Molecular Spectroscopy

11

Click here to load reader

description

This Presentation is made by Miss. Tahira Khalid, a student of N.E.D University.

Transcript of Molecular Spectroscopy

Page 1: Molecular Spectroscopy

Molecular Spectroscopy

BY,

TAHIRA KHALID

1

Page 2: Molecular Spectroscopy

Spectroscopy

Construction Of Spectrophotometer

Instrumentations Deviations

Absorbance

Absorption Laws

Colorimeter

Construction Of Colorimeter

Other Instruments

2

Page 3: Molecular Spectroscopy

Spectrophotometry involves the use of a spectrophotometer. A spectrophotometer is a photometer that can measure intensity as a function of the light source wavelength. Important features of spectrophotometers are spectral bandwidth and linear range of absorption or reflectance measurement.

A spectrophotometer is commonly used for the measurement of transmittance or reflectance of solutions, transparent or opaque solids, such as polished glass, or gases. However they can also be designed to measure the diffusivity on any of the listed light ranges that usually cover around 200 nm - 2500 nm using different controls and calibrations. Within these ranges of light, calibrations are needed on the machine using standards that vary in type depending on the wavelength of the photometric determination.

3

Page 4: Molecular Spectroscopy

DOUBLE BEAM SPECTROPHOTMETER

SINGLE BEAM SPECTROPHOTOMETER

Page 5: Molecular Spectroscopy

SOURCE DETECTORMONOCHROMATOR CUVETTE

Source must be:

• Continuous

• Stable

• Intense

Cuvette

must be

made up

of silica or

quartz

More the

grooves present

in the

monochromator,

efficient will the

monochromator

5

Page 6: Molecular Spectroscopy

STRAY DEVIATION:

Some external source is entering like sunlight

VOLTAGE FLUXUATION:

Voltage is varying continuously which effects the sensitivity of the spectrophotometer

CHANGE IN SECSITIVITY OF DETECTOR:

With the passage of time sensitivity of the spectrophotometer decreases

POLYCHROMATIC RADIATION:

Light passing through the sample is polychromatic because the monochromator is not working properly

UNEXPECTED CHEMICAL REACTION:

The given sample is dissociated after some time

Concentration of the given sample is changed

6

Page 7: Molecular Spectroscopy

For each wavelength of light passing through the spectrometer, the intensity of the light passing through

the reference cell is measured i.e. Io

The intensity of the light passing through the sample cell is also measured for that wavelength - given the

symbol, I.

If I is less than Io, then obviously the sample has absorbed some of the light. A simple bit of maths is then

done in the computer to convert this into something called the absorbance of the sample - given the

symbol, A.

7

Page 8: Molecular Spectroscopy

LAMBERT’S LAW:

According to this law

“ Absorbance is directly proportional to the breadth of the cuvette

i.e. A α b

BEER’S LAWS:

According to this law

“ Absorbance is directly proportional to the concentration of the sample

i.e. A α C

By combining both laws we get

A α bC

A= ε bC

Where ε is the molar absorptivity

where “l” is the breadth of the cuvette

8

Page 9: Molecular Spectroscopy

A Colorimeter is a device used

in colorimetry. In scientific fields the

word generally refers to the device

that measures the absorbance of particular wavelengths of light by a

specific solution. This device is most

commonly used to determine

the concentration of a

known solute in a given solution by

the application of the Beer-Lambert

law, which states that the

concentration of a solute is

proportional to the absorbance.

Page 10: Molecular Spectroscopy

The essential parts of a colorimeter are:

A light source (often an ordinary low-voltage filament lamp)

An adjustable aperture

A set of colored filters

A cuvette to hold the working solution

A detector (usually a photoresistor) to measure the transmitted light

A meter to display the output from the detector

In addition, there may be:

A voltage regulator ,to protect the instrument from fluctuations in main voltage.

A second light path, cuvette and detector. This enables comparison between the working

solution and a "blank", consisting of pure solvent, to improve accuracy.

10

Page 11: Molecular Spectroscopy

FiltersChangeable optics filters are used in the colorimeter to select the wavelength of light which the solute absorbs the most, in order to maximize accuracy. The usual wavelength range is from 400 to 700 nanometers (nm). If it is necessary to operate in the ultraviolet range (below 400 nm) then some modifications to the colorimeter are needed. In modern colorimeters the filament lamp and filters may be replaced by several light-emitting diodes of different colors.

CuvettesIn a manual colorimeter the cuvettes are inserted and removed by hand. An automated colorimeter (as used in an AutoAnalyzer) is fitted with a flawcell through which solution flows continuously.

Output

The output from a colorimeter may be displayed by an analogue or digital meter and may be shown as transmittance (a linear scale from 0-100%) or as absorbance (a logarithmic scale from zero to infinity). The useful range of the absorbance scale is from 0-2 but it is desirable to keep within the range 0-1 because, above 1, the results become unreliable due to scattering of light.

11