Model ZR22G, ZR402G, and ZR202GDirect In Situ Zirconia Oxygen Analyzersand High temperature Humidity Analyzers
GS 11M12A01-E-H 4th Edition
The basic analyzer consists of a direct insertion type probe and aconverter that provides an analog output for control or monitoringpurposes. In addition a digital signal is superimposed on thecurrent output (HART) for optimization of maintenance. Theconverter controls also the (semi) automatic calibration when usedwith the calibration units.The detector and the converter are combined in one unit (ZR202)for simplified piping and wiring or the converter is mountedseparately (ZR402)Various accessories are available for optimal installation in difficultapplications ( dust filter, flame arrester, probe protector, hightemperature probe adapter, probe supporter) and for ease ofmaintenance and calibration.The analyzer is most commonly used to monitor the Oxygen levelin combustion air to allow optimal burner control to avoid airpollution and save energy. The direct insertion installation isespecially attractive from a maintenance point of view: samplingsystems are not required. The analyzer can also be used to monitor the Humidityconcentration at high temperatures in air with water as found inhumidifying processes like Bakeries or in indirectly heated dryers asfound in Paper machines.
FEATURES Direct measurement: No sampling required Field replaceable unique Zirconia cell, made to last. Heavy duty sensor heater Rebuildable probe construction Various methods for reference air to optimize accuracy
versus ease of installation Automatic compensation of duct pressure possible. Remote maintenance and set-up by digital communication
(HART)
GeneralSpecifications
SYSTEM CONFIGURATION
ZR22G ZR402G
ZR202G
Instrument air
Calibration gas pressure regulator
Air Set
Needle valveflowmeter
Reference gas
Calibration gas
Check valve ~
Zero gas cylinder
Model ZR40H Auto Calibration unit
Model ZR22G Separate type Zirconia Oxygen/ High temperature Humidity Analyzer, Detecor
Model ZR402G Converter
Calibration gasunit case
EXA ZR402G
Conduit *1
*2
*3
*4
100 to 240 V ACContact inputAnalog output, contact outputDigital output (HART)
Signal (6-core shield cable)
Heater (2-core shield cable)
~ 100 to 240 V ACAuto Calibration unitZ R20H
Reference gas
Calibration gas
Contact inputAnalog output, contact outputDigital output (HART)
Air Set
Instrument air
Calibration gasunit case
Calibration gas pressure regulator
Zero gas cylinder
Model ZR202G Intgrated type Zirconia Oxygen/ High temperature Humidity Analyzerwith auto calibration (ZR202G----A-----A)
2
GS 11M12A01-E-H
Process gas temperature 0 to 600 ºC ( 700 ºC with inconel bolts) Process gas temperature 0 to 1400 ºC
MountingInsertionlength
General-use Probe Application
Horizontalto
vertical
Horizontal
Vertical
0.4to2m
3m
• Boiler• Heating furnace• Low dust level
• For pulverized coil boiler with gas flow velocity 10 m/s or more• Cement kiln• High dust level
3m
0.4to2m
Horizontalto
vertical
1.0to2m
Horizontalto
vertical
• Boiler• Heating furnace• Low dust level
• Black liquid recovery boiler• Cement kiln• High dust level
Vertical 3m
High temperature detector
• Heating furnace• 0 to 5.0kPa
- 0.5 to 0kPa
General-purposedetector ZR22G-
General-purpose detector with probe supporter ZO21R
General-purpose detectorwith protector and filterwith: ZO21R K9471 UA
General-purpose detectorwith filter K9471 UA
ZR22G-015ZO21P-HNeedle valvenot supplied
ZR22G-015ZO21P-HE7046EC
* Note: Downward oriented mounting preferred
*
Examples of system component selection based on sample gas condition
3
GS 11M12A01-E-H
STANDARD SPECIFICATIONS (Oxygen analyzer)
General specificationsMeasurement range : 0.01 to 100-vol% O2Output signal : 4- 20 mA DC, maximum load 550 OhmOutput span : 0-5 vol% and for zero suppression any
span with minimum of 0.3 x concentration at 4 mA
Digital Communication: Frequency Shift Key following HART
protocol; 250- 550 Ohm, depending onnumber of field devices (multi drop mode)
Warm-up time : 20 minutes
Performance specificationsRepeatability : ± 0.5% FS for set range <25 vol%
± 1% FS for set range >25 vol%Linearity : ± 1% FS for set range <25 vol%
± 3% FS for set range <50 vol%± 5% FS for set range <100 vol%
Long term stability : ± 2% FS per month for zero and spanResponse time : 90% within 5s measured from calibration
gas inlet
Conditions : Dry instrument air used as reference air Pressure fluctuations < 0 ± 0.49 kPaMeasurements > 2 weeks after firstinstallation
1. ZR402 Remote Converter of Zirconia analyzerDisplay : Graphical display 1/4 VGA (320 x 240
dots) with touch panelCurrent output : (2) current outputs 4- 20 mA DC with
HART on mA1: max. load 550ΩContact output : (4) contacts max. load 3A @ 30 VDC and
3A @ 250VAC resistive load with leakagecurrent <3 mA with adjustable hysteresis(0- 9.9vol%) and delay function (0- 250s)One contact output is fail safe, 3 contactsselectable NO or NC
Additional Contacts : (2) for live output to calibrating unit (zerogas and span gas)
Contact functions : HIGH, LOW, FAIL, HOLD, CAL, rangeswitch answer back), Warm-up, Start blow-back, Flame out signal (answer back), Calgas pressure (answer back), High Temp.
