USER MANUAL
AVL DITEST MCS 120
MODULE CONDITIONING SYSTEM
Ident number: AT7953E
Revision: 02
Issue: 2020-04
Software version:
Data may change without notice. All data valid at time of printing.
PASSION INNOVATES FUTURE
AVL DiTEST GmbH
Alte Poststrasse 156
8020 Graz
AUSTRIA
Tel: +43 316 787-0
Fax: +43 316 787-1460
www.avlditest.com
Copyright © 2020 AVL DiTEST GMBH, all rights reserved.
The content of this publication may not be reproduced in any way or forwarded to third parties, either in part or in full, without the prior written consent of AVL DiTEST. This publication was created with due care such that AVL DiTEST is not liable for any remaining errors or omissions or for any damages arising therefrom.
AVL DiTEST MCS 120 Warning and Safety Notes
User Manual I
Warning and Safety Notes
This user manual contains important warning and safety notices that must be observed by the user. The product is intended only for the highly specific use described in the user manual. The most important prerequisites and safety measures for the use and operation of the product are also described to ensure faultless operation. No warranty can be given and no liability is assumed for applications beyond the described use, irrespective of observance of the necessary prerequisites and safety measures. The product may only be used and operated by personnel who, based on their qualifications, are capable of adhering to the necessary safety measures during use and operation. Only accessories and consumables supplied by AVL DiTEST or approved by AVL DiTEST may be used. The measurement results obtained from the product in question depend not only on correct functioning of the product, but also on a series of general conditions. The results delivered by the product must therefore be evaluated by a specialist (e.g. plausibility check) before further measures are taken on the basis of a delivered measurement. Settings and maintenance work on open devices while still live may only be performed by trained specialists who are aware of the associated danger. The product may only be repaired in the factory of origin or by specialists specifically trained to perform such repair. When using the product, it must be ensured by a specialist that the test object or test system is not brought into any operational state that could result in damage to goods or endangerment of people.
Warning and Safety Notes AVL DiTEST MCS 120
II User Manual
AVL DiTEST MCS 120 Special Safety
User Manual III
General Safety
Danger to life by electric potential on vehicles with high voltage systems
Deadly high voltages are present on the HV energy store (HV battery) and on parts connected to it!
Make sure no-one can come into contact with the connections on the HV battery, connecting cables of the HV battery or other parts under high voltage!
The battery to be tested should be removed from the vehicle. If the measurements are made on an open but still installed in the vehicle battery, the vehicle must be disabled and secured.
Danger to life by electric potential on vehicles with high voltage systems
Please note that work, in particular dismantling, on high-voltage vehicles may only be carried out by suitably trained and authorized personnel.
Danger to life by electric potential on vehicles with high voltage systems
For work under tension special safety and training conditions apply!
In every situation make sure that these are observed!
Danger to life by electric potential on vehicles with high voltage systems
Make sure that all necessary safety precautions have been taken before you start measuring.
Check the HV insulation gloves for suitability before using them!
WARNING
When connecting the cell module, make sure the IR temperature sensor is not covered and ensure direct visibility between the sensor and the module.
DANGER
DANGER
DANGER
DANGER
Special Safety AVL DiTEST MCS 120
IV User Manual
AVL DiTEST MCS 120 IMPORTANT SAFETY INSTRUCTIONS
User Manual V
IMPORTANT SAFETY INSTRUCTIONS
1. Read all instructions.
2. Care must be taken as burns can occur from touching hot parts.
3. Do not operate equipment with a damaged cord or if the equipment has been dropped or damaged - until it has been examined by a qualified serviceman.
4. Do not let cord hang over edge of table, bench or counter or come in contact with hot manifolds or moving fan blades.
5. If an extension cord is necessary, a cord with a current rating equal to or more than that of the equipment should be used. Cords rated for less current than the equipment may overheat. Care should be taken to arrange the cord so that it will not be tripped over or pulled.
