Plates form and function
-
Upload
gaurav-deshwar -
Category
Health & Medicine
-
view
1.413 -
download
5
description
Transcript of Plates form and function
PLATES: FORMS & FUNCTIONS Dr. Gaurav Deshwar Junior resident III Department of
orthopaedics Sarojini Naidu
Medical College, Agra
History Plates for fixation of long bone fractures were
first recorded by Hansmann, of Heidelberg University, Germany in 1886.One of his original plate sets is seen below .Already then the instruments where listed. the white label says: Attention Do not lose anything.
History Hansmann’s plates were: Bent at the end to protude through the
skin Attched to bone by screw with long
shanks that projected outside the soft tissues.
History Since 1958, AO has devised a
family of plates for long bone fractures, starting with a round holed plate.
In 1969 the Dynamic Compression Plate was developed.
In 1994 LC DCP was created. In 2011 LCP with combination
holes has come into use.
Principle of AO
1. Anatomical Reduction. 2. Stable internal fixation. 3. Preservation of Blood supply 4. Early active pain free mobilisation
Biomechanical Aspect of AO Technique
Neutralization Plate or Protection Plate Compression Plating Lag screw Tension Band Principle Intra Medullary Nailing External Fixation.
Plates : Introduction Bone plates are like internal splints
holding together the fractured ends of a bone.
A bone plate has two mechanical functions. It transmits forces from one end of a bone to the other, bypassing and thus protecting the area of fractures. It also holds the fracture ends together while maintaining the proper alignment of the fragments throughout the healing process.
Plate : Form and Function
To understand how changes in the design of plates has evolved to meet the needs of the patient
To understand how you can use a plate in several different ways to achieve different types of fixation
Plate : Form
DCP 4.5 narrow and broad types DCP 3.5 LC-DCP 3.5 and 4.5 LCP Reconstruction plate 3.5 and 4.5 mm Semitubular - 1/3rd Tubular Plate PC-FIX (Point contact fixator) LISS
Orthopaedic alloy
DCP - 3.5 and 4.5
First introduced in 1969 by Danis
Revolutionary concept of compression
plating
Featured a new hole designed for axial
compression
Broad 4.5 for Femur & Narrow 4.5 for
Humerus & Tibia
DCP 3.5 for Forearm, Fibula, Pelvis &
Clavicle
Problems with DCP
Unstable fixation leads to fatigue &
failure
Strict adherence to principles of
compression
Compromised blood supply due to
intimate contact with underlying cortex
“Refractures” after plate removal
LC-DCP Represents a design change Overcome problems with DCP Plate footprint reduced Minimized kinking at screw
holes, more countourable, reduced plate fatigue at
hole Allows more inclination of screw
in longitudinal plane and transverse
plane.
Tubular plates
3.5 system - 1/3rd Tubular 4.5 system - Semitubular Limited stability Collared hole
Lateral malleolus Distal ulna / Olecranon Distal humerus
Reconstruction plates
Deep notches between holes
Accurate contouring in any plane
Pelvis Acetabulum Distal humerus Clavicle Olecranon
LCP – Locking Compression Plate
LCP
Latest in the evolution
“ Internal fixator ” Combination of
locking screw with conventional screw
Extraperiosteal location of plate
LCP
Combines advantages of DCP principle and locking head principle.
Flexibility of choice within a single implant. Screw hole have been specially designed to accept either:
cortical screw and locking screw
Traditional plating techniques produced stability by:
Compression the plate to the bone surface
Engaging both cortices.thereby producing a rectangular hoop with two bicortical screws.
The locking screws, by achieving angular stability within the plate holes are able to produce a similar hoop with just two unicortical screws.
Shown – LCP used as internal fixator to bridge multifragmentary diaphyseal fracture zone.
In this example : Locking compression plate
is used. Standard cortical and
cancellous screws are used as a traditional plate.
LISS System
Preshaped plates with self drilling self tapping screws with threaded heads.
Through a small incision (using this jig ) plate is slid along the bone surface. position of plate and wire are checked radiologically before insertion of metaphyseal screw .
