Quality Assurance for the Small Water Lab - rusa-or.orgAssurance+for+the+small+lab.pdfQuality Manual...
Transcript of Quality Assurance for the Small Water Lab - rusa-or.orgAssurance+for+the+small+lab.pdfQuality Manual...
Department of Environmental Quality
Quality Assurance for the Small Water Lab
SOUTHERN OREGON WATER AND WASTEWATER SHORT SCHOOL
8/16/2018
Scott Hoatson – Oregon DEQ [email protected]
Department of Environmental Quality
Agenda – Not necessarily in this order
• Introduction• References• Appropriate Methods• Quality Assurance
• QA Expectations• Quality System• Documents and Records• Data Reporting
• Sampling and Storage
• QA / QC Requirements
Department of Environmental Quality
Purpose of Monitoring
• The goal of the NPDES and SDWA programs is to preserve and protect water quality and human health.
• Facilities are required to self monitor and report to the State.
• Data must be scientifically valid and defensible and of known and acceptable precision and accuracy.
Department of Environmental Quality
Where Do I Find Information ??
• Ecfr.gov
– 40 CFR Part 136.3 Identification of Test Procedures
– 40 CFR Part 136.7 Quality Assurance and Quality Control.
– 40 CFR Part 122.21 Application for a permit
• Handbook for the Analytical Quality Control In Water and Wastewater Laboratories (HB) – EPA-600/4-79-019, March 1979
Department of Environmental Quality
More References
• NPDES Compliance Inspection Manual (CIM) - EPA 305-X-04-001, July 2004
• QA Guidance for Self-Monitoring Laboratories (NPDES and WPCF) DEQ09-LAB-0071-QAG 4/25/2013
Department of Environmental Quality
QA References
• Field Sampling Reference Guide – DEQ86-LAB-0002-QAG 08/15/18
• DEQ– Quality Manual Template –
Department of Environmental Quality
40 CFR 136.3
– https://www.ecfr.gov
Table I A – Approved Microbiology methods
Table I B – Approved Inorganic Methods
Table I C – Approved Organic Methods
Table I D - Approved Pesticide Methods
Table IE - Approved Radiologic Methods
READ THE FOOTNOTES!!!!!!
Department of Environmental Quality
40 CFR 136.3 Part Dieux
• Table IF - Approved Methods for Pharmaceutical Pollutants
• Table IG - Approved Methods for Pesticide Active Ingredients
• Table IH - Approved Microbiological Methods for Ambient Water
• Table II - Required Containers, Preservation Techniques, and Holding Times
Department of Environmental Quality
Compliance Testing
• Must use methods in 40 CFR 136.3• Standard Methods references
– New editions– You need to get:
• A new book or buy individual methods• Subscribe to Standard methods on-line
Department of Environmental Quality
2017 Method Update Rule
• EPA re-evaluates methods approx. every 5 years. Goes for public comment before finalizing
• Signed - 8/7/2017 – Incorporated into 40 CFR part 136
• Updates Standard Methods – 22nd – 23rd
Editions• Fecal Coliforms by IDEXX Colilert 18• New MDL Procedure
Department of Environmental Quality
APPROPRIATE METHODS
Department of Environmental Quality
40 CFR part 136.3 Table 1 A
Department of Environmental Quality
40 CFR part 136.3 Table 1 B
Department of Environmental Quality
Very Outdated – Retire Them
Department of Environmental Quality
21 Edition – 2005, 22 Edition 201223rd Edition - 2017
Department of Environmental Quality
What about Hach Methods?
Department of Environmental Quality
Nitrate Example
• Hach Method 10206 Approved– Hach TNT 835 (test kit) is NOT approved by
itself• The approved methods have associated
quality control requirements
• Test Kit “procedure” does not.
Department of Environmental Quality
Hach Company TNTplusTM 830/831/832 Ammonia Method 10205
• “This method is equivalent to EPA 350.1, 351.1, and 351.2 for the purposes of regulatory reporting of Ammonia (as nitrogen) and Total Kjeldahl Nitrogen.”
• EPA said chemistry is equivalent (2007 EPA Memo)
• Hach method 10205 not in CFR. OK to cite 350.1 for Ammonia IF….
Department of Environmental Quality
IF…
• You follow what is in 10205, not just what is in TNT 830/831.
• You should have a copy of EPA 350.1 (you can easily get from the internet. – Google or other another search engine
Department of Environmental Quality
In Summary
• There is a Significant difference between the Hach Method and Test Kit instructions.
• Hach products are acceptable for use in many compliance tests provided The Method is followed.
