Radcliffe Cardiology · This system allows the use of any 0.014" coronary guidewire appropriate for...

8
Summer 2017 • Promotional Supplement www.ICRjournal.com Next Generation FFR Microcatheter Technology Proceedings of presentations at EuroPCR 2017, Paris, France. Session summaries by Katrina Mountfort Medical Writer, Radcliffe Cardiology Radcliffe Cardiology Lifelong Learning for Cardiovascular Professionals

Transcript of Radcliffe Cardiology · This system allows the use of any 0.014" coronary guidewire appropriate for...

Summer 2017 • Promotional Supplement www.ICRjournal.com

Next Generation FFR Microcatheter Technology

Proceedings of presentations at EuroPCR 2017, Paris, France.

Session summaries by Katrina Mountfort Medical Writer, Radcliffe Cardiology

Radcliffe CardiologyLifelong Learning for Cardiovascular Professionals

Euro PCR summary (M016) FINAL.indd 1 18/09/2017 21:24

2

Proceedings of presentations at EuroPCR 2017

INTERVENTIONAL CARDIOLOGY REVIEW

AbstractFractional flow reserve (FFR) is the mainstay of functional haemodynamic assessment of coronary artery lesions, guiding decisions in

percutaneous coronary interventions (PCI). The RXi® rapid exchange FFR system, featuring an ultrathin monorail pressure microcatheter

(Navvus™) has the potential to simplify PCI procedures. Data from two studies sponsored by ACIST Medical Systems evaluating the clinical

utility of the microcatheter system were presented at EuroPCR, which took place over 16–19 May 2017 in Paris. Early data from the FFR-

Stent Evaluated at Rotterdam Cardiology Hospital (FFR-SEARCH) registry have indicated that post-PCI, almost half of patients have FFR

values below 0.90 even when stent placement appears fine on angiography. This registry is noteworthy for including a high proportion of

unstable patients. The Assessment of Catheter-based Interrogation and Standard Techniques for Fractional Flow Reserve measurement

(ACIST-FFR) study has shown that the microcatheter system provides a modestly lower FFR value compared with the traditional pressure

wire, and an independent predictor of a difference between the two is the physiological severity of the lesion as measured by the Navvus

microcatheter, meaning that the clinical impact of the difference is minimal for most measurements. These findings add to the growing

body of evidence in support of the microcatheter FFR system and have prompted further research into optimising procedures.

KeywordsFractional flow reserve, microcatheter, coronary artery disease

Citation: Interventional Cardiology Review, 2017;Promotional Supplement:2–7 DOI: 10.15420/icr.2017:12:2.S1:

Next Generation FFR Microcatheter Technology

Proceedings of presentations at EuroPCR 2017, Paris, France.

Expert reviewers: Roberto Diletti MD, PhD1 and Matthew J Price MD2

1. Thoraxcenter, Erasmus University, Rotterdam, the Netherlands; 2. Director, Cardiac Catheterization Laboratory, Scripps Clinic, La Jolla, California, USA

Fractional flow reserve (FFR) assesses the reduction in flow resulting

from a coronary artery stenosis. It is an essential tool in the evaluation

of the physiological significance and proper treatment of coronary

stenosis in the catheterisation laboratory since many stenotic lesions

identified at angiography are of intermediate severity, and their

impact on myocardial perfusion cannot be determined accurately

from angiography alone.1 Multiple studies have demonstrated the

efficacy and safety of FFR,2 including the 2009 Fractional Flow

Reserve Versus Angiography for Multivessel Evaluation (FAME)

study.3,4 However, the technique is underutilised: a 2012 report of

the National Cardiovascular Data Registry reported that only 6.1  %

of patients in the US underwent FFR assessment prior to percutaneous

coronary intervention (PCI) for intermediate coronary stenosis

(40–70 % stenosis).5

Most FFR measurements are performed using a 0.014" guidewire

incorporating a distal pressure sensor. This is more difficult to guide

through tortuous lesions. In addition, assessment of diffuse disease or

multiple lesions requires wire pullback, which results in the loss of wire

position. As the use of FFR is increasing worldwide and increasingly

complex lesions are being treated, there is a need for improved

technologies to improve speed, accuracy and ease of use. In addition,

post-stenting FFR has been proposed as an effective tool to assess

the effect of treatment on coronary flow. However, FFR wiring after

stenting requires additional time and could lead to wire placing behind

the stents struts.

