Mechanism of Ionization in Matrix-Assisted Laser Desorption Ionization (MALDI):
On the coupling of ion-exchange chromatography to surface-activated chemical ionization in the...
-
Upload
simone-cristoni -
Category
Documents
-
view
215 -
download
2
Transcript of On the coupling of ion-exchange chromatography to surface-activated chemical ionization in the...
RAPID COMMUNICATIONS IN MASS SPECTROMETRY
Rapid Commun. Mass Spectrom. 2008; 22: 2134–2138
ublished online in Wiley InterScience (www.in
PRCM
Letter to the Editor
Scheme 1. (a) M3G, (b) M6G, and (c) most abundant fragment ion structures. The
To the Editor-in-Chief
Sir,
On the coupling of ion-exchange
chromatography to surface-activated
chemical ionization in the analysis of
highly polar metabolites in diluted
urine samples
Ion-exchange chromatography (IEC) is
a widely employed technique used in
the analysis of highly polar com-
pounds.1 This technique allows the
analysis of those compounds that
usually are not/poorly retained on
reversed-phase (RP) chromatographic
columns. Basically, the analytes are
retained by means of a binding inter-
action between the analyte and the
stationary phase and are eluted by
using a solution containing ions that
compete with the analyte in the
column binding or by pH gradient.
This chromatographic approach is
rarely coupled to mass spectrometry
atmospheric pressure ionization (API)
sources due to the occurrence of
in-source discharge phenomena typi-
cally observed in the presence of high
concentrations of buffer and salts
needed for the high voltages of API
sources (electrospray ionization (ESI;
4000–6000 V) and atmospheric pres-
sure chemical ionization (APCI;
1000–6000 V)). Moreover, the presence
of salts and buffer can lead to matrix
ionization suppression effects. In other
words the analytes are not ionized.
These in-source discharge effects lead
to analyte degradation with conse-
quent severe loss of sensitivity.2 Thus,
RP chromatography is used to avoid
these problems.3–5 However, con-
ditions required to retain highly
polar/ionic compounds in RP mode
(i.e. high percentage of water in the
mobile phase) account for poor ioniz-
ation efficiency in ESI mode and, even
worse, for relevant matrix effects
(signal suppression/background en-
hancement) due to coeluting matrix
interferents. For these reasons, a pre-
purification step (e.g. solid-phase
extraction (SPE)), is usually necessary
in order to avoid the matrix effect.4,5
In previous investigations, a surface-
activated chemical ionization (SACI)6,7
source was employed to analyze com-
pounds in matrix sample extracts
without the need for pre-purification.8
The ionization mechanism of SACI is
based on the interaction of the eluent
neutral species with a bipolar moment
(e.g. H2O or neutral salts CH3COONa)
with a metallic surface. A low potential
terscience.wiley.com) DOI: 10.1002/rcm.3590
m/z values of the precursor and fragment io
difference between the surface and the
inlet to the mass spectrometer (50–
700 V) is used so as to create a reactive
environment that makes possible
the production of the analyte ions.9
The low potential difference leads to
better focusing of the analyte ions
inside the mass spectrometer and
allows the process to work in the
presence of salts and buffers avoiding
in-source discharge phenomena and
the consequent degradation in spec-
trum quality.6,7,9 Moreover, working at
high flow rates the cloud of neutral
reactive species surrounding the sur-
face strongly increases leading to high
ns are also reported.
Copyright # 2008 John Wiley & Sons, Ltd.
Figure 1. (a) APCI, (b) ESI, and (c) SACI direct infusion full scan spectra of
morphine-3-glucoronide. The solution concentration was 50 ng/mL. The direct infu-
sion flow rate was 10mL/min.