• Also See QC requirements in 40 CFR 136.7
Department of Environmental Quality
Quality Assurance
• QA refers to a total program for ensuring the reliability of data by utilizing administrative and technical procedures and policies regarding personnel, resources, and facilities. - CIM
• Quality Control (QC) are the activities related to ensuring a specific process meets requirements of project, program or data user.
Department of Environmental Quality
Purpose of QA/QC
• ALL ABOUT TRACEABILITY:
– Ability to definitively demonstrate compliance
– Recreate the sampling and analytical activities
– Documentation should be able to stand on its own merit without explanation.
– Data of known and documented quality.
– Demonstration of laboratory competency
Department of Environmental Quality
Quality Assurance Expectations (CIM)
• QA is required for all functions bearing on environmental measurements and includes activities such as:– project/study definition– sample collection and tracking– laboratory analysis– data validation analysis, reduction, and reporting – documentation– data storage systems.
Department of Environmental Quality
40 CFR 136.7 Quality assurance and quality control
• The permittee/laboratory shall use suitable QA/QC procedures when conducting compliance analyses with any part 136 chemical method or an alternative method specified by the permitting authority.
• These QA/QC procedures are generally included in the analytical method or may be part of the methods compendium for approved Part 136 methods from a consensus organization.
Department of Environmental Quality
40 CFR 136.7 continued
• The permittee/laboratory shall follow these QA/QC procedures, as described in the method or methods compendium. If the method lacks QA/QC procedures, the permittee/laboratory has the following options to comply with the QA/QC requirements:
Department of Environmental Quality
40 CFR 136.7 continued
• (a) Refer to and follow the QA/QC published in the “equivalent” EPA method for that parameter that has such QA/QC procedures;
• (b) Refer to the appropriate QA/QC section(s) of an approved part 136 method from a consensus organization compendium;
Department of Environmental Quality
40 CFR 136.7 continued
• (c)(1) Incorporate the following twelve quality control elements, where applicable, into the laboratory's documented standard operating procedure (SOP) for performing compliance analyses when using an approved part 136 method when the method lacks such QA/QC procedures. One or more of the twelve QC elements may not apply to a given method and may be omitted if a written rationale is provided indicating why the element(s) is/are inappropriate for a specific method
Department of Environmental Quality
QA/QC in the CFR – 40 CFR 136.7
• Included in most methods
• Use QA/QC procedures in Compendium methods (e.g. Standard Methods, ASTM, etc)
• If not in above, incorporate 12 items
Department of Environmental Quality
The Dirty Dozen (40 CFR 136.7)
1. Demonstration of Capability
2. Method Detection Limit
3. Method Blanks (MB)
4. Laboratory Control Sample (LCS)
5. Matrix Spike (MS) and MS Duplicate (MSD)
6. Internal Standards, surrogates, tracers
Department of Environmental Quality
The Dirty Dozen (40 CFR 136.7) – cont.
7. Calibration and verification (Initial (ICV) and Continuing (CCV))
8. Control Charts (trend analysis)
9. Corrective action (root cause analysis)
10.QC Acceptance Criteria
11.Batch definitions (may drive QC frequency)
12.Frequency for conducting QC activities
Department of Environmental Quality
Before You Start Running Samples
Department of Environmental Quality
Method Validation
• Method Selection– Method appropriate to meet your needs– In our case methods found in 40 CFR 136.3
(NPDES) or 40 CFR 141 (SDWA)– www.ecfr.gov
• Develop Procedure (ATP)– 40 CFR 136.5
• Traceability of Measurements
• Show that it works (Validate)
Department of Environmental Quality
Traceability Components
• Analyst• Date/time• Instrument• Standards• Reagents• Calibration• Analytical QC• Sample QC• Records Management
Department of Environmental Quality
Method Validation
• Calibration– Titrations – Standardization of Titrant– Gravimetric – Balance Calibration– Microbiology
• Autoclave• Volumetric equipment• Water baths• Ovens• UV instruments
– Most other methods – Multipoint calibration
Department of Environmental Quality
Method Validation
• Detection and Quantitation levels– MDL (method detection limit) – 40 CFR 136
Appendix B Revision 2– Reporting Limits (Quantitation Limits)
• Precision and Bias determination– Can also be used for Demonstration of capability
for the one analyst.• Demonstration of Capability
– Analyst specific
Department of Environmental Quality
Reporting (QL) / Detection Limits (DL)
Detection Limit - What does it mean?• Within Calibration Range.
