Model ZR22G, ZR402G, and ZR202G Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers GS 11M12A01-E-H 4th Edition The basic analyzer consists of a direct insertion type probe and a converter that provides an analog output for control or monitoring purposes. In addition a digital signal is superimposed on the current output (HART) for optimization of maintenance. The converter controls also the (semi) automatic calibration when used with the calibration units. The detector and the converter are combined in one unit (ZR202) for simplified piping and wiring or the converter is mounted separately (ZR402) Various accessories are available for optimal installation in difficult applications ( dust filter, flame arrester, probe protector, high temperature probe adapter, probe supporter) and for ease of maintenance and calibration. The analyzer is most commonly used to monitor the Oxygen level in combustion air to allow optimal burner control to avoid air pollution and save energy. The direct insertion installation is especially attractive from a maintenance point of view: sampling systems are not required. The analyzer can also be used to monitor the Humidity concentration at high temperatures in air with water as found in humidifying processes like Bakeries or in indirectly heated dryers as found in Paper machines. FEATURES Direct measurement: No sampling required Field replaceable unique Zirconia cell, made to last. Heavy duty sensor heater Rebuildable probe construction Various methods for reference air to optimize accuracy versus ease of installation Automatic compensation of duct pressure possible. Remote maintenance and set-up by digital communication (HART) General Specifications SYSTEM CONFIGURATION ZR22G ZR402G ZR202G Instrument air Calibration gas pressure regulator Air Set Needle valve flowmeter Reference gas Calibration gas Check valve ~ Zero gas cylinder Model ZR40H Auto Calibration unit Model ZR22G Separate type Zirconia Oxygen / High temperature Humidity Analyzer, Detecor Model ZR402G Converter Calibration gas unit case Conduit *1 *2 *3 *4 100 to 240 V AC Contact input Analog output, contact output Digital output (HART) Signal (6-core shield cable) Heater (2-core shield cable) ~ 100 to 240 V AC Auto Calibration unit Z R20H Reference gas Calibration gas Contact input Analog output, contact output Digital output (HART) Air Set Instrument air Calibration gas unit case Calibration gas pressure regulator Zero gas cylinder Model ZR202G Intgrated type Zirconia Oxygen / High temperature Humidity Analyzer with auto calibration (ZR202G----A-----A)

Transcript of General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen...

Page 1: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

Model ZR22G, ZR402G, and ZR202GDirect In Situ Zirconia Oxygen Analyzersand High temperature Humidity Analyzers

GS 11M12A01-E-H 4th Edition

The basic analyzer consists of a direct insertion type probe and aconverter that provides an analog output for control or monitoringpurposes. In addition a digital signal is superimposed on thecurrent output (HART) for optimization of maintenance. Theconverter controls also the (semi) automatic calibration when usedwith the calibration units.The detector and the converter are combined in one unit (ZR202)for simplified piping and wiring or the converter is mountedseparately (ZR402)Various accessories are available for optimal installation in difficultapplications ( dust filter, flame arrester, probe protector, hightemperature probe adapter, probe supporter) and for ease ofmaintenance and calibration.The analyzer is most commonly used to monitor the Oxygen levelin combustion air to allow optimal burner control to avoid airpollution and save energy. The direct insertion installation isespecially attractive from a maintenance point of view: samplingsystems are not required. The analyzer can also be used to monitor the Humidityconcentration at high temperatures in air with water as found inhumidifying processes like Bakeries or in indirectly heated dryers asfound in Paper machines.

FEATURES Direct measurement: No sampling required Field replaceable unique Zirconia cell, made to last. Heavy duty sensor heater Rebuildable probe construction Various methods for reference air to optimize accuracy

versus ease of installation Automatic compensation of duct pressure possible. Remote maintenance and set-up by digital communication

(HART)

GeneralSpecifications

SYSTEM CONFIGURATION

ZR22G ZR402G

ZR202G

Instrument air

Calibration gas pressure regulator

Air Set

Needle valveflowmeter

Reference gas

Calibration gas

Check valve ~

Zero gas cylinder

Model ZR40H Auto Calibration unit

Model ZR22G Separate type Zirconia Oxygen/ High temperature Humidity Analyzer, Detecor

Model ZR402G Converter

Calibration gasunit case

EXA ZR402G

Conduit *1

*2

*3

*4

100 to 240 V ACContact inputAnalog output, contact outputDigital output (HART)

Signal (6-core shield cable)

Heater (2-core shield cable)

~ 100 to 240 V ACAuto Calibration unitZ R20H

Reference gas

Calibration gas

Contact inputAnalog output, contact outputDigital output (HART)

Air Set

Instrument air

Calibration gasunit case

Calibration gas pressure regulator

Zero gas cylinder

Model ZR202G Intgrated type Zirconia Oxygen/ High temperature Humidity Analyzerwith auto calibration (ZR202G----A-----A)

Page 2: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

2

GS 11M12A01-E-H

Process gas temperature 0 to 600 ºC ( 700 ºC with inconel bolts) Process gas temperature 0 to 1400 ºC

MountingInsertionlength

General-use Probe Application

Horizontalto

vertical

Horizontal

Vertical

0.4to2m

3m

• Boiler• Heating furnace• Low dust level

• For pulverized coil boiler with gas flow velocity 10 m/s or more• Cement kiln• High dust level

