Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

27
Effect of True Area Sample Size on Variability in Endothelial Cell Density (ECD) Results Jackie Hai and Vivian Xue HAI Laboratories, Inc.

description

Presentation of our abstract included in the Cornea Society / EBAA 2010 Fall Educational Symposium. Complete set of notes here: http://www.hailabs.com/2010/10/specular-microscopy-series-effect-of-true-area-sample-size-on-variability-in-endothelial-cell-density/

Transcript of Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Page 1: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Effect of True Area Sample Size on Variability in Endothelial Cell Density

(ECD) Results

Jackie Hai and Vivian XueHAI Laboratories, Inc.

Page 2: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Purpose

Evaluate the effect of sample area size on ECD measurement results and variability

Determine a recommended minimum sample size by comparing the results of cell counts performed within larger areas vs smaller areas

Page 3: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Sample Selection Criteria True area required for all samples

Images captured at high resolution

Low instances of pathology or abnormality

Random selection with variety of ECD, polymegathism, pleomorphism within normal range

Page 4: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Untrue Area: excessive concave curvature (circled)

Page 5: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Untrue Area: incorrect focal depth (cells out of focus)

Page 6: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

True Area: flat, focused image with clear cell borderlines

Page 7: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

True Area: flat, focused image with clear cell borderlines

Page 8: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Low Resolution Thumbnails(160x120 pixels)

Page 9: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Sample Area: 39981.10μm2

Cells Counted: 113Density: 2826

High Resolution Analysis(640x480 pixels)

Page 10: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Sample Area: 44843.25μm2

Cells Counted: 113Density: 2520

High Resolution Analysis(640x480 pixels)

Page 11: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Sample Selection Criteria True area required for all samples

Images captured at high resolution

Low instances of pathology or abnormality

Random selection with variety of ECD, polymegathism, pleomorphism within normal range

Page 12: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Counting Method

Variable frame counting method

Single-field (40,000μm2 – 60,000μm2)

Multiple-field (20,000μm2 – 25,000μm2)

Page 13: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Single-field variable frame count (60,000μm2)

HAI CAS / EB 2.10

Page 14: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Multiple-field variable frame count 1 of 3 (23,000μm2)

HAI CAS / EB 2.10

Page 15: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Multiple-field variable frame count 2 of 3 (22,000μm2)

HAI CAS / EB 2.10

Page 16: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Multiple-field variable frame count 3 of 3 (22,500μm2)

HAI CAS / EB 2.10

Page 17: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Results

ECD ranged from 1700 – 3500 cells/mm2

Result groupings: Class A (large samples): 40,000μm2 - 70,000μm2 Class B (small samples): 20,000μm2 - 25,000μm2

Standard Deviation: Class A (large samples): 6.27-63.11 (mean 26.3) Class B (small samples): 17.22-155.32 (mean 75.0)

Page 18: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability
Page 19: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Results

Standard Error (S.D. / √N): Class A (large samples): 3.14-31.56 (mean 13.2) Class B (small samples): 9.94-89.67 (mean 43.3)

Precision (S.E. / Mean): Class A (large samples): 0.0052 Class B (small samples): 0.0156

Page 20: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability
Page 21: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Results

In every instance: Class A Class B(40,000–70,000μm2) < (20,000–25,000μm2) Mean ECD Mean ECD

RPD ranged from 3.7% – 13.2%

Mean RPD of 7.9%

Page 22: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Conclusion

Larger sample sizes (>40,000μm2) yielded lower standard error and more precise cell count results

Cell counts performed in smaller sample sizes produce significantly higher ECD than larger sample sizes

Page 23: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Discussion

Rate of Change 1,000,000__ in Cell Density = Sample Area(cells per mm2) (in μm2)

Page 24: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability
Page 25: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Discussion

Sample size was based on area

Samples based solely on # of cells yield smaller sample size in high-density cornea

In smaller samples, ECD increases at a higher rate for every extra cell counted

Page 26: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Discussion

Smaller sample sizes compound possibility for increased margin of error Multiple small samples have longer total

perimeter than a single large sample

Human error factors Image quality Cell identification Tracing of cell boundaries

Page 27: Specular Microscopy Methodology: Effect of Sample Area Size on ECD Variability

Recommendations1. Capture images at high resolution and

maximum screen size (640 x 480 pixels).

2. Select the largest sample area possible on the screen for cell counting.

3. If there are folds or distortions, select the largest contiguous true area possible.

4. For high resolution images free of distortions, a minimum selection of 50,000μm2 should always be possible.