployed from the SIO pier for a three week period.

    Salinity, temperature, and depth were measured using a CTD

    (Microcat, Seabird, Inc.). Discrete samples were collected for

    DIC, AT, pH, and salinity. The SAMI-pHs were programmed

    to measure pH every 30 min.

    The SAMI-pH instruments went through a number ofpreparation steps in order to safely package them for deployment

    off the pier. A copper cage was placed around the exposed SAMI

    flow cell, fiber optics, temperature probe, and pump-valve

    housing in order to protect the SAMI and provide fouling

    resistance. The reagent bag was placed in a PVC box clamped to

    the SAMI housing. The SAMI was thenplacedin a stainless steel

    cage for protection and to provide solid anchor points. A 100-

    foot serial communication cable was attached to the SAMI to

    monitor the instrument performance. The deployment took place

    from April 27, 2006 through May 25, 2006. Water depth at the

    end of the pier is 8 m and the instruments were placed at 4 m

    depth on opposite sides of the same piling (2 m apart), with anorth and south orientation to obtain the same exposure to east

    west tidal and wave motion.

    4. Results and discussion

    4.1. Indicator pH

    The pH perturbation is strongly dependent upon the

    difference between the sample and indicator pH. To determine

    the performance sensitivity to indicator pH, a 5.0103 mol

    kg1 indicator solution was adjusted to pHs of 8.2, 8.0, 7.8, and

    7.5 (0.1) using 1 M HCl or NaOH. Over this range, indicatorpH did not significantly influence accuracy and precision using

    23M.P. Seidel et al. / Marine Chemistry 109 (2008) 1828

  • 7/27/2019 Sensor for pH DeGrandpre.pdf

    7/11

    the linear extrapolation method; however, the best fits (highest

    R2) were found with the lowest pH solution (Table 2). Based on

    these results, we continued to use an indicator solution pH of

    7.5, the pH chosen during the early stages of testing. Although

    it makes intuitive sense to use a pH close to the sample pH to

    minimize the perturbation, the results here suggest relative

    insensitivity to this parameter.

    4.2. Flow cell gradient

    Initially, we were concerned that the gradient of indicator

    concentration in the SAMI-pH optical flow cell could lead to

    systematic errors because Beer's Law assumes a homogeneous

    concentration. The validity of this assumption with the 17 l

    optical cell was examined. The absorbance gradient between two

    consecutive 50 l pump pulses is estimated from data in Fig. 2.

    We split thisabsorbance gradient into30 separate slices along the

    flow cell and converted the absorbances to transmittance. A

    corrected absorbance and pH were then calculated from theaverage of these thirty transmittance values. This procedure

    estimates a +0.0004 pH unit error in the actual pH of the sample.

    Therefore, the absorbance gradient is nota significant problemin

    the SAMI-pH; however, it could be more important for systems

    that use long (10 cm) pathlength cells such as liquid core

    waveguides (Liu et al., 2006) if the sample and indicator are not

    well mixed.

    4.3. Laboratory precision and accuracy

    In the Martz et al. (2003) design, precision was found to

    be 0.004 pH units and was controlled by blank drift, lampstability, and microscopic bubbles. The new design precision

    was quantified over a wide range of sample pH (Table 3). The

    average precision is 0.0007 pH units (n =165). We attribute

    the improved precision to reduced bubble trapping in the z-

    cell, a broader mixing curve due to the static mixer, and

    implementation of frequent blank measurements.

    The accuracy for the two SAMI-pH instruments was

    determined by analyzing Hood Canal seawater (S=31, t=25 C)

    every 20 min for an 8-hour period followed by analysis of the

    remaining sample on the Cary UV/Vis (n=7). The average pH

    measured by SAMI-47 and SAMI-51 were 7.91930.0005 and

    7.91590.0006, respectively. The average pH measured on the

    Cary spectrophotometer was 7.9180 0.0008 resulting in an

    accuracy, relative to the Cary, for SAMI-47 and SAMI-51 of

    +0.0013 and 0.0021, respectively (n=7).

    4.4. SAMI-pH in situ testing

    During the initial in situ tests in the laboratory seawater tank,

    the average offset between SAMI-47 and SAMI-51 for 45tankmeasurements was 0.0017 units over a pH range of 8.19

    8.21 and temperature from 22.1 to 22.2 C (data not shown).

    Benchtop spectrophotometric pH measurements agreed to within

    +0.0005 for SAMI-47 and 0.0015 for SAMI-51 (n =12).

