Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) •...

67
Sterilization and Quality Assurance Procedures Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved Quality Assurance Procedures Eric S. Kastango October 10, 2013

Transcript of Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) •...

Page 1: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Sterilization andQuality Assurance Procedures

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights ReservedCopyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

Sterilization andQuality Assurance Procedures

Eric S. KastangoOctober 10, 2013

Page 2: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

The essence of quality assuranceis demonstrating that you are reallydoing what you say you are doing.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

The essence of quality assuranceis demonstrating that you are reallydoing what you say you are doing.

Page 3: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Standard of Sterility

• <1> Injections– Parenteral articles are prepared….to ensure they meet

pharmacopeial requirements for sterility, pyrogens,...– Sterility Tests – Preparations for injection meet the

requirements under Sterility Tests <71>

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• <1> Injections– Parenteral articles are prepared….to ensure they meet

pharmacopeial requirements for sterility, pyrogens,...– Sterility Tests – Preparations for injection meet the

requirements under Sterility Tests <71>

Page 4: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

What is “Sterility”?

• “Free from bacteria or other microorganisms”– American Heritage's Definition of Sterility

• “Within the strictest definition of sterility, a specimen would bedeemed sterile only when there is complete absence of viablemicroorganisms from it.”– <1211> Sterilization and Sterility Assurance of Compendial Articles

• Is it possible to demonstrate complete absence of microorganismsfrom a CSP?

• Absolute sterility can’t be demonstrated without the completedestruction of every article from the lot of CSPs.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• “Free from bacteria or other microorganisms”– American Heritage's Definition of Sterility

• “Within the strictest definition of sterility, a specimen would bedeemed sterile only when there is complete absence of viablemicroorganisms from it.”– <1211> Sterilization and Sterility Assurance of Compendial Articles

• Is it possible to demonstrate complete absence of microorganismsfrom a CSP?

• Absolute sterility can’t be demonstrated without the completedestruction of every article from the lot of CSPs.

4

Page 5: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Critical Concepts of Sterilization

• Sterility Assurance Level (SAL) is the probability of a non-sterileitem making it through the validated sterilization process.

• Items terminally sterilized by moist or dry heat, irradiation, orchemical sterilants have a SAL of 10-6

– 1 nonsterile item per 1 million items sterilized

• Items prepared aseptically prepared with a 0.22 micron filter havea SAL of 10-3

– 1 nonsterile item per 1 thousand items sterilized

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Sterility Assurance Level (SAL) is the probability of a non-sterileitem making it through the validated sterilization process.

• Items terminally sterilized by moist or dry heat, irradiation, orchemical sterilants have a SAL of 10-6

– 1 nonsterile item per 1 million items sterilized

• Items prepared aseptically prepared with a 0.22 micron filter havea SAL of 10-3

– 1 nonsterile item per 1 thousand items sterilized

Page 6: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Verification of Accuracy and Sterilization

• The quality (sterility and accuracy) of the CSP is directly related toensuring that methods used to compound the sterile preparationachieve the desired goal of purity, potency and sterility.

• CSPs that require some form of terminal sterilization:– either by filtration

– Steam

– ionizing radiation (not the subject of this session)

MUST be verified to ensure that each CSP is void of microbialcontamination.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• The quality (sterility and accuracy) of the CSP is directly related toensuring that methods used to compound the sterile preparationachieve the desired goal of purity, potency and sterility.

• CSPs that require some form of terminal sterilization:– either by filtration

– Steam

– ionizing radiation (not the subject of this session)

MUST be verified to ensure that each CSP is void of microbialcontamination.

Page 7: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Verification of sterilization

• High-risk level compounding must achieve sterility– (nonsterile sterile)

• Methods of sterilization– Autoclave or Dry-Heat oven sterilization cycles

• Autoclave (121°C for 12-30 min @ 15-20 PSI)• Dry-heat

– 150°C (300°F) for 2.5 hours– 140°C (285 °F) for 3 hours

• Use of Biological Indicators (BIs)

– Filtration methodology– FDA approved/challenged filters– Filter integrity testing

– Ionizing radiation

The stated temperaturesand cycle times must beindependently qualifiedby user

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• High-risk level compounding must achieve sterility– (nonsterile sterile)

• Methods of sterilization– Autoclave or Dry-Heat oven sterilization cycles

• Autoclave (121°C for 12-30 min @ 15-20 PSI)• Dry-heat

– 150°C (300°F) for 2.5 hours– 140°C (285 °F) for 3 hours

• Use of Biological Indicators (BIs)

– Filtration methodology– FDA approved/challenged filters– Filter integrity testing

– Ionizing radiation

The stated temperaturesand cycle times must beindependently qualifiedby user

Page 8: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Responsibility of Compounding Supervisor

• The licensed healthcare professionals who supervise compoundingshall be responsible for determining that the selected sterilizationmethod both sterilizes and maintains the strength, purity, quality,and packaging integrity of CSPs.– See Methods of Sterilization under Sterilization and Sterility Assurance

of Compendial Articles 1211

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• The licensed healthcare professionals who supervise compoundingshall be responsible for determining that the selected sterilizationmethod both sterilizes and maintains the strength, purity, quality,and packaging integrity of CSPs.– See Methods of Sterilization under Sterilization and Sterility Assurance

of Compendial Articles 1211

Page 9: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Methods of Sterilization

• Sterilization method must bothsterilize and maintain thestrength, purity, and packagingintegrity

• Methods– Steam– Dry Heat– Filtration

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Sterilization method must bothsterilize and maintain thestrength, purity, and packagingintegrity

• Methods– Steam– Dry Heat– Filtration

Image courtesy of www.tuttnauer.com

Page 10: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Steam (Moist Heat) Sterilization

• Autoclaves are used to sterilizeheat-stable oil-free liquids,glassware, utensils andmedical instruments.

