Resolving Volume-Restricted...
Transcript of Resolving Volume-Restricted...
COLUMN WATCH
Total peak shape analysis
SAMPLE PREPARATION
PERSPECTIVES
Extraction technologies in food
safety studies
QUESTIONS OF QUALITY
Data integrity metrics for
chromatography
CE
LEBRATING OUR 30th YEA
RCEL
EB
RATING OUR 30th YEAR
��
1987-20171987-2017
��
��
����
��
��
December 2017 Volume 30 Number 12
www.chromatographyonline.com
The benefits of CE–MS for analyzing limited amounts of polar
and charged biological samples
Resolving Volume-Restricted Metabolomics
Q
D
c
www.restek.com/rxi Pure Chromatography
Leverage the outstanding inertness, low bleed, and high
reproducibility of Rxi® 3-in-1 technology to gain:
• Accurate Data
• The Right Results Fast
• Maximized Instrument Uptime
Put Rxi® columns to work in your lab.
Visit www.restek.com/rxi and order yours today.
Rxi® columns come
with an unbeatable
guarantee.
That is Restek®
Pure Satisfaction.
wyatt.com
© 2017 Wyatt Technology and Leo Cullum from cartoonbank.com. All rights reserved. All trademarks and registered trademarks are properties of their respective holders.
That’s the problem with relying on the assumptions of column calibration to characterize macromolecules.
You don’t really know if your standards and your samples are the same! Perhaps that’s why every major
pharmaceutical and biotechnology company, as well as most federal regulatory agencies are switching from
relative methods to Wyatt Technology’s absolute measurements. We’ll show you how to end your dependence
on reference standards, in order to make you and
your products a whole lot more convincing. See our
webinar on Absolute Biophysical Characterization
at wyatt.com/Absolute.
LC•GC Europe December 2017654
Editorial Policy:
All articles submitted to LC•GC Europe
are subject to a peer-review process in association
with the magazine’s Editorial Advisory Board.
Cover:
Original materials courtesy: Nik Merkulov/
Shutterstock.com
Columns662 SAMPLE PREPARATION PERSPECTIVES
The Use of Extraction Technologies in Food Safety
Studies
Changling Qiu and Doug Raynie
Traditional extraction methods for food samples, such as liquid–
liquid extraction and Soxhlet extraction, are often time-consuming
and require large amounts of organic solvents. Therefore, one of
the objectives of analytical food safety studies currently has been
the development of new extraction techniques that can improve
the accuracy and precision of analytical results and simplify the
analytical procedure.
670 COLUMN WATCH
Peak Shapes and Their Measurements: The Need and the
Concept Behind Total Peak Shape Analysis
M. Farooq Wahab, Darshan C. Patel, and Daniel W. Armstrong
A new “total peak shape analysis” approach is suggested that
facilitates detection and quantifi cation of concurrent fronting and
tailing in peaks. Several remediation approaches are proposed to
chromatographers, which can help them analyze and improve the
peak shapes.
679 QUESTIONS OF QUALITY
Data Integrity Metrics for Chromatography
Mark E. Newton and R.D. McDowall
This article discusses metrics for monitoring data integrity within
a chromatography laboratory, from the regulatory requirements to
practical implementation.
Departments686 Products
688 Events
689 The Applications Book
COVER STORY
658 Resolving Volume-Restricted Metabolomics Using
Sheathless Capillary Electrophoresis–Mass Spectrometry
Rawi Ramautar
Advances that signifi cantly
improved the performance of
capillary electrophoresis–mass
spectrometry for in-depth
metabolic profi ling of limited
sample amounts are considered.
Attention is also devoted to
various technical aspects that
still need to be addressed to
make capillary electrophoresis–
mass spectrometry a viable
approach for volume-restricted
metabolomics.
December | 2017
Volume 30 Number 12
.
A major challenge in bioanalysis and metabolomics is the
profiling of (endogenous) metabolites in volume-restricted
and often scarce biological samples. Though the analytical
techniques currently used in metabolomics are powerful, the
need for relatively large amounts of sample for pre-analytics
and injection prevent their use in many biomedical and clinical
applications. For example, metabolic profiling of small-volume
biological samples, such as cerebrospinal fluid (CSF) from
mouse models, spheroids or microtissues, liquid biopsies,
and samples from microfluidic organ-on-a-chip systems,
is seriously hindered by the current analytical technologies
because of the limited sample material. The standard
analytical techniques also do not allow a maximum amount of
biochemical information from valuable and scarcely available
biological samples to be obtained because the material is
completely consumed for a single metabolomics measurement.
Despite important developments in analytical separations
technology over recent years, the analysis of small-volume
biological samples remains a challenging task. Therefore, there
is a critical need for the introduction of analytical methods to
allow highly sensitive metabolic profiling of volume-restricted or
mass-limited biological samples.
Capillary electrophoresis–mass spectrometry (CE–MS) can
be considered an attractive microscale analytical method for
metabolomics because in CE nanolitre injection volumes are
used from microlitre sample amounts or less in the injection
vial (1–3). Therefore, CE–MS is highly suited for the profiling
of metabolites in ultrasmall biological samples, as has been
recently demonstrated for the analysis of CSF from mice and
extracts from small tissues or a single cell (4–7). Moreover,
CE (denoting here capillary zone electrophoresis [CZE])
separates compounds based on differences in their intrinsic
electrophoretic mobility, which is dependent on the charge and
size of the analyte and, therefore, well-suited for the analysis of
polar and charged metabolites. As the separation mechanism
of CE is fundamentally different from chromatographic-based
separation techniques, a complementary view on the
composition of (endogenous) metabolites present in a
given biological sample is provided. In comparison with
chromatographic-based methods the separation efficiency
of CE is very high because there is no mass transfer betwe
phases, and under perfect experimental conditions the only
source of band broadening in CE is from longitudinal diffus
Over the past few years, various new CE–MS approaches
been developed that show a strong potential for improving
sensitivity and metabolic coverage in metabolomics (8). T
article will focus on advances that significantly improved t
analytical performance—especially with regards to impro
the metabolic coverage—of CE–MS for volume-restricted
metabolomics studies.
Sheathless CE–MS Using a Porous Tip Spraye
CE is essentially a low-flow microscale separation techn
that reaches its optimal separation performance at very
flow rates—typically in the range of 20 nL/min to 100 nL
min—depending on the pH of the separation buffer wh
a bare fused-silica capillary. A high separation resolutio
obtained in CE by merely separating the compounds b
Resolving Volume-Restricted
Metabolomics Using Sheathless
Capillary Electrophoresis–
Mass Spectrometry
Rawi Ramautar, Division of Systems Biomedicine and Pharmacology, Leiden Academic Centre for Drug Research,
Leiden University, The Netherlands
The analytical toolbox used in present-day metabolomics encounters difficulties for the analysis of
limited amounts of biological samples. Therefore, a significant number of crucial biomedical and clinical
questions cannot be addressed by the current metabolomics approach. Capillary electrophoresis–
mass spectrometry (CE–MS) has shown considerable potential for the profiling of polar and charged
metabolites in volume-restricted or mass-limited biological samples. This article considers advances
that significantly improved the performance of CE–MS for in-depth metabolic profiling of limited sample
amounts. Attention is also devoted to various technical aspects that still need to be addressed to make
CE–MS a viable approach for volume-restricted metabolomics.
KEY POINTS
t� There is a strong need for microscale analytica
technology and workflows for volume-restricte
metabolomics.
t� CE–MS using a sheathless porous tip interfac
attractive tool for highly sensitive analysis of p
charged metabolites.
t� New pre-analytics are needed in combination
CE–MS to effectively profile metabolites in ult
samples.
-$r($�&VSPQF
658
Introducing the new Agilent Infi nityLab LC Purifi cation Solutions!
A complete and scalable portfolio based on a single platform
gives you the choice to tailor a system to meet your lab’s current
and future needs. Team up with Agilent and get on top of your
analytical and preparative-scale LC purifi cation challenges.
Live effi ciency. Live purifi cation.
www.agilent.com/chem/livepreplc
#WeLiveEffi ciency #Effi cientUHPLC
NEW PURIFICATION TEAM
WE LIVE EFFICIENCY
© Agilent Technologies, Inc. 2017
LC•GC Europe December 2017656
‘Like’ our page LCGC Join the LCGC LinkedIn groupFollow us @ LC_GC
Published by
UBM Americas
Vice President/Group
Publisher
Mike Tessalone
Editorial Director
Laura Bush
Editor-in-Chief
Alasdair Matheson
Managing Editor
Kate Mosford
Associate Editor
Lewis Botcherby
Sales Manager
Oliver Waters
Sales Executive
Liz Mclean
Senior Director, Digital
Media
Michael Kushner
Webcast Operations
Manager
Kristen Moore
Project Manager
Vania Oliveira
Digital Production Manager
Sabina Advani
Managing Editor Special
Projects
Kaylynn Chiarello-Ebner
kaylynn.chiarello.ebner@
ubm.com
Art Director
Dan Ward
Subscriber Customer Service
Visit (chromatographyonline.com)
to request or change a
subscription or call our customer
service department on
+001 218 740 6877
Hinderton Point,
Lloyd Drive,
Cheshire Oaks,
Cheshire,
CH65 9HQ, UK
Tel. +44 (0)151 353 3500
Fax +44 (0)151 353 3601
Daniel W. Armstrong
University of Texas, Arlington, Texas, USA
Günther K. Bonn
Institute of Analytical Chemistry and
Radiochemistry, University of Innsbruck,
Austria
Deirdre Cabooter
Department of Pharmaceutical and
Pharmacological Sciences, University of
Leuven, Belgium
Peter Carr
Department of Chemistry, University
of Minnesota, Minneapolis, Minnesota,
USA
Jean-Pierre Chervet
Antec Scientific, Zoeterwoude, The
Netherlands
Jan H. Christensen
Department of Plant and Environmental
Sciences, University of Copenhagen,
Copenhagen, Denmark
Danilo Corradini
Istituto di Cromatografia del CNR, Rome,
Italy
Hernan J. Cortes
H.J. Cortes Consulting,
Midland, Michigan, USA
Gert Desmet
Transport Modelling and Analytical
Separation Science, Vrije Universiteit,
Brussels, Belgium
John W. Dolan
LC Resources, McMinnville, Oregon,
USA
Anthony F. Fell
Pharmaceutical Chemistry,
University of Bradford, Bradford, UK
Attila Felinger
Professor of Chemistry, Department of
Analytical and Environmental Chemistry,
University of Pécs, Pécs, Hungary
Francesco Gasparrini
Dipartimento di Studi di Chimica
e Tecnologia delle Sostanze
Biologicamente Attive, Università “La
Sapienza”, Rome, Italy
Joseph L. Glajch
Momenta Pharmaceuticals, Cambridge,
Massachusetts, USA
Jun Haginaka
School of Pharmacy and Pharmaceutical
Sciences, Mukogawa Women’s
University, Nishinomiya, Japan
Javier Hernández-Borges
Department of Chemistry
(Analytical Chemistry Division),
University of La Laguna
Canary Islands, Spain
John V. Hinshaw
Serveron Corp., Hillsboro, Oregon,
USA
Tuulia Hyötyläinen
VVT Technical Research of Finland,
Finland
Hans-Gerd Janssen
Van’t Hoff Institute for the Molecular
Sciences, Amsterdam, The Netherlands
Kiyokatsu Jinno
School of Materials Sciences, Toyohasi
University of Technology, Japan
Huba Kalász
Semmelweis University of Medicine,
Budapest, Hungary
Hian Kee Lee
National University of Singapore,
Singapore
Wolfgang Lindner
Institute of Analytical Chemistry,
University of Vienna, Austria
Henk Lingeman
Faculteit der Scheikunde, Free University,
Amsterdam, The Netherlands
Tom Lynch
BP Technology Centre, Pangbourne, UK
Ronald E. Majors
Analytical consultant, West Chester,
Pennsylvania, USA
Debby Mangelings
Department of Analytical Chemistry
and Pharmaceutical Technology, Vrije
Universiteit, Brussels, Belgium
Phillip Marriot
Monash University, School of Chemistry,
Victoria, Australia
David McCalley
Department of Applied Sciences,
University of West of England, Bristol, UK
Robert D. McDowall
McDowall Consulting, Bromley, Kent, UK
Mary Ellen McNally
DuPont Crop Protection,Newark,
Delaware, USA
Imre Molnár
Molnar Research Institute, Berlin, Germany
Luigi Mondello
Dipartimento Farmaco-chimico, Facoltà
di Farmacia, Università di Messina,
Messina, Italy
Peter Myers
Department of Chemistry,
University of Liverpool, Liverpool, UK
Janusz Pawliszyn
Department of Chemistry, University of
Waterloo, Ontario, Canada
Colin Poole
Wayne State University, Detroit,
Michigan, USA
Fred E. Regnier
Department of Biochemistry, Purdue
University, West Lafayette, Indiana, USA
Harald Ritchie
Trajan Scientific and Medical, Milton
Keynes, UK
Koen Sandra
Research Institute for Chromatography,
Kortrijk, Belgium
Pat Sandra
Research Institute for Chromatography,
Kortrijk, Belgium
Peter Schoenmakers
Department of Chemical Engineering,
Universiteit van Amsterdam, Amsterdam,
The Netherlands
Robert Shellie
Trajan Scientific and Medical,
Ringwood, Victoria, Australia
Yvan Vander Heyden
Vrije Universiteit Brussel, Brussels,
Belgium
Editorial Advisory Board
Subscribe on-line at
www.chromatographyonline.com
The Publishers of LC•GC Europe would like to thank the members of the Editorial Advisory Board
for their continuing support and expert advice. The high standards and editorial quality associ-
ated with LC•GC Europe are maintained largely through the tireless efforts of these individuals.
LCGC Europe provides troubleshooting information and application solutions on all aspects of
separation science so that laboratory-based analytical chemists can enhance their practical
knowledge to gain competitive advantage. Our scientific quality and commercial objectivity
provide readers with the tools necessary to deal with real-world analysis issues, thereby
increasing their efficiency, productivity and value to their employer.
SUBSCRIPTIONS: LC•GC Europe is free to qualifi ed readers in Europe. To apply for a free subscription, or to change your name or address, go to
www.chromatographyonline.com, click on Subscribe, and follow the prompts.
To cancel your subscription or to order back issues, please email your request to
[email protected], putting LCE in the subject line.
Please quote your subscription number if you have it.
MANUSCRIPTS: For manuscript preparation guidelines, visit www.chromatographyonline.com or
call the Editor, +44 (0)151 353 3500. All submissions will be handled with reasonable care, but
the publisher assumes no responsibility for safety of artwork, photographs or manuscripts. Every
precaution is taken to ensure accuracy, but the publisher cannot accept responsibility for the
accuracy of information supplied herein or for any opinion expressed.
DIRECT MAIL LIST: Telephone: +44 (0)151 353 3500.
Reprints: Reprints of all articles in this issue and past issues of this publication are available
(250 minimum). Contact Brian Kolb at Wright’s Media, 2407 Timberloch Place, The Woodlands, TX
77380. Telephone: 877-652-5295 ext. 121. Email: [email protected].
© 2017 Advanstar Communications (UK) Ltd. All rights reserved. No part of this
publication may be reproduced in any material form (including photocopying or storing it in any
medium by electronic means and whether or not transiently or incidentally to some
other use of this publication) without the written permission of the copyright owner
except in accordance with the provisions of the Copyright, Designs & Patents Act
(UK) 1988 or under the terms of a licence issued by the Copyright Licensing
Agency, 90 Tottenham Court Road, London W1P 0LP, UK. Applications for the
copyright owner’s permission to reproduce any part of this publication should be
forwarded in writing to Permissions Dept, Hinderton Point, Lloyd Drive, Ellesmere
Port, Cheshire, CH65 9HQ. Warning: The doing of an unauthorized act in relation to a
copyright work may result in both a civil claim for damages and criminal prosecution.
10% Post
Consumer
Waste
For product information: www.vwr.com
Range of complementary stationary phases
Powerful method development software
Full training and consulting
Highly robust instrument
Solve HPLC method development challenges
U[UVGOCVKECNN[�CPF�GHƂEKGPVN[�s�WUKPI�#%'�EQNWOPU�
- providing chromatographers with more choices
for alternative selectivity, without compromising
stability or robustness.
%QODKPGF�YKVJ�%JTQO5YQTF#WVQ���CPF�
%JTQOCUVGT7NVTC�4U��OCZKOKUG�GHƂEKGPE[�KP�
your method development processes.
Need a demonstration? Email: [email protected]
FULLY AUTOMATED
METHOD
DEVELOPMENT
A major challenge in bioanalysis and metabolomics is the
profiling of (endogenous) metabolites in volume-restricted
and often scarce biological samples. Though the analytical
techniques currently used in metabolomics are powerful, the
need for relatively large amounts of sample for pre-analytics
and injection prevent their use in many biomedical and clinical
applications. For example, metabolic profiling of small-volume
biological samples, such as cerebrospinal fluid (CSF) from
mouse models, spheroids or microtissues, liquid biopsies,
and samples from microfluidic organ-on-a-chip systems,
is seriously hindered by the current analytical technologies
because of the limited sample material. The standard
analytical techniques also do not allow a maximum amount of
biochemical information from valuable and scarcely available
biological samples to be obtained because the material is
completely consumed for a single metabolomics measurement.
Despite important developments in analytical separations
technology over recent years, the analysis of small-volume
biological samples remains a challenging task. Therefore, there
is a critical need for the introduction of analytical methods to
allow highly sensitive metabolic profiling of volume-restricted or
mass-limited biological samples.
Capillary electrophoresis–mass spectrometry (CE–MS) can
be considered an attractive microscale analytical method for
metabolomics because in CE nanolitre injection volumes are
used from microlitre sample amounts or less in the injection
vial (1–3). Therefore, CE–MS is highly suited for the profiling
of metabolites in ultrasmall biological samples, as has been
recently demonstrated for the analysis of CSF from mice and
extracts from small tissues or a single cell (4–7). Moreover,
CE (denoting here capillary zone electrophoresis [CZE])
separates compounds based on differences in their intrinsic
electrophoretic mobility, which is dependent on the charge and
size of the analyte and, therefore, well-suited for the analysis of
polar and charged metabolites. As the separation mechanism
of CE is fundamentally different from chromatographic-based
separation techniques, a complementary view on the
composition of (endogenous) metabolites present in a
given biological sample is provided. In comparison with
chromatographic-based methods the separation efficiency
of CE is very high because there is no mass transfer between
phases, and under perfect experimental conditions the only
source of band broadening in CE is from longitudinal diffusion.
Over the past few years, various new CE–MS approaches have
been developed that show a strong potential for improving the
sensitivity and metabolic coverage in metabolomics (8). This
article will focus on advances that significantly improved the
analytical performance—especially with regards to improving
the metabolic coverage—of CE–MS for volume-restricted
metabolomics studies.
Sheathless CE–MS Using a Porous Tip SprayerCE is essentially a low-flow microscale separation technique
that reaches its optimal separation performance at very low
flow rates—typically in the range of 20 nL/min to 100 nL/
min—depending on the pH of the separation buffer when using
a bare fused-silica capillary. A high separation resolution is
obtained in CE by merely separating the compounds based
Resolving Volume-Restricted Metabolomics Using SheathlessCapillary Electrophoresis–Mass SpectrometryRawi Ramautar, Division of Systems Biomedicine and Pharmacology, Leiden Academic Centre for Drug Research,
Leiden University, The Netherlands
The analytical toolbox used in present-day metabolomics encounters difficulties for the analysis of limited amounts of biological samples. Therefore, a signifi cant number of crucial biomedical and clinical questions cannot be addressed by the current metabolomics approach. Capillary electrophoresis– mass spectrometry (CE–MS) has shown considerable potential for the profi ling of polar and charged metabolites in volume-restricted or mass-limited biological samples. This article considers advances that signifi cantly improved the performance of CE–MS for in-depth metabolic profi ling of limited sample amounts. Attention is also devoted to various technical aspects that still need to be addressed to make CE–MS a viable approach for volume-restricted metabolomics.
Ph
oto
Cre
dit: N
ik M
erk
ulo
v/S
hu
tte
rsto
ck.c
omKEY POINTS
• There is a strong need for microscale analytical
technology and workflows for volume-restricted
metabolomics.
• CE–MS using a sheathless porous tip interface is an
attractive tool for highly sensitive analysis of polar and
charged metabolites.
• New pre-analytics are needed in combination with
CE–MS to effectively profile metabolites in ultrasmall
samples.
LC•GC Europe December 2017658
on their electrophoretic mobilities, that is, under (near-)zero
electroosmotic flow conditions. The inherently low flow rates of
CE are also useful with regards to the electrospray ionization
(ESI) mechanism. In ESI, smaller droplets are generated
under low flow separation conditions, which results in a more
efficient desolvation and an improved transfer of ions to the MS
system (9–11). Moreover, at very low flow rates (≤20 nL/min) ion
suppression is significantly reduced, resulting in an improved
concentration sensitivity (9), which is important for ultratrace
detection of metabolites in limited sample amounts.