– QL – specified in permits – share with lab– LOQ – Limit of Quantitation– RL – Reporting Limit
• Estimated – Needs to be Qualified– LOD – Limit of Detection– MDL - Method Detection Limit
Department of Environmental Quality
MDLs
40 CFR 136 Appendix B Revision 2
• New procedure in 2017 MUR. • Takes blank background levels
consideration
Department of Environmental Quality
Data Qualifiers
• Data Qualifiers provide useful information to data users. Lab should not be hiding problems.
• Not always bad. e.g. Tight control limits may lead to more qualifiers
Pitfalls• Lab does not use qualifiers
Department of Environmental Quality
Quality Assurance
• Quality Control Procedures– Calibration Verifications– Positive Controls (LCS, MS, SRM)– Negative Controls(Blanks, non-target bacteria,
etc.)
• Proficiency Testing
• Data Review
Department of Environmental Quality
Quality Control (QC) Elements
Department of Environmental Quality
Laboratory Quality Control (QC) Elements
• Positive Control – Show process performs correctly using a known test subject. – Lab Control Sample– Known coliforms
• Negative Controls – Shows that process doesn’t interfere with determination.– Blanks – Contamination– Non-coliform bacteria – False positives
Department of Environmental Quality
Quality Control (QC) Elements
• BLANKS
• DUPLICATES
• STANDARD REFERENCE MATERIALS (good for validating method or DOC)
• BLANK SPIKES (Lab control samples)
• SPIKE SAMPLES
• SPLIT SAMPLES AND CERTIFICATION PROGRAMS (eg. DMRQA)
Department of Environmental Quality
QC Elements - Negative Controls
• Blanks– Equipment Blanks– Instrument Blanks– Method Blanks (Includes sterile growth media)– Trip Blanks (Volatile Organics)
• Non-Target Bacteria– Media supports growth – Differentiate between types of bacteria.
Department of Environmental Quality
QC Elements - Positive Controls
• Accuracy is determined by analyzing reference standards or spiked samples and is expressed as standard deviation form the theoretical value or average percent recovery.– Laboratory Control Samples– Matrix Spikes– SRM (Standard Reference Material)– Acceptance limits are in referenced method or
calculated from Control Charts.
Department of Environmental Quality
QC Elements – Positive Controls
• Precision: Measure of the degree of agreement among replicate analyses of a sample.
• Determined by running duplicates and is expressed as standard deviation, range, or relative percent difference (RPD).– Lab Duplicates (batch), Split Samples
(Periodic), Duplicate Spikes (if duplicates not performed)
Department of Environmental Quality
QA/QC Compliance inspection
• PRECISION, ACCURACY, CONTROL– Multiple replicates:
• blanks,• duplicates, • spikes, • and splits?
– Plotted precision and accuracy control charts?
Department of Environmental Quality
Quality Manual
• Should support the laboratory data with an adequate documentation program that provides valid records of the control measures applied to all factors bearing on the final results of investigations.
• Because of the importance of laboratory analyses in determining practical courses of action that may be followed, quality assurance programs to insure the reliability of the water and wastewater data are essential.
Department of Environmental Quality
Quality Manual (CIM)
• The 40 Code of Federal Regulations (CFR) Section 122.41(e) (conditions applicable to all permits), requires adequate laboratory and process controls, including appropriate QA procedures.
• Each permittee's laboratory must have a QA/QC program.
Department of Environmental Quality
Quality Manual Cont (CIM)
• The laboratory must document the QA program in a written QA/QC manual and the lab should make it available to all personnel responsible for sample analyses.
• The manual must clearly identify the individuals involved in the QA program and document their responsibilities.
Department of Environmental Quality
Quality Assurance Manual
• Simply Put: Documentation of the laboratories quality assurance program activities.
• It should be a living document and not just written and put on a shelf
• Review annually to make sure it reflects laboratory practices – only update if needed.