Letter to the Editor 2135
ionization efficiency. This allows the
direct analysis of diluted biological
samples (e.g. urine, hair extracts)
by liquid chromatography/mass spec-
trometry (LC/MS) and tandem
mass spectrometry (LC/MS/MS)
approaches.8,10–13 Looking at the
results, this technology seems to be
in principle highly effective when used
in conjunction with ion-exchange
chromatography (IEC) due to its ability
to work in the presence of salts and
buffers. It must be emphasized that
this chromatographic approach per-
mits the avoidance of the use of organic
solvent (methanol, acetonitrile) for the
analyte elution, with consequent
reductions in costs and lower health
risks for operators.
In this application, two highly
hydrophilic morphine metabolites
(morphine-3-glucoronide and morphine-
6-glucoronide), weakly retained using
the classical RP columns, have been
analyzed in urine samples by
using cation-exchange chromatog-
raphy (CEC) coupled to ESI, APCI
and SACI sources to verify and com-
pare the compatibility of these
ionization approaches with the IEC
techniques. Moreover, different chro-
matographic conditions were investi-
gated to reduce the matrix effect and
consequently to analyze directly urine
samples, avoiding pre-purification
steps.
Standard morphine-3-glucuronide
(M3G; Scheme 1(a)), morphine-6-
glucuronide (M6G; Scheme 1(b)), for-
mic acid and ammonium formate were
purchased from Sigma Aldrich (Milan,
Italy).
The direct infusion MS and MS/MS
spectra were achieved by analyzing a
50 ng/mL aqueous analyte solution.
The direct infusion sample flow was
10mL/min
Negative urine samples were spiked
with a known amount of standard
compounds (50 ng/mL for each ana-
lyte) and used to test the stability of the
developed LC-CEC/MS/MS method.
Positive urine samples were provided
from drug-addicted subjects. These
solutions were diluted (1:5) with dis-
tilled water and directly analyzed by
means of the LC-CEC/ESI- and
SACI-MS/MS approaches. A volume
of 10mL of the diluted analyte solution
was injected for each analysis.
Copyright # 2008 John Wiley & Sons, Ltd.
An Ultimate 3000 apparatus (Dio-
nex, Sunnyvale, CA, USA) was used
for liquid chromatography (LC).
The chromatographic column was a
cation-exchange BioBasic SCX
(50� 2.1 mm, 5mm, 300 A). The HPLC
gradient was performed using two
eluents at a flow rate of 250mL/min.
The eluents were: (A) H2O and
(B) H2Oþ 0.5% formic acid þ100 mmol/L ammonium acetate.
Solution B (5%) was maintained for
5 min, then a linear gradient was used
passing from 5 to 100% of B in 3 min.
Solution B was maintained for 10 min
at 100%, and then re- equilibrated back
to the starting conditions after 2 min.
Thus, chromatographic analysis was
Rapi
performed in 20 min, but the mass
chromatogram acquisition time was
set to 27 min to allow the chromato-
graphic column to re-equilibrate. A
divert valve was used to direct the
eluent flow to waste for the first 7 min
of the analysis; then it was switched to
direct the eluent flow to the mass
spectrometer for another 10 min and
after which the valve was switched
again to the waste position.
The ESI, APCI and SACI sources
were employed to obtain direct infu-
sion spectra and LC/MS/MS chroma-
tograms.
MS inlet capillary voltage was 4000 V.
The entrance capillary temperature was
2508C. The flow of nebulizing gas
d Commun. Mass Spectrom. 2008; 22: 2134–2138
DOI: 10.1002/rcm
Figure 2. (a) APCI, (b) ESI, and (c) SACI direct infusion full scan spectra of
morphine-3-glucoronide obtained after background noise subtraction. The solution
concentration was 50 ng/mL. The direct infusion flow rate was 10mL/min.
2136 Letter to the Editor
(nitrogen) was 9.0 L/min. The drying
gas flow rate was 6.0 L/min.
APCI vaporizer temperature was
4008C and the entrance capillary
temperature was 1508C. The APCI
needle current was 4000 nA. The flow
rate of nebulizing gas (nitrogen) was
9.0 L/min. The drying gas flow rate
was 5.0 L/min. The MS inlet capillary
voltage was set to 4000 V.
