Installation and Configuration Manual for VMP-300-based* Instruments and Boosters *VMP300-based instruments are: SP-200 SP-240 SP-300 VSP-300 VMP-300 HCV-3048

Transcript of Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment...

Page 1: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Installation and

Configuration Manual

for VMP-300-based* Instruments

and Boosters

*VMP300-based instruments are: SP-200 SP-240 SP-300 VSP-300 VMP-300 HCV-3048

Page 2: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review
Page 3: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments sConfiguration Manual

i

Certificate Of Conformity

We certify that all goods detailed below have been inspected and tested, and unless otherwise stated conform with the order of the customer, the drawings and specifications of Bio-Logic Science Instruments SAS. Instruments used to calibrate this equipment is traceable to NIST Standards. CE certificates available in paragraph 12.

Model

Chassis serial number

SP-200 chassis

SP-240 chassis

SP-300 chassis

VSP-300 chassis

VMP-300 chassis

HCV-3048 chassis

The Certificates of Conformity of the channel boards, booster channels, cell cables, and accessories (IS1, MP-MEA, Analog Ramp Generator) are available on request. Firmware of EC-Lab®: ....... Signed, for and on behalf of Bio-Logic Science Instruments SAS: Date:

Page 4: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

ii

Contact Bio-Logic Instruments

ITS users can find Bio-Logic contact or representative for warranty, service or technical support by clicking on the following link: http://www.bio-logic.info/ask-for-a-quote-contact-us/

French office Bio-Logic SAS Web: www.bio-logic.info Phone: +33 476 986 831 Fax: +33 476 986 909 Address: 1 rue de l'Europe Claix 38640 France

USA office Bio-Logic USA, LLC Contact: Bill Eggers Phone: +1 865-769-3800 Fax: +1 865-769-3801 Address: P.O.Box 30009 Knoxville 37930 USA

Indian office Bio-Logic Science Instruments Pvt. Ltd Contact: Samuel Varghese Phone: +912225842128 Address: 304,Orion Business Park, Next to Cine Wonder, G. B. Road, Thane(W) Mumbai 400 607 India

Representatives : see www.bio-logic.info

Page 5: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments sConfiguration Manual

iii

Safety and Legal Information

To avoid injury to an operator the safety precaution given bellow, and throughout the manual, must be strictly adhered to, whenever the equipment is operated. Only advanced user can use the instrument. Bio-Logic SAS accepts no responsibility for accidents or damages resulting from any failure to comply with this precaution.

Intended use of the equipment

The VMP-300 based instruments are intended for electrical measurements only. They should be used for no other purpose.

The VMP-300 based instruments have been designed for exclusive use with the VMP-300 modules. The VMP-300 based instruments are electrical laboratory equipment intended for use by professionals and in laboratories, commercial and light-industrial environments (indoor). Instrumentation and accessories shall not be connected to humans.

Instructions for use To avoid injury to an operator the safety precautions given below and throughout the manual must be strictly adhered to whenever the equipment is operated. Only advanced users may use the instrument. Bio-Logic SAS accepts no liability for accidents or damage resulting from any failure to comply with these precautions.

Grounding

WARNING! The instrument should be connected to the safety ground (Protective Earth). The continuity of the ground connection should be checked periodically. Any interruption of the Protective Earth (PE) conductor outside the instrument could result in personal injury. Use only the IEC320-C13 power cords which have been supplied with the instrument and designed for the appropriate current rating (10 A max) and including a protective earth conductor.

Atmosphere and Temperature

- Do not operate in an Explosive Atmosphere - Ensure that the ventilation grids are not obstructed on the front and rear sides of the VMP-300 based instruments. An external cleaning of the air filter can be performed with a vacuum cleaner if necessary (see Maintenance chapter), to prevent excess ambient temperature (The equipment produces heat), please consult our specialists to discuss the best location in your lab for the instrument (avoid small &closed spaces).

Page 6: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

iv

Avoid Unsafe Equipment

The equipment may be unsafe if any of the following statements apply: - Equipment shows visible damage, - Equipment has failed to perform an intended operation, - Equipment has been stored in unfavourable conditions, - Equipment has been subjected to physical stress.

If there is any doubt as to the service ability of the equipment, do not use it. Have it properly checked out by a qualified service technician.

Live Conductors

When the equipment is connected to its measurement inputs or supply, the opening of covers or removal of parts could expose live conductors. Only qualified personnel, who should refer to the relevant maintenance documentation, must do adjustments, maintenance or repair.

Equipment Modification

The equipment described in this manual has been built with approval safety components and has been designed in accordance with safety standard IEC61010. It has been supplied in a safe condition. To avoid introducing safety hazards, never install non-standard parts in the equipment, or make any unauthorized modification. To ensure safety, always return the equipment to Bio-Logic SAS for service and repair.

Page 7: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments sConfiguration Manual

v

General Safety Information

Class I

This product is a Safety Class 1 instrument, which means it has a protective earth terminal. That terminal must be connected to earth ground through a power source equipped with an earth ground. Before installation or operation, check the power supply and review this guide for safety warnings and instructions.

The equipment is not isolated (the chassis ground is connected to the Earth safety).

Warranty and liability claims in the event of injury or material damage are excluded when they are the result of one of the following.

Improper use of the device,

Improper installation, operation or maintenance of the device,

Operating the device when the safety and protective devices are defective and/or inoperable,

Non-observance of the instructions in the manual with regard to transport, storage or installation,

Unauthorized structural alterations to the device,

Unauthorized modifications to the system settings,

Inadequate monitoring of device components subject to wear,

Improperly executed and unauthorized repairs,

Unauthorized opening of the device or its components,

Catastrophic events due to the effect of foreign bodies.

ONLY QUALIFIED PERSONNEL should to install this equipment Refer to the Hardware Installation Chapter for more information.

WARNING

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and fixed.

CAUTION

A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and fixed.

Page 8: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

vi

Symbols Used

NOTE: The following symbols may appear within the documentation or on the appliance.

Caution This symbol is intended to alert the user to the presence of important operating and maintenance (servicing) instructions in the literature accompanying the appliance.

Dangerous Voltage This symbol is intended to alert the user to the presence of uninsulated dangerous voltage within the product’s enclosure that may be of sufficient magnitude to constitute a risk of electric shock to persons.

This Marking is used to indicate that the equipment is compliant with the European Standards about Safety, Electro-Magnetic Compatibility and The RoHS - Restriction of Hazardous Substance (no SVHCs) - Directives The IEC standards used open the principle of mutual recognition in order to eliminate technical barriers to the free movement of goods.

Waste Electrical and Electronic Equipment (WEEE) This Marking is used to indicate that these Electrical and Electronic Equipments should be collected when arrival at their end of live.

Page 9: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments sConfiguration Manual

1

Table of contents Certificate Of Conformity ......................................................................................... i Contact Bio-Logic Instruments ................................................................................ ii Safety and Legal Information ................................................................................. iii General Safety Information ..................................................................................... v

1. Introduction to the product range ............................................................................... 5

1.1 Instruments description .......................................................................................... 6

1.1.1 Single channel/Multichannels/internal boosters ........................................... 6 1.1.1.1 SP-240/SP-300 slot numbering ................................................................... 7 1.1.1.2 VSP-300 slot numbering .............................................................................. 7 1.1.1.3 Horizontal version ........................................................................................ 7 1.1.1.4 Vertical version ............................................................................................ 7 1.1.1.5 VMP-300 slot numbering ............................................................................. 8

1.1.2 External booster: HCV-3048 ........................................................................ 8

1.2 New board installation in an existing instrument ................................................... 11

1.3 Software features ................................................................................................. 12

2. Getting started ............................................................................................................ 13

2.1 Hardware installation ............................................................................................ 13

2.1.1 Preparing the physical installation ............................................................. 13 2.1.2 Safety emergency shutdown ..................................................................... 13 2.1.3 Ventilation requirement .............................................................................. 13

2.1.3.1 Potentiostats/galvanostats ......................................................................... 13 2.1.3.2 HCV-3048 ................................................................................................. 15

2.2 Unpacking and inspection .................................................................................... 15

2.3 Electrical equipment requirements ........................................................................ 15

2.4 Safety precautions ................................................................................................ 16

2.5 Communication connections ................................................................................. 16

2.5.1 Direct USB connection .............................................................................. 16 2.5.2 Direct Ethernet connection ........................................................................ 16 2.5.3 Network connections ................................................................................. 17

3. Software installation on the computer ...................................................................... 18

3.1 EC-Lab® software installation ............................................................................... 18

3.2 EC-Lab® Express software/OEM package installation .......................................... 22

3.3 Errors during the installation ................................................................................. 22

3.4 PC installation and configuration .......................................................................... 23

3.4.1 TCP/IP installation and configuration ......................................................... 23

3.5 USB driver installation .......................................................................................... 25