• Lethality is achieved bydenaturing proteins ofmicroorganisms.

• Equipment must have systemsto control temperature andcycle time.

• The item is exposed tosaturated steam under hightemperature and pressureconditions.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Autoclaves are used to sterilizeheat-stable oil-free liquids,glassware, utensils andmedical instruments.

• Lethality is achieved bydenaturing proteins ofmicroorganisms.

• Equipment must have systemsto control temperature andcycle time.

• The item is exposed tosaturated steam under hightemperature and pressureconditions.

Page 11: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

How to load an autoclave

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

http://www.acad.polyu.edu.hk/~mmktlau/ME429/autoclave_2.gif

Image: http://ehs.unc.edu/

Page 12: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Time to Achieve Equivalent Microbial Lethality atDifferent Exposure Temperatures

(0.13 Minutes )

( 0.9 Minutes )

12 Minutes

80 Hours

643 Hours

10 20 30 40 50 60 100 300 500 700MINUTES HOURS

176° F (80° C)

200° F (93° C)

212° F (100° C)

270° F (132° C)

285° F (140° C)

250° F (121° C)

321 Hours

(0.13 Minutes )

( 0.9 Minutes )

12 Minutes

270° F (132° C)

285° F (140° C)

250° F (121° C)

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

Used with permission from Steris

(0.13 Minutes )

( 0.9 Minutes )

12 Minutes

80 Hours

643 Hours

10 20 30 40 50 60 100 300 500 700MINUTES HOURS

176° F (80° C)

200° F (93° C)

212° F (100° C)

270° F (132° C)

285° F (140° C)

250° F (121° C)

321 Hours

12 Minutes

80 Hours

643 Hours

10 20 30 40 50 60 100 300 500 700MINUTES HOURS

176° F (80° C)

200° F (93° C)

212° F (100° C)

250° F (121° C)

321 Hours

Page 13: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Chemical and Biological Indicators

• Biological Indicator (BIs) aresystems that use a heat-resistant spore-formingmicroorganism (B.stearothermophilus or similar)

• Biological Indicators (BIs) areused to determine whether asterilizer has delivered a lethal(effective) cycle.– They should be used with every

cycle and in conjunction with time-temperature data generated fromthe sterilizer

• Autoclave tape does not meetthis requirement.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Biological Indicator (BIs) aresystems that use a heat-resistant spore-formingmicroorganism (B.stearothermophilus or similar)

• Biological Indicators (BIs) areused to determine whether asterilizer has delivered a lethal(effective) cycle.– They should be used with every

cycle and in conjunction with time-temperature data generated fromthe sterilizer

• Autoclave tape does not meetthis requirement.

Page13

Unprocessed Positive Control Processed

Image adapted from: Autoclave Training; Universityof Kentucky. 2005.

Page 14: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Steam Sterilization (continued)

• The effectiveness of steamsterilization is verified:– Using confirmation

methods such astemperature sensingdevices-thermocouples

• Temperature mapping– Hot and cold spots

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• The effectiveness of steamsterilization is verified:– Using confirmation

methods such astemperature sensingdevices-thermocouples

• Temperature mapping– Hot and cold spots

Image courtesy of http://www.ellab.com/

Page 15: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Steam Sterilization (continued)

• Biological Indicators– Need to incubate after exposure to steam cycle to determine

effectiveness.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

Image courtesy of http://www.confirmmonitoring.com/spore.asp/

Page 16: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Inspector Evaluation Metrics: Steam Autoclave

• Who operates the equipment?– See training records

• How is the equipment operated?– See PnPs or SOPs on its use

• How do you know the autoclave cycle waseffective?– What system is used to control temperature and

cycle time• Print out of time/temperature

– Typical temp: 121°C for at least 12 minutes (drugdependent)

– What type of monitoring and controlling sensorsare used (e.g., mercury-in-glass thermometer,thermocouple, RTD, pressure gauge)?

– Use of biological indicator (tape is not acceptable)• How is the equipment maintained, cleaned,

services, and calibrated

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Who operates the equipment?– See training records

• How is the equipment operated?– See PnPs or SOPs on its use

• How do you know the autoclave cycle waseffective?– What system is used to control temperature and

cycle time• Print out of time/temperature

– Typical temp: 121°C for at least 12 minutes (drugdependent)

– What type of monitoring and controlling sensorsare used (e.g., mercury-in-glass thermometer,thermocouple, RTD, pressure gauge)?

– Use of biological indicator (tape is not acceptable)• How is the equipment maintained, cleaned,

services, and calibrated

Page 17: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Dry Heat Sterilization and Depyrogenation

• Used for oils and powders

• Designed to provide heated filteredair evenly distributed throughout thechamber by a blower device (air mustcirculate around items)

• Requires higher temperature andlong exposure times

• Oven should be equipped with asystem for controlling temperatureand exposure period

• Effectiveness must be verified usingappropriate biological indicators andtemperature sensing devices

http://sterisure.com/product

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Used for oils and powders

• Designed to provide heated filteredair evenly distributed throughout thechamber by a blower device (air mustcirculate around items)

• Requires higher temperature andlong exposure times

• Oven should be equipped with asystem for controlling temperatureand exposure period

• Effectiveness must be verified usingappropriate biological indicators andtemperature sensing devices

http://www.despatch.com/lac.aspx

Page 18: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Dry Heat Depyrogenation

• Used on equipment made of glass or metal– Rinse with Sterile Water prior to use

• Requires a much higher temperature than sterilization: 250°C for30 minutes

• Validation of a depyrogenation cycle requires inoculation ofbacterial endotoxin