6. Always unplug equipment from electrical outlet when not in use. Never use the cord to pull the plug from the outlet. Grasp plug and pull to disconnect.
7. Let equipment cool completely before putting away. Loop cord loosely around equipment when storing.
8. To reduce the risk of fire, do not operate equipment in the vicinity of open containers of flammable liquids (gasoline).
9. Adequate ventilation should be provided when working on operating internal combustion engines.
10. Keep hair, loose clothing, fingers, and all parts of body away from moving parts.
11. To reduce the risk of electric shock, do not use on wet surfaces or expose to rain.
12. Use only as described in the manual. Use only manufacturer´s recommended accessories.
13. ALWAYS WEAR SAFETY GLASSES. Everyday eyeglasses only have impact resistant lenses, they are NOT safety glasses.
SAVE THESE INSTRUCTIONS
IMPORTANT SAFETY INSTRUCTIONS AVL DiTEST MCS 120
VI User Manual
AVL DiTEST MCS 120 Content
User Manual VII
Content
Warning and Safety Notes ........................................................................... I
General Safety ............................................................................................ III
IMPORTANT SAFETY INSTRUCTIONS ....................................................... V
1 General .............................................................................................. 1 1.1 General Notes ..................................................................................................................... 1 1.2 EC Declaration of Conformity/CE mark .............................................................................. 2 1.3 Certificate of Calibration...................................................................................................... 4 1.4 Designated Use .................................................................................................................. 6 1.5 Field of application .............................................................................................................. 6
2 Components ...................................................................................... 7 2.1 Cable sets ......................................................................................................................... 16 2.1.1 USB-cable (USB 2.0 A auf USB 2.0 B Rugged) EX7069 ................................................. 16 2.1.2 Cable Set plus minus, X590 ............................................................................................. 16 2.1.3 Cable, analog module, X590 ............................................................................................ 16 2.1.4 Cable, CAN bus module ................................................................................................... 17 2.1.5 SPI-UART-Adapter ........................................................................................................... 18
3 Commissioning ............................................................................... 19 3.1 Installation of the DSS Software ....................................................................................... 19 3.2 Firmware Update .............................................................................................................. 19 3.3 Hardware .......................................................................................................................... 20
4 Performing a Measurement ............................................................ 23 4.1 Start MCS 120 .................................................................................................................. 23 4.1.1 Safety test ......................................................................................................................... 32 4.1.2 Self-test ............................................................................................................................. 35 4.1.3 Deviceinfo ......................................................................................................................... 37 4.1.4 Viewing all reports ............................................................................................................. 39 4.1.5 Result report directory....................................................................................................... 41 4.1.6 Entering a list of operator names ...................................................................................... 42 4.1.7 Switch off .......................................................................................................................... 42
5 Troubleshooting .............................................................................. 43
6 Maintenance and care..................................................................... 45 6.1 Optical check .................................................................................................................... 45 6.2 Cleaning ............................................................................................................................ 45
7 Scope of delivery ............................................................................ 47
8 Technical data ................................................................................. 49 8.1 Operating data .................................................................................................................. 49
9 Index ............................................................................................... 51
Content AVL DiTEST MCS 120
VIII User Manual
AVL DiTEST MCS 120 General
User Manual 1
1 General
1.1 General Notes
The Module Conditioning System MCS 120, here after named MCS 120, is able to charge and discharge single cell modules of high voltage batteries from E-cars and hybrid cars. To guarantee an overall equal load of all cell modules on a HV-battery, the new cell module has to be charged / discharged to meet this load of the whole battery.
The conditioned „balanced“ cell module replaces the defect cell module.
MCS 120 is technically identical to MCS (Module Conditioning System). Differences exist only in the appearance of the imprints (front plate and type plate) and in the design of the user interface.
The MCS 120 is designed to use in car shops which have to do HV battery repairs. To use this device you must have a special qualification. See below.