3m

0.4to2m

Horizontalto

vertical

1.0to2m

Horizontalto

vertical

• Boiler• Heating furnace• Low dust level

• Black liquid recovery boiler• Cement kiln• High dust level

Vertical 3m

High temperature detector

• Heating furnace• 0 to 5.0kPa

- 0.5 to 0kPa

General-purposedetector ZR22G-

General-purpose detector with probe supporter ZO21R

General-purpose detectorwith protector and filterwith: ZO21R K9471 UA

General-purpose detectorwith filter K9471 UA

ZR22G-015ZO21P-HNeedle valvenot supplied

ZR22G-015ZO21P-HE7046EC

* Note: Downward oriented mounting preferred

*

Examples of system component selection based on sample gas condition

Page 3: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

3

GS 11M12A01-E-H

STANDARD SPECIFICATIONS (Oxygen analyzer)

General specificationsMeasurement range : 0.01 to 100-vol% O2Output signal : 4- 20 mA DC, maximum load 550 OhmOutput span : 0-5 vol% and for zero suppression any

span with minimum of 0.3 x concentration at 4 mA

Digital Communication: Frequency Shift Key following HART

protocol; 250- 550 Ohm, depending onnumber of field devices (multi drop mode)

Warm-up time : 20 minutes

Performance specificationsRepeatability : ± 0.5% FS for set range <25 vol%

± 1% FS for set range >25 vol%Linearity : ± 1% FS for set range <25 vol%

± 3% FS for set range <50 vol%± 5% FS for set range <100 vol%

Long term stability : ± 2% FS per month for zero and spanResponse time : 90% within 5s measured from calibration

gas inlet

Conditions : Dry instrument air used as reference air Pressure fluctuations < 0 ± 0.49 kPaMeasurements > 2 weeks after firstinstallation

1. ZR402 Remote Converter of Zirconia analyzerDisplay : Graphical display 1/4 VGA (320 x 240

dots) with touch panelCurrent output : (2) current outputs 4- 20 mA DC with

HART on mA1: max. load 550ΩContact output : (4) contacts max. load 3A @ 30 VDC and

3A @ 250VAC resistive load with leakagecurrent <3 mA with adjustable hysteresis(0- 9.9vol%) and delay function (0- 250s)One contact output is fail safe, 3 contactsselectable NO or NC

Additional Contacts : (2) for live output to calibrating unit (zerogas and span gas)

Contact functions : HIGH, LOW, FAIL, HOLD, CAL, rangeswitch answer back), Warm-up, Start blow-back, Flame out signal (answer back), Calgas pressure (answer back), High Temp.

Contact inputs : (2) programmable for Cal gas pressure,Range switching, Cal start, Blow backstart, system failure.

Analog input : 4- 20 mA @24 VDC for Temperature inputAmbient temperature : -20 to + 55 ºCStorage temperature : -30 to + 70 ºCHumidity : 10 to 95%Enclosure : IP65, NEMA 4XAltitude : Below 2000 mPower supply : 100- 240 VAC, –15% + 10%, 50/60 Hz ±10%Power consumption : Maximum 300 W, normal operation 100WCE conformance : Conforms to EN61010-1: 1993; EN61326: 1998Cable entry : (8) inlets M20x 1.5 other connections on

requestInstallation : Pipe mounting DN50/ 2" stand pipe or Wall

mountMaterial : Cast Aluminum, Polyurethane coating, off-

white (case) and moss green (cover)Weight : 6 kgDistance from detector

: maximum resistance 10 Ohm 2-way.(250m for one mm2 wire)

2. ZR202 Integral Zirconia AnalyzerDisplay : 6-digit LCDOperation : 3 optical switches that can be operated

without opening the coverAnalog output : 4- 20 mA, maximum load 550 Ohm.Digital communication: HART on mA: 250- 550 OhmContact output : (2) contacts: one is fail safe, Normally open

(NO)Contact input : (2) contactsWeight : 3 kg + 3 kg (flange PN10DN50) + 4 kg/m

insertionInsertion length : 0.4, 1.0, 1.5 and 2.0 m

All other specifications of ZR22 and ZR402 are applicable forZR202 as well.

3. ZR22G Detector of Zirconia AnalyzerSample gas temperature

: 0- 600 ºC for standard version600- 700 ºC for version with Inconel bolts700- 1400 ºC for high temperature version(0.15 m) with High Temperature probeadapter.

Ambient temperature: -20 to +150 ºC

Sample gas pressure : 0 ± 5 kPa for standard models withreference air by natural convection orinstrument air, -5 to 250 kPa for versionwith pressure compensation constructionPressure fluctuations influence the accuracyunless pressure compensation is applied

Insertion length : 0.15, 0.4, 1.0, 1.5, 2.0 m other insertionlengths on request

Material of probe : Stainless Steel 316 for standard detectorStainless steel 310S for high temperatureprobe (<1000 C) Silicium Carbide for hightemperature probe (< 1400 C)

Exposed materials : SUS 316 (detector), SUS 304 (flange),Hastelloy B (sensor flange), ZirconiumOxide, Platinum, (Inconel)

Reference air system: Natural convection; instrument air orpressure compensation

Instrument air : Dry and clean, recommended pressure200 kPa, flow rate 1 Nl/min.

Construction : Heater and thermocouple unit replaceable(specify insertion length) General purpose,IP66/NEMA4X only with pressurecompensation version

Terminal box : Cast Aluminum, Polyurethane coating off-white (box) and moss green (cover)

Connection gas lines: R1/4 or 1/4" FNPTWiring connection : M20 x 1.5. Other connections on requestMounting : Horizontal to vertical downward. Horizontal

mounting of long detectors (>2.0 m) is onlypossible using probe supporter or probeprotector. (<3m)

Weight : 1 +3 kg (flange DIN PN10DN50) + 4 kg/minsertion.

