Nitrosamine Detection in Meat
By Beryl Ombaso,
Chemistry 4101,
Fall 2008
1
BACKGROUND
• Nitrite is added to meat both for preservation
and color enhancement.
• Amines in the meat react with the nitrite to
form nitrosamines which are known
carcinogens.1
NO2 + H+ HNO2
2HNO2 N203 + H20
N203 + R2NH R2N.NO + HNO2
(a nitrosamine)
2
Tumorgenic levels of Nitrosamines
• USDA recommended exposure level-10 ppb.2
Nitrosamine TD50, mg/kg body weight/day
Expected levels in typical sample, µg/kg
Nitrosodimethylamine 0.01 1.90 – 2.37
Nitrosodiethylamine 0.11 0.33 – 0.54
Nitrosopyrrolidine 2.10 2.32 – 8.94
Nitrosopiperidine 1.30 0.5 – 1.31
Nitrosodibutylamine 0.69 0.16 – 0.24
3
Methods for Nitrosamine detectionMethod Limit of
DetectionLinear range
Special notes
Spectrophotometry3 0.78ppb 25 to 2000 ppb
Derivatization with Griess reagent.
Electroanalysis4 6.0 x 10-8 mol L-1
2 x 10-6 to 1.35 x 10-5 mol L-1
Determine total nitrosamines
HPLC/Fluorescence5 0.04 ppb 0.04 to 1.05 ppb.
Derivatization with Dansyl chloride.
4
Proposed Method for Nitrosamine Detection
• Gas Chromatography with tandem MS would be most ideal.6
Nitrosamines are highly volatile and thermally stable.
Limit of detection – 0.01 ppb.
Linear range – 0.01 to 10 ppb.
Percent recovery – 70-80 %.
Efficient separation and ease of use.
5
Instrumentation
6
SAMPLE PREPARATION
• Interfering species: amines, amino acids.
• SPE limits amount of interfering species.
• Blank sample consists of 0.1N NaOH.
7
Experimental conditions6
• Combination of Extrelut-Florisil cartridge for SPE.• Column – 30 m, 0.25 mm i.d, 0.25 μm film thickness
containing 1:6 cyanopropylphenyl:methyl polysiloxane.• Carrier gas – Helium gas at 1.0 mL/min.• Oven – initial temp. 500C, final temp. 2600C• Injection volume- 2.0 μl• Outlet pressure – 54.8 kPa• Electrospray ionization at 70eV.• Reagent for chemical ionization- ammonia gas.• Quadrupole mass selective detectors for analysis.
8
Sample Chromatogram
9
Sample Mass Spectrum
• Quantification accomplished by using calibration curve obtained from standards.
• Standards obtained easily from manufacturer.
10
Equipment model
• HP 5890A/5970 GC/MS• Cost: $ 9,950• Oven temp: up to 4500C• Compact size• Power supply: 120V• Resolution: 0.1amu• S/N: 20:1 • Scan speed: 1500 amu/sec• Mass range:10 – 800 amu• Gentechscientific.com
11
References1. Mirvish, S.; Cancer letters, 1995, 93, 17-48.
2. Jay, J.; Modern Food Microbiology, 6th Ed, 2000, pg 261.
3. Perez, E. L.; Rios, A.; Valcarcel, M.; J Anal. Chem, 2001, 371, 891-895.
4. Oliveira, R. T.; Salazar-Banda, G. R.; Machado, S. A.; Avaca, L. A.; Electroanalyis, 2008, 20, 396-401.
5. Cardenes, L.; Ayala, J. H.; Venerando, G.; Afonso, A. M.; J. Liq. Chrom. & Rel. Technol., 2002, 25, 977-984.
6. Jurtchenko, S.; Tenno, T.; Molder, U.; Reinik, M.; Proc. Estonian Acad. Sci. Chem., 2002, 51, 169-184.
12