Please follow the regulation valid to your country!
Read this operating manual and follow especially the safety hints!
General AVL DiTEST MCS 120
2 User Manual
1.2 EC Declaration of Conformity/CE mark
The manufacturer hereby declares (fig 1-1) that the MCS is in conformity with the regulations of the following EC directives, including all the relevant modifications
2014/30/EU Council Directive on the approximation of the laws of the Member States relating to electromagnetic compatibility
2014/35/EU Council Directive on the approximation of the laws of the Member States related to electrical equipment designed for use within certain voltage limits
2011/65/EU Directive on the restriction of the use of certain hazardous Substances in electrical and electronic equipment
The following standard conformities are available at the MCS:
SAFETY: EN 61010-1 Safety requirements for electrical equipment IEC 61010-1 for measurement, control, and laboratory use – UL 61010-1 Part 1: General requirements CAN/CSA-C22.2 No. 61010-1-12
EN 61010-2-30 Safety requirements for electrical equipment IEC 61010-2-30 for measurement, control, and laboratory use. UL 61010-2-30 Particular requirements for testing and measuring circuits CAN/CSA-C22.2 No. 61010-2-30
UL 201 Standard for Garage Equipment UL 60950-1 Information Technology Equipment
EMV / EMC: EN 61326-1 Electrical equipment for measurement, control and IEC 61326-1 laboratory use – EMC requirements – Part 1: General
requirements
FCC Part-15 Subpart-B Part 15—Radio Frequency Devices Subpart B— Unintentional Radiators
RoHS EN 50581 Technical documentation for the assessment of electrical and electronic products with respect to the restriction of hazardous substances
AVL DiTEST MCS 120 General
User Manual 3
EC Declaration of Conformity/CE mark
Fig. 1-1 EC Declaration of Conformity/CE mark
General AVL DiTEST MCS 120
4 User Manual
1.3 Certificate of Calibration
The manufacturer hereby declares (fig 1-2) that the device delivered with this operating manual does not require any calibration within the first 24 month after its delivery. Subsequent calibrations should be carried out every 12 months.
AVL DiTEST MCS 120 General
User Manual 5
Module Conditioning System MCS 120 Bestellnummer: BO7971
Das Module Conditioning System MCS 120 wurde unter Einhaltung aller Vorgaben nach der jeweils gültigen Prüfvorschrift erfolgreich getestet und hat in einwandfreiem Zustand unser Haus verlassen.
In den ersten 24 Monaten nach der Auslieferung des Geräts ist keine
Kalibrierung erforderlich.
The Module Conditioning System MCS 120 was tested successfully
with compliance to all specified values and under the actual test
procedure and left our facilities in perfect condition.
During the first 24 month after delivery of the device, calibration is not
required.
AVL DiTEST ist nach ISO 9001 zertifiziert!
AVL DiTEST is accredited according to ISO 9001!
AVL DiTEST GmbH - Alte Poststrasse 156 - A-8020 Graz
Fig. 1-2 Certificate of initial calibration
General AVL DiTEST MCS 120
6 User Manual
1.4 Designated Use
It’s only allowed to use the MCS 120 in that way, described in this manual.
WARNING
The Housing should only be opened by AVL DiTEST Service.
The product described has been developed, manufactured and checked according to the relevant safety standards. If the safety instructions are observed, the start-up is carried out as stipulated, the device is used for the intended purpose and the recommended maintenance and care is also observed, then in normal cases there is no danger regarding damages to property or for the health of persons associated with the MCS 120.
1.5 Field of application
The MCS 120 is designed for charging / discharging of single cell modules of a HV battery to a certain load level.
