Measurement And
Instrumentations
DR. Eng. Samir Elshamy
DR. Eng. Samir Elshamy1
![Page 2: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/2.jpg)
Definition of TermsReadability of an instrument:This term indicates the closeness with which the scale of the instrument
may be read; an instrument with a 12-cm scale would have a higher readability than an instrument with a 6-cm scale and the same range.
The least count:is the smallest difference between two indications that can be detected on the instrument scale.
DR. Eng. Samir Elshamy2
![Page 3: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/3.jpg)
The sensitivity of an instrument:is the ratio of the linear movement of the pointer on an analog instrument to the change in the measured variable causing this motion.For example, a 1-mV recorder might have a 25-cm scale length. Its sensitivity would be 25 cm/mV
DR. Eng. Samir Elshamy3
![Page 4: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/4.jpg)
Accuracy and Precision
DR. Eng. Samir Elshamy4
![Page 5: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/5.jpg)
The accuracy of an instrument
indicates the deviation of the reading from a known input. Accuracy is
frequently expressed as a percentage of full-scale reading, so that a 100-kPa
pressure gage having an accuracy of 1 percent would be accurate within ±1
kPa over the entire range of the gage.
DR. Eng. Samir Elshamy5
![Page 6: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/6.jpg)
The precision of an instrument
indicates its ability to reproduce a certain reading with a given accuracy. As an
example of the distinction between precision and accuracy, consider the
measurement of a known voltage of 100 volts (V) with a certain meter. Four
readings are taken, and the indicated values are 104, 103, 105, and 105 V. From
these values it is seen that the instrument could not be depended on for an
accuracy of better than 5 percent (5 V), while a precision of ±1 percent is
indicated since the maximum deviation from the mean reading of 104 V is only
1 V. DR. Eng. Samir Elshamy6
![Page 7: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/7.jpg)
DR. Eng. Samir Elshamy7
![Page 8: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/8.jpg)
Calibration
The calibration of all instruments is important, for it affords the opportunity to check the instrument against a known standard and subsequently to reduce errors in accuracy. Calibration procedures involve a comparison of the particular instrument with either(1)a primary standard, (2) a secondary standard with a higher accuracy than the instrument to be calibrated(3) a known input source.
DR. Eng. Samir Elshamy8
![Page 9: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/9.jpg)
For example, a flow meter might be calibrated by
(1) comparing it with a standard flow-measurement facility
of the National Institute for Standards and Technology (NIST),
(2) comparing it with another flow meter of known accuracy,
(3) directly calibrating with a primary measurement such as
weighing a certain amount of water in a tank and recording the
time elapsed for this quantity to flow through the meter.
DR. Eng. Samir Elshamy9
![Page 10: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/10.jpg)
The Generalized Measurement System منظام القياس المعم
Most measurement systems may be divided into three parts:1. A detector-transducer stage, which detects the physical variable
and performs either a mechanical or an electrical transformation to convert the signal into a more usable form.
2. Some intermediate stage, (Signal conditioning)which modifies the direct signal by amplification, filtering, or other means so that a desirable output is available.
3. A final or terminating stage, which acts to indicate, record, or control the variable being measured. The output may also be digital or analog. DR. Eng. Samir Elshamy10
![Page 11: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/11.jpg)
Bourdon-tube pressure gage
1) The bourdon tube is the detector-transducer stage because it converts the pressure signal into a mechanical displacement of the tube.
2) The intermediate stage consists of the gearing arrangement, which amplifies the displacement of the end of the tube so that a relatively small displacement at that point produces as much as three-quarters of a revolution of the center gear.
3) The final indicator stage consists of the pointer and the dial arrangement, which, when calibrated with known pressure inputs, gives an indication of the pressure signal impressed on the bourdon tube.
DR. Eng. Samir Elshamy11
![Page 12: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/12.jpg)
DR. Eng. Samir Elshamy12
![Page 13: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/13.jpg)
DR. Eng. Samir Elshamy
Analysis of Experimental DataIn this section we present a discussion of some of the types of
errors that may be present in experimental data and begin to
indicate the way these data may be handled.
13
![Page 14: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/14.jpg)
DR. Eng. Samir Elshamy
Types of Errors
Random
Systematic
Gross error
Environmental
Observational
Instrumental
14
![Page 15: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/15.jpg)
DR. Eng. Samir Elshamy
An experimental error is an experimental error. If the experimenter knew what
the error was, he would correct it and it would no longer be an error. In other
words, the real errors in experimental data are those factors that are always
vague to some extent and carry some amount of uncertaintyعدم اليقين.
Our task is to determine just how uncertain a particular observation may be
and to devise a consistent way of specifying the uncertainty in analytical form.
A reasonable definition of experimental uncertainty may be taken as the
possible value the error may have. This uncertainty may vary a great deal
depending on the circumstances of the experiment. Perhaps it is better to
speak of experimental uncertainty instead of experimental error because the
magnitude of an error is always uncertain. Both terms are used in practice,
however, so the reader should be familiar with the meaning attached to the
terms and the ways that they relate to each other.
Error Analysis on a Commonsense Basis تحليل الخطأ علي أساس المنطق
15
![Page 16: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/16.jpg)
DR. Eng. Samir Elshamy16
Consider the calculation of electric power from
P = EIwhere E and I are measured asE = 100 V ± 2 VI = 10 A ± 0.2 AThe nominal value of the power is 100×10 = 1000W. By taking the worst possible variations in voltage and current, we could calculatePmax = (100 + 2)(10 + 0.2) = 1040.4 WPmin = (100 − 2)(10 − 0.2) = 960.4 WThus, using this method of calculationthe uncertainty in the power is= +4.04 % & −3.96 %.
![Page 17: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/17.jpg)
DR. Eng. Samir Elshamy17
Static error (A) :
is defined as the difference between the best measured value
and the true value A=A1-A
Where A1=measured value & A =true value
Relative static error(uncertainty)= (A/A)
![Page 18: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/18.jpg)
DR. Eng. Samir Elshamy18
The result R is a given function of the independent variables x1, x2, x3, . . . , xn. Thus, R = R(x1, x2, x3, . . . , xn)Let ωR be the uncertainty in the result and ω1, ω 2, . . . , ωn be the uncertainties in the independent variables.
If this relation is applied to the electric-power relation of the previous section, the expected uncertainty is 2.83 percent instead of 4.04 percent.
![Page 19: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/19.jpg)
DR. Eng. Samir Elshamy19
Example 3.1 :UNCERTAINTY OF RESISTANCE OF A COPPER WIRE.
The resistance of a certain size of copper wire is given as
R = R0[1 + α(T − 20)]
where R0 = 6 ± 0.3 percent is the resistance at 20◦C,
α = 0.004◦C−1 ± 1 percent is the temperature coefficient of
resistance, and the temperature of the wire is T = 30±1◦C.
Calculate the resistance of the wire and its uncertainty.
![Page 20: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/20.jpg)
DR. Eng. Samir Elshamy20
![Page 21: Types of Error](https://reader030.fdocuments.us/reader030/viewer/2022020311/577c84b31a28abe054b9f9e3/html5/thumbnails/21.jpg)
Thanks for your kind attention