Particle Image Velocimetry (PIV) Calibration Presentation
description
Transcript of Particle Image Velocimetry (PIV) Calibration Presentation
Particle Image Velocimetry (PIV) Calibration PresentationA step-by-step guide to calibrating a custom
PIV system using PIVlab 1.2.
ContentsEquipmentConfigurationPhysical ProcedurePIVlab 1.2 Analysis ProcedureResultsSources of ErrorConclusionsFuture Directions
EquipmentHarvard Syringe Pump Model 33Two, plastic 60ml syringes with inside
diameter (ID)of of 26mm2m of clear rubber hose with an ID of 4.5mmT-valveClear, glass or plastic tubing with of ID of
6.25mm and outside diameter (OD) of 7.7mmBuret ClampSiliconePIV system
Configuration
Syringe Pump Procedure1. Accurately mark the glass tube at 32.6mm
(calculated with the ID using πr²h) above the opening. This is 1ml in the glass tube.
2. Silicone all tubing joints, as not to allow any air to enter or exit the tube system.
3. After silicone is cured, fill the PIV tank with water to desired level.
4. Insert the glass tube into the water, slightly under the water’s surface.
5. Using a level, plumb the glass tube to make it vertical.
Syringe Pump Procedure cont.6. Program the diameter of the syringes and
set pump to Continuous run7. Program the syringe pump to 50ml/hr.,
which calculates to 0.026m/s (which is the maximum pumping rate for these syringes)
8. Turn on PIV system9. Press “Run/Stop” on syringe pump10. Record vortex11. After experimentation, film a known
distance for a calibration image
PIVlab 1.2 Analysis Procedure: load images with sequencing style 1-2, 2-3, 3-4,…
Select region of interest (ROI) just below the glass pipe opening
Image pre-processing, enable contrast-limited adaptive histogram equalization (CLACHE) and highpass filtering (highpass) CLACHE enhances contrast in the imageHighpass sharpens the image and removes
background signals
PIV settings: set interrogation area ([px] in both dimensions) to 24 and Step ([px] spacing between the centers of interrogation area) in both dimensions to 12
Analysis: analyze all frames
Post processing: set standard deviation (stdev) filter threshhold [n*stdev] to 8. Apply to all frames
Calibration: load calibration image and select reference distance along with time step
Plot: set display parameter to velocity magnitude, smooth data to 20% and apply to all frames
A few velocity calculation options…-0- frame (represented as -0-) is the moment
the water is completely expelled from the tube into the aquarium
Velocity is calculated by programming the flow rate (i.e. .833ml/s)*distance fluid traveled (i.e. 0.0326m) = 0.026m/s² . Record the maximum velocity/frame
1. A running average: take 20 slides after -0- and average the velocity calculations (smoothing data)
2. Only look at 1 frame and average the same frame for each run
Extractions: polyline analysis A polyline is a line segment that calculates each
individual vector’s velocity. This is represented in a graph with the Y-axis as velocity and the X-axis as the distance on the line. The polyline can be drawn through the regions of highest velocity denoted by the color bar.
Set parameter to Velocity magnitude and Type to Polyline
Draw the line across the region with the highest velocity determined by the color-bar
Plot dataRecord the velocity for that frameRepeat this process to take a running average
Extractions: polyline analysis
Volume of a torus is 2π²(D/2)r².
Mass is calculated using the density of H2O at room temp.
Smooth the data to 100%.
Torus volume
Torus picture courtesy of: http://blog.teachersource.com/2012/01/
Measure distance and angle: measure radius (D/2) from core to core
To insure accurate measurement of radius r, take 6 measurements from the same side of the torus half from the core to the outside wall and average them: radius 1
Radius 2…
Radius 3… repeat 3 more times
Results
Results cont.
Run
1ml High Velocity (0.026m/s)
1ml Low Velocity (0.016m/s)
1ml High Velocity Torus Volume (ml)
1ml Low Velocity Torus Volume (ml)
High Velocity Acceleration (m/s²)
High Velocity Force (N)
High Velocity Work (J)
Low Velocity Acceleration (m/s²)
Low Velocity Force (N)
Low Velocity Work (J)
1 0.027 0.015 1.029 0.978 5.21 0.00534 3.20E-05 3.25 0.00316 1.90E-05
2 0.026 0.021 0.986 1.015 5.22 0.00513 3.07E-05 3.24 0.00327 1.96E-05
3 0.020 0.023 0.995 1.021 5.24 0.00520 3.12E-05 3.25 0.00330 1.98E-05
4 0.027 0.019 1.028 1.023 5.22 0.00535 3.21E-05 3.26 0.00332 1.99E-05
5 0.025 0.017 1.040 0.998 5.23 0.00542 3.25E-05 3.25 0.00323 1.94E-05
Ave. 0.025 0.019 1.016 1.001
SE 0.0011 0.0013 0.0105 0.0085
t -0.434 2.549 1.519 0.861
p <0.05 <0.05 <0.05 <0.05
Ave. Accel. (m/s²) 5.22 3.25Ave. Force (N) 0.00529 0.00326Ave.Work (J) 3.17E-05 1.95-05
Possible Sources of ErrorThe Harvard Apparatus Twin Syringe Pump
Model 33 has an accuracy within ±0.35% and reproducibility within ±0.1%
The clear rubber hose may be expanding and contracting due to thermal fluctuations
Convection currents with the experimental system
ConclusionsAll measurements using this 2D PIV system
are not statistically different from the expected values suggested by the syringe pump apparatus
Although the Reynolds number (Re) for the syringe pump experiment is 162 and 100 for 0.026m/s and 0.016m/s, respectively, and the Re estimate for aquatic leaping in Rana catesbeiana is 16100, this method can be used to calculate force, work, and power of amphibian propulsion in an aquatic medium
Future DirectionsTo use a static glass tube from the syringes to
keep the water pressure constantTo use larger syringes with a faster syringe
pump will allow vortices to be produced at higher Re ranges, possibly within the frog aquatic leap velocities
To produce a wider range of flow velocitiesConduct experiment in a more temperature
stable area
Acknowledgements Undergraduate Students Duane Barbano Erik Dillingham Antonia Tallante Nicholas Gengler Russell Nelson Maxwell Wheeler
Graduate Students Krysta Powers, M.S. student Sang Hoon Yeo, Ph.D. student Kari Taylor, M.S. student
Post-docs Cinnamon Pace, Ph. D. Jenna Monroy, Ph.D.
FacultyKiisa Nishikawa, Physiology, NAUStan Lindtsted, Physiology, NAUAlice Gibb, Physiology, NAUTed Uyeno, Biomechanics, VSUBrent Nelson, Mechanical Engineering, NAUDavid Lee, Biomechanics, UNLV
FundingNAU VPRNAU GSGSigma Xi GIAR Grant G20111015158992
PersonalMr. Pippiens