DRM/ZKF8KM2/030915/01 DRM/ZFK8KM2/030915/01
ZFK8/ZKM-2 IN-SITU ZIRCONIA ANALYSERZFK8/ZKM-2 Zirconia Oxygen Gas Analyzers are used as combustion controllers in combustion facilities and industries which require high energy consumption. The oxygen concentration control ensures complete combustion, thus reducing CO2/SOx/NOx emissions and helping prevent air pollution.
FEATURES- In-situ design with integrated detector filter- Max. Gas Temperature: • -10°C to +600°C (flow rate version) • -10°C to +1500°C (ejector version)- IP55 transmitter protection (options: IP66 or IP67)- Enhanced safety design with integrated and remote power isolation fuctions, including: • Thermocouple failure • Emergency power shut off • Password protected menu functions- High speed response (4 to 7 seconds)- Modular detector design allow easy field replacement of zirconia element
SPECIFICATIONS
ZFK8/ZKM-2Repeatability +/- 0.5% FS or less
Drift +/- 2.0% FS or less per monthResponse Time 4 - 7 seconds
Mounting Flange JIS5K 65A (80A for high dust)Tube Length Options 300mm to 1000mm
Display 4 Digit backlit LCDMeasured Gas Pressure -3 to +3 kPA
Analogue Outputs 4-20mA DC or 0-1V DC
Contact Output (Option) RS-485 Modbus or RS235 Modbus
Power Supply 100 to 240V, 50/60HzConsumption Approx. 100VA
For more information, please feel free to contact our sales team
(Contact details viewable below)
Datasheet continues overleaf
DRM/ZFK8KM2/030915/01 DRM/ZFK8KM2/030915/01
Model Method Gas & Ranges
ZFK8 Detector
ZKM Controller
Single Gas O2 (ZrO2):0-2/50% (in 1 vol% O2 steps)
Zero Gas:0.25 to 2.0% O2
Span Gas:20.6 to 21.0% O2
Typical Applications
Crematorium Combustion ControlBoiler MonitoringPower Generation
Petroleum RefiningPetrochemicalsManufacturing
Ejector Version.
Flow Tube Version.
Easy Change Detector.
NOTES: