Optimizing your handpiece Optimizing your handpiece life!life!

By The Dentist’s Choice

Page 2: Optimizing your handpiece life!

The following lessons will The following lessons will help you help you maximizemaximize the life the life of your handpieceof your handpiece

o Lesson 1 – Proper In-Office Settingso Lesson 2 – Sterilization Procedureso Lesson 3 – Lubrication

Page 3: Optimizing your handpiece life!

Lesson 1Lesson 1oProper In-Office Settings

oHandpiece InventoryoOperatory Air PressureoSterilization Equipment (Autoclave or StatIM)oLubrication Equipment

Page 4: Optimizing your handpiece life!

Handpiece InventoryHandpiece Inventory

o Having the correct number of handpieces in your office is very importantoAll Manufacturers require 3 handpieces per

operatory to optimize turbine lifeoThis typically allows for 1 handpiece ready for use

by the doctoro1 handpiece in the sterilizero1 handpiece cooling to room temperature

o Please take a few minutes to count your handpieces in inventory

Page 5: Optimizing your handpiece life!

Operatory Air PressureOperatory Air Pressure

o Make sure your operatory’s air pressure is set correctly

o As a rule of thumb, set it to 40 p.s.i.o For complete accuracy, refer to the p.s.i.

chart included on this maintenance CD

*The KaVo station pictured at left shows the unit Air Pressure with both a Dial faceplate and also with a Digital reading

Page 6: Optimizing your handpiece life!

Sterilization - AutoclavesSterilization - Autoclaves

o Set to Correct TemperatureoAlways between 121°C and 135°CoSetting Autoclave temp above 135°C will cause

internal handpiece damage

*As noted in the picture above, most handpieces carry a 135°C warning

Page 7: Optimizing your handpiece life!

Sterilization - StatSterilization - StatIMIMo The StatIM sterilization units have an

automatic Handpiece-Cycle that we’ll visit in Lesson 2

Page 8: Optimizing your handpiece life!

Lubrication SettingsLubrication Settings

o Spray Lubricationo Pen Oilerso Dedicated Automatic Lubricators

Page 9: Optimizing your handpiece life!

Spray LubricationSpray Lubrication

Affix correct Nozzle◦ When using Spray Lubricants, always have the correct

nozzle for the handpiece being lubricated

* In the picture at left, we are correctly using a Fixed Back-End nozzle to lubricate a Fixed Back-End handpiece

Page 10: Optimizing your handpiece life!

Spray Lubricants & NozzlesSpray Lubricants & Nozzleso Euro-Lube Handpiece Lubricant

o Ideal for your Highspeed Handpieces

o Full-Synthetic Handpiece Lubricant & Cleanero Great for your Electric AttachmentsoAlso ideal for your Slowspeed Attachments

o Nozzleso We carry a full line of Nozzles for every handpiece model

Page 11: Optimizing your handpiece life!

Pen OilersPen Oilers

o Typically, Pen Oilers are used exclusively on Slowspeed handpieceso They are simple to use and you should always have a

few in your office as a back-up to your highspeed lubrication systems

Page 12: Optimizing your handpiece life!

Automatic LubricationAutomatic LubricationStationsStationso Check lubricant reservoir

o If it’s low, please fill it to the top line in the tanko Check that lubricant is flowing through the

machineo Do this by occasionally placing a piece of gauze in with the

handpieces (If working properly, after the lubricant cycle, the gauze should be damp)

Page 13: Optimizing your handpiece life!

Lesson 2Lesson 2oSterilization Procedures

oAutoclaveoStatIM

Page 14: Optimizing your handpiece life!

Sterilization - AutoclavesSterilization - Autoclaveso Always use Autoclave Bags

oUsing bags will protect the handpiece from external steam and prevent internal rust build-up

o Place bags paper-side upo This will allow water vapor to escape the bag and will

also prevent overheating

*Step 1: Place handpiece in Autoclave bag, then seal the open end

*Step 2: Place bags paper-side up onto the Autoclave tray

Page 15: Optimizing your handpiece life!

Sterilization - AutoclavesSterilization - Autoclaveso CoolingoOnce the Autoclave cycle is complete, simply

crack the door to the unit open approximately one inch

o Then allow the contents to cool down slowly over one-half hour

o You may then remove the handpieces safely

Page 16: Optimizing your handpiece life!

Important Cooling WarningImportant Cooling Warning

o Always allow handpieces to cool down naturally

o Never submerge handpiece in liquido Do not place handpieces under running

water

Page 17: Optimizing your handpiece life!

Sterilization - StatSterilization - StatIMIMo Always use Sterilization Bags

oUsing bags will protect the handpiece from external steam and prevent internal rust build-up

o Place bags paper-side upo This will allow water vapor to escape the bag and will

also prevent overheating

*Step 1: Place handpiece in Sterilization bag, then seal the open end

*Step 2: Place bags paper-side up onto StatIM cassette

Page 18: Optimizing your handpiece life!

Sterilization - StatSterilization - StatIMIM

o Be sure to select the correct setting on your StatIMo As noted below, it is the “Hollow Wrapped” cycle

Page 19: Optimizing your handpiece life!

