Next-Generation Ultra Fast LC/MS/MS
Fully Compatible with HPLC and UHPLC
Triple quadrupole mass spectrometer plays very important role in trace analysis of complex matrices such as residual pesticide
in foods, contaminants in environment, drug concentration in blood and screening of abused drugs. In recent years, UHPLC
system has been developed realizing high speed and high resolution, however, analysis speed of mass spectrometer has
bottlenecked in combination of UHPLC and triple quadrupole mass spectrometer. Our LCMS-8030 meets your requirement by
combining Ultra Fast MRM Acquisition and Ultra Fast Polarity Switching. LCMS-8030 enjoys maximum benefit of UHPLC
performance and offers analytical speed beyond comparison.
U l t r a F a s t L C p e r f o r m a n c eWith a pressure range up to 130 MPa, high-speed injection, overlapping injection and highly efficient gradient mixing, Nexera UHPLC enables ultra-high speed and ultra-high resolution analysis.
U l t r a F a s t M S S p e e d● Fastest multiple reaction monitoring (MRM) transition
times available today (dwell times of 0.8 msec and pause times of 1 msec)
● High speed multi-analyte detection with 555 MRM transitions in one second
● Unsurpassed polarity switching speed (polarity switching time of 15 msec)
S y s t e m I n t e g r a t i o nBy bringing together the world’s highest performance LC system and the fastest triple quadrupole mass spectrometer available today, Shimadzu delivers a new approach to ultra fast mass spectrometry detection. Robust, simple and Ultra Fast.
GCMS-QP2010 UltraGCMS-QP2010 SE
LCMS-IT-TOF iMScope MALDI-7090GCMS-TQ8030 LCMS-8030 LCMS-2020LCMS-8040 LCMS-8050
Figure1. Standard Chromatogram of 226 Pesticides in Two Minutes
Figure 2
0.25 0.50 0.75 1.00 1.25 1.50 1.75 min0
0.5
1.0
1.5
×105
Inte
nsi
ty
Low speed
Product Ion scanMRM
scan startthreshold
High speed
01.00 1.25 min
2.0
4.0
6.0
8.0
×105
Inte
nsi
ty
×105
0.0100 200 300 m/z
1.0
2.0
3.0
108.9 218.1
261.1
399.3
Inte
nsi
ty
N
N
H
C
OOCH3
OCH3
N
N O O CH2 CH3
H
S
O
O
Compound Name m/z Time (0.885 min – 1.462 min)
Fusion of Ultra Fast MRM Acquisition and Ultra Fast Polarity Switching
U l t r a F a s t M a s s S p e c t r o m e t r y D e t e c t i o n
An Ultra Fast Scan Speed of 15,000 u/sec Results in a Wealth of Information
With high-resolution, high peak capacity
separations, mass spectrometry detection needs
to be Ultra Fast to acquire sufficient data points
across a peak to deliver accurate and reliable
quantitation. Coupled with Nexera, the
LCMS-8030 not only delivers the fastest MRM
acquisition times available today, with dwell
times of 1 msec, it also acquires data with a
polarity switching time of just 15 msec.
Figure 1 illustrates a 2-minute elution of 226
pesticides using the LCMS-8030 with the
Nexera UHPLC system. Ultra Fast polarity
switching and Ultra Fast MRM analysis times
deliver reliable and accurate quantitation.
Figure 2 illustrates a 1.5-minute elution of 29 pesticides using the LCMS-8030 with the Nexera UHPLC
system. Ultra Fast polarity switching and Ultra Fast MRM transitions deliver reliable and accurate
quantitative results. Furthermore, Synchronized Survey Scan technology, utilizing a high-speed scanning
rate of 15,000 u/sec, allows full spectrum scans within a series of MRM measurements, providing
confirmation of target compounds with information-rich product ion spectra.
Introducing the Next Generation of Triple Quadrupole Technology
To provide best-in-class performance for Ultra Fast data acquisition rates in MS/MS detection, we have developed UFsweeper technology. This technology efficiently accelerates ions out of the collision cell, dramatically minimizing cross talk and shortening MRM analysis time to the lowest possible level. Simply put, the LCMS-8030 delivers higher data quality at higher data acquisition speeds.
4
UFsweeper Technology Effectively Accelerates Ions Out of the Collision Cell
555 MRM Transitions in One Second is Now Possible!
patent pending
UFsweeper is a unique technology created by
Shimadzu that delivers unparalleled
efficiency and speed. UFsweeper accelerates
ions out of the collision cell by forming a
pseudo-potential surface. The result is higher
CID efficiency and Ultra Fast ion transport to
reduce the sensitivity losses and cross-talk
that are observed on other systems.
