Kalibrasi Mesin Anestesi - Datex Ohmeda

24
Mesin Anestesi Aespire SE PT. Utama Sarana Medika

description

kalibrasi

Transcript of Kalibrasi Mesin Anestesi - Datex Ohmeda

Mesin Anestesi Aespire SE

PT. Utama Sarana Medika

Service Checkout

Service mode Calibrations :

• O2 Calibration

• Zero Flow and Airway Sensors

• Adjust Drive Gas Regulator

• Airway Sensor Span

• PEEP Valve Calibration

• Inspiratory Valve Calibration

• Pressure Sensitivity

Self tests • powerup tests

• continuous tests

• periodic tests

Powerup tests

• Sequential watchdog

• Logical watchdog

• Data RAM walking pattern test

• FLASH ROM CRC verification

• PEEP Valve test

• PEEP Safety Valve test

Continuous tests

These tests are run continuously during normal operation and alarms are associated with each test.

• Supply voltage checks

• Battery voltage checks

Periodic tests

These tests are run every 30 seconds during normal operation.

• CPU Test

• Display RAM walking pattern test

• Data RAM walking pattern test

• FLASH ROM CRC verification

• Inspiratory control valve DAC and voltage feedback

• PEEP control valve DAC and voltage feedback

Service Mode

Service Mode

About Ventilator

Alarm Log

Error Log

Language

User Settings

System Configuration

Calibrations

O2 Calibrations

Zero Flow and Airway Sensors

Adjust Drive Gas Regulator

Airway Sensor Span

Airway Sensor Span

procedure calculates a gain coefficient for the airway pressure transducer.

PEEP Valve Calibration

Inspiratory Valve Calibration

Pressure Sensitivity

Terima Kasih