Contact inputs : (2) programmable for Cal gas pressure,Range switching, Cal start, Blow backstart, system failure.
Analog input : 4- 20 mA @24 VDC for Temperature inputAmbient temperature : -20 to + 55 ºCStorage temperature : -30 to + 70 ºCHumidity : 10 to 95%Enclosure : IP65, NEMA 4XAltitude : Below 2000 mPower supply : 100- 240 VAC, –15% + 10%, 50/60 Hz ±10%Power consumption : Maximum 300 W, normal operation 100WCE conformance : Conforms to EN61010-1: 1993; EN61326: 1998Cable entry : (8) inlets M20x 1.5 other connections on
requestInstallation : Pipe mounting DN50/ 2" stand pipe or Wall
mountMaterial : Cast Aluminum, Polyurethane coating, off-
white (case) and moss green (cover)Weight : 6 kgDistance from detector
: maximum resistance 10 Ohm 2-way.(250m for one mm2 wire)
2. ZR202 Integral Zirconia AnalyzerDisplay : 6-digit LCDOperation : 3 optical switches that can be operated
without opening the coverAnalog output : 4- 20 mA, maximum load 550 Ohm.Digital communication: HART on mA: 250- 550 OhmContact output : (2) contacts: one is fail safe, Normally open
(NO)Contact input : (2) contactsWeight : 3 kg + 3 kg (flange PN10DN50) + 4 kg/m
insertionInsertion length : 0.4, 1.0, 1.5 and 2.0 m
All other specifications of ZR22 and ZR402 are applicable forZR202 as well.
3. ZR22G Detector of Zirconia AnalyzerSample gas temperature
: 0- 600 ºC for standard version600- 700 ºC for version with Inconel bolts700- 1400 ºC for high temperature version(0.15 m) with High Temperature probeadapter.
Ambient temperature: -20 to +150 ºC
Sample gas pressure : 0 ± 5 kPa for standard models withreference air by natural convection orinstrument air, -5 to 250 kPa for versionwith pressure compensation constructionPressure fluctuations influence the accuracyunless pressure compensation is applied
Insertion length : 0.15, 0.4, 1.0, 1.5, 2.0 m other insertionlengths on request
Material of probe : Stainless Steel 316 for standard detectorStainless steel 310S for high temperatureprobe (<1000 C) Silicium Carbide for hightemperature probe (< 1400 C)
Exposed materials : SUS 316 (detector), SUS 304 (flange),Hastelloy B (sensor flange), ZirconiumOxide, Platinum, (Inconel)
Reference air system: Natural convection; instrument air orpressure compensation
Instrument air : Dry and clean, recommended pressure200 kPa, flow rate 1 Nl/min.
Construction : Heater and thermocouple unit replaceable(specify insertion length) General purpose,IP66/NEMA4X only with pressurecompensation version
Terminal box : Cast Aluminum, Polyurethane coating off-white (box) and moss green (cover)
Connection gas lines: R1/4 or 1/4" FNPTWiring connection : M20 x 1.5. Other connections on requestMounting : Horizontal to vertical downward. Horizontal
mounting of long detectors (>2.0 m) is onlypossible using probe supporter or probeprotector. (<3m)
Weight : 1 +3 kg (flange DIN PN10DN50) + 4 kg/minsertion.
4
GS 11M12A01-E-H
4. ZO21P-H High Temperature Probe Adapterfor ZR22G detector
Description : The ZO21P-H high temperature probeadapter consists of a Stainless steel TEEpiece with JIS5k32AFF flange forconnection to the ZR22G-015 detector.The process side of the TEE is blinded offwith a flange DIN PN10DN50 with an R _thread for mounting of the sampling probe.The other side of the TEE is blinded offwith a flange and supplies a R1/4 or 1/4"FNPT connection for ejector or valve.
Sample temperature : 0-1400 ºC for SiC sampling probe0- 800 ºC for SUS 310S sampling probe
Sample pressure : -0.5 to + 5.0 kPa. For negative samplepressure an auxiliary ejector assembly isnecessary.
Insertion length : 1.0, 1.5 mProcess connection : DIN PN10 DN50 flanged connectionMaterial : SUS 316 (adapter), SUS 304 (flange), SiC
or SUS310S (sampling probe)Mounting : Vertically downward sampling probe with
zero ± 5 degree from vertical (SiC samplingprobe. Horizontal mounting with SUS310Ssampling probe is possible
Weight : 7 kg for 1m version; 8 kg for 1.5m version
5. E7046EC/E7046EN Auxiliary ejector forZo21P-H probe adapter
Ejector assembly : Inlet air pressure 29 to 68 kPa, airconsumption 30- 40 Nl/min, sample flowrate 3- 7 Nl/min, material SUS304,connection Rc 1/4 or 1/4" NPT, tubeconnection compression fitting for 6 mmOD copper or SS tubing.
Pressure gauge : 0- 100 kPa G, materials in contact with gasSUS316, case material coated Aluminum.