UFsweeperConventional
design
UFsweeper efficiently acceleratesions out of the collision cell without losing momentum.
Ions lose momentum due to collision with gas.
Dwell time
100 msec
10 msec
1 msec
%RSD
0.48
0.79
1.92
×104
00 0.5 1.0 1.5 2.0 2.5 3.0 3.5 min
0.5
1.0
1.5
Alprazolam m/z 309 >281, FWHM = 3 sec, Pause time 1 msec, n = 6
Dwell time1 msec
Dwell time10 msec
Dwell time100 msec
In the UFsweeper collision cell, there is no ion loss
even at 1 msec dwell times as ions are accelerated
from the pressurized collision cell without losing
momentum. In addition, higher RF power
capability exceptionally minimizes the pause time
between each MRM transition. For
the first time, true high-
throughput analysis without
sacrificing ion intensity is a reality.
5
Atmospheric ionization source
Shimadzu engineering results in the rapid transfer of the wide m/z range of ions from the Qarray to the octopole region, enabling Ultra Fast MRMs and product ion scans.
Ion opticsOur unique Qarray ion guide was designed to effectively focus ions over a wide m/z range by overlapping multiple electric fields (Patent pending).
Qarray
The transmission efficiency between octopoles is dramatically improved with a larger aperture size, supported by a triple inlet turbo molecular pump.
Octopole
Desolvation line: Easy maintenance without breaking vacuum
U l t r a F a s t M a s s S p e c t r o m e t r y D e t e c t i o n
Delivering Results – Quickly and Accurately
The ion optics on the LCMS-8030 have been
fully optimized for ion transport to the
detector using a series of RF and DC ion
guides before Q1. This design has a
significant impact on challenging matrices.
For example, in the case of a 10 ppb
pesticides mixture spiked into tomato
extract, 1400 individual 1μl aliquots were
injected over a period of 2 days. The result is
an outstanding 2.19% reproducibility.
00
0.5
1
1.5
2
2.5
3
200 400 600 800 1000 1200
Rel
. I
nte
nsi
ty
Injection
Chromafenozide 10ng/mL in tomato
1400 analyses (47 hours)
Internal standard %RSD 2.19%
Ultra Fast Speed Combined with Femtogram Detection
Drawing on our groundbreaking research and hands-on experience, the LCMS-8030 has been designed to achieve outstanding performance quicker than ever before. Bringing together advanced ion acceleration technology and high-sensitivity detection systems, we have created one of the most advanced triple quadrupole mass spectrometry systems currently available.
6
Detector
UFsweeper collision cell
Utilizing over 30 years experience in quadrupole research, development, and manufacture, Shimadzu delivers high-performance hyperbolic mass filters to its current LC/MS products. Even at the maximum scan rate (15,000 u/sec), excellent sensitivity and resolution are maintained.
Quadrupole rods
01.00 1.25 1.50 1.75 2.00 min
0.5
1.0
1.5
2.0×106
In
ten
sity
Dummy transition
m/z 700.0 >165.0
01.00 1.25 1.50 1.75 2.00 min
0.5
1.0
1.5
2.0×106
Verapamil 1 μg/mL
m/z 455.2 >165.0
Pause Time 3msec
Dwell Time 3msec
In
ten
sity
100
50
01.50 1.75 min
×10,000
Below 0.003%
Area%RSD(n=6)
7.75
3.36
3.11
1.43
0.79
1.29
0.31
0.70
0.58
0.40
Accuracy%
116.35
109.37
95.38
90.55
94.15
97.02
99.98
96.92
99.83
100.48
Conc.ng/mL
0.0058
0.0108
0.0477
0.0907
0.4708
0.9702
4.9995
9.6907
49.9108
100.4682
Std. Conc.ng/mL
0.005
0.01
0.05
0.1
0.5
1
5
10
50
100
Malachite greenm/z 329.2 > 313.15
R2= 0.99990530.005 –100ng/mL
Are
a
0.00.0
1.0
2.0
3.0
4.0
5.0
×106
25.0 50.0Concentration, ng/mL
75.0 100
UFsweeper Technology
Linearity - Dynamic Range with Ultra Fast Mass Spectrometry
UFsweeper technology is a unique approach to
accelerating MRM and ion fragmentation. Even at
high-speed data acquisition rates, cross talk and background
noise are eliminated.