Department of Environmental Quality
It doesn’t have to be huge
But you want it to have value
Department of Environmental Quality
Quality Manual Templates
• DEQ Website: www.oregon.gov/deq/FilterDocs/QAManualTemplate.docx (borrowed by permission from VA)
• DEQ Lab: http://www.oregon.gov/deq/FilterDocs/DEQ91LAB0006QMP.pdf
• TNI – Compliant to 2009 TNI Standard available in MS Word http://www.nelac-institute.org/content/NELAP/howto.php
• Via Google: Search: “Environmental Laboratory Quality Manual Template”)
Department of Environmental Quality
Quality Manual
• Describes the overall quality program• Documents the Laboratory’s quality policies
and procedures• Provides a tool for ensuring personnel are
both knowledgeable of and committed to these policies and procedures
• Demonstrates management commitment
Department of Environmental Quality
Bottom Line
• Generate data of “Known and Documented Quality”
Department of Environmental Quality
3 Fundamental Concepts
• Traceability• Process Improvement• Preventative Action
Department of Environmental Quality
QM Table of Contents
• Introduction – Scope/Policy• Organization and Management
– Who can sign reports, QA documents• Position Descriptions / Roles• Document Control and Record Management• Subcontracting of Environmental Tests• Purchasing
Department of Environmental Quality
QM Table of Contents – Part 2
• Control of Non-Conforming Work• Improvement• Corrective Action• Preventive Action• Control of Records• Audits• Management Reviews• Data Integrity• Personnel
Department of Environmental Quality
QM Table of Contents Part 3
• Accommodations and Environmental Conditions (aka Facilities)
• Environmental Methods and Method validation
• Calibrations• Measurement Traceability• Sample Collection• Sample Handling• Quality Assurance• Reporting Results
Department of Environmental Quality
Control of Documents
• Describes the overall review, approval and control process– Can be electronic– May refer to a separate SOP.– Only current versions in use and accessible to all
• Standard Operating Procedures (SOPs)– All procedures documented
• Quality Manual– Documentation of the quality system
Department of Environmental Quality
Documented Procedures
• Sampling • Equipment calibration and maintenance• Analytical methods• QA/QC activities (40 CFR 136.7)• Data handling and reporting
• May be in electronic format (e.g. pdf)
Department of Environmental Quality
Sampling Procedures
• Collection and analysis schedule for permitted parameters (and non-permitted)
• Collection Locations
• Sample Types (Grab, composite, sampler instructions) 40 CFR part 122.21
• Field Filtration – Dissolved metals
Department of Environmental Quality
Sample Handling Procedures
• Proper preservation, containers, holding times, AND proper sample location,
• Security
• Chain-of-Custody• Proper storage
• Documentation
Department of Environmental Quality
Facilities and Equipment Procedures
• Operating instructions for equipment such as balances, meters, incubators, and samplers.
• Adequate equipment (type and specifications)
• Maintenance schedules on major equipment • Cleaning (glassware, micro apparatus, etc)• Daily balance checks and annual service• Ovens • Daily water checks
Department of Environmental Quality
Facilities and Equipment Procedures
• Pure Water? – Daily Conductivity Checks• Adequate equipment? • Equipment maintenance and calibration?
(Maintain log and schedule)• Clean area?• Adequate Standards and Reagents within
shelf-life? – Maintain Log• Documentation?
Department of Environmental Quality
Laboratory Procedures
• EPA Approved procedures (40 CFR Part 136) Specified in Permit? – Written SOPs – Necessary unless following
method to the letter.– Calibration and maintenance– QA/QC Procedures used and adequate– Data Review/Approval
• Contract Laboratory Information• Documentation?
Department of Environmental Quality
Proficiency Testing
• Blind Check Samples to assess routine performance.– Run like your regular samples– Do not run at same time as Check PT samples
• Major Facilities – Require DMRQA Studies
• Recommended for others.– Water Pollution
Department of Environmental Quality
Proficiency Testing
• Corrective action required for failures– Why did it fail
– What did you do to correct the problem• Does it prevent recurrence?
– Was any other data affected?
– Must demonstrate you are on control
Department of Environmental Quality
Data Handling
• Data sheets (or log books) must be dated, initialed by analyst, legible, and written in ink – Drawing a single line through the incorrect
recording should make any corrections, insertion of corrected result, initialed and dated, and contain a comment (NO CORRECTION FLUID OR TAPE)
Department of Environmental Quality
Data Handling and Reporting
– ROUNDING to nearest EVEN number?• EXAMPLES: 4.45 = 4.4 and 4.35 = 4.4
– Significant Figures for each analysis (either 2 or 3)?
– Cross-checking calculations?– Correct formulas – Have spreadsheets been
verified and locked?– Proper Control Chart Use– 3 Years of Records
Department of Environmental Quality
Data Handling
• Data sheets should include calibration measurements, incubator temperature records, and the time samples were placed in and removed from incubators, ovens, etc.
• Data sheets should include equations used for deriving final test results.
• Second Level Review – Another pair of eyes.• Qualify for QC failures, HT, preservation
etc*.
Department of Environmental Quality
Control of Records
Record Management
• General Definition: The practice of maintaining records of an organization from the time they are created up to their eventual disposal.
Department of Environmental Quality
What are Records?
• Raw Data files• Support equipment records• Standard Traceability log records• Calibration records• Other log books• Electronic files (instrument files)• Emails• Quality records: internal audits, management reviews,
corrective actions and preventative actions.• Etc.