Sterilization - StatSterilization - StatIMIM

o Following the Sterilization CycleoRemove Sterilization Cassette from the StatIM

and place it on your work surfaceoAllow the contents to cool naturally to room

temperature (approximately 30 minutes)

Page 20: Optimizing your handpiece life!

Important Cooling WarningImportant Cooling Warning

o Always allow handpieces to cool down naturally

o Never submerge handpiece in liquido Do not place handpieces under running

water

Page 21: Optimizing your handpiece life!

Lesson 3Lesson 3oLubrication

oLube Free HandpiecesoHighspeed HandpiecesoSlowspeed HandpiecesoElectric Attachments

Page 22: Optimizing your handpiece life!

LubricationLubrication

o Please follow the procedures on the upcoming slides as closely as possible

o Proper lubrication is the most important factor in lengthening handpiece life

Page 23: Optimizing your handpiece life!

Lube Free Handpieces Lube Free Handpieces o No lubrication is requiredo To clean chuck

oSpray a small amount of cleaner onto a proxy brush; then insert the brush into the chuck to clean

o To clean exterioroSimply wipe the outer surface with cotton swabs

containing a small amount of alcohol

Page 24: Optimizing your handpiece life!

Highspeed HandpiecesHighspeed Handpieceso Lubricate with your The Dentist’s Choice Euro-

Lube prior to sterilizingo Be sure to use the proper lubricant nozzle for your handpiece

and also affix a Spray Guard to the handpiece head to avoid splatter

o Spray lubricant through the handpiece for a 1 second bursto Purge with air for 15 to 30 secondso Sterilize

*Step 1: Push handpiece securely into lubricant nozzle

*Step 2: Push lubricant activator cap for 1 second

Page 25: Optimizing your handpiece life!

Highspeed HandpiecesHighspeed Handpieceso Following sterilization, lubricate once moreo You can lengthen the life of your handpieces up

to 40% by lubricating them after sterilization as welloSimply spray a 1 second burst of lubricant through the

handpieceo Purge on an air station or operatory for 15 to 30 seconds

before use by doctor

Page 26: Optimizing your handpiece life!

Slowspeed HandpiecesSlowspeed Handpieces

o Using your Pen Oiler, place 3 drops of oil into the Drive-Air tube (see below)

o Run the handpiece for 10 to 20 secondso To clean exterior, simply wipe the surface with a cotton

swab containing a small amount of alcoholo Sterilize the Handpiece

* The arrows point to the Drive-Air tube for the 3 different slowspeed back ends you may encounter

Page 27: Optimizing your handpiece life!

Slowspeed HandpiecesSlowspeed Handpieceso Lubrication for slowspeed handpieces that fit

onto a swivel couplero Like most highspeed handpieces, you will need to affix

the proper nozzle for your model of handpiece to a can of lubricant

oAs mentioned earlier, we have a full line of lubricant nozzles for every handpiece make

Page 28: Optimizing your handpiece life!

Slowspeed HandpiecesSlowspeed Handpieceso Lubricate your Slowspeed prior to sterilizing

o We recommend using our Full-Synthetic Lubricant & Cleanero Using the correct nozzle, spray lubricant through the handpiece

for a 1 to 2 second bursto Now run the handpiece for 10 to 20 secondso Sterilize the Handpiece

*Step 1: Push handpiece securely into lubricant nozzle

*Step 2: Push lubricant activator cap for 1 to 2 seconds

Page 29: Optimizing your handpiece life!

Electric AttachmentsElectric Attachmentso Lubricate ALL Electric Attachments with The

Dentist’s Choice Full-Synthetic Handpiece Lubricant & CleaneroA synthetic lube is necessary for Electric Attachments

due to their high speed and high number of internal moving parts

*Notice all the internal gears and bearings. Our Synthetic Lubricant will coat these internal components, greatly reducing wear and noise while lengthening handpiece life

Page 30: Optimizing your handpiece life!

Electric AttachmentsElectric Attachmentso To lubricate, affix the Attachment nozzle to your can of Full-

Synthetic Lubricant & Cleanero Insert the nozzle firmly into the back of the Attachment and spray

for a 2 second bursto While holding upside down, insert the Attachment onto a motor

unit and run for 10 secondso Sterilize the Attachment

Page 31: Optimizing your handpiece life!

Maintenance SummaryMaintenance Summaryo Check Handpiece Inventory

o Remember – 3 Handpieces per operatory (minimum)o Sterilization

o Set Autoclave Temperature to max of 135°Co Use sterilization pouches with paper side upo Let handpieces cool naturally

o Air Pressureo Set operatory air pressure to 40 p.s.i.o Refer to the handpiece p.s.i. chart for exact settings

o Lubricationo Lube Free – No lubrication requiredo Highspeed – Use Euro-Lube for 1 second with proper nozzleo Slowspeed – Place 3 drops of oil into Drive-Air tube with Pen Oilero Electric Attachments – Use Full-Synthetic Lubricant & Cleaner for 1 seconds with

proper nozzle