In a conventional CE system, both ends of the separation
capillary are immersed in buffer vials to which electrodes
are added to provide a high voltage gradient. To couple CE
to MS, the outlet vial must be replaced by an interface to
close the electrical circuit and to provide contact with the ESI
stream. Therefore, a CE–MS interface needs to apply voltage
to the capillary outlet while maintaining independent CE and
ESI electrical circuits. A sheath-liquid interface and various
other interfacing techniques have been developed to allow
the coupling of CE to MS. Until now, most CE–MS-based
metabolomics studies have been performed with a sheath-liquid
interface (12). CE–MS approaches using a sheath-liquid interface
for metabolomics were first developed by Soga and co-workers
(13,14). The sheath-liquid interface, originally designed by
Smith and co-workers (15), has been used for a broad range
of bioanalytical applications with acceptable analytical figures
of merit. However, the sheath-liquid is generally provided at a
flow rate between 2 μL/min to 10 μL/min, thereby significantly
diluting the CE effluent and resulting in compromised detection
sensitivities for metabolomics applications. Moreover, this flow
rate is not compatible with nano-ESI-MS. However, an important
benefit of the sheath-liquid interface is that the composition can
be modified to improve the ionization efficiency without affecting
the selectivity and efficiency of the electrophoretic separation
(16,17). The influence of these agents on metabolomics studies
by CE–MS still needs to be explored. Overall, as both CE
and ESI-MS perform optimally at low flow-rate conditions, the
coupling of CE to MS should preferably be performed via an
interface that effectively makes use of the intrinsically low flow
separation property of CE and the improved ESI efficiency under
these conditions.
At present, the development of new or improved interfacing
designs for CE–MS and the evaluation of their potential for
bioanalysis and metabolomics remains an active area of
research (18–25). So far, the most encouraging results for
volume-restricted metabolomics studies have been obtained by
CE–MS using a sheathless porous tip interface. In this design,
which was developed by Moini (19), a porous capillary tip
inside a cylindrical metal tube maintains electrical contact via
transport of ions and electrons through the capillary wall while
spraying the CE effluent into the nano-ESI-MS system. The
sheathless porous tip design is especially useful for interfacing
narrow (<30 μm internal diameter [i.d.]) capillaries and for low
flow-rate (<20–30 nL/min) nano-ESI-MS analyses (26).
In-Depth Metabolic Profi ling of Volume-Restricted SamplesThe capabilities of CE–MS using a sheathless porous tip
interface for global profiling of cationic metabolites in human
urine have been assessed (27). Low nanomolar limits of
detection (LODs) were obtained for a wide range of metabolite
classes in human urine by using only an injection volume of
9 nL. As a follow-up to this work, the potential of sheathless
CE–MS for metabolic profiling of very minute amounts of
biological samples was assessed. To this end, mouse CSF
was selected as a typical example of a biological material only
available in minute quantities, that is, only a few microlitres
can be obtained under proper conditions (4). Figure 1 shows
a metabolic profile obtained by sheathless CE–MS for mouse
CSF after only a 1:1 dilution with water, thereby fully retaining
sample integrity. By using an injection volume of 45 nL from a
vial containing only 2 μL of a 1:1 diluted CSF, more than 300
compounds could be observed. As only 45 nL of the sample
was consumed, the proposed method allows multiple analyses
on a single highly valuable and precious mouse CSF sample,
enabling repeatability studies and, even more interesting,
analysis of the same sample at different separation conditions
to further increase the metabolic coverage. Performing multiple
analyses on a single, scarce biological sample is not possible
with conventional analytical techniques.
Apart from volume-restricted biological samples, such as
mouse CSF, there is currently a strong interest in analytical
tools capable of providing highly sensitive metabolic profiles
for microscale cell culture samples. The performance of
sheathless CE–MS has therefore been assessed for the
profiling of cationic metabolites in an extract from approximately
10,000 HepG2 cells. Figure 2 shows that a reasonable number
of metabolites could be detected by sheathless CE–MS using
an injection volume of 60 nL, which in this case equals the
aliquot of only 5 HepG2 cells. An improvement of the injection
part is needed to further enhance the detection sensitivity of
the sheathless CE–MS method, however, the results obtained
so far clearly indicate the strong potential of the method for
metabolic profiling of limited sample amounts.
Recently, the utility of the sheathless CE–MS method
was also examined for the profiling of anionic metabolites in
biological samples (28), using the same separation conditions
used for the profiling of cationic metabolites—only the MS
detection and CE separation voltage polarity were switched.
A wide range of anionic metabolite classes could be profiled
under these conditions, including sugar phosphates,
nucleotides, and organic acids. An injection volume of 20 nL
resulted in nanomolar detection limits, which corresponded to
659www.chromatographyonline.com
Ramautar
50
1
2
3
4
5
x106
10 15 20 25
Migration time (min)
Inte
nsi
ty (
cou
nts
)Figure 1: Metabolic profile (m/z 65–1000) of mouse cerebrospinal
fluid obtained with sheathless CE–MS using a porous tip
sprayer. Conditions: separation buffer: 10% acetic acid (pH 2.2);
electrophoretic separation at +30 kV; sample injection: circa
45 nL. Adapted with permission from reference 4.
a significant enhancement when compared to the micromolar
detection limits typically obtained with classical sheath-liquid
CE–MS methods. Structural isomers of phosphorylated sugars
as well as isobaric metabolites could be selectively analyzed
by the proposed sheathless CE–MS method without using
any derivatization. The approach was used for the profiling of
anionic metabolites in extracts from the glioblastoma cell line.
Around 100 compounds were found for an injection volume of
20 nL, corresponding to the content of approximately 400 cells.
Future DirectionsThe examples highlighted here clearly indicate the ability
of sheathless CE–MS using a porous tip sprayer for
in-depth metabolic profiling of minute amounts of sample
material. Although (sub-)nanomolar detection limits can
be obtained for a wide range of polar compounds by only
using nanolitre injection volumes, as required for trace-level
analysis of metabolites in limited sample amounts, the next
important step is to examine the utility of sheathless CE–MS
for large-scale volume-restricted metabolomics studies.
Hirayama and co-workers have recently shown that a single
sheathless porous tip capillary can be used for more than
180 successive runs of a 10-fold diluted human urine sample
(29). Still, the long-term performance of sheathless CE–MS for
volume-restricted metabolomics needs to be assessed in more
extended studies analyzing large numbers of diverse biological
samples.
When using a highly sensitive microscale analytical
method for in-depth metabolic profiling, attention should also
be paid to the design of reliable sampling techniques for
volume-restricted or mass-limited samples. Nemes
and co-workers have recently developed a microprobe
single-cell CE–MS method, which integrates capillary
microsampling, microscale metabolite extraction, and
CE–MS analysis, for in situ mass spectrometric profiling of
metabolites in single cells (7). It is anticipated that microprobe
single-cell CE–MS will allow metabolomics studies in larger
populations of single cells and other model organisms. Proper
sample pretreatment strategies are needed for the selective
extraction of polar and charged metabolites from minute
amounts of sample. In this context, it would be interesting to
evaluate strategies like in-line or on-line solid-phase extraction
(SPE)-CE–MS or to explore the possibilities of electrodriven
sample pretreatment techniques, such as electromembrane
extraction or three-phase electroextraction (30,31). Overall,
further developments in this area may result in a more viable
CE–MS approach for probing the polar metabolome in
volume-restricted samples.
AcknowledgementsRawi Ramautar would like to acknowledge the financial support
of the Veni grant scheme of the Netherlands Organization for
Scientific Research (NWO Veni 722.013.008).
Confl ict of InterestThe author has no other relevant affiliations or financial
involvement with any organization or entity with a financial
interest in or financial conflict with the subject matter or
materials discussed in the manuscript apart from those
disclosed.
LC•GC Europe December 2017660
Ramautar
Time (min)
Inte
nsi
ty (
cou
nts
)
9e4
8e4
7e4
6e4
5e4
4e4
3e4
2e4
1e4
0e013 14 15 16 17
m/z 284,1157
m/z 76,0462
m/z 118,0943
m/z 130,0945
m/z 192,1707
m/z 90,0624
m/z 133,0693
m/z 120,0888
m/z 120,0735
m/z 106,0575
m/z 134,0532
m/z 150,0674
m/z 116,0786
m/z 156,0862
m/z 132,1103
m/z 147,1219
m/z 130,0582
m/z 147,0856
m/z 182,0921
m/z 182,0588
m/z 166,0963
m/z 241,0558
m/z 137,0543
m/z 175,1296
m/z 132,0852
m/z 537,9050
m/z 148,0695
18 19 20 21 22 23 24 25 26 27
Figure 2: Multiple extraction ion electropherograms for a selected number of metabolites obtained with sheathless CE–MS using a porous
tip sprayer in an extract of HepG2 cells. Conditions: separation buffer: 10% acetic acid (pH 2.2); electrophoretic separation at +30 kV; sample
injection: circa 60 nL.
References(1) W. Zhang, T. Hankemeier, and R. Ramautar, Curr. Opin. Biotechnol.
43, 1–7 (2017).(2) N.L. Kuehnbaum and P. Britz-McKibbin, Chem. Rev. 113, 2437–2468
(2013).(3) A. Slampova and P. Kuban, J. Chromatogr. A 1497, 164–171 (2017).(4) R. Ramautar, R. Shyti, et al., Anal. Bioanal. Chem. 404, 2895–2900
(2012).(5) J.X. Liu, J.T. Aerts, S.S. Rubakhin, X.X. Zhang, and J.V. Sweedler,
Analyst 139, 5835–5842 (2014).(6) R.M. Onjiko, S.A. Moody, and P. Nemes, Proc. Natl. Acad. Sci. USA
112, 6545–6550 (2015).(7) R.M. Onjiko, E.P. Portero, S.A. Moody, and P. Nemes, Anal Chem. in
press (2017).(8) P.W. Lindenburg, R. Haselberg, G. Rozing, and R. Ramautar,
Chromatographia 78, 367–377 (2015).(9) A. Schmidt, M. Karas, and T. Dulcks, J. Am. Soc. Mass Spectrom. 14,
492–500 (2003).(10) M. Wilm and M. Mann, Anal. Chem. 68, 1–8 (1996).(11) G.A. Valaskovic, N.L. Kelleher, and F.W. McLafferty, Science 273,
1199–1202 (1996).(12) A. Hirayama, M. Wakayama, and T. Soga, TrAC Trends in Analytical
Chemistry 61, 215–222 (2014).(13) T. Soga, Y. Ohashi, Y. Ueno, H. Naraoka, M. Tomita, and T. Nishioka,
J. Proteome Res. 2, 488–494 (2003).(14) T. Soga, Y. Ueno, H. Naraoka, Y. Ohashi, M. Tomita, and T. Nishioka,
Anal. Chem. 74, 2233–2239 (2002).(15) R.D. Smith, C.J. Barinaga, and H.R. Udseth, Anal. Chem. 60,
1948–1952 (1988).(16) T.J. Causon, L. Maringer, W. Buchberger, and C.W. Klampfl, J.
Chromatogr. A 1343, 182–187 (2014).(17) G. Bonvin, S. Rudaz, and J. Schappler, Anal. Chim. Acta 813, 97–105
(2014).(18) E.J. Maxwell, X. Zhong, H. Zhang, N. van Zeijl, and D.D. Chen,
Electrophoresis 31, 1130–1137 (2010).(19) M. Moini, Anal. Chem. 79, 4241–4246 (2007).(20) R. Wojcik, O.O. Dada, M. Sadilek, and N.J. Dovichi, Rapid Commun.
Mass Spectrom. 24, 2554–2560 (2010).(21) X. Guo, T.L. Fillmore, Y. Gao, and K. Tang, Anal. Chem. 88,
4418–4425 (2016).
(22) S.B. Choi, M. Zamarbide, M.C. Manzini, and P. Nemes, J. Am. Soc. Mass Spectrom. 28, 597–607 (2017).
(23) T.T. Nguyen, N.J. Petersen, and K.D. Rand, Anal. Chim. Acta 936, 157–167 (2016).
(24) V. Gonzalez-Ruiz, S. Codesido, J. Far, S. Rudaz, and J. Schappler, Electrophoresis 37, 936–946 (2016).
(25) J. Krenkova, K. Kleparnik, J. Grym, J. Luksch, and F. Foret, Electrophoresis 37, 414–417 (2016).
(26) J.M. Busnel, B. Schoenmaker, et al., Anal. Chem. 82, 9476–9483 (2010).
(27) R. Ramautar, J.M. Busnel, A.M. Deelder, and O.A. Mayboroda, Anal. Chem. 84, 885–892 (2012).
(28) C. Gulersonmez, S. Lock, T. Hankemeier, and R. Ramautar, Electrophoresis 37, 1007–1014 (2016).
(29) A. Hirayama, M. Tomita, and T. Soga, Analyst 137, 5026–5033 (2012).(30) A. Gjelstad and K.F. Seip, Bioanalysis 7, 2133–2134 (2015).(31) R.J. Raterink, P.W. Lindenburg, R.J. Vreeken, and T. Hankemeier,
Anal. Chem. 85, 7762–7768 (2013).
Rawi Ramautar strongly believes in the power of analytical
technology to contribute to a better understanding of
biochemical mechanisms underlying diseases. He therefore
studied both pharmacochemistry and analytical sciences
at the Vrije Universiteit of Amsterdam (The Netherlands) to
have the necessary background to undertake a Ph.D. in this
direction. In 2010, he completed his Ph.D. on the development
of capillary electrophoresis–mass spectrometry methods
for metabolomics at Utrecht University (The Netherlands).
Intrigued by metabolomics for disease prediction and
diagnosis, Rawi switched to the Leiden University Medical
Center (The Netherlands) to broaden his horizon on this topic.
Currently, he is a principal investigator (tenured) at the Leiden
Academic Center for Drug Research of the Leiden University
where his group is developing microscale analytical workflows
for volume-restricted biomedical problems.
www.biotech.se
INNOVATIVE PRODUCTS FOR FLUIDIC SYSTEMS
HEADQUARTER ° BIOTECH AB ° SWEDEN ° TEL +46 (0)300-56 91 80 ° [email protected]�USA ° BIOTECH USA LLC ° TEL 612-703-5718 ° [email protected] ° BIONIK INC. ° +81 545-38-9125 ° [email protected]
GO WITH A HIGH FLOWINTRODUCING THE ALL NEW
DEGASI PREP+
Biotech in co-operation with IDEX H&S are proud to announce
the world’s first in-line, membrane degasser ready to use with
aggressive media and organic solvents, while maintaining
flow-rates up to 150ml/min and above.* Available configura-
tions: Stand alone, OEM open frame and 1-4 channels.
The world’s first high flow rate, in-line
degasser, for organic solvents
* 150ml/min is only a guideline value, the degasser will in many
application work with flow rates well above this guideline. The
degasser comes with 2-channels, where each channels recom-
mended flow rate is 75 ml/min, connecting these chambers in a
series will allow for a combined flow rate of 150 ml/min.
HIGH
FLOW RATE
75 ml/min
Per channel
661www.chromatographyonline.com
Ramautar
662 LC•GC Europe December 2017
SAMPLE PREPARATIONPERSPECTIVES
Food safety has become a top
concern in our society. In general
the public is increasingly concerned
about the safety of the food products
they consume every day as more and
more food contamination incidents
and widespread recalls arise. Such
incidents have alerted the authorities
and the public that more efforts and
deeper investigations are needed.
As a result, reliable and efficient
methods for food safety analyses are
required. Even with modern detection
techniques, because of the low
concentrations of contaminants and
complicated food matrices, efficient
sample preparation is necessary.
Traditional extraction methods
for food samples, such as liquid–
liquid extraction (LLE) and Soxhlet
extraction, are often time-consuming
and require large amounts of
organic solvents. Therefore, one
of the objectives of analytical food
safety studies has currently been
the development of new extraction
techniques that can improve the
accuracy and precision of analytical
results and simplify the analytical
procedure.
Because of increased concerns
for food safety, attention is given to
developing methods for determination
of contaminants and other harmful
substances from food samples. The
analysis of food samples is usually a
complicated procedure involving many
steps. It requires extensive sample
extraction before further analysis.
Sample extraction is a crucial step
in food sample analysis because it
can affect the concentration of the
analyte and the cleanliness of the
sample. Traditional sample extraction
techniques used in food safety
studies are based on the suitable
choice of solvents and the use of
heat and agitation to improve the
solubility of the desired compounds
and the mass transfer (1), like in
Soxhlet, liquid–liquid, and shake-flask
extraction. Some of the traditional
extraction techniques can require
a great deal of time—Pedersen
and Olsson (2) performed Soxhlet
extraction of acrylamide from potato
chips, and it took seven days to get
a complete extraction. Frenich and
coworkers (3) reported a method
for the determination of residues of
organochlorine and organophosphorus
pesticides using Soxhlet extraction.
This extraction method involved
laborious steps with the use of large
amounts of solvent. Analysis of
polychlorinated biphenyls (PCBs),
polychlorinated dibenzo-p-dioxins
(PCDDs), and polychlorinated
dibenzofurans (PCDFs) in butter
based on three different liquid–liquid
extraction methods was studied
by Ramos and coworkers (4). The
reported methods also involved
time-consuming and large solvent
consumption steps. These traditional
extraction techniques are quite
laborious, time-consuming, and involve
large quantities of organic solvents,
which are flammable, expensive, and
generate hazardous waste.
In recent years, several new
extraction techniques emerged
as alternatives to the conventional
sample preparation methods,
including microwave-assisted
extraction (MAE), ultrasound-assisted
extraction (UAE), accelerated
solvent extraction (ASE) (also
known as pressurized solvent
extraction [PSE]), supercritical fluid
extraction (SFE), and solid-phase
microextraction (SPME). These new
extraction techniques have numerous
advantages over traditional extraction
methods, including shortened
extraction times, reduced solvent
consumption, reduced cost, and
automation. In this column, we will
provide an overview of these newer
extraction techniques that have been
applied to food safety studies.
Liquid–Liquid ExtractionLiquid–liquid extraction is the most
widely used method for the extraction
of analytes from aqueous food
samples. In LLE, the sample
is distributed or partitioned between
two immiscible solvents in which the
analyte has different solubilities. The
solution containing the analyte must
be immiscible with the solvent used
to extract the analyte. The
main advantages of this method
are the wide availability of solvents
and the use of low-cost apparatus.
However, low recoveries, limited
selectivity, and time-consuming
procedures limit LLE. A variety of
microscale variants of LLE have been
The Use of Extraction Technologies in Food Safety StudiesChangling Qiu1 and Douglas E. Raynie2, 1Department of Chemistry and Biochemistry, University of Texas at Arlington, Arlington,
Texas, USA, 2Sample Preparation Perspectives Editor
Traditional extraction methods for food samples, such as liquid–liquid extraction and Soxhlet extraction, are often time-consuming and require large amounts of organic solvents. Therefore, one of the objectives of analytical food safety studies currently has been the development of new extraction techniques that can improve the accuracy and precision of analytical results and simplify the analytical procedure.
663www.chromatographyonline.com
SAMPLE PREPARATION PERSPECTIVES
reported and food applications of
these have been recently reviewed
(5).
Solid-Phase ExtractionIn SPE, the sample passes over the
stationary phase (solid phase). The
analytes separate according to the
degree to which each component
is partitioned or adsorbed by the
stationary phase. The analytes may
favourably adsorb to the solid phase
or they may remain in the liquid
phase. If the analytes are adsorbed,
a stronger eluting solvent selectively
Liquid–liquid extraction is the most widely used method for the extraction of analytes from aqueous food samples. The sample is distributed or partitioned between two immiscible solvents in which the analyte has different solubilities.
highly capable and
easy to use whether
5 or 500 samples
sample preparation for allanalytical laboratories
jÒANALYTICALÒPRECISIONÒjÒSIMPLEÒ7ORK¹OWÒSOFTWAREÒjÒ
• robotic syringe exchange •
adaptable rack/vial platform •
jÒMICROÒ30%ÒANDÒμLTERÒCARTRIDGESÒjÒWIDEÒRANGEÒ
of syringes, tools and accessories •
Effortless Accuracy Consistent Results
Automation for Every Laboratory
WWW�EPREP�COM�AU�E0REP ��� ��� ����Ò����ÒÒÒÒÒÒINFO EPREP�COM�AU
jÒANAL
jÒM
Effortless Accuracy Co
Automation for Eve
(a) (b)
(c)
Figure 1: Schematics of: (a) the fibre SPME extraction procedure, (b) thermal desorption in a GC injection port, and (c) solvent desorption using an SPME interface.
664 LC•GC Europe December 2017
SAMPLE PREPARATION PERSPECTIVES
desorbs the analytes. If the analytes
remain in the liquid phase, they
can be collected and prepared
properly for further analysis, provided
the interferents are bound to the
solid-phase sorbent.
Effective separation by SPE can
be achieved by choosing suitably
selective solid-phase sorbent
and eluting solvents. With proper
selection of the sorbent and solvents,
SPE is capable of being used for
gases, solids, and liquids. However,
the primary area of application of
SPE is in the selective extraction and
enrichment of liquids samples. SPE
is used widely in the environmental,
pharmaceutical, biological, clinical,
forensic science, and food and
beverage areas.
SPE is also used to concentrate
and clean a sample before using
a chromatographic or other
analytical method. SPE has very
extensive applications in food
safety studies because of its low
cost, good selectivity, small solvent
consumption, and high recovery.
However, long sample preparation
times and multistep procedures are
among its disadvantages.
Solid-Phase MicroextractionSPME, developed by Pawliszyn
and coworkers (6) in 1990, involves
the use of a fibre coated with
suitable stationary phase extracting
material for the removal of analytes
of interest from the sample. The
sample molecules adsorb onto the
fibre and subsequently desorb into
the gas chromatograph’s injection
port for analysis. It is a simple, fast,
inexpensive, and efficient extraction
method that has been applied to both
headspace and aqueous sample
analysis with great sensitivity and
selectivity (7).