3.5.1 Windows XP installation ............................................................................ 25 3.5.2 Windows Seven, Eight, 10 & Vista installation ........................................... 27 3.5.3 Uninstall USB drivers ................................................................................. 28

4. Connecting the PC to the instrument ........................................................................ 31

4.1 Network parameter configuration with the Ethernet connection ............................ 31

4.2 Connecting to the instrument using EC-Lab software ......................................... 31

4.2.1 IP address modification of the instrument .................................................. 33

4.3 Connection using EC-Lab Express software ....................................................... 34

4.3.1 IP address modification of the instrument .................................................. 35

Page 10: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

2

4.4 Windows Security Alert ......................................................................................... 37

4.5 Firmware Upgrading with EC-Lab software ......................................................... 37

4.6 Firmware Downgrading with EC-Lab software .................................................... 37

5. Connection to the cell ................................................................................................ 41

5.1 Cell cable to the instrument .................................................................................. 41

5.1.1 Cell cable and potentiostat/galvanostat board association ......................... 41 5.1.2 Boards description ..................................................................................... 41

5.1.2.1 Potentiostat/galvanostat/ZRA board .......................................................... 41 5.1.2.2 Booster boards .......................................................................................... 42

5.1.2.2.1 1A/48V booster board .......................................................................... 42 5.1.2.2.2 2A/30V; 4A/14V; 10A/5V booster board ............................................... 42

5.1.3 HCV-3048 booster chassis ........................................................................ 43

5.2 Connecting the cell cable to your cell. .................................................................. 43

5.2.1 Cables description when used as standalone potentiostat or with booster boards 43 5.2.2 Standard connection.................................................................................. 45

5.2.2.1 Standard three-electrode connection ......................................................... 45 5.2.2.2 Two-electrode connection ......................................................................... 46 5.2.2.3 Four-electrode connection ......................................................................... 46

5.2.3 CE to Ground connection mode ................................................................ 47 5.2.4 WE to Ground connection mode ................................................................ 47 5.2.5 High voltage control connection mode ....................................................... 47 5.2.6 Floating mode............................................................................................ 48

5.3 Specific connection for high current boosters. ...................................................... 48

5.3.1 Connection to the cell with 1A/48V booster ............................................... 48 5.3.2 Connection to the cell with 2A/30V, 4A/14V and with 10A/5V boosters ..... 49 5.3.3 Connection to the cell with HCV-3048 external booster ............................. 50

5.3.3.1 Standard connection.................................................................................. 50 5.3.3.2 Additional power cables/connector can be purchased ............................... 53

5.4 Auxiliary inputs/outputs (DB9) .............................................................................. 54

6. Advanced features ...................................................................................................... 57

6.1 Bandwidth ............................................................................................................ 57

6.1.1 Bandwidth selection of a standard channel ................................................ 57 6.1.2 Bandwidth selection of a channel connected to ULC and/or one booster ... 57

6.2 Floating mode....................................................................................................... 58

6.3 Filtering ................................................................................................................ 58

6.4 External device control and recording ................................................................... 59

6.4.1 General description ................................................................................... 59 6.4.2 Rotating electrodes control ........................................................................ 61

6.4.2.1 Connections .............................................................................................. 61 6.4.3 Temperature control .................................................................................. 63 6.4.4 Electrochemical Quartz Crystal Microbalance coupling ............................. 63

6.5 Virtual potentiostat ................................................................................................ 65

7. Options & Accessories .............................................................................................. 66

7.1 Analog Ramp Generator (ARG) ............................................................................ 66

7.2 Isolation System (IS1) .......................................................................................... 66

7.3 Current Collector (CC5) ........................................................................................ 68

7.4 MP-MEA option .................................................................................................... 69

Page 11: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments sConfiguration Manual

3

7.4.1 Description ................................................................................................ 70

7.5 Test boxes for calibration check and user training ................................................ 70

7.5.1 Dummy Cell 2 (DC2) ................................................................................. 70 7.5.2 Dummy cell for booster .............................................................................. 71 7.5.3 Test Box 2 ................................................................................................. 71 7.5.4 Test Box 3 ................................................................................................. 72

7.6 Temperature probe ............................................................................................... 72

7.7 Battery holders ..................................................................................................... 72

7.7.1 Battery Holder: BH-1 ................................................................................. 72

7.8 Labview VIs .......................................................................................................... 73

7.9 Electrochemistry accessories ............................................................................... 73

8. Validation and calibration .......................................................................................... 74

8.1 Channel validation ................................................................................................ 74

8.1.1 Channel board without EIS option ............................................................. 75 8.1.2 Channel board with EIS option .................................................................. 76 8.1.3 Validation of the ultra-low current module. ................................................. 78 8.1.4 Validation of the booster board .................................................................. 78

8.2 Channel calibration with EC-Lab® software........................................................... 79

8.2.1 Potentiostat/galvanostat board calibration ................................................. 79 8.2.2 Booster board calibration ........................................................................... 84 8.2.3 HCV-3048 booster ..................................................................................... 87

9. Maintenance ................................................................................................................ 91

9.1 Calibration ............................................................................................................ 91

9.1.1 SP-200/SP-240/SP-300/VSP-300/VMP-300 .............................................. 91 9.1.2 HCV-3048 external booster ....................................................................... 91

9.2 Ventilation ............................................................................................................ 91

9.3 Cleaning ............................................................................................................... 91

9.3.1 Ventilation grids ......................................................................................... 91 9.3.2 Air filters cleaning ...................................................................................... 91 9.3.3 Fuses: ....................................................................................................... 92

10. Technical Specifications ........................................................................................ 94

10.1 Equipment Ratings ............................................................................................... 94

10.1.1 Electrical & Mechanical specifications ....................................................... 94 10.1.2 Environmental ........................................................................................... 94

10.2 Channel specifications .......................................................................................... 94

10.2.1 Channel board ........................................................................................... 94 10.2.2 Ultra Low Current ...................................................................................... 96 10.2.3 1A/48V booster .......................................................................................... 97 10.2.4 2A/30 V booster ......................................................................................... 98 10.2.5 4A/14V booster .......................................................................................... 99 10.2.6 10 A/ 5V booster ...................................................................................... 100 10.2.7 HCV-3048 used with HV-48 electrometer ................................................ 101 10.2.8 MP-MEA .................................................................................................. 103

10.3 PC requirements ................................................................................................ 103

11. Troubleshooting ................................................................................................... 105

11.1 Data saving ........................................................................................................ 105

11.2 PC Disconnection ............................................................................................... 105

Page 12: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

4

11.3 Effects of computer save options on data recording ........................................... 105

12. EC declaration of conformity ............................................................................... 106

12.1 SP-200 ............................................................................................................... 106

12.2 SP-240/SP-300 .................................................................................................. 107

12.3 VSP-300 ............................................................................................................. 108

12.3.1 Horizontal VSP-300 ................................................................................. 108 12.3.2 Vertical VSP-300 ..................................................................................... 109

12.4 VMP-300 ............................................................................................................ 110

12.5 HCV-3048 .......................................................................................................... 111

13. Glossary ................................................................................................................ 112

14. Index ...................................................................................................................... 114

Page 13: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

5

1. Introduction to the product range

Historically, the first of our potentiostats was designed to study intercalated compounds with long experiment times due to slow diffusion coefficients. It was a multichannel instrument that worked in either galvanostatic mode or potentiostatic mode (each channel was devoted to one of either mode). Then the interest for multichannel potentiostats increased with battery testing, corrosion study & biosensors development. So, new versatile instruments have been developed to address all these applications. This was the first generation of the Bio-Logic instruments, based on the VMP3 technology. Since 2009, a new generation of instrument has been introduced with the SP-300. This generation offers highest current & voltage sensitivity and also the best impedance specification of the market. This is the generation based on the SP-300 technology. This family of instrument includes SP-200, SP-240, SP-300, VSP-300 and VMP-300 instruments. This range of instruments is called VMP300-based instruments. The instruments consist of a single control and communication board associated with one or more potentiostat/galvanostat/ZRA channel board(s), depending on the configuration. For example, there is one or two channel(s) for the SP series, six channels for VSP-300, and sixteen channels for VMP-300. Once started, the on-board computer fully controls the entire experiment. Therefore, if the host computer (PC, Mac, etc.) fails, or the connection to the Bio-Logic instrument is interrupted, the experiment will continue to be performed and data collected by the instrument. Data can be retrieved from the instrument when connection to the host computer is re-established. All units are designed to be multi-user instrument. Several users may simultaneously connect to the same instrument, offering great flexibility for our multichannel devices. These computers can be connected to the instrument through an Ethernet or USB connection. All the instruments

are controlled by EC-Lab Additionally, all the instruments can be coupled with ultra-low current option and have slots (except SP-200 and SP-240) which can accept other modules such as booster board or additional potentiostat/galvanostat/ZRA board.

