Registry Part 3

28
Image Formation Part III Nuclear Magnetic Resonance Amanda Golsch BSc, RT(R)(MR)

description

Part three of my MRI review session

Transcript of Registry Part 3

Page 1: Registry Part 3

Image Formation

Part IIINuclear Magnetic Resonance

Amanda Golsch BSc, RT(R)(MR)

Page 2: Registry Part 3

What does the N really mean in NMR?

• The term nuclear in Nuclear Magnetic Resonance refers to the nucleus of an atom. However, this term is slightly intimidating due to many patients envisioning something radioactive. Therefore, the name was changed to Magnetic Resonance Imaging.

Page 3: Registry Part 3

The Atom• The atom has a nucleus that contains protons and neutrons.• Certain nuclei have properties that give them magnetic characteristics. One of the properties is an

uneven mass number (protons + neutrons). The uneven mass number makes the substance MR active.• The hydrogen atom is 80% abundant in the human body (water/fat) and contains one proton. These

attributes make it MR active.• The proton has mass, a positive charge, and spins on its axis. Moving charged particles make a

magnetic field, and the spinning motion of a positively charged particle creates a magnetic field around the proton.

• The proton’s magnetic field is termed the magnetic moment.

Page 4: Registry Part 3

Parallel and Antiparallel• Hydrogen protons have a magnetic

moment and therefore behave in a certain manner when they are placed in an externally applied magnetic field.

• These protons can have two energy states.

– Low energy state- These are the protons/spins that align or are in parallel with the magnetic field also known as the B0 field. The circle labeled A demonstrates a spin that is in the low energy parallel state.

– High energy state- These are the protons/spins that are aligned anti-parallel, or against the direction of the B0 field. The circle labeled B demonstrates a spin in a high energy antiparallel state.

A

B

Page 5: Registry Part 3

Question 1• T or F: Spins in the low energy state align parallel with the B0 magnetic

field.

Page 6: Registry Part 3

Parallel and Antiparallel Continued…

• There will be a slight majority or spin excess in the low energy parallel state. This is known as thermal equilibrium.

• During thermal equilibrium, the magnetic moments of spins in the high energy state cancel the effect of those in the low energy state.

• Since there is a spin excess or greater number of spins in the low energy state, their magnetic moments add to make a magnetic field known as the net magnetization.

• The net magnetization is expressed as a vector. The net magnetization vector is aligned parallel with the B0 field and represents the spin excess and is also know as the proton density.

Page 7: Registry Part 3

Spins and the X,Y,Z Coordinate System

• Y= Anterior to Posterior• X= Left to Right• Z= Head to Foot• In the image, the low energy spins

are aligned along the Z axis. During excitation the net magnetization vector will be transferred onto the X, Y plane.

Page 8: Registry Part 3

Precessional Frequency and Calculation

• Hydrogen protons spin, align, and precess about the axis of the static magnetic field.

• Precession has been compared to that of a spinning/wobbling top.• The frequency of precession is important for MR imaging.• Precessional Frequency can be calculated using the Larmor Equation

• Gyromagnetic Ratio 1H=42.6 MHz/T. This is the ratio of the spin angular momentum of the proton to it’s magnetic moment.

Page 9: Registry Part 3

Question 2• At 1.5 T, what is the resonant frequency of a hydrogen proton?

– A. 52.5 MHz– B. 63.9 MHz– C. 72.9 MHz– D. 99.8 MHz

Page 10: Registry Part 3

The RF Pulse• RF energy, that is used for MRI is low energy non-ionizing electromagnetic

radiation.• The frequency of the RF wave is determined by the Larmor equation.• The radiofrequency is transmitted in short bursts called the RF pulse.

Page 11: Registry Part 3

What happens when magnetized protons are

subjected to an RF pulse?1. The spins precess in phase, and low energy spins enter the high energy

state. This causes the NMV to move into the transverse plane.2. The hydrogen spins that precess in phase causes the NMV to precess.3. The continuous RF pulse causes some spins in the low energy state to

absorb energy from the RF field and move into the high energy state.4. As more spins absorb the energy from the RF field, their spin states

change, and the NMV tips outward away from the longitudinal Z axis and through the X/Y transverse plane. The X/Y transverse plane is situated 90 degrees from the Z longitudinal plane. This is where the term 90 degree RF pulse comes from.

Page 12: Registry Part 3

How to achieve resonance• In order to achieve excitation or resonance, the transmitted RF pulse must

match the frequency of precession.– In order to achieve the excitation of water at 1.5T the transmitted RF pulse must

be 63.9 MHz. Why?

– 42.6 MHz/T * 1.5T = 63.9 MHz

Page 13: Registry Part 3

Question 3• In order to get signal in MRI a radiofrequency pulse is used to:

– A. Relax Spins– B. Align Spins– C. Reduce Spins– D. Excite Spins

Page 14: Registry Part 3

Question 4• After the application of the RF pulse spins:

– A. Are reduced– B. Move into the high energy state– C. Relax– D. Attract

Page 15: Registry Part 3

FID• In order to collect signal, there must be transverse magnetization that is in

phase. The signal is detected by the receiver coil when the NMV is rotated through the coil.