AVL DiTEST MCS 120 Components
User Manual 7
2 Components
Transport box closed
Fig. 2-1 Transport box closed
(1) Latch
(2) Carrying handle
(3) Latch
(4) Latch
(5) Carrying handle
(6) Latch
1 2 3
6 5 4
Components AVL DiTEST MCS 120
8 User Manual
Transport box opened, with MCS 120
Fig. 2-2 Transport box opened
(1) MCS 120
(2) Accessories
(3) Transport box, bottom part
1 2
3
AVL DiTEST MCS 120 Components
User Manual 9
Transport box opened, with MCS 120
Fig. 2-3 Transport box opened, with MCS 120
(1) MCS 120
(2) Transport box, bottom part
1
2
Components AVL DiTEST MCS 120
10 User Manual
Transport box opened, MCS 120, cell module
Fig. 2-4 Adapter und transport box open (top view).
(1) Cell module
(2) MCS
(3) Transport box, (bottom part)
1 2
3
3
AVL DiTEST MCS 120 Components
User Manual 11
Transport box opened, MCS 120, cell module connected
Fig. 2-5 Transport box opened, MCS 120, cell module connected
(1) Cable “minus” (black)
(2) Cable “plus” (red)
(3) MCS 120
(4) Transport box, bottom part
(5) Cable “CAN”
(6) Cell module
1 2 3
6 5 4
Components AVL DiTEST MCS 120
12 User Manual
MCS 120, rear view
Fig. 2-6 MCS 120, rear view
(1) Carrying handle
1
AVL DiTEST MCS 120 Components
User Manual 13
MCS 120, side view right
Fig. 2-7 MCS 120, side view right
(1) Air inlet grill
(2) USB-connector (rugged) for connection of the USB cable
(3) Power supply inlet. Connection for power cord
WARNING
Use the included mains cord set (power cord)! If the mains cord set has to be changed note the following requirements strictly!
On 240 V mains grid the mains cord set must be designed for a continuous current of 10 A, on 120V mains grid the mains cord set must be designed for a continuous current of 20 A. The mains cord set must have a C19 plug for the connection to the MCS suitable for the above current ratings. Connect MCS 120 only to protective contact sockets!
Set up the MCS 120 in a way that there is always free access to the power separator! (Power cord from the mains socket).
Version USA: Only oil-resistant cables may be used!
WARNING
Set up the MCS 120 so that adequate ventilation is guaranteed, fans and air vents are not obstructed and the air must circulate freely!
3 2 1
Components AVL DiTEST MCS 120
14 User Manual
MCS 120, side view left
Fig. 2-8 MCS 120, side view left
(1) Specification plate
(2) Air inlet grill
WARNING
Set up the MCS 120 so that adequate ventilation is guaranteed, fans and air vents are not obstructed and the air must circulate freely!
2 1
AVL DiTEST MCS 120 Components
User Manual 15
MCS 120, front view
Fig. 2-9 MCS, front view
(1) Connector „plus“ (red)
(2) Connector “analogue”
(3) LEDs
Green Charging / discharging is running
Blue Device on and ready
Blue Device is idle
Red Charging / discharging is running or an error has happened
(4) On / Off button
Green Power module ON (active mode)
Red Power module OFF (idle mode)
(Only the modules are switched on/off)
(5) Connector “CAN”
(6) IR temperature sensor
(7) Carrying handle
(8) Connector “minus” (black)
1 2 3 4 5 6 7 8
Components AVL DiTEST MCS 120
16 User Manual
2.1 Cable sets
2.1.1 USB-cable (USB 2.0 A auf USB 2.0 B Rugged) EX7069
This cable connects the MCS, socket USB, with the PC, socket USB.
Fig. 2-10 USB-Cable USB 2.0 A to USB 2.0 B
2.1.2 Cable Set plus minus, X590
Plus & minus cable with rotatably mounted M6 knurled screw.
Fig. 2-11 Cable Set
2.1.3 Cable, analog module, X590
Cable to connect the analog cell voltage module taps.
(without CSC (Cell-Supervision-Circuit).