Fig. 1: SP-300 instrument in bipot configuration with the 10 A booster board & the Ultra Low Current (ULC) electrometer.

The aim of this manual is to guide the user through the instrument’s installation and configuration. This manual is composed of several chapters. The first is a general description of the instruments. The second and third parts describe how to install the software and how to configure the computer. The fourth and the fifth parts concern the installation and configuration

Page 14: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

6

of the instrument and how to connect the instrument to the computer. The sixth chapter is dedicated to the cell connection in different configurations. The seventh chapter deals with the advanced features. Useful accessories are described in the eighth part. Finally calibration, maintenance and specifications are shown in the two last parts. NOTE: WHEN AN USER RECEIVES A NEW UNIT FROM THE FACTORY, THE SOFTWARE AND FIRMWARE

ARE INSTALLED AND UP TO DATE. THE INSTRUMENT IS READY FOR USE. IT DOES NOT NEED TO BE

UPGRADED.

1.1 Instruments description

1.1.1 Single channel/Multichannels/internal boosters

Depending on the instrument selected, the number of channels and the option (low current/boosters) may be different. Limit of instrument configuration:

VSP-300 horizontal chassis accepts a maximum of 4 booster boards (10 A and 2A) and these boosters have to be inserted in the slots #1 to #4 (not #5 and #6).

VSP-300 vertical chassis accepts a maximum of 4 booster boards (10 A and 2 A) and these boosters can be inserted in any slot.

Because of power limitation, VMP-300 chassis accepts a maximum of 12 booster boards if powered with 110Vac electrical network. They can be inserted in any slot.

The numbering of the slots is explained hereafter (Fig.2 to Fig. 4) A description of the instruments is as follows:

Table 1: General description of the instruments.

Type of instrument

Name Communication

board Calibration

board

Available slots in the instrument

Other Module (number of slot

used by the module)

Single Channel

SP-200 √ √ 1 ULC (0) and/or ARG (0)

SP-240 √ √ 2

4A/14Va (1) and ULC (0) and/or ARG (0)

SP-300 √ √ 2

1A/48V (1) and/or 2A/30V (1) and/or 4A/14V (1) and/or 10A/5V (1) and/or ULC (0) and/or ARG (0)

Multichannel

VSP-300 √ √ 6

1A/48V (1) and/or 2A/30V (1) and/or 4A/14V (1) and/or 10A/5V (1) and/or ULC (0) and/or ARG (0)

VMP-300 √ √ 16

1A/48V (1) and/or 2A/30V (1) and/or 4A/14V (1) and/or 10A/5V (1) and/or ULC (0) and/or ARG (0)

a provided by default ULC: Ultra Low Current ARG: Analog Ramp Generator

Page 15: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

7

1.1.1.1 SP-240/SP-300 slot numbering

It is from the left to right. The slot 1 is the slot the closest to the communication board. The other slot is dedicated to the booster board or the second potentiostat board.fff

1.1.1.2 VSP-300 slot numbering

1.1.1.3 Horizontal version

Fig. 2: VSP-300 numbering.

1.1.1.4 Vertical version

Fig. 3: Vertical VSP-300 numbering.

5

6 3

2

1

4

Page 16: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

8

1.1.1.5 VMP-300 slot numbering

Fig. 4: VMP-300 numbering.

1.1.2 External booster: HCV-3048

The HCV-3048 is a high power external booster that can deliver a maximum of 30 A at a voltage of 48 V. Due to the high power, it requires specific ventilation and high velocity fans and requires its own external chassis. Even if this booster has a specific chassis it cannot work independently and must be connected to a pstat/gstat board on a VMP-300 family instrument. The booster purpose is only to provide the power i.e. the current while the pstat/gstat will handle the voltage measurement and control of the set values. The connection between the pstat/gstat and the booster as well as the connection to the cell will be explained in Chapter 6.Connecting this booster to pstat/gstat allows the user to access a voltage range of 48 V in addition to the ranges already available with the pstat/gstat. This voltage is available in standard control mode as well as for any impedance measurements. This requires the use of a specific cell cable called HV-48 for which the voltage available on is 48 V. This cell cable will also be described in Chapter 6. The user will also have access to three current ranges: 30 A, 3 A and 300 mA. The standard current ranges of the pstat/gstat will not be available. More information are given in Chapter 10.

1 2 3 4 5 6 7 8 16

Page 17: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

9

Fig. 5: HCV-3048 front panel.

Table 2: Description of the components of the front panel of the HCV-3048.

Item Description Item Description

1 Front fans for heat sink To extract heat of the heat sink of the power board. Air flow goes from front and bottom to rear.

4 Emergency Stop This button can be pushed when necessary. It will open the relays that connect the power supply of the instrument to the cell. The Status will turn red.

2 LIO (Liaison Interface Option) In and Out (SubD15 connectors) Used to connect to the potentiostat/galvanostat board and to another HCV-3048 when used in parallel. Up to 4 boosters can be connected in parallel, increasing the current ability up to 120 A.

5 Power sockets The 6 mm banana + and – power sockets allow to connect 10 mm2 power cables that can pass a maximum current of 70 A. The maximum voltage difference between these two cables must be lower than 55 V.

3 Status LED Status Description

Off o The instrument is off line (= turned off).

Green o The instrument is connected to the PC and can be readily used.

Red o The instrument is not yet connected to the PC. o The instrument has stopped due to a major error. Major errors: - The Emergency Stop button is pushed. - A TTL signal is sent to the Open In BNC connector on the back panel. - The AC/DC power supply of the instrument is under failure status. - The instrument is overheating.

Note: Exhaustive LED behavior is explained in more details in section 10.2.7.

Page 18: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

10

Fig. 6: HCV-3048 back panel.

Table 3: Description of the components of the back panel of the HCV-3048.

Item Description Item Description

1 Rear ventilation grids for heat sink This is where the air and heat will come out.

5 Power socket Standard C19 socket to power the booster.

2

Open In It is a high impedance BNC input that can be remotely accessed to trigger the Emergency Stop. It will open the relays of the power circuit. The Status LED will turn red.

6 Power switch

3 Fuses 2 x 16 AF This is where the protection fuses are located. Each is a 16 A Fast fuse.

7 The back panel also features additional safety information as well as the serial number of the instrument.

4 Rear ventilation grids for power supply This is where the air and heat from the power supply will come out.

WARNING 1) When HCV-3048 are in parallel, the BNC Open In inputs must also be connected such that all boosters stop at the same time when the Emergency Stop button is pushed or when an external device triggers an Emergency Stop via Open In. 2) This is a high input impedance as long as the Emergency Button is not pushed. If it is pushed with a connected instrument, the outer conductor of the BNC connector will be shorted to Ground and can damage the output of the connected instrument.

Page 19: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

11

Fig. 7: HCV-3048 back panel additional information.

Fig. 8: HCV-3048 bottom panel.

Table 4: Description of the component of the bottom panel of the HCV-3048.

Item Description

1 Bottom filter for power supply This is where the air cooling the power supply is suck up. It has a filter that needs to be cleaned and/or changed. See Chapter 9 Maintenance.

1.2 New board installation in an existing instrument

When a user orders new boards (channel board, low current board or a booster board), they can install them themselves. The procedure for this operation is described in the corresponding service note. With the newly provided boards, the latest software version must always installed. This can be downloaded from our website. The board installation procedure consists of 4 steps:

1- Install the new software version on the computer and on the instrument firmware in order to have the old unit and the new boards in the same software version.

2- Power off the unit and unplug it. 3- Install the new boards in the chassis, plug in the unit and power it on.

4- Calibrate the new boards with EC-Lab.

Page 20: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

12

The low current boards are coupled with a channel board, so a low current board must ALWAYS be calibrated with the corresponding channel board. The board installation can be validated in two steps:

Firstly, it is recommended to calibrate the channel with the calibration tool which is available in the “Tools” menu of EC-Lab®. The procedure is described in 8 and also in the Technical Note #18.

If a further check is needed, it is possible to validate the boards thanks to the dummy cells provided with the board(s).

1.3 Software features

In general, any electrochemical experiment is composed of open circuit sequences and/or galvanic sequences performed while measuring the potential, and/or imposed potential sequences performed while measuring the current. Common electrochemical techniques, such as Cyclic Voltammetry (CV), Chronopotentiometry (CP), etc. are obtained by combination of these elementary sequences. These combinations appear in EC-Lab as flow-diagrams for easy visual description for the user. At various points within any experimental sequence, conditional tests can be performed on the working electrode potential or current, the counter electrode potential, or the external parameters. These conditional tests force the experiment to go to the next step, to loop to a previous sequence, or end the sequence or experiment. The application software package provides useful protocols for general electrochemistry, corrosion, batteries, super-capacitors and custom applications. Standard graphic functions such as re-scaling, zoom, linear and log scales are available. Standard processed files can be created at the user's convenience upon running an experiment for the purpose of real time display of the experiments in progress. Post processing is also possible using built-in options to create variables at the user's convenience, such as derivative or integral values, etc. Raw data and processed data can be exported as standard ASCII text files.

