Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1...

13
1 Lab Manual KEYENCE LASER SENSOR DISPLACMENT Laser Head LK-H150 / Controller LK-G5001 / LK-Navigator 2 Created by Huu-Tu Nguyen Lab of Electromechanical Energy Department of Precision Engineering. National Chung Hsing University October 15, 2013

Transcript of Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1...

Page 1: Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1 This manual will introduce you how to setup to measure vibration at micrometer scale

1

Lab Manual

KEYENCELASER SENSOR DISPLACMENT

Laser Head LK-H150 / Controller LK-G5001 / LK-Navigator 2

Created by

Huu-Tu Nguyen

Lab of Electromechanical EnergyDepartment of Precision Engineering.

National Chung Hsing UniversityOctober 15, 2013

Page 2: Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1 This manual will introduce you how to setup to measure vibration at micrometer scale

2

Contents

1. Objectives ...............................................................................3

2. Instruments .............................................................................3

3. Overview of laser head LK-H150 ...........................................4

4. Overview of LK-Navigator 2 ..................................................5

5. Procedure ................................................................................. 7

Page 3: Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1 This manual will introduce you how to setup to measure vibration at micrometer scale

3

1. Objectives1.1 This manual will introduce you how to setup to measure vibration at micrometer

scale using laser displacement sensor LK-H150.1.2 Understanding the control parameters of monitoring software LK-Navigator 2, so

that you can measure vibration in different unit scales (mm, µm).2. Instruments

2.1 Laser head LK-H150

2.2 Controller LK-G5001

2.3 Monitoring software LK-Navigator 2

Page 4: Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1 This manual will introduce you how to setup to measure vibration at micrometer scale

4

3. Overview of LK-H150.

Dimension in () indicate the reference distance and measurement range for specularreflection mounting

Page 5: Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1 This manual will introduce you how to setup to measure vibration at micrometer scale

5

4. Overview of LK-Navigator 2LK-Navigator 2 is software for configuring parameters and monitoring the LK-G5000

series controller. It is used by connecting a PC and the controller to configure parametersand monitor operating status.

Operation settings4.1 LK-Navigator 2 allows you to upload the settings in the controller to the PC for

viewing and editing. It features copy and paste functions for configuring multiplecontrollers and setting programs in a short amount of time.

4.2 The settings on the PC can be downloaded and reflected in the controller.LK-Navigator 2 communicates with the controller at the simple press of a button.

4.3 Settings can be loaded from and saved on the PC. Settings saved on the PC can beused to restore the settings in the controller in the event the settings are erased.

The following items can be set from the PC.1. Head settings: measurement modes (standard, transparent, translucent,...),

reflection type, range, etc.2. OUT settings: calculation method, tolerance, offset, display unit,...3. Common settings: sampling cycle, data storage, analog ouput,...4. Environment settings: number of used heads, RS-232 settings, ethernet,...

Page 6: Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1 This manual will introduce you how to setup to measure vibration at micrometer scale

6

Monitoring functionsThese two items can be monitored.1. Measurement value.2. Received light waveform.The screen shown below is a display example of the "Measurement value display".

Data storageThe controller has a data storage function for storing measurement values.The PC can be used to control these functions:1. Start/stop/clear the data storage.2. Configure the data storage method.3. Upload/download settings between the controller and PC.4. Upload storage data from the controller.5. Read and save storage data on the PC.6. Display waveform of up to 12 data series downloaded from the controller.7. Display calculations on data between user-set cursors.

Page 7: Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1 This manual will introduce you how to setup to measure vibration at micrometer scale

7

Note: for more information, please read the ‘LK-Navigator 2 User’s manual’ book of Keyence Corp. (file “LK-H2-M-E.pdf”in the installation source).

5. Procedure1. Setup the instruments as the following figure:

Be connecting the laser head with controller.Measurement value is monitored from the computer by LK-Navigator 2 software.

Page 8: Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1 This manual will introduce you how to setup to measure vibration at micrometer scale

8

2. Star LK-Navigator on the computer.

3.Select ‘Read Controller Settings’ (default) and click OK.

4. Adjusted the parameter setting.

Step1: sets the sampling cycle

Page 9: Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1 This manual will introduce you how to setup to measure vibration at micrometer scale

9

Step2: sets the measurement unit.

Step 3: Set the measurement type Averages time: sets the number of averaging measurements if “Type”is set“Moving -average”.

Set the to tolerance comparator.

Page 10: Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1 This manual will introduce you how to setup to measure vibration at micrometer scale

10

Step 4: send settings to controller

5. Check the reading signal by selecting ‘View Received Light Waveform’.

Tip: Adjust the laser beam to the object until the signal is higher than 512.

Page 11: Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1 This manual will introduce you how to setup to measure vibration at micrometer scale

11

6. Go to‘Data Storage’function for displaying measurement results.Step1: click in Data Storage

Step2: click in Data Storage Setting such as in photo below.

Page 12: Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1 This manual will introduce you how to setup to measure vibration at micrometer scale

12

Step3: Amount of Data Store: sets the number of data point to store Store cycle: sets the interval for data storage

Step 4: start storage and read data from controller to PC

Page 13: Keyence disp sensor LK-H150-SOPweb.nchu.edu.tw/~daw/Manual/LK-H150-SOP.pdf · 3 1. Objectives 1.1 This manual will introduce you how to setup to measure vibration at micrometer scale

13

This window can display the following data:1. Storage data read from the controller.2. Storage data saved on the PC.

There are function buttons:Data Storage Setting: sets the method for storing measurement data (amout of datastored, sampling cycle).Storage Data Readout: loads storage data from the controller.Start Storage: starts storing data on the controller.Stop Storage: stops storing data on the controller.Clear Stored Data: clears the data stored on the controller.Save in file:saves the storage data to a file (*.csv file).Read from file: loads storage data from a file (*.csv file).

A basic data storage flow:

Start Storage

Stop Storage

Clear Stored Data

Data Storage Setting

Storage Data Readout

Save in File

Readout