Page 4: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

4

GS 11M12A01-E-H

4. ZO21P-H High Temperature Probe Adapterfor ZR22G detector

Description : The ZO21P-H high temperature probeadapter consists of a Stainless steel TEEpiece with JIS5k32AFF flange forconnection to the ZR22G-015 detector.The process side of the TEE is blinded offwith a flange DIN PN10DN50 with an R _thread for mounting of the sampling probe.The other side of the TEE is blinded offwith a flange and supplies a R1/4 or 1/4"FNPT connection for ejector or valve.

Sample temperature : 0-1400 ºC for SiC sampling probe0- 800 ºC for SUS 310S sampling probe

Sample pressure : -0.5 to + 5.0 kPa. For negative samplepressure an auxiliary ejector assembly isnecessary.

Insertion length : 1.0, 1.5 mProcess connection : DIN PN10 DN50 flanged connectionMaterial : SUS 316 (adapter), SUS 304 (flange), SiC

or SUS310S (sampling probe)Mounting : Vertically downward sampling probe with

zero ± 5 degree from vertical (SiC samplingprobe. Horizontal mounting with SUS310Ssampling probe is possible

Weight : 7 kg for 1m version; 8 kg for 1.5m version

5. E7046EC/E7046EN Auxiliary ejector forZo21P-H probe adapter

Ejector assembly : Inlet air pressure 29 to 68 kPa, airconsumption 30- 40 Nl/min, sample flowrate 3- 7 Nl/min, material SUS304,connection Rc 1/4 or 1/4" NPT, tubeconnection compression fitting for 6 mmOD copper or SS tubing.

Pressure gauge : 0- 100 kPa G, materials in contact with gasSUS316, case material coated Aluminum.

Needle valve : Material SUS316, connection R 1/4 or 1/4"NPT

6. ZO21R Probe Protector for Zirconia OxygenAnalyzer

Used when sample gas flow velocity is approx. 10m/sec or moreand dust particles wears the detector in cases such as pulverizedcoal boiler of fluidized bed furnace (or burner) to protect thedetector from wearing by dust particles. When probe insertionlength is 2.5m or more and horizontal installation, specify theZO21R-L-uuu-u*B to reinforce the probe.

Insertion Length : 1.05m, 1.55m, 2.05m.Flange : JIS 5K 65A FF equivalent. ANSI CLASS

150-4-FF (without serration) equivalent orDIN PN10-DN50-A equivalent. However,flange thickness is different.

Material : SUS316, SUS304 (Flange)Weight : 1.05m; Approx. 6/10/8.5kg (JIS/ANSI/DlN),

1.55m; Approx. 9/13/11.5kg (JIS/ANSI/DIN),2.05m; Approx. 12/16/14.5kg (JIS/ANSI/DIN)

Installation : Bolts, nuts, and washers are provided fordetector, probe adapter and process-sideflange.

7. Filter for Oxygen Analyzer K9471UAThis filter is used to protect the cell from corrosive dustcomponents or high velocity dust in recovery boilers and cementkiln. Measured gas flow rate is needed to be 1m/sec or more toreplace gas inside zirconia sensor.

Mesh : 30 microns Material : Carborundum (Filter), SUS316Weight : Approx. 0.2 kg

8. L9852CB/K9471UN Stop ValveThe stop valve is mounted on the calibration gas line.It is attached when the suffix code (/SV) is selected for theZirconia Oxygen Analyzer/High-temperature Humidity Analyzerprove ZR22G or the Zirconia Oxygen Analyzer/High-temperatureHumidity Analyzer ZR202G.

Connection : Rc 1/4 or 1/4 FNPT Material : SUS/BrassWeight : Approx. 80 g

9. K9292DN/K9292DS Check Valve This is used to prevent entry of process gas into calibration gasline. Purpose is the same as stop valve, but is convenient, as itdoes not need to be opened or closed for calibration.Mount directly on calibration gas inlet of detector in place of stopvalve. However as source pressure of 50 kPa G or more isneeded.When option code “/CV” of the ZR22G or the ZR202G isspecified, check valve is provided.

Connection : Rc1/4 or 1/4FNPTMaterial : SUS304Pressure : Between 70 kPa G or more 350 kPa G or

lessWeigh : Approx. 40g

*10. ZA8F Flow Setting Unit Used when instrument air is provided. This unit controls flowrates of calibration gas and reference gas and consists offlowmeter and flow rate control valve.

FIowmeter : Calibration gas; 0.1 to 1.0 l/min. Reference

air; 0.1 to 1.0 l/min.

Construction : Dust-proof and rainproof construction

Case Material : SPCC, Dark-green (Munsell 2.0 GY 3.1/0.5

or equivalent)

Painting : Baked epoxy resin

Tube Connections : Rc 1/4 or 1/4 FNPT

Reference Air pressure: Clean air supply of measured gas

pressure+approx. 50 kPa G (pressure at

inlet of the auto-calibration unit)

Air Consumption : Approx. 1.5 l/min

Weight : Approx. 2kg

Note : Used instrument air for span calibration

gas, if without instrument air is used,

contact YOKOGAWA.

* Recommended calibration gas = 1% Oxygen in Nitrogen FlowRate 0.6 ±0.1 Nl/min

Page 5: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

5

GS 11M12A01-E-H

*11. ZR40H Auto-calibration Unit (for Separatetype)

This unit is used when the instrument air is provided and theauto-calibration unit is attached. Solenoid valves are provided as standard equipment.

Reference air pressure: Clean air supply of measured gas

pressure+approx. 150 kPa G (pressure atinlet of the auto-calibration unit)

Air consumption : Approx. 1.5 l/minWeight : Approx. 3.5 kg

*12. ZR20H Auto-calibration Unit (for Integratedtype

Used when automatic calibration is specified for the integratedtype and instrument air is provided.Equipped with the analyzer when automatic calibration isspecified in the suffix code of the ZR202G Integrated type byselecting either “-A (Horizontal mounting)” or “-B (Verticalmounting)”. The ZR20H should be arranged when auto-calibration is to be required after the ZR202H has been installed.Ask Yokogawa service station for its mounting.