SACI vaporizer temperature was
4008C and the MS capillary tempera-
ture was 2508C. The surface voltage
was 50 V. The surface temperature, as
measured by an optical pyrometer,
was 1108C. The flow rate of nebulizing
gas (nitrogen) was 9.0 L/min. The
drying gas flow rate was 2.0 L/min.
The MS inlet capillary was set to 650 V.
Themassspectrawereobtainedusing
an HCTultra ion trap mass analyzer
(Bruker Daltonics, Bremen, Germany).
The maximum number of ions per scan
was 100 000, with an average of three
microscans per spectrum. The ion
charge control (ICC) was on.
LC/ESI and SACI mass chromato-
grams were obtained using the
tandem mass spectrometry (MS/MS)
approach. The isolation width of the
ions was three mass units. The collision
energy was 100% of its maximum
value (2 V peak-to-peak). The mass
spectra were acquired in positive ion
mode.
The signal/noise (S/N) ratio was
calculated using the root mean square
(RMS) algorithm. An extensive descrip-
tion of this S/N calculation approach
has been reported elsewhere.14 Hystar
software (Bruker Daltonics) wasused to
acquire the data and DataAnalysis
software (Bruker Daltonics) was used
to elaborate them.
Preliminary results were performed by
direct infusion of two 50 ng/mL of M3G
(Scheme 1(a)) and M6G (Scheme 1(b))
aqueous solutions. APCI, ESI and SACI
direct infusion spectra of M3G are
shown in Figs. 1(a), 1(b) and 1(c). It
must be emphasized that the direct
infusion full scan and MS/MS spectra
of M6G exhibit the same behaviour
observed for M3G but the ionization
efficiency is about 10 times higher. In
the case of APCI (Fig. 1(a)) only an
abundant ion at m/z 286 was detected
at this concentration level due to in-
source compound fragmentation
(Scheme 1(c)). Thus, the compound is
Copyright # 2008 John Wiley & Sons, Ltd.
fully fragmented in the ionization
source, probably due to the highly
destructive effect of the corona dis-
charge, with consequent loss of speci-
ficity. In the case of ESI (Fig. 1(b)) an
abundant ion ([MþH]þ) at m/z 462
(counts/s 6.0� 105; S/N ratio of the
[MþH]Rion: 18) was detected. In the
case of SACI (Fig. 1(c)), a high ioniz-
ation efficiency was achieved (counts/
s 8� 105; S/N ratio of the [MþH]Rion:
33) even if the fragment ion at m/z 286,
probably due to in-source thermal
decomposition or to the well-known
surface-induced dissociation (SID)
effect,16 was observed. To better com-
pare the delta in ionization efficiency of
the different sources noise subtraction
Rapi
was applied to all samples. Figures 2
(a)– 2(c) show the APCI, ESI and SACI
spectra, respectively, obtained after
noise subtraction. As can be seen, even
in this case the higher signal intensity
was achieved using SACI (Fig. 2(c);
7.5� 104) with respect to that achieved
using ESI (Fig. 2(b); 5.5� 104). How-
ever, even without background noise
subtraction, as usual a lower chemical
noise was observed in the mass spec-
trum obtained under SACI conditions.
Thus, ESI and SACI were chosen,
due to their best performance, in
order to develop a LC-CEC/MS/MS
approach to analyze these polar com-
pounds. The ESI and SACI cation
exchange – ion extraction – tandem
d Commun. Mass Spectrom. 2008; 22: 2134–2138
DOI: 10.1002/rcm
Figure 3. Cation exchange – ion extraction – tandem mass chromatograms
obtained of morphine-3-glucoronide and morphine-6-glucoronide by using (a) ESI
(water standard solution), (b) SACI (water standard solution), and (c) SACI (addicted
subject urine sample). A volume of 10mL was injected for each analysis.