The user can find more information about EC-Lab software in the software manuals, available in the “Help” menu of the software.

It is assumed that the user is familiar with Microsoft Windows© and knows how to use the

mouse and keyboard to access the drop-down menus.

Page 21: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

13

2. Getting started

2.1 Hardware installation

2.1.1 Preparing the physical installation

Questions YES NO

Has the incoming electrical power been taken into account ? (1)

Have the environmental specifications been taken into account ? (Room temperature, room humidity, pollution degree, EMC environment...) (1)

Have the electrical connections cables been supplied?

(1): See 10.1 Equipment Ratings If all the answers are YES then you can proceed with the installation.

2.1.2 Safety emergency shutdown

As described in Part 1.1.2, HCV-3048 is a high power instrument and features an Emergency

Stop button. This button will This button will open the output of the power at the power

sockets. WARNING When HCV-3048 are used in parallel to increase the delivered current, it is possible to simultaneously open the relay of the power output on all the boosters parallelized with only one Emergency Stop by connecting the BNCs in the back of all four instruments. The connection can be made by using BNC T-connector on the Open In of all boosters except the main one on which the potentiostat is connected (See 5.1.3 for more information on connections).

2.1.3 Ventilation requirement

As our instruments are delivering power, they produce heat which needs to be evacuated to maintain the temperature inside of the instrument as constant as possible. Precautions must be taken to avoid blocking the air flow. Blocking the air flow will increase the temperature of the instrument, hence reducing the lifetime of the components as well as impairing its capabilities. All the specifications and capabilities listed in this manual are guaranteed under recommended operation, which means normal air flow

2.1.3.1 Potentiostats/galvanostats

In all our instruments (except the vertical VSP-300) the air flows from the bottom or the front to the back of the instrument (Fig. 9). For the vertical VSP-300 the air flows from the right to the left when facing the instrument (Fig. 10).

Page 22: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

14

Fig. 9: Air flow for SP-200/SP-240/SP-300/Horizontal VSP-300 and VMP-300 chassis.

WARNING There should be sufficient space at the rear of the instrument to prevent the hot air from reentering the instrument. A minimum distance of 30 cm should be observed. It must also be ensured that nothing can block the air flow at the bottom of the instrument. Something as harmless as a sheet of paper can reveal itself quite dangerous if it is stuck underneath the instrument, on the ventilation grids, blocking the air flow. For the vertical VSP-300 it must be ensured that sufficient space if left on both sides of the instrument. Again a minimum distance of 30 cm should be observed.

Fig. 10: Air flow for vertical VSP-300

Page 23: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

15

2.1.3.2 HCV-3048

WARNING In HCV-3048, the air flows from front to the rear of the instrument (Fig. 13-14). It should be placed at a minimum of 30 cm from a wall. Otherwise, the lifetime and the specifications of the instrument are not guaranteed.

Fig. 11: Air flow for HCV-3048

WARNING As a general rule, the instruments should not be seated on an inflammable surface. Please prefer inert surface such as glass or stone. This does not apply for instruments connected in parallel, where they would be vertically stacked.

2.2 Unpacking and inspection

. VSP-300, VMP-300 and HCV-3048 are heavy instruments (22, 30 and 26 kg, respectively) and necessitate two persons to unpack and carry them, to avoid crashing and health hazard.

2.3 Electrical equipment requirements

Please find below the electrical specifications of the instruments:

Table 5: Electrical input specifications of the instruments described in this manual.

Instrument Input

Voltage Range /Vac

Frequency range/Hz

Power max /W

Fusesa

SP-200 [90;264] [50;60] 350 2 x 4 AF

SP-240 [90;264] [50;60] 350 2 x 4 AF

SP-300 [90;264] [50;60] 350 2 x 4 AF

VSP-300 [90;264] [50;60] 650 2 x 5 AF

VSP-300 (Vertical)

[90;264] [50;60] 650 2 x 5 AF

VMP-300 [90;264] [50; 60] 1500 2 x 10 AF

HCV-3048 [85;264] [47;40] 2000 2 x 16 AF

Page 24: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

16

2.4 Safety precautions

WARNING To avoid electrical shocks:

The equipment must be connected to a protective ground.

The equipment must be disconnected from the power source before it is opened.

To avoid electrostatic shocks:

Every lead from the instrument to the cells (or amplifier to cells) must be connected either to an electrochemical cell or to the testing box.

2.5 Communication connections

NOTE: When multichannel instruments (SP-300, VSP-300 or VMP-300) and/are used or when measurements require fast sampling rates, use of the Ethernet connection is highly recommended. Depending on your local installation, you can use a direct connection (one PC to one instrument) or a network connection (one or several PCs to one or several instruments). By default; the IP address of the instrument is 192.109.209.128 or 192.168.0.1.

2.5.1 Direct USB connection

This connection can be done easily using the USB connection cable. One end must be connected on the instrument communication board and the other one on the control unit of the computer.

2.5.2 Direct Ethernet connection

Connect the computer directly to the instrument can be done with the Ethernet cable. The Ethernet cable is provided with the instrument ans should have either green or black plugs on its ends. The IP addresses of both devices must be in the same network. This means that the first three groups of numbers in the IP address must be the same. For example, 192.109.209.128 or 192.168.0.1 for the instrument and 192.109.209.127 or 192.168.0.2 for the computer, respectively.

Fig. 12: Direct connection: one instrument to one PC.

Page 25: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

17

2.5.3 Network connections

Several PCs can be connected to the same instrument through the network. WARNING: check IP addresses before connection to avoid any IP conflicts (see TCP/IP configuration chapter).

Fig. 13: Ethernet cable for network connection (1 INSTRUMENT, several PCs).

Alternatively, a single PC can control several INSTRUMENTs through the network.

Fig. 14: Network connection: one PC to several instruments.

Page 26: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

18

3. Software installation on the computer

Before turning on the instrument, it is recommended to first install the EC-Lab® software and/or EC-Lab® express. For this operation, insert the CD-Rom in the computer. The installation is automatically launched with the “Autorun” function. The first software to be installed is EC-Lab®. If your computer does not have a CD-Rom drive, all software can be downloaded from the Bio-Logic website at: http://www.bio-logic.info/electrochemistry-ec-lab/downloads/

3.1 EC-Lab® software installation

Fig. 15: EC-Lab® software installation. Step 1.

Fig. 16: EC-Lab® software installation. Step 2.

Page 27: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

19

Fig. 17: EC-Lab® software installation. Step 3 and 4.

During the installation, two main folders are created.

“EC-Lab.exe”, the USB drivers, TeraTermPro and other files are located in the in the “Program Files” directory (Fig. 18 top).

“Batch”, “Data” (include the data samples), “Documents” (include manuals and getting started document), “Newsletter”, “Settings” (includes the default settings files), “Temp” folders and also the “EC-Lab.ini” & “VMPerr.txt” files (includes the initial settings) are installed in the “Documents” directory (Fig. 18 bottom).

Page 28: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

20

Fig. 18: EC-Lab® software installation. Step 5. (Top): Installation in the “Program Files” Directory. (bottom): Installation in the “Documents” Directory.

Page 29: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

21

Fig. 19: EC-Lab® software installation. Step 6.

Fig. 20: EC-Lab® software installation. Step 7.

Page 30: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

22

Fig. 21: EC-Lab® software installation. Step 8.

At the end of the installation, the instrument will be detected via USB is and ready to be connected and used. See 4: Connecting the PC to the instrument to connect to the instrument with the software.

3.2 EC-Lab® Express software/OEM package installation

The EC-lab® Express software package is optional software which offers the user some enhanced features and control over the regular EC-Lab® software. Installation of EC-lab® Express is not required to operate the instrument provided that EC-Lab® is installed. The installation of EC-lab® Express software and the OEM package is done in exactly the same way as for EC-Lab®. Please see the above section for the installation. NOTE: EC-lab® Express doesn’t support the VMP-300.

3.3 Errors during the installation

During the installation an error may occur if you are not the administrator of the computer. In this case it is necessary to contact your network administrator for the installation.

Fig. 22: Error during the installation.