• Technically “Records” include all of the above plus “Documents”
Department of Environmental Quality
Record Management
• Records management and storage– Identification– Collection– Archives– Filing– Access– Disposal
Department of Environmental Quality
Records ManagementTNI 2009 V1M2 4.13.3: Control of Records
• The laboratory shall retain all records for a minimum of five (5) years from generation of the last entry in the records. (Drinking Water – 10 years)
• Records that are stored only on electronic media shall be supported by the hardware and software necessary for their retrieval.
• The laboratory shall establish a record keeping system that allows the history of the sample and associated data to be readily understood through the documentation.
40 CFR 122.41 and 40 CFR 503 – 5 year retention
Department of Environmental Quality
Records Management System
• This system shall produce unequivocal, accurate records that document all laboratory activities such as:– laboratory facilities, equipment– analytical methods and data – related laboratory activities, such as
• sample receipt, • sample preparation• data verification• inter-laboratory transfers of samples and/or
extracts.
Department of Environmental Quality
Do you have all of the Records?
Department of Environmental Quality
Fundamental Question
• Based on the documented information, can you say with confidence that the permitted facility is in compliance or not?
Department of Environmental Quality
Subcontracting Work
• When do you use subcontract labs?
• How do you communicate your requirements to the subcontractor?
• Give them copy of permit!!
• Be sure they are clear on QLs and method requirements (don’t report SW 846 methods)
Department of Environmental Quality
DO NOT USE the Yellow Pages Approach
• The Customer is ultimately responsible for the site / discharge / project.
• Not the Lab!!
Department of Environmental Quality
Third Party Labs (Sub Contracting)
• Request report with QC (with qualifiers)
• Recommend using accredited labs– Receive Audits every two years– Analyze 2 PT samples per year– Meet a known standard of documentation
• If not accredited– Recommend visiting the lab yourself– Recommend asking for a “Raw Data Package” at
least once.
Department of Environmental Quality
Communication Pitfalls
• Not giving requirements to the lab until after the project is over or not at all.– Lab may or may not have met your
requirements (wrong method, QL)– Have to backpedal to try and salvage data– OR resample and retest.
Department of Environmental Quality
Ethics and Data Integrity
• What are your data integrity policies?
• Do you perform data integrity training?
• How would you handle a reported violation?
• Notification of clients/regulators if data is adversely impacted?
Department of Environmental Quality
Self Evaluation Practices
• Audits/Assessments
• Corrective Actions
• Preventive Action
• Management Review
Department of Environmental Quality
Audits/Assessments
• Internal - Self Monitoring• External - Outside Monitoring (DEQ,
ORELAP, etc)• Performance Audits – Check samples (PT
samples)
Department of Environmental Quality
Findings
• Purpose – To get better and NOT to point blame
• What do you do with audit results?
• Document evaluation of cause and correction
• Can incorporate into Corrective Action process.
Department of Environmental Quality
Corrective Actions
• Documented process
• How do you handle and document identified corrective actions– Paper– Spreadsheet– Data base
Department of Environmental Quality
Corrective Action Process
• Identify Corrective Action
• Investigate for Cause
• Identify Corrective Action– Correction vs Corrective Action
• Implement Corrective Action
• Verify Effectiveness
Department of Environmental Quality
Technical Corrective Actions
• QC failures– Reanalysis– Data Qualification
• Non-symptomatic random failures require no further action
• Review of failures over time may indicate need for Corrective actions
Department of Environmental Quality
Reporting Results
• QC failures must be noted – Data user needs to know potential limitations of
the data• Qualifiers are preferred (Required for NELAP)• Narration is acceptable
• Subcontract labs must be identified
• Report Amendments– Identify as “revised” or “supplement” or “…”and
reason for change– Indicate what it is replacing or supplementing.
Department of Environmental Quality
Need Help ?
• OELA – www.oelaonline.com– 2017 presentations available
• Linkedin.com – Discussion groups– “Water and Wastewater Laboratory
Professionals”– “Water and Wastewater Professionals”
• Call fellow plant operator
• DEQ – Permit Compliance Inspector
Department of Environmental Quality
DEQ Website
DEQ Home Page: http://www.oregon.gov/DEQ
DEQ QA Guidance for Self Monitoring Laboratories (NPDES, WPCF)http://www.oregon.gov/deq/FilterDocs/QAguidanceSML.pdf
Department of Environmental Quality
Department of Environmental Quality
If Time Permits
Department of Environmental Quality
Sampling and Sample Handling/Storage
Department of Environmental Quality
– Sampling Technique can contribute up to 75% uncertainty to the final result.