• A three log reduction of E. Coli endotoxin is required to have aneffective depyrogenation cycle

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Used on equipment made of glass or metal– Rinse with Sterile Water prior to use

• Requires a much higher temperature than sterilization: 250°C for30 minutes

• Validation of a depyrogenation cycle requires inoculation ofbacterial endotoxin

• A three log reduction of E. Coli endotoxin is required to have aneffective depyrogenation cycle

Page 19: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Chemical and Biological Indicators

• Biological Indicator (BIs) aresystems that use a heat-resistantspore-forming microorganism(Bacillus atrophaeus or similar)

• Biological Indicators (BIs) are usedto determine whether a sterilizerhas delivered a lethal (effective)cycle.– They should be used with every

cycle and in conjunction withtime-temperature data generatedfrom the sterilizer

• Autoclave tape does not meet thisrequirement.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Biological Indicator (BIs) aresystems that use a heat-resistantspore-forming microorganism(Bacillus atrophaeus or similar)

• Biological Indicators (BIs) are usedto determine whether a sterilizerhas delivered a lethal (effective)cycle.– They should be used with every

cycle and in conjunction withtime-temperature data generatedfrom the sterilizer

• Autoclave tape does not meet thisrequirement.

Page19

http://www.mesalabs.com/mesastrip/

Page 20: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Biological Indicators

SterilizationMethod BI microorganism BI incubation

Steam Geobacillus stearothermophilus Seven days at 50-55°C

Dry-Heat Bacillus atrophaeus Seven days at 30-35°C

Dry-HeatDepyrogenation

Inoculate glassware or vials with aminimum of 5,000 EU of E. colilipopolysaccharide.

All inoculated glassware placed mustdemonstrate, at minimum, a three-log reduction in endotoxin.

N/A

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

Dry-HeatDepyrogenation

Inoculate glassware or vials with aminimum of 5,000 EU of E. colilipopolysaccharide.

All inoculated glassware placed mustdemonstrate, at minimum, a three-log reduction in endotoxin.

Page 21: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Polling Question

The use of biological indicators is not required with eachbatch of drug that is terminally sterilized by either steam ordry-heat?

1. True

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

1. True2. False

Kate
Highlight
Page 22: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Inspector Evaluation Metrics: Dry Heat Oven• Who operates the equipment?

– See training records• How is the equipment operated?

– See PnPs or SOPs on its use• How do you know the dry-heat cycle was effective?

– What system is used to control temperature and cycletime?• Print out of time/temperature

– Typical temp: 250° C for at least 30 minutes (drugdependent)

– What type of monitoring and controlling sensors areused (e.g., mercury-in-glass thermometer,thermocouple, RTD, pressure gauge)?

– Use of biological indicator (tape is not acceptable)• Different BI for steam vs. dry heat

• How is the equipment maintained, cleaned, services,and calibrated?

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Who operates the equipment?– See training records

• How is the equipment operated?– See PnPs or SOPs on its use

• How do you know the dry-heat cycle was effective?– What system is used to control temperature and cycle

time?• Print out of time/temperature

– Typical temp: 250° C for at least 30 minutes (drugdependent)

– What type of monitoring and controlling sensors areused (e.g., mercury-in-glass thermometer,thermocouple, RTD, pressure gauge)?

– Use of biological indicator (tape is not acceptable)• Different BI for steam vs. dry heat

• How is the equipment maintained, cleaned, services,and calibrated?

Page 23: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Inspector Evaluation Metrics: Sterilization

Selected Work Practices Inspector Evaluation Metrics

Sterilization-Steam(autoclave) and Dry-HeatOven Sterilization methods

• Inspect equipment

Maintained, clean and located in a suitable area?

• Are compounded solutions pre-filtered prior toautoclaving?

• How are the supplies wrapped/packaged prior tosterilizing?

• Equipment has controls for controlling temperatureand cycle time

Evidence: Print-out of time/temperature?

Linked to batch?

• Are biological indicators used during the cycle (onetime or every time?)

• Are written CSP specific steam autoclave and dry-heat over PnP maintained?

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Inspect equipment

Maintained, clean and located in a suitable area?

• Are compounded solutions pre-filtered prior toautoclaving?

• How are the supplies wrapped/packaged prior tosterilizing?

• Equipment has controls for controlling temperatureand cycle time

Evidence: Print-out of time/temperature?

Linked to batch?

• Are biological indicators used during the cycle (onetime or every time?)

• Are written CSP specific steam autoclave and dry-heat over PnP maintained?

Page 24: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Filtration

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

An electron microscope image of a bacterial community, on a filter.Image courtesy of http://www.livescience.com/18397-marine-microbes-big-implications.html

Page 25: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Filtration

• Process for removing particulate matter from a liquid• Removes particles by entrapment in the channels of the filter,

adsorption to the filter by chemical or electrical attraction, bysieving, or size exclusion

• Most frequently, filters work by a combination of thesemechanisms

• SAL for sterilizing a solution by filtration is normally accepted as1:1000 (10-3)

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Process for removing particulate matter from a liquid• Removes particles by entrapment in the channels of the filter,

adsorption to the filter by chemical or electrical attraction, bysieving, or size exclusion

• Most frequently, filters work by a combination of thesemechanisms

• SAL for sterilizing a solution by filtration is normally accepted as1:1000 (10-3)

Page 26: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Filtration

• Sterile filters used to sterilizeCSPs shall be:– Pyrogen-free and have a nominal

porosity of 0.2 μm or 0.22 μm

– Certified by the manufacturer toretain at least 107 microorganismsof a strain of Brevundimonasdiminuta per cm2 of filter surfacearea

• The filter dimensions and liquidmaterial to be sterile-filtered shallpermit the sterilization process tobe completed rapidly, without thereplacement of the filter duringthe process.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Sterile filters used to sterilizeCSPs shall be:– Pyrogen-free and have a nominal

porosity of 0.2 μm or 0.22 μm

– Certified by the manufacturer toretain at least 107 microorganismsof a strain of Brevundimonasdiminuta per cm2 of filter surfacearea

• The filter dimensions and liquidmaterial to be sterile-filtered shallpermit the sterilization process tobe completed rapidly, without thereplacement of the filter duringthe process.