    The data collected during the 22 d deployment off Scripps

    pier are shown in Fig. 3. A total of 1005pH measurements were

    made by each SAMI. The pH was calculated using the linear

    extrapolation method described above and shown in Fig. 2.

    Sample regression plots for SAMI-47 and SAMI-51 at two

    different pH values are shown in Fig. 4. All in situ and sample

    data were corrected to the same in situ temperature using a

    temperature coefficient of 0.0015 pH units per 0.1 C

    calculated using the salinity and DIC/AT found at the site. The

    resulting pH ranged from 8.42 to 7.74 with changes as large as

    0.6 pH units over diurnal and tidal cycles (Fig. 3).

    The time-series show large tidal swings and associated

    temperature and salinity changes (Fig. 3). The depth ranged from

    3.5 to 5.5 m and temperature changes N6 C were observed

    during the tidal oscillations. The temperature alsoincreased from

    a daily high of 1516 C at the start of the deployment to 18

    19 C at the end of the deployment. The presence of large lateral

    and vertical temperature gradients was confirmed using the

    temperature records from SAMI-47 and SAMI-51. The two

    SAMI-pH temperatures differed by up to 2 C during periods of

    rapid temperature fluctuations. These times correspond to

    Table 2

    Evaluation of different indicator pHs

    [mCP]

    (mol

    kg1)

    Indicator

    pH

    Average

    pH

    pH

    precision

    Slope

    (pH kg

    mol1)

    R2 pH

    perturbation

    0.005 8.2 7.8890 0.0004 222 0.98 0.011

    0.005 8.0 7.8884 0.0004 40 0.79 0.0020.005 7.8 7.8886 0.0004 166 0.97 0.008

    0.005 7.5 7.8886 0.0007 366 1.00 0.018

    The average seawater sample pH was determined using the pH

    perturbation method described in the text (n =5). Slope is the average

    of the five slopes generated from the pH versus total indicator plot. The

    R2 value is the average for the five measurements. As expected, the pH

    perturbation is larger (larger positive or negative slope) when the pH of

    the indicator is further away from the sample pH (to within the

    tolerance of the uncertainty in the indicator pH). The perturbation is the

    change in pH of the sample due to 5105 mol kg1 total indicator

    concentration (the indicator concentration in the indicator reservoir

    was 5 103 mol kg1). The indicator pH was determined using a pH

    electrode and is only accurate to within 0.1 pH units. Measurementtemperature was 25 C.

    Table 3

    The pH precision for measurements made over the typical seawater

    range

    pH Precision 95% confidence Measurements

    7.6862 0.0005 0.0002 23

    7.7833 0.0009 0.0003 507.9159 0.0005 0.0002 25

    7.9193 0.0006 0.0002 25

    8.0260 0.0007 0.0003 17

    8.2029 0.0009 0.0004 25

    Samples are CRMs and Hood Canal seawater adjusted to varying pH

    values using concentrated HCl or NaOH. Measurement temperature

    was 25 C.

    24 M.P. Seidel et al. / Marine Chemistry 109 (2008) 1828

  • 7/27/2019 Sensor for pH DeGrandpre.pdf

    8/11

    significant pH differences between the SAMIs, presumably due

    to large pH gradients associated with the temperature (water

    mass) differences. Therefore, any time the measured temperature

    difference between SAMI-51 and SAMI-47 wasN0.2 C, the pH

    measurements were removed from the intercomparison between

    the two instruments (122 measurements were removed). Using

    the remainingdata, the agreementbetween SAMI-47 andSAMI-

    51 was 0.00420.0126 (n =883) for the 22 d period (Fig. 5).

    Tightening the temperature difference criteria to 0.1C improved

    the comparison to 0.00400.011 and removed another 116

    measurements. Filtering both time-series using a 2 h average

    further reduces the scatter between the two SAMIs to 0.0084.

    The lab measurements were more precise than this (Table 3) and

    it is likely that the large spatial gradients, particles in the flow cell

    (see below), and small response time differences, e.g., due to

    different flushing times, degraded the precision between the two

    instruments. The between-instrument precision is notably larger

    during periods when pH was rapidly changing, i.e., at the

    beginning and end of the deployment (compare Figs. 3 and 5).

    Early on, the pH rate of change was as large as 0.016 min1 and

    often largerthan 0.005 min

    1

    . Looking at the pH dilution curves,SAMI-51 flushed indicator1530 s more slowly than SAMI-

    47 due to flow rate differences (pump efficiency), which, based

    on the pH rates of change, could account for a significant portion

    of the pH differences.