– Incorrect Sampling can cause headaches for the lab. It can even cause sample analysis to be cancelled!
– Sampling is often overlooked in procedures, training, and defining roles and responsibilities.
• Document, Document, Document!– If done correctly, you can use sampling to add
statistical defensibility and to increase the value of your data!
• Sampling Procedures
Why is Sampling Important?
Department of Environmental Quality
•From the Sample Collector’s Perspective•Writing and Understanding the Sampling Procedures
•Scheduling with Laboratory Needs in Mind!•Using the correct Supplies and Equipment•Having the right Preservatives•Preparing for the unknown (extra supplies, extra time, etc.)
•Using Established and Communicated Procedures
•Training and Technique
•Knowing your sample locations – should be in the permit
Sampling Procedures
Department of Environmental Quality
• Effluent sampling should be representative of final discharge after all processing (e.g. after chlorination)
• Composites can be used for most tests EXCEPT (these need to be Grab) (40 CFR 122.21)
• Microbiological samples• pH, Temperature• DO, Cl2• Oil and Grease• Cr6+, Sulfite• Phenolics• VOCs
• Special handling if compositing (not automatic sampler) for VOC, phenolics, sulfide.
Grab Vs Composites
Department of Environmental Quality
• Need to keep cold during compositing period
• Frequency usually flow based unless constant flow then it may be time based. (should be specified in permit)
• May be conducted manually or automatically
• The sample holding time starts at the end of the composite cycle.
24 Hour Composites
Department of Environmental Quality
Loss of Analyte
• How we sample, preserve and store the sample can result in loss of analyte.– Volatile Organics
• Compositing• Acid Preservation (e.g. 2 CVE, Acrolein )
– Aromatic Compounds• Microbiological activity
– Semi-Volatile Organics• Breakdown from heat or sunlight
– Cr+6, NO2• Reduction or Oxidation
Department of Environmental Quality
– Physicals• Solids, pH, Conductivity, Alkalinity, Turbidity, BOD• 1000 mL Poly (Temperature Preserved)
– Temperature Preservation is < 6o C unless “SAME DAY”» Same day still needs to show an attempt at cooling. The laboratory will
document whether it is “on ice”» Blue Ice is no longer recommended as an effective shipping coolant. » Bagging the Ice will keep any cooler leaks from causing problems
– More volume might be required if doing several of the above analyses.
• Physicals Concerns– Homogeneity
» What is an indicator of a non-homogenous sample?» Visual» Analytical (Method Duplicates, Field Duplicates)
– Shorter Holding Times
Sampling Supplies, Preservatives, and Volumes
Department of Environmental Quality
– Nutrients• Total Phosphorous, Total Kjeldahl Nitrogen, Ammonia,
Nitrate/Nitrite– Typically Preserved with H2SO4 and Temperature < 6o C– Unpreserved = short holding time (48 hours)– 500 mL
• Orthophosphate, Anions (Sulfide, Chloride, Bromate, Nitrite)– Filtered in the field (0.45 µm filter)– Filtration Apparatus might be checked with a Field/Equipment
Blank
Sampling Supplies, Preservatives, and Volumes
Department of Environmental Quality
– Metals• Metals must be collected in 250 or 500 mL Poly bottles• Waters preserved with Nitric Acid (HNO3) to pH < 2 (no
thermal preservation is required!) (except Hg)– Total Recoverable Metals– Dissolved Metals
• Must be filtered in the field
Sampling Supplies, Preservatives, and Volumes
Department of Environmental Quality
• E. Coli, Total Coliform, Fecal, Enterococcus• Grab samples are collected in sterile polyethylene or borosilicate
glass bottles at discharge after disinfection. – Note: If at all possible, use the sample bottle directly to take the sample. Do not
use a scoop or other transfer device as they are susceptible to contamination. – Short holding-time (Results must be qualified if not met)– Thermal Preservation required (< 10oC)– Treated Water? Dechlorination by Sodium Thiosulfate (Na2S2O3)
• Note: Sodium sulfite (Na2SO3) that is used for dechlorination for the BOD5 method is toxic to gram-negative microbes such as coliform and therefore MUST NOT be used for dechlorination of coliform samples
• Approved Method???• Manufacturer’s CLAIM some equipment/methods are approved by EPA but
you still must check! – It might be approved only if you do an ATP study
Sampling Supplies, Preservatives, and Volumes
Department of Environmental Quality
Total Residual Chlorine
• Grab samples are to collected at discharge from chlorine contact chamber. Samples must be analyzed immediately.