Page 31: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

23

3.4 PC installation and configuration

IT IS HIGHLY RECOMMENDED TO ASK FOR ASSISTANCE FROM YOUR NETWORK ADMINISTRATOR. NOTE: IP ADDRESSES OF THE INSTRUMENT(S) AND THE COMPUTER(S) MUST REMAIN THE SAME

THROUGHOUT AN ENTIRE EXPERIMENT OR COMMUNICATION BETWEEN INSTRUMENT AND PC WILL

NOT BE MAINTAINED. THE AUTOMATIC WINDOWS UPDATE BE MUST DISABLED IN ORDER TO AVOID

ANY IP ADDRESS CHANGE WHILE RUNNING AN EXPERIMENT. THIS IS ESPECIALLY CRITICAL FOR

PROLONGED EXPERIMENTS.

3.4.1 TCP/IP installation and configuration

The instrument uses the TCP/IP (Transfer Control Protocol / Internet Protocol) to exchange data with the PC. This protocol uses IP addresses to identify hosts on a network, so you will need 2 IP numbers, one for the instrument and one for the PC. For a direct connection between the instrument and the PC, you can use the following numbers (default factory settings):

192.109.209.127 or 192.168.0.2 (PC)

192.109.209.128 or 192.168.0.1 (INSTRUMENT)

If you connect the PC and the instrument to your local network, you need to ask your system administrator for 2 VALID IP NUMBERS FOR YOUR INTRANET (and the sub-net mask and the gateway numbers if necessary). Note: 1- Before the installation of the TCP/IP protocol, your Ethernet board must be properly installed on your computer.

2- With Windows Vista, it is recommended to replace the default IP addresses with new ones even if the instrument is directly connected to the computer. Windows Vista does not accept universal IP addresses. You can use the following ones 192.109.209.127 for the PC and 192.109.209.128 for the instrument. The TCP/IP protocol must be installed on the PC computer to establish the connection with the instrument. If your computer is connected to a network, the TCP/IP protocol may already be installed. In that case the computer already has an IP address (obtained automatically). When the computer is connected directly to the instrument, it is necessary to give a static IP address to the computer. The following section describes how to give a static IP address to the computer: The procedure is show for Windows XP©, but the preocdure is similar for other windwos Operating System. 1- In the Control Panel, double click on the Network Connections icon. Then the Local Area

Connection window appears. Right click on the name and choose “Status” to see the computer IP address in the network.

Page 32: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

24

2- On the “General” tab click on Properties. This will load the window in Fig. 23.

Fig. 23: Local Area connection status.

3- Select Internet Protocol (TCP/IPv4

or TCP/IPv6) and click on the Properties button. The window in Fig. 24.

Fig. 24: Network window.

Page 33: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

25

4- At this point in the installation

(Fig. 25), the user has to activate the “Use the following IP address” box.

WARNING: THERE MIGHT BE ANOTHER

TCP/IP PROTOCOL INSTALLED CALLED

"TCP/IP DISTANT ACCESS", DO NOT CLICK

ON THIS LINE!

5- Enter the PC IP address, DO NOT

ENTER A NETWORK MASK (it will automatically be added) and click on the OK button.

WARNING: IP ADDRESSES MUST BE UNIQUE

IN A NETWORK 6- Restart the PC. Now the PC and the

INSTRUMENT are in the same network.

Fig. 25: TCP/IP properties window.

3.5 USB driver installation

The instrument can use a USB connection to exchange data with the PC. This connection requires USB drivers to be installed in the computer operating system. Installation of the drivers will vary depending upon the operating system of the computer.

We highly recommend that the user works with at least Windows 2000 to control the potentiostat through a USB connection.

For other Windows versions, the user will probably have to specify where to find the driver on the CD-Rom. In this window select the automatic installation of the software.

3.5.1 Windows XP installation

The way to proceed to install USB drivers is described below for Windows XP Pro. After connecting the instrument to the computer with the USB cable power on the instrument. When the user powers on the instrument, Windows automatically detects a new USB device. Then the following installation window appears:

Page 34: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

26

Fig. 26: USB device installation window. Step 1.

In this window, select “No, not this time” and click on “Next”. The following window is displayed:

Check that the EC-Lab installation CD is in the CD drive. If it is, Windows XP will automatically search on the CD, find the USB driver and complete the installation.

Page 35: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

27

Fig. 27: USB device installation window. Step 2 to 4.

Click on Finish. The potentiostat can now be connected to the computer through the USB connection. It is not necessary to restart the computer after this installation.

3.5.2 Windows Seven, Eight, 10 & Vista installation

With Windows seven, eight, 10 and Vista systems, the USB driver is automatically installed when the instrument is detected. The following message is displayed at the end of the installation:

Page 36: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

28

Fig. 28: USB device installation for Seven and Vista.

3.5.3 Uninstall USB drivers

For this operation, please open the folder: C:\Progrm Files\Bio-Logic\USBIO as described in the picture below. Double click on “uninstall” to proceed.

Fig. 29: Uninstall USB driver. Step 1.

The uninstall wizard is launched. Click on the “Uninstall” button to proceed.

Page 37: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

29

Fig. 30: Uninstall USB driver. Step 2.

Fig. 31: Uninstall USB driver. Step 3.

Page 38: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

30

Fig. 32: Uninstall USB driver. Step 4.

When you click on the “Finish” button the USB driver is completely uninstalled.

Page 39: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

31

4. Connecting the PC to the instrument

4.1 Network parameter configuration with the Ethernet connection

The Ethernet connection on the communication board is a 10/100 baseT compatible with every network. The USB connection is also integrated on this board. When it is installed on a Local Area Network (via the Ethernet connection), the instrument is automatically detected by the computers of the network. It becomes very easy to select an instrument in the network and modify its IP address via the Ethernet connection. This is possible with a MAC Address (set at the factory on the communication board) even if the instrument is not in the same network as the computer (before being connected together). All new instruments are delivered with the following IP address commonly used as default: 192.109.209.128 or 192.168.0.1. You can either manage your instrument directly with the computer (direct connection with the cable) or change the instrument’s IP address to add the instrument in your local network. The way to proceed is the same in both cases. The first step is the detection of the new instrument by a computer (directly or via the network). The second step is the IP address change before the connection, either to have both the instrument and the computer in the same LAN or to make a small network consisting of only the instrument and the PC.

Note: - to switch between EC-Lab and EC-Lab Express software, the instrument has to be switched off and restarted.

4.2 Connecting to the instrument using EC-Lab software

The procedure to connect your computer directly or via the network to the instrument is as follows.

1. Launch EC-Lab software V10.30 or higher

Fig. 33: Main window of EC-Lab®.

Page 40: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

32

2. In the “Devices” frame, click on the “+” button to add the instrument of interest to the list. Only one session of EC-Lab® is needed to control several instruments.

Fig. 34: “Devices” frame for connection.

Note: This step is only required the first time. The instruments configured previously are saved and will be displayed in the list of device each time EC-Lab® is opened. In that case, go directly to step 6.

3. The window shown in Fig. 35 appears. Click on “Refresh” to see the instruments present on your network. Note: The automatic search frame shows the name of the instruments detected with their serial numbers (#). A MAC address is given to each instrument at the factory. The MAC address is used to detect the instrument in a LAN even if its IP Address is not valid in the network. It is also used for the detection of instruments connected by USB.

Fig. 35: “New Device” window to select and add an instrument to the current devices.

4. Select the instrument and click on the “Select” button.

Note: If the IP address of the instrument is not valid, it has to be changed (see next section).

5. The instrument selected appears in the list displayed in the “Device” frame. Then, the

connection is established automatically and the circle to the left of the device name will turn green.

Page 41: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

33

Fig. 36: “Devices” frame for connection with the new device.

Note: if the instrument is already in the list, the user has to select the instrument in the “Device”

frame and then click on the “connect” button . It is possible to remove a device by clicking on the “-“ button.

When the connection is established “Connected” is displayed in the connection status of the “Connection” window. One can see the “Connection status” with the device type and the instrument’s IP Address.

4.2.1 IP address modification of the instrument

If the IP address of the instrument is not valid i.e. two IP addresses are the same or not in the appropriate network, the IP address has to be changed. In the “Devices” frame, click on the “+” button to open the “New Device” window. On the “New Device” window select the desired instrument and click on “Modify”. The following window then appears:

Fig. 37: “Advanced Ethernet settings” of the “New Device” window used to change the instrument IP Address.

Modify the IP Address to have a valid address in your network. Repeat this procedure with the Gateway and click on “Modify device”. A “configuration changed” message is displayed in green when the instrument receives the new IP address. Several “Bip” sounds are emitted by

Page 42: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

34

the instrument indicating that the communication board is reinitialized with the new IP Address. “Configuration changed” appears at the bottom of this window.

Fig. 38: New configuration.

Then click “OK” to display the “New Device” window where you have to click “Refresh” to refresh the window and select your instrument IP address. Now the instrument is ready for use. Note that it is possible to communicate with the instrument from another subnet with the following ports 23455 (broadcast), 23456, 23457 and 23458.