SPME is most effective when
coupled to gas chromatography
(GC). It has also been used with high
performance liquid chromatography
(HPLC) separations (8). Figure 1
shows the SPME device (9). It
consists of a fibre bonded to a
stainless steel plunger and installed
in a holder. The fibre, which is coated
with a suitable stationary phase, is
immersed in the sample or exposed
to the headspace above the sample.
Analytes in aqueous samples are
extracted by direct immersion, where
To fume
hood
IR
temperature
sensors
Air flow
Solvent
detector
Magnetic
stirring
Microwave
field
Direct
temperature
sensor
Multimode
microwave
cavity
Figure 3: Schematic diagram of instrumentation for microwave-assisted extraction (24).
Figure 2: Steps in the original and official versions of QuEChERS sample preparation for pesticide residues in food commodities.
665www.chromatographyonline.com
SAMPLE PREPARATION PERSPECTIVES
analytes partition between the aqueous sample and
the fibre coating. When equilibrium is reached, the fibre
is removed and exposed to the injection port of a gas
chromatograph for analysis. Headspace analysis can be
used for the extraction of volatile or semivolatile analytes
from solid, liquid, or gaseous samples. In the headspace
extraction mode, the analytes first partition between
the sample and the headspace, then the analytes are
adsorbed by the fibre, which is then inserted directly into
the injection port of a GC system (10).
Since SPME is an equilibrium extraction technique,
several factors influence the extraction efficiency, such
as fibre-coating thickness and characteristics, sample
size, vial size, and adsorption and desorption conditions
(temperature and time) (10). To achieve successful
quantitative analysis, it is vital that each of these variables
is constant between analyses.
SPME has become increasingly popular in the analysis
of volatile and semivolatile compounds because of its
advantages over conventional extraction methods. It is a
simple, effective, and low-cost technique. The extraction
combines sampling, isolation, and concentration in one
step. SPME is also considered environmentally friendly
because of the elimination of organic solvents. The
technique has been widely applied to food samples (7,11–
13). One of these recent reviews (11) provided a summary
of fibre coatings, while another (13) discussed techniques
that are considered variants of SPME.
Shake-Flask ExtractionThe most common approach for extraction from solids
is conventional liquid-solid extraction, in the form of
shake-flask extraction. Shake-flask extractions can be
easily performed by putting a sample into a flask, adding
solvent, and agitating. After extraction, the extract is
separated from the solid residue by filtration. Shake-flask
extraction requires minimal glassware, small amounts of
organic solvent, and is comparatively fast (10–50 min).
It is one of the oldest and most widely used extraction
methods. However, because of its poor recovery and low
efficiency, the application is limited.
Soxhlet ExtractionSoxhlet extraction is a traditional extraction technique
for many food samples. It was originally designed for the
extraction of a lipid from a solid material by Franz von
Soxhlet in 1879 (14). The technique uses a specialized
piece of glass apparatus where the solid sample is placed
and is semicontinuously extracted with a sub-boiling
solvent. Though Soxhlet extraction is simple, standard,
and robust, there are disadvantages (15). Soxhlet
extraction usually requires long extraction times (8–24 h)
and large amounts of solvent. The operation lacks
automation, but several samples can be extracted in
parallel.
QuEChERSQuEChERS (quick, easy, cheap, effective, rugged, and
safe) extraction has become a very attractive sample
extraction method for various food samples. This method
was developed by Lehotay and Anastassiades in 2003
for the analysis of pesticides in vegetables and fruits (16).
Now, QuEChERS has been widely used in pharmaceutical,
clinical, and environmental analysis including steroids,
hormones, acetaminophen, acrylamide, perfluorinated
compounds, polycyclic aromatic hydrocarbons, alkaloids,
mycotoxins, and other applications. Overall, this procedure
has two main steps: extraction with a solvent and
partitioning salts, and clean up with dispersive solid-phase
extraction (dSPE). Figure 2 displays the sequence of
events in QuEChERS in the original and the AOAC and
European official methods (17). The QuEChERS method
has many advantages over traditionally used techniques.
QuEChERS provides accurate analytical results with high
recoveries, saves time and labour, reduces hazardous
solvent consumption and waste disposal, and uses less
laboratory glassware with a minimal number of steps.
Rajczak and Tuzimski (18) recently reviewed QuEChERS,
including food applications.
Ultrasound-Assisted ExtractionUltrasound-assisted extraction is also employed in food
safety studies for the extraction of contaminants or
bioactive components from food materials. The principle
of UAE is based on the propagation of ultrasound pressure
Sample extraction is a crucial step in food sample analysis because it can affect the concentration of the analyte and the cleanliness of the sample.
666 LC•GC Europe December 2017
SAMPLE PREPARATION PERSPECTIVES
waves and resulting in a cavitation
phenomena. Ultrasound waves are
elastic waves that have a frequency
above the threshold of human
hearing, approximately 20 kHz. The
extraction mechanism involves two
steps, diffusion through the cell walls
and releasing the cell content after
the walls are disrupted (19). In one
configuration, the sample is immersed
in an ultrasonic bath with a solvent
and subjected to ultrasonic radiation.
Higher energy extractions use a
horn or probe device. Ultrasound
waves create bubbles in the solvent
and produce high local negative
pressure that can cause the collapse
of cavitation bubbles. The collapse
of cavitation bubbles near cell walls
produces cell disruption, and as a
result, solvent penetrates into the cells
and causes the release of extractable
compounds. The ultrasound waves
can also facilitate the diffusion process
and increase mass transfer.
UAE can reduce extraction time
and solvent consumption, thus
resulting in higher extraction
rates and good extraction efficiency.
Compared to other extraction
techniques, UAE is simple, fast,
productive, inexpensive, and
capable of operating with many
samples at one time. UAE usually
provides good results for food
samples (20–22). The benefits
for using UAE for food samples
include enhancement of extraction
yield or rate and extraction of
heat-sensitive bioactive and food
components under lower processing
temperatures. Food components
such as antioxidants, phenols,
aromas, carotenoids, anthocyanins,
and oils can be isolated from fruits
and vegetables, herbs and spices,
and seeds using UAE (22).
Microwave-Assisted ExtractionMicrowave-assisted extraction is an
extraction technique that combines
microwave and traditional solvent
extraction. Its use in food safety
analysis has become one of the more
common and low-cost extraction
methods today. Typically, a microwave
system includes a microwave
power generator, a waveguide for
transmission, a resonant cavity, and
a power supply (23). The microwave
power generator is a magnetron. At
the common microwave frequency
of 2.45 GHz, electromagnetic
energy is conducted from the
magnetron to the cavity using a
waveguide. The sample and solvent
placed inside the resonant cavity is
subjected to microwave energy. This
arrangement is outlined in Figure 3
(24). After typically 5–30 min, the
extraction is complete, and the extract
can be filtered and prepared for
analysis.
Compared to other extraction
techniques, an important advantage of
MAE is the extraction rate acceleration
resulting from the high temperatures
employed. Therefore, short extraction
times are obtained. Other advantages
include reduced solvent consumption
and improved extraction yield and
product quality. On the other hand,
disadvantages include an additional
filtration step needed to remove the
solid residue after the extraction, the
poor efficiency of microwaves when
the solvents are nonpolar and volatile,
and the use of high temperatures
that might degrade heat-sensitive
compounds. The stated requirement
that analytes or samples must be
polar to absorb microwave energy
is usually ignored with food analysis
because of the typically high water
content.
MAE has been applied to a
diverse range of sample types (soils,
sediments, sewage sludge, plants,
food). MAE is employed extensively
in the extraction of pesticides,
pigments, bioactive compounds
from vegetables, plants, and
natural products as an alternative to
traditional techniques of extraction
(25,26).
Compared to other extraction techniques, an important advantage of MAE is the extraction rate acceleration resulting from the high temperatures employed.
Solvent
Mixing
valve
Relief
valve
Solvent
Oven
Pump
Extraction cell
Static valve
Collection bottle
Solvent
Figure 4: Schematic diagram of accelerated solvent extraction instrumentation.
www.chromatographyonline.com
SAMPLE PREPARATION PERSPECTIVES
Supercritical Fluid ExtractionSupercritical fluid extraction uses a fluid phase having
unique properties between a gas and a liquid to effect
the solubilization of solutes. Compared to traditional
liquid solvents, supercritical fluids have lower viscosities
and high diffusivities, thus allowing more-efficient mass
transfer of solutes from sample matrices. SFE can be
operated in two modes, off-line and on-line (27). In the
on-line mode, the SFE instrument is coupled directly to
the analytical instrument, such as with SFE–GC. Off-line
SFE focuses on the sample preparation only, for analytical
purposes or on a larger scale to either remove unwanted
components from a product or collect desired components
(28).
An SFE system contains reservoirs for the supercritical
fluid and cosolvent, a thermostated extraction cell, a
restrictor to maintain flow and pressure, and a collection
vial. Typically, the supercritical fluid is pumped to a
heating zone, where it is heated to supercritical conditions.
It then passes into the extraction cell, where it rapidly
diffuses into the sample and dissolves the components
to be extracted. The dissolved components flow from the
extraction cell into a collection vial. The supercritical fluid
can then be condensed and recycled, or discharged to
atmosphere.
The most commonly used supercritical fluid is carbon
dioxide, which has a critical point of 31.3 °C and 72.8 bar.
This low critical temperature and pressure allows
extraction to occur near room temperature and at a
mild pressure. Carbon dioxide is inexpensive, nontoxic,
nonflammable, inert, and a good solvent for nonpolar
molecules. In general, supercritical carbon dioxide
extraction has a very wide range of applications, such as
in food, cosmetics, pharmaceutical, environmental, and
other related industries. Pesticides, organic pollutants,
fats and lipids, flavours, and natural bioactive components
are all classes of compounds that can be separated and
extracted from food samples (29). One attractive feature
for the use of carbon dioxide in food analysis is that it is
“generally regarded as safe” by the U.S. Food and Drug
Administration.
Accelerated Solvent Extraction Accelerated solvent extraction is a fast and automatic sample
extraction technique that uses elevated temperatures and
pressures with liquid solvents to obtain fast and efficient
extractions. It allows a high extraction efficiency with a small
volume of solvent (10–40 mL) and a short extraction time
(5–20 min).
ASE is mostly applicable to solid or semisolid samples
that can be held in the extraction cell during extraction.
A schematic of the ASE apparatus is presented in
Figure 4 (30). With ASE, a solvent or a mixture of
solvents is pumped into an extraction cell containing
the sample, which is then brought to elevated pressure
and temperature for extraction. Following extraction, the
sample is purged with compressed gas and prepared for
analysis. Application of ASE in food safety studies has
been reported for the extraction of various compounds
and contaminants like residual pesticides, fats and
lipids, food additives, and microbial contaminants in food
samples (31).
Optimization of various parameters in ASE, including
solvent, temperature, pressure, extraction cycles,
and time, is necessary to achieve good efficiency,
quantification, and reproducibility. For an efficient
extraction, the solvent must be able to solubilize the
desired analyte while keeping the sample matrix intact.
Most organic solvents and buffered aqueous solutions can
be used in ASE, so the need for extraction and the cost
of the solvent should be considered when developing a
method. ASE uses high temperatures to accelerate the
extraction processes. As the temperature is increased,
solvent diffusivity increases and viscosity decreases,
increasing the extraction rate. The temperatures used
in most ASE applications are in the 50–150 °C range.
Changing pressure has little impact on ASE extraction,
as the main effect of pressure is to maintain the solvent
in its liquid state. Most accelerated solvent extractions
use 1500 psi as the standard operating pressure. Static
extraction cycles are used to introduce fresh solvent
during the extraction process, which maintains favourable
extraction equilibrium. Time is also optimized to obtain a
complete and efficient extraction.
Thermal DesorptionThermal desorption is a well-known sample introduction
technique for GC determination of volatile or semivolatile
organic compounds. For gaseous samples, volatile
organic compounds are collected onto a sorbent, then
thermally desorbed for GC analysis, while volatile and
ROMIL-UpS™ Ultra Purity Solventsfor critical analytical applications
Delivering the ultimate quality for hyphenated techniquesROMIL-UpS™ Ultra Purity Solvents have been developed to meet the demands of today’s hyphenated techniques. Every batch is manufactured, use-tested and packaged to ensure the best possible performance time after time.
Purity. Reliability. Consistency.
When it works well, it’s pure chemistry.
And when it’s pure chemistry, it’s ROMIL.
Discover the world of ROMIL foryourself at www.romil.com
SAMPLE PREPARATION PERSPECTIVES
semivolatile analytes in solid samples
can be determined directly by
thermal desorption.
Thermal desorption has numerous
benefits for analysis of trace-level
volatile and semivolatile organic
compounds. Thermal desorption
performs sample collection and
concentration at the same time. The
use of sorbents enables accurate
and efficient analyses of volatile
organic compounds in large sample
volumes even at low concentration.
Thermal desorption uses heat instead
of solvent to desorb analytes from
the sorbent and transfers the entire
sample to a GC system for analysis.
This technique enables a complete,
fast and solvent-free desorption of
the analytes. Thermal desorption is
a flexible, efficient, and convenient
sample introduction method. It
has myriad applications, including
fragrances and flavours.
Static Headspace ExtractionHeadspace extraction is usually
defined as a vapour-phase extraction,
involving the partitioning of analytes
between a nonvolatile liquid or solid
phase and the vapour phase above
the liquid or solid. In this process, the
sample is placed in a sealed glass vial
with a septum-lined cap. The vial is
then heated to a specific temperature
so that the volatile compounds diffuse
into the headspace above the sample.
After the equilibrium is reached,
the analytes in the headspace are
collected with a gas-tight syringe
and injected into a GC system for
analysis. The extraction of volatile
and semivolatile organic compounds
in solid, liquid, and gas samples
can be achieved by headspace
analysis. This extraction technique is
simple and fast, and it can provide
acceptable sensitivity. Common
applications include analyses of
flavour compounds in beverages and
food products.
Purge-and-Trap ConcentrationThe purge-and-trap method is a
dynamic headspace technique that
involves the purging of inert gas
through a liquid or solid sample,
followed by trapping of the volatile
analytes on a sorbent and desorption
into a GC system for separation and
identification. This method uses
the inert gas to strip the volatile
analytes from the sample matrix
and concentrate them on a sorbent.
Purge-and-trap concentration reduces
matrix effects and increases sensitivity.
This sampling method has been used
extensively in food analysis (32,33).
ConclusionsThe various extraction methods
described here provide an overview of
methods that are widely used in food
safety analysis. Conventional methods
such as Soxhlet extraction,
liquid–liquid extraction, and
Thermal desorption is a flexible, efficient, and convenient sample introduction method. It has myriad applications, including fragrances and flavours.
CHROMacademy is the world’s largest eLearning website
for analytical scientists, containing 1000’s of interactive
learning topics.
Lite members have access to less than 5% of our content.
Premier members get so much more!
We have 1000’s of eLearning topics
SPEMSGCHPLC IR BLS BIO
Find out more about CHROMacademy Premier membership contact:
Glen Murry on +1 732.346.3056 | e-mail: [email protected]
Peter Romillo: +1 732.346.3074 | e-mail: [email protected]
www.chromacademy.com
powered by
669www.chromatographyonline.com
SAMPLE PREPARATION PERSPECTIVES
shake-flask extraction are laborious,
require the use of large amount
of solvents and tedious extraction
steps, and have limited applications.
Modern extraction methods such as
those presented here have numerous
advantages when compared to the
traditional methods, such as shortened
extraction time, reduced solvent and
energy consumption, and improved
extraction efficiency. They are usually
considered to be green techniques
and have been used extensively for
determination of various contaminants
and harmful substances in food
samples.
References(1) L. Wang and C.L. Weller, Trend Food
Sci. Technol. 17, 300–312 (2006).
(2) J.R. Pedersen and J.O. Olsson, Analyst
128, 332–334 (2003).
(3) A. Garrido Frenich, J.L. Martínez
Vidal, A.D. Cruz Sicilia, M.J. González
Rodríguez, and P. Plaza Bolaños, Anal.
Chim. Acta 558, 42–52 (2006).
(4) L. Ramos, E. Eljarrat, L.M. Hernández,
J. Rivera, and M.J. González,
Chemosphere 38, 3141–3153 (1999).
(5) A. Lawal, G.H. Tan, and A.M. Ali, J.
AOAC Intl. 99, 1383–1394 (2016).
(6) C.L. Arthur and J. Pawliszyn, J. Anal.
Chem. 62, 2145–2148 (1990).
(7) H. Kataoka, H.L. Lord, and J. Pawliszyn,
J. Chromatogr. A880, 35–62 (2000).
(8) J. Aulakh, A. Malik, V. Kaur, and P.
Schmitt-Kopplin, Crit. Rev. Anal. Chem.
35, 71–85 (2005).
(9) M.E.C. Queiroz and F.M. Lanças, LCGC
Europe 18, 145–154 (2005).
(10) A.J. King, J.W. Readman, and J.L.
Zhou, Environ. Geochem. Health 25,
69–75 (2003).
(11) C.-H. Xu, G.-S. Chen, Z.-H. Xiong,
Y.-X. Fan, X.-C. Wang, and Y. Liu, TrAC
Trend. Anal. Chem. 80, 12–29 (2016).
(12) J. Li, Y.-B Wang, K.-Y. Li, Y.-Q. Cao,
S. Wu, and L. Wu, TrAC Trend. Anal.
Chem. 72, 141–152 (2015).
(13) T.A. Souza-Silva, E. Gionfriddo, and J.
Pawliszyn, TrAC Trend. Anal. Chem. 71,
236–248 (2015).
(14) F. Soxhlet, Polytechnisches J. 232,
461–465 (1879).
(15) M.L. De Castro and L. Garcıa-Ayuso,
Anal. Chim. Acta 369, 1–10 (1998).
(16) M. Anastassiades, S.J. Lehotay, D.
Štajnbaher, and F.J. Schenck, J. AOAC
Intl. 86, 412–431 (2003).
(17) S.J. Lehotay, R.E. Majors, and M.
Anastassiades, LCGC North Am. 28,
504–516 (2010).
(18) T. Rejczak and T. Tuzimski, Open
Chem. 13, 980–1010 (2015).
(19) D. Veli̷koviDž, D. MilenoviDž, M. RistiDž,
and V. VeljkoviDž, Ultrasonics Sonochem.
13, 150–156 (2006).
(20) Z. J. Dolatowski, J. Stadnik, and D.
Stasiak, Acta Sci. Pol., Technol. Aliment
6, 89–99 (2004).
(21) K. Vilkhu, R. Mawson, L. Simons, and D.
Bates, Innov. Food Sci. Emerg. Techn.
9, 161–169 (2008).
(22) F. Chemat, N. Rombaut, A.-G. Sincaire,
A. Meullemiestre, A.-S. Fabiano-
Tixier,and M. Abert-Vian, Ultrasonics
Sonochem. 34, 540–560 (2017).
(23) E. Thostenson and T.-W. Chou,
Composites Part A: Appl. Sci. Manufac.
30, 1055–1071 (1999).
(24) B. McManus, M. Horn, B. Lockerman,
S. Smith, and G. LeBlanc, LCGC North
Am. 31(11), 22–27 (2013).
(25) H. Wang, J. Ding, and N. Ren, TrAC
Trend. Anal. Chem. 75, 197–208 (2016).
(26) L. Angilillo, M.A. Del Nobile, and A.
Conte, Curr. Opin. Food Sci. 5, 93–98
(2015).
(27) J.W. King, J. Chromatogr. Sci. 27,
355–364 (1989).
(28) C. Turner, C.S. Eskilsson, and E.
Björklund, J. Chromatogr. A 947, 1–22
(2002).
(29) M. Herrero, J.A. Mendiola, A. Cifuentes,
and E. Ibáñez, J. Chromatogr. A 1217,
2495–2511 (2010).
(30) A. Kettle, LCGC North Am. 31(11),
28–33 (2013).
(31) P. Vazquez-Craig and Y. Rico, TrAC
Trend. Anal. Chem. 71, 55–64 (2015).
(32) J.G. Wilkes, E.D. Conte, Y. Kim, M.
Holcomb, J.B. Sutherland, and D.W.
Miller, J. Chromatogr. A 880, 3–33
(2000).
(33) L. Zoccolillo, L. Amendola, C. Cafaro,
and S. Insogna, J. Chromatogr. A 1077,
181–187 (2005).
Changling Qiu is a postdoctoral
associate in the Department of
Chemistry and Biochemistry at the
University of Texas at Arlington,
where she is investigating
vacuum-UV detection for capillary
GC. She earned degrees in food
engineering at universities in China
and completed her Ph.D. in chemistry
at South Dakota State University,
studying extraction development for
food safety applications.
Douglas E. Raynie is Department
Head and Associate Professor in
the Department of Chemistry and
Biochemistry at South Dakota State
University. His research interests
include green chemistry, alternative
solvents, sample preparation, high
resolution chromatography, and
bioprocessing in supercritical fluids.