Letter to the Editor 2137
mass chromatograms of M3G and
M6G obtained by injecting 10mL of a
50 ng/mL standard solution, corre-
sponding to 500 pg injected, in water
are shown in Figs. 3(a) and 3(b). The
collisionally generated MS/MS frag-
ment ion at m/z 286 was monitored
(Scheme 1(c)). As can be seen, the two
compounds are detected only under
SACI conditions (retention times
8.42 min for M3G and 10.11 min for
M6G). This phenomenon is probably
due to the ion suppression effect
obtained under cation-exchange
elution conditions. In fact, the two
compounds elute with 100% of mobile
phase B (H2Oþ 0.5% formic acidþ100 mmol/L ammonium formate). It
must be emphasized that, by using ESI
and working with 100% B buffer
conditions, a high in-source discharge
Copyright # 2008 John Wiley & Sons, Ltd.
current is achieved (40000 nA; instru-
ment current limit) while in the case of
SACI a definitely lower value (24 nA) is
achieved. This behaviour suggests that
the degradation of analytes is achieved
under high discharge conditions, as
observed in the APCI case. On the
other hand it is possible that a higher
matrix ionization suppression effect
takes place under ESI conditions.
Moreover, the chromatographic peaks
of the two isomers are well resolved
under CEC conditions. Thus, they can
be separately quantified even if they
exhibit the same fragmentation beha-
viour.
Thus, SACI was chosen, due to its
sensitivity and its ability to operate in
the presence of high amounts of salts
and buffers,6,7,9 to analyze the selected
morphine metabolites, in diluted urine
Rapi
samples, without purifying the sample
by SPE, usually employed when work-
ing under RP conditions.
Figure 3(c) shows the LC-CEC/
SACI-MS/MS ion extraction mass
chromatogram obtained by analyzing
the real positive urine sample. The
urine was diluted 1:5 using doubly
distilled water before the analysis. A
volume of 10mL of the diluted mixture
was injected. As can be seen under
SACI conditions, the two compounds
were clearly detected.
Finally, in order to evaluate matrix
effects,8,15 20 different negative urine
solutions were then spiked with each
analyte at 50 ng/mL, and analyzed as
previously described. The % chroma-
tographic peak area variation obtained
by comparing urine and water sample
analyte areas was calculated using
Eqn. (1):
AV ¼ ½ðAu � AwÞ=Aw� � 100 (1)
where AV is the % area variation of the
compounds (M3G or M6G), Au and Aw
are the compound chromatographic
peak areas obtained analyzing urine
and water samples, respectively. The
results clearly show that the chromato-
graphic peaks % area variation among
the selected 20 samples in urine with
respect to that achieved by analyzing
water samples was between �0.5 and
�8%. Thus, a low ion suppression effect
seems to be present. However, in order
to further investigate this phenomenon
a post-column analyte (M3G and M6G)
infusion method was employed.17 Even
in this case the matrix effect observed
was negligible.
In conclusion, SACI technology was
shown to be fully compatible with
the cation-exchange chromatographic
approach in the analysis of M3G and
M6G while APCI and ESI leads to
higher in-source sample fragmentation
due to higher in-source discharge or to
an ionization matrix suppression
effect. These results have been
achieved mainly thanks to the low
potential (50 V) at which SACI oper-
ates, strongly reducing the in-source
discharge phenomenon (ion current
25 nA) that is present when working
under APCI and ESI conditions (about
40000 nA) in the presence of cation
exchange elution phase buffers. On the
other hand, ESI does not lead to
in-source fragmentation effects under
d Commun. Mass Spectrom. 2008; 22: 2134–2138
DOI: 10.1002/rcm
2138 Letter to the Editor
pure water solvent conditions but
exhibits a lower ionization efficiency
with respect to SACI. In fact, it
provides a lower signal intensity and
higher chemical noise, reflected in a
lower S/N ratio. Moreover, the SACI
approach allows the use of inexpensive
and non-toxic organic solvents. There-
fore, this technique allows a new way
to obtain the separation and analysis of
highly polar metabolite in urine
samples.
Future developments will be
focused on the application of the
LC-CEC/SACI-MS/MS technique in
the screening of other addictive drug
metabolites in urine samples. In fact
this technique could be used to
advantage for other highly polar
metabolites because of the high poten-
tial of the LC ion exchange technique
coupled to SACI mainly in terms of
chromatographic resolution, matrix
effect reduction and sensitivity.
