Practical Absorption Spectrometry: Ultraviolet Spectrometry Group
Mass Spectrometry by ANITHA SRI
-
Upload
anitha-sri -
Category
Documents
-
view
222 -
download
0
Transcript of Mass Spectrometry by ANITHA SRI
-
8/3/2019 Mass Spectrometry by ANITHA SRI
1/54
OVERVIEW OFMASS SPECTROMETRY
M. Anitha Sri (Y11MPH448)
I/II M.Pharmacy, Industrial pharmacy
CHALAPATHI INSTITUTE OF
PHARMACEUTICAL SCIENCES.
-
8/3/2019 Mass Spectrometry by ANITHA SRI
2/54
CONTENTS Introduction
Instrumentation
Mass Spectrum
Resolution Determination of molecular formula
Data analysis and interpretation
Applications
2
-
8/3/2019 Mass Spectrometry by ANITHA SRI
3/54
INTRODUCTION A mass spectrometer is an instrument that measures the masses
of individual molecules. Three Basic functions:
1. creating gaseous ion fragments from the samples
2. separating them according to their mass-to-charge ratio
3. records the relative abundance of each ionic species
present
Also known as positive ion spectra or line spectra.
3
-
8/3/2019 Mass Spectrometry by ANITHA SRI
4/54
Block diagram of Components of Mass Spectrometer
-
8/3/2019 Mass Spectrometry by ANITHA SRI
5/54
INSTRUMENTATION
Inlet system
Ion source
Electrostatic accelerating system
Magnetic field Ion separator
Ion collector and Detector
Vacuum system
5
-
8/3/2019 Mass Spectrometry by ANITHA SRI
6/54
INLETSYSTEM
Direct vapor inlet
Direct insertion probe
Gas chromatography(GC-MS)
Liquid chromatography(LC-MS)Particle Beam Interface
Thermospray Interface
Electrospray Interface
Desorption techniques(FAB and LSIMS)
6
-
8/3/2019 Mass Spectrometry by ANITHA SRI
7/54
DIRECT VAPOUR INLET
Gases or volatile liquidsMethod is Molecular leak or Molecular pumping
The sample can be introduced through a septum port or
through a valve port.
7
-
8/3/2019 Mass Spectrometry by ANITHA SRI
8/54
DIRECT INSERTION PROBE
8
Solids and liquid samples.
Autoprobe.
-
8/3/2019 Mass Spectrometry by ANITHA SRI
9/54
GC-MS
Most common technique for
introducing samples.
Several different interface
designs are used to connect
these two instruments.
The MS coupled to the GC
should be capable of high
resolution.
Highly specific.
9
-
8/3/2019 Mass Spectrometry by ANITHA SRI
10/54
10
-
8/3/2019 Mass Spectrometry by ANITHA SRI
11/54
LC-MS
Used for Thermo labile compounds.
Several interfaces are used to connect LC and MS.
11
-
8/3/2019 Mass Spectrometry by ANITHA SRI
12/54
PARTICLE BEAM INTERFACE
Solvent is removed from an
aerosol of LC effluent
The resulting analyte is
analysed in the ion source
Known as MAGIC
(Monodisperse Aerosol
Generator Interface for
Chromatography)
12
-
8/3/2019 Mass Spectrometry by ANITHA SRI
13/54
THERMOSPRAY
Involves simply heating the
tip of the entry tube to
promote vaporisation.
Through the centre of the
stainless steel tube, passes asmall diameter tube which
carries the column eluent.
The tube projects slightly
beyond the end of the heater
cap which is situated in acartridge heater together with
a thermocouple.
13
-
8/3/2019 Mass Spectrometry by ANITHA SRI
14/54
ELECTRO SPRAY
14
Sample is dissolved in
a solvent and pumped
through a narrow
capillary.Voltage is applied to
the capillary tip and
the sample is dispersed
into an aerosol, aidedby a coaxially
introduced nebulising
gas.
-
8/3/2019 Mass Spectrometry by ANITHA SRI
15/54
ESI
15
The charged droplets diminish
in size by solvent evaporation
assisted by a flow of drying
gas.
Eventually charged sample
ions, free from solvent, are
released from the droplets,
which pass through the orificeinto an intermediate vacuum
region and from these through
a small aperture into the
analyser of the MS.