He earned his Ph.D. in 1990 at
Brigham Young University under the
direction of Milton L. Lee.
method-compliant
instrumentation & accessories
for VOC and SVOC emissions
testing from:
ɵ Building materials
ɵ Spray polyurethane foam
ɵ Automotive components
ɵ Consumer products
ɵ Medical devices
World leaders in
material testingfor product &
thermal desorption
Talk to Markes about
Find out morehttp://chem.markes.com/XR
LC•GC Europe December 2017670
COLUMN WATCH
Why do separation scientists care
about peak shapes so much? The
analysis of peak shape is critical during
the development and optimization
of synthetic approaches of new
stationary phases as well as the quality
of packing for all stationary phases
(1). To the end-user, Gaussian peaks
are more amenable to integration
than non-Gaussian shapes, provide
higher detection sensitivity, and allow
a higher number of peaks within a
given run time—that is, increased
peak capacity. Peak distortion
may also indicate a closely eluted
component. From an instrumentation
designer’s perspective, obtaining a
narrow dispersion of analyte peaks
is an indication of a well-behaving
instrument. Unfortunately, peak shapes
encountered in practice are rarely
ideal in the majority of the separation
modes. The shape of the peak can be
affected by factors such as the column
packing, secondary interactions of the
analyte with the stationary phase, the
connection tubing from the injector
to the detector inlet, the detector
sampling rate, and the nature of the
digital filter (mathematical elimination
of noise) in the chromatographic
software (2). These factors result in a
different efficiency, retention time, and
selectivity or lower overall resolution
than expected for the particle size of a
given support. Table 1 summarizes the
origins of undesired peak shapes with
their underlying causes.
How to Make an Initial Assessment of an Experimental Peak Shape The simplest way to assess the
quality of the chromatographic
signal is to visually inspect the peak
for mirror image symmetry and a
measurement of its width W, which
can be measured at a given height.
These measurements allow the
calculation of two chromatographic
figures of merit: the theoretical plates
(N) and the peak asymmetry. Like an
engine’s horsepower, N indicates the
“horsepower” of a chromatography
column. A larger value of N indicates
that the column can produce narrower
peak widths and can separate more
peaks in a given time window. The
features of a typical peak obtained
in chromatography are labelled
in Figure 1 (3). Figure 1(a) shows
how N can be calculated. Similarly,
Figure 1(b) shows how peak shapes
are measured at various heights using
two popular quantities: the United
States Pharmacopeia (USP) tailing
factor and the asymmetry factor. If the
users assume that peaks are visually
Gaussian (the judgement depends on
the user), the equation to calculate N is
given by equation 1:
N = a 2t
R
W [1]
where tR is the retention time, and W
is the width of a peak at given height.
Because it is possible to measure W at
Peak Shapes and Their Measurements: The Need and the Concept Behind Total Peak Shape AnalysisM. Farooq Wahab1, Darshan C. Patel2, and Daniel W. Armstrong1, 1University of Texas at Arlington, Texas, USA, 2AbbVie
Inc., North Chicago, Illinois, USA
Gaussian peak shapes in chromatography are indicative of a well-behaved system. Such peak shapes are highly desirable from the perspective of column packing technology. From an analyst’s point of view, Gaussian peaks provide improved sensitivity (lower detection limits) and allow ease of quantitation. In practice, one can obtain peaks that tail, front, or concurrently front and tail for reasons such as column packing issues, chemical and kinetic effects, and suboptimal high performance liquid chromatography (HPLC) system plumbing and detector settings. Here, we discuss a number of approaches for peak shape measurement that are available in modern chromatography software, along with their advantages and drawbacks. A new “total peak shape analysis” approach is suggested that facilitates detection and quantifi cation of concurrent fronting and tailing in peaks. Several remediation approaches are proposed that can help chromatographers analyze and improve peak shapes.
LC•GC Europe December 2017672
COLUMN WATCH
various heights, the factor a is adjusted
accordingly. Unfortunately, even for
extremely high efficiency columns,
perfect Gaussian peaks are rarely
observed. Because we are interested
in measuring the actual peak shapes,
the method of moments is the most
accurate measure of peak properties.
It involves slightly tedious calculations,
but several computer data systems
(CDS) readily allow users to calculate
moments as shown in Figure 2.
N =2
m1
m2 [2]
In equation 2, m1 is the centroid or
centre of gravity of the peak, and m2
is the second moment or variance of
distribution of the analyte in time. The
definitions are provided in Figure 2. The
definition of N described in equation 2
does not assume any peak shape,
because the centroid and variance
can be determined for any peak shape
encountered in chromatography (that
is, fronting, tailing, split, shouldering,
horned, and so forth). The main
drawback of moments is that they are
very sensitive to peak start (t1) and
peak end (t2), and noise in the signal
S(t) (Figure 2). Secondly, moments
also depend on the data sampling rate.
Typically, the signal-to-noise ratio (S/N)
should be 200 or above for obtaining
reliable moment values (4). All the first
three moments can be calculated in
Microsoft Excel by estimation of the
equations shown in Figure 2 (4). Slightly
incorrect peak integration could lead to
poor precision; therefore, the moment
analysis for total shape analysis is too
extensive and sensitive for routine work,
despite their easy availability in modern
data acquisition and analysis software.
Let us compare the width-based
measurements of the two peaks
obtained on a 15 × 0.46 cm,
2.7-μm dp C18 core–shell column.
The manufacturer’s quality control
test reports an exceptionally high
plate number (39,000 per column).
When tested, indeed the Gaussian
efficiencies are 38,000 and 39,700
for uracil and phenol, respectively
(equation 1). However, N values
obtained from moment analysis are
24,000 and 26,000, respectively, as
calculated by Agilent’s ChemStation
software (equation 2). There is a
surprising difference of more than
10,000 plates, proving the point that
even the most efficient columns today
Table 1: Origin of peak distortions in chromatography (thermodynamic, kinetic, peak processing, and fluid dynamical reasons)
Origin of Peak Distortion Peak Shape Phenomenological Cause Correctable by User?
Thermodynamic and Kinetic Origins
Stationary-phase
characteristics: carbon
load, particle or pore
size distribution
Variable
Thermodynamics–kinetics;
higher bonding coverage
may slow kinetics; wide
particle size distribution
results in poor bed structure
No: Stationary phase
synthesis issues
Sample concentration
effects (analyte and
other components)
Tailing, fronting, split
tailing and split fronting,
retention time shift
When analyte concentration
exceeds the adsorption
capacity of the stationary
phase/mobile phase
Yes: Lower analyte
concentration
Solvent mismatch between
diluent and the mobile phase
Possibly distorted or split
peaks, viscous fi ngering
The elution strength of the
diluent is signifi cantly stronger
than the mobile phase or
possible immiscibility issues
Yes: Match the diluent
composition with the
mobile phase
Frictional effects of
the mobile phase
Peak distortion, shoulders
in worst case
Radial frictional heating of the
column at high fl ow rates
Yes: Use narrow inner
diameter columns or decrease
the fl ow rate; use still-air–
based temperature control
Fluid Mechanics
Slurry packing process
of columns
All shapes possible because
of axially and radially
heterogeneous bed
Suspension rheology
during column packing
under pressure
No: Can only be changed
by the column packer
Injector–connection tubing,
frits, detector design
Tailing, peak broadening, and
shoulders (in case there are
dead volumes or clogged frits)
Fluid dynamics
Partly yes: Use the shortest
possible and narrower
inner diameter tubing,
low-volume fl ow cell,
zero-dead-volume fi ttings
Mathematical Processing of the Chromatographic Data
Digital fi lters in the
instrument’s software
Symmetric widening or tailing,
never fronting; can produce
dips on chromatogram;
can deceptively make
peaks look more symmetric
(with a Gaussian fi lter)
Mathematical operations
on the raw chromatogram
to decrease noise level
No: The fi lters are
mathematics embedded
in the software; can be
circumvented by collecting
analog output and users
can apply smoothing fi lter of
their choice to improve S/N
673www.chromatographyonline.com
COLUMN WATCH
up with an improved measure to assess
peak shapes. There are several ways
to measure peak asymmetry, many of
which are included in chromatography
data acquisition software for reporting.
The definitions of various peak shape
measurements are shown in Table 2.
Figure 3 shows that USP tailing factor
of peak 1 is 1.22, the asymmetry factor
do not produce pure Gaussian peaks. If
the peaks were perfectly Gaussian, the
plate numbers from equations 1 and 2
would match. Chromatographers are
used to looking at large numbers for
efficiency; most column manufacturers
use the simplified calculation of
efficiency since the method of moments
often results in very low values that do
not reflect favourably on a column’s
performance. The efficiency only
carries the information about the width
of a peak, and nothing about the nature
of its entire shape.
With chiral separations, even
when the peaks may have very
high efficiency, the peaks are often
asymmetric. In the unusual case of
β-blockers or compounds such as
hydantoins on chiral columns, the peaks
have a slight ascending fronting as well
as a tailing. Both of these shapes can
originate because of column packing
or kinetic band broadening effects, or
both. We have aptly named such peaks
Eiffel Tower peaks because the top is
mostly very narrow, yet the ascent and
the descent are rather bent like the
Eiffel Tower (Figure 3) (5). The Gaussian
efficiencies of peaks 1 and 2 in Figure 3
are 10,000 and 6400, respectively;
however, keep in mind that these peaks
are not even visually Gaussian. The
plates are vastly overestimated and
the efficiency by moments tells us the
actual efficiencies are 3300 and 1800
for peaks 1 and 2, respectively. These
points again show that we need to come
100
(a) (b)
tR
tR
Wi
Wi
W
WN
a
a
a
b
b
a
W0.5
W0.5
W0.05
W0.05
W3σ
W3σ
W4σ
W4σ
W5σ
W5σ
Wtan
Wtan
f0.05
2f0.05
Tangent
4
5.54
9
16
25
16
Inflection
Method
2
FWHM
3σ
4σ
5σ
=
==
Asymmetry factor
USP tailing factor
% P
eak h
eig
ht
% P
eak h
eig
ht60.7
10
5
50
32.4
13.4
4.4
Figure 1: (a) Illustration of the features of a Gaussian peak profile. The theoretical
plates can be calculated for various heights with an appropriate factor a.
(b) Definitions of the commonly used measures of peak shape.
Pure
•
•
•
•
•
LC•GC Europe December 2017674
COLUMN WATCH
is 1.33, the symmetry is 1.72, and the
moment-based measure called skew
is 1.64. All of these numbers tell us that
the peaks are tailing in a net fashion.
These methods assign a unique number
to a given peak, an approach that may
not present the full picture. Indeed, they
only indicate the contributions to the
asymmetry that are in excess. A careful
examination of the chromatographic
peaks in Figure 3 reveals that tailing is
coupled with fronting, which is rarely
detected and never quantified. The
USP tailing (T) is the most common
measurement and is required by the
U.S. Food and Drug Administration
(FDA). The FDA recommends a tailing
factor of ≤2. A peak shape T of ~2 is
visually very asymmetric and deformed.
The Concept of Total Peak Shape AnalysisHow can we detect peak deformations
throughout the entire peak rather than
rely on a single value? Such an analysis
for the whole of the peak is beneficial
in troubleshooting the peak shape
problems highlighted in Table 1. We
developed two simple tests to study
complete peak shapes graphically (5).
Both methods are intuitive and can be
used with Microsoft Excel. The first one
is the derivative test and the second
one is the Gaussian test.
The Derivative Test: Taking the
derivative with respect to time of a
given peak is the most straightforward
approach to assess total symmetry
and peak shapes. If S is the
chromatographic signal, then the
derivative is
=S
2 –S
1dSdt t
2–t
1 [3]
Equation 3 merely states that we find
the difference between two consecutive
signal values (S2 and S1) and divide it
by the sampling interval. Like moment
analysis, this peak shape test does
not preassume any peak model.
Table 2: Peak shape measurements and their availability in chromatography data acquisition software
Names Defi nition Data Acquisition Software Details
USP tailing factor
T = W0.05/2f0.05
where W0.05 = peak width
at 5% peak height, and f0.05
= distance from the leading
edge of the peak to the peak
maxima at 5% peak height
Universally present in all
major software. Symmetry
factor (JP) or (EP) is
identical with the tailing
factor (USP). Chromeleon
also calls it “skewness”.
1, perfect symmetry;
<1, net fronting
>1, net tailing
Measured at 5% height
Asymmetry
T = b0.1/a0.1
where a0.1 = distance
from leading edge of the
peak to the peak maxima
at 10% peak height, and
b0.1 = distance from peak
maxima to the trailing edge
at 10% peak height
OpenLAB, Empower,
Atlas, Chromeleon,
ChromNAV, ChemStation
1, perfect symmetry;
<1, net fronting
>1, net tailing
Measured at 10% height
Symmetry
m1+m
2
m3+m
4
mi is the ith moment, where i =
1, represents mean; i = 2
represents the variance; i = 3,
vertical symmetry; and i = 4
is a measure of the
compression or stretching of
the peak along a vertical axis
ChemStationProprietary formula
based on moments (6)
Skew (7)m
3
m23/2
ChemStation, Empower,
many software programs
use their own equations.
0, perfect symmetry
<0, net fronting
>0, net tailing
with respect to the centroid
(mean retention time)
Derivative test =S
2 –S
1dSdt t
2–t
1
Chromeleon
Test of peak symmetry,
no shape assumed
Same absolute value of the
maximum and minimum
indicates the peak is
perfectly symmetric
Gaussian test
Graphical representation,
divides the peak in half and
assumes the top 15–20%
of the peak is a perfect
Gaussian; measures residuals
New approach suggested for
graphical peak shape analysis
Minimum fronting and
tailing residuals
675www.chromatographyonline.com
COLUMN WATCH
Thus any peak shape can be analyzed. The sampling
interval can be determined as follows: If the data are
sampled at 160 Hz (160 data points per second), the
sampling interval is 1/160 s or 1/(60×160) min. The only
requirements of the derivative test is to ensure a high
sampling rate (80 Hz and above), to have low response
time settings (< 0.1 s), and a high signal-to-noise ratio.
When time and the derivative are plotted on the same
axis as the original chromatogram, the derivative
intersects the x-axis at the same position as the
maximum of the original peak (Figure 4). Figure 4(a)
shows how the derivative of a pure Gaussian peak
(ideal chromatography) would look. The maximum and
minimum values are identical, which would be true for
any symmetric peak. If a peak has a very slight tail as in
Figure 4(b), then the left maximum has a larger absolute
value than the right minimum. This tailing is coming from
a very short column (0.5 cm × 0.46 cm). As indicated in
Table 1, the tailing is originating from the column packing
as well as from extracolumn effects. If there is concurrent
fronting or tailing, then the absolute values will depend
on which half of the peak is dominating, as we will show
later. The gist of the derivative test is that if the absolute
values of the maxima and minima of the first derivative
of a peak do not match, the slope of the leading edge of
the peak differs from the slope of the trailing edge and
reveals the presence of asymmetry. For the derivative
test, the excess magnitude of the positive end indicates
tailing and the excess magnitude of the negative end
suggests a fronting element to the peak. The derivative
test is a very sensitive test to detect the peak asymmetry,
and it does not rely on the user to choose peak start
and end time, which makes it independent of integration
errors.
The Gaussian Test: The Gaussian test is a graphical
and quantitative approach for analyzing the total peak
shape and its departure from an ideal Gaussian shape. It
concurrently determines the extent of a peak’s deviation
from a perfect Gaussian form on the leading and the
trailing edge and allows detection and quantitation of
fronting and tailing peaks. As discussed earlier, Table 2
compares what measurement methods are available in
chromatographic software for analyzing experimental
peaks. Ideal chromatography peaks typically follow
the Gaussian peak profile G(t) and amplitude A can be
modelled using equation 4:
=(t
–t
R)
–G (t) A exp
2
[4]
The standard deviation or σ of a peak can be obtained at
any peak height using equation 5:
wH
2√2ln(1H
(
[5]
If we know the experimental standard deviation and
retention time, we can construct a complete Gaussian
peak using equation 4. This method makes the critical
assumption that even in the most distorted peaks, the
upper regions of a peak (80% peak height or above)
follow the Gaussian peak shape; the peak shape
LC•GC Europe December 2017676
COLUMN WATCH
analysis is approached from the top
in this method. The idea behind the
Gaussian test is as follows:
1. Normalize the experimental peak
height to unity. This process simplifies
the visualization and calculations.
2. The top section of most
chromatographic peaks is almost
an ideal Gaussian shape—for
example, the >80% height (Figure 1).
To confirm this hypothesis, the
standard deviation should be
extracted at other heights (for
example, 85%, 90%). The σ
values should closely match.
3. Extract the standard deviation of
the experimental peak at a given
height (>80% peak height) from
equation 5, and determine the peak
maximum (that is, the retention time).
4. Plot a pure Gaussian peak, using
equation 4 with the retention
time and standard deviation
extracted from the experimental
peak. Graphically, superimpose
the ideal peak on a real peak.
5. Find the differences at each point
of the pure Gaussian peak and
the experimental peak. These
values are called residuals. Plot
the residuals on the same graph
against retention time. The residuals
show both if the peak fronts or if
there is a shoulder, or if the peaks
tail and how much they tail.
6. Express the percentage of fronting
residual and tailing residual as the
fraction of the sum of all residuals.
Herein we assume that S/N is high,
and the baseline is not drifting. We
wish to emphasize here that this test is
not simply a curve-fitting procedure,
where the only goal is to minimize
the residuals by the method of least
squares. Although we fit the curve on
the experimental peak, mathematical
constraints are placed in this process
(see steps 2 and 3). To make the
technique easily accessible, a useful
Excel template with prefilled formulas is
available (5) that ultimately automates
the derivative and the Gaussian
test. In the Excel file, the user simply
pastes the retention time as well as a
chromatographic signal (absorbance,
fluorescence, refractive index, and so
forth) and allows Excel to perform the
rest of the calculations.
Figure 5 provides examples where
we show the utility of this Gaussian
test. During initial column packing
experiments, a core–shell material
was producing peak shapes that
were seemingly Gaussian, but only
deceptively symmetric. The column
efficiency was acceptable (for a 5-cm,
2.7-μm column, ~8000–9000 plates).
The USP tailing factor was 0.93
(Figure 5[a]), which can indicate a
relatively symmetric peak shape
with a net fronting; however, a visual
examination showed that there is
coupled fronting and tailing, which
had to be eliminated experimentally
by optimizing the packing conditions.
Modern columns are packed as a
slurry of particulates suspended
in an appropriate solvent. The
concentration of the slurry and the
nature of the solvent affects the
peak shape of the packed column.
440
390
340
290
240
190
140
90
40
-10
1 1.5 2
Retention time (min)
Ab
so
rb
an
ce
5-Methyl-5-phenylhydantoin
USP tailing factor
Asymmetry factor
Symmetry
Skew
1.22
1.33
0.72
1.64
1.44
1.71
0.59
2.32
o
o
NH
NH
2.5
1
1
2
2
Figure 3: Comparison of single valued measures of peak shapes of the enantiomers
of 5-methyl-5-phenyl hydantoin on a chiral column. The USP tailing factor, symmetry,
and skew were obtained using Agilent Chemstation software. The asymmetry factor
was obtained using Microsoft Excel. Manually calculated values may slightly differ,
however the trend is same. Eiffel Tower–shaped peaks were obtained in this case. A
visual examination shows that the peaks have fronting and tailing attributes yet the
single value descriptors only indicate tailing.
The Concept of Moments
Meaning
Peak area
Mean
S(t) dt
(t – m1)2 S(t)
dt
dtVariance
∫
0
190Plates 1. Uracil
38,000
24,000Nmoments
t1
t2
NG
(W0.5
) 39,700
26,000
1
2. Phenol
2
140
=
tS (t)
m0
m0
t1
t2
∫t1
t2
=
=
90A
bso
rb
an
ce
40
-10
0.2 0.4 0.6 0.8 1
Retention time (min)
m0
m1
m2
Calculated by
Figure 2: A comparison of different indicators of peak width. The calculation of
column efficiency using two different approaches for calculating theoretical plates;
NG assumes that the experimental peaks are perfect Gaussian peaks and efficiency is
measured using peak width at half height (W0.5). Moment analysis considers the exact
peak shape. The discrepancy in plates Nmoments shows that even high-efficiency
columns may not produce ideal Gaussian peaks. See moment equations on the left.
S refers to the chromatographic signal. The peak start and the end times are indicated
by t1 and t2.
677www.chromatographyonline.com
COLUMN WATCH
In Figure 5(a), the Gaussian test
(steps 1 to 5) was performed on this
initial packing condition with a low
slurry concentration; the derivative
test confirms the presence of peak
distortion (maximum = 84.839 and
minimum = −82.822). Only the full
peak shape analysis identifies the
problematic regions of the entire peak.
The presence of residuals shows
significant areas of fronting plus tailing
elements despite an indication from
the USP tailing factor that the peak
only fronts. The contribution to peak
distortion is 58% from fronting and
42% from tailing. In another set of
conditions, the slurry concentration
was increased for optimization. As
shown in Figure 5(b), the fronting
element has become negligible and
the left side of the peak has nearly
a perfect Gaussian character. The
value of the derivative still detects
asymmetry (maximum = 86.598,
minimum = –80.282). The peak
shape distortion from the Gaussian
tests shows 10% contribution from
fronting and 90% from tailing. Although
a visual inspection may lead users
to assume that the original peak
shown in Figure 5(a) is desirable over
the optimized condition shown in
Figure 5(b) based on the USP tailing
factor, the Gaussian test disproves this
assumption by showing the problematic
1
0.8
0.6
Sig
nal S (
mA
U)
Sig
nal S (
mA
U)
0.4
0.2
170 3000
2500
2000
-2000
1500
-1500
1000
-1000
500
-500
+1.44
+2876(b)(a)
-1377-1.44
0 Fir
st d
eriv
ativ
e
Fir
st d
eriv
ativ
e
150
130
110
90
70
50
30
10
-10
0
-0.2
15.0 17.5 20.0 22.5 25.0
Time (s) Time (s)
-2.0
-1.5
-1.0
-0.5
0.0
0.5
1.0
1.5
2.0
0.0 0.4 0.8 1.2 1.6 2.0
Figure 4: Illustration of a derivative test for testing peak symmetry (a) for a perfect
Gaussian peak and (b) for a peak obtained on a 0.5-cm column at 5 mL/min on a
custom-built chromatographic system. Note how the absolute values of the maximum
and minimum match for a symmetric peak as in (a). Any symmetric peak shape will
show the same values of maximum and minimum.