Equipping a true high-speed pulse-counting detector and conversion dynode system has resulted in
astonishing data acquisition rates even with polarity switching. The development of a unique semi-floating
high-voltage power supply realizes ultra-high speed polarity switching (15 msec).
7
Easy Maintenance Identical to the LCMS-2020
22
33
11
DL(Desolvation Line)
Optimizing System Performance
U l t r a F a s t M a s s S p e c t r o m e t r y D e t e c t i o n
The LCMS-8030 heated ion source works with the most challenging samples, delivering robust,
high-sensitivity detection using ESI, APCI or our dual probe ionization interface.
System maintenance for the ion source is simplicity itself. Cleaning the heated desolvation capillary is quick
and maintains system vacuum to provide greater uptime and usability.
Simplicity
The LCMS-8030 achieves best-in-class performance for ultra fast mass spectrometry
detection and opens new opportunities for all application areas.
Using LabSolutions software to optimize the power of the Nexera UHPLC and
LCMS-8030 accelerates results, simply.
8
MRM Synchronization optimizes MRM cycle times for overlapping retention time windows and helps to simplify method
development and routine analysis with multiple MRM experiments. MRM retention time windows can partially or
completely overlap; however, the cycle time for each analyte is constant, ensuring high-speed data sampling across a peak.
MRM Synchronization delivers optimized peak sampling and quantitative data for each retention time window even with
high-resolution chromatography.
MRM Synchronization
Quantitation Browser for Effective Multianalyte Quantitation
Single-vendor Solution Provides Seamless Operation
The combination of Shimadzu’s LCMS-8030 and Nexera UHPLC brings together the latest hardware on a
single platform for the next generation of Ultra Fast technology.
The unified platform provides
unmatched qualitative and quantitative
analysis, increased productivity, and
accelerated workflows for high-
throughput data analysis. Also, all
software operations are handled
seamlessly, reducing PC conflicts and the
need for user intervention.
The Quantitation Browser in
LabSolutions LCMS software
provides intuitive functionality
that contributes to more efficient
data processing. With the
Quantitation Browser, peak
information, quantitative results,
and statistical calculations of a
series of data can be rapidly
viewed in a single window.
9
When performing quantitation, system parameters must first be optimized. Shimadzu offers
various method packages to eliminate the need for MRM determination and chromatographic
method development.
Open Access Sample Logging for Triple Quadrupole LCMS
With sample logging now
made easy, Quan Solution
allows users unfamiliar
with LCMS software the
capability to perform
LC/MS/MS analysis using
pre-set methods. Simple
screens allow users to
access the power of a
triple quadrupole LCMS
without extensive
training.
Residual PesticidesNo. of compounds: 167
Veterinary DrugsNo. of compounds: 42
Water Quality AnalysisNo. of compounds: 44 (Golf Course Pesticides), 32 (Pesticides Targeted for Water Quality Control)
Drugs of AbuseNo. of compounds: 286
Rapid Toxicology Screening SystemNo. of compounds: 106
Primary MetabolitesNo. of compounds: 55
Lipid MediatorsNo. of compounds: 130
Po w e r f u l Wo r k f l o w S o l u t i o n s
Sample registration by selecting the standard samples, unknown samples, and QC samples at the specified positions
Login
Quantitative Reports Received Via E-mail
Sample Vial Selection
Analysis
Quan Solution SoftwareOptional Open Solution Series Software
10
Changing between ESI, APCI and DUIS interface is a completely tool-free operation.
APCI-8030 DUIS-8030
Corona needle
ESI probe
Sample injection
Corona needle
Profiling Solution software analyzes a huge amount of data from chromatograms, extracting every peak
from multiple data files and creating a data table, which is required for multivariate analysis. Profiling
Solution supports both LC/MS and GC/MS instruments and allows multivariate analysis with various
analytical methods.
Profiling Solution Ver. 1.1
Optional Ion Source
Data Processing Software for Profiling
Detected
Peak
An
alyticalin
strum
ent
Mu
ltiple d
ataD
ata table
Mu
ltivariatean
alysis softw
are
Data file
11
LCM
S-8030
Printed in Japan 3655-07305-50ANS
Company names, product/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation or its affiliates, whether or not they are used with trademark symbol “TM” or “®”.Third-party trademarks and trade names may be used in this publication to refer to either the entities or their products/services. Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.
For Research Use Only. Not for use in diagnostic procedures. The contents of this publication are provided to you “as is” without warranty of any kind, and are subject to change without notice. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication.
© Shimadzu Corporation, 2013www.shimadzu.com/an/
iMScope (page 2) may not be sold in your country. Please contact us to check the availability of iMScope in your country.