Fig. 2-12 Cable, symbol photo
AVL DiTEST MCS 120 Components
User Manual 17
2.1.4 Cable, CAN bus module
Cable to connect the digital cell voltage module via CAN bus.
(with CSC (Cell-Supervision-Circuit).
Fig. 2-13 Cable, symbol photo
WARNING
Use only the referenced accessories from AVL DiTEST!
WARNING
Use the included mains cord set (power cord)!
If the mains cord set has to be changed note the following requirements strictly!
On 240 V mains grid the mains cord set must be designed for a continuous current of 10 A, on 120 V mains grid the mains cord set must be designed for a continuous current of 20 A.
The mains cord set must have a C19 plug for the connection to the MCS 120 suitable for the above current ratings.
Connect MCS 120 only to protective contact sockets!
Set up the MCS 120 in a way that there is always free access to the power separator!
(Power cord from the mains socket).
Version USA: Only oil-resistant cables may be used!
Components AVL DiTEST MCS 120
18 User Manual
2.1.5 SPI-UART-Adapter
The cable is used to connect modules with TPL communication.
Fig. 2-14 SPI-UART-Adapter
AVL DiTEST MCS 120 Commissioning
User Manual 19
3 Commissioning
3.1 Installation of the DSS Software
CAUTION
Please install the DSS-Software before connecting the MCS 110 to the PC.
To install the DSS-Software you will need admin rights!
To install the DSS Software note the document AT7260 “DSS installation and update”.
At the point “Select the application to be installed” select MCS 120 Module Conditioning System”.
3.2 Firmware Update
Note the documentation AT7260, "DSS Installation and Update".
Commissioning AVL DiTEST MCS 120
20 User Manual
3.3 Hardware
CAUTION
Temperature compensation
When delivering the MCS 120 in the winter, the inside of the device must be prevented from becoming moist. For this reason, we request that you allow the device to stand for approx. 3 hours at the location from which it is to be operated before connecting it or switching it on!
WARNING
Set up the MCS 120 so that adequate ventilation is guaranteed, fans and air vents are not obstructed and the air must circulate freely!
Set up the MCS 120 in a way that there is always free access to the power separator! (Power cord from the mains socket).
1. Open the latches (1), (2), (4) and (5) and remove the cover (3).
Fig. 3-1 MCS 120
1 2 3
5 4
AVL DiTEST MCS 120 Commissioning
User Manual 21
2. Take the basic device (2) from the transport / adapter case (1) and put it as shown right beside the case (3).
Fig. 3-2 MCS 120
3. Connect the MCS 120 to the tester via an USB-cable. Make sure that the screw cap on the USB Rugged plug (1) is tight.
Fig. 3-3 MCS 120
1 2
2
1
Commissioning AVL DiTEST MCS 120
22 User Manual
4. Connect the power cord to the mains connector (1) and the main plug into the wall outlet.
WARNING
Use the included mains cord set (power cord)!
If the mains cord set has to be changed note the following requirements strictly!
On 240 V mains grid the mains cord set must be designed for a continuous current of 10 A, on 120 V mains grid the mains cord set must be designed for a continuous current of 20 A.
The mains cord set must have a C19 plug for the connection to the MCS 120 suitable for the above current ratings.
Connect MCS 120 only to protective contact sockets!
Set up the MCS 120 in a way that there is always free access to the power separator! (Power cord from the mains socket).
Version USA: Only oil-resistant cables may be used!
Fig. 3-4 MCS 120
5. The MCS 120 is ready for operation.
1
AVL DiTEST MCS 120 Performing a Measurement
User Manual 23
4 Performing a Measurement
4.1 Start MCS 120
1. Start the DSS Manager by selecting: Start│Programs│DiTEST│DSS.
2. Select Diagnostic│Module Blancer.
Fig. 4-1 Screen DSS Manager
Performing a Measurement AVL DiTEST MCS 120
24 User Manual
3. The start screen appears. Click on Charge/Discharge.
Fig. 4-2 Startscreen
AVL DiTEST MCS 120 Performing a Measurement
User Manual 25
4. This is followed by a screen with safety instructions.
WARNING
Read and observe the safety instructions on the screen!