Needle valve : Material SUS316, connection R 1/4 or 1/4"NPT
6. ZO21R Probe Protector for Zirconia OxygenAnalyzer
Used when sample gas flow velocity is approx. 10m/sec or moreand dust particles wears the detector in cases such as pulverizedcoal boiler of fluidized bed furnace (or burner) to protect thedetector from wearing by dust particles. When probe insertionlength is 2.5m or more and horizontal installation, specify theZO21R-L-uuu-u*B to reinforce the probe.
Insertion Length : 1.05m, 1.55m, 2.05m.Flange : JIS 5K 65A FF equivalent. ANSI CLASS
150-4-FF (without serration) equivalent orDIN PN10-DN50-A equivalent. However,flange thickness is different.
Material : SUS316, SUS304 (Flange)Weight : 1.05m; Approx. 6/10/8.5kg (JIS/ANSI/DlN),
1.55m; Approx. 9/13/11.5kg (JIS/ANSI/DIN),2.05m; Approx. 12/16/14.5kg (JIS/ANSI/DIN)
Installation : Bolts, nuts, and washers are provided fordetector, probe adapter and process-sideflange.
7. Filter for Oxygen Analyzer K9471UAThis filter is used to protect the cell from corrosive dustcomponents or high velocity dust in recovery boilers and cementkiln. Measured gas flow rate is needed to be 1m/sec or more toreplace gas inside zirconia sensor.
Mesh : 30 microns Material : Carborundum (Filter), SUS316Weight : Approx. 0.2 kg
8. L9852CB/K9471UN Stop ValveThe stop valve is mounted on the calibration gas line.It is attached when the suffix code (/SV) is selected for theZirconia Oxygen Analyzer/High-temperature Humidity Analyzerprove ZR22G or the Zirconia Oxygen Analyzer/High-temperatureHumidity Analyzer ZR202G.
Connection : Rc 1/4 or 1/4 FNPT Material : SUS/BrassWeight : Approx. 80 g
9. K9292DN/K9292DS Check Valve This is used to prevent entry of process gas into calibration gasline. Purpose is the same as stop valve, but is convenient, as itdoes not need to be opened or closed for calibration.Mount directly on calibration gas inlet of detector in place of stopvalve. However as source pressure of 50 kPa G or more isneeded.When option code “/CV” of the ZR22G or the ZR202G isspecified, check valve is provided.
Connection : Rc1/4 or 1/4FNPTMaterial : SUS304Pressure : Between 70 kPa G or more 350 kPa G or
lessWeigh : Approx. 40g
*10. ZA8F Flow Setting Unit Used when instrument air is provided. This unit controls flowrates of calibration gas and reference gas and consists offlowmeter and flow rate control valve.
FIowmeter : Calibration gas; 0.1 to 1.0 l/min. Reference
air; 0.1 to 1.0 l/min.
Construction : Dust-proof and rainproof construction
Case Material : SPCC, Dark-green (Munsell 2.0 GY 3.1/0.5
or equivalent)
Painting : Baked epoxy resin
Tube Connections : Rc 1/4 or 1/4 FNPT
Reference Air pressure: Clean air supply of measured gas
pressure+approx. 50 kPa G (pressure at
inlet of the auto-calibration unit)
Air Consumption : Approx. 1.5 l/min
Weight : Approx. 2kg
Note : Used instrument air for span calibration
gas, if without instrument air is used,
contact YOKOGAWA.
* Recommended calibration gas = 1% Oxygen in Nitrogen FlowRate 0.6 ±0.1 Nl/min
5
GS 11M12A01-E-H
*11. ZR40H Auto-calibration Unit (for Separatetype)
This unit is used when the instrument air is provided and theauto-calibration unit is attached. Solenoid valves are provided as standard equipment.
Reference air pressure: Clean air supply of measured gas
pressure+approx. 150 kPa G (pressure atinlet of the auto-calibration unit)
Air consumption : Approx. 1.5 l/minWeight : Approx. 3.5 kg
*12. ZR20H Auto-calibration Unit (for Integratedtype
Used when automatic calibration is specified for the integratedtype and instrument air is provided.Equipped with the analyzer when automatic calibration isspecified in the suffix code of the ZR202G Integrated type byselecting either “-A (Horizontal mounting)” or “-B (Verticalmounting)”. The ZR20H should be arranged when auto-calibration is to be required after the ZR202H has been installed.Ask Yokogawa service station for its mounting.
Construction : Dust-proof and rainproof construction:NEMA4X/IP67 (excluding flowmeter)
Mounting : Mounted on ZR202G, no vibrationMaterials : Body: Aluminum alloy, Piping: SUS316,
SUS304, Flowmeter: MA (Methacrylateresin)
Finish : Polyurethane corrosion-resistance coating,Case: Frosty white CC21, Cover: Deep-seamoss green CC32
Piping Connection : Refer to Model and Suffix CodesPower Supply : 24V DC (from ZR202G), Power
consumption: 1.3 WReference Air Pressure
: Sample gas pressure + Approx. 150 kPa(690 kPa max.), (Pressure at inlet of auto-calibration unit)
Air Consumption : Approx. 1.5 l/minWeight : Approx. 2 kgAmbient Temperature
: -20 to +55°C, no condensing and freezingAmbient Humidity : 0 to 95%RHStorage Temperature: -30 to +65°C
13. Dust Protector ZH21BThis protector is designed to protect the detector Tip fromingress of dust (i.e., to prevent combustible materials fromentering the detector cell) where humidity measurements aremade under dusty environements.