Construction : Dust-proof and rainproof construction:NEMA4X/IP67 (excluding flowmeter)

Mounting : Mounted on ZR202G, no vibrationMaterials : Body: Aluminum alloy, Piping: SUS316,

SUS304, Flowmeter: MA (Methacrylateresin)

Finish : Polyurethane corrosion-resistance coating,Case: Frosty white CC21, Cover: Deep-seamoss green CC32

Piping Connection : Refer to Model and Suffix CodesPower Supply : 24V DC (from ZR202G), Power

consumption: 1.3 WReference Air Pressure

: Sample gas pressure + Approx. 150 kPa(690 kPa max.), (Pressure at inlet of auto-calibration unit)

Air Consumption : Approx. 1.5 l/minWeight : Approx. 2 kgAmbient Temperature

: -20 to +55°C, no condensing and freezingAmbient Humidity : 0 to 95%RHStorage Temperature: -30 to +65°C

13. Dust Protector ZH21BThis protector is designed to protect the detector Tip fromingress of dust (i.e., to prevent combustible materials fromentering the detector cell) where humidity measurements aremade under dusty environements.

Insertion length : 0.428 mFlange : JIS 5 K 80 A FF equivalent or ANSI 4 B

150 LB FF equivalent (however, thickness isdifferent)

Materials of Flange : SUS 316, SUS 304, Stainless SteelWeight : approximately 6/8.5 kg (JIS/ANSI)Mounting : mounted on the detector, and process

flange with bolts and associated nuts andwashers

* Recommended calibration gas = 1% Oxygen in Nitrogen FlowRate 0.6 ±0.1 Nl/min

14. ZR20-CAL – Standard Local Calibration unitEasy to use and lightweight portable unit for calibration gassupply consisting of span gas (air) and zero gas.Included in this set:

- 2 span gas alucan light weight bottles (one as a spare)- 2 zero gas alucan light weight bottles (one as a spare)- 2 constant flow regulators with quick connectors, set to 0.5 L/min- 1 m tubing with quick connectors and nipple for direct connection

to the sensor- Holder for gas bottles- Carrying case

Capacity 0.5 literFilled pressure 120 bar (60 liter gas)Composition Span gas: 21% O2 (air), zero gas: 1%

Oxygen in nitrogenWeight approx. 8 kgMaterial Holder PVCMaterial Bottles Aluminium

Page 6: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

6

GS 11M12A01-E-H

Configurations : All configurations are possible, but insertion length of 0.4m is most common

Options : ZH21B Dust protector for high dust environment. Insertion length 0.416 m flange JIS 5k-80 or ANSI 150# 4",material SUS 316, Carborundum (SiC), SUS 304 (flange), weight 6 kg

Standard specifications for use as HumidityAnalyzer

Description : The Zirconia Oxygen analyzer can also beused to monitor the absolute humidity in anenvironment where standard air is mixedwith water vapor. Standard air consists of79% Nitrogen and 21% oxygen. If the air ismixed with Water vapor, the vol% ofoxygen decreases as the vol% of Waterincreases. The Zirconia based Oxygenanalyzer is therefore very useful as ananalyzer for moisture content of the samplegas. Both the ZR402 and the ZR202 canbe software switched from Oxygen analysisto Humidity analysis.

Sample composition : The sample gas may only contain a mixtureof Oxygen, Nitrogen and Water.

Performance : The performance specifications aredetermined by the primary measurement:the measurement of Oxygen. The possibleerrors in humidity units can be calculatedfrom the possible errors in the Oxygenanalysis.

Range : 1. Percent by volume: range 0- 100%,minimum span 0- 25 vol%2. Weight Ratio: kg H2O/ kg dry air: range1- 1.00, min span 0- 0.200 kg/kg dry air3. Dew point: range 0- 100 ºC4. Relative Humidity: range 0- 100%

Calculation

air

0% 100%water

100 %

O2

H2O

N2

0%100%

0%

x

a

Calculation Dew Point, calculation rH

Page 7: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

7

GS 11M12A01-E-H

ZR402G

Converterthread

-M M20x1.5 mm

Separate type Zirconia Oxygen/ High temperature Humidity Analyzer, Converter

Model Suffix code Optioncode Description

Tag plates *1 For humidity measurements, be sure to specify /HS options.*2 Specify either /SCT or /PT option code.*3 Will b l bl i th f t

-E-G

Display

-A

-E

/HS

/H

Instruction manual

Options

/SCT/PT

Set for Humidity Analyzer (*1)

Hood (*3)

English

EnglishGerman

Always -A

Stainless steel tag plate (*2)Printed tag plate (*2)

MODEL AND SUFFIX CODES

1. Separate type Zirconia Oxygen / Hightemperature Humidity Analyzer, Converter

ZR202G

ANSI CLASS150-4-RF (Equivalent)DIN PN10-DN50-A (Equivalent)

-C-E

-040-070-100-150-200

-S-C

0.4 m0.7 m1.0 m1.5 m2.0 m

-C-E

Rc 1/4-R

M20x1.5 mm-M

-A

-E

/C

Set for Humidity Analyzer (*4)/HS

/CV/SV

Printed tag (*7)/PT

Dust Filter (*6)/F1

Auto Calibration No auto calibration unit mountedAuto calibration unit mountedVertical mounting

-N-A-B

integrated type Zirconia Oxygen/ High temperature Humidity Analyzer

Model Suffix code Optioncode Description

Length

Wetted material SUS316Stainless steel with Inconel calibration gas tube (*7)

Flange(*2)

Reference air Natural convectionPressure compensation

Gas Thread

Connection box thread

Instruction manual English

Always -A

Options

Inconel bolt (*3)

Check valve (*5)Stop valve (*5)

Durethane coating/D

*1 For the horizontal installed probe whose insertion length is 2.5 meters or more, use the Probe Protector. Be sure to specify ZO21R-L--. Specify the flange suffix code either -C or -K.*2 The thickness of the flange depends on its dimensions.*3 Inconel probe bolts are used. Use this option for high temperature use (ranging from 600 to 700 C). *4 For humidity measurements, be sure to specify /HS options. Pressure compemsation of reference air can not be selected.*5 Specify either /CV or /SV option code.*6 Not used with the high-temperature humidity analyzer.*7 Recommended if measured gas contains corrosive gas like chlorine.