AcknowledgementsThe authors thank Bruker Daltonics for sup-port. This work was supported by theINGENIO project (Finlombarda spa). Theauthors thank Ms. Karim Jacqueline AmayaMendoza for her essential support.
Copyright # 2008 John Wiley & Sons, Ltd.
Simone Cristoni1*, Sara Crotti1, LorenzoZingaro1, Luigi Rossi Bernardi2, RossellaGottardo3, Lucia Politi4, Aldo Polettini3,
and Franco Tagliaro31
ISB – Ion Source & Biotechnologies,Milan, Italy
2Multimedica Laboratories, Milan,Italy
3Universita degli Studi di Verona,Facolta di Medicina e Chirurgia, Dip.
to di Medicina e Sanita Pubblica–Sez.di Medicina Legale, Verona, Italy
4Universita degli Studi di Pavia, Dip.todi Medicina e Sanita Pubblica, Pavia,
Italy
*Correspondence to: S. Cristoni, ISB – IonSource & Biotechnologies, Via Fantoli 16/15, 20138 Milan, Italy.E-mail: [email protected]/grant sponsor: INGENIO project(Finlombarda spa).
REFERENCES
1. Fritz JS. J. Chromatogr. A 2005; 1085: 8.2. Hao C, March RE. J. Mass Spectrom.
2001; 36: 509.3. Siegel MM. Curr. Topics Med. Chem.
2002; 2: 13.4. Murphy CM, Huestis MA. J. Mass
Spectrom. 2005; 40: 1412.5. Murphy CM, Huestis MA. J. Mass
Spectrom. 2005; 40: 70.6. Cristoni S, Rubini S, Bernardi LR.
Mass Spectrom. Rev. 2007; 26: 645.7. Cristoni S, Bernardi LR, Biunno I,
Tubaro M, Guidugli F. Rapid Com-mun. Mass Spectrom. 2003; 17: 1973.
Rapi
8. Cristoni S, Bernardi LR, Gerthoux P,Mocarelli P, Brambilla M, Gonella E,Guidugli F, Cantu M. Rapid Commun.Mass Spectrom. 2006; 20: 2376.
9. Cristoni S, Bernardi LR, Guidugli F,Tubaro M, Traldi P. J. Mass Spectrom.2005; 40: 1550.
10. Cristoni S, Basso E, Gerthoux P,Mocarelli P, Gonella E, BrambillaM, Crotti S, Bernardi LR. Rapid Com-mun. Mass Spectrom. 2007; 21: 2515.
11. Cristoni S, Cantu M, Bernardi LR,Gerthoux P, Mocarelli P, BrambillaM, Gonella E, Guidugli F. J. MassSpectrom. 2005; 40: 1609.
12. Cristoni S, Cantu M, Bernardi LR,Gerthoux P, Gonella E, BrambillaM, Cavalca V, Zingaro L, GuidugliF. Rapid Commun. Mass Spectrom.2005; 19: 1231.
13. Cristoni S, Bernardi LR, Gerthoux P,Gonella E, Mocarelli P. Rapid Com-mun. Mass Spectrom. 2004; 18: 1847.
14. Wentzell PD, Brown CD. Signal pro-cessing in analytical chemistry. InEncyclopedia of Analytical Chemistry.John Wiley: Chichester, 2000; 9764.
15. Mallet CR, Lu Z, Mazzeo JR. RapidCommun. Mass Spectrom. 2004; 18: 49.
16. Jo SC, Cooks RG. Eur. J. Mass Spec-trom. 2003; 9: 237.
17. De Nardi C, Bonelli F. Rapid Commun.Mass Spectrom. 2006; 20: 2709.
Received 24 December 2007Revised 24 April 2008
Accepted 25 April 2008
d Commun. Mass Spectrom. 2008; 22: 2134–2138
DOI: 10.1002/rcm