• The signal that is induced in the receiver coil is known as the FID or free induction decay.

• The FID decays rapidly after the application of the RF pulse. This is due to a shrinking NMV as it rotates through the X/Y plane. Why does the NMV shrink?

– The NMV shrinks because the spins lose phase coherence after the RF pulse is removed.

– The peaks show us where the NMV is in phase or perpendicular to the receiver coil. The troughs show us where the NMV is out of phase and perpendicular to the receiver coil. Notice how the FID decays over time.

Page 16: Registry Part 3

FID Continued• At the same time, yet independently, spins move from the high energy

state back into the low energy state. This ultimately leads to the NMV regrowing in the Z axis.

Page 17: Registry Part 3

Question 5• In order to detect an MRI signal there must be:

– A. Transverse magnetization that is zero– B. Longitudinal magnetization that is non zero– C. Longitudinal magnetization that is out of phase– D. Transverse magnetization that is in phase

Page 18: Registry Part 3

Question 6• The signal induced in the receiver coil following the RF pulse is known as:

– A. Gradients– B. FID– C. FOV– D. FUN

Page 19: Registry Part 3

Chemical Shift• Chemical shift is a difference in precessional frequencies. It is field

strength dependent.• Let’s take a look at the chemical shift between fat and water.• Do you know what the chemical shift between fat and water is in ppm?

– 3.5ppm. This number is constant.

What about at 1.5T?1.5T * 42.6MHz/T=63.8663.86=1ppmTherefore, 63.86 Hz * 3.5= 224Hz 224Hz is the difference in frequency between fat and water at 1.5T

Page 20: Registry Part 3

Contrast

Part IVAmanda Golsch BSc,RT(R)

(MR)

Page 21: Registry Part 3

Tissue Characteristics• The properties of human tissue (fat and water) contribute to the image

contrast that we see in MR images.• These properties include proton density, T1 and T2 relaxation times, and

flow/motion.

Page 22: Registry Part 3

Relaxation• During the relaxation period that occurs after the RF pulse, there are

two simultaneous yet independent processes that occur1. T1 Relaxation or Longitudinal Recovery2. T2 Relaxation or Transverse Decay

Page 23: Registry Part 3

Transverse Decay• T2 or spin-spin relaxation occurs when the RF pulse is removed. It is due

to an exchange of energy between the spins which results in dephasing.• T2 relaxation is the loss or decay of transverse magnetization.• The decay occurs exponentially and can be defined as the time it takes for

63% of a tissues transverse magnetization to decay.• T2 decay can’t be reversed or corrected and different tissues have

different T2 times.• Fat = Short T2 decay times• Water = Long T2 decay times• T2 time is measured in milliseconds

Page 24: Registry Part 3

Question 7• T2 relaxation is the time it takes for

– A. 55% of a tissue’s transverse magnetization to decay.– B. 63% of a tissue’s transverse magnetization to decay.– C. 75% of a tissue’s transverse magnetization to decay.– D. 99% of a tissue’s transverse magnetization to decay.

Page 25: Registry Part 3

T2* and T2 Decay• Inhomogeneities within the magnetic field and within tissues result in rapid

dephasing. In addition, the hydrogen in fat and water precess at different frequencies (chemical shift).

• T2* is the combination of T2, field inhomogeneities, and chemical shift. All of these factors result in the rapid loss of transverse magnetization.

• The FID decays at a rate based on T2*• T2 is evaluated on spin echo images.• T2* is evaluated on gradient echo images. Note that there are more

susceptibility artifacts on T2* images. Why?

– There is no 180 degree refocusing pulse that helps to correct field inhomogeneities and chemical shift.

Page 26: Registry Part 3

T1 Relaxation• T1 relaxation occurs at the same time as T2 decay yet it is an independent

process.• The longitudinal magnetization regrows along the Z axis as high energy

spins return to a low energy state.• It is defined as the time it takes for 63% of a tissue’s longitudinal

magnetization to recover.• T1 relaxation is also referred to as spin-lattice relaxation because there is a

loss of the spins energy to the surrounding molecular lattice.• Fat=Short T1 times.• Water=Long T1 times• T1 recovery can also be dependent on field strength and temperature.• Increased field strength=Lengthens T1 recovery time.• Increasing a tissues temperature=Lengthens T1 recovery time.

Page 27: Registry Part 3

Spin Density• Spin density represents the number of hydrogen protons in a given volume

of tissue.• The proton density is related to the spin excess or the NMV.• CSF has the highest proton (spin) density.

Page 28: Registry Part 3

Flow and Motion• Flowing blood and motion can effect image contrast.• Flow/motion can cause phase ghosting artifacts as well as extraneous

signals in images.

– This will be discussed in further detail in a later presentation.