LISS-Less Invasive Stabilization System
LISS
Plate: Function
Each time a plate is used the surgeon
determines how a plate will function. Plates
can be used in four different ways: Neutralization/protection Compression Buttress Tension Band
Neutralization Plate
A neutralization plate acts as a ""bridge". It transmits various forces from one end of the bone to the other, bypassing the area of the fracture. Its main function is to act as a mechanical link between the healthy segments of bone above and below the fracture. Such a plate does not produce any compression at the fracture site.
A plate used with combination with lag screw is also a neutralisation plate lag screw produce compression and stability. neutralisation plate merely protects the lag screw, allowing mobilization of the extremity. Lag screw generates forces of 3000 N.
Term protection plate expresses the true function.
it is crucial to use a plate that is long enough so that at least three bicortical screw can be inserted in to each main fragment.
The most common clinical application of the neutralization plate is to protect the screw fixation of a short oblique fracture, a butterfly fragment or a mildly comminuted fracture of a long bone, or for the fixation of a segmental bone defect in combination with bone grafting.
Compression plate A compression plate produces a
locking force across a fracture site to which it is applied. The effect occurs according to Newton's Third Law (action and reaction are equal opposite). The plate is attached to a bone fragment. It is then pulled across the fracture site by a device, producing tension in the plate. As a reaction to this tension, compression is produced at the fracture site across which the plate is fixed with the screws. The direction of the compression force is parallel to the plate
Role of compression
Reduction of the space between the bone fragments to decrease the gap to be bridged by the new bone.
Compaction of the fracture to force together the interdigitating spicules of bone and increase the stability of the construct.
Protection of blood supply through enhanced fracture stability.
Friction, which at the fracture surfaces resists the tendency of the fragments to slide under torsion or shear. This is advantageous as plates are not particularly effective in resisting torsion.
Static compression between two fragments maintained over several weak and does not enhance bone resorption and necrosis.
Interfragmentary compression leads to absolute stability but has no direct influence on bone biology or fracture healing.
METHODS OF ACHIEVING COMPRESSION
With tension devise By overbanding With dynamic compression principle
(DCP/LC-DCP) By contouring plate Additional lag screw thro plate
Compression with external devise
it is recommended for fractures of the femur or humeral shaft, when the gap to be closed exceeds 1–2 mm, as well as for the compression of osteotomies and nonunions.
After fixation of the plate to one main fragment, the fracture is reduced and held in position with a reduction forceps. The tension device is now connected to the plate and fixed to the bone by a short cortex screw.and then after comression another fragment is fixed to plate.
Application of the articulated tension device
In oblique fractures the tension device must be applied in such a way that the loose fragment locks in the axilla if compression is produced.
This figure demonstrates the tension device applied in the wrong position
Compression with overbanding
If a straight plate is tensioned on a straight bone, a transverse fracture gap will open up due to the eccentric forces acting on the opposite side.
If the plate is slightly prebent prior to the application (a), the gap in the opposite cortex will disappear as compression is built up (b), so that finally the whole fracture is firmly closed and compressed (c).
Compression plating
Compression through plate
- DCP / LC-DCP
Dynamic compression principle:a The holes of the plate are shaped like an inclined and transverse cylinder.b–c Like a ball, the screw head slides down the inclined cylinder.d–e Due to the shape of the plate hole, the plate is being moved horizontally when the screw is driven home.f The horizontal movement of the head, as it impacts against the angled side of the hole, results in movement of the plate and the fracture fragment already attached to the plate by the first screw (1). This leads to compression of the fracture.
After insertion of one compressing screw, it is only possible to insert one further screw with compressing function in the same fragment. Movement of the plate pushes the first compression screw against the side of the screw hole and prevents further movement. When the second screw is tightened, the first has to be loosened to allow the plate to slide on the bone, after which it is retightened.
Screw holes allow 1mm compression
Additional compression with 1 more eccentric screw before locking first screw
The oval shape of the holes allows 25° inclination of the screws in the longitudinal plane, and up to 7° inclination in the transversal plane
Contouring Plates
Straight plates often need to be contoured prior to application to fit the anatomy of the bone. This is best done with hand-held bending pliers, the bending press, or bending irons. Special flexible templates are available that can be modeled to the bone surface. Repeated bending back and forth should be avoided, as this weakens the plate.