    Importantly, the offset is small and there is no apparent drift

    between the instruments during the deployment (Fig. 5). There

    was also no systematic trend in the SAMI difference over the

    pH and temperature range during the deployment (Fig. 6).

    These results were further verified by comparison with

    Fig. 3. In situ SAMI-pHs, salinity, temperature and depth during the 22

    d deployment off SIO pier.

    Fig. 4. pH versus total indicator concentration [mCP]T for two measurements from SAMI-47 and SAMI-51 during deployment off SIO pier. Theresulting pH (y-intercept) and R2 values are given in each plot.

    Fig. 5. Comparison between SAMI-47 and SAMI-51 during the 22

    d deployment calculated from Fig. 3 data. The average offset is

    0.0042 0.0126 (n =883). See text for more discussion.

    25M.P. Seidel et al. / Marine Chemistry 109 (2008) 1828

  • 7/27/2019 Sensor for pH DeGrandpre.pdf

    9/11

    measurements of pH on discrete samples collected off the pier

    on April 30 and May 3. SAMI-47 was deployed for both of

    these days while SAMI-51 was deployed on May 2 and was

    only compared to the pH values on May 3. Good agreement

    was achieved between the SAMI-pH instruments and bench-

    top spectrophotometric pH measurements with an average pH

    offset of +0.0013 (n =8) for SAMI-47 and 0.0043 (n =3) for

    SAMI-51.

    After May 3, pH values were calculated from DIC and ATfrom discrete samples using CO2SYS (Lewis and Wallace,

    1998) and the dissociation constants of Mehrbach et al. (1973)

    refit by Dickson and Millero (1987). Using these pH values, the

    average offset for both SAMI-47 and SAMI-51 are nearly

    identical at 0.0348 and 0.0330 pH units (n =14). Use of other

    CO2 equilibrium constants in CO2SYS (Lewis and Wallace,1998) did not improve the offset. Examination of the discrete

    sample data pointed to a possible error in the AT values. In

    surface waters, AT often follows a conservative relationship with

    salinity (Millero et al., 1998); however, AT of the discrete

    samples varied by 40 mol kg1 while the sample salinities

    were very similar (33.54 0.05). Using an approximate relation-

    ship between salinity and AT for California coastal waters

    (AT=2150+44 [S31.25]) (G. Friederich, pers. comm., 2006),

    the expected AT for the salinity range observed is 2249

    2252 mol kg1 compared to the measured range of 22162254

    (14) mol kg1. Using the salinity-predicted AT and the

    measured DIC, the pH errors reduce to 0.014 and 0.012 pH unitsfor SAMI-47 and SAMI-51, respectively. Uncertainty in the pKa

    (Eq. (4)), due to extrapolation to the in situ temperatures, may

    also contribute to the pH uncertainty.

    Although the instruments were only in the water for a short

    time, barnacle and algal growth completely covered the

    instruments. One concern in such an intense fouling environ-

    ment is that, in drawing in a sample for analysis, particles can

    occlude the optical path or clog the plumbing. The 780 nm

    reference channel light throughput for the two instruments

    during the 22 d deployment suggest that particles may have been

    drawn in at times (Fig. 7). The fluctuating intensity for SAMI-51

    from May 8

    9 may be due to small particles entering the tubingand optical cell. The return to the original intensity value shows

    that the particles eventually flushed out. SAMI-47 shows a

    different trend with the intensity slowly drifting downward with

    only 40% of the original intensity at the end of the deployment.

    The resulting behavior was most likely due to lamp drift or

    particulate matter entering the flow cell and slowly reducing the

    light throughput. No particles were evident upon inspection of

    the inside of the flow cell, however. The change in light

    throughput shown in Fig. 7 was accounted for by obtaining blank

    measurements for each pH analysis, i.e., for one set of data as

    shown in Fig. 2. To verify the need for frequent blanks, pH

    values for both instruments were calculated using individual

    blank values and an initial blank value over a 32-hour section of

    the field deployment data (May 2022). Using individual blank

    values over this time-period, the average difference between

    SAMI-47 and SAMI-51 pH valueswas 0.0017 0.0149pH units(n =64), while using a single initial blank value gave an average

    offset of 0.0070 0.0246 pH units. No drift was observed

    between the two instruments (Fig. 5), even with the large

    changes in light throughput (Fig. 7), demonstrating the

    robustness of the technique.

    Fig. 6. pH difference between SAMI-47 and SAMI-51 versus temperature (A) and pH (B). The plot shows that the SAMIs respond very similarly over

    the range of temperature and pH found during the deployment.