Insertion length : 0.428 mFlange : JIS 5 K 80 A FF equivalent or ANSI 4 B
150 LB FF equivalent (however, thickness isdifferent)
Materials of Flange : SUS 316, SUS 304, Stainless SteelWeight : approximately 6/8.5 kg (JIS/ANSI)Mounting : mounted on the detector, and process
flange with bolts and associated nuts andwashers
* Recommended calibration gas = 1% Oxygen in Nitrogen FlowRate 0.6 ±0.1 Nl/min
14. ZR20-CAL – Standard Local Calibration unitEasy to use and lightweight portable unit for calibration gassupply consisting of span gas (air) and zero gas.Included in this set:
- 2 span gas alucan light weight bottles (one as a spare)- 2 zero gas alucan light weight bottles (one as a spare)- 2 constant flow regulators with quick connectors, set to 0.5 L/min- 1 m tubing with quick connectors and nipple for direct connection
to the sensor- Holder for gas bottles- Carrying case
Capacity 0.5 literFilled pressure 120 bar (60 liter gas)Composition Span gas: 21% O2 (air), zero gas: 1%
Oxygen in nitrogenWeight approx. 8 kgMaterial Holder PVCMaterial Bottles Aluminium
6
GS 11M12A01-E-H
Configurations : All configurations are possible, but insertion length of 0.4m is most common
Options : ZH21B Dust protector for high dust environment. Insertion length 0.416 m flange JIS 5k-80 or ANSI 150# 4",material SUS 316, Carborundum (SiC), SUS 304 (flange), weight 6 kg
Standard specifications for use as HumidityAnalyzer
Description : The Zirconia Oxygen analyzer can also beused to monitor the absolute humidity in anenvironment where standard air is mixedwith water vapor. Standard air consists of79% Nitrogen and 21% oxygen. If the air ismixed with Water vapor, the vol% ofoxygen decreases as the vol% of Waterincreases. The Zirconia based Oxygenanalyzer is therefore very useful as ananalyzer for moisture content of the samplegas. Both the ZR402 and the ZR202 canbe software switched from Oxygen analysisto Humidity analysis.
Sample composition : The sample gas may only contain a mixtureof Oxygen, Nitrogen and Water.
Performance : The performance specifications aredetermined by the primary measurement:the measurement of Oxygen. The possibleerrors in humidity units can be calculatedfrom the possible errors in the Oxygenanalysis.
Range : 1. Percent by volume: range 0- 100%,minimum span 0- 25 vol%2. Weight Ratio: kg H2O/ kg dry air: range1- 1.00, min span 0- 0.200 kg/kg dry air3. Dew point: range 0- 100 ºC4. Relative Humidity: range 0- 100%
Calculation
air
0% 100%water
100 %
O2
H2O
N2
0%100%
0%
x
a
Calculation Dew Point, calculation rH
7
GS 11M12A01-E-H
ZR402G
Converterthread
-M M20x1.5 mm
Separate type Zirconia Oxygen/ High temperature Humidity Analyzer, Converter
Model Suffix code Optioncode Description
Tag plates *1 For humidity measurements, be sure to specify /HS options.*2 Specify either /SCT or /PT option code.*3 Will b l bl i th f t
-E-G
Display
-A
-E
/HS
/H
Instruction manual
Options
/SCT/PT
Set for Humidity Analyzer (*1)
Hood (*3)
English
EnglishGerman
Always -A
Stainless steel tag plate (*2)Printed tag plate (*2)
MODEL AND SUFFIX CODES
1. Separate type Zirconia Oxygen / Hightemperature Humidity Analyzer, Converter
ZR202G
ANSI CLASS150-4-RF (Equivalent)DIN PN10-DN50-A (Equivalent)
-C-E
-040-070-100-150-200
-S-C
0.4 m0.7 m1.0 m1.5 m2.0 m
-C-E
Rc 1/4-R
M20x1.5 mm-M
-A
-E
/C
Set for Humidity Analyzer (*4)/HS
/CV/SV
Printed tag (*7)/PT
Dust Filter (*6)/F1
Auto Calibration No auto calibration unit mountedAuto calibration unit mountedVertical mounting
-N-A-B
integrated type Zirconia Oxygen/ High temperature Humidity Analyzer
Model Suffix code Optioncode Description
Length
Wetted material SUS316Stainless steel with Inconel calibration gas tube (*7)
Flange(*2)
Reference air Natural convectionPressure compensation
Gas Thread
Connection box thread
Instruction manual English
Always -A
Options
Inconel bolt (*3)
Check valve (*5)Stop valve (*5)
Durethane coating/D
*1 For the horizontal installed probe whose insertion length is 2.5 meters or more, use the Probe Protector. Be sure to specify ZO21R-L--. Specify the flange suffix code either -C or -K.*2 The thickness of the flange depends on its dimensions.*3 Inconel probe bolts are used. Use this option for high temperature use (ranging from 600 to 700 C). *4 For humidity measurements, be sure to specify /HS options. Pressure compemsation of reference air can not be selected.*5 Specify either /CV or /SV option code.*6 Not used with the high-temperature humidity analyzer.*7 Recommended if measured gas contains corrosive gas like chlorine.