5. Pay close attention to the safety instructions and confirm with F8 Next.
Fig.4-3 Screen with safety instructions
Performing a Measurement AVL DiTEST MCS 120
26 User Manual
6. Choose a cell module type from the drop down list (2) and enter a serial number (3) Input the nominal cell voltage (4). Optional with barcode reading pen: Scan the bar code label attached to the cell module. Field „Scan-code“ (1) shows the scanned cell module information data. Module type (2) and the serial (3) are filled in automatically.
7. Confirm by clicking F8 Next.
Fig. 4-4 Screen „Select cell module“
AVL DiTEST MCS 120 Performing a Measurement
User Manual 27
8. The next screen guides you through the connection process. It is important to do all connecting work in the shown order.
WARNING
Execute pt. 2 and 3 in the shown order!
WARNING
When connecting the cell module, make sure the IR temperature
sensor is not covered and ensure direct visibility between the
sensor and the module.
9. Continue by clicking F8 Next.
Click on the button to get a picture schowing where to connect the cable.
Fig. 4-5 Screen „Connecting cell module“
WARNING
Caution, all contact surfaces must be clean.
After connecting, ensure that all cables are firmly fixed.
Performing a Measurement AVL DiTEST MCS 120
28 User Manual
Remarks to Connecting a cell module
Cable „plus“, „minus“ and cable „CAN“:
Fig. 4-6 Connect a cell module
To disconnect the "plus" and "minus" cables, pull on the plug:
Fig. 4-7 Connecting a cell module
Close the cover:
Fig. 4-8 Close the cover
+
-
CAN
AVL DiTEST MCS 120 Performing a Measurement
User Manual 29
10. The cell module gets charged/discharged now. Section 1 shows all relevant data. By clicking F6 Stop you can stop the charging / discharging process.