Image courtesy of http://www.pall.com

Page 27: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Filters

• Size and configuration shouldaccommodate the volume beingfiltered to permit completefiltration without clogging

• 25 mm disk filter should filter nomore than 100 mL

• Filter and housing should bephysically and chemicallycompatible with the product tobe filtered and capable ofwithstanding the temperature,pressures and hydrostatic stressimposed on the system

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Size and configuration shouldaccommodate the volume beingfiltered to permit completefiltration without clogging

• 25 mm disk filter should filter nomore than 100 mL

• Filter and housing should bephysically and chemicallycompatible with the product tobe filtered and capable ofwithstanding the temperature,pressures and hydrostatic stressimposed on the system

Image courtesy of http://www.millipore.com

Page 28: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Bubble-Point (Filter Integrity Test)

• A test to ensure the integrity offilters used for sterilizingcompounded solutions.

• The Bubble Point Test is based onthe fact that liquid is held in acapillary tube by surface tension.

• The minimum gas pressurerequired to force liquid out of thetube is a direct function of tubediameter

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• A test to ensure the integrity offilters used for sterilizingcompounded solutions.

• The Bubble Point Test is based onthe fact that liquid is held in acapillary tube by surface tension.

• The minimum gas pressurerequired to force liquid out of thetube is a direct function of tubediameter

Images courtesy of http://www.millipore.com

Page 29: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Filter Integrity Testing Millex® / Sterivex™ Integrity Tester

Bubble Point Procedure:1. Wet the filter with the appropriate

fluid, typically water for hydrophilicmembranes or an alcohol/watermixture for hydrophobic membranes.

2. Pressurize the system to about 80% ofthe expected bubble point pressurewhich is stated in manufacturer’sliterature.

3. Slowly increase the pressure until rapidcontinuous bubbling is observed at theoutlet.

4. A bubble point value lower than thespecification is an indication of thefollowing:• fluid with different surface tension

than the recommended test fluid• integral filter, but wrong pore size• high temperature• incompletely wetted membrane• non-integral membrane or seal

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

Images courtesy of http://www.millipore.com

Bubble Point Procedure:1. Wet the filter with the appropriate

fluid, typically water for hydrophilicmembranes or an alcohol/watermixture for hydrophobic membranes.

2. Pressurize the system to about 80% ofthe expected bubble point pressurewhich is stated in manufacturer’sliterature.

3. Slowly increase the pressure until rapidcontinuous bubbling is observed at theoutlet.

4. A bubble point value lower than thespecification is an indication of thefollowing:• fluid with different surface tension

than the recommended test fluid• integral filter, but wrong pore size• high temperature• incompletely wetted membrane• non-integral membrane or seal

Page 30: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Polling Question

Conducting filter integrity testing can be used as analternative quality release test to sterility testing?

1. True

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

2. False

Kate
Highlight
Page 31: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Inspector Evaluation Metrics: Filtration

• Is the filter a sterilizing grade filter?

• How was the filter membrane selected?

– Aqueous (hydrophilic)

– Non-aqueous (hydrophobic)

• What is the maximum filtration volume perfilter?

– See PnPs or SOPs on its use

• Are filters used to sterilize solutions integritytested (Bubble-point)?

• Are there specific written procedures forperforming a filter integrity test?

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Is the filter a sterilizing grade filter?

• How was the filter membrane selected?

– Aqueous (hydrophilic)

– Non-aqueous (hydrophobic)

• What is the maximum filtration volume perfilter?

– See PnPs or SOPs on its use

• Are filters used to sterilize solutions integritytested (Bubble-point)?

• Are there specific written procedures forperforming a filter integrity test?

Page 32: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Probability

“The sterility of a lot purported to be sterile istherefore defined in probabilistic terms, wherethe likelihood of a contaminated unit or article isacceptably remote.”

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

Page 33: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Sterility Testing Requirements

• High risk CSPs– Batches of more than 25

individual CSPs– Any high risk CSP that has

been exposed longer than 12hours at 2º to 8º or longerthan 6 hours at warmer than8º

• All risk level CSPs– When USP <797> BUD limits

are exceeded

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• High risk CSPs– Batches of more than 25

individual CSPs– Any high risk CSP that has

been exposed longer than 12hours at 2º to 8º or longerthan 6 hours at warmer than8º

• All risk level CSPs– When USP <797> BUD limits

are exceeded

Page 34: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Sterility Testing

• USP <71> Sterility Tests states:– “These Pharmacopeial procedures are not by themselves designed to

ensure that a batch of product is sterile or has been sterilized. This isaccomplished primarily by method suitability of the sterilizationprocess or of the aseptic processing procedure.”

– Is an industrial test used to detect incidents of gross contamination.

– It should be recognized that the USP sterility test might not detectmicrobial contamination if present in only a small percentage of thefinished articles in the batch.

– Current knowledge offers no nondestructive alternatives forascertaining the microbiological quality of every finished article in thelot.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• USP <71> Sterility Tests states:– “These Pharmacopeial procedures are not by themselves designed to

ensure that a batch of product is sterile or has been sterilized. This isaccomplished primarily by method suitability of the sterilizationprocess or of the aseptic processing procedure.”