Stainless steel tag plate (*7)/SCT
2. Integrated type Zirconia Oxygen / High temperature Humidity Analyzer, Converter
8
GS 11M12A01-E-H
3. Separate type Zirconia Oxygen / High Temperature Humidity Analyzer, Detectors
Model
ZR22G
Flange(*3)
ANSI CLASS150-4-RF (Equivalent)DIN PN10-DN50-A (Equivalent)JIS 5K-32-FF (Equivalent) (for high temperature use) (*4)
-C-E-Q
Separate type Zirconia Oxygen/ High temperature Humidity Analyzer, Detector
Length
Wetted material
-015-040-070-100-150-200
-S-C
0.15 m (for high temperature use) (*1)0.4 m0.7 m1.0 m1.5 m2.0 m
SUS316Stainless steel with Inconel calibration gas tube (*10)
Suffix code Optioncode Description
Reference air Natural convectionPressure compensation
-C-P
Gas Thread Rc 1/4-R
Connection box thread
Instruction manual
M20 x1.5 mm-M
Always -A-A
English-E
Options
Valves
Filter
Tag plates
*1 Used with the ZO21P High temperature Probe Adapter. Select flange (-Q).*2 When installing horizontally the probe whose insertion length is 2.5 meters or more, use the Probe Protector. Be sure to specify ZO21R-L--. Specify the flange suffix code either -C or -K.*3 The thickness of the flange depends on its dimensions.*4 Not used in conjunction with —P (pressure compensation) for reference air.*5 Inconel probe bolts are used. Use this option for high temperature use (ranging from 600 to 700 C). *6 Specify either /CV or /SV option code.*7 Not used with the high temperature humidity analyzer.*8 Specify either /SCT or /PT option code.*9 No water proof,avoid rain. Operating maximun temperature is 80C.*10 Recommended if measured gas contains corrosive gas like chlorine.Note: ZR22G can not be connected to Model ZA8C, AV8C, HA400 converter.
Inconel bolt (*5)/C
Durethane coating/D
Check valve (*6)Stop valve (*6)
/CV/SV
Dust Filter (*7)/F1
Stainless steel tag plate (*8)Printed tag plate (*8)
/SCT/PT
9
GS 11M12A01-E-H
Joint: RC 1/4, Material: SUS316
Joint: 1/4 NPT, Material: SUS316
DescriptionPart No.
L9852CB
G7016XH
Joint: RC 1/4, Material: SUS304
Joint: 1/4 NPT, Material: SUS304
Part No.
K9292DN
K9292DS
Description
ZR40H
-R
-M
Rc 1/4
Model Suffix code Option code Description
Automatic calibration unit for ZR402G
Gas piping connection
Wiring connection
- -A
20 mm (M20 x 1.5)
Always -A
ZR20H
-R
-E
Rc 1/4
Model Suffix code Option code Description
Automatic calibration unit for ZR202G*1
Gas piping connection
Reference air *2
- -A
Instrument air
Always -A
-P Pressure compensated
Mounting -A Horizontal mounting
-B Vertical mounting
*1 Ask Yokogawa service station for additional mounting of ZR20H to the preinstalled ZR202G.*2 Select the appropriate reference air of ZR20H according to the one of ZR202G.
11. Automatic Calibration Unit for Separatetype Analyzer (Needs instrument air.)
12. Automatic Calibration Unit for Integratedtype Analyzer (Needs instrument air.)
8. Stop Valve for Calibration-gas line
9. Check Valve for Calibration-gas line
1.0 m1.5 m
DIN PN10-DN50-A equivalentANSI Class 150-4-RF
ZO21P
SiCSUS 310S
-100-150
*A
-E-A
-H
-A-B
High Temperature Probe Adapter
Material
Insertionlength
Flange
Style AStyle code
Model Suffix code Optioncode Description
ZH21B
* The flange thickness varies. Specify the probe ZR22G-040.
DescriptionOption codeSuffix codeModel
Dust protector (0 to 600C)
0.409 mInsertion length -040
Flange
Style code *B
-A ANSI Class 150-4B FF equivalent *
Style B
ZA8F
-J-A
Model Suffix code Option code Description
Flow setting unit
JointWith 1/4" NPT adapter
Style B
Rc 1/4
Style code *B
Part No.
E7046AL
E7046BB
E7046AP
E7046AQ
SiC, insertion length 1.0 m
SiC, insertion length 1.5 m
SUS310S, insertion length 1.0 m
SUS310S, insertion length 1.5 m
Description
RC 1/4 6 / 4 TUBE joint: SUS304
Part No.
E7046EC
Description
ZO21R
DIN PN10-DN50-A equivalent
-L Probe Protector(0 to 700 8C)
-100-150-200
1.05 m (3.5ft)1.55 m (5.1ft)2.05 m (6.8ft)
Insertionlength
Flange ( *1)
*1 Thickness of flange depends on dimensions of flange.
ANSI Class 150-4-RF equivalent
Model Suffix code Optioncode Description
*B Style BStyle code
-E-A
Part No.
K9471UA
Description
Filter
4. Adapter for High temperature Probe ofseparate type Oxygen Analyzer
Note: For this high-temperature use probe adapter, be sure tospecify the ZR22G probe of it’s insertion length 0.15meters.
