NIRCam ETU Testing at LMATC, Palo Alto Kailash C. Sahu.
-
Upload
laurence-flowers -
Category
Documents
-
view
213 -
download
0
Transcript of NIRCam ETU Testing at LMATC, Palo Alto Kailash C. Sahu.
NIRCam ETU Testing at LMATC, Palo Alto
Kailash C. Sahu
What’s NIRCam?
• NIRCam is the near-infrared camera (0.6-5 microns) for JWST
– Dichroic used to split range into short (0.6-2.3m) and long (2.4-5) sections
– Nyquist sampling at 2 and 4m– Coronagraphic capability for both short and
long wavelengths– Low-resolution spectroscopic capability in
the LW channel.
• NIRCam is the wavefront sensor– Must be fully redundant
2 Channels Per Module
• Each module has two channels (0.6 to 2.3 m and 2.4 to 5 m)
– 7 wide band filters (4 SW, 3 LW) for deep surveys
– Survey efficiency is increased by observing the same field at long and short wavelength simultaneously
• Pixel scale: SW: 0.032”/pixLW: 0.064”/pix
Long wavelength channelShort wavelength channel
Mo
du
le A
Mo
du
le B
2.2’
Focus and alignment
mechanism
Coronagraphocculting masks
First fold mirror
Collimator lens group
Dichroic beamsplitter
Long wave filter wheel assembly
Long wave camera lens
group
Long wave focal plane housing
Short wave filter wheel assembly
Short wave camera lens group
Short wave fold mirror
Pupil imaging lens assembly
Short wave focal plane housing
Light from OTE
Pick-off Mirror
Module A
NIRCam ETU/WFS Testing
• 1-9 April at LMATC (Lockheed Martin Advanced Techonology Center), Palo Alto
• First ~3 days (after the “Red Chamber” attained operating temperature of ~38K) spent on SCA/Assembly testing, rest on WFS testing (only with the SW channel).
• People from STScI: • Massimo Robberto, • Elizabeth Barker, • Kailash Sahu (WIT).• George Hartig, • Erin Elliot (TEL)
NIRCam ETU/WFS Testing
• Tests were generally very successful– Operated with the flight software over eight days with no crashes or major problems– Wavefront sensing components were demonstrated to work in the NIRCam pupil
Main problems: - Transfer of FITS files often needed manual intervention.- PIL (Pupil Imaging Lens) had some alignment problems.
“First Light”Fuzzy because of chamber vibrations, tests were done with “pulsed light”
Structure in this image is due to the known poor alignment of the OMA and ETU at this early stage.
Weak Lens Images
The shapes are very close to predicted images.Ghosts are due to reflections in the OMA optics, not NIRCam
DHS Spectra
Spectra are at the correct angle.Known emission line in the super continuum source being used for illumination.
A spectrum extracted from the previous image. The “notches” in the filter used in series with the DHS are apparent.
Flat Fields from Internal Red Chamber Lamps
Detector Linearity from Chamber Lamp Data
Sample ramps from independent exposures Black: Without linearity correction
Blue: After linearity correction
These data are the first chance we had to check linearity with ASIC + SCA. Overall linearity is good (better than 5%) and is dominated by SCA, not ASIC. Some systematic effects at “low” count rates need to be investigated further.
(a) Over-saturated point source image taken during ETU testing.Estimated counts (from the ghost): 650,000 to 1 million DN.
(b) Exposure taken immediately after (a). Counts at the peak of the point source: ~650 DN (< 0.1%)(c) Exposure taken ~4 minutes later. Peak counts: ~80 DN.
Detector Latency
(a) (b)(c)
SUMMARY
• Initial SCA and WFS testing results are very encouraging.
• The software operated relatively smoothly, but problems were identified which will be rectified in time for FM testing.
• Participations from all the groups (GSFC, LMATC, UA and STScI) were coherent and complimentary.
• All the ETU Test data are being archived at STScI through SID archive.
Sample Images from the Focus Sweeps
These are not organized in any fashion and some were taken at different wavelengths.
PIL Image of DHS
Blue areas are due to the missing pieces in the DHS. Segment gaps and the secondary supports are evident.