OIL TEST SET - Motwane TEST SET.pdf · Oil Test vessel with electrode gap adjustment facility....
Transcript of OIL TEST SET - Motwane TEST SET.pdf · Oil Test vessel with electrode gap adjustment facility....
OIL TEST SET
www.motwane.com
Analog Oil Test Set
Manual cum motorized operation
Safety door & Zero start interlock
Memory effect analog voltmeter
Oil Test vessel with electrode gap adjustment facility.
Digital Oil Test Set
Motorized operation with 2KV/s ramp rate
Safety door & Zero start interlock
Digital display with Hold facility
Voltage Calibrator for onsite calibration
Fully Automatic Oil Test Set
Fully automatic microprocessor based kit
Preprogrammed test sequence as per IEC 60156 / IS6792/ASTM
Customized preprogrammed test sequences available
Large LED display with touchpad for alphanumeric data entry
Up to 100 test results storage facility
USB interface & Software facility
Inbuilt Thermal printer for test report
Voltage Calibrator for onsite calibration
The Motwane Manufacturing Pvt.Ltd. offer a wide range of Oil Breakdown voltage test kits to check integrityof insulating oil used in Transformers, Capacitors, XLPE cables and Switchgear for safe operation and maintenance. OTS series comes in three distinct models - OTS100A, a preprogrammed fully automatic kit ; Motorized ( Digital ) OTS100DS kit and OTS100 an analog motorized cum manually operated kit. Designed with unmatched quality and improved reliability and safety having earth & zero start, safety door interlock ensuring high degree of protection.
Technical Specifications
The Motwane Manufacturing Company Pvt. Ltd.Gyan Baug, Motwane Road, Nashik, Maharashtra, India.Tel. No.: +91-253-2463752 / 53 Fax.:+91-253-2463197 Toll Free:1800 - 233 - 7766E-mail: [email protected] visit : www.motwane.com
Designed and Manufactured by:
Notes1.The Instrument is accompanied with Test & calibration sheet. 2.Test Facilities can be provided at the factory with the available test set-ups only. 3. The Motwane has a policy of continuous product improvement. We therefore reserves the right to change design and specification without notice. 4.Stated accuracies are valid from 10% of the range to 95% of the range. 5. Accuracy Specified for temperature range of 25°C ± 5°C & 55% RH ±10%.
Model
Type
Mains Supply
Voltage Accuracy
Voltage resolution
Voltage rise rate
Display
Temperature measurement
Test Standards
Data storage
PC Interface
Protection
safety standard
MOT-OTS60/80/100 MOT - OTS60DS/80DS/100DS MOT -OTS60A/80A/100A
Output Voltage ( in KV )
software for report
Printer
Tripping time
Environmental condition
Dimension ( L X W X H - in mm )
Weight (in Kgs )
Calibrator
Standard Accessories
Manual & Motorized ( Analog )
230V,50Hz ±10%
60KV/80KV/100KV
±1.5%
2KV
2 KV/S
<10ms
Analog
X
X
IS 6792
X
X
X
X
Safety Door & Zero start interlock & Overload protection
X
-10 °C to +55 °C,<90% RH, Non condensing
540X520X435
50
X
1. Test cell with electrodes2. Go-No-Go Gauge3. User manual4. Calibration Certificate
Motorized ( Digital )
230V,50Hz ±10%
60KV/80KV/100KV
±2%
0.1KV
2 KV/S
<10ms
LED
X
X
Yes
Ambient Temperature & Humidity
Fully Automatic
230V,50Hz ±10%
60KV/80KV/100KV
±1%
0.1KV
0.5 - 5KV/S
<10ms
LCD
IEC60156/IS6792
X
X
X
X
IEC60156/IS6792/ASTM
Yes
100 Test Results
USB
Yes
Safety Door & Zero start interlock & Overload protection
Safety Door & Zero start interlock & Overload protection
IEC601010-1,EMC 61326-1X
-10 °C to +55 °C,<90% RH, Non condensing
-10 °C to +55 °C,<90% RH, Non condensing
465x510x445 465x510x445
53 48
Yes - Optional Yes - Optional
1. Test cell with electrodes2. Stirrer3. Go-No-Go Gauge4. User manual5. Calibration Certificate
1. Test cell with electrodes2. Stirrer3. Go-No-Go Gauge4. User manual5. Calibration Certificate
Customize setting (Stand by ,Stirring time, Intermediate stand time ,Trip current, Sequence repetition & Voltage rise rate)