High temperature Probes (Spare Parts)
5. Auxiliary Ejector for High-temperature Useof separate type Oxygen Analyzer
7. Filter for Zirconia Oxygen Analyzers
13. Dust Protector for High-temperatureHumidity Analyzers
10. Flow setting unit for manual calibration(Needs instrument air.)
6. Probe Protector for Zirconia Oxygen Analyzer
Model DescriptionZR20-cal Standard Calibration Unit
Spare parts DescriptionZR20-zero Spare Zero Gas bottleZR20-span Spare Span Gas bottle
14. ZR20-CAL – Standard Local Calibration unit
10
GS 11M12A01-E-H
1 to 6 (Panel Thickness)
EXA ZR402G 4 - 6 holesfor Wall mounting
4-R8 to R10or4-C5 to C8
4 - 6 holes
8-G1/2, 8-1/2NPT etc(Wiring connection)
( for wall mounting)
F08_01.EPS
2-inch mounting pipe
Wall mounting Panel Cut-out
100
108228
280
126.
5
57.3
136.3 54.7
111
36
120.
2
(1/2
NP
T)
190 183+20
274
+2
0
126.
5
1. Model ZR402G Separate type Zirconia Oxygen/ High temperature Humidity Analyzer, Converter
11
GS 11M12A01-E-H
153 to 164
125
L
170252 to 265 49
48.5
338 to 351
25
t
122
L= 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 (m)
4-G1/2,2-1/2NPT etc.Cable connection port
Display side
1563
2564 49
3424
25
t
PIPING:A
PIPING:B
L
L= 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 (m)
4-G1/2,2-1/2NPT etc.Cable connection port
50
.8
Rc1/4 or 1/4NPTReference air inlet
Rc1/4 or 1/4NPTCalibration gas inlet
Reference air outlet
Stopvalve
PIPINGABBABBAABBBBA
ANSI CLASS150-2-RF equivalentANSI CLASS150-3-RF equivalentANSI CLASS150-4-RF equivalentDIN PN10-DN50-A equivalentDIN PN10-DN80-A equivalentDIN PN10-DN100-A equivalentJIS 5K-65-FFJIS 10K-65-FFJIS 10K-80-FFJIS 10K-100-FFJPI CLASS150-4-RF equivalentJPI CLASS150-3-RF equivalentWestinghouse
152.4190.5228.6165200220155175185210229190155
120.6152.4190.5125160180130140150175
190.5152.4127
4 - 194 - 198 - 194 - 188 - 188 - 184 - 154 - 198 - 198 - 198 - 194 - 19
4 - 11.5
A B C19242418202014181818242414
tFlange
5
0.8
Rc1/4 or 1/4NPTReference air inlet
Rc1/4 or 1/4NPTCalibration gas inlet
123
Terminal side
C
Flange
Flange
B
C
Flange
A
A
B
2. Model ZR202G Integrated type Zirconia Oxygen/ High temperature Humidity Analyzers
Model ZR202G...-P (with pressure compensation) Integrated type Zirconia Oxygen / High temperatureHumidity Analyzers
12
GS 11M12A01-E-H
25
t285 4
50
.8
12
7
85
L
48
1563 69
L=0.15, 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 3.6, 4.2, 4.8, 5.4 (m)
C
Reference gas outlet
25 48
153 to 164 69
Rc1/4 or 1/4NPTReference air inlet
2-G1/2,2-1/2NPT etc.Cable connection port
Rc1/4 or 1/4NPTCalibration gas inletFlange
PIPING : A
PIPING:B
Rc1/4 or 1/4NPTReference air inlet
2-G1/2, 2-1/2NPT etc.Cable connection port
B
A
C
Flange
Flange
B
A
Rc1/4 or 1/4NPTCalibration gas inlet
Flange
Stop Valve
EXTERNAL DIMENSIONS
3. Model ZR22G Separate type Zirconia Oxygen / High temperature Humidity Analyzer, Detectors
Model ZR22G...-P (with pressure compensation) Separate type Zirconia Oxygen / High temperatureHumidity Analyzer, Detectors
PIPINGANSI CLASS150-2-RF equivalentANSI CLASS150-3-RF equivalentANSI CLASS150-4-RF equivalentDIN PN10-DN50-A equivalentDIN PN10-DN80-A equivalentDIN PN10-DN100-A equivalentJIS 5K-65-FFJIS 10K-65-FFJIS 10K-80-FFJIS 10K-100-FFJPI CLASS150-4-RF equivalentJPI CLASS150-3-RF equivalentWestinghouse
152.4190.5228.6165200220155175185210229190155
120.6152.4190.5125160180130140150175
190.5152.4127
4 - 194 - 198 - 194 - 188 - 188 - 184 - 154 - 198 - 198 - 198 - 194 - 19
4 - 11.5
A B C19242418202014181818242414
ABBABBAABBBBA
tFlange
(m)
GS 11M12A01-E-H
13
R1/2(Note2)
Approx. 351
180 Flange (Thickness)JIS 5K 32A FF equivalent
Gasket (Thickness 3)
Detector(ZR22G)
Flange <1>
52 over
30
60.5
Reference air inlet <2>
Calibration gas inlet<3>
High temperatureProbe SiC pipe(GK material)
110
115
4225
170
App
rox,
215
(Ins
etio
n le
ngth
) (N
ote1
)
App
rox,
100
69
Approx. 48
C
(Note 1) 1=1000 or 1500 (mm)(Note 2) Sample gas outlet
(if the sample gas pressure is negative, connect the auxiliary ejector.)