The GOLD Standard in
Mycotoxin Testing
High quality reference materials are a must for accurate and reliable results in any analysis.
With Biopure™, Romer Labs offers the broadest range of high quality mycotoxin reference materials currently available.
The stars of the Biopure™ product line are the fully labeled 13C internal standards. Available for all regulated mycotoxins, these calibrants rely on a patented technology, assuring the highest standard of analytical accuracy.
DGOLDGOLDGGGGOOLLDDGGGGGOOOOOLLLDDDDDDDDDDD
For more information visit www.romerlabs.com
LC•GC Europe December 2017678
COLUMN WATCH
regions of the peak with concurrent
fronting and tailing.
Indeed, where resolution from a
closely eluted impurity or a low-level
analyte eluted before the main peak
is of concern, the peak shape shown
in Figure 5(a) may conceal the smaller
impurity despite its seemingly more
symmetric peak shape. In such a case,
despite a larger USP tailing factor, the
peak shape shown in Figure 5(b) would
be desired because it may facilitate
improved resolution of the impurity; the
main peak has practically no fronting.
Note that since uracil is an early eluted
analyte on a narrow-bore column, the
persistent presence of tailing has its
origins in extracolumn connections plus
a nonoptimized slurry solvent.
ConclusionsAn ideal Gaussian peak shows that the
system (column and the instrument)
are well-behaved. When a peak is
asymmetric, single-valued descriptors
of peak shapes such as USP tailing,
asymmetry factor, symmetry, and
skew can be inadequate because
they do not give a complete picture of
the overall peak shape. Researchers
engaged in instrument design and
stationary-phase development need to
analyze peak shapes during synthesis
and column packing. The derivative
test is based on the concept that if a
peak is symmetric, their inflection points
will be mirror images. The Gaussian
test superimposes a Gaussian
model on a normalized peak with
its set of constraints and shows the
problematic regions of the peak. The
standard deviation is extracted from
the upper section (>80% height) of
the peak rather than the conventional
half-height approach. This total
peak shape analysis approach can
uncover the concurrent fronting and
tailing in peaks. The origin, underlying
phenomenological cause, and possible
remedies are highlighted.
References(1) M.F. Wahab, D.C. Patel, R.M.
Wimalasinghe, and D.W. Armstrong, Anal.
Chem. 89, 8177–8191 (2017).
(2) M.F. Wahab, P.K. Dasgupta, A.F. Kadjo,
and D.W. Armstrong, Anal. Chim. Acta
907, 31–44 (2016).
(3) B.A. Bidlingmeyer and F.V. Warren, Anal.
Chem. 56, 1583A–1596A (1984).
(4) P.G. Stevenson, H. Gao, F. Gritti, and
G. Guiochon, J. Sep. Sci. 36, 279–287
(2013).
(5) M.F. Wahab, D.C. Patel, and D.W. Armstrong,
J. Chromatogr. A 1509, 163–170 (2017).
(6) Agilent Technologies, “Evaluating
System Suitability CE, GC, LC and
A/D ChemStation” Revisions: A.03.0x-
A.08.0x.
(7) E. Grushka, J. Phys. Chem. 76,
2586–2593 (1972).
M. Farooq Wahab is currently a
research engineering scientist at the
University of Texas at Arlington (Texas,
USA). He received a postdoctoral
fellowship with Professor Armstrong
after completing his Ph.D. at the
University of Alberta. His research
interests include fundamentals of
separation science, ultrahigh-efficiency
phases, hydrophilic interaction
chromatography, supercritical fluid
chromatography, data processing, and
subsecond separations in LC. For Excel
template and correspondence contact:
Darshan C. Patel received his
Ph.D. from the University of Texas at
Arlington, where he developed new
technologies for ultrahigh-throughput
LC under the direction of Dr. Armstrong.
In his current role, as a senior scientist
at AbbVie Inc., in Process Research
and Development, he focuses on the
development of process analytical
technologies for in-process sampling
and automation.
Daniel W. Armstrong is the Welch
Distinguished Professor of Chemistry
at the University of Texas at Arlington.
Professor Armstrong has received
more than 30 national and international
research and teaching awards.
His current interest involves chiral
recognition, macrocycle chemistry,
synthesis and use of ionic liquids,
separation science, and mass
spectrometry. He has more than 600
publications, one book, and 30 patents.
For correspondence and a copy of the
Excel template email [email protected]
David S. Bell is a manager in
Separations Technology and Workflow
Development at MilliporeSigma (formerly
Sigma-Aldrich/Supelco). With a B.S.
degree from SUNY Plattsburgh (New
York, USA) and a Ph.D. in analytical
chemistry from The Pennsylvania
State University (Pennsylvania, USA),
Dave spent the first decade of his
career within the pharmaceutical
industry performing analytical method
development using various forms of
chromatography and electrophoresis.
During the past 20 years, working
directly in the chromatography industry,
Dave has focused his efforts on the
design, development, and application
of stationary phases for use in HPLC
and hyphenated techniques. In his
current role at MilliporeSigma, Dr. Bell’s
main focus has been to research,
publish, and present on the topic of
molecular interactions that contribute
to retention and selectivity in an array
of chromatographic processes. Direct
correspondence to: LCGCedit@ubm.
com
Retention time (min)
0.60
1.0
0.8
0.6
0.4
0.2
0.0
-0.2
1.0
0.8
0.6
0.4
0.2
0.0
-0.2
1.00
0.95
0.90
0.85
0.80
0.75
0.70
0.65
0.60
0.55
0.50
0.45
0.40
0.35
0.30
0.25
0.20
0.15
0.10
0.05
0.00 Re
sid
ua
ls
Re
sid
ua
ls
1.00
0.95
0.90
0.85
0.80
0.75
0.70
0.65
0.60
0.55
0.50
0.45
0.40
0.35
0.30
0.25
0.20
0.15
0.10
0.05
0.00
0.65 0.70
Normalized
peak
First derivative
Max = 84.839
Min = -82.822
First derivative
Max = 86.598
Min = -80.282
Gaussian
test
Ga
ussia
n t
est
(a) (b)
Ga
ussia
n t
est
Gaussian
test
Normalized
peak
0.75 0.80 0.60 0.65 0.70 0.75 0.80
Retention time (min)
Figure 5: The Gaussian test applied on examples of (a) asymmetric peak (which
deceptively appears symmetric) obtained from the use of a nonoptimal (2.3% w/v)
slurry concentration and (b) improvement of peak shape by using an optimal (16%
w/v) slurry concentration during column packing. The experimental peaks were
obtained using 5 cm × 0.3 cm columns packed with native 2.7-μm superficially porous
particles at different slurry concentrations. The black curve shows the raw data, and
the superimposed blue dashed lines show the ideal peak shape. The residuals in red
show the regions where the experimental peak does not match the ideal peak.
679www.chromatographyonline.com
QUESTIONS OF QUALITY
Key performance indicators
(KPIs) or metrics are used by
many laboratories to measure the
operational and quality performance
of processes and the laboratory
itself: If you can’t measure it, you
can’t control or manage it. Metrics
are also a requirement of the ISO
9001 quality standard and six sigma
initiatives for continuous improvement
of processes. In this instalment of
“Questions of Quality”, we look at
the regulatory requirements for data
integrity metrics so that laboratory
managers can understand how their
laboratories are performing with
respect to data integrity and identify
where they may have potential
problems. For those that don’t
understand laboratory metrics, let us
start with a primer.
Understanding Laboratory MetricsConsider a laboratory analysis, a
supervisor or manager needs to
know how many samples are being
analyzed and how long it takes to
analyze them. They need to know
if there are any bottlenecks and
if they have enough resources
assigned for the job. Metrics can
provide this information to measure a
process or activity. If used correctly,
metrics can help staff understand
the performance measures that a
laboratory is judged against. Some
common laboratory metrics are
shown in Table 1.
Metrics Must be Generated AutomaticallyA key requirement for collection of
metrics is that the process must be
automatic. Why is this? The simple
reason is that if humans are used
to collect and collate the data
manually it becomes an error-prone,
tedious, and labour-intensive
process, and could also be subject to
falsification.
Automatic metric generation is the
only way to create metrics that are
timely, accurate, and repeatable. This
is where suppliers of chromatography
data systems and other laboratory
informatics applications could help
by implementing applications that
generate both general and data
integrity metrics automatically.
FDA Quality Metrics GuidanceThe Food and Drug Administration
(FDA) are also focusing on quality
metrics as a way of identifying
facilities that have lower risks. The
Agency issued two draft guidance
for industry documents on quality
metrics in July 2015 and November
2016. To monitor the performance
of QC laboratories, the Agency has
selected out-of-specification (OOS)
results. The metric chosen was
invalidated out-of-specification rate
(IOOSR), which is defined as the
number of OOS test results for lot
release and long-term stability testing
invalidated (1).
From a laboratory perspective,
knowing the OOS rate is an important
criterion for both quality and data
integrity. One of the key questions
to ask when auditing or inspecting a
laboratory is the OOS results for the
past six months. However, answering
“we don’t have any” can result in a
regulatory surprise:
Since beginning manufacturing
operations in 2003, your
firm has initiated a total
of 0 out-of-specification
investigations for finished
product
FDA 483 Observation,
November 2014.
As analytical procedures are
subject to variation we expect
to see not only OOS but also
out-of-expectation (OOE) and
out-of-trend (OOT) results. There
is a specific EU GMP regulation
that requires trending of laboratory
results:
6.9 Some kinds of data
(e.g. tests results, yields,
environmental controls) should
be recorded in a manner
permitting trend evaluation.
Any out of trend or out of
specification data should be
addressed and subject to
investigation (2).
6.16 The results obtained
should be recorded. Results of
parameters identified as quality
attribute or as critical should be
Data Integrity Metrics for ChromatographyMark E. Newton1 and R.D. McDowall2, 1Eli Lilly and Company, Lilly Corporate Center, Indianapolis, Indiana, USA, 2R.D. McDowall Ltd, Bromley, Kent, UK
The authors discuss metrics for monitoring data integrity within a chromatography laboratory, from the regulatory requirements to practical implementation.
Key performance indicators (KPIs) or metrics are used by many laboratories to measure the operational and quality performance of processes and the laboratory itself: If you can’t measure it, you can’t control or manage it.
LC•GC Europe December 2017680
QUESTIONS OF QUALITY
trended and checked to make
sure that they are consistent
with each other…… (2).
What would be an OOS rate in
a regulated laboratory? If you
don’t know there might be some
issues during the next inspection.
Why Metrics for Data Integrity?We now need to ask why data
integrity metrics are important.
Put simply, it is now a regulatory
expectation as we can see from the
PIC/S guidance PI-041 in Table 2
(3). There is an expectation that
data integrity metrics of processes
and systems are collected for
management review. Of course, there
is the implicit expectation to act if the
metrics indicate an activity or trend
that has the potential to compromise
data integrity.
Metrics Lead Behaviour?The second row of Table 2 states that
you should be careful when selecting
a KPI or metric so as not to result in
a culture in which data integrity is
lower in priority. Let us consider this
statement further. Take the example
of turnaround time (TAT) in Table 1.
Imagine that the laboratory target
for TAT is set for 10 days and over
the past few months the target has
been missed and the laboratory
manager is not pleased. The sample
you are starting to analyze has been
in the laboratory for nine days and
your analysis will take two days to
perform. You will miss the TAT target
unless…
This is where quality culture,
management, and data integrity
collide and it is where the PIC/S
guidance caution about metrics
comes into play—can some metrics,
intended to monitor activities,
become the means of inducing
behaviours that compromise data
integrity?
One other factor to consider
here: As you monitor some actions,
they will improve because they get
attention, but this can be at the
expense of other activities that
are not being monitored. Be
aware of what you do not monitor
as well, for example, measuring
metrics on quality control production
TAT can cause stability tests to be
neglected.
Overview of Data Integrity Metrics in An OrganizationThis is an evolving subject but
the aim in this column is an
overview of data integrity metrics
that could be generated as part of
a data integrity programme of work
as well as for routine work. Let us
look at where in an organization
metrics for data integrity could
be generated (Figure 1). In this
column, we will not consider
production manufacturing metrics.
Although focused in a quality control
laboratory, the same principles
apply to an analytical development
laboratory in R&D. The scope of
data integrity metrics can cover
Automatic metric generation is the only way to create metrics that are timely, accurate, and repeatable.
Data IntegrityMetrics
Data IntegrityMetrics
Data Governance orData Integrity
Policy
• DI Policy Status• Number of Staff Trained• Understanding Measured• Audit Findings & Management Review
• Computer Systems and Manual Processes Assessed• Short-Term Remediation Completed Versus Plan• Long-Term Solutions
• Process and Analytical Dev• Manufacturing• Quality Control• Outsourcing CMO and CRO
• DI Audits• Audit Findings and CAPA• For Cause Data Integrity Audits• Data Integrity Investigations
Assessment andRemediation
R&D andProduction
Batches
Quality AssuranceOversight
Figure 1: Scope of data integrity metrics within an organization.
681www.chromatographyonline.com
QUESTIONS OF QUALITY
the four main areas within an
organization:
• Data governance such as
data integrity policy(ies) and
• Development and production
activities including outsourcing
of laboratory analysis;
• Quality assurance oversight.
the associated training;
• Assessment and remediation of
laboratory manual processes
and computerized systems;
24th - 26th January 2018, Cardiff, UK
www.ilmexhibitions.com/HTC
HTC 15 will take place in the historic city of Cardiff and will cover all fundamental aspects, instrumental developments and applications of hyphenated chromatography techniques. These include hyphenation between chromatography and new detection systems, multi column (multi-dimensional) chromatography, and the coupling between sample preparation and separation techniques. Many aspects of chromatography are addressed, including automation, miniaturisation, and micro fabricated analytical devices, high pressure and high temperature LC.
Besides the core scientifi c programme, the symposium will host an attractive technical exhibition where companies active in the fi eld will present their latest developments in instrumentation, software and applications related to the hyphenation of separation, detection and sample preparation techniques topped with technical seminars.
Tel: +44 (0) 1727 858840 s�%MAIL��[email protected]
REGISTRATION NOW OPEN15th International Symposium on Hyphenated Techniques in Chromatography and Separation Technology
Supporting Organisations:
Overall Process Data Integrity MetricsTurnaround Time (TAT)
% Right First Time
% Out of Specification (OOS) & Invalidated OOS (IOOSR) Results
Sampling Preparation Acquisition Reviewing ReportingIntegration and
Calculation
Sample Preparation
Records
Manual Data
Input to CDS
Multiple Runs
Identical Data
Duplicate Files
Second Person Review
Focus on Work Correctly Performed
Focus on Potential Poor Data Management Practices
% Manual
Integration or
Intervention
Time to Review
Analyst Throughput
Post Release Activity
Figure 2: Some data integrity metrics for laboratory processes.
LC•GC Europe December 2017682
QUESTIONS OF QUALITY
DI Policies and Assessment and Remediation of Processes and SystemsThe first two areas from Figure 1 to
consider for data integrity metrics are
data integrity policies and procedures
and assessment and remediation of
processes and systems.
Data integrity policies and
procedures should include the
following:
• DI policy outlining company
expectations, behaviours, and
culture;
• Good documentation practices SOP
covering paper, hybrid and electronic
processes and systems;
• Second person review SOP including
audit trail and e-records review;
• Updating current procedures
to include validation processes
that ensure the integrity of
records and detection of
improper actions;
• Metrics for policies and procedures
in Table 3 focus on the training
and its effectiveness. This is
a subject where a read and
understand approach is not tenable
and demonstrable evidence of
understanding is required.
Assessment of processes and
systems should cover the following
activities:
• Assessment of new systems for
integrity gaps prior to purchase;
• Identification of existing
computerized systems and
paper processes for assessment
including access databases and
spreadsheets and their listing in an
inventory;
• Processes and systems are
prioritized by risk and record impact;
• Assessment of systems and paper
process;
• Data flow mapping to highlight
generated records and their
vulnerabilities;
• For each process and system there
Table 1: Some common laboratory key performance indicators (KPIs) or metrics
Metric Measurement
Turnaround time
(TAT)
• Time of sample receipt to delivery of final report or certificate of
analysis
• If required, further subdivision is possible, for example:
Time between receipt and start of analysis
Time for the actual analysis
Time for review of results
Time for report writing
Chromatograph
use
• Percentage utilization per instrument (weekends included or excluded)
• Machine breakdowns: time off-line
• Samples analyzed between maintenance periods
Injections • Number of injections per chromatograph per run
• Peaks analyzed with automated versus manual integration
Table 2: Requirements for data integrity metrics from PIC/S PI-041 guidance (3)
PIC/S PI-041 Section
Requirements
6.4 Modernizing
the pharmaceutical
quality system
6.4.2 The company’s Quality Management System should be able
to prevent, detect, and correct weaknesses in the system or their
processes that may lead to data integrity lapses.
The company should know their data life cycle and integrate the
appropriate controls and procedures such that the data generated
will be valid, complete, and reliable.
Specifically, such control & procedural changes may be in the
following areas:
Quality metrics and reporting to senior management
Regular
management
review of quality
metrics
6.5.1 There should be regular management reviews of quality metrics,
including those related to data integrity, such that significant issues are
identified, escalated, and addressed in a timely manner.
Caution should be taken when key performance indicators are selected
so as not to inadvertently result in a culture in which data integrity is lower
in priority.
Table 3: Data integrity metrics for policies and assessment and remediation of paper
processes and computerized systems
DI Subject Area Suggested Metrics
Data governance
integrity policy
and associated
procedures
• DI policy status: drafting, issued, in-revision versus schedule
• Staff training in DI policies and procedures
Staff training: number completed versus plan
Number of staff passing first time
• Metrics on HR issues related to DI
Assessment of
processes and
systems
• Number of systems and processes, risk priority
• Percentage of high-, medium-, and low-risk systems and process
assessments completed versus plan
Remediation of
processes and
systems
• Percentage of short-term quick fixes completed versus plan: high-,
medium-, and low-risk processes and systems
• Longer term remediation projects completed versus plan: high-,
medium-, and low-risk processes and systems
• Manual processes: percentage of uncontrolled blank forms replaced
by a computerized process or control over master templates and
controlled blank forms
As you monitor some actions, they will improve because they get attention, but this can be at the expense of other activities that are not being monitored.
683www.chromatographyonline.com
QUESTIONS OF QUALITY
should be short- and long-term
remediation plans;
• The metrics shown in Table 3 are
focused on the completion of
assessments for high-, medium-,
and low-risk systems.
Remediation plans executed for
processes and systems:
• Implementation order of both
short- and long-term remediation
should be based on inventory
(total) risk;
• Short-term quick fixes to remediate
critical risks to records, for
example, eliminate shared user
identities or restrict access to
data directories to meet ALCOA+
criteria;
• Longer term remediation (for
example, update the system to
have a database or effective audit
trail, or replace hybrid systems with
electronic systems);
• Replacement of manual processes
using uncontrolled blank forms with
either a computerized process, or
controlled master templates and
blank forms with reconciliation.
Table 4: Some data integrity metrics for laboratory processes
DI Subject Area Suggested Metrics
Reprocessed
sample analysis • Sample results that were released, withdrawn, and released again.
(This is usually done for errors discovered after release)
Injections • Run times longer than average
• Run times shorter than average
Exported runs
• A missing metric from some chromatography data systems is a list
of all runs exported to LIMS or ELN systems. This missing piece
can permit users to run a sample multiple times and release their
“favourite”. A report of runs with 1–2 injections per vial can help
detect some of this behaviour, but in general, accounting for all
chromatography injections is an issue.
Runs stopped or
aborted
• This metric could indicate behaviour where the analyst does not like
what they see as they look at the run in progress, so they stop data
collection.
Table 5: Data integrity metrics for quality assurance oversight
DI Subject Area Suggested Metrics
QA oversight
metrics
• DI audits progress against plan for scheduled and unscheduled
audits
• Number of DI investigations
• CAPAs raised from audits and investigations
Number and classification of findings
Number of CAPAs closed versus planned completion
CAPA effectiveness on follow-up
Might consider some time metrics here: for example, average, min, max
time to release test results, sorted by method
Our Nano- and Micro-UHPLC gradient pumps support a vast range of split-less �ow rates. Methods can be scaled from standard- to micro-, cap-, and nano-HPLC while using the same pump all the way, thereby achieving more than 700x sensitivity gain. Highest-resolution linear piston drives and �ow sensors allow for very precise gradient control even at the highest pressures and lowest �ow rates. Thanks to the unique pump head design
www.prolab.ch/uhplc
Christoph-Merian-Ring 31CH-4153 ReinachSwitzerlandwww.prolab.chwww.prolab.ch/ZircoFit
The Fitting System of Choice forMaximum Flexibility.
