INDUCTION INITIAL PHASE ONCE THE TARGET MAC VALUE OF 0.8 TO 1 HAS BEEN REACHED CARRIER GASES * From a patient age of > 6 months; All values are clinically-tested orientation values, however in individual cases, they must be adapted according to the individual response by the patient, the performance characteristics of the machines used and the operational requirements. Schematic Procedure* of Minimal-Flow Anaesthesia with Carrier Gases PDF-Nr: PDF-8088 Isoflurane Sevoflurane Desflurane AIR O2 N2O AIR O2 N2O Oxygen: 1 L/min for 1-8 min., 100% Oxygen: 1 L/min, 40% Air: 3 L/min Fresh Gas Flow: 4 L/min Fresh Gas Flow: 4.4 L/min Oxygen: 1.4 L/min, 32% Nitrous Oxide: 3 L/min Oxygen: 100% to 0.25-0.35 L/min NOTE: Change of anesthetic agent concentration Oxygen: 0.3 L/min, 68% Air: 0.2 L/min Fresh Gas Flow: 0.5 L/min Fresh Gas Flow: 0.5 L/min Oxygen: 0.3 L/min, 60% Nitrous Oxide: 0.2 L/min 60-80% 35-40% 30-40% O 2 FRESH GAS FLOW SETTINGS INSPIRATORY OXYGEN CONCENTRATION STABILIZATION (VOL%): VAPORIZER SETTINGS (VOL%): REDUCE FRESH GAS FLOW: INCREASE VAPORIZER SETTINGS (VOL%): Pre-oxygenation with 100% oxygen at 6 L/min for 1-3 minutes with a face mask Intravenous administration of the hypnotic or inhalation induction Analgesia and relaxation (NOTE: Oxygen & Oxygen/Air Mixtures may require 20% higher dosing of induction opioid) Endotracheal intubation or insertion of a larynx mask Connection of patient to the circuit system OXYGEN OXYGEN/ AIR MIXTURE OXYGEN/ NITROUS OXIDE O 2 O 2 N N O DES SEVO ISO 5-6% 12% 5-6% DES SEVO ISO 2.5% 6% 3.5% DES SEVO ISO 1-1.5% 4-6% 2-2.5% DES SEVO ISO N/A N/A N/A DES SEVO ISO 5% 8% 5% DES SEVO ISO 2.5% 5-7.5% 3-3.5%

Transcript of O O O - DraegerN/AN N/A /A ISO SEVO DES 5% 5% 8% ISO SEVO DES 2.5% 3-3.5% 5-7.5%. Created Date:...

Page 1: O O O - DraegerN/AN N/A /A ISO SEVO DES 5% 5% 8% ISO SEVO DES 2.5% 3-3.5% 5-7.5%. Created Date: 6/6/2017 4:43:12 PM ...

INDUCTION

INITIAL PHASE

ONCE THE TARGET MAC VALUE OF 0.8 TO 1 HAS BEEN REACHED

CARRIER GASES

* From a patient age of > 6 months; All values are clinically-tested orientation values, however in individual cases, they must be adapted according to the individual response by the patient, the performance characteristics of the machines used and the operational requirements.

Schematic Procedure* of Minimal-Flow Anaesthesia with Carrier Gases

PD

F-N

r: PD

F-80

88

Isoflurane

Sevoflurane

Desflurane

AIRO2 N2O

AIRO2 N2O

Oxygen:

1L/min for 1-8min.,100%

Oxygen:

1L/min,40%

Air:

3L/min

FreshGas Flow:

4L/min

FreshGas Flow:

4.4L/min

Oxygen:

1.4L/min,32%

NitrousOxide:

3L/min

Oxygen:

100% to 0.25-0.35L/minNOTE: Change of

anesthetic agent concentration

Oxygen:

0.3L/min,68%

Air:

0.2L/min

FreshGas Flow:

0.5L/min

FreshGas Flow:

0.5L/min

Oxygen:

0.3L/min,60%

NitrousOxide:

0.2L/min

60-80% 35-40% 30-40%

O2

FRESH GAS FLOW SETTINGS

INSPIRATORY OXYGEN CONCENTRATION STABILIZATION (VOL%):

VAPORIZER SETTINGS (VOL%):

REDUCE FRESH GAS FLOW:

INCREASE VAPORIZER SETTINGS (VOL%):

Pre-oxygenation with 100% oxygen at 6 L/min for 1-3 minutes with a

face mask

Intravenous administration of the hypnotic or inhalation

induction

Analgesia and relaxation

(NOTE: Oxygen & Oxygen/Air Mixtures may

require 20% higher dosingof induction opioid)

Endotracheal intubation or insertion of a larynx mask

Connection of patient to the circuit system

OXYGEN OXYGEN/AIR MIXTURE

OXYGEN/NITROUS OXIDEO2 O2

NNO

DESSEVOISO

5-6% 12%5-6%

DESSEVOISO

2.5% 6%3.5%

DESSEVOISO

1-1.5% 4-6%2-2.5%

DESSEVOISO

N/A N/AN/A

DESSEVOISO

5% 8%5%DESSEVOISO

2.5% 5-7.5%3-3.5%