11. After the charging / discharging is finished please click F8 Next.
Fig. 4-9 Screen „charging/discharging“
Performing a Measurement AVL DiTEST MCS 120
30 User Manual
12. A screen appears which shows you how to disconnect the cell module. Please follow the exact order shown.
13. Continue with F8 Next.
Fig. 4-10 Screen „Disconnect cell module“
AVL DiTEST MCS 120 Performing a Measurement
User Manual 31
14. On this screen the report is shown. F4 Print Starts a print out of the report. F5 Edit The report can be edited.
15. By pressing F8 Next the charge / discharge process is finished.
Fig. 4-11 Screen „Report“
Performing a Measurement AVL DiTEST MCS 120
32 User Manual
4.1.1 Safety test
1. Start the DSS Manager by selecting: Start│Programs│DiTEST│DSS. 2. Select Diagnostic│Module Balancer.
Fig. 4-12 DSS Manager
3. The start screen appears: Select Safety test.
Fig. 4-13 Start screen
AVL DiTEST MCS 120 Performing a Measurement
User Manual 33
4. Connect the cell monitoring cable and close the cover. Continue with F8 Next.
Fig. 4-14 Safety test
5. Open the cover and disconnect the monitoring plug. Continue with F8 Next.
Fig. 4-15 Safety test
Performing a Measurement AVL DiTEST MCS 120
34 User Manual
6. Is the safety test passed successfully, the following screen appears. With F8 Next to the start screen.
Fig. 4-16 Safety test
AVL DiTEST MCS 120 Performing a Measurement
User Manual 35
4.1.2 Self-test
1. Start the DSS Manager by selecting: Start│Programs│DiTEST│DSS. 2. Select Diagnostic│Module Balancer.
Fig. 4-17 DSS Manager
3. The start screen appears: Select Selftest.
Fig. 4-18 Start screen
Performing a Measurement AVL DiTEST MCS 120
36 User Manual
4. Observe the screen, and disconnect the cables shown.
Fig. 4-19 Selftest
5. The self-test is performed and the result displayed. Return to the Start screen with F8 Next.
Fig. 4-20 Selftest
AVL DiTEST MCS 120 Performing a Measurement
User Manual 37
4.1.3 Deviceinfo
1. Start the DSS Manager by selecting: Start│Programs│DiTEST│DSS. 2. Select Diagnostic│Module Balancer.
Fig. 4-21 DSS Manager
3. The start screen appears: Select Device info.
Fig. 4-22 Start screen
Performing a Measurement AVL DiTEST MCS 120
38 User Manual
4. A screen appears with the device data. With F8 Next to the home screen.
Fig. 4-23 Device info
AVL DiTEST MCS 120 Performing a Measurement
User Manual 39
4.1.4 Viewing all reports
With this function you can manage the result protocols.
Click on Result Management.
Fig. 4-24 Result management
Source Select Local / Network / Saved Results.
Period Select Day, Week, Month, Year or All.
You can flip forwards and backwards using the arrows.
Report type Select Proof / Diagnosis Logs etc.
Search text Choose between “None”, “Manufacturer” or “Licence number”. In the Search text field, you can narrow the selection down further.
Return to “Results Management” with Quit.
Performing a Measurement AVL DiTEST MCS 120
40 User Manual
Individual logs can be selected by checking a box.
F2 Choose Selects all logs.
F3 Discard Discards the log selection.
F4 Print Prints the selected logs.
F5 Delete Deletes the selected logs.
F6 Export Exports the selected logs.
F5 Directory Opens a window to choose the save location. Choose a location and then click OK. F8 Next Starts the export.
F7 Archive Archives the selected logs. The archived logs will be deleted from the internal database.
F5 Directory Opens a window to choose the save location. Choose a location and then click OK. F8 Next Starts the archiving.
<< Cancel Quits the results display.
AVL DiTEST MCS 120 Performing a Measurement
User Manual 41
4.1.5 Result report directory
Result protocols are stored for future use at the default location: “C:\Users\Public\Documents\xxxx“.
This location can be changed:
Click on ExtrasSettingsAVL DiGate480.
Fig. 4-25 Changing the Report protocol directory
Result storage for backup by year and month
If this item is activated, can be selected by clicking on the path of the
network drive on which are stored the results (XML and PDF format).
With Test the connection to the network drive can be tested.
Directory for transmission to external applications
---
Performing a Measurement AVL DiTEST MCS 120
42 User Manual
4.1.6 Entering a list of operator names
Using this function you can manage a list of operator names.
The list appears in operation mode charge/discharge.
Click on ExtrasSettingsPersons.
Clicking New opens an input window into which you can enter the
name of a new tester.
Return to “Tester” with OK.
Clicking Delete deletes the blue-highlighted operator name.
Clicking Edit opens an input window in which you can edit the
blue-highlighted tester name.
Return to “Tester” with OK.
Clicking Current Operator selects the blue-highlighted operator name
(or clicking the desired operator name).
Clicking Responsible Operator selects the blue-highlighted tester name
(or clicking the desired operator name).
4.1.7 Switch off
The MCS 120 has no on/off switch.
With the button, see fig. 2-10, only the power modules are switched off.
The MCS 120 enters Idle, Sleep mode.
To disconnect the MCS 120 from the power grid press the red button (Power module off) and disconnect the power cable from the wall outlet.
AVL DiTEST MCS 120 Troubleshooting
User Manual 43
5 Troubleshooting
Maintenance
DANGER
Danger to life by electrical shock
Maintenance is only allowed by AVL DiTEST!
Do not open the MCS 120, because exposed parts can have dangerous voltages!
