Iongantry @ HIT
Prof. Dr. Thomas HabererScientific-technical Director
Heidelberg Iontherapy Center
![Page 2: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/2.jpg)
GoalThe key element to improve the clinical
outcome is local control!
entrance channel:• low physical dose• low rel. biol. effiency
tumour:• high physical dose• high rel. biol. effiency
![Page 3: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/3.jpg)
Rasterscan Method
scanning offocussedion beamsin fastdipole magnets
active variationof the energy,focus andintensity in theaccelerator andbeam lines
Haberer et al., NIM A , 1993
![Page 4: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/4.jpg)
final sumseparate optimizationMulti-field optimization
simultaneous optimization
![Page 5: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/5.jpg)
Scanning Ion Gantry / Rationale
TPS studies (beam scanning only) give:
• Benefits for ~20% of skull base tumors• Spinal/cervical chordoma benefit significantly (robustness of plans)• Pancreatic/retroperitoneal tumors can be treated with a vertical beam
but with a gantry improved treatments can be realized • Some lung tumor situations may benefit• No benefit for prostate/pelvic tumors was found
![Page 6: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/6.jpg)
Th. Haberer
Motivation Gantry
Advantage of a rotating
beamline
Pancreas, supine position via gantry advantageous
![Page 7: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/7.jpg)
Scanning Ion Gantry / Requirements
Clinical: • Isocentric set-up and a fixed floor• Identical field size in all beamlines of 20 cm x 20 cm• Integration of fluoroscopy systems in two planes (IGRT, organ movement)Technical / financial:• Normal conducting elements (field quality of about 10-4 in 90-degree
bending magnet, price, 330 days 24/7 op.)• Full rotation (clinical workflow, minor saving)• Barrell-type (less bending than cork-screw)• Scanning upstream to the last bending (radius vs. weight)• Edge focussing (large SAD)• Truss-based structure
![Page 8: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/8.jpg)
Gantry Design Study @ GSI
barrel-type
box-girderconstruction20mm walls (center)
675 to
max. lateraldeformation 0.3mm
temp. coefficient0.2mm / oC SEAG/ACCEL/GSI
![Page 9: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/9.jpg)
GantrySegment
Prototype @ GSI
test of scanning gantrysegment at GSI
![Page 10: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/10.jpg)
Design for HIT
MT Aerospace
• isocentric barrel-type
• world-wide firstion gantry
• 2D beam scanningupstream to finalbending, almostparallel due to edgefocussing
• ± 180o rotation3o / second
• 13m diameter25m length600 to rotating(145 to magnets)
MT Mechatronics
![Page 11: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/11.jpg)
HIT-Gantry: Building Interface
MT Mechatronics
![Page 12: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/12.jpg)
HIT-Gantry: Maintenance Position
MT Mechatronics
![Page 13: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/13.jpg)
Integration of 90°-bending Magnet
MT Mechatronics
![Page 14: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/14.jpg)
High Precision Component Positioning
MT Mechatronics
![Page 15: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/15.jpg)
Medtechintegration of • robotic patient positioner• digital X-ray systems (30Hz)• stereotactic equipment
(PET-upgrade available)
MT Mechatronics
![Page 16: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/16.jpg)
Patient Environment / Nozzle
Siemens Medical Solutions
• challenging interfaces
• balance functionalityand workflow withsafety
• collision avoidance:coupled controls of gantry and robotsoft bumperslaser curtains
• isocenter definition:room-fixed vs.rotating reference
MT Mechatronics
![Page 17: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/17.jpg)
Patient Environment / Nozzle
![Page 18: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/18.jpg)
Emergency Analysis
best case worst case
Siemens Medical Solutions
In case of severe problems, how fast do we have to be (20s … 60s)?
![Page 19: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/19.jpg)
Assembly and Mounting
![Page 20: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/20.jpg)
Gantry: first beam
at theisocenter
January 4th, 2008
![Page 21: Iongantry @ HIT - PTCOG](https://reader034.fdocuments.us/reader034/viewer/2022050108/626cc7eb75f6fa46eb3d76f4/html5/thumbnails/21.jpg)
Thank you for your attention !
(Intensity modulated raster scan, 12C at 430 Mev/u, October 15th 2007)