OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS
description
Transcript of OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS
![Page 1: OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS](https://reader036.fdocuments.us/reader036/viewer/2022070406/568141f6550346895dadd381/html5/thumbnails/1.jpg)
OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS
V. Jazbinsek, J. Luznik, S. Mieke* and Z. Trontelj
Institute of Mathematics, Physics and Mechanics, University of Ljubljana, Ljubljana, Slovenia
*Physikalisch-Technische Bundesanstalt, Berlin, Germany
![Page 2: OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS](https://reader036.fdocuments.us/reader036/viewer/2022070406/568141f6550346895dadd381/html5/thumbnails/2.jpg)
Introduction
Non-invasive blood pressure (NIBP) measuring devices are based on recording the arterial pressure pulsations in an inflated cuff wrapped around a limb during the cuff pressure deflation.
The recorded NIBP data contains the pressure pulses in the cuff, called oscillometric pulses, superimposed on the cuff pressure deflation.
![Page 3: OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS](https://reader036.fdocuments.us/reader036/viewer/2022070406/568141f6550346895dadd381/html5/thumbnails/3.jpg)
Objective to separate the deflation and
external artifacts from the pressure pulses and to obtain the typical shapes of pressures changes in the cuff due to those artifacts
![Page 4: OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS](https://reader036.fdocuments.us/reader036/viewer/2022070406/568141f6550346895dadd381/html5/thumbnails/4.jpg)
Measurements
EU-project “Simulator for NIBP”
Compressor for the cuff inflation and a sensor for pressure detection, built in a PC (LODE, NL)
25 healthy volunteers took part in this research
![Page 5: OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS](https://reader036.fdocuments.us/reader036/viewer/2022070406/568141f6550346895dadd381/html5/thumbnails/5.jpg)
Measurement Measured data
Arterial pressure pulses, called oscillometric pulses, superimposed on the cuff pressure deflation.
Filtered pulsesDeflation can be removed from data by the digital high pass filtering. Subtraction of negative envelope (curve defined by blue dots) leads to
Oscillometric pulsesThe waveform describing the pulse amplitudes (red dots) as a function of the cuff pressure is a base for automated determination of the systolic and diastolic pressures.
![Page 6: OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS](https://reader036.fdocuments.us/reader036/viewer/2022070406/568141f6550346895dadd381/html5/thumbnails/6.jpg)
• Segmentation of data (red) into pulses (blue vert. lines)
• The deflation signal is calculated by the interpolation of data between subsequent segment borders (green)
• Subtraction of deflation from the measured data to obtain only pulses with positive deflections (oscillometric pulses - black)
Data presentation
![Page 7: OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS](https://reader036.fdocuments.us/reader036/viewer/2022070406/568141f6550346895dadd381/html5/thumbnails/7.jpg)
Mesurements: tapping on the cuff with a plastic tube every 2 s
Segmentation of artefact was obtained from filtered data (band pass
2.5-250 HZ) by selecting 0.7 s time interval around the artefact peak.
Extraction of tapping artifact
![Page 8: OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS](https://reader036.fdocuments.us/reader036/viewer/2022070406/568141f6550346895dadd381/html5/thumbnails/8.jpg)
Typical shapes of high frequency artifacts
Averaging of tapping artifacts
![Page 9: OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS](https://reader036.fdocuments.us/reader036/viewer/2022070406/568141f6550346895dadd381/html5/thumbnails/9.jpg)
Projection into normalized heart beat view – reference waveform
![Page 10: OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS](https://reader036.fdocuments.us/reader036/viewer/2022070406/568141f6550346895dadd381/html5/thumbnails/10.jpg)
Subtraction of reference waveform
a)3rd recording
b)Subtracted reference using waveform optimization (WO)
c) Subtracted reference using single beat optimization (SBO)
![Page 11: OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS](https://reader036.fdocuments.us/reader036/viewer/2022070406/568141f6550346895dadd381/html5/thumbnails/11.jpg)
Extraction of fist closing artifactMesurements: fist closing every 5 s
![Page 12: OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS](https://reader036.fdocuments.us/reader036/viewer/2022070406/568141f6550346895dadd381/html5/thumbnails/12.jpg)
Typical shapes of low frequency artifacts
![Page 13: OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS](https://reader036.fdocuments.us/reader036/viewer/2022070406/568141f6550346895dadd381/html5/thumbnails/13.jpg)
Conclusion I
High frequency artifacts (tapping, vibrations) can be exctracted by applying digital fitering
![Page 14: OSCILLOMETRIC NONINVASIVE BLOOD PRESSURE MEASUREMENTS: THE INFLUENCE OF EXTERNAL ARTIFACTS](https://reader036.fdocuments.us/reader036/viewer/2022070406/568141f6550346895dadd381/html5/thumbnails/14.jpg)
Conclusion II
Low frequency artifacts (moving) extraction require:
• Transformation into heartbeat view to obtain the reference signal.
• Reference can be applied as a template to extract a normal oscillometric waveform used for automated determination of systolic and diastolic pressures.
• Subtraction of normalized oscillometric waveform from contaminated data.
• Averaging of extracted artifacts to obtain typical shape.