Easily assemble your custom UHPLC capillaries in minutes
Fit 10 to 100 �m ID, 365 �m OD fused silica capillaries into 1/16" ports of valves and columns
Starter Kit containing ZircoFit sleeves, crimping tool, capillaries, and �ttings in a handy tool case
20,000PSI
UH
PLC PRESSU
RE
-R
A T E D
-
Watch the short demoon www.zirco�t.com
The World's Finest UHPLC Pumps.
using active valves, it is possible to completely automate all functions, individually pre-compress eluents for short equilibration times, and perform diagnostic and maintenance tasks with no manual intervention. Complete with comprehensive LC/MS system interface options, it is a truly outstanding UHPLC system with no match on the market today.
LC•GC Europe December 2017684
QUESTIONS OF QUALITY
Metrics for progress of short-term
and long-term remediation versus
plans are shown in Table 3.
Laboratory Data Integrity MetricsWe can see some of the data integrity
metrics that could be generated in
Figure 2.
Preliminary Considerations:
• Most data integrity reports can
provide only points of concern. It
is rare where 1 record = 1 data
integrity breach. You must carefully
select what areas will be monitored
by metrics because too many
can bury you in data. Start small,
evaluate the data, and adjust as
necessary.
• Not all metrics are equal; therefore,
their review should not be equal.
• It is nearly impossible to generate
and analyze metrics for manual
or hybrid activities. For example,
you could detect missing files
in a system by comparing the
instrument use log (paper) against
a report of sample IDs from all
runs stored on a computer system.
However, this approach is too
tedious to be done on a regular
basis and is best left for either a
data integrity audit or investigation.
In contrast, sample IDs in a
laboratory information management
system (LIMS) or electronic
laboratory notebook (ELN) could
be compared to the data files on
laboratory data systems using an
automated script. Owing to the
time involved in their preparation,
manual reports should be limited
to specific scenarios over a
specific time period (for example,
detecting fraud from a single
suspected individual as part of an
investigation).
• There are some activities that are
just difficult to catch with a report.
For example, once a user changes
the naming convention on repeat
injections, you will probably not
detect it on a routine report. And
trying to catch everything that
could be done is like testing every
possible branch of a program. It
could be done if you had infinite
resources.
• Some things can’t be reported.
For example, chromatographers
can delete data on a balance or
other simple analytical system.
The data existed outside the system
and so the records are gone. This
is where data capture to a central
informatics system can improve
the situation compared with the
following citation: “He said that
he recalibrated the balance and
prepared new documentation, and
subsequently discarded the original
record. Furthermore, we learned
that additional original calibration
records of other balances had
similarly been discarded” (4).
• Metrics could also include
assessment of the performance of
individuals, for example, measuring
the time to complete a method
for several different people and
looking for someone working too
quickly. This is where regulatory
compliance can clash with the
works council—who are concerned
with protecting the rights of
workers— in some countries.
The balance between ensuring
data integrity of the organization
producing and marketing
pharmaceutical products needs
to be balanced with the rights of
individuals. However, the regulatory
expectation is that processes must
ensure the integrity of records
generated.
Some metrics for laboratory
processes are shown in Table 4
and cover the main areas of
chromatographic analysis.
Quality Assurance Data Integrity (DI) MetricsThe main areas for quality assurance
oversight in data integrity are DI
audits and investigations and the
resulting corrective and preventive
actions (CAPAs) that are raised
following them. Typically, there will
be:
• DI audits following a schedule
of both announced and
unannounced visits for all
processes, systems, and areas
of an organization. The reason
for unannounced audits is to see
a better picture of work carried
out although there is a risk of
disrupting work;
• Data integrity investigations raised
either by an audit or inspection
finding or a staff concern;
• CAPAs arising from audits or
investigations, close out versus
planned completion, and how
effective each action plan was.
Management Review of DI MetricsData integrity metrics need to be
reviewed by management because
they are responsible for the whole of
the quality management system. The
review should be formal with minutes
kept of the meeting and action items
raised and their progress monitored.
This is especially true for high risk or
impact systems—along with rapid
implementation of short-term fixes
to ensure any major data integrity
gaps are remediated. Demonstrable
progress is important and management
activity in this area is best evidenced
by actions and not words. Management
review and follow-up emphasizes the
importance of data integrity to the
organization and ensures that process
and resource bottlenecks are exposed
and removed.
It’s Déjà Vu All Over Again! For those with short memories, data
integrity is the third major IT systems
improvement program that has faced
the pharmaceutical industry over the
past 20 years, the other two being
Year 2000 (Y2K) and electronic
records and signatures (Part 11)
assessment and remediation. Is
the pharmaceutical industry going
to make the same mistakes again?
Let us explore this question. The
Y2K program was simply replacing
applications and operating systems
that could handle dates past 31st
December 1999. Typically, it was a
case of updating, rather than process
improvement, to complete the work
before the deadline; this was a
technical project with a fixed due
date.
In contrast, a 21 CFR 11
assessment and remediation program
was an opportunity to upgrade and
provide substantial business benefit
by changing the business process to
Data integrity metrics need to be reviewed by management because they are responsible for the whole of the quality management system.
685www.chromatographyonline.com
QUESTIONS OF QUALITY
use electronic signatures and eliminate paper. However,
not many laboratories took advantage of this approach
and simply replaced noncompliant with technically
compliant software.
Is the industry going to repeat the Part 11 behaviour?
Reading the various guidance documents (3, 5–9),
you see the storm clouds on the horizon; tight and
bureaucratic control of blank forms and discouraging use
of hybrid systems. The message is clear—get rid of paper
or control it rigorously. The cost of electronic management
is steady or declining, but the cost of using paper records
is rising. Consider not just the physical storage cost but
also the time to access reports and trend data—paper’s
management cost is considerable and highly labour
intensive. Conversion to electronic records is the only
option for the pharmaceutical industry.
An alternative view for data integrity remediation
is seeing it as a second chance to get Part 11 right
by looking at the intent rather than the letter of the
regulation. Seen in this way, industry can both align
with regulators and position themselves for
productivity—and data integrity—improvements in their
processes.
However, many organizations complain that that this
will cost money. Yes, but what is the impact on the
organization’s cash flow if every batch can be released
a few days earlier? Do the sums and then put in your
upgraded systems project proposals.
SummaryMetrics can be used to monitor for some potential
integrity issues but not all. Care should be taken
to ensure that metrics do not drive behaviours that
compromise data integrity. To be sustainable,
accurate, and timely, collection of metrics must be
automated.
It is essential that metrics enable management
monitoring of data integrity because this is a key part
of success for an overall data governance and data
integrity programme in an organization. However, not
all metrics are equal, they need to be chosen carefully.
Use long-term data integrity remediation as an
opportunity to also improve productivity in laboratory
processes.
References(1) US Food and Drug Administration, Guidance for Industry
Submission of Quality Metrics Data, Revision 1 (FDA,
Rockville, Maryland, USA, 2016).
(2) EudraLex, Volume 4 Good Manufacturing Practice
(GMP) Guidelines, Chapter 6 Quality Control (European
Commission, Brussels, Belgium, 2014).
(3) PIC/S, PI-041 Draft Good Practices for Data Management
and Integrity in Regulated GMP / GDP Environments.
Pharmaceutical Inspection Convention/Pharmaceutical
Inspection Co-Operation Scheme, Geneva, Switzerland (2016).
(4) FDA Warning Letter Sun Pharmaceuticals, (Food and Drug
Administration, Rockville, Maryland, USA, 2014).
(5) MHRA, GMP Data Integrity Definitions and Guidance for Industry
2nd Edition. Medicines and Healthcare Products Regulatory
Agency, London, UK (2015).
(6) US Food and Drug Administration, Draft Guidance for Industry
Data Integrity and Compliance with cGMP (Silver Spring,
Maryland, USA, 2016).
(7) WHO, Technical Report Series No.996 Annex 5 Guidance on
Good Data and Records Management Practices. World Health
Organization, Geneva, Switzerland (2016).
(8) MHRA, GxP Data Integrity Definitions and Guidance for
Industry, Draft version for consultation July 2016. Medicines
and Healthcare Products and Regulatory Agency,
London, UK, 2016.
(9) EMA, Questions and Answers: Good Manufacturing Practice:
Data Integrity. 2016; Available from: http://www.ema.europa.eu/
ema/index.jsp?curl=pages/regulation/general/gmp_q_a.jsp&mid
=WC0b01ac058006e06c#section9.
Mark E. Newton is a laboratory informatics QA
representative at Eli Lilly and Company, Indianapolis,
Indiana, USA. He is also a co-lead of the ISPE/GAMP Data
Integrity Interest Group.
“Questions of Quality” editor Bob McDowall is Director
at R.D. McDowall Ltd., Bromley, Kent, UK. He is also a
member of LCGC Europe’s editorial advisory board. Direct
correspondence about this column should be addressed
to the editor-in-chief, Alasdair Matheson, at alasdair.
It is essential that metrics enable management monitoring of data integrity because this is a key part of success for an overall data governance and data integrity programme in an organization.
LC•GC Europe December 2017686
PRODUCTS
Procainamide glycan labelling
Procainamide labelling permits glycan
identifi cation by either MS or
(U)HPLC, and because of its improved
ionization efficiency compared to 2AB
labelling it can permit identifi cation of
minor glycans (<1% relative peak area)
by ESI-MS. Ludger’s procainamide
labelling kit, LT-KPROC-VP24, is
suitable for N-glycans, O-glycans,
GSL-glycans, heparin, or any
sugar with a reducing terminus.
Procainamide-labelled purifi ed glycan
standards are also available.
https://www.ludger.com/procainamide
Ludger Ltd, Oxford, UK.
FID gas station
The VICI fl ame ionization detector (FID)
gas station combines the reliability
of the VICI DBS hydrogen and zero
air generators into one compact and
convenient package. Available in high
and ultrahigh purity for all GC detector
and carrier gas applications. The
generator is available in two styles,
fl at for placement under a GC, or the
Tower. Available in H2 fl ow ranges up to
1 L/min and 11 bar.
www.VICIDBS.com
VICI AG International, Schenkon, Switzerland.
Thermal desorber
Designed for material emissions-,
fl avour-, and air analysis, the TD
3.5+ processes 3.5” tubes or Gerstel
plus tubes for enhanced recovery
and sensitivity. The liner-in-liner
design replaces the transfer line
reducing analyte loss and memory
effects. Up to 240 samples can
be processed automatically. In
combination with DHS 3.5+, dynamic
headspace up to 1 L volume is
performed.
www.gerstel.com
Gerstel GmbH & Co. KG,
Mülheim an de Ruhr, Germany.
(U)HPLC method development
DryLab 4 is a (U)HPLC method
development and optimization
software that predicts
chromatograms under a much
wider range of experimental
conditions than would be
possible in the laboratory,
according to the company.
DryLab can reportedly
quickly and easily determine how a specifi c separation
behaves as the user simultaneously varies multiple
method parameters, such as pH, temperature, buffer
concentration, and many more.
www.molnar-institute.com
Molnár-Institute, Berlin, Germany.
HILIC columns
Hilicon has launched its novel
iHILIC-Fusion, iHILIC-Fusion(+), and
iHILIC-Fusion(P) UHPLC/HPLC HILIC
columns, based on spherical silica
or polymer particles. According to
the company, their surface-bonding
technologies provide customized
selectivity, high separation efficiency,
and ultra-low column bleeding. The columns are
versatile and can be used for the analysis of polar
compounds in “-omics” research, pharmaceutical
discovery, food and beverage, clinical diagnostics, and
environmental monitoring.
www.hilicon.com
Hilicon AB, Umeå, Sweden.
Microchambers
Markes’ Micro-Chambers offer
a compact, stand-alone unit for
the rapid, method-compliant
screening of chemicals emitted
from a wide variety of products,
materials, and foodstuffs. They are used in numerous
industries from automotive, building products, and
consumer goods, to food and drink. For more information
please watch this video:
http://chem.markes.com/UCTE
Markes International Ltd., Llantrisant, UK.
687www.chromatographyonline.com
PRODUCTS
MALS detector
The μDAWN is,
according to the
company, the world’s
fi rst multi-angle light
scattering (MALS)
detector that can be
coupled to any UHPLC
system to determine
absolute molecular weights and sizes of polymers, peptides,
and proteins or other biopolymers directly, without resorting
to column calibration or reference standards. The WyattQELS
Dynamic Light Scattering (DLS) module, which measures
hydrodynamic radii “on-the-fl y”, reportedly expands the
versatility of the μDAWN.
www.wyatt.com
Wyatt Technology, Santa Barbara, California, USA.
HPLC system
The VWR Hitachi Chromaster delivers highly
reliable results, according to the company.
The high precision delivered by the pump,
the low carryover and high precision of the
autosampler, the stability of the column oven,
and the sensitivity of the detectors combined
contribute to this reliability. Furthermore, the
modular nature of the Chromaster system
allows a multitude of confi gurations to suit all
needs.
https://vwr.com/chromaster
VWR International GmbH, Darmstadt,
Germany.
Portable detectors
Biotech in co-operation with
Runge are proud to announce
a unique series of portable
pocket-sized detectors for liquid
chromatography. The product
line consists of a photometer, a
fl uorimeter, and a conductivity
meter that can be combined
depending on the situation.
Light sources, fi lter, and measuring cell can be adapted
to changing tasks. With these devices you can take the
detector to the measurement site and not vice versa.
https://www.biotech.se/products/detectors/mikron/
Biotech AB, Onsala, Sweden.
LC solvents
Romil-UpS ultra LC solvents
have been developed to
meet the demands of today’s
hyphenated techniques.
Each batch is produced using
exacting purifi cation techniques,
use-tested, and supplied in
bottles specially pre-treated to maintain ionic impurities
at trace levels, to ensure the best possible performance,
according to the company.
www.romil.com
Romil Ltd, Cambridge, United Kingdom.
Full details upon request
Please send applications by email to: [email protected]
Vacancy for Analytical HPLC Sales Engineer
Verulam Scientific are an expanding specialist sales and service company offering state of the art analytical instruments in laboratory environments. Due to new opportunities we have a vacancy for a technical sales engineer responsible for promoting instrument sales in electrochemistry, molecular imaging and sample preparation techniques.
Responsibilities:t� �.BOBHFNFOU�PG�FYJTUJOH�DVTUPNFST�BOE�HFOFSBUJPO�PG�OFX�BDDPVOUTt� �%JSFDU�SFTQPOTJCJMJUZ�GPS�TBMFT�OFHPUJBUJPOT�BOE�MJBJTPO�XJUI�TVQQMJFSTt� �4VQQPSU�UIF�EFWFMPQNFOU�PG�DVTUPNFS�BQQMJDBUJPOTt� �1BSUJDJQBUJPO�BU�DPOGFSFODFT�BOE�FYIJCJUJPOTt� �7JTJUJOH�DVTUPNFS�TJUFT�PO�B�SFHVMBS�CBTJT�UP�HJWF�TBMFT�QSFTFOUBUJPOT �
perform instrument demonstrations and support customers on the use of their equipment
t� �%FWFMPQNFOU�PG�OFX�CVTJOFTT�WFOUVSFT
Requirements:t� �4VJUBCMF�DBOEJEBUFT�TIPVME�IPME�B�EFHSFF�PS�FRVJWBMFOU�JO�
chemistry, biology or related disciplinet� �"U�MFBTU�UXP�ZFBST�PG�FYQFSJFODF�XPSLJOH�XJUI�)1-$�JOTUSVNFOUBUJPO�
in a laboratory and/or sales environmentt� �"WBJMBCJMJUZ�UP�USBWFM�UISPVHIPVU�UIF�6,�BOE�PDDBTJPOBMMZ�PWFSTFBT�
for training and business meetings
We offer you:t� �4BMBSZ�CBTFE�PO�FYQFSJFODF�CVU�JO�UIF�SFHJPO�b�����,�05&t� �"�DPNQFUJUJWF�QBDLBHF�PG�CFOFmUT �JODMVEJOH�DBS�BMMPXBODF �
laptop and phone
HPLC
The classified directory from -$r($�&VSPQF To advertise in this section call Elizabeth McLeantoday at +44 (0) 151 353 3527 or email [email protected] to reserve your space
Supplies & Services
LC•GC Europe December 2015688
EVENT NEWS
24–26 January 201815th International Symposium
on Hyphenated Techniques In
Chromatography and Separation
Technology (HTC-15)
Cardiff City Hall, Cardiff, UK
E-mail: [email protected]
Website: www.ilmexhibitions.com/
htc
18–21 February 201834th International Symposium
on Microscale Separations and
Bioanalysis
Rio de Janeiro, Brazil
E-mail: [email protected]
Website: www.msb2018.org
26 February–1 March 2018Pittcon Conference and Expo
(Pittcon 2018)
Orange County Convention Center,
Orlando, Florida, USA
E-mail: [email protected]
Website: https://pittcon.org/
pittcon-2018/
12–14 March 2018The 10th Conference of the World
Mycotoxin Forum
Amsterdamn, The Netherlands
E-mail: WMF@bastiaanse-communi-
cation.com
Website: www.worldmycotoxinforum.
org
10–13 April 201826th International Trade Fair for
Laboratory Technology, Analysis,
Biotechnology, and Analytica
Conference (Analytica 2018)
Munich, Germany
E-mail: [email protected]
Website: www.analytica.de/index-2.
html
17–18 April 2018International Scientific
Conference Ion Chromatography
and Related Techniques
Zabrze, Poland
E-mail: rajmund.michalski@ipis.
zabrze.pl
Website: http://ipis.pan.pl/en/
Please send any upcoming event
information to lewis.botcherby@
ubm.com
42nd International Symposium on Capillary Chromatography and the 15th GC×GC Symposium
The 42nd International Symposium on
Capillary Chromatography (ISCC) and 15th
GC×GC Symposium will be held at the Palazzo
dei Congressi, Riva del Garda, Italy, from
13–18 May 2018.
The International Symposium on Capillary
Chromatography (ISCC) has established its
reputation as a leading forum for microcolumn separation techniques. From
the first meeting in Hindelang in 1975, the most important developments in
capillary gas chromatography (GC), microcolumn liquid chromatography (LC),
and electromigration techniques have been presented in this symposium
series. The format and the atmosphere of the 42nd meeting will be similar to
the previous meetings, however, this year there will be a particular emphasis
on mass spectrometry (MS).
The six-day event will feature recent findings from leading academic and
industrial experts in the form of lectures and posters revealing the most
recent advances in pressure- and electrodriven microcolumn separation
techniques, and comprehensive two-dimensional gas chromatography
(2D GC). As previously mentioned a particular emphasis will be directed
to comprehensive separation technologies that are combined with various
forms of MS from unit-mass to high resolution, and from single- to hybrid
analyzers. The conference will offer sessions on capillary GC, microcolumn
LC, electromigration methods, and microfabricated analytical systems,
which are expected to cover lab-on-a-chip, column technology, coupled
and multidimensional techniques, comprehensive techniques, hyphenated
techniques, sampling and sample preparation, trace analysis, and automation.
Application sessions include environmental applications, energy, petrochemical,
and industrial applications, biomedical and pharmaceutical applications, and
the analysis of natural products, food, flavours, and fragrances.
Workshop seminars from instrument manufacturers and an extensive
exhibition of instrumentation, accessories, and supplies will run in parallel to
the scientific programme.
At the meeting, the 2018 Marcel Golay Award will be presented in
recognition of outstanding contributions in the field of separation science.
The Leslie Ettre Award will be presented to a young scientist for research
on capillary GC applied to environmental or food analyses. The Giorgio
Nota Award will be presented to a scientist in recognition of a lifetime of
achievement in capillary LC. The John Phillips Award will be awarded to
individuals who have made outstanding contributions to the field of GC×GC
analysis. The GC×GC Lifetime Achievement Award honours an experienced
GC×GC scientist who has made significant contributions to the field.
Scholarships for young researchers will also be promoted by the Interdivisional
Group of Separation Science of the Italian Chemical Society (Italy) and its
partners.