Stainless steel tag plate (*7)/SCT

2. Integrated type Zirconia Oxygen / High temperature Humidity Analyzer, Converter

Page 8: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

8

GS 11M12A01-E-H

3. Separate type Zirconia Oxygen / High Temperature Humidity Analyzer, Detectors

Model

ZR22G

Flange(*3)

ANSI CLASS150-4-RF (Equivalent)DIN PN10-DN50-A (Equivalent)JIS 5K-32-FF (Equivalent) (for high temperature use) (*4)

-C-E-Q

Separate type Zirconia Oxygen/ High temperature Humidity Analyzer, Detector

Length

Wetted material

-015-040-070-100-150-200

-S-C

0.15 m (for high temperature use) (*1)0.4 m0.7 m1.0 m1.5 m2.0 m

SUS316Stainless steel with Inconel calibration gas tube (*10)

Suffix code Optioncode Description

Reference air Natural convectionPressure compensation

-C-P

Gas Thread Rc 1/4-R

Connection box thread

Instruction manual

M20 x1.5 mm-M

Always -A-A

English-E

Options

Valves

Filter

Tag plates

*1 Used with the ZO21P High temperature Probe Adapter. Select flange (-Q).*2 When installing horizontally the probe whose insertion length is 2.5 meters or more, use the Probe Protector. Be sure to specify ZO21R-L--. Specify the flange suffix code either -C or -K.*3 The thickness of the flange depends on its dimensions.*4 Not used in conjunction with —P (pressure compensation) for reference air.*5 Inconel probe bolts are used. Use this option for high temperature use (ranging from 600 to 700 C). *6 Specify either /CV or /SV option code.*7 Not used with the high temperature humidity analyzer.*8 Specify either /SCT or /PT option code.*9 No water proof,avoid rain. Operating maximun temperature is 80C.*10 Recommended if measured gas contains corrosive gas like chlorine.Note: ZR22G can not be connected to Model ZA8C, AV8C, HA400 converter.

Inconel bolt (*5)/C

Durethane coating/D

Check valve (*6)Stop valve (*6)

/CV/SV

Dust Filter (*7)/F1

Stainless steel tag plate (*8)Printed tag plate (*8)

/SCT/PT

Page 9: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

9

GS 11M12A01-E-H

Joint: RC 1/4, Material: SUS316

Joint: 1/4 NPT, Material: SUS316

DescriptionPart No.

L9852CB

G7016XH

Joint: RC 1/4, Material: SUS304

Joint: 1/4 NPT, Material: SUS304

Part No.

K9292DN

K9292DS

Description

ZR40H

-R

-M

Rc 1/4

Model Suffix code Option code Description

Automatic calibration unit for ZR402G

Gas piping connection

Wiring connection

- -A

20 mm (M20 x 1.5)

Always -A

ZR20H

-R

-E

Rc 1/4

Model Suffix code Option code Description

Automatic calibration unit for ZR202G*1

Gas piping connection

Reference air *2

- -A

Instrument air

Always -A

-P Pressure compensated

Mounting -A Horizontal mounting

-B Vertical mounting

*1 Ask Yokogawa service station for additional mounting of ZR20H to the preinstalled ZR202G.*2 Select the appropriate reference air of ZR20H according to the one of ZR202G.

11. Automatic Calibration Unit for Separatetype Analyzer (Needs instrument air.)

12. Automatic Calibration Unit for Integratedtype Analyzer (Needs instrument air.)

8. Stop Valve for Calibration-gas line

9. Check Valve for Calibration-gas line

1.0 m1.5 m

DIN PN10-DN50-A equivalentANSI Class 150-4-RF

ZO21P

SiCSUS 310S

-100-150

*A

-E-A

-H

-A-B

High Temperature Probe Adapter

Material

Insertionlength

Flange

Style AStyle code

Model Suffix code Optioncode Description

ZH21B

* The flange thickness varies. Specify the probe ZR22G-040.

DescriptionOption codeSuffix codeModel

Dust protector (0 to 600C)

0.409 mInsertion length -040

Flange

Style code *B

-A ANSI Class 150-4B FF equivalent *

Style B

ZA8F

-J-A

Model Suffix code Option code Description

Flow setting unit

JointWith 1/4" NPT adapter

Style B

Rc 1/4

Style code *B

Part No.

E7046AL

E7046BB

E7046AP

E7046AQ

SiC, insertion length 1.0 m

SiC, insertion length 1.5 m

SUS310S, insertion length 1.0 m

SUS310S, insertion length 1.5 m

Description

RC 1/4 6 / 4 TUBE joint: SUS304

Part No.

E7046EC

Description

ZO21R

DIN PN10-DN50-A equivalent

-L Probe Protector(0 to 700 8C)

-100-150-200

1.05 m (3.5ft)1.55 m (5.1ft)2.05 m (6.8ft)

Insertionlength

Flange ( *1)

*1 Thickness of flange depends on dimensions of flange.