• Approved Methods– Standard Methods (see CFR for year of issuance)– Iodometric Method I (SM 4500-Cl B) or II (SM 4500-CL C) – Amperometric Titration (SM 4500-CL D) – Amperometric Low level Titration (SM 4500-CL E) – DPD Ferrous Titration (SM4500-CL-F) – DPD Colorimetric (SM4500-CL-G) – **Historically, EPA and DEQ have approved DPD kits in
special cases.
Department of Environmental Quality
– Semi-Volatiles• Amber Glass containers (light sensitive)
• Temperature Preserved (< 6o C)
• Volume is an important consideration– Does the lab need to run Matrix Quality Control on your
project? Would you like them to?– Usually one set of 3 Amber jars for a site per event
Sampling Supplies, Preservatives, and Volumes
Department of Environmental Quality
• Volatile Organic Compounds (VOCs)– These are collected in 40 mL vials (pre-preserved with HCl) in
sets of 3. – Transport Blanks are required! – Cannot be analyzed if there is ANY headspace in the container.
Sampling Supplies, Preservatives, and Volumes
Department of Environmental Quality
– Technique– Contamination – Preservative Quality and Expiration– Field Instrument Calibration– Cleaning between Sites
• Keeping your Sampling Gear Uncontaminated– GPS– Camera– Gloves– Safety Equipment
• Vehicle AND Chemical Safety
Other Sampling Considerations
Department of Environmental Quality
Contamination
• As we drive detection levels down and seek out more analytes, the potential for influence from contamination increases.
• Examples– Mercury– Ethanol– DEET– PAHs
Department of Environmental Quality
Contamination
• Possible sources– Sampling materials and equipment
• Bailers, autosamplers, etc. • Gloves
– Sample bottles• Certified clean (but not for ultra trace levels)• Some plastics have PAH contamination
– Reagents (Preservatives)• Buffers can affect other analytes
– Clothes, lotions, etc.
Department of Environmental Quality
Quality Control
• Types of Quality Control– Laboratory
• This is performed with every analysis as part of the method.– Field
• This pertains to the specific project and can add defensibility to your data.
• Will your project benefit from additional QC?– Split Samples
• Independent Verification of Accuracy.• You could do the same analysis over and over again with
passing QC but still be inaccurate
Department of Environmental Quality
• Field Duplicates– Must be collected within 5 minutes of each
other from the SAME sampling location –define in your QAM
– Common acceptance criteria is 20% RPD or a difference of < LOQ. This can vary by method and matrix. (i.e. Organic solid matrix may be as high as 50% RPD)
• LOQ (Limit of Quantitation) = MRL (Minimum Reporting Limit)
Types of Field Quality Control
Department of Environmental Quality
Needed to ensure positive hits are real• Permit Compliance• Effluent Characterization
• Types– Equipment – Dissolved parameters, – Transfer - Identify possible contamination from field
environment– Transport (Trip) - Volatiles– Common Acceptance Criteria is <LOQ, < ½ LOQ, 1/10
action level (Depending on data use).
Field Quality Control - Blanks
Department of Environmental Quality
• Split Samples are taken as one sample and then “split” to send one to the regular laboratory (or analyzed in your own) and one to a different laboratory. – Which parameters would you like to take split samples for?– Do you have procedures for splitting the sample to ensure
homogeneity?• Compare the two results
– Things to consider:• Both labs using same method?• Both labs have similar Reporting Limits?
Split Sampling
Department of Environmental Quality
• What does “custody” mean?– A sample is considered “in custody” if:
• In someone’s physical possession or view OR• Locked up or kept in a secured location only available to authorized personnel• All changes in sample possession must be fully and completely documented, with the
date, time, and persons relinquishing and receiving the samples on the appropriate chain-on-custody record.
• Chain of Custody is “cradle to grave”– There must be a record of each person who was responsible for the
sample at every part of the sampling and analysis process.
Custody Procedures
Department of Environmental Quality
Sample Receipt
• Preservation check• Holding time check• Compare to COC• Log-in – LIMs, Log book, etc• Unique sample number
• Sample Acceptance Policy?
Department of Environmental Quality
Sample Handling and Storage
• Procedure for sub-sampling if done by lab• Storage
– Refrigeration– Temperature monitoring– How long
• Sample Shipping• Sample Disposal
Department of Environmental Quality
Specific Methods
• BOD SM 5210• TSS SM 2540D• pH EPA 150.1 , SM 4500-H+ B• E. Coli Coliert, Colilert -18 (SM 9223B)• 40 CFR 136.7• SM 1020 – QA/QC - general• SM 2020 – QA/QC – 2000 Series• SM 4020 – QA/QC – Refers back to 1020• SM 9020 – QA/QC – 9000 Series
Department of Environmental Quality
BOD - SM 5210B
• Storage and Pretreatment– Holding Time– Residual chlorine test– Initial pH
• Adjustment– Initial DO
• Supersaturation?– Temperature
• During composite collection• Storage
Department of Environmental Quality
BOD – Equipment
• Reagents and Glassware• Incubator• DO Meter
– Membrane Probe– Luminescence Probe
Department of Environmental Quality
BOD – Sample Seeding
• What is the Seed?– Added to samples that don’t have sufficient
biochemical activity to depletion criteria• Who needs to use Seed?