The ACIST RXi® rapid exchange system (ACIST Medical Systems)

features an ultrathin monorail microcatheter (Navvus™; ACIST Medical

Systems) with an optical pressure sensor located close to the distal

catheter tip (see Figure 1). This system allows the use of any 0.014"

coronary guidewire appropriate for the patient’s anatomy. It enables

repeated pullbacks and advancements, as well as the ability to move

the sensor up and down the artery to determine the site of change in

pressure gradient, without losing wire position.6,7 Furthermore, it allows

an easy and fast assessment of post-stenting FFR.

However, the larger diameter of the catheter (0.022" mean diameter,

compared with the 0.014" guidewire) could influence FFR measurement.

The ACIST Diagnosis of Coronary Arterial Disease with a Rapid Exchange

Monorail Pressure Sensor for the Measurement of Fractional Flow

Reserve (ACCESS-NZ) study compared the Navvus/RXi system with a

conventional pressure wire and found that fewer patients had clinically

significant (>±0.03) drift with the Navvus catheter (13 %) than with the

pressure wire (33 %) when FFR was measured with both systems.8

On-going studies are evaluating the clinical utility of the Navvus

microcatheter systems, including the FFR-Stent Evaluated at Rotterdam

Cardiology Hospital (FFR-SEARCH) registry and the Assessment of

Catheter-based Interrogation and Standard Techniques for Fractional

Flow Reserve measurement (ACIST–FFR) study. This article describes

two presentations given at EuroPCR (16–19 May 2017, Paris) that

discussed the latest findings of these studies. n

Euro PCR summary (M016) FINAL.indd 2 18/09/2017 21:24

Next Generation FFR Microcatheter Technology

INTERVENTIONAL CARDIOLOGY REVIEW 3

Proceedings of presentations at EuroPCR 2017

Intracoronary measurement of myocardial FFR is a simple, reliable

and reproducible index of the physiological significance of coronary

stenosis. In addition, FFR assessment can be used to guide PCI,

improving procedural outcomes. However the impact of FFR values

after PCI on clinical outcomes remains unclear.

The FFR-SEARCH study aimed to determine the impact of post-PCI

FFR values on long-term clinical outcomes in a prospective all-comer

registry study.9 All patients who underwent successful PCI were

eligible. An FFR assessment using the ACIST RXi system and Navvus

microcatheter was performed after each PCI when the angiographic

result was considered acceptable by the operator. Wire access to the

vessel was maintained and a pressure sensor FFR microcatheter was

inserted over the previously used coronary guide-wire. The Navvus

microcatheter allows FFR assessment after crossing the lesion with

any 0.014" guidewire of choice. A platinum marker band is located 2.5

mm from the tip and a fibre-optic sensor 2.5 mm from the marker band

(5 mm from the tip).7 This gives a profile comparable to 0.022" diameter

at the lesion site (see Figure 1).

The study screened 1,512 patients undergoing PCI. Of these, 512

patients were excluded: 156 because of unstable disease, 148 because

of operator decision, 129 because vessels were too small and 79 for

other reasons. Of the 1,000 patients who underwent assessment using

the Navvus microcatheter, the wire did not cross the lesion in 28 cases,

equipment failure occurred in 11 cases and two patients had an adverse

response to adenosine. Of the remaining patients, 1,165 lesions had a

successful post-PCI FFR measurement. Failure to cross the lesion was

reported in 14 procedures, nine patients were not assessed because of

unstable disease, 22 lesions were not measured because of operator

decision and 28 for other reasons. In 109 lesions the vessel was too

small and there was one case of equipment failure. The baseline

characteristics were reported on 1000 patients and procedural results

were reported on 959 patients with 1,165 lesions. The average age at

baseline was 64.6 ± 11.8 years and patients were predominantly male

(72.5 %). Of note, clinical presentations were evenly distributed between

stable angina (30.4 %), unstable angina/non ST-elevation MI (36.7 %) and

acute MI (32.9 %), i.e. >60 % of the patients involved were unstable.