ANSI Class 150-4-RF equivalent

Model Suffix code Optioncode Description

*B Style BStyle code

-E-A

Part No.

K9471UA

Description

Filter

4. Adapter for High temperature Probe ofseparate type Oxygen Analyzer

Note: For this high-temperature use probe adapter, be sure tospecify the ZR22G probe of it’s insertion length 0.15meters.

High temperature Probes (Spare Parts)

5. Auxiliary Ejector for High-temperature Useof separate type Oxygen Analyzer

7. Filter for Zirconia Oxygen Analyzers

13. Dust Protector for High-temperatureHumidity Analyzers

10. Flow setting unit for manual calibration(Needs instrument air.)

6. Probe Protector for Zirconia Oxygen Analyzer

Model DescriptionZR20-cal Standard Calibration Unit

Spare parts DescriptionZR20-zero Spare Zero Gas bottleZR20-span Spare Span Gas bottle

14. ZR20-CAL – Standard Local Calibration unit

Page 10: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

10

GS 11M12A01-E-H

1 to 6 (Panel Thickness)

EXA ZR402G 4 - 6 holesfor Wall mounting

4-R8 to R10or4-C5 to C8

4 - 6 holes

8-G1/2, 8-1/2NPT etc(Wiring connection)

( for wall mounting)

F08_01.EPS

2-inch mounting pipe

Wall mounting Panel Cut-out

100

108228

280

126.

5

57.3

136.3 54.7

111

36

120.

2

(1/2

NP

T)

190 183+20

274

+2

0

126.

5

1. Model ZR402G Separate type Zirconia Oxygen/ High temperature Humidity Analyzer, Converter

Page 11: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

11

GS 11M12A01-E-H

153 to 164

125

L

170252 to 265 49

48.5

338 to 351

25

t

122

L= 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 (m)

4-G1/2,2-1/2NPT etc.Cable connection port

Display side

1563

2564 49

3424

25

t

PIPING:A

PIPING:B

L

L= 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 (m)

4-G1/2,2-1/2NPT etc.Cable connection port

50

.8

Rc1/4 or 1/4NPTReference air inlet

Rc1/4 or 1/4NPTCalibration gas inlet

Reference air outlet

Stopvalve

PIPINGABBABBAABBBBA

ANSI CLASS150-2-RF equivalentANSI CLASS150-3-RF equivalentANSI CLASS150-4-RF equivalentDIN PN10-DN50-A equivalentDIN PN10-DN80-A equivalentDIN PN10-DN100-A equivalentJIS 5K-65-FFJIS 10K-65-FFJIS 10K-80-FFJIS 10K-100-FFJPI CLASS150-4-RF equivalentJPI CLASS150-3-RF equivalentWestinghouse

152.4190.5228.6165200220155175185210229190155

120.6152.4190.5125160180130140150175

190.5152.4127

4 - 194 - 198 - 194 - 188 - 188 - 184 - 154 - 198 - 198 - 198 - 194 - 19

4 - 11.5

A B C19242418202014181818242414

tFlange

5

0.8

Rc1/4 or 1/4NPTReference air inlet

Rc1/4 or 1/4NPTCalibration gas inlet

123

Terminal side

C

Flange

Flange

B

C

Flange

A

A

B

2. Model ZR202G Integrated type Zirconia Oxygen/ High temperature Humidity Analyzers

Model ZR202G...-P (with pressure compensation) Integrated type Zirconia Oxygen / High temperatureHumidity Analyzers

Page 12: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

12

GS 11M12A01-E-H

25

t285 4

50

.8

12

7

85

L

48

1563 69

L=0.15, 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 3.6, 4.2, 4.8, 5.4 (m)

C

Reference gas outlet

25 48

153 to 164 69

Rc1/4 or 1/4NPTReference air inlet

2-G1/2,2-1/2NPT etc.Cable connection port

Rc1/4 or 1/4NPTCalibration gas inletFlange

PIPING : A

PIPING:B

Rc1/4 or 1/4NPTReference air inlet

2-G1/2, 2-1/2NPT etc.Cable connection port

B

A

C

Flange

Flange

B

A

Rc1/4 or 1/4NPTCalibration gas inlet

Flange

Stop Valve

EXTERNAL DIMENSIONS

3. Model ZR22G Separate type Zirconia Oxygen / High temperature Humidity Analyzer, Detectors

Model ZR22G...-P (with pressure compensation) Separate type Zirconia Oxygen / High temperatureHumidity Analyzer, Detectors

PIPINGANSI CLASS150-2-RF equivalentANSI CLASS150-3-RF equivalentANSI CLASS150-4-RF equivalentDIN PN10-DN50-A equivalentDIN PN10-DN80-A equivalentDIN PN10-DN100-A equivalentJIS 5K-65-FFJIS 10K-65-FFJIS 10K-80-FFJIS 10K-100-FFJPI CLASS150-4-RF equivalentJPI CLASS150-3-RF equivalentWestinghouse

152.4190.5228.6165200220155175185210229190155

120.6152.4190.5125160180130140150175

190.5152.4127

4 - 194 - 198 - 194 - 188 - 188 - 184 - 154 - 198 - 198 - 198 - 194 - 19

4 - 11.5

A B C19242418202014181818242414

ABBABBAABBBBA

tFlange

(m)

Page 13: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

GS 11M12A01-E-H

13

R1/2(Note2)

Approx. 351

180 Flange (Thickness)JIS 5K 32A FF equivalent

Gasket (Thickness 3)

Detector(ZR22G)

Flange <1>

52 over

30

60.5

Reference air inlet <2>

Calibration gas inlet<3>

High temperatureProbe SiC pipe(GK material)

110

115

4225

170

App

rox,

215

(Ins

etio

n le

ngth

) (N

ote1

)

App

rox,

100

69

Approx. 48

C

(Note 1) 1=1000 or 1500 (mm)(Note 2) Sample gas outlet

(if the sample gas pressure is negative, connect the auxiliary ejector.)