    Fig. 7. % transmittance (relative light throughput) at 780 nm. This

    wavelength is used for the correction of lamp drift, bubbles, particles,

    and any other factors that affect light throughput between blank

    measurements. The % transmittance is calculated relative to the initialthroughput at the start of the deployment.

    26 M.P. Seidel et al. / Marine Chemistry 109 (2008) 1828

  • 7/27/2019 Sensor for pH DeGrandpre.pdf

    10/11

    5. Conclusions

    Multiple steps were undertaken to improve the SAMI-

    pH performance. The filter-based detection system im-

    proved the signal-to-noise and consequently, the absor-

    bance precision. Molar absorptivities were corrected forthe presence of other indicator forms by using the pH,

    concentration, and absorbance values of the acidic and

    basic solutions. With these changes, the agreement be-

    tween the Cary UV/Vis and SAMI-pH instruments in a

    laboratory setting was improved from 0.003 to +0.0017.

    Optimization of various procedures, including the pH

    perturbation correction and mixing of the indicator and

    sample, improved the precision to 0.0007 compared to

    the previous precision of 0.004 of the Martz et al. (2003)

    design.

    The 22 d deployment off Scripps pier collected fromSAMI-47 and SAMI-51 collected the longest known in

    situ spectrophotometric seawater pH time-series to date.

    The excellent agreement between the two SAMI-pH

    instruments and the Cary UV/Vis is very promising;

    however, further tests are needed to determine internal

    consistency with other inorganic carbon parameters.

    The SAMI-pH can be used for studies related to

    pressing contemporary carbon cycle questions such as

    the oceanic uptake of atmospheric CO2 and the impacts

    of ocean acidification. The instrument can be deployed

    on many different platforms including Volunteer Obser-

    ving Ships, permanent moorings, drifters, autonomousprofilers, and AUVs. Development of sensors for DIC

    and AT is needed to make possible fully autonomous

    characterization of the inorganic carbon system, i.e.,

    through combined measurements of pH or pCO2 with

    DIC orAT.

    Acknowledgments

    We thank Cory Beatty (University of Montana) for lab

    and field assistance, Todd Martz (formerly UM, now at

    MBARI) for discussions and Charles Coughran (SIO) fordeployment of instruments. Financial support was pro-

    vided by NSF grant OCE-0327763 and a NSF-EPSCoR

    fellowship to M. Seidel.

    References

    Bellerby, R.G.J., Olsen, A., Johannessen, T., Croot, P., 2002. A high

    precision spectrophotometric method for on-line shipboard sea-

    water pH measurements: the automated marine pH sensor (AmpS).

    Talanta 56, 6169.

    Byrne, R.H., Breland, J.A., 1989. High precision multiwavelength pH

    determinations in seawater using cresol red. Deep-Sea Res. 36,

    803810.

    Byrne, R.H., McElligott, S., Feely, R.A., Millero, F.J., 1999. The role

    of pHT measurements in marine CO2-system characterization.

    Deep-Sea Res. 46, 19851997.

    Chierici, M., Fransson, A., Anderson, L.G., 1999. Influence of m-cresol

    purple indicator additions on the pH of seawater samples: correction

    factors evaluated from a chemical speciation model. Mar. Chem. 65,

    281290.Clayton, T.D., Byrne, R.H., 1993. Spectrophotometric seawater

    pH measurements: total hydrogen ion concentration scale calibra-

    tion of m-cresol purple and at sea results. Deep-Sea Res. 40,

    21152129.

    Clayton, T.D., Byrne, R.H., Breland, J.A., Feely, R.A., Millero, F.J.,

    Campbell, D.M., Murphy, P.P., Lamb, M.F., 1995. The role of pH

    measurements in modern oceanic CO2-system characterizations:

    precision and thermodynamic consistency. Deep-Sea Res. II 42,

    411429.

    DeGrandpre, M.D., Hammar, T.R., Smith, S.P., 1995. In situ

    measurements of seawaterpCO2. Limnol. Oceanogr. 40, 969975.

    Dickson, A.G., Millero, F.J., 1987. A comparison of the equilibrium

    constants for the dissociation of carbonic acid in seawater media.

    Deep-Sea Res. 34, 17331743.Dickson, A.G., 2001. Reference materials for oceanic CO2 measure-

    ments. Oceanog. 14, 2122.

    Dickson, A.G., Afghan, J.D., Anderson, G.C., 2003. Reference

    materials for oceanic CO2 analysis: a method for the certification

    of total alkalinity. Mar. Chem. 80, 185197.