– Is an industrial test used to detect incidents of gross contamination.

– It should be recognized that the USP sterility test might not detectmicrobial contamination if present in only a small percentage of thefinished articles in the batch.

– Current knowledge offers no nondestructive alternatives forascertaining the microbiological quality of every finished article in thelot.

Page 35: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Sterility Testing

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

Cundell AM. “Review of the Media Selection and Incubation Conditions for the Compendial Sterility and Microbial LimitTests,” Pharm.Forum 28 (6), 2034–2041.

Page 36: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Sterility Testing (Membrane Filtration)

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

Millipore Equinox Steritest System

Page 37: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Flow of the Sterility Test1. Media and Bacteriostasis/Fungistasis Testing (Method Suitability)2. Eliminate any bacteriostatis/fungistatic properties3. Determine number of articles, quantity from each, to test4. Incubate the samples5. Examine test articles for signs of growth6. Examine suspect tubes microscopically for signs of growth7. Subculture if necessary8. Write the report

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

1. Media and Bacteriostasis/Fungistasis Testing (Method Suitability)2. Eliminate any bacteriostatis/fungistatic properties3. Determine number of articles, quantity from each, to test4. Incubate the samples5. Examine test articles for signs of growth6. Examine suspect tubes microscopically for signs of growth7. Subculture if necessary8. Write the report

Reminder: All Compendial Microbiological Test Methods, including Sterility Tests, areClassical Growth Based Methods

Page 38: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Polling Question

The preferred method of sterility testing is which of thefollowing?

1. Direct inoculation2. Filter integrity testing3. Membrane filtration4. Something else

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

1. Direct inoculation2. Filter integrity testing3. Membrane filtration4. Something else

20%

Kate
Highlight
Page 39: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Sterility Testing

• “Passing” a sterility test does not guarantee that every unit in thatbatch is sterile.

• The use of two types of medias is required.– TSB and FTM

• Two incubations are required for the medias• Membrane filtration is the preferred method of sterility testing• BUDs are not universal and must be verified by each vendor• Must be based on sterility testing according to USP 71 or other

procedures, methods or processes that have been proven to beequivalent or superior with statistical significance

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• “Passing” a sterility test does not guarantee that every unit in thatbatch is sterile.

• The use of two types of medias is required.– TSB and FTM

• Two incubations are required for the medias• Membrane filtration is the preferred method of sterility testing• BUDs are not universal and must be verified by each vendor• Must be based on sterility testing according to USP 71 or other

procedures, methods or processes that have been proven to beequivalent or superior with statistical significance

Challenge external testing labs and vendors on how they accept samplesless than the quantities prescribed in USP 71, Table 3.

Challenge external testing labs and vendors on how they accept samplesless than the quantities prescribed in USP 71, Table 3.

Page 40: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Number of Articles to be Tested in Relation to theNumber of Articles in the Batch (From USP<71>)

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

Page 41: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Minimum Quantity to be Used for Each Medium(from USP <71>)

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

Page 42: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Sterility Testing Quick ReferenceKey Element Description

Correct # units Per Chapter <71>, Table 3

Correct volume /unit Per Chapter <71>, Table 2

Correct method • Membrane Filtration: pool all samples and run throughsingle filter

• Must have justification if using Direct Inoculation: 1:1 (unittested and broth used)

• Other methods (not in <71>) IF verification demonstratesequivalence to USP <71>

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Membrane Filtration: pool all samples and run throughsingle filter

• Must have justification if using Direct Inoculation: 1:1 (unittested and broth used)

• Other methods (not in <71>) IF verification demonstratesequivalence to USP <71>

Method Suitability Testingperformed

Determines if the sterility testing method is valid for aparticular type of CSP and that the drug does not interferewith the sterility test method

Correct Media Used • Fluid Thioglycollate Media (FTM) Incubated for 14 days (20-25°C) anaerobic and aerobic bacteria

• Soybean Casein Digest Media (SCDM) Incubated for 14 days(30-35°C) for both aerobic bacteria and fungi

Page 43: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Inspector Evaluation Metrics: Sterility Testing

• Does the pharmacy have detailedpolicies, procedures, forms orelectronic methods ofdocumentation about all aspects ofsterility testing?

• Does the pharmacy comply withUSP Chapter <71>?

• Does the compounding locationhave a written procedure forimmediate recall of the dispensedCSPs in the event of any evidenceof microbial growth in the testspecimens?

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Does the pharmacy have detailedpolicies, procedures, forms orelectronic methods ofdocumentation about all aspects ofsterility testing?

• Does the pharmacy comply withUSP Chapter <71>?

• Does the compounding locationhave a written procedure forimmediate recall of the dispensedCSPs in the event of any evidenceof microbial growth in the testspecimens?

Page 44: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Inspector Evaluation Metrics: Sterility Testing

• CSPs are not dispensed prior to thetime their final sterility test resultsare available.

• Does the compounding locationhave a written procedure requiringdaily observation of the incubatingsterility test specimens (if the CSPis dispensed prior to the finalsterility test results?

• Does the pharmacy have aprocedure for the immediate recallof the dispensed CSPs in the eventof any evidence of microbialcontamination?

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• CSPs are not dispensed prior to thetime their final sterility test resultsare available.

• Does the compounding locationhave a written procedure requiringdaily observation of the incubatingsterility test specimens (if the CSPis dispensed prior to the finalsterility test results?

• Does the pharmacy have aprocedure for the immediate recallof the dispensed CSPs in the eventof any evidence of microbialcontamination?