To encourage scientific exchange and friendship building, the scientific
programme will be enhanced with the well-known “Riva Social Programme”,
which consists of a welcome reception, cocktail party, classical concert,
wine and cheese evening, and farewell cocktails. Considering the interest in
comprehensive techniques, the 15th GC×GC Symposium will be organized
during the same period to allow scientists to attend both meetings. The 15th
GC×GC Symposium will start on 13 May 2018 with a course presented
by experts in the field covering the fundamental aspects of comprehensive
techniques and a plenary session on 14 May 2018. For both meetings,
abstracts for consideration as lecture or poster presentations can be submitted
on-line at: http://www.chromaleont.it/iscc
For more information please visit www.chromaleont.it/iscc or e-mail: iscc@
chromaleont.it
EEVVEENNTT NNEEWWSSEVENT NEWS
Ph
oto
Cre
dit:
Lu
ka
sz S
zw
aj/sh
utt
ers
toc
k.c
om
LC•GC Europe December 2017688
THEAPPLICATIONS
BOOK
December 2017
www.chromatographyonline.com
CONTENTS
THE APPLICATIONS BOOK
Environmental
691 Analysis of Organophosphates in Lithium Ion Battery Electrolytes by HILIC–ESI-MS
Jonas Henschel1, Martin Winter1,2, Sascha Nowak1, and Wen
Jiang3, 1MEET Battery Research Center, University of Münster, 2Forschungszentrum Jülich GmbH, IEK-12, Helmholtz Institute Münster
(HI MS), 3HILICON AB
693 Overcome the Memory Effect in Dioxins Extraction with ETHOS X Powered by FastEX-24
Diego Carnaroglio and Matteo Volpi, Milestone Srl
Food and Beverage
694 Determination of Chloramphenicol in Honey with Automated Solid-Phase Extraction and HPLC–MS/MS Detection
Hans Rainer Wollseifen, Johannes Brand, and Detlef Lambrecht,
MACHEREY-NAGEL GmbH & Co. KG
696 Determination of Ephedra Alkaloids and Synephrine in Dietary Supplements via Strong Cation-Exchange SPE and LC–MS/MS DetectionBrian Kinsella, UCT, LLC
Medical/Biological
697 Monitoring Emissions From Respiratory Medical Devices in Accordance With ISO 18562-3
David Barden and Gareth Roberts, Markes International
699 Definitive EtG/EtS LC–MS/MS Analysis: A Rugged 4-Min Method for High-Throughput Laboratories
Justin Steimling and Frances Carroll, Restek Corporation
Cover Photography: Shutterstock.com
690 THE APPLICATIONS BOOK – DECEMBER 2017
Photo
Cre
dit:
Melic
a/S
huttesto
ck.c
om
Photo
Cre
dit:
Will
yam
Bra
db
err
y/S
hutters
tock.c
om
Photo
Cre
dit: ke
ren
-seg
/Shutters
tock.c
om
Photo
Cre
dit:
T.S
um
aeth
o/S
hutters
tock.c
om
Photo
Cre
dit: A
bel Tum
ik/S
hutters
tock.c
om
CONTENTSCONTENTS
THE APPLICATIONS BOOK – DECEMBER 2017 691
ENVIRONMENTAL
New separation techniques for the analysis of polar and ionic
analytes have aroused great interest in the fi eld of metabolomics and
environmental investigation in the past two decades. Hydrophilic
interaction liquid chromatography (HILIC) is a promising tool to
address this challenge. HILIC separation is based on the polarity
of analytes, which generally show stronger retention with increasing
polarity according to the HILIC separation mechanism. Furthermore,
the high content of organic solvent in the mobile phase leads to
good ionization properties in the electrospray ionization (ESI), and
consequently enhances the detection sensitivity by hyphenated
mass spectrometry (MS) detector.
Analysis of organophosphates as the degradation products
of pesticides, flame retardants, plasticizers, or batteries is very
important for safety evaluation because their chemical structures
are similar to chemical warfare agents with high toxic potential
(1). Lithium ion batteries possess electrolytes in cells that usually
consist of different formulations of linear and cyclic organic
carbonates and hexafluorophosphate (LiPF6) conducting salts.
One of the major issues is the degradation of LiPF6 as a result
of thermal and chemical instability of the P-F bond. Subsequent
reactions lead to battery performance loss. The presence of
traces of moisture and elevated temperatures in a battery cell
promote the decomposition of PF6 to the reactive intermediate
PF5 after a reaction cascade culminating in fluorinated and
alkylated phosphates (3). Therefore, it is vital to monitor all the
organophosphates in batteries, which helps to better understand
battery performance loss and to avoid its potential threat to the
environment during the disposal process.
Organophosphates cover a wide spectra of ionic and high polar
species as well as nonpolar analytes. Multiple separation techniques
such as gas chromatography (GC), ion chromatography (IC), or
reversed-phase liquid chromatography (LC) are often required for
the comprehensive determination of the organophosphates and
their formed degradation products (1,2). However, we aimed to
develop an LC method using either HILIC or RP column that leads
to a simultaneous separation of all organophosphorus species
by one chromatography run for our study. In this application,
we demonstrate the separation potential of polar and ionic
organophosphate species with an iHILIC®-Fusion(+) column
packed with charge modulated hydroxyethyl amide silica. A
mixed interaction, for example, hydrophilic partitioning and ionic
interactions, may be involved in the HILIC separation mechanism
depending on the chemical properties of the analytes.
Experimental
LC–MS system 1: Thermo Ultimate 3000 LC system and AB SCIEX
3200 QTrap® equipped with an ESI source, operated in ESI(+) and
ESI(-) mode for analysis of standards.
Analysis of Organophosphates in Lithium Ion
Battery Electrolytes by HILIC–ESI-MS Jonas Henschel1, Martin Winter1,2, Sascha Nowak1, and Wen Jiang3, 1MEET Battery Research Center, University of Münster, 2Forschungszentrum Jülich GmbH, IEK-12, Helmholtz Institute Münster (HI MS), 3HILICON AB
TEP
DBP DEP DMP
DFP
TMP DMFP
O
O
O
O P
O
O
O
O P
O
O
O
O P
O
F
O
O P
O
F
F
O P
–
O
O
O
O P–
O
O
O
O P–
–
100
50
0
0 4 8
Time (min)
DEP DMP
DBP
DFP
TMP
TEP
DMFP
Rel. in
ten
sit
y (
%)
12 16
Figure 1: Chemical structures of organophosphate standards used in the study.
Figure 2: Extracted ion chromatograms of organophosphate standards in HILIC separation with iHILIC®-Fusion(+).
692 THE APPLICATIONS BOOK – DECEMBER 2017
ENVIRONMENTAL
LC–MS system 2: Shimadzu Nexera X2 LC system and Shimadzu
LC-MS-IT-TOF™ equipped with an ESI source, operated in ESI(+)
mode for analysis of thermally treated battery electrolyte.
Column: 150 × 2.1 mm, 3.5-μm iHILIC®-Fusion(+) (P/N
100.152.0310, HILICON AB)
Gradient elution 1: A) acetonitrile; B) deionized water-1 M
ammonium formate, pH 2.9 (99.5:0.5); 0–3.5 min (90:10) A–B;
3.5 to 6.5 min, gradient elution from (90:10) A–B to (50:50) A–B;
6.5–14.5 min, (50:50) A–B.
Gradient elution 2: A) acetonitrile; B) deionized water-1 M
ammonium acetate, pH 6.1 (99:1); 0–3.5 min (95:5) A–B; 3.5 to
8 min, gradient elution from (95:5) A–B to (75:25) A–B; 8–16 min,
(75:25) A–B.
Flow rate: 0.3 mL/min
Column temperature: 40 °C
Injection volume: 10 μL
Organophosphates: Trimethyl phosphate (TMP), triethyl phosphate
(TEP), dimethyl phosphate (DMP), diethyl phosphate (DEP),
dibutyl phosphate (DBP), difl uorophosphate (DFP), dimethyl
fl uorophosphates (DMFP). Their detailed chemical structures are
illustrated in Figure 1. A 50 μM measure of TMP, TEP, DMP, and
DEP, 100 μM of DFP and DBP, and 350 μM of DMFP were prepared
in 90:10 acetonitrile–water solution.
Battery electrolyte: 1 M LiPF6 in ethylene carbonate–ethyl methyl
carbonate–vinylene carbonate (30:70:+2 w%, LP 572, BASF) with
2v% water addition was stored at 80 °C for 14 days.
Results and Conclusion
As shown in Figure 2, seven standards of organophosphates
can be simultaneously separated and determined using an
iHILIC®-Fusion(+) column hyphenated to ESI-MS detection. The
nonpolar organophosphates TEP, TMP, and DMFP were eluted fi rst.
This is expected due to weaker interaction with the HILIC stationary
phase through the hydrophilic partioning. The other four ionic species,
DFP, DBP, DEP, and DMP, show stronger retention and good peak
shapes by hydrophilic partitioning and ionic interaction.
The applicability of the iHILIC®-Fusion(+) column for the study
of battery electrolyte degradation products was investigated and is
illustrated in Figure 3 and Table 1. The hyphenation of HILIC with an
IT-TOFTM-MS system allowed us to separate and elucidate structural
information of several formed organophosphates in thermally treated
battery electrolyte.
This work emphasizes the potential application of using
HILIC–ESI-MS methods for the analysis of organophosphates in the
battery industry and for environmental investigation.
References
(1) V. Kraft et al., RSC Advances 6, 8–17 (2016).
(2) I. Van der Veen and J. de Boer, Chemosphere 88, 1119–1153 (2012).
(3) C. Campion et al., J. Electrochem. Soc. 152(12), A2327–A2334 (2005).
HILICON ABTvistevägen 48, SE-90736 Umeå, Sweden
Tel.: +46 (90) 193469
E-mail: [email protected]
Website: www.hilicon.com
Rel. in
ten
sit
y (
%)
Time (min)
100
50
0
0 4 8 12 16
TEP
155.046
171.041
183.077
199.072
213.089
Figure 3: Extracted ion chromatograms of possible degradant organophosphate products in lithium ion battery electrolyte.
Table 1: Retention times (tR) and suggested structures of
degradation products from LiPF6 electrolyte
tR (min) m/z (M+H)+ Suggested Structure
1.3 213.089
1.3 183.077
1.3 199.072
8.8 199.072
9.1 155.046
9.4 171.041
9.7 171.041
13.2 171.041
THE APPLICATIONS BOOK – DECEMBER 2017 693
ENVIRONMENTAL
pressure reactor. This
approach ensures
the lowest analytical
b l a n k s a n d
completely avoids
carryover between
runs.
Instrumentation
Milestone’s new
ETHOS X microwave
extraction system
can extract organic
c o n t a m i n a n t s
from soils, in full
compliance with EPA
3546 (100–115 °C
and 50–150 psi). Disposable glass vials and contactless temperature
control in all positions makes it a unique and innovative solution for the
extraction of contaminants from soils, providing unmatched ease of use
and low running costs. The ETHOS X is capable of processing up to
30 g of sample per vessel (up to 24 samples simultaneously), thereby
improving the limit of quantitation (LOQ) for analysis. The handling is very
easy because the sample is weighed directly into the disposable glass
vial, 1:1 hexane–acetone or CH2Cl
2–acetone is added, and the vessel is
loaded into the FastEX-24 rotor. After 10–20 min of microwave heating,
the sample is ready to be fi ltered and analyzed by gas chromatography
(GC). ETHOS X provides full data traceability of the entire process.
Results
Table 1 shows recoveries of all the molecules in the range of
80–120% with great reproducibility (RSD%). It also shows the
blank values of the following run with new disposable glass vials (all
the blanks were below 0.065 μg/kg).
Conclusion
The ETHOS X enables simultaneous dioxins extraction of up to 24
samples (from weighing to fi ltration steps) in less than 1 h offering
superior productivity than any other technology in the market. In
addition, the use of contactless temperature control and disposable
glass vials fully solves constant issues in dioxins determination by
avoiding tedious cleaning, memory effect, and
carryover. The ETHOS X with all its unique features
fully addresses the needs of environmental
laboratories in terms of productivity, ease of use,
running costs, extraction quality, and is fully
compliant with EPA 3546 method.
Microwave extraction becomes faster and easier with the new
Milestone ETHOS X with its FastEX-24. This application note
shows the use of Milestone technology for extraction of dioxins
from environmental matrices, with emphasis on one of the most
common challenge of this application: the carryover effect. The
unique design of the FastEX-24 rotor with disposable glass vials
allows easy and effi cient dioxin extraction, and other organic
pollutants, to be performed, avoiding any memory effect. The
FastEX-24 simplifi es the routine pollutants extraction process
and provides superior productivity at lower costs.
Dioxin is known as one of the most toxic chemicals, a serious
public health threat. According to the US Environmental Protection
Agency (EPA) reports, dioxin and dioxin-like chemicals have been
associated with adverse health effects in the population.
The analysis of this class of pollutants is fundamental for
environmental and human health protection, but because of the
persistent nature of those compounds, the sample preparation
and analysis is challenging for many extraction techniques. Many
solutions are available for the extraction of dioxins such as Soxhlet,
automated Soxhlet, sonication, pressurized liquid extraction (PLE),
and closed microwave vessels. A common and very critical issue
in all sample preparation techniques is dioxin extraction and the
subsequent memory effect or carryover in the extraction cells.
ETHOS X equipped with the unique FastEX-24 rotor is specifically
designed to overcome the memory effect and cleaning by using
disposable glass vials as reaction vessels, which are placed into a
Overcome the Memory Effect
in Dioxins Extraction with
ETHOS X Powered by FastEX-24
Diego Carnaroglio and Matteo Volpi, Milestone Srl
Milestone SrlVia Fatebenefratelli, 1/5 24010 Sorisole (BG), Italy
Tel.: +39 035 573857
E-mail: [email protected]
Website: www.milestonesrl.com
Figure 1: ETHOS X microwave extraction system.
MILESTONE
Table 1: Recovery (n = 4) of PCDD and PCDF from sandy
soil standard reference material BCR®-529 (2 g) and blank
value after extraction with replaced disposable glass vial
AnalyteCertifi ed Value
(μg/kg)
Ethos X
Recovery (%)
RSD
(%)
Blank Value
(μg/kg)
2,3,7,8 -TCDD 4500±0.6 94 3.4 0.0185
1,2,3,7,8-PeCDD 440±0.05 117 2.8 0.02122
1,2,3,4,7,8-HxCDD 1220±0.21 106 3.1 0.00312
1,2,3,6,7,8-HxCDD 5400±0.9 85 2.1 0.02198
1,2,3,7,8,9-HxCDD 3000±0.4 84 1.9 0.06412
2,3,7,8-TCDF 78±0.013 96 2.7 0.02524
1,2,3,7,8-PeCDF 145±0.028 80 3.5 0.01939
2,3,4,7,8-PeCDF 360±0.07 91 2.6 0.01889
1,2,3,4,7,8-HxCDF 3400±0.5 100 1.9 0.03112
1,2,3,6,7,8-HxCDF 1090±0.15 99 3.8 0.00633
1,2,3,7,8,9-HxCDF 22±0.010 82 3.6 0.04043
2,3,4,6,7,8-HxCDF 370±0.05 120 2.2 0.05186
694 THE APPLICATIONS BOOK – DECEMBER 2017
FOOD AND BEVERAGE
A sensitive determination method with automated
solid-phase extraction (SPE) is presented and the
recovery rates are compared with manual extraction. The
automated extraction was developed with the SPE robot
FREESTYLETM SPE module (LCTech GmbH). According to the
European Regulation (EC) No 470/2009, CHROMABOND®
HLB was applied for SPE (1). A high performance liquid
chromatography tandem mass spectrometry (HPLC–MS/MS)
method with the biphenyl phase NUCLEODUR® π2 was used
for the subsequent analysis.
Chloramphenicol is a broad-spectrum antibiotic against
Gram-positive and Gram-negative bacteria. It has many side
effects. Thus, it is only used nowadays as a reserve antibiotic (2).
For warranty of the food safety, many countries declare limits for
pharmaceuticals in animal products. The necessary determination
methods demand a high-performance solid-phase extraction
(SPE) to reliably detect and quantify residues in trace level with
the subsequent analysis.
Automated Solid-Phase Extraction (3)
Sample pretreatment: Weigh out 5 g of honey. Add 4 mL water
and shake rigorously for 30 s. Spike with 1 mL standard solution
(chloramphenicol , chloramphenicol-d5 , c = 5 ng/mL in methanol)
and shake rigorously for 30 s. Add 15 mL ethyl acetate and shake
rigorously for 30 s. Centrifuge at 3000 rpm for 10 min. Take 12 mL
of supernantant and evaporate extracts to dryness at 40 °C under
a stream of nitrogen. Redissolve residue in 10 mL water.
Extraction:
SPE robot: FREESTYLETM SPE module, LCTech
SPE column: CHROMABOND® HLB, 3 mL, 200 mg,
MACHEREY-NAGEL REF 730924
Column conditioning: 3 mL methanol (dispensing speed 1 mL/min),
then 5 mL dest. water (dispensing speed 1 mL/min)
Sample application: 9 mL pretreated sample (dispensing speed
3 mL/min over sample loop)
Washing: 10 mL dest. water (dispensing speed 3 mL/min)
Drying: 100 mL air (dispensing speed 100 mL/min)
Elution: 5 mL 80:20 (v/v) ethyl acetate–methanol
Determination of Chloramphenicol in Honey with Automated
Solid-Phase Extraction and HPLC–MS/MS DetectionHans Rainer Wollseifen, Johannes Brand, and Detlef Lambrecht, MACHEREY-NAGEL GmbH & Co. KG
1.00e5
9.00e4
8.00e4
7.00e4
6.00e4
5.00e4
4.00e4
3.00e4
2.00e4
1.00e4
0.0
0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0
Time (min)
Max 27e5 cps. (min)
Inten
sit
y (
cp
s)
8.0 9.0 10.0 11.0 12.0 13.0 14.0
Figure 1: Chromatogram of spiked honey sample (1 μg/kg).
THE APPLICATIONS BOOK – DECEMBER 2017 695
FOOD AND BEVERAGE
Drying: 100 mL air (dispensing speed 100 mL/min)
Evaporation: under nitrogen, 40 °C
Reconstitution: in 1 mL 95:5 (v/v) dest. water–acetonitrile
Manual Solid-Phase Extraction
After identical sample pretreatment as used for automated
solid-phase extraction, the sample was manually extracted by a
vacuum manifold under comparable SPE conditions.
Subsequent Analysis: HPLC–MS/MS (4)
HPLC column: EC 150/2 NUCLEODUR® π2, 5 μm, MACHEREY-NAGEL
REF 760624.20
Eluent A: dest. water
Eluent B: acetonitrile
Gradient: 5–95% B in 7.5 min, 95% B for 1 min, 95–5% B in 1 min,
5% B for 5 min
Flow rate: 0.30 mL/min
Temperature: 35 °C
Injection volume: 5 μL
MS/MS detection: API 5500 (AB Sciex), ion source: ESI, negative
ionization mode, scan type: MRM (chloramphenicol: [M-H]- 320.9,
152.0 [Q1], 256.0 [Q2]; chloramphenicol-d5: [M-H]- 325.9, 156.6
[Q1], 261.8 [Q2]), curtain gas: 35 psig, ion spray voltage: -4500 V,
temperature: 450 °C, nebulizer gas: 45 psig, turbo gas: 45 psig,
CAD medium.
Results
Solid-phase extraction of chloramphenicol from the matrix honey
with CHROMABOND® HLB results in an outstanding recovery.
The recovery rates presented in Table 1 show that higher
recovery rates up to 90% with very good reproducibility (< 5%)
are achieved by automation with the SPE robot FREESTYLETM
SPE module. A correction with an internal standard is especially
advantageous for the manual sample preparation. Figure 1
shows the chromatogram of a sample extract from spiked honey.
A definite identification and quantification is possible as a result
of a clear separation of matrix interferences from the analyte
peak.
The liquid chromatographic method could be accelerated by using
a shorter HPLC column because of good retention of chloramphenicol
by the additional π-π-interactions of NUCLEODUR® π2.
Conclusion
A determination method for chloramphenicol in honey has been
developed, which receives very good results—especially by the
automation of solid-phase extraction.
References
(1) Regulation (EC) No 470/2009 of the European Parliament and of the
Council.
(2) T. Tomaszewski, Br Med J. 384–2, 388–392 (1951).
(3) Application No. 306350, MACHEREY-NAGEL, available from www.mn-net.
com/apps
(4) Application No. 128140, MACHEREY-NAGEL, available from www.mn-net.
com/apps
MACHEREY-NAGEL GmbH & Co. KGNeumann-Neander-Str. 6 – 8, 52355 Düren, Germany
Tel.: +49 24 21 969 0 Fax: +49 24 21 969 199
E-mail: [email protected] Website: www.mn-net.com
Table 1: Comparison of recovery rates from automated to manual solid-phase extraction
Recovery Rates (%) ChloramphenicolChloramphenicol
(Corrected by Int. Standard)Chloramphenicol-d
5
Automated SPE 86.8 ± 4.4 (n = 4) 90.8 ± 4.6 (n = 4) 95.6 ± 4.5 (n = 4)
Manual SPE 74.6 ± 2.7 (n = 3) 92.9 ± 3.4 (n = 3) 80.3 ± 4.7 (n = 3)
696 THE APPLICATIONS BOOK – DECEMBER 2017
FOOD AND BEVERAGE
Ephedra alkaloids are phenethylamines that occur naturally
in plants, including the herb Ma Huang used in traditional
Chinese medicine. Ephedra alkaloids are potent CNS stimulants
and also have a sympathomimetic effect on the peripheral
nervous system. The aim of this study was to develop a
multi-analyte procedure for the extraction, cleanup, and
quantifi cation of the ephedra alkaloids in functional foods
and natural products. High capacity strong cation-exchange
SPE cartridges were used for the isolation of ephedrine,
pseudoephedrine, norephedrine, norpsuedoephedrine,
methylephedrine, and synephrine from dietary supplements.
Procedure
1. Sample Pretreatment
a) Weigh 1 ± 0.1g of dietary supplement into a 15 mL polypropylene
centrifuge tube. For this study, a SPEX® SamplePrep®GenoGrinder®
was used to pulverize the tablets.
b) Add 10 mL of 1% formic acid to each sample.
c) Shake or vortex sample for 15 min to fully extract the ephedra
alkaloids. Ensure samples are fully dissolved.
c) Centrifuge the sample for 10 min at ≥3000 × g and 4 °C.
2. SPE Procedure
SPE Conditioning
a) Add 2 × 4 mL of methanol to CUBCX1HL56 SPE cartridge.
b) Add 4 mL of ultrapure water.
c) Add 4 mL of 1% formic acid.
Note: Do not let the cartridge go dry otherwise repeat steps a)
through c).
Sample Extraction
a) Load supernatant from step 1.
b) Allow sample to percolate through the cartridge or apply a
vacuum if necessary (adjust vacuum for fl ow of 1–3 mL per minute).