Troubleshooting AVL DiTEST MCS 120
44 User Manual
AVL DiTEST MCS 120 Maintenance and care
User Manual 45
6 Maintenance and care
6.1 Optical check
Carry out a regular optical check of the MCS 120, USB cable and measuring cables with the test adapters.
Check for damage and dirt.
CAUTION
Damaged Parts (power cord, adapter and USB cables) have to be exchanged immediately!
6.2 Cleaning
If the housing is dirty and you want to clean it, please use a dry cloth.
CAUTION
Before cleaning the MCS 120, switch off the Power module and remove the mains plug!
Remove all adapter cables!
Be sure that no liquid is flowing into the housing!
Maintenance and care AVL DiTEST MCS 120
46 User Manual
AVL DiTEST MCS 120 Scope of delivery
User Manual 47
7 Scope of delivery
Standard delivery:
Part ID number
Module Conditioning System MCS incl. power cable (GB)
VS
Transport box
VS
USB cable (USB 2.0 A - USB 2.0 B Rugged)
EX7069
Cable Plus-Minus, X590
Plus & Minus cable with knurled nut M6
VS
Cable analogue module, X590
Analogue module cable adapter
(without CSC Cell-supervision-Circuit)
VS
Cable set bus modules, X590
Module adapter cable for modules with CAN bus (with CSC Cell-Supervision-Circuit)
VS
User manual
AT7953E
Scope of delivery AVL DiTEST MCS 120
48 User Manual
Spare parts:
Part ID number
USB cable (USB 2.0 A - USB 2.0 B Rugged)
EX7069
Cable Plus-Minus, X590
Plus & Minus cable with knurled nut M6
Cable analogue module, X590
Analogue module cable adapter
(without CSC Cell-Supervision-Circuit)
Cable set bus modules, X590
Module adapter cable for modules with CAN bus (with CSC Cell-Supervision-Circuit)
Further customer-specific cable sets on request.
AVL DiTEST MCS 120 Technical data
User Manual 49
8 Technical data
8.1 Operating data
AC-Power grid
Nominal voltage range
Power supply
Frequency
Power consumption
100 … 240 V~
50 ... 60 Hz
2150 W
Charging/discharging power
2 - 75 V
-48 - +80 A
max. 2000 W
Cell monitoring (charge/Discharge)
Depends on type of cell module via analog measurements and via the CAN interface. Monitored values are voltages, currents and temperatures.
External interface
USB
USB 2.0-Rugged
Ambient conditions Operating
Transport and storage
Ambient
temperature
Relative humidity
Ambien
temperature Relative humidity
0 to +40 °C
10 to 80 %
non condensing
-20 to +55 °C
10 to 80 %,
non condensing
Disposal:
This AVL DiTEST product is a high-quality electrical and electronic device that must not be disposed of in the household waste.
When disposing of this system, you must comply with your local legal requirements!
Technical data AVL DiTEST MCS 120
50 User Manual
AVL DiTEST MCS 120 Index
User Manual 51
9 Index
C Cable Set plus minus, X590 16
Cable sets 16
Cable, analog module, X590 VII, 16
Cable, CAN bus module 17
Certificate of Calibration 4
Cleaning 45
Commissioning 19
Components 7
D Designated Use 6
Deviceinfo 37
E EC Declaration of Conformity/ CE mark 2
Entering a list of operator names 42
F Firmware Update 19
G General 1
General Notes 1
General Safety III
H Hardware 20
I IMPORTANT SAFETY INSTRUCTIONS V
Installation of the DSS Software 19
M Maintenance 43
Maintenance and care 45
O Operating data 49
Optical check 45
R Result report directory 41
S Safety test 32
Self-test 35
SPI-UART-Adapter 18
Switch off 42
T Technical data 49
Troubleshooting 43
U USB-cable (USB 2.0 A auf USB 2.0 B Rugged) EX7069 16
V Viewing all reports 39
W Warning and Safety Notes I
Index AVL DiTEST MCS 120
52 User Manual