Page 45: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Polling Question

The two media used for sterility testing are:

1. FTM & LAL2. LAL & TSB3. TSB & FTM4. TSA & TSB

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

1. FTM & LAL2. LAL & TSB3. TSB & FTM4. TSA & TSB

Kate
Highlight
Page 46: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Critical Concepts in Pyrogen Testing

• Endotoxins produced by gram-negative bacteria and foundwithin the cell membraneaccount for an estimated 99%of the pyrogens found on thesurface of medical devices

• The term "pyrogen," literally"heat generating," refers toany substance—microbial orotherwise—which wouldinduce a temperature risewhen introduced into a patient

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Endotoxins produced by gram-negative bacteria and foundwithin the cell membraneaccount for an estimated 99%of the pyrogens found on thesurface of medical devices

• The term "pyrogen," literally"heat generating," refers toany substance—microbial orotherwise—which wouldinduce a temperature risewhen introduced into a patient

Page 47: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Why is pyrogen testing important?

• Endotoxins are produced by both gram-positive and -negativebacteria and fungi.– Endotoxins from gram-negative bacteria are more potent and

represent a serious threat to patient safety and well-being.

• Intrathecal CSPs contaminated with endotoxins have beenimplicated in cases of both septic and aseptic meningitis, inaddition to shock and death.– The intrathecal space does not have the same immunological and

biological defense mechanisms as the intravenous system.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Endotoxins are produced by both gram-positive and -negativebacteria and fungi.– Endotoxins from gram-negative bacteria are more potent and

represent a serious threat to patient safety and well-being.

• Intrathecal CSPs contaminated with endotoxins have beenimplicated in cases of both septic and aseptic meningitis, inaddition to shock and death.– The intrathecal space does not have the same immunological and

biological defense mechanisms as the intravenous system.

Page 48: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Why is pyrogen testing important?

• The official endotoxin limitsare 5 EU/kg/hr or 350 EU/totalbody/hour for drugs andbiologicals

• Drugs for intrathecaladministration have a muchlower endotoxin limit:– 0.2 EU/kg/hr

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• The official endotoxin limitsare 5 EU/kg/hr or 350 EU/totalbody/hour for drugs andbiologicals

• Drugs for intrathecaladministration have a muchlower endotoxin limit:– 0.2 EU/kg/hr

Image: Retrieved from Medscape on October 3, 2013.

Official endotoxin limits wereestablished by the FDA

Official endotoxin limits wereestablished by the FDA

Page 49: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Sources of pyrogens• The choice of components used in CSPs is a critical step, especially

during preparing intrathecal injections.• Commercially available components from manufacturers are

typically sterile and pyrogen-free.– Nonsterile components used in high-risk level compounding may not

pyrogen-free.

• Water is the main source of pyrogens.• Endotoxins are a type of pyrogen and are stable over long periods

of time.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• The choice of components used in CSPs is a critical step, especiallyduring preparing intrathecal injections.

• Commercially available components from manufacturers aretypically sterile and pyrogen-free.– Nonsterile components used in high-risk level compounding may not

pyrogen-free.

• Water is the main source of pyrogens.• Endotoxins are a type of pyrogen and are stable over long periods

of time.

Page 50: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Sources of pyrogens (continued)

• Glassware, devices, or utensils that have been washed with tapwater and then used to compound high-risk level CSPs can serve asa reservoir of endotoxins.– These pyrogens can be transferred to solutions prepared in beakers or

utensils during the compounding process.

• Filtration using typical 0.22 micron filters or moist-heatsterilization will not eliminate endotoxins after introduction into asolution.– Pall Posidyne® filter is a 0.2 micron positively charged nylon filter

membrane that can retain endotoxins in solution.

• Properly depyrogenating glassware and utensils by using a dry-heat oven can eliminate the introduction of pyrogens.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Glassware, devices, or utensils that have been washed with tapwater and then used to compound high-risk level CSPs can serve asa reservoir of endotoxins.– These pyrogens can be transferred to solutions prepared in beakers or

utensils during the compounding process.

• Filtration using typical 0.22 micron filters or moist-heatsterilization will not eliminate endotoxins after introduction into asolution.– Pall Posidyne® filter is a 0.2 micron positively charged nylon filter

membrane that can retain endotoxins in solution.

• Properly depyrogenating glassware and utensils by using a dry-heat oven can eliminate the introduction of pyrogens.

Page 51: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Bacterial Endotoxin Test (BET)

• Historically, pyrogen testing ofdrugs was performed usingrabbits.

– This method took longer andwas less sensitive.

• This method was replaced withthe LAL method.

– LAL is an acronym for LimulusAmebocyte Lysate which isbased on clotting properties ofthe horseshoe crab’s blood.

• BET or LAL is a test method forestimating the concentration ofbacterial endotoxins.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Historically, pyrogen testing ofdrugs was performed usingrabbits.

– This method took longer andwas less sensitive.

• This method was replaced withthe LAL method.

– LAL is an acronym for LimulusAmebocyte Lysate which isbased on clotting properties ofthe horseshoe crab’s blood.

• BET or LAL is a test method forestimating the concentration ofbacterial endotoxins.

Page 52: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

BET (continued)

• There are several methods ofperforming this test:– Gel-clot technique: gel-clot

formed in presence ofendotoxin.

– Photometric• Turbidimetric method:

presence of turbidity isobserved

• Chromogenic method:color change is observed

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• There are several methods ofperforming this test:– Gel-clot technique: gel-clot

formed in presence ofendotoxin.

– Photometric• Turbidimetric method:

presence of turbidity isobserved

• Chromogenic method:color change is observed

Page 53: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

BET Summary

• BET appears simple, but it is a complex procedure and requiresmethod validation for the individual drug being tested.

• BET has several requirements in order to be a validated test. Theyare:– All testing solutions have a neutral pH

– All testing solutions have a low salt or solute concentration to allowan optimal reaction between the LAL reagent and endotoxins.