Where possible, FFR was performed at least 20 mm from the most

distal stent then a pullback was performed to obtain pressure

gradients on the distal and proximal stent edge border as well as to

assess overall drift at the ostium. The average drift value was low,

0.011 ± 0.015, and the average procedural time per lesion was 5.0 ±

1.4 minutes, which included the full duration of the adenosine infusion,

showing that FFR can be used quickly to evaluate PCI results. Failure

to cross the post-PCI lesion occurred in 42 of 1,207 (3.5 %) attempted

lesions, mostly because of tortuosity and calcification, and there were

no Navvus-related complications reported.

Dr Diletti presented preliminary outcome data from this study. Patient-

level analysis showed that almost half (48 %) of patients had an FFR

below 0.90. Lesion level assessment showed that in 43.3 % of cases,

the FFR was ≤0.90. In 7.7 % of lesions post-stenting FFR was below the

threshold for ischaemia (i.e. ≤0.80). The number of lesions per 0.01 FFR

increment was 660 lesions (57 %) with FFR >0.90; 505 lesions (43 %)

≤0.90; 231 lesions (20 %) ≤0.85 and 90 lesions (8 %) ≤0.80 (see Figure 2).

The cut-off of 0.90 could help guide procedural optimisation.

The primary outcome measure was major adverse coronary events

(MACE), including all-cause mortality, Q-wave MI, target lesion

revascularisation, target vessel revascularisation, stent thrombosis and

any revascularisation at 30 days. Of 798 (89 %) patients with complete

30-day follow-up, MACE occurred in only 15 (1.9 %). The MACE rate for

Routine Fractional Flow Reserve Measurement after Percutaneous Coronary Intervention: The FFR-SEARCH Study

Presented by Roberto Di lett i

Thoraxcenter, Erasmus University, Rotterdam, the Netherlands

Figure 1: The Navvus Microcatheter

Figure 2: The FFR-SEARCH Study: Number of Lesions Per 0.01 FFR Increment

Figure 3: The FFR-SEARCH Study: MACE at 30-day Follow-up

Transducer5mm from tip

Platinum markerband2.5 mm

from trp Femoral markers80/100 cm

Working length 150 cm(total length 335 cm)

00.6 0.7 0.8

FFR value(N = 1165 lesions)

Num

ber

of

lesi

ons

• 660 lesions (56.7 %) > 0.90• 505 lesions (43.3 %) ≤ 0.90• 231 lesions (19.8 %) ≤ 0.85• 90 lesions (7.7 %) ≤ 0.80

0.9 1.0 1.1

20

40

60

80

100

Source: Diletti, 20179

Source: Diletti, 20179

90 %0 10

MACE de�ned as: all-cause mortality, MI, any revascularisation

P = 0.792FFR ≤ 0.80FFR 0.81–0.85FFR 0.86–0.90FFR > 0.90

20 30

92 %

94 %

96 %

98 %

100 %

Euro PCR summary (M016) FINAL.indd 3 18/09/2017 21:24

Proceedings of presentations at EuroPCR 2017

INTERVENTIONAL CARDIOLOGY REVIEW4

The use of FFR guidance to identify and treat functionally significant

coronary lesions has been shown to reduce the risk of MACE and

save healthcare resources compared with angiographic guidance

alone.4 Two types of FFR technologies are commercially available.

Pressure wire technology involves a specially constructed 0.014"

wire. Microcatheter technology employs a low-profile catheter with

a pressure sensor incorporating fibre-optic technology into the distal

end, giving a profile comparable to 0.022" diameter at the lesion

site. The latter is more convenient and may overcome some of the

limitations associated with conventional pressure wire systems, but

the larger profile may influence coronary haemodynamics, an effect

that might also depend on lesion and vessel characteristics. There is

a need to investigate these unresolved issues in a large, multicentre,

prospective study including a wide range of vessel and lesion types

and utilising an independent core laboratory.

The ACIST-FFR prospective, open label study aimed to assess the

differences between FFR measured by a microcatheter and commercially

available 0.014" pressure wires in the setting of clinically relevant

vessel diameters and lesion lengths.10,11 A Navvus microcatheter was

used and compared to two different commercially available pressure

wire systems: Abbott-St. Jude Medical and Philips-Volcano. The primary

endpoint was the difference in agreement between microcatheter and

pressure wire FFR as assessed by Bland-Altman analysis. The core

laboratory (Cardiovascular Research Foundation, New York) assessed

FFR and provided quantitative coronary angiography. Independent

analyses were performed by Stanford University.

