2008/2/27 1
FDA/INDUSTRY STATISTICS WORKSHOP: FDA/INDUSTRY STATISTICS WORKSHOP: Washington, D.C. Sept. 29, 2006Washington, D.C. Sept. 29, 2006
Statistical Issues in Medical Device Statistical Issues in Medical Device TrialsTrials
George Koustenis, FDA-CDRHDavid Breiter, Boston ScientificRoseann White, Abbott VascularGeorge Woodworth, Univ. of Iowa
![Page 2: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/2.jpg)
2008/2/27 2
The Use of Objective Performance Criteria The Use of Objective Performance Criteria (OPC) in Medical Device Evaluation(OPC) in Medical Device Evaluation
FDA/Industry Statistics WorkshopSeptember 29, 2006Washington, D.C.
George KoustenisDivision of Biostatistics
Office of Surveillance and BiometricsCenter for Devices and Radiological Health
U.S. Food and Drug Administration
![Page 3: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/3.jpg)
2008/2/27 3
OBJECTIVEOBJECTIVE
The purpose of this presentation is to provide a brief introduction to the possible use of Objective Performance Criteria (OPC) in the evaluation of medical devices during the regulatory approval process.
![Page 4: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/4.jpg)
2008/2/27 4
DEFINITION OF OPCDEFINITION OF OPC
Fixed Target(s)Objective and Meaningful StandardProvides Comparison in Evaluating Safety and EffectivenessUsually a RateSurrogate for Control GroupBenchmark for Minimally Acceptable ValuesNot a Control Group
![Page 5: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/5.jpg)
2008/2/27 5
CONTROLS CONTROLS
Standard Comparison for an Experimental TreatmentGroup of Patients with Same Condition, Demographics, & Prognostic ValuesBy Controlling for as Many Variables as Possible – Any Differences are Presumed Due to New Intervention
![Page 6: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/6.jpg)
2008/2/27 6
TYPES OF CONTROLSTYPES OF CONTROLS
Concurrent Randomized Controlled Trial (RCT)The Gold StandardEffectively Minimizes BiasEffectively Balances Demographics & UnknownsSupports Basic Assumptions of Standard Statistical Methodology
![Page 7: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/7.jpg)
2008/2/27 7
Historical ControlsHistorical Controls
Compares Current Therapy & Patients Against Others Studied in Previous InvestigationsNot Randomized to Current ConditionsMay Be Too Far Removed in TimeVery Difficult to Validate the Data from Historical TrialsTend to Produce More Positive Results
![Page 8: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/8.jpg)
2008/2/27 8
ESTIMATES OF OPCESTIMATES OF OPC’’SS
Necessarily Driven by Historical DataRequires Appropriate Pooling of Different InvestigationsOPC’s Inherit All of Problems Seen with Historical ControlsEven Sophisticated Pooling/Analytical Techniques Cannot Eliminate Basic Problems
![Page 9: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/9.jpg)
2008/2/27 9
USE OF OPCUSE OF OPC’’s In CDRHs In CDRH
Some Current Use of OPC’sCardiac Ablation CathetersReplacement Heart ValvesOphthalmicsHip Replacement Systems
![Page 10: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/10.jpg)
2008/2/27 10
USE OF OPCUSE OF OPC’’s In DEVICE s In DEVICE TRIALSTRIALS
Given that the use of OPC represents a significant departure from the standard scientific approach to the design and analysis of medical device clinical trials, the question now becomes: when might it be appropriate to use this type of non-control comparison in the medical device approval process?
