THE COBALT BOMB!!!DR REJIL RAJAN
![Page 2: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/2.jpg)
TELETHERAP
Y
•Source of radiation is at a distance from the patient
ISOTOPIC TELETHE
RAPY
•Treatment machines incorporating γ ray sources for use in external beam radiotherapy
![Page 3: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/3.jpg)
HISTORY
Until 1951, all isotope machines produced were tele-radium units ( radium bomb ).
The source to skin distance was usually not greater than 10 cm in
these machines.
Major drawbacks of these machines were-i) high risk of radiation hazard due to radon gas leak produced as a by product, ii) high cost of radium, iii) large self absorption, iv) low exposure rate constant and v) low output.
ERA OF TELE-RADIUM AND TELE-CESIUM
![Page 4: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/4.jpg)
Bryant Symons radium "bomb" at Westminster Hospital, London, England, in the 1930s.
![Page 5: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/5.jpg)
For tele-cesium units the source to skin distance is 20cm to 40cm.
They have not been very popular because of relatively low exposure rate constant and low specific activity.
![Page 6: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/6.jpg)
THE COBALT BOMBDr. Harold E. Johns - Invented the cobalt-60 unit in
the 1950s.
First Tele-Cobalt machines –Saskatoon Cancer
Clinic , Canada and the Victoria Hospital, London
![Page 7: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/7.jpg)
COBALT-60 V/S RADIUM-226 AND CESIUM-137
![Page 8: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/8.jpg)
PICKER MACHINE
Fixed head
SSD Technique
![Page 9: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/9.jpg)
JANUS Co-60 machine. Ulrich Henschke, Bronx, New York
Co-60 source of about 2000 Ci. Two specially designed rooms were built adjacent to each other
Roman MythologyTwo heads looking in opposite directionsThe gift to see both future and past
![Page 10: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/10.jpg)
Iso-centric Cobalt-60 Teletherapy Machine, Atomic Energy of Canada, Ltd.
Isocentric
SSD : 60cm
Source activity :low
Output low Time taken MORE
![Page 11: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/11.jpg)
The main components of a teletherapy machine are:
1. radioactive source;
2. source housing, including beam collimator and source movement mechanism;
3. a gantry and stand in isocentric machines or a housing support assembly in stand-alone machines;
4. a patient support assembly
5. a machine console.
COMPONENTS
![Page 12: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/12.jpg)
![Page 13: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/13.jpg)
![Page 14: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/14.jpg)
Cylindrical double-welded stainless-steel capsule
Diameter ranging from 1.0 to 2.0 cm
Discs / pellets
Positioned in the cobalt unit with its circular end facing the patient
COBALT – 60 SOURCE
![Page 15: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/15.jpg)
DECAY
![Page 16: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/16.jpg)
• The Cobalt-60 source is produced by irradiating ordinary, stable 59Co with neutrons in a nuclear reactor.
•The nuclear reaction is represented as
• The nuclei of 60Ni will be in the excited states following this decay and the de-excite to the ground state by emitting two γ ray photons of energy 1.17 MeV and 1.33 MeV in cascade.
• The decay half-life is 5.26 years and the average photon energy is 1.25 MeV.
![Page 17: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/17.jpg)
The housing for the source is called the “source head”.
It consists of a steel shell filled with lead for shielding purposes and device for bringing the source in front of an opening in the head from which the useful beam emerges.
Also a heavy metal alloy sleeve is provided to form an additional primary shield when the source is in the OFF position.
SOURCE HOUSING
![Page 18: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/18.jpg)
![Page 19: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/19.jpg)
A number of methods have been developed for moving the source from OFF position to ON position-
1. Source mounted on a rotating wheel inside the source head to carry the source from OFF to On position
2. Source mounted on a heavy metal drawer is moved horizontally by pneumatic system through a hole running through the source head. In the ON position the source faces the aperture for the treatment beam and in the OFF position the source moves to its shielded location and the light source mounted on the same drawer occupies the ON position of the source.
ON OFF
![Page 20: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/20.jpg)
3. Mercury is allowed to flow into a container immediately below the source to shut OFF the beam.
4. Source is fixed in front of the aperture and the beam can be turned ON and OFF by a shutter consisting of heavy metal jaws.
![Page 21: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/21.jpg)
![Page 22: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/22.jpg)
Collimators provide beams of desired shape and size.
Collimators of teletherapy machines provide square and rectangular radiation fields typically ranging from 5 × 5 to 35 × 35 cm2 at 80 cm from the source
The rotational movement of the collimator is continuous, and it can rotate 360° about its own axis.
The collimator system can move to any position when the gantry is rotated.
COLLIMATORS
![Page 23: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/23.jpg)
The gantry can rotate by 360°. The rotational movement of the gantry is motorized and controlled in two directionscontinuously; its rotation speed can be adjusted.
Teletherapy machines are most often mountedisocentrically, allowing the beam to rotate about the patient at a fixed SAD. They can be used either as fixed field machines or rotation units.
GANTRY
![Page 24: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/24.jpg)
Treatment Bed has motorized movements Horizontal Vertical Lateral
Table Top – 90 rotation to each side Base - 110 rotation to each side
PATIENT SUPPORT ASSEMBLY
![Page 25: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/25.jpg)
Control Console is situated outside the bunkerInterlocks present on the console for 1. Air Pressure2. Door3. Head Lock -Treatment Head has a swivel movement of +/- 1804. OFF Shield5. Treatment Mode6. Wedge Filter7. Tray Interlock8. Timer
CONTROL CONSOLE
![Page 26: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/26.jpg)
Timer The prescribed target dose is delivered with the help of two treatment timers: Primary Timer - the primary timer actually controls the treatment time. Secondary Timer - accounts for the source movement from OFF to ON position and again to OFF position (shutter error).
Source ON/OFF Indicator – Red- ON Green- OFF Amber- TRANSIT
![Page 27: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/27.jpg)
• γ rays constitute the useful treatment beam.
• The β particles are absorbed in the cobalt metal and the stainless steel capsules resulting in the emission of bremsstrahlung x-rays and a small amount of characteristic x-rays.
TREATMENT BEAM
![Page 28: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/28.jpg)
• The lower energy γ rays produced by the interaction of the primary γ radiation with the source itself, the surrounding capsule, the source housing and the collimator system are also contaminants of the treatment beam.
• The scattered components of the beam contribute significantly ( approx. 10% to the total intensity of the beam.
![Page 29: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/29.jpg)
COBALT 60 V/S LINAC
![Page 30: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/30.jpg)
![Page 31: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/31.jpg)
![Page 32: THE COBALT BOMB](https://reader036.fdocuments.us/reader036/viewer/2022081421/587ac0a21a28abc0478b6c29/html5/thumbnails/32.jpg)
BHABHATRON - The indigenous Cobalt-60 Teletherapy Machine has a capacity of 200 RMM source and its source to iso-centre distance is 80 cm. The system has unique user-friendly features and fully closable collimator for improved radiation safety.