The inclusion criteria were patients with stable ischaemia; a vessel

with a Thrombolysis in Myocardial Infarction flow of 3; a de novo lesion

that a physician has considered to have a clinical indication for FFR

measurement; and operator-assessed reference vessel diameter (RVD)

of the target lesion ≥2.25 mm. Exclusion criteria were acute ST-elevation

or non-ST-elevation MI, New York Heart Association Class 4 severe heart

failure; a target vessel with an angiographically visible or suspected

thrombus; a target lesion within a bypass graft; angiographic evidence

of a dissection prior to initiation of pressure wire measurements and a

target vessel containing excessive tortuosity or calcification.

The pressure wire was delivered, equalised and advanced across the

lesion, a resting aortic pressure ratio (Pd/Pa) measured, then

the Navvus microcatheter was advanced over the pressure wire,

equalised, advanced past the lesion and Pd/Pa and FFR were

measured simultaneously with both systems. The microcatheter drift

was assessed; it was then withdrawn and FFR and drift were measured

with the pressure wire alone.

The study enrolled 245 patients at 11 sites in the US. Of these, 13

subjects consented but FFR was not measured, FFR was measured but

no drift assessed in nine and FFR was attempted but not measured in

13 because of failure to cross or device malfunction. The remaining

210 subjects underwent paired FFR assessment. Of these, 41 were

excluded by the core laboratory either for ventricularisation, dampening

of the proximal pressure tracings and drift from one or both of the

systems without proper re-measurement. This left a primary cohort of

169 patients who met the core laboratory criteria.

The baseline patient characteristics were typical of PCI study

populations: average age was 68 years and 37  % had a history of

diabetes. Baseline angiographic characteristics were as follows: mean

diameter stenosis was 47  %, i.e., angiographically indeterminate

lesions. The average lesion length was 15.3  mm and the average

RVD of the target lesion was 2.8  mm. Of the enrolled target lesions,

30  % were <2.5 mm in diameter. Therefore small vessels were well

represented in this study – prior microcatheter FFR clinical studies had

only included vessels with RVD ≥2.5 mm. The mean pressure wire FFR

was 0.83, again representative of a patient population with a clinical

indication for FFR.

Compared with the pressure wire, the microcatheter gave consistent

and modestly lower FFR measurements with an average bias of −0.022

(see Figure 4). In terms of the secondary endpoint, the difference in

pressure wire FFR with and without the microcatheter on the pressure

wire, the average bias was −0.03. A sensitivity analysis of all site-reported

data (i.e., all patients with paired FFRs) showed that the bias was similar

to that of the primary analysis cohort (−0.025). A strong correlation was

seen between the microcatheter and the pressure wire FFR (Pearson

Primary Results of the Assessment of Catheter-Based Interrogation and Standard Techniques for Fractional Flow Reserve Measurement Study:

The ACIST-FFR Study

Presented by Matthew Price

Cardiac Catheterization Laboratory Scripps Clinic, La Jolla, CA, USA

each subgroup at 30 days did not differ significantly between groups:

FFR ≤0.80 2.8  %, FFR 0.81–0.85 2.6  %, FFR 0.86–0.90 2.0  % and FFR

>0.90 1.5  % (see Figure 3). Although not statistically significant, it is

interesting to observe the initial trend of the lower the post-PCI FFR

value, the higher the reported MACE rate at 30 days. The study design

intended to provide a clear message of clinical outcomes at 2 years.

Dr Diletti summarised the key points of the presentation as follows:

• Microcatheter-based FFR post-PCI allows a safe and rapid

physiological assessment of the treatment result.

• A relevant percentage of the stented vessels have a final FFR <0.90.

• PCI optimisation by microcatheter-based FFR may improve

clinical outcome.