4.3 Connection using EC-Lab Express software

1) Launch EC-Lab Express software.

Fig. 39: Main window of EC-Lab® Express.

Page 43: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

35

2) In the “device” frame, all the instruments available (connected by USB, direct Ethernet or through a LAN) are displayed. The window shown in Fig. 40 below appears. Click on “Refresh” to see the instruments present on your network. Note: The automatic search frame shows the name of the instruments detected with their serial numbers (#). A MAC address is given to each instrument at the factory. The MAC address is used to detect the instrument in a LAN even if its IP Address is not valid in the network. It is also used for the detection of instruments connected by USB.

Fig. 40: Device connection window.

3) Select the instrument of interest and click on the “Select” button. Note: If the IP address of the instrument is not valid, it has to be changed (see next section).

Fig. 41: Device connection window (selected instrument).

4) Click on “Connect”. The “Connect” button turns into a “Disconnect” button and

becomes green to show the connection has been established.

4.3.1 IP address modification of the instrument

If the Ethernet connection requires changing the instrument’s IP address, select the instrument and click on “Modify”. The following window appears:

Page 44: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

36

Fig. 42: “Advanced Ethernet settings” of the “New Device” window used to change the instrument IP Address.

Modify the IP Address to have a valid address in your network. Repeat this procedure with the Gateway and click on “Modify device”. Then the new IP Address is sent to the instrument and a “configuration changed” message appears in green. Several “Bip” sounds are emitted by the instrument indicating that the communication board is reinitialized with the new IP Address. “Configuration changed” appears at the bottom of this window.

Fig. 43: New configuration.

Then click “OK” to display the “New Device” window and “Refresh” to refresh the window and select your instrument’s IP address. Click on the “Select” button. Now the instrument is connected and ready for use.

Page 45: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

37

4.4 Windows Security Alert

When the user tries to find an instrument in the network or by USB, the software will use a broadcast that may be stopped by windows firewall. In this case click on the “Allow access” button:

Fig. 44: “Windows security Alert” window.

4.5 Firmware Upgrading with EC-Lab software

When the user receives a new unit from the factory, the software and firmware (in the instrument) are installed and upgraded. The instrument is ready for use. It does not need to be

upgraded. However, when a new EC-Lab version is released (with new protocols or improvements) the firmware has to be updated and installed by the user.

4.6 Firmware Downgrading with EC-Lab software

It is possible to downgrade the firmware of the instrument in an advanced tool available in EC-

Lab. This procedure should only be done by an advanced user and in very special cases. Please contact technical support for assistance. WARNING:

If the user downgrades the firmware of the instrument, corresponding version of EC-

Lab must be used. For example, if the firmware of the instrument is the V10.20, the

user must control the instrument with EC-Lab V10.20.

Make sure that the older version supports the particular instrument.

Instruments Lowest compatible version

SP-200 V10.00 SP-240 V10.17 SP-300 V10.00 VSP-300 V10.17 VMP-300 V10.30

The firmware downgrade procedure is as follows: NOTE:

the example shows a downgrade from V10.23 to V10.12 but the procedure will be the same for other versions.

Page 46: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

38

1) Make sure that both versions are installed on the computer (here it is V10.12 and V10.23).

Fig. 45: Both versions on the computer.

2) Connect the instrument

with the latest software version (here: V10.23). The version # is listed in the header of EC-Lab window.

Fig. 46: connection under the latest version.

3) Open “Firmware Upgrade/Downgrade” tool in the “Tools” menu.

Fig. 47: Firmware Upgrade/Downgrade tool in the “Tools” menu.

4) The window displays the pathway of the current firmware flash (here: V10.23).

NOTE: For V10.33 and later, the flash file is located in the “Program files” directory.

Page 47: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

39

Fig. 48: Loading of the flash.

5) Click on the “Browse” button to select to the flash file (EC0368.flash) of the older version (here: V10.12). Click on “Open” when the flash file is selected.

NOTE: if the wrong file is selected, the communication board may be damaged.

Fig. 49: Flash location in the EC-Lab folder.

6) The pathway of

the flash of the old version (V10.12) is now shown in the “Firmware Upgrade/Downgrade” window.

Fig. 50: “Firmware Upgrade/Downgrade” window.

7) Click on the “Load”

button. A warning message shows up. Click on “OK” to continue the downgrade.

Fig. 51: Warning message.

Page 48: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

40

8) The downgrade process starts and “Upgrading...” appears on the bottom of the window (but the firmware is really downgraded).

Fig. 52: Flash loading.

Fig. 53: Upgrading message on the status bar.

9) Close the “firmware downgrading/upgrading” window and EC-Lab V10.23. The downgrading procedure is now completed and you can connect the older version to the instrument.

10) Open the older version of EC-Lab V10.12 and connect to the instrument.

Page 49: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

41

5. Connection to the cell

To connect the instrument to your experiment (the “cell”) a cell cable is used. The connection of the cell cable to the instrument varies depending on whether or not a current booster or low current option is to be used. Section 6.1 describes the various ways in which the cell cable can be connected to the instrument. Section 6.2 describes the various ways in which the cell cable can be connected to the experiment, depending on the type of system you have (e.g., battery, 3-electrode electrochemical cell, etc.) and the information you wish to obtain.

5.1 Cell cable to the instrument

5.1.1 Cell cable and potentiostat/galvanostat board association

To keep the high end specifications especially for EIS measurement, the cable and board are calibrated together, so user has to use the boards with the associated cable (same serial number). When potentiostat/galvanostat board(s) is (are) purchased with the chassis, each cable has a label indicating the channel number i.e. the number of the slot where the potentiostat/galvanostat board is inserted. If additional boards are purchased afterwards, user has to plug the appropriate cable to board. The serial number indicated on the electrometer (blue box) are the same as the serial number indicated on the label stick on the board. Note for ULC cable, the boards, the cable and the electrometer have to have the same serial number. NOTE: When the board is plugged into the chassis, it is possible to get the serial number of the board by using the calibration tool (more info on the calibration routine at the end of the manual).

Fig. 54: Serial number information via the “Channel calibration” procedure.

5.1.2 Boards description

Each board has several connectors on the front panel but each board has been designed to avoid any confusion (different types and different genders for each connector).

5.1.2.1 Potentiostat/galvanostat/ZRA board

Three connectors are available on the front panel of the potentiostat/galvanostat/ZRA board:

DB-9 connector for auxiliary analog/digital signals and to connect to the calibration board,

DB-15 connector for the connection to the calibration board,

Sub mixed DB-25 connector to be connected to the cell cable (standard cable or ULC cable).

Page 50: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

42

Note that when the channel board is running, the green LED blinks.

5.1.2.2 Booster boards

There are two kinds of booster boards. The booster boards which needn’t any additional cable to carry the current (1A/48V) and the boosters which need a cable to carry the current (2A/30V; 4A/14V; 10A/5V). The 2A/30V; 4A/14V; 10A/5V boosters can be connected in parallel to increase the maximum current. NOTE: It is possible to connect in parallel only the same boosters, not possible to mix 4A and 10 A booster for example.

Fig. 55: Booster boards.

5.1.2.2.1 1A/48V booster board

DB-15 female connector for connection to the calibration board,

DB-15 male connector for connection to the potentiostat board.

The 1A/48V booster increases the unit’s compliance and the cell polarization to +/- 48 V. This option is compatible with the Ultra-Low current option. This combination of options allows user to get a current autoranging from 1 pA to 1 A. This option can be used both with EIS (maximum value 2 MHz) and standard channel boards. Once the booster is connected to the potentiostat board, connection to the cell is the same as without the booster 1A/48V connected.

5.1.2.2.2 2A/30V; 4A/14V; 10A/5V booster board

DB-15 female connector for connection to the calibration board,

DB-15 male connector for connecting the potentiostat/galvanostat board to the booster board,

DB-9 connector for connecting the power cable P1 and P2 leads. The 2A/30V, 4A/14V and 10A/5V boosters have similar boards with one DB-15 female connector, one DB-15 male connector and one DB-9 male connector. These boosters are connected to channel board and to calibration board in the same way. The booster connection is as follows:

Connect it to the standard channel with the DB15 male – DB15 female cable

The power cable with P1 and P2 leads of the kit has to be connected to the booster board via DB9 connector.

Note that the connection to the cell is the following:

P1 lead of the standard or ultra-low current (ULC) board cable has to be connected to the P1 lead of the booster option.

P2 lead of the standard or ultra-low current (ULC) board cable has to be connected to the P2 lead of the booster option.

Page 51: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

43

Then, connection to the cell with the booster option is the same as the standard or ULC board connection.