-
8/3/2019 Mass Spectrometry by ANITHA SRI
16/54
IONISATION TECHNIQUES
Name Ionising agent
Electron Impact(EI) Energetic electronsChemical Ionisation (CI) Reagent gaseous ions
Field Ionisation (FI) High potential electrode
Field Desorption (FD) High potential electrode
Electro Spray Ionisation (ESI) High electrical field
Matrix assisted Laser Desorption
Ionisation (MALDI)
Laser beam
Fast Atom Bombardment (FAB) Energetic atomic beamSecondary ion Mass
Spectrometry (SIMS)
Energetic beam of ions
Thermospray Ionisation (TS) High temperature
16
-
8/3/2019 Mass Spectrometry by ANITHA SRI
17/54
ELECTRON IMPACT
17
-
8/3/2019 Mass Spectrometry by ANITHA SRI
18/54
CHEMICAL IONISATION
Chemical interaction between reagent gas ions and analytemolecule.
Two-step process.
CH4 + e- = CH4
+ + 2e-
Secondary ions of reagent gas are produced, which react with
the analyte molecules. The mechanism may be proton transfer, hydride abstraction or
charge transfer.
CH4+ + MH = CH4 + MH
+
CH5+ + MH = CH4 + MH2
+
CH3+ + MH = CH4 + M
+
Reagent gases: Argon, Helium, Nitrogen.18
-
8/3/2019 Mass Spectrometry by ANITHA SRI
19/54
FIELD IONISATION
19
Ions are formed under the
influence of high electric field
produced by applying high
voltages.
On the surface of fine tube,
many hundreds of projecting
carbon microtips are present.
These extract the electron from
the sample and ionise the
sample molecules.
-
8/3/2019 Mass Spectrometry by ANITHA SRI
20/54
FAST ATOM BOMBARDMENT
20
High energy primary
beam is directed at a
target surface to obtain
high yield of secondaryions.
Primary beam may be
ions, electrons, photons
or neutral atoms. SIMS may be
dynamic or static.
-
8/3/2019 Mass Spectrometry by ANITHA SRI
21/54
MALDI
Two step process.
Desorption is triggered
by a laser beam.
The second step isionization.
Nitrogen laser of 337 nm
wavelength is used.
Sinapinic acid is used asmatrix for proteins and -
cyano-4-
hydroxycinnamic acid for
peptides.21
-
8/3/2019 Mass Spectrometry by ANITHA SRI
22/54
CHOOSINGAN IONISATIONTECHNIQUE
Information desired Ionization techniqueDepth profiling Fast atom bombardment/secondary ion
mass spectroscopyChemical speciation/component
analysis (fragmentation desired) Electron impactMolecular species identification of
compounds soluble in common
solventsElectrospray ionization
Molecular species identification of
hydrocarbon compounds Field ionizationMolecular species identification of
high molecular weight compoundsMatrix assisted laser desorption
ionizationMolecular species identification of
halogen containing compoundsChemical ionization (negative mode)
22
-
8/3/2019 Mass Spectrometry by ANITHA SRI
23/54
ELECTROSTATIC ACCELERATINGSYSTEM
23
The positive ions formed in the ionisation chamber
are accelerated by pairs of accelerator plates to
impart velocities to the ions.
Ions are sorted acc. to m/e ratio based on 3
properties: energy, velocity and momentum.
The beam from the slits of these plates consists of
a collimated ribbon of ions having equal energies.
K.E = eV = m1v12 = m2v2
2.
-
8/3/2019 Mass Spectrometry by ANITHA SRI
24/54
MAGNETIC FIELD
eV = mv2
F = HeV
HeV = mv2/r
m/e = H2r2/2V
24
e = charge
m= mass
v = velocity
V = voltage
F = Magnetic force
H = Magnetic field strength
r = radius
-
8/3/2019 Mass Spectrometry by ANITHA SRI
25/54
ION SEPARATOR
Single Focussing
Double Focussing
Cycloidal
Quadrupole TOF
MS/MS
Radio Frequency
25
-
8/3/2019 Mass Spectrometry by ANITHA SRI
26/54
SINGLE FOCUSSING
26
-
8/3/2019 Mass Spectrometry by ANITHA SRI
27/54
SINGLE FOCUSSING
27
-
8/3/2019 Mass Spectrometry by ANITHA SRI
28/54
DOUBLE FOCUSSING
28
-
8/3/2019 Mass Spectrometry by ANITHA SRI
29/54
CYCLOIDAL
29
-
8/3/2019 Mass Spectrometry by ANITHA SRI
30/54
QUADRUPOLE
30
-
8/3/2019 Mass Spectrometry by ANITHA SRI
31/54
TIMEOF FLIGHT
The time-of-flight (TOF) mass analyzer separates ions in time as they travel downa flight/drift tube.
This is a very simple mass spectrometer that uses fixed voltages and does not
require a magnetic field. The greatest drawback is that TOF instruments have
poor mass resolution, usually less than 500.