In the subsequent audience discussion, some experts at EuroPCR

questioned whether the FFR was detecting suboptimal stent

placement, or if other factors such as microcirculatory problems

might be contributing. Others questioned the next step to optimising

procedures. The upcoming FFR REACT randomised study aims to

evaluate the clinical impact of intravascular-ultrasound-directed FFR

optimisation of PCI. n

Euro PCR summary (M016) FINAL.indd 4 18/09/2017 21:24

Next Generation FFR Microcatheter Technology

INTERVENTIONAL CARDIOLOGY REVIEW 5

The increased use of FFR – particularly in multivessel disease and tandem

lesions – has necessitated more efficient and accurate techniques, and

it is increasingly important to understand the benefits and limitations

of available FFR technologies. The use of microcatheter-based FFR

technology can overcome the limitations of pressure wire technology

such as accessibility in challenging anatomy, maintaining wire position,

pressure-measurement drift and ease of obtaining post-intervention

FFR. The microcatheter system can thus simplify PCI procedures as

well as avoid technical errors. Preliminary data from the FFR-SEARCH

registry has indicated that post-PCI, almost half of patients have FFR

values below 0.90 even when stent placement appears adequate on

angiography. The ACIST-FFR trial has shown that the clinical impact of

the difference between microcatheter and conventional pressure wire

FFR measurements is minimal in most cases.

These data have added to the body of clinical evidence supporting the

use of microcatheter-based FFR. Furthermore, while FFR has previously

been considered a diagnostic tool to assess whether a lesion has to be

treated or not, these studies have led to an expanded view of FFR as

a tool to understand and evaluate the results of PCI procedures. These

studies have stimulated further research into optimising procedures

and should facilitate the more widespread use of FFR. n

Summary and Concluding Remarks

coefficient 0.901, p<0.001). Since the bias is less at FFR values within

the grey zone and higher, the clinical impact of the difference may be

minimal. For a threshold of 0.80 or less, the diagnostic agreement was

81 %. There were only five cases (2.9 %) where the pressure wire FFR

exceeded 0.80 but the microcatheter FFR was <0.75.

Because of the large study cohort, it was possible to identify predictors

of bias. On univariate analysis, smaller RVD tended to predict greater

bias and longer lesion length was a significant predictor of greater

bias. However, using multivariate analysis incorporating microcatheter

FFR, only the FFR measured by the microcatheter was found as the

independent predictor of bias. Mean signal drift was similar between

the two technologies, but clinically significant drift (defined as >0.03)

was numerically higher with the pressure wire (7.4 % vs. 3.5 %, p=0.10;

see Figure 5).

In summary:

• A pressure-monitoring microcatheter provides a modestly lower

FFR value compared with a traditional pressure wire, with an

average bias of −0.02.

• Using univariate analysis, smaller RVD and greater lesion length

tend to predict a lower FFR measured by the microcatheter.

• Using multivariate analysis, lower FFR as measured by the

microcatheter is the only predictor of a greater difference between

the two measurements.

• Since the bias is less at higher FFR values (e.g., the grey zone and

above), the clinical impact of this difference for those scenarios is

likely minimal.

This study confirms and expands previous findings and provides robust

evidence for the diagnostic capabilities of microcatheter FFR. n

Figure 4: The ACIST-FFR Study Microcatheter Versus Pressure Wire FFR, Core Lab Values, by Bland Altman Analysis

Figure 5: ACIST-FFR Study: Magnitude and Rate of Drift

-0.15

0.4 0.5 0.6 0.7 0.8

Mean FFR

N = 169

Upper 95% LOA: 0.071

Lower 95% LOA: -0.12

Bias: -0.022(95%CI: -0.029,-0.015)

FFR

Diff

eren

ce

(Mic

roca

thet

er-P

W)

0.9 1.0

-0.10

-0.05

0.00

0.05

0.10

0.15

0.20

Source: Fearon et al., 201711

Source: Fearon et al., 201711

Mean Signal Drift

Pressure WireMicrocatheter

0.015 0.015

3.5 %

7.4 %

p=0.10

Clinically Signi�cant Drift (>0.03)

Euro PCR summary (M016) FINAL.indd 5 18/09/2017 21:24

Proceedings of presentations at EuroPCR 2017

INTERVENTIONAL CARDIOLOGY REVIEW6

1. Kern MJ, Lerman A, Bech JW, et al. Physiological assessment of coronary artery disease in the cardiac catheterization laboratory: a scientific statement from the American Heart Association Committee on Diagnostic and Interventional Cardiac Catheterization, Council on Clinical Cardiology. Circulation 2006;114:1321–41. DOI: 10.1161/circulationaha.106.177276; PMID: 16940193.