– Chlorinated– pH Adjusted– Non biological Wastewater– Any with insufficient microbial activity
• Seed Correction Factors
Department of Environmental Quality
BOD - QC
• Unseeded Blank (dilution water) – Depletion < 0.2 mg/L
• Glucose Glutamic Acid (GGA)– Acts as LCS– Every 20 samples or weekly (minimum)– 198 +/- 30.5 mg/L (approx 15%)
Department of Environmental Quality
More BOD - QC
• Minimum residual DO – 1.0 mg/L
• Minimum DO depletion– 2.0 mg/L
• Example:– Initial DO: 8.2 mg/L– Final DO: 7.1– Depletion <2.0 mg/L
• Use a different dilution that meets criteria
Department of Environmental Quality
BOD – Other Things to Remember
• Sample bottles must be water-sealed before incubation
• Dilution water should be distilled or deionized
• At least 2 dilutions for each sample• Temperature monitor the incubator during
incubation period
• If criteria not met – Qualify, Qualify, Qualify
Department of Environmental Quality
E. Coli SM 9223B
• Collection– Homogenize before “pouring off”
• Holding Time to set-up– < 8 hours from sample collection
• Bottles Lot Tested? Documented?– Fluorescence– Positive and Negative Controls
• Media Lot Tested? Documented?– Expiration, Traceability– Positive and Negative Controls
Department of Environmental Quality
E. Coli SM 9223B
• Equipment– Incubator checks– Thermometer calibration
• Records– Times Documented (In and Out)
• Quantification– Test Sealer with dye
Department of Environmental Quality
E. Coli SM 9223B
• Look for fluorescence with a 6-watt, 365-nm UV light within 5 inches of the sample in a dark environment.– UV bulbs fade over time– SM 9020 (3) Lab Equipment and Instrument.
• Clean with cloth/ethanol - monthly• UV light meter – test quarterly > 70% intensity• Test sterilization
Department of Environmental Quality
E. Coli SM 9223B
• Comparators– These expire just like media
• Yellow colonies that do not fluoresce are positive for Total coliform.
• Duplicate Count? Duplicate Analyst?
Department of Environmental Quality
Total Suspended Solids (TSS) – SM 2540D
• Sample Storage and Pretreatment– Storage Temperature < 6oC– Holding Time – 7 days
Samples MUST be shaken or stirred prior to pouring for analysis.
Department of Environmental Quality
TSS - Equipment
• Glass fiber filters – Grades listed in method• Analytical Balance – weigh to 0.1 mg (4
place)• Desiccator that has good desiccant (use
color indicator dessicant)• Drying Oven – maintain 103o-105oC
– Record temp when placing into and removing from oven
Department of Environmental Quality
TSS - QC
• Reagent Blank – each batch• Duplicate – per batch or 10% frequency• Laboratory Control Sample
– Old DMRQA or other PTs– Purchase– Diatomaceous earth
• SM 1020 – QA/QC - general• SM 2020 – QA/QC
Department of Environmental Quality
TSS – General topics
• Pre-wash filters* or purchase pre-washed and weighed
• Dried and re-weighed to constant weight– If not, lab can establish drying time (and
document how it was determined)– Sample filters must be dried for at least an hour.
• Residue should be 1-200mg. – use more or less sample volume to achieve
Department of Environmental Quality
pH – SM 4500 H+ B, EPA 150.1
• Sample Storage and Pretreatment– Samples collected in plastic or glass containers
– Analysis within 15 minutes of collection!
Department of Environmental Quality
pH Equipment
• pH meter calibrated or verified each day• Probe stored in proper solution• Measurement to include temperature
correction?• pH meter accurate to at least 0.1 pH unit• Non-sealed electrodes refilled prior to use.
Department of Environmental Quality
pH QC and General items
• Uncover filling hole before use• Calibrate per manufacture
– Using 4,7, and 10 buffers– Record the calibration slope
• Stir/swirl sample while taking reading• Replicates 10% frequency and readings with
0.2 pH units.• Record date and time of analysis