Fig. 56: Cell connection with 2A, 4A or 10A boosters

NOTE: It is possible to use the ULC cable presented in the next paragraph. Note that if the potentiostat board is connected to 2 A, 4 A or 10 A booster and to a ULC cable, the lowest Irange is 1 µA. The lowest Irange of the ULC cable is not reachable.

5.1.3 HCV-3048 booster chassis

The connection of the HCV-3048 to the potentiostat/galvanostat board is done using the DB15 or LIO cable. When wanting to parallelize the HCV-3048 to increase the current, each chassis is connected to the other using the DB15 or CLIO cable. Also the BNC Open In in the back of the instrument should also be connected such that when the Emergency Stop button is pushed on one instrument the power of all boosters is shut.

Fig. 57: Connection of the HCV-3048 external boosetrs with the controlling pstat/gstat

5.2 Connecting the cell cable to your cell.

5.2.1 Cables description when used as standalone potentiostat or with booster boards

Two types of cable, standard and Ultra Low Current (ULC) can be connected to the DB25 connector of the potentiostat/galvanostat/ZRA boards. The difference between the two cables is their specifications. The current ranges available for the standard board is 10 nA to 500 mA (10 nA and 100 nA range are gains) and for the ULC, it is from 1 pA to 500 mA (1 pA and 10 pA are gains).

Page 52: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

44

WARNING: This probe with its cable, such as the standard cable, can only be plugged or unplugged on the channel board when the instrument is powered OFF.

The Ultra-Low Current option is composed of a cable associated with a very sensitive low current probe. This option uses no slot in the instrument chassis and the low current cable replaces the standard cable. This option includes current ranges from 100 nA down to 100 pA with additional gains extending the current ranges to 10 pA and 1 pA. The resolution on the lowest range is 76 aA. This option is compatible with another option the 1 A/48 V booster providing then current autoranging from 1 pA to 1 A.

Fig. 58: Ultra Low current cable This option can be used both with EIS (maximum value 3 MHz) and standard channel boards. The Ultra-low current option can also be used 2A, 4A and 10A boosters. The lowest Irange is 1µA. This option can be calibrated using EC-Lab® software for DC measurements. For AC measurements, and in order to keep very good specifications at high frequencies (up to 3 MHz), it is necessary to calibrate the AC part at the factory with its associated channel board. Both cables offer 6 terminals after the electrometer. Each of this lead is ended with stackable 2 mm bananas for the connection to the electrochemical cell. The ULC electrometer offers also a Guard connector (red socket of 2 mm) which allows user to protect the cell from external perturbations. A channel has 6 leads connections plus the 2 leads coming from the booster board to the electrochemical cell. Four are used in the cell control loop (2 for the current and 2 for the potential) while the 5th lead permits simultaneous recording of a supplementary cell potential. Additionally, a 6th “ground” lead is provided for cell shielding purposes or particular cell arrangements. To be easily identified, each lead has an associated color and label as follows:

S1: RED – Sense 1 for the control and measurement of the Working electrode potential.

S2: WHITE – Sense 2 for the control and measurement of the Reference electrode potential.

S3: BLUE – Sense 3 for the control and measurement of the Counter electrode potential.

P1: RED – Power 1 for the control and measurement of current flowing through the Working electrode.

P2: BLUE – Power 2 for the control and measurement of current flowing through the Counter electrode.

GND: BLACK – Ground

The internal structure of the channel board offers two working modes: grounded and floating. To switch from one mode to the other, no connection change is required. A channel has the ability to link up with 2, 3, or 4 electrodes in different configurations depending on the electrochemical cell. Ewe and Ece are measured as follows:

Page 53: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

45

Ewe = S1 - S2 Ece = S3 - S2

Notes: CAUTION Disconnection of the cell cable without switching off the instrument can damage the electrometer located on the cable (either on the standard cable or on the ultra-low current option) WARNING Voltage limits for standard and ULC cables:

- In Grounded mode: the potential between the grounded lead (black lead) and the terminals should not exceed 50 V. The potential between the grounded lead and the earth not exceed 10 V.

- In Floating mode, the potential between the ground lead (black lead) and the terminals should not exceed 50 V. The potential between the ground lead and the earth should not exceed 100 V.

Fig. 59: voltage limits in grounded and floating mode.

The current (defined in the positive direction) crosses the electrochemical cell from P1 to P2. Three typical standard configurations are explained below.

5.2.2 Standard connection

5.2.2.1 Standard three-electrode connection

In the standard three-electrode connection mode typically used in analytical electrochemistry or corrosion experiments, the working electrode is connected to S1+P1. The counter-electrode is connected to S3+P2, and the reference electrode is connected to S2.

Fig. 60: Standard three-electrode connection for a classical metal-solution interface.

Page 54: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

46

Another three-electrode connection with a reference electrode can be done, for example in battery applications. This connection allows the user to record/control the positive and negative electrodes of the battery simultaneously. For this, the following connection has to be done:

- Connection of the positive electrode (WE) to S1+P1,

- Connection of the negative electrode (CE) to S2+P2,

- Connection of the reference electrode (REF) to S3.

Fig. 61: Alternative three-electrode connection.

In the instrument, potential regulation is done between S1 and S2. Therefore, the total potential of the battery will be displayed by default. The other parameters, such as the potential of the positive and the negative electrodes versus the reference electrode, can be displayed by ticking the boxes Ece and Ewe-Ece in the “Cell Characteristics” window. In the data file, the following rows will be displayed:

- Ewe related to S1-S2 i.e. total potential of the battery, - Ece related to S3-S2 i.e. – negative electrode potential vs. Reference, - Ewe - Ece related to S1-S3 i.e. positive electrode potential vs. Reference

electrode. It is then possible to plot the change of potential (positive, negative, total) as a function of time or state of charge (SOC).

5.2.2.2 Two-electrode connection

In the two-electrode connection mode, the positive electrode of the device of interest (battery, fuel cell, supercapacitor, ….etc) is connected to S1+P1. The potential control or measurement is performed between S1 and S2, and the controlled or measured current crosses the cell from P1 to P2. So the negative electrode has to be connected to S2+S3+P2. In order to study the positive battery and negative electrode materials, the user inserts a reference electrode. Then a three-electrode assembly is required (refer to the previous part).

Fig. 62: Two-electrode connection to a battery cell.

5.2.2.3 Four-electrode connection

In the four-electrode connection mode, the user has the ability to record the liquid-liquid interface potential (Ece), also called Kelvin probe connection. In this connection mode, S1 should always be connected to WE (or to the positive electrode) for proper cell isolation. However, to avoid an IR drop in connections, it is recommended to connect S1 directly to the cell electrode and not to the P1 cable.

Fig. 63: Four-electrode connecti-on for a liquid – liquid interface.

Page 55: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

47

5.2.3 CE to Ground connection mode

This connection mode is chosen in the software “Advanced settings” window. Then the connections must be done in a special way, in connecting the ground cable to the CE electrode of the cell, as shown below:

Fig. 64: CE to Ground connection mode.

5.2.4 WE to Ground connection mode

This connection mode is chosen in the software “Advanced settings” window. Then the connections must be done in a special way, in connecting the S1 and the Ground cable to the WE of the cell and S3 and P1 to the CE electrode of the cell, as shown below:

Fig. 65: CE to Ground connection mode.

5.2.5 High voltage control connection mode

This connection mode is chosen in the software “Advanced settings” window. Then the connections must be done in a special way, in connecting the S2 and the P3 cable to the WE of the cell and S1 and P1 to the CE electrode of the cell, S2 is connected to RE. Connection is shown below: NOTE: when this mode is selected the impedance techniques are no more available.

Fig. 66: High voltage control connection mode. Left: +/- 48V for the 1A/48V booster. Right: +/- 30V for the 2A/30V booster.

Page 56: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

48

5.2.6 Floating mode

The instrument’s floating mode is selected in the advanced settings window of the EC-Lab® software. Both the standard mode and the floating mode can be used without modifying the cell connection.

Fig. 67: Floating or Grounded mode selection.

The floating mode allows the user to work with grounded cells, autoclaves, pipelines, glove boxes, etc. When the channel board is in floating mode, the red LED is not bright as it is when the channel board is in Grounded mode. Note that some specifications of the instrument can be modified using the floating mode. Using the floating mode for measurements will require taking some measurement precautions. Indeed, the cell connections should be isolated from the ground such as other apparatus that can be connected to the cell (multi-tester, voltmeter …).

5.3 Specific connection for high current boosters.

5.3.1 Connection to the cell with 1A/48V booster

The connection is similar to the connection described in the previous paragraph. The only difference is that the potentiostat/galvanostat (with/without EIS) board has to be connected to the 1A/48 V booster with the DB15 connection cable.

Fig. 68: Channel board connection to the cell with the 1A/48V booster.