127

85

Measurement gas outlet

Flange providedby customer

Unit: mm

60.5A

B

A

Pipe hole (2- G1/2, 2-1/2NPT, etc)

Needle valve

<1>

<2>

Pressure gauge

Nozzle (Note1)

R1/2Ejector assembly Detector

<1> Rc1/4 or 1/4 NPT<2> 6/ 4 or 1/4 inch copper tube (stainless) with ejector to connect<3> Rc1/4

4020Tee<1>

Blow<1>

43

App

rox.

67

App

rox.

88

39

App

rox.

70

38

Ful

l ope

n he

ight

(Note1) The connector of ejector assembly is a dedicated connector with nozzle function.

Measurement gas inlet

4. Model ZO21P Adapter for High temperature Probe of separate type Oxygen Analyzer

5. Model E7046EC, E7046EN Auxiliary Ejector for High temperature Use of separate type OxygenAnalyzer

<1> <2> <3>Rc1/4 or ø6/ø4 or 1/4” Rc1/41/4 NPT copper tube

(stainless) withejector to connect

Page 14: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

14

GS 11M12A01-E-H

51

32

10

Caborundum filter

Increase of insertion length

Flange<1>

JIS 5K-65-FFequivalent

155

228.6

130

190.5

4- 15

8- 19

5

12

A B C t

ANSI CLASS150-4-FF equivalent

40

50

D

D

C

B

B

Flange <1>(with bolts, nuts, gasket and washer)

gasket (t1.5)

SUS316

Dimensions ofholes on opposing surface

A

Gas flow

Washer (12)Mounting nut (M12)

l (Insert length)

l=1050,1550,2050

6. Model ZO21R Probe Protector for Zirconia Oxygen Analyzers

7. Model K9471UA Filter for Oxygen Analyzer

K9292DN : Rc 1/4(A),R 1/4(B)K9292DS : 1/4NPT(A),1/4NPT(M)(B)

Approx. 19 Approx. 54

55(F

ull o

pen

leng

th)

40

Rc1/4 or 1/4NPT

43

8. L9852CB/ K9471UN Stop Valve forCalibration-gas line

9. K9292DN/ K9292DS Check Valve forCalibration-gas line

43

55

Page 15: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

15

GS 11M12A01-E-H

6 hole

Reference air outlet Zero gas outlet Zero gas inlet Instrument air inlet

4-Rc1/4

20 2040 40

Piping connectionport

180

235.

870

35

268

222.

87

140

REFERENCE CHECK

REFERENCE SPAN ZERO

F20.EPS

200

10. Model ZA8F Flow setting unit for manual calibration

35 3535 35

Spangas inlet

Page 16: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

16

GS 11M12A01-E-H

Unit: mm

OCK

140

250

42

26

41.2 41.2

54 71.526

49.5

1222

3

4616

3040 47.5 25

102

MA

X58

90 116.5

6.5

*1 with four ISO M6 screws can wall-mount

*1 4-

(wiring inlet is at same position on rear)

Flowmeter

Setting Valve forcalibration gas

Setting Valve forreference air

wiring inlet ; 2-G1/2,Pg13.5,M2031.5 or 1/2NPT(Female)

calibration gas outlet Rc1/4 or 1/4 NPT(Female)

reference air inlet

Rc1/4 or 1/4 NPT(Female)zero gas inlet

Rc1/4 or 1/4 NPT(Female)Rc1/4 or 1/4 NPT(Female)reference air outlet

2B pipe mounting example

11. Model ZR40H Automatic Calibration Unit for Separate type Analyzer

ZR402G Converter ZR40H Automatic Calibration unit

AC-Z

AC-C

AC-S

Zero

Span

EV1

EV2

CHECKOUT

ZERO GAS IN AIR IN

Instrument air Approx.1.5 l/min.

*2 Needle valve is supplied as accessory with flow meter

flow meter

flow meter

REFOUT

*2 *2solenoid valveEV1,2

EV1 EV2

Page 17: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

17

GS 11M12A01-E-H

zero gas inlet

Rc1/4 or 1/4NPT(Female)

reference air inlet

Rc1/4 or 1/4NPT(Female)

terminal box side

span gas inlet

Rc1/4 or 1/4NPT(Female)

zero gas inlet

Rc1/4 or 1/4NPT(Female)

Span gas inlet

Rc1/4 or 1/4NPT(Female)

reference air inlet

Rc1/4 or 1/4NPT(Female)

166.5

166.5

258

60

244

214 MAX44

4516

0

66.540 40

66.540 40

180 MAX44

Vertical mounting on the ZR202G (-B)

Horizontal mounting on the ZR202G (-A)Unit: mm

display side

12. Model ZR20H Automatic Calibration Unit for Integrated type Analyzer

t

C

C

B

B

A

B

76.3

72

A

180

228.5

B

145

190.5

C

4-19

8-19

t

12

12

D

40

50

Unit: mm416 (Insertion length)

Install facing upwards.Flange

JIS 5 K FF equivalent

ANSI 4 B 150 LB FF

13. Model ZH21B Dust Protector for High temperature Humidity Analyzers Unit: mm

JIS 5K-80-A-FF equivalent

428

D

Page 18: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

18

GS 11M12A01-E-H

1FG

2AO1(+)

3AO1(-)

4AO2(+)

5AO2(-)

6CJ(+)