![Page 11: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/11.jpg)
2008/2/27 11
CITED ADVANTAGES Of OPCCITED ADVANTAGES Of OPC’’ss
Smaller Sample SizeStandard Value for All SponsorsSave Time and MoneyEasier to ExecuteI.E. – “Least Burdensome”
![Page 12: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/12.jpg)
2008/2/27 12
DISADVANTAGES Of OPCDISADVANTAGES Of OPC’’ss
All the Problems Associated with Historical ControlsProblems with Validity of Data & AnalysisProblems with Advances in Practice of MedicineMay be Disagreement on Final OPC ValueProblems with Single Arm TrialsSelection BiasMay Not See Pre-Post Benefits
![Page 13: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/13.jpg)
2008/2/27 13
DISADVANTAGES Of OPCDISADVANTAGES Of OPC’’ss
May Not Be Least BurdensomeSmaller Sample Size? N=100 or 150?Time and Resource Intensive to DevelopSets a Minimum StandardSuperiority?Older and Older DataProtect and Promote the Public Health?
![Page 14: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/14.jpg)
2008/2/27 14
WHEN MIGHT OPCWHEN MIGHT OPC’’s Be USED?s Be USED?
History of OPC’s in Medical Devices
Non-Experimental or Quasi-Experimental Designs May be Valid in Certain Situations
![Page 15: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/15.jpg)
2008/2/27 15
MINIMUM REQUIREMENTS MINIMUM REQUIREMENTS WHEN OPCWHEN OPC’’s MIGHT Be USED s MIGHT Be USED
Great Deal is Known About the Natural History of the Disease or ConditionUnderlying Patient Population is Well Described & Relatively StableExtensive Clinical History & Experience with This Device
![Page 16: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/16.jpg)
2008/2/27 16
MINIMUM REQUIREMENTS MINIMUM REQUIREMENTS WHEN OPCWHEN OPC’’s MIGHT Be USED s MIGHT Be USED (cont.)(cont.)
Stable and Well Known Standard of CareAppropriate Ancillary Technology is Relatively StableNo Significant New Questions of Safety or Effectiveness Consensus Among FDA, Industry, Clinical, Academic and Patient CommunitiesExpectation of Significantly Positive Treatment Effect
![Page 17: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/17.jpg)
2008/2/27 17
HOW SHOULD OPCHOW SHOULD OPC’’s BE s BE DERIVED? DERIVED?
Data DrivenRigorous And Scientifically Valid Methodologies Valid DatabasesAppropriate Statistical ModelingAppropriately Designed and Powered Pivotal Trial Periodical Re-Evaluation and Updating the OPC’s
![Page 18: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/18.jpg)
2008/2/27 18
OTHER IMPORTANT OPC OTHER IMPORTANT OPC ISSUESISSUES
All Parties Must Have a Clear Understanding of the OPC in Medical Device Trials – Success and FailureOPC’s Already in Limited Use in Device
TrialsNo Current General Policy of OPC Use in CDRHNeed for Comprehensive, Coherent, and Consistent OPC Policy for Device Trials
![Page 19: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/19.jpg)
2008/2/27 19
WHAT SHOULD THIS POLICY WHAT SHOULD THIS POLICY CONTAIN?CONTAIN?
Clear Definitions of All Appropriate TermsEstablished Guidance on Minimum Standards for a Device or Device ClassSpecific Roles Played by FDA, Industry, Clinical Community, Academia and Other Interested PartiesProvisions for Periodic Updating of OPCSpecific Guidance on Methodology to Derive an OPCUnambiguous Policy Regarding Failure to Meet OPC
![Page 20: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/20.jpg)
2008/2/27 20
SUMMARYSUMMARY
An introduction to the history and use of Objective Performance Criteria (OPC) in medical device evaluationAdvantages and disadvantages of OPC’sWhen and how OPC might be used & developedDiscussed the need for developing a CDRH policy regarding use of OPC in medical device trialsGeneral outline of this policy is presented in an effort to begin dialogue between CDRH, industry, clinical community & other interested parties
![Page 21: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/21.jpg)
2008/2/27 21
REFERENCES REFERENCES
Please see accompanying brief commentary
Request electronic copy from [email protected]
![Page 22: Statistical Issues In Medical Device Trials](https://reader034.fdocuments.us/reader034/viewer/2022051818/54982bf4b47959654d8b5352/html5/thumbnails/22.jpg)
2008/2/27 22
Q&AQ&A