    DOE, 1994. In: Dickson, A., Goyet, C. (Eds.), Handbook of Methods

    for the Analysis of the Various Parameters of the Carbon Dioxide

    System in Seawater. version 2.0.

    Feely, R.A., Sabine, C.L., Lee, K., Berelson, W., Kleypas, J., Fabry, V.J.,

    Millero, F.J., 2004. Impact of anthropogenic CO2 on the CaCO3system in the oceans. Science 305, 362366.

    Gabriel, M.D., Forja, J.M., Rubio, J.A., Gomez-Parra, A., 2005.

    Temperature and salinity dependence of molar absorptivities of

    thymol blue: application to the spectrophotometric determination

    of pH in estuarine waters. Cienc. Mar. 31(B), 309318.

    Ingle, J.D., Crouch, S.R., 1988. Spectrochemical Analysis. Prentice

    Hall, New Jersey.

    Lewis, E., Wallace, D.W.R., 1998. Program developed for CO2 system

    calculations. ORNL/CDIAC-105, Carbon Dioxide Information

    Analysis Center, Oak Ridge National Laboratory, U.S. Department

    of Energy, Oak Ridge, TN.

    Liu, X., Want, Z.A., Byrne,R.H., Kaltenbackher,E.A., Bernstein, R.E.,

    2006. Spectrophotometric measurements of pH in-situ: laboratory

    and field evaluations of instrumental performance. Environ. Sci.

    Technol. 40, 50365044.

    Martz, T.R., Carr, J.J., French, C.R., DeGrandpre, M.D., 2003. A

    submersible autonomous sensor for spectrophotometric pHmeasurements of natural waters. Anal. Chem. 75, 18441850.

    Martz, T.R., 2005. A tracer monitored titration for seawater total

    alkalinity. Ph.D. Thesis, University of Montana.

    Mehrbach, C., Culberson, C.H., Hawley, J.E., Pytkowicz, R.M., 1973.

    Measurements of the apparent dissociation constants of carbonic

    acid in seawater at atmospheric pressure. Limnol. Oceanogr. 18,

    897907.

    Millero, F.J., 1995. Thermodynamics of the carbon dioxide system in

    the oceans. Geochim. Cosmochim. Acta. 59, 661677.

    Millero, F.J., Lee, K., Roche, M., 1998. Distribution of alkalinity in the

    surface waters of the major oceans. Mar. Chem. 60, 111130.

    Murphy, P.P., Feely, R.A., Wanninkhof, R., 1998. On obtaining high-

    precision measurements of oceanic pCO2 using infrared analyzers.

    Mar. Chem. 62, 103115.

    27M.P. Seidel et al. / Marine Chemistry 109 (2008) 1828

  • 7/27/2019 Sensor for pH DeGrandpre.pdf

    11/11

    Nakano, Y., Kimoto, H., Watanabe, S., Harada, K., Watanabe, Y.W.,

    2006. Simultaneous vertical measurementsof in situ pH and CO2 in

    the sea using spectrophotometric profilers. J. Oceanogr. 62, 7181.

    Orr, J.C., et al., 2005. Anthropogenic ocean acidification over the

    twenty-first century and its impact on calcifying organisms. Nature

    437, 681686.

    Peng, T., Wanninkhof, R., Feely, R.A., 2003. Increase of anthropo-genic CO2 in the Pacific Ocean over the last two decades. Deep-

    Sea Res. II 50, 30653082.

    Sabine, C.L., Feely, A.F., Gruber, N., Key, R.M., Lee, K., Bullister, J.L.,

    Wanninkhof, R., Wong, C.S.,Wallace,W.R., Tilbrook, B., Millero, F.J.,

    Peng, T.H., Kozyr, T.O., Rios, A.F., 2004. The oceanic sink for

    anthropogenic CO2. Science 305, 367371.

    Tapp, M., Hunter, K., Currie, K., Mackaskill, B., 2000. Apparatus for

    continuous-flow underway spectrophotometric measurement of

    surface pH. Mar. Chem. 72, 193202.

    Yuan, S., DeGrandpre, M., 2006. Comparison between two detection

    systems for fiber-optic chemical sensor applications. Appl.Spectrosc. 60, 465470.

    Zhang, H., Byrne, R.H., 1996. Spectrophotometric pH measurements

    of surface seawater at in-situ conditions: absorbance and

    protonation behavior of thymol blue. Mar. Chem. 52, 1725.

    28 M.P. Seidel et al. / Marine Chemistry 109 (2008) 1828