31
-
8/3/2019 Mass Spectrometry by ANITHA SRI
32/54
MS/MS
32
Hybrid MS.
The two analysers are separated by a field free collision chamber,
which contains an inert gas.
-
8/3/2019 Mass Spectrometry by ANITHA SRI
33/54
RADIO FREQUENCY
33
An assembly of grids is
employed to select ions acc. to
their velocities.
Alternative grids are connected
to a radiofrequency source and
the other grids are connected to
a steady potential.
It is simple in construction and
doesnt require a magnet.
-
8/3/2019 Mass Spectrometry by ANITHA SRI
34/54
IONCOLLECTORAND DETECTOR
Detection of ions is based on their charge
Detectors monitors the ion current, amplifies it and
the signal is transmitted to the data system where it
is recorded in the form of mass spectra.
Types of Detectors:
Faraday Cup Collector.
Electron Multiplier
Channel Electron Multiplier ArrayThe detection is either by pulse counting or analog
measurement.
34
-
8/3/2019 Mass Spectrometry by ANITHA SRI
35/54
FARADAY-CUP
35
Ions enter the cup and transfer
their charge to the cup.
Secondary electrons are
generated.
No. of secondary electrons
generated depends on several
factors:
mass of ionsenergy of ions
charge on the ions
Angle of incidence
material of cup
nature of the ion.
-
8/3/2019 Mass Spectrometry by ANITHA SRI
36/54
ELECTRON MULTIPLIER
36
A metal plate called
conversion dynode that
converts the impinging ionsto electrons is present.
Ion beams strikes the
conversion dynode.
Secondary electrons are
produced by the electron
multiplier.
-
8/3/2019 Mass Spectrometry by ANITHA SRI
37/54
-
8/3/2019 Mass Spectrometry by ANITHA SRI
38/54
VACUUMSYSTEM
All mass spectrometers operate at very low pressure (high vacuum).
This reduces the chance of ions colliding with other molecules in the
mass analyzer. Any collision can cause the ions to react, neutralize,
scatter, or fragment. All these processes will interfere with the mass
spectrum.
To minimize collisions, experiments are conducted under high
vacuum conditions, typically 10-2 to 10-5 Pa (10-4 to 10-7 torr)
depending upon the geometry of the instrument.
This high vacuum requires two pumping stages. The first stage is a
mechanical pump that provides rough vacuum down to 0.1 Pa (10-3
torr). The second stage uses diffusion pumps or turbo molecular
pumps to provide high vacuum.
38
-
8/3/2019 Mass Spectrometry by ANITHA SRI
39/54
GENERAL FRAGMENTATION PATTERNS
Simple Direct cleavage
Retro-Diels Alder Reaction
Hydrogen Transfer Rearrangement
Mc lafferty rearrangement
39
-
8/3/2019 Mass Spectrometry by ANITHA SRI
40/54
MC-LAFFERTYREARRANGEMENT
40
Involves intramolecular migration of -
hydrogen from electron rich center to electron
deficit center followed by cleavage at position
resulting in the formation of neutral alkene.
Common in ketones, esters and carboxylic acids
-
8/3/2019 Mass Spectrometry by ANITHA SRI
41/54
TYPESOFIONS
Molecular ion
Fragment ions
Rearrangement ions
Multiply charged ions
Negative ions
Metastable ions
Pseudomolecular or Quasi ions
41
-
8/3/2019 Mass Spectrometry by ANITHA SRI
42/54
METASTABLEIONS
Ions formed in the analyser after moving away from
the ionisation chamber.
Gives broad bands.
Formed at non-integral mass numbers.
Mass of metastable ion is calculated by:
m* = m22/m1
m* = mass of metastable ionm1 = mass of molecular ion
m2 = mass of daughter ion42
-
8/3/2019 Mass Spectrometry by ANITHA SRI
43/54
DERIVITISATION
For some ionisation techniques, the compound should bederivitised before being analysed.
Derivitisation is the use of chemicals to modify theanalyte, usually to reduce its polarity. Often OH and NHgroups are reacted with silylating reagents, or acetic
anhydride, to form compounds with O-Si, N-Si, O-C orN-C bonds instead.
The derivative then lacks the ability to form hydrogenbonds and is more volatile than the analyte was.
Mass spectrometry is a gas phase technique; irrespective
of the nature of the sample the analysis is on gaseousions, hence the need for volatility.