• Compounded drugs usually require significant dilution to preventinterference and risk of false-negative results.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• BET appears simple, but it is a complex procedure and requiresmethod validation for the individual drug being tested.

• BET has several requirements in order to be a validated test. Theyare:– All testing solutions have a neutral pH

– All testing solutions have a low salt or solute concentration to allowan optimal reaction between the LAL reagent and endotoxins.

• Compounded drugs usually require significant dilution to preventinterference and risk of false-negative results.

Page 54: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Inspector Evaluation Metrics: Bacterial Endotoxin

• Is bacterial endotoxin (pyrogen)testing on all high risk level CSPs,(except those for inhalation andophthalmic administration)performed?

• Does the pharmacy have a specificpolicy and procedure for bacterialendotoxin testing?

– Does it includes the description ofthe procedure?

– Have specific endotoxin unit limitsbased on USP Endotoxin Test < 85>been developed?

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Is bacterial endotoxin (pyrogen)testing on all high risk level CSPs,(except those for inhalation andophthalmic administration)performed?

• Does the pharmacy have a specificpolicy and procedure for bacterialendotoxin testing?

– Does it includes the description ofthe procedure?

– Have specific endotoxin unit limitsbased on USP Endotoxin Test < 85>been developed?

Page 55: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Polling Question

The reagent for the Bacterial Endotoxins test is which of thefollowing?

1. FTM2. LAL3. TSB4. TSA

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

1. FTM2. LAL3. TSB4. TSA

Kate
Highlight
Page 56: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Quality Release Checks And Tests: Documentation

• Does the pharmacy have awritten release checks and testsPnP used to evaluate?

• Were the correct ingredientsused?

• Were the weights and measurescorrect?

• Verified by review of batchdocumentation

• Signatures or initials ofcompounders can be traced backto signature log

• Date of preparation andspecimen label retained andattached to compoundingdocumentation

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Does the pharmacy have awritten release checks and testsPnP used to evaluate?

• Were the correct ingredientsused?

• Were the weights and measurescorrect?

• Verified by review of batchdocumentation

• Signatures or initials ofcompounders can be traced backto signature log

• Date of preparation andspecimen label retained andattached to compoundingdocumentation

Page 57: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Quality Release Checks and Tests

• Physical inspection provides very basic information about the CSP.• Primary concern is that the CSP contains the ingredients specified

in the original prescription.• The final preparation still should be evaluated for:

– Container leaks and integrity: physical examination of the finalprepared container visually inspected for leaks, holes or othercontainer-closure breaches;

– Particulates in solution: physical examination of the solution for thepresence of mobile, randomly sourced, extraneous substances otherthan gas;

– Solution color, volume, and odor (if possible) and

– Phase separation (oiling, creaming, or cracking)

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Physical inspection provides very basic information about the CSP.• Primary concern is that the CSP contains the ingredients specified

in the original prescription.• The final preparation still should be evaluated for:

– Container leaks and integrity: physical examination of the finalprepared container visually inspected for leaks, holes or othercontainer-closure breaches;

– Particulates in solution: physical examination of the solution for thepresence of mobile, randomly sourced, extraneous substances otherthan gas;

– Solution color, volume, and odor (if possible) and

– Phase separation (oiling, creaming, or cracking)

Page 58: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Quality Release Testing: Equipment

• How does the pharmacy ensure thatequipment, apparatus, and devicesused to compound a CSP beconsistently capable of operatingproperly and within acceptabletolerance limits?

• Are written procedures outliningrequired equipment calibration, annualmaintenance, monitoring for properfunction, and controlled procedures foruse of the equipment and specifiedtime frames for these activities areestablished and followed?

– Routine maintenance andfrequencies shall be outlined in theseSOPs.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• How does the pharmacy ensure thatequipment, apparatus, and devicesused to compound a CSP beconsistently capable of operatingproperly and within acceptabletolerance limits?

• Are written procedures outliningrequired equipment calibration, annualmaintenance, monitoring for properfunction, and controlled procedures foruse of the equipment and specifiedtime frames for these activities areestablished and followed?

– Routine maintenance andfrequencies shall be outlined in theseSOPs.

Page 59: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Quality Release Testing: Equipment

• Are results from the equipment calibration, annual maintenancereports, and routine maintenance are kept on file for the lifetimeof the equipment?

• Are personnel prepared through an appropriate combination ofspecific training and experience to operate or manipulate anypiece of equipment, apparatus, or device they may use whenpreparing CSPs?

– Training includes gaining the ability to determine whether any item ofequipment is operating properly or is malfunctioning.

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Are results from the equipment calibration, annual maintenancereports, and routine maintenance are kept on file for the lifetimeof the equipment?

• Are personnel prepared through an appropriate combination ofspecific training and experience to operate or manipulate anypiece of equipment, apparatus, or device they may use whenpreparing CSPs?

– Training includes gaining the ability to determine whether any item ofequipment is operating properly or is malfunctioning.

Page 60: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Inspector Evaluation Metrics: Quality Release

• Does the pharmacy a formal writtenpolicy for release checks and tests forCSPs?

• Is there evidence that CSPs areinspected for particulate matter?

• Is there evidence that the pharmacycalibrates and maintains theirequipment in good working condition?

• Is there evidence that staff areappropriately trained on the use andtroubleshooting of equipment?

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Does the pharmacy a formal writtenpolicy for release checks and tests forCSPs?

• Is there evidence that CSPs areinspected for particulate matter?

• Is there evidence that the pharmacycalibrates and maintains theirequipment in good working condition?

• Is there evidence that staff areappropriately trained on the use andtroubleshooting of equipment?