7CJ(-)

8TC(+)

9TC(-)

10CELL

(+)

11CELL

(-)12FG

13DI-1

14DI-2

15DI-C

16AI(+)

17AI(-)

18AC-

ZERO

19AC-

SPAN

20AC-

COM

21FG

22FG

CELL(+)

CELL(-)

TC(+)

TC(-)

CJ(+)

CJ(-)

23DO-1

24DO-1

25DO-2

26DO-2

27DO-3

28DO-3

29DO-4

30DO-4

31HTR

32HTR

33L

34N

35G

36FG HTR HTR

100-240V AC50/60 Hz

Model ZR402G Separate type Zirconia Oxygen Analyzer/ Converter

Model ZR22G Separate type Zirconia OxygenAnalyzer, Detector

Analog output 2

Analog output 1

4-20 mA DC4-20 mA DCDigital output

Contact input 2

Contact input 1

Contact output 1

Contact output 2

Contact output 3

Contact output 4

Solenoid valve for span gas for automatic calibration

Solenoid valve for zero gas for automatic calibration

Flow setting unit*1

*1

*1

1DI-1

2DI-2

3DI-C

4DO-1

5DO-1

6DO-2

7DO-2

8FG

9AO(+)

10AO(-)

11L

12N

13G

14FG

Model ZR202G Integrated type Zirconia Oxygen / High temperature Humidity Analyzer

Contact input 1Contact output 2Contact output 1

Analog output 4-20 mA DC

100 to 240 V AC, 50 or 60 Hz

Digital output

Contact input 2

*1

*1 Ground resistence is 100 ohm or less.

WIRING CONNECTIONS

Page 19: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

19

GS 11M12A01-E-H

Inquiry Sheet for Models ZR22G, ZR402G, and ZR202G Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers

vol%O2,

kg/kg,

C,

kPa,

m/sec,

g/Nm3,

ppm,

ppm,

ppm,

ppm,

Vol%H2O

Please place checkmarks in the appropriate boxes and fill in the necessary information in the blanks.

Customer

Destination of delivery

Plant name

Measurement points

2. Process conditions

2.1 Measurement gas components

2.2 Oxygen concentration

Moisture contents

2.3 Temperature

2.4 Pressure

2.5 Gas flow

2.6 Dust type, Size

2.7 Corrosive gas

2.8 Combustible gas

2.9 Others

3. Installation site conditions

3.1 Ambient temperature

3.2 Vibration

3.3 1 Probe installation location

3.4 Instrument air supply

3.5 Converter location

3.6 Cable length between probe and converter

3.7 Calibration method

4. Quotation data

Probe Model ZR22G General-use Probe Refer to the Probe Configuration for probe selection.Model ZO21P-H High temperature Use Probe Adapter

Model ZR402G Separate type Analyzer, Converter

Model E7046EC/E7046EN Auxiliary Ejector for high temperature use.

Model ZH21B Dust Protector for high temperature humidity analyzer.

Model ZO21R Probe Protector for Oxgen Analyzer

1. General information

Nor. Min. Max.

Nor.

Nor.

Nor.

Nor.

Nor.

Min.

Min.

Min.

Min.

Min.

Max.

Max.

Max.

Max.

No gas Gas , quantity

m quantity

No gas Gas , quantity

, quantity

, quantity

1. Around Probe temp. from to C, 2. Around Converter temp. from to C

No vibration Vibration

Furnace Stack Others

2 Probe position Horizontal Vertical Others

Indoor Outdoor Covered

3 Probe insertion length (m) 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0, Others

4 Flange JIS ANSI Others

Cannot be used. Can be used. kPa

Indoor Outdoor Covered (under roof)

meters

Manual Automatic

Quotation Qantity Description

Options (for general use)

Model K9471UA Filter

Model ZR202G Integrated type Zirconia Oxygen / High temperature Humidity Analyzer

Model ZO21S Standard Gas Unit Select any one of Model ZO21S, ZA8F, ZR40H .

Model ZA8F Flow Setting Unit

Model ZR40H Automatic Calibration Unit Not required if probe options are specified.

L9852CB/K9471UN Stop Valve

K9292DN/K9292DS Check Valve

G7011XF/E7040EL Air Set

G7001ZC Zero Gas Cylinder

G7013XF/G7014XF Pressure Regulator

Type of analyzer : Oxygen Analyzer High temperature Humidity Analyzer

Sparate type Integrated type

Object : indication record control alarm

Fuel : gas oil coal

Power requirements V AC Hz

Page 20: General Specifications Direct In Situ Zirconia Oxygen ... · Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers ... 1 +3 kg (flange DIN ... Analyzers

GS 11M12A01-E-HSubject to change without notice Printed in The Netherlands, 04-401 (A) QCopyright©

YOKOGAWA EUROPE B.V.Databankweg 203821 AL AMERSFOORTThe NetherlandsTel. +31-33-4641 611Fax +31-33-4641 610E-mail: [email protected]

YOKOGAWA CORPORATION OF AMERICA2 Dart RoadNewnan GA 30265United StatesTel. (1)-770-253-7000Fax (1)-770-251-2088E-mail: [email protected]

YOKOGAWA ELECTRIC ASIA Pte. Ltd.5 Bedok South RoadSingapore 469270SingaporeTel. (65)-241-9933Fax (65)-241-2606E-mail: [email protected]

YOKOGAWA HEADQUARTERS9-32, Nakacho 2-chome,MusashinoshiTokyo 180JapanTel. (81)-422-52-5535Fax (81)-422-55-1202E-mail: [email protected]

Yokogawa has an extensive sales anddistribution network. Please refer to the European web-site(www.yokogawa-europe.com) to contactyour nearest representative.