Plate contouring steps Twisting the plate-The anteromedial surface of the
tibial shaft twists internally approximately 20° as it approaches the medial malleolus.The first step of plate contouring is to twist the plate so it matches the tibial surface upon which it will lie. If the plate is bent before it is twisted, the process of twisting will alter the bend that has been created.
Matching the curvature Depending upon the plate location, more or less
bending of the plate will be required to match the contour of the intact (or reduced) bone. Much of the medial tibial shaft is quite straight, so that little bending is required. However, the distal medial surface has a significant concavity, with a typical radius of curvature of 20 cm as illustrated.Such a 20 cm radius can be drawn on a sterile drape and used as a template for plates to be used in this location.
Bending the plate The plate can be bent with bending irons
alone, but it is preferable to bend with a bending press, because the press gives more control.In either case, the bending is done in small steps to produce a smooth contour. Contouring only takes place over the distal 10-12 cm of the plate. When finished, the plate should match the 20 cm radius of curvature.
Buttress Plate
A buttress is a construction that resists axial load by applying force at 90° to the axis of potential deformity
Used in metaphyseal/epiphyseal shear or split fractures
For application of a buttress plate, the first screw must be eccentric to prevent sliding of the plate.
Tension band principle
Frederic Pauwels observed that a curved, tubular structure under axial load always has a compression side as well as a tension side. Under vertical pressure the curved femur creates a tension force laterally and a compression force medially A plate positioned on the side of tensile forces neutralizes them at the fracture site provided there is cortical contact opposite to this plate. In case of a cortical defect, the plate will undergo bending stresses and eventually fail due to fatigue.
Dynamic and static tension band
a) tension band principle on a fracture of the patella. Upon knee-flexion the distraction forces are converted to compression.
In the olecranon fracture the figure-of-eight wire loop acts as a tension band upon flexion of the elbow
tension band principle at the proximal humerus with an avulsion of the greater tubercle
Tension band principle to the medial malleolus example of static tension band
Tension band principal
The following prerequisites are essential: a) Bone or a fracture pattern that is able
to withstand compression.b) An intact cortical buttress on the opposite side of the tension band element.c) Solid fixation that withstands tensile forces.
Antiglide Concepts
• In this model black plate is secured by three whitescrews distal to the blue fracture line.
• The fracture is oriented such that displacement from axial loading requires the proximal portion to move to the left.
• The plate acts as a buttress against theproximal portion, prevents it from “sliding”and in effect prevents displacement froman axial load.
• If this concept is applied to an intraarticular fracture component it is usually referred to as abuttress plate, and when applied to a diaphyseal fracture it is usually referred to as an antiglideplate.
Bone-implant composite
Interdependence of bone and implant in contributing to stability
Intact femur- support axial load of 850 kg
Transverse fracture of mid shaft, with plate on lateral, tension cortex – withstand upto about 800 kg.
Similar fixation with gap in medial cortex will fail under a load of about 60 kg
Plated gap in shaft- buckle under a 20 kg load.
Fixation with an empty screw hole directly leval with a single plane fracture resulted in early fatigue failure due to
Movement Stress concentration at the weak point. If a single plane fracture is spilnted
with a plate even with axial interfragmentary compression,the fracture will open at the cortex oppsite to plate due to elasticity of the plate.
This instability is avoided by-lag screw and prestressd by overbending and by incorporate bone graft medially.
Relative stability : Biological osteosynthesis
In multiplaner fracture complex,use of technique to achieve absolute rigid stability can jeopardize the fragment biology and failure of healing and fixation. Application of plate over top of wedge can damage the vascularity.
To avoid these complication of absolute stability has developed the concept of relative stability.
In such a fraacture if the comminuted zone is bridged in a manner that the main diaphyseal fragments are:
Aligned Correctly matched for rotation Out of length Undisturbed intermediate
fragments heal rapidly by formation of external callus in response to interfragmentary motion.
Addition of bone graft will ensure rapid bone healing.
AO Organisation
☻ Philosophies and techniques of treatment will change with time
☻ The philosophy that we exist to improve the care given to our patients will last forever
THANK YOU