Page 61: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Quality Management System

• Does the pharmacy have a formaland written QualityAssurance/PerformanceImprovement Plan for all aspectsof compounding activities?

– includes specific monitoringand evaluation activities; detailson how results are reported;and delineation of the personsresponsible.

• Are Adverse events and defectsrelated to CSPs are reportedthrough the FDA’s MedWatchand/or Quantros programs?

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Does the pharmacy have a formaland written QualityAssurance/PerformanceImprovement Plan for all aspectsof compounding activities?

– includes specific monitoringand evaluation activities; detailson how results are reported;and delineation of the personsresponsible.

• Are Adverse events and defectsrelated to CSPs are reportedthrough the FDA’s MedWatchand/or Quantros programs?

Page 62: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Quality Management System

• Does the pharmacy have a consistent process to document, trend,and ascertain the effectiveness of a corrective action plan toresolve issues when:– a desired patient outcome is not achieved;

– an Action Limit is exceeded;

– or other operational variance is noted?

• Examples: CAPA, QRE/Incident Report– Corrective and Preventative Action

• Does the compounding location have a written procedurerequiring notification of the physician and patient to whom apotentially contaminated CSP was administered?

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Does the pharmacy have a consistent process to document, trend,and ascertain the effectiveness of a corrective action plan toresolve issues when:– a desired patient outcome is not achieved;

– an Action Limit is exceeded;

– or other operational variance is noted?

• Examples: CAPA, QRE/Incident Report– Corrective and Preventative Action

• Does the compounding location have a written procedurerequiring notification of the physician and patient to whom apotentially contaminated CSP was administered?

Page 63: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Example of CAPA Form

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

Page 64: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Inspector Evaluation Metrics: Quality Management

• Does the pharmacy a formaland written quality assuranceplan?

• Does the pharmacy documentpatient complaints?

• Does the pharmacy respondappropriately to complaintsand is there evidence that aninvestigation was conductedand resolution documented?

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Does the pharmacy a formaland written quality assuranceplan?

• Does the pharmacy documentpatient complaints?

• Does the pharmacy respondappropriately to complaintsand is there evidence that aninvestigation was conductedand resolution documented?

Page 65: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Patient Counseling

• Does the pharmacy have a training program includes didactic andpractical hands on training of patients and caregivers with successfulreturn demonstration of the required skills for administration of CSPs?

• The specific required competencies are formally assessed andpatients/caregivers must demonstrate competency before they areallowed to administer CSPs independent of supervision by healthcareprofessionals?

• Does the pharmacy have a formal training program that ensures thatpatients/caregivers achieve the required competencies for safe storage,handling, administration, monitoring, emergency actions and disposal ofCSPs?

• Does the pharmacy provide written materials to patients/caregivers tosupplement verbal training?

• Does the pharmacy have procedures/system in place for patients andcaregivers to use to report questions or concerns relative to CSPs theyreceive from the compounding location?

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Does the pharmacy have a training program includes didactic andpractical hands on training of patients and caregivers with successfulreturn demonstration of the required skills for administration of CSPs?

• The specific required competencies are formally assessed andpatients/caregivers must demonstrate competency before they areallowed to administer CSPs independent of supervision by healthcareprofessionals?

• Does the pharmacy have a formal training program that ensures thatpatients/caregivers achieve the required competencies for safe storage,handling, administration, monitoring, emergency actions and disposal ofCSPs?

• Does the pharmacy provide written materials to patients/caregivers tosupplement verbal training?

• Does the pharmacy have procedures/system in place for patients andcaregivers to use to report questions or concerns relative to CSPs theyreceive from the compounding location?

Page 66: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

Redispensing CSPs, Storage and Transport

• Common practice in hospitals (also known as recycling)

• Originally assigned BUD dates are not changed and CSPs are notredispensed unless changes are supported by the originally assignedBUD.

• Is there absolute certainty that proper storage conditions have beenmaintained during the time they were outside of pharmacy’s control?

• Is there evidence of a systematic process to evaluate when unopened,returned CSPs are safe to redispense?

• Is there evidence that packaging maintains physical integrity, sterility,stability and purity of CSPs?

• Is there evidence that the storage of finished CSPs and drug componentsis separate from food storage and from any specimen storage (if occursonsite)?

• Does the pharmacy have evidence that the methods used to transportCSPs to the patient prevent damage and maintain appropriatetemperatures during transit?

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

• Common practice in hospitals (also known as recycling)

• Originally assigned BUD dates are not changed and CSPs are notredispensed unless changes are supported by the originally assignedBUD.

• Is there absolute certainty that proper storage conditions have beenmaintained during the time they were outside of pharmacy’s control?

• Is there evidence of a systematic process to evaluate when unopened,returned CSPs are safe to redispense?

• Is there evidence that packaging maintains physical integrity, sterility,stability and purity of CSPs?

• Is there evidence that the storage of finished CSPs and drug componentsis separate from food storage and from any specimen storage (if occursonsite)?

• Does the pharmacy have evidence that the methods used to transportCSPs to the patient prevent damage and maintain appropriatetemperatures during transit?

Page 67: Sterilization and Quality Assurance Procedures and Quality... · – (no nsterile sterile) • Methods of sterilization – Autoclave or Dry-Heat oven sterilization cycles • Autoclave

“Absence of Evidence Does Not EqualEvidence of Absence”

Dr. David Hussong(FDA and the USP Microbiology and Sterility

Assurance Expert Committee)

Copyright © 2008-2014 ClinicalIQ, LLC® - All Rights Reserved

“Absence of Evidence Does Not EqualEvidence of Absence”

Dr. David Hussong(FDA and the USP Microbiology and Sterility

Assurance Expert Committee)