Wash Cartridge
a) Add 2 × 4 mL of 0.1% formic acid.
b) Add 2 × 4 mL methanol.
c) Dry under vacuum for ≈30 s to remove excess solvent.
Elution
a) Elute the ephedra alkaloids using 8 mL of methanol containing
2% ammonium hydroxide.
b) Evaporate off the methanol solvent at 40 °C under a gentle
stream of nitrogen until it reaches a volume of ≈1 mL.
c) Add 1 mL of aqueous mobile phase (10 mM ammonium acetate)
d) Evaporate off any remaining methanol*
e) Vortex the samples for 1 min and fi lter through a 0.2 μM syringe
fi lter directly into a HPLC vial.
*Ephedra alkaloids are similar to amphetamines, which are known to be
volatile compounds. Therefore, extra care was taken during the evaporation
step to avoid any potential loss in recovery that may occur during this step.
Results
Conclusion
A method was successfully developed for the extraction, cleanup, and
quantifi cation of the ephedra alkaloids and synephrine in functional
foods and natural products. Strong cation-exchange SPE was used to
isolate the phenethylamines from the complex sample
matrix, which consisted of nine herbal extracts containing
a high concentration of calcium, caffeine, and additional
excipients. A high capacity SPE sorbent was used because
it offers better retention than standard SCX sorbent.
Determination of Ephedra Alkaloids and
Synephrine in Dietary Supplements via Strong
Cation-Exchange SPE and LC–MS/MS DetectionBrian Kinsella, UCT, LLC
UCT, LLC2731 Bartram Road, Bristol, Pennsylvania 19007, USA
Tel: (800) 385 3153
E-mail: [email protected] Website: www.unitedchem.com
NL: 3.35E4TICF: + C ESI SRMms2 150.142(91.019-91.021,107.009-107.011)MS Standard_10ppb
0
100
80
60
40
20
0
100
80
60
40
20
0
100
80
60
40
20
0
100
80
60
40
20
0
100
80
60
40
20
0
2 4 6 8 10
Time (min)
Re
lati
ve
Ab
un
da
nce
17.66
14.74
13.72
11.43
9.86
13.818.30
11.449.91
8.95
6.72
4.20
5.59
3.31
2.12RT: 0.00- 19.99 SM 5G
12 14 16 18
NL: 2.36E3TICF: + C ESI SRMms2 152.141(91.019-91.021,115.019-115.021)MS Standard_10ppb
NL: 6.83E3TICF: + C ESI SRMms2 166.173(91.009-91.011,115.019-115.021)MS Standard_10ppb
NL: 2.46E3TICF: + C ESI SRMms2 169.173(90.989-90.991,114.999-115.001)MS Standard_10ppb
NL: 2.94E3TICF: + C ESI SRMms2 180.175(90.989-90.991,147.039-147.041)MS Standard_10ppb
Figure 1: Chromatogram of a 10 ng/mL standard.
Table 1: Extraction and analytical materials
CUBCX1HL56 High-load benzenesulfonic acid – 500 mg / 6 mL
SLPFPP100ID21-3UMSelectra® PFPP HPLC column
100 × 2.1 mm, 3-μm
SLPFPPGDC20-3UMSelectra® PFPP guard column
10 × 2.1 mm, 3-μm
THE APPLICATIONS BOOK – DECEMBER 2017 697
MEDICAL/BIOLOGICAL
This study describes the monitoring of potentially harmful
volatile organic compounds (VOCs) emitted from respiratory
medical devices, by pumped sampling and thermal
desorption–gas chromatography–mass spectrometry
(TD–GC–MS) analysis in accordance with ISO 18562 part 3.
Emissions from two sets of face-mask supply tubing and
three nasal cannulas were compared, and all were found to
emit VOCs at levels that may give cause for concern.
Respiratory medical devices (RMDs) are widely used in medical
practice, but the potential for polymeric components in the gas
pathways to release harmful volatile organic compounds (VOCs) is
giving rise to concern that prolonged use could be detrimental to
patient health. This is the driving force behind ISO 18562 (1), which
was released in March 2017. The predecessor to ISO 18562 (ISO
10993) did not cover VOC emissions directly, and was partly based
on animal implant tests of uncertain scientifi c value. In contrast, the
new method is based on tests performed directly on the device, and
does not involve animal experimentation.
The ultimate outcome of testing in accordance with ISO 18562 is
to assess the patient’s exposure to chemicals from the RMD during
their treatment (expressed in units of μg/day), which is then related
to thresholds of toxicological concern (TTCs) specified in ISO 18562.
Devices requiring testing under ISO 18562 include medical devices
and accessories containing gas pathways, which includes ventilators,
anaesthesia workstations, oxygen-conserving equipment, nebulizers,
gas monitors, masks, mouthpieces, breathing tubes, and breathing
accessories.
Part 3 of ISO 18562 deals with testing for emissions of VOCs, and
references two analytical options: ISO 16000-6 for sorbent-tube
sampling and ASTM D5466 for canister sampling. Both are
well-established and popular for air monitoring and related applications,
but generally speaking, tube-based methods such as ISO 16000-6 are
preferred for investigating emissions from products and materials, for
reasons of wider analyte range and also greater ease of use. Analysis
in accordance with the tube-based standard ISO 16000-6 is therefore
the focus of this study.
Experimental
Three sets of face-mask supply tubing (A, B, C), and two nasal
cannulas (D, E) were assessed. In brief, VOC emissions were
collected on a 3½″ × ¼″ sorbent tube after being swept from the
RMD by a fl ow of clean air, followed by analysis of the tubes using
a TD100-xr™ thermal desorption tube autosampler connected to
a GC–MS system. Compounds were identifi ed by screening the
chromatograms against a 540-component library of compounds
typically encountered in polymeric consumer goods and
construction materials. Full analytical conditions are available in
Markes International’s Application Note 132 (http://chem.markes.
com/RMD).
Results and Discussion
VOC Profi les: As an example of a typical profi le, Figure 1 shows the
analysis of Sample A (face-mask supply tubing). The large number
of compounds released is immediately apparent, as is the presence
of a broad hydrocarbon/oil-like response, which was confi rmed as
a real feature by its absence from a sampling run without the RMD
in position.
Across the five samples A–E, screening of the chromatographic
profiles indicated the presence of between 45 and 60 components
from the 540-component library. The most abundant components
are shown in Figure 2.
Monitoring Emissions From Respiratory Medical
Devices in Accordance With ISO 18562-3David Barden and Gareth Roberts, Markes International
Ab
un
dan
ce (
x 1
09 c
ou
nts)
Retention time (min)
3
2
1
0
2 4 6 8 10 12 14 16 18 20 30
10396
92
90
89
86
84
7950
2
2
50
79
84
86
89
90
92
96
103
Acetone
Cyclohexanone
2-Ethylhexan-1-ol
n-Undecane
1,2,4,5-Tetramethylbenzene
n-Dodecane
n-Tridecane
n-Tetradecane
Diethyl phthalate
Bis(2-ethylhexyl) phthalate
Figure 1: VOC profi le of a sample of face-mask supply tubing. Selected peaks are labelled.
698 THE APPLICATIONS BOOK – DECEMBER 2017
MEDICAL/BIOLOGICAL
Two compounds dominate the profiles of all five samples:
Cyclohexanone is used as a binding agent for polymer fittings,
while 2-ethylhexan-1-ol is a precursor in the synthesis of phthalate
plasticizers. The large number of other compounds present includes
solvents, such as dichloromethane (now declining in use as a binding
agent, but present in all five samples), methyl ethyl ketone (sometimes
used in conjunction with cyclohexanone, and found in Samples B and
D), and toluene (present in all five samples).
Phthalates were also present in all five samples, with diethyl
phthalate and bis(2-ethylhexyl) phthalate being confidently identified.
The latter was the least volatile of the components, eluting as the last
peak. A specification for phthalate-free medical tubing is available,
but these results show that at least some manufacturers are not yet
producing phthalate-free products.
TVOC Values and Exposures: ISO 18562-3 is primarily concerned
with the “dose to the patient of a substance or sum of substances”
within three exposure categories, which are: “Limited” (≤24 h),
“Prolonged”(>24 h and <30 days), and “Permanent” (≥30 days).
The sampling time of 1 h and sampling rate of 0.1 L/min used in
the current study means that the results are of greatest relevance to
breathing volumes of newborns and to devices used in the “Limited”
category.
To assess how Samples A–E perform within this category, total VOC
(TVOC) values were determined by integrating the entire chromatogram
(Table 1). Extrapolating these 1-h TVOC values to 24 h on the basis of
a separately determined decay curve allowed derivation of cumulative
exposure values for the first 24 h of use. Values are in the range
1370–7000 μg, which exceeds three times the TTC for adults of
360 μg/day, and is over 500 times the TTC for newborns of ~2.5 μg/day.
Conclusions
This study shows that the sorbent sampling tubes and
TD–GC–MS analytical equipment used is versatile and fully
compliant with the requirements of ISO 18562-3 (and of the
cited method ISO 16000-6). In particular, the thermal desorption
instrument used to pre-concentrate the vapours is highly sensitive
and easily automated, and in combination with GC–MS is able to
reliably identify compounds released from the gas pathways of
respiratory medical devices.
The results obtained indicate that some respiratory medical devices
emit a broad range of VOCs, at levels that indicate potential issues
with manufacturing quality. The use of low flow rates, and the fact
that the products were sampled immediately after being removed
from the packaging, means that these results represent a “worst-case
scenario” for VOC emissions, but nevertheless one that may be typical
of products used for the care of certain patients.
References
(1) ISO 18562: Biocompatibility evaluation of breathing gas pathways in
healthcare applications, International Organization for Standardization,
2017. Part 3 (“Tests for emissions of volatile organic compounds (VOCs)”)
is available at www.iso.org/standard/62894.html.
For the complete study, including a full table of results, download
Markes International’s Application Note 132 (http://chem.markes.
com/RMD).
Markes InternationalGwaun Elai Medi-Science Campus, Llantrisant, Wales, UK
Tel.: +44 (0)1443 230935
E-mail: [email protected] Website: www.markes.com
80
60
40
20
0
A B C D E
Sample
Ma
ss (
μg
to
lue
ne
eq
uiv
ale
nts)
Tolu
ene
Cyclo
hexanone
Cum
ene
o-C
ym
ene
2-E
thylh
exan-1
-ol
Aceto
phenone
n-U
ndecane
n-T
ridecane
n-D
odecane
Bis(2
-eth
ylh
exyl)
phth
ala
te
Figure 2: Comparison of the masses of the 10 compounds present at levels above 2000 ng in any of the samples.
Table 1: Estimated TVOC exposures in μg (toluene
equivalents) over the first 1 and 24 h of use of Samples
A–E, based on integration of the entire chromatogram and a
50% 24-h decay rate
A B C D E
1 h 371 172 429 75 122
24 h 6777 3131 7830 1369 2234
THE APPLICATIONS BOOK – DECEMBER 2017 699
MEDICAL/BIOLOGICAL
Methods for monitoring alcohol consumption biomarkers EtG
and EtS are generally limited by poor retention and coelution
with matrix interferences, as well as by long analysis times and
short column lifetimes. The dilute-and-shoot EtG/EtS LC–MS/MS
analysis developed here using the novel Raptor EtG/EtS column
easily resolves EtG and EtS from matrix interferences, providing
consistent, accurate results for high-throughput labs testing
human urine samples for alcohol consumption.
Ethyl glucuronide (EtG) and ethyl sulfate (EtS) are important
biomarkers of alcohol use. The detection of these metabolites offers
many advantages for abstinence monitoring, including a three-day
detection window, good stability in properly stored specimens, and
analytical specifi city. However, typical methods for EtG/EtS analysis
have several shortfalls: poor resolution of EtG and EtS from matrix
components, long run times that limit sample throughput, and short
column lifetimes.
In this study, a simple dilute-and-shoot method was developed,
validated, and applied to patient samples for EtG/EtS LC–MS/MS
analysis in human urine. The method was developed on a Raptor
EtG/EtS column because it provides the retention characteristics that
are needed to consistently elute the target analytes away from matrix
interferences. In addition, it is rugged and long-lasting, which is
benefi cial to high-throughput laboratories.
Experimental Conditions
Standard and Sample Preparation: Human urine (alcohol free) was
fortifi ed with EtG and EtS to prepare calibration standards and QC
samples. The concentration of calibration standards ranged from
50–5,000 ng/mL for both analytes. Four QC levels were prepared at
50, 150, 750, and 4000 ng/mL.
Aliquots (50 μL) of urine were diluted with 950 μL of the working
internal standard (100 ng/mL EtG-d5 and 25 ng/mL EtS-d5 in
0.1% formic acid in water). Samples were vortexed at 3500 rpm for
10 s to mix followed by centrifugation at 3000 rpm for 5 min at 10 °C.
Additional double blanks were extracted for column equilibration.
Defi nitive EtG/EtS
LC–MS/MS Analysis: A
Rugged 4-Min Method for
High-Throughput LaboratoriesJustin Steimling and Frances Carroll, Restek Corporation
Time (min)
0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0
MatrixInterference
MatrixInterference
MatrixInterference
MatrixInterference
EtG
EtG
EtS
Injection 1
Injection 1,000
EtS
Figure 1: Rugged Raptor EtG/EtS columns provide reliable results for EtG/EtS LC–MS/MS analysis allowing more samples to be analyzed between column changes.
Table 1: Analyte transitions
Analyte Precursor Ion Product Ion Product Ion
EtG-d5 225.9 84.9 –
EtG 220.8 84.9 74.8
EtS-d5 129.7 97.7 –
EtS 124.7 96.8 79.7
LC–MS/MS Analysis: In order to ensure good response, peak
shape, and retention time consistency, the analytical column
was conditioned prior to use with 30 matrix injections that were
run through the full gradient program. Instrument parameters for
EtG/EtS LC–MS/MS analysis are shown below and the analyte
transitions are given in Table 1.
Analytical column: Raptor EtG/EtS, 2.7 μm, 100 mm × 2.1 mm
(cat.# 9325A12)
Guard column: UltraShield UHPLC precolumn fi lter, 0.2 μm frit
(cat.# 25809)
Mobile phase A: 0.1% formic acid in water
Mobile phase B: 0.1% formic acid in acetonitrile
Gradient Time (min) %B
0.00 5
2.50 35
2.51 5
4.00 5
Flow rate: 0.5 mL/min
Injection volume: 10 μL
Column temp.: 35 °C
Ion mode: Negative ESI
Results
Chromatographic Performance: As shown in Figure 1, a fast,
4-min EtG/EtS LC–MS/MS analysis was obtained from the direct
(dilute-and-shoot) injection of supernatant. Both EtG and EtS are
clearly resolved from matrix interferences, making accurate peak
identifi cation an easy task. Even after 1000 sample injections on a
Raptor EtG/EtS column, all chromatographic peaks maintained the
700 THE APPLICATIONS BOOK – DECEMBER 2017
MEDICAL/BIOLOGICAL
Restek Corporation110 Benner Circle, Bellefonte, Pennsylvania 16823, USA
Tel.: (800) 356 1688 Fax: (814) 353 1309
Website: www.restek.com
Time (min)
0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0
EtG
(overlay of 5,000 ng/mL standard)
Matrix double
blank injection with
post-column EtG
infusion.
Figure 2: EtG elutes outside the matrix suppression zones (double blank injection with post-column EtG infusion).
initial peak shape, retention time, and response.
Linearity: Using linear 1/x weighted regression for EtG and EtS, both
compounds showed good linearity with r2 values of 0.999 or greater
over a calibration range of 50–5000 ng/mL. Accurate calibration
down to 50 ng/mL allows low-level positive results to be reported with
confi dence.
Accuracy and Precision: Precision and accuracy analyses were
performed on three different days. The method accuracy was
demonstrated to be within 6.3% of the nominal concentration
for all QC levels for both EtG and EtS. The %RSD was
1.28–9.19% and 4.01–6.82% for intra- and inter-run,
respectively, at the QC LLOQ. The %RSD was 0.675–7.78% and
1.00–4.99% for intra- and inter-run, respectively, at the QC low,
mid, and high levels, indicating good method precision for EtG/EtS
LC–MS/MS analysis (Table 2).
Matrix Effect: Samples prepared in matrix show approximately 80%
(EtG) and 100% (EtS) of the signal obtained when samples are
prepared in solvent across all QC levels, demonstrating that matrix
effect is minimal. The sensitivity and consistency of the EtG response
at the LLOQ is only possible because retention on the Raptor EtG/EtS
column was optimized so that EtG elutes outside the zones of matrix
suppression (Figure 2). Because the optimal EtG elution time was
established here during method development, clinical labs adopting
the method can be confi dent that matrix suppression has already
been minimized. This, in combination with the established calibration
range and LLOQ, demonstrates that the method provides excellent
sensitivity for EtG, which is generally harder to detect at low levels
than EtS.
Sample Analysis: The robustness of the method was further tested by
monitoring fi ve patient samples with positive results for both EtG and EtS
in the lower end of the linear range across multiple instrument platforms
and nine lots of Raptor EtG/EtS columns. The precision of the results
Table 2: Accuracy and precision of QC samples
QC LLOQ (50 ng/mL) QC Low (150 ng/mL)QC Mid
(750 ng/mL)
QC High
(4000 ng/mL)
Analyte
Average
Conc.
(ng/mL)
Average
Accuracy
(%)
%RSD
Average
Conc.
(ng/mL)
Average
Accuracy (%)%RSD
Average
Conc.
(ng/mL)
Average
Accuracy
(%)
%RSD
Average
Conc.
(ng/mL)
Average
Accuracy (%)%RSD
EtG 51.2 102 6.82 143 95.2 4.99 749 99.8 3.68 3949 98.7 2.04
EtS 46.9 93.7 4.01 143 95.6 1.42 762 102 1.00 3958 98.9 1.59
Table 3: Inter-run precision of patient samples across
multiple instrument platforms and nine column lots
Patient
Sample
Average EtG
Concentration
(ng/mL)
%RSD
Average EtS
Concentration
(ng/mL)
%RSD
A 216 6.13 78.0 3.56
B 1,167 4.81 300 3.24
C 98.2 9.76 82.4 4.01
D 319 8.33 233 3.63
E 247 11.2 163 8.70
(n = 9) was found to range from 3.24–11.2% for both analytes over
multiple days and sample preparations, indicating excellent robustness
and ease of method transfer (Table 3). High-throughput labs in particular
will benefi t from implementing this rugged EtG/EtS LC–MS/MS analysis
method as highly consistent performance is seen across a wide range
of scenarios.
Conclusions
It was demonstrated that the Raptor EtG/EtS column is excellent for the
rapid and accurate analysis of EtG and EtS in human urine. Isobaric
matrix interferences are easily resolved, preventing issues with peak
identifi cation and quantitation. In addition, superior sensitivity for EtG
is achieved because it elutes outside the zones of matrix suppression
that are typically observed with dilute-and-shoot assays. With a fast
and simple sample preparation procedure and only four minutes
of chromatographic analysis time, the EtG/EtS LC–MS/MS method
established here provides accurate, high-throughput monitoring of
alcohol consumption. For the full article, visit www.restek.com and
enter “CFAN2736” in the search box.
L+,/,& �)XVLRQ�
DGYDQFHV�+,/,&�VHSDUDWLRQV�LQ�8+3/&�DQG�+3/&
HILICONHILICON
� &KDUJH�PRGXODWHG�K\GUR[\HWK\O�DPLGH�DPLGH�+,/,&�FROXPQV
� &RPSOHPHQWDU\�VHOHFWLYLWLHV�IRU�SRODU�K\GURSKLOLF�DQDO\WHV
� ([FHOOHQW�GXUDELOLW\�DQG�XOWUD�ORZ�EOHHGLQJ�IRU�/&�06
� 9HUVDWLOH�FROXPQV�IRU�μ2PLFVμ�VWXGLHV
� L+,/,& �)XVLRQ�DQG�L+,/,& �)XVLRQ�����S+����������������DQG����P� �
� L+,/,& �)XVLRQ�3���S+����������P��
)RU�PRUH�LQIRUPDWLRQ�
(PDLO��LQIR#KLOLFRQ�FRP�_�:HEVLWH��ZZZ�KLOLFRQ�FRP��+,/,&21�$%��$OO�ULJKWV�UHVHUYHG��_�L+,/,& �LV�D�UHJLVWHUHG�WUDGHPDUN�RI�+,/,&21�$%��6ZHGHQ
�
L+,/,& �)XVLRQ�
L+,/,& �)XVLRQ����
L+,/,& �)XVLRQ�3��
6LOLFD�EDVHG 6LOLFD�EDVHG 3RO\PHU�EDVHG
� �
www.gerstel.com
Not seeing the forest for the ...?
Get a clear view of the big picture and relevant details: Use intelligently automated GERSTEL solutions with PrepAhead productivity to get your work done in a transparent, efficient and precise manner.
Sample preparation for GC/MS and LC/MS analysis:s�Liquid/liquid extractions�Clean-up and matrix removals�Analyte concentrations�Derivatization and liquid additions�Generation and addition of standards
Our solvent-free, solvent reduced andμL-scale sample prep techniquesimprove LOQs and provide aclearer view while reducing costand environmental impact. What can we do for you?
Derivatization, Adding Standards
SPE, Filtration, Clean Up
Automated Liner Exchange ALEX (QuEChERS)
SPME, SBSE, Thermal desorption
Headspace & Dynamic Headspace (DHS)
Sample Prep by mouse-click (MAESTRO)