127
85
Measurement gas outlet
Flange providedby customer
Unit: mm
60.5A
B
A
Pipe hole (2- G1/2, 2-1/2NPT, etc)
Needle valve
<1>
<2>
Pressure gauge
Nozzle (Note1)
R1/2Ejector assembly Detector
<1> Rc1/4 or 1/4 NPT<2> 6/ 4 or 1/4 inch copper tube (stainless) with ejector to connect<3> Rc1/4
4020Tee<1>
Blow<1>
43
App
rox.
67
App
rox.
88
39
App
rox.
70
38
Ful
l ope
n he
ight
(Note1) The connector of ejector assembly is a dedicated connector with nozzle function.
Measurement gas inlet
4. Model ZO21P Adapter for High temperature Probe of separate type Oxygen Analyzer
5. Model E7046EC, E7046EN Auxiliary Ejector for High temperature Use of separate type OxygenAnalyzer
<1> <2> <3>Rc1/4 or ø6/ø4 or 1/4” Rc1/41/4 NPT copper tube
(stainless) withejector to connect
14
GS 11M12A01-E-H
51
32
10
Caborundum filter
Increase of insertion length
Flange<1>
JIS 5K-65-FFequivalent
155
228.6
130
190.5
4- 15
8- 19
5
12
A B C t
ANSI CLASS150-4-FF equivalent
40
50
D
D
C
B
B
Flange <1>(with bolts, nuts, gasket and washer)
gasket (t1.5)
SUS316
Dimensions ofholes on opposing surface
A
Gas flow
Washer (12)Mounting nut (M12)
l (Insert length)
l=1050,1550,2050
6. Model ZO21R Probe Protector for Zirconia Oxygen Analyzers
7. Model K9471UA Filter for Oxygen Analyzer
K9292DN : Rc 1/4(A),R 1/4(B)K9292DS : 1/4NPT(A),1/4NPT(M)(B)
Approx. 19 Approx. 54
55(F
ull o
pen
leng
th)
40
Rc1/4 or 1/4NPT
43
8. L9852CB/ K9471UN Stop Valve forCalibration-gas line
9. K9292DN/ K9292DS Check Valve forCalibration-gas line
43
55
15
GS 11M12A01-E-H
6 hole
Reference air outlet Zero gas outlet Zero gas inlet Instrument air inlet
4-Rc1/4
20 2040 40
Piping connectionport
180
235.
870
35
268
222.
87
140
REFERENCE CHECK
REFERENCE SPAN ZERO
F20.EPS
200
10. Model ZA8F Flow setting unit for manual calibration
35 3535 35
Spangas inlet
16
GS 11M12A01-E-H
Unit: mm
OCK
140
250
42
26
41.2 41.2
54 71.526
49.5
1222
3
4616
3040 47.5 25
102
MA
X58
90 116.5
6.5
*1 with four ISO M6 screws can wall-mount
*1 4-
(wiring inlet is at same position on rear)
Flowmeter
Setting Valve forcalibration gas
Setting Valve forreference air
wiring inlet ; 2-G1/2,Pg13.5,M2031.5 or 1/2NPT(Female)
calibration gas outlet Rc1/4 or 1/4 NPT(Female)
reference air inlet
Rc1/4 or 1/4 NPT(Female)zero gas inlet
Rc1/4 or 1/4 NPT(Female)Rc1/4 or 1/4 NPT(Female)reference air outlet
2B pipe mounting example
11. Model ZR40H Automatic Calibration Unit for Separate type Analyzer
ZR402G Converter ZR40H Automatic Calibration unit
AC-Z
AC-C
AC-S
Zero
Span
EV1
EV2
CHECKOUT
ZERO GAS IN AIR IN
Instrument air Approx.1.5 l/min.
*2 Needle valve is supplied as accessory with flow meter
flow meter
flow meter
REFOUT
*2 *2solenoid valveEV1,2
EV1 EV2
17
GS 11M12A01-E-H
zero gas inlet
Rc1/4 or 1/4NPT(Female)
reference air inlet
Rc1/4 or 1/4NPT(Female)
terminal box side
span gas inlet
Rc1/4 or 1/4NPT(Female)
zero gas inlet
Rc1/4 or 1/4NPT(Female)
Span gas inlet
Rc1/4 or 1/4NPT(Female)
reference air inlet
Rc1/4 or 1/4NPT(Female)
166.5
166.5
258
60
244
214 MAX44
4516
0
66.540 40
66.540 40
180 MAX44
Vertical mounting on the ZR202G (-B)
Horizontal mounting on the ZR202G (-A)Unit: mm
display side
12. Model ZR20H Automatic Calibration Unit for Integrated type Analyzer
t
C
C
B
B
A
B
76.3
72
A
180
228.5
B
145
190.5
C
4-19
8-19
t
12
12
D
40
50
Unit: mm416 (Insertion length)
Install facing upwards.Flange
JIS 5 K FF equivalent
ANSI 4 B 150 LB FF
13. Model ZH21B Dust Protector for High temperature Humidity Analyzers Unit: mm
JIS 5K-80-A-FF equivalent
428
D
18
GS 11M12A01-E-H
1FG
2AO1(+)
3AO1(-)
4AO2(+)
5AO2(-)
6CJ(+)
7CJ(-)
8TC(+)
9TC(-)
10CELL
(+)
11CELL
(-)12FG
13DI-1
14DI-2
15DI-C
16AI(+)
17AI(-)
18AC-
ZERO
19AC-
SPAN
20AC-
COM
21FG
22FG
CELL(+)
CELL(-)
TC(+)
TC(-)
CJ(+)
CJ(-)
23DO-1
24DO-1
25DO-2
26DO-2
27DO-3
28DO-3
29DO-4
30DO-4
31HTR
32HTR
33L
34N
35G
36FG HTR HTR
100-240V AC50/60 Hz
Model ZR402G Separate type Zirconia Oxygen Analyzer/ Converter
Model ZR22G Separate type Zirconia OxygenAnalyzer, Detector
Analog output 2
Analog output 1
4-20 mA DC4-20 mA DCDigital output
Contact input 2
Contact input 1
Contact output 1
Contact output 2
Contact output 3
Contact output 4
Solenoid valve for span gas for automatic calibration
Solenoid valve for zero gas for automatic calibration
Flow setting unit*1
*1
*1
1DI-1
2DI-2
3DI-C
4DO-1
5DO-1
6DO-2
7DO-2
8FG
9AO(+)
10AO(-)
11L
12N
13G
14FG
Model ZR202G Integrated type Zirconia Oxygen / High temperature Humidity Analyzer
Contact input 1Contact output 2Contact output 1
Analog output 4-20 mA DC
100 to 240 V AC, 50 or 60 Hz
Digital output
Contact input 2
*1
*1 Ground resistence is 100 ohm or less.