2. Christou MA, Siontis GC, Katritsis DG, et al. Meta-analysis of fractional flow reserve versus quantitative coronary angiography and noninvasive imaging for evaluation of myocardial ischemia. Am J Cardiol 2007;99:450–6. DOI: 10.1016/j.amjcard.2006.09.092; PMID: 17293182.

3. Tonino PA, De Bruyne B, Pijls NH, et al. Fractional flow reserve versus angiography for guiding percutaneous coronary intervention. N Engl J Med 2009;360:213–24. DOI: 10.1056/NEJMoa0807611; PMID: 19144937.

4. van Nunen LX, Zimmermann FM, Tonino PA, et al. Fractional flow reserve versus angiography for guidance of PCI in

patients with multivessel coronary artery disease (FAME): 5-year follow-up of a randomised controlled trial. Lancet 2015;386:1853–60. DOI: 10.1016/s0140-6736(15)00057-4; PMID: 26333474.

5. Dattilo PB, Prasad A, Honeycutt E, et al., Contemporary patterns of fractional flow reserve and intravascular ultrasound use among patients undergoing percutaneous coronary intervention in the United States: insights from the National Cardiovascular Data Registry, J Am Coll Cardiol 2012;60:2337–9. DOI: 10.1016/j.jacc.2012.08.990; PMID: 23194945.

6. Briguori C, Nishida T, Adamian M, et al. Assessment of the functional significance of coronary lesions using a monorail catheter. J Invasive Cardiol 2001;13:279–86. PMID: 11287712.

7. Diletti R, Van Mieghem NM, Valgimigli M, et al. Rapid exchange ultra-thin microcatheter using fibre-optic sensing technology for measurement of intracoronary fractional flow reserve. EuroIntervention 2015;11:428–32. DOI: 10.4244/eijy15m05_09; PMID: 26013582.

8. Menon M, Jaffe W, Watson T, et al. Assessment of coronary fractional flow reserve using a monorail pressure catheter: the first-in-human ACCESS-NZ trial. EuroIntervention 2015;11:257–63. DOI: 10.4244/eijv11i3a51; PMID: 26196752.

9. Diletti R, on behalf of the Thoraxcenter Investigators: van Bommel R, Masdjedi K, van Zandvoort L, et al. Routine fractional flow reserve measurement after percutaneous coronary intervention: the FFR-SEARCH study. Presented at: EuroPCR, May 18 2017, Paris, France.

10. NCT02577484. Assessment of Catheter-based Interrogation and Standard Techniques for Fractional Flow Reserve Measurement (ACIST-FFR), https://clinicaltrials.gov/ct2/show/NCT02577484 (accessed 29 August 2017).

11. Fearon WF, Chambers JW, Seto AH, et al FFR: Trials and new techniques. Primary Results of the Assessment of Catheter-based Interrogation and Standard Techniques for Fractional Flow Reserve Measurement Study: The ACIST-FFR Study. Presented at: EuroPCR, May 16-19 2017, Paris.

Euro PCR summary (M016) FINAL.indd 6 18/09/2017 21:24

INTERVENTIONAL CARDIOLOGY REVIEW 7

Euro PCR summary (M016) FINAL.indd 7 18/09/2017 21:24

Bracco Group

ACIST, ACIST RXi and Navvus are trademarks of ACIST Medical Systems, Inc., registered in the US. ACIST Medical Systems, Inc., reserves the right to modify the specifi cations and features described herein, or discontinue manufacture of the product described at any time without prior notice or obligation. Please contact your authorized ACIST representative for the most current information. © 2017 ACIST Medical Systems, Inc. All Rights Reserved. P/N: 0617.630.01

For more information from ACIST Medical Systems, Inc., visit our website: www.acist.com

1. Price M. Primary results of the assessment of catheter-based interrogation and standard techniques for fractional fl ow reserve measurement study: The ACIST-FFR Study, paper presented at: EuroPCR 2017, May 16-19, 2017; Paris, France.

2. Compared with Navvus® FFR MicroCatheter before and after processing improvement in benchtop testing. Data on fi le at ACIST. May not be indicative of clinical performance.

Learn more at www.acistffr.com

Complete your FFR toolkit with the only FFR MicroCatheter.Navvus®, the proven1 alternative to pressure wires.

More deliverable2

18%Reduced lesionentry profi le2

More crossable2

35%Reduced

crossing force2

NOW

Euro PCR summary (M016) FINAL.indd 8 18/09/2017 21:24