43
-
8/3/2019 Mass Spectrometry by ANITHA SRI
44/54
MASS SPECTRUM
The mass spectrum is presented in terms of ion abundance vs. m/eratio (mass)
The most abundant ion formed in ionization gives rise to the tallest
peak on the mass spectrumthis is the base peak
44
base peak, m/e
43
All other peak intensities are relative to the base peak as a
-
8/3/2019 Mass Spectrometry by ANITHA SRI
45/54
All other peak intensities are relative to the base peak as a
percentage
If a molecule loses only one electron in the ionization process,
a molecular ion is observed that gives its molecular weightthis is designated as M+ on the spectrum
45
M+,m/e 114
In most cases, when a molecule loses a valence electron, bonds
-
8/3/2019 Mass Spectrometry by ANITHA SRI
46/54
, ,
are broken, or the ion formed quickly fragment to lower energy
ions.
The masses of charged ions are recorded as fragment ions by
the spectrometerneutral fragments are not recorded !
46
fragment ions
-
8/3/2019 Mass Spectrometry by ANITHA SRI
47/54
RESOLUTION
47
Adjacent peaks must be clearly separated.
The valley between the two adjacent peaks should not be more
than 10% of the height of the larger peak.
R = Mn/Mn - Mm
-
8/3/2019 Mass Spectrometry by ANITHA SRI
48/54
DETERMINATIONOF MOLECULAR FORMULA
Nitrogen Rule
Rule Of Thirteen
When a molecular mass, M+, is known, a base formula can be
generated from the following equation:
M/13 = n + r/13
the base formula being: CnHn+r
Index of Hydrogen Deficiency:
HDI = n-r+2 / 2 Ring rule:
For the molecule CwHxNyOz, R = w + 1 + y-x/248
-
8/3/2019 Mass Spectrometry by ANITHA SRI
49/54
DATA ANALYSISFROMMASSSPECTRUM
The molecular ion peak in aromatic compounds is
relatively much intense.
Conjugated olefins show more intense molecular ion
peak as compared to the corresponding non-
conjugated olefins with same no. of unsaturation.
The relative abundance of the saturated hydrocarbon
is more than the corresponding branched chain
compound. In aromatic compounds, the substituent groups like -
OH, -OR, -NH2 increase the relative abundance and
NO2, -CN decrease the relative abundance.49
-
8/3/2019 Mass Spectrometry by ANITHA SRI
50/54
Absence of molecular ion peak in the mass
spectrum means that the compound under
examination is highly branched or tertiary alcohol.
In case of Chloro or Bromo compounds, isotope
peaks(M+ + 2) are also formed along with the
molecular ion peak.
Isotope peak is not observed when Fluorine or
Iodine atom is present in the compound.
50
-
8/3/2019 Mass Spectrometry by ANITHA SRI
51/54
COMPUTERISEDMATCHINGOFSPECTRA
WITHSPECTRALLIBRARIES
The computer can compare a mass spectrum it has
determined with the spectra in the databases of the
libraries.
The output is a table called HIT LIST.
Hit list includes the name of each compound that the
computer has used for matching, its molecular weight,
molecular formula, probability that the spectrum of the
test compound matches the spectrum in the data base.
The probability is determined by the no. of peaks andtheir intensities that can be matched.
51
-
8/3/2019 Mass Spectrometry by ANITHA SRI
52/54
APPLICATIONS
Determination of molecular mass and structure.
Determination of Isotopic abundance.
Distinction between isomers.
Determination of Ionisation potential and BondDissociation energies.
Detection of presence of impurities.
Identification of unknown compound.
52
-
8/3/2019 Mass Spectrometry by ANITHA SRI
53/54
REFERENCES
U.S.P. Y.R. SHARMA. ELEMENTARY ORGANIC SPECTROSCOPY,
PRINCIPLES AND CHEMICAL APPLICATIONS. 4th ed. S.CHAND.
2007.
D.A.SKOOG, F.J. HOLLER, S.R.CROUCH. PRINCIPLES OF
INSTRUMENTAL ANALYSIS. 6th
ed. THOMPSON BROOKS. 2007. D.L.PAVIA, G.M.LAMPMAN, G.S.KRIZ. INTRODUCTION TO
SPECTROSCOPY. 3rd ed. THOMPSON BROOKS. 2001.
G.R.CHATWAL, S.K.ANAND. INSTRUMENTAL METHODS OF
CHEMICAL ANALYSIS. 5th ed. HIMALAYA PUBLISHING HOUSE.
2002.
H.HWILLARD, L.L.MERRITT, J.A.DEAN, F.A.SETTLE.
INSTRUMENTAL METHODS OF ANALYSIS. 7th ed. CBS
PUBLISHERS.
-
8/3/2019 Mass Spectrometry by ANITHA SRI
54/54
THANK YOU