WIRING CONNECTIONS
19
GS 11M12A01-E-H
Inquiry Sheet for Models ZR22G, ZR402G, and ZR202G Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers
vol%O2,
kg/kg,
C,
kPa,
m/sec,
g/Nm3,
ppm,
ppm,
ppm,
ppm,
Vol%H2O
Please place checkmarks in the appropriate boxes and fill in the necessary information in the blanks.
Customer
Destination of delivery
Plant name
Measurement points
2. Process conditions
2.1 Measurement gas components
2.2 Oxygen concentration
Moisture contents
2.3 Temperature
2.4 Pressure
2.5 Gas flow
2.6 Dust type, Size
2.7 Corrosive gas
2.8 Combustible gas
2.9 Others
3. Installation site conditions
3.1 Ambient temperature
3.2 Vibration
3.3 1 Probe installation location
3.4 Instrument air supply
3.5 Converter location
3.6 Cable length between probe and converter
3.7 Calibration method
4. Quotation data
Probe Model ZR22G General-use Probe Refer to the Probe Configuration for probe selection.Model ZO21P-H High temperature Use Probe Adapter
Model ZR402G Separate type Analyzer, Converter
Model E7046EC/E7046EN Auxiliary Ejector for high temperature use.
Model ZH21B Dust Protector for high temperature humidity analyzer.
Model ZO21R Probe Protector for Oxgen Analyzer
1. General information
Nor. Min. Max.
Nor.
Nor.
Nor.
Nor.
Nor.
Min.
Min.
Min.
Min.
Min.
Max.
Max.
Max.
Max.
No gas Gas , quantity
m quantity
No gas Gas , quantity
, quantity
, quantity
1. Around Probe temp. from to C, 2. Around Converter temp. from to C
No vibration Vibration
Furnace Stack Others
2 Probe position Horizontal Vertical Others
Indoor Outdoor Covered
3 Probe insertion length (m) 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0, Others
4 Flange JIS ANSI Others
Cannot be used. Can be used. kPa
Indoor Outdoor Covered (under roof)
meters
Manual Automatic
Quotation Qantity Description
Options (for general use)
Model K9471UA Filter
Model ZR202G Integrated type Zirconia Oxygen / High temperature Humidity Analyzer
Model ZO21S Standard Gas Unit Select any one of Model ZO21S, ZA8F, ZR40H .
Model ZA8F Flow Setting Unit
Model ZR40H Automatic Calibration Unit Not required if probe options are specified.
L9852CB/K9471UN Stop Valve
K9292DN/K9292DS Check Valve
G7011XF/E7040EL Air Set
G7001ZC Zero Gas Cylinder
G7013XF/G7014XF Pressure Regulator
Type of analyzer : Oxygen Analyzer High temperature Humidity Analyzer
Sparate type Integrated type
Object : indication record control alarm
Fuel : gas oil coal
Power requirements V AC Hz
GS 11M12A01-E-HSubject to change without notice Printed in The Netherlands, 04-401 (A) QCopyright©
YOKOGAWA EUROPE B.V.Databankweg 203821 AL AMERSFOORTThe NetherlandsTel. +31-33-4641 611Fax +31-33-4641 610E-mail: [email protected]
YOKOGAWA CORPORATION OF AMERICA2 Dart RoadNewnan GA 30265United StatesTel. (1)-770-253-7000Fax (1)-770-251-2088E-mail: [email protected]
YOKOGAWA ELECTRIC ASIA Pte. Ltd.5 Bedok South RoadSingapore 469270SingaporeTel. (65)-241-9933Fax (65)-241-2606E-mail: [email protected]
YOKOGAWA HEADQUARTERS9-32, Nakacho 2-chome,MusashinoshiTokyo 180JapanTel. (81)-422-52-5535Fax (81)-422-55-1202E-mail: [email protected]
Yokogawa has an extensive sales anddistribution network. Please refer to the European web-site(www.yokogawa-europe.com) to contactyour nearest representative.