P1 + S1

P2 + S3

S2

Page 57: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

49

5.3.2 Connection to the cell with 2A/30V, 4A/14V and with 10A/5V boosters

With the 2A/30V, 4 A/14 V and 10A/5V boosters, the connection with the cell cable of the potentiostat/galvanostat board remains the same, but the P1 and P2 (4 mm) leads coming from the booster board have also to be connected to the cell. The booster board needs to be connected to the channel board using the DB15 connection cable. The 2A/30V, 4A/14V and 10A/5V boosters can be connected in parallel, so it is possible to increase the current ability of the channel board by using in parallel numerous 2 A, 4 A or 10A boosters. For example, if 12 boosters of 10A are connected to one potentiostat/galvanostat board, user can reach a current of 120 A. Limit of instrument configuration:

VSP-300 chassis accepts a maximum of 4 booster boards and these boosters have to be inserted in the slots #1 to #4 (not #5 and #6).

Because of power limitation, VMP-300 chassis accepts a maximum of 12 booster boards if powered with 110Vac electrical network. They can be inserted in any slot.

Fig. 69: Channel board connection to the cell with one 2A/30V booster.

P1 booster lead

P1 + S1

P2 + S3

S2

P2 booster lead

Page 58: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

50

Fig. 70: Channel board connection to the cell with fifteen boosters.

Note: for clarity the P1 and P2 leads coming from the 10A booster cables are not displayed in the scheme but all the P1 and P2 leads of 10A boosters have to be connected to the cell trough the current collector.

5.3.3 Connection to the cell with HCV-3048 external booster

5.3.3.1 Standard connection

Two different cell cables have to be used: .The voltage cell cable also named HV-48 cable connects the pstat/gstat to the cell and measures the potential. It is only composed of two voltage senses coming from an electrometer similar to the electrometer of the ULC cable. These voltage senses are protected 4 mm banana leads.

Fig. 71: Cell cable when used with the HCV-3048 external booster with 4 mm protected banana leads

..............

..............

Current collector

Page 59: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

51

.The current to the cell is provided by the booster using two 10 mm2 highly flexible Cu-leads with silicone insulation with 6 mm safety sockets and plugs with snap-in lock at both ends that fit into the 6 mm safety flush-mounting receptacles with snap-in lock on the front panel of the HCV-3048.

Fig. 72: Connection of the pstat to the HCV-3048 boosters and cells

Fig. 73: HCV-3048 2.5 m standard power cables with 6 mm safety sockets.

WARNING It must be ensured when using boosters in parallel that the BNC cables in the back of the instruments are also connected, such that when the emergency stop is triggered, either by pushing on the front panel button or by a logic signal (TTL) coming from an external device. When using several boosters, the power cables are all connected to the same point on the stack, as well as the voltage measurement/control. As the maximum power per booster can reach 2 kW, each HCV-3048 must be connected to one separate power grid socket. Please be aware of the following limitations for the voltage measurement leads (HV-48) and the power cables or current measurement/control leads (HCV-3048).

Page 60: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

52

WARNING In this mode, the voltage on the + power cable can be equal to 100 Vdc, which is dangerous.

Page 61: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

53

6 mm safety flush-mounting receptacles with snap-in lock and M6 threaded stud (Fig. 74) are provided to connect the 6 mm standard cable to the battery or battery pack.

Fig. 74: 6 mm power socket.

5.3.3.2 Additional power cables/connector can be purchased

Additional power cables are available:

Flat contact clamps for fast and safe contact on flat conductor bars (max. 10mm thick). Highly flexible lead with Ø 6mm spring-loaded safety socket with snap-in lock, the same as in the standard power cables.

Fig. 75: Flat contact clamps.

M6-threaded pliers are for fast and safe contact of threaded bolts on motor terminal boards. Highly flexible lead silicone insulated with Ø 6mm safety socket. A maximum of 40 A can pass through the M6 thread.

Fig. 76: M6 threaded single-collet pliers.

Screw-on adapter for M8 threaded bolts. The axial connection is made by means of a Ø 6mm socket with locking device or a Ø 4mm safety plug.

Page 62: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

54

Fig. 77: Screw-on adapters for M8 threaded bolts.

Clamping contact adapter for M12 hex. screw heads. It comes together with S-SL/PF10x38 safety fuse cable. The axial connection is made by means of a Ø 6mm coupling socket with locking device.

Fig. 78: Clamping contact adapter for M12 hexagonal screw heads.

5.4 Auxiliary inputs/outputs (DB9)

The following figure shows the structure of the DB9F connector and the different pins that can be used as auxiliary inputs/outputs.

Page 63: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

55

Trigger In and Trigger Out are techniques programmed into the experiment protocol. . Trigger In can be used to start or stop an electrochemistry technique. For this, another instrument must send a trigger to the instrument. Trigger Out can be sent at the beginning or the end of the experiment. It is used to start or stop an experiment on another instrument. The Trigger signals have high and low levels: + 3.5 V < Trigger high level < + 5 V 0 V < Trigger low level < 0.8 V

Fig. 79: Structure of the DB9 connector.

Analog Out is used to control an external device (1 mA max).

Analog In1 and Analog In2 are general analog inputs used for auxiliary signal (voltage) recording. Activate Record Aux1 and Record Aux2 in the cell characteristics window to record these variables in the data file. ±10 V max, accuracy: 16 bits resolution with automatic ± 2.5 V, ± 5 V, ± 10 V ranges for Analog In1, ±10V for Analog In2, Input impedance: 1 MΩ/GND + 10 kΩ in series.

E monitor is a voltage analog output that can be used to inspect Ewe with an oscilloscope. The output signal is reversed with respect to the software measured values.

I monitor is an voltage analog output that can be used to inspect I with an oscilloscope. The current to voltage conversion depends on the Irange selected:

Current Range S/W Current Reading Std Cable I mon Output

ULC Cable I mon Output

1A ±500mA (max) ±0.5V ±0.5V

100mA ±100mA ±1V ±1V

10mA ±10mA ±1V ±1V

1mA ±1mA ±1V ±1V

100uA ±100uA ±1V ±0.9V

10uA ±10uA ±1V ±1V

1uA ±1uA ±1V ±0.9V

100nA ±100nA ±0.1V ±0.9V

10nA ±10nA ±0.01V ±0.9V

1nA ±1nA x ±0.9V

100pA ±100pA x ±1V

10pA ±10pA x ±0.1V

1pA ±1pA x ±0.01V

OPEN In is an external Trigger signal (active low) that opens relays to Open Circuit and stops the experiment performed on the cell. This may be used as an emergency stop of the experiment that is triggered by an external monitor/event.

Ground is tied to the earth.

Page 64: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

56

For convenience we have designed a 50 cm cable that connects to the DB9 connector of a channel on one end with 8 BNCs on the other end. The available signals are: E monitor, I monitor, Trigger In, Trigger Out, Analog In1, Analog In2 and Analog Out. Another cable is also available, it is the IS-1 option. This cable is described in detail in the paragraph 8.

Fig. 80: DB9-8BNC cable.

Page 65: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

57

6. Advanced features

6.1 Bandwidth

6.1.1 Bandwidth selection of a standard channel

Nine bandwidth possibilities are available for the instrument. These bandwidths represent the stability factors (SF) for a good measurement; if the bandwidth value is not well adjusted, the potentiostat will oscillate. Selection of the bandwidth well adapted to the cell is mainly done considering the response time of the studied cell and its capacity. Table 6 gives information to help the user select the bandwidth most adapted to the studied cell in the required experimental conditions. Note that the values of capacity are given for a cell purely capacitive. The time constant (TC) is calculated considering the capacity of the cell and the resistance of the current range used for the measurement, i.e. TC = RC. Cells located in the green part of the table can be studied with all the bandwidths (1 to 9). Note however that the most suitable bandwidth is underlined in the table. For the cells located in the white part of the table, fewer bandwidth choices are available, whereas for the cells located in the red part, selection of the bandwidth has to be done by the trial and error method depending on the cell’s particulars.

Table 6: Help for bandwidth selection for a standard board considering the capacity and the time constant (TC) of the studied cell, but also the experimental conditions. Considering these factors the most appropriate bandwidth (SF) is given.

1 µA/ 1 MΩ

10 µA/ 100 kΩ

100 µA/ 10 kΩ

1 mA/ 1 kΩ

10 mA/ 100 Ω

100 mA/ 10 Ω

1 A/ 1 Ω

TC SF TC SF TC SF TC SF TC SF TC SF TC SF

10 pF 10 µs 6/7 1 µs 7/8 100 ns 1-9 10 ns 1-9 1 ns 1-9 100 ps 1-9 10 ps 1-9

100 pF 100 µs 5 10 µs 6/7 1 µs 7/8 100 ns 1-9 10 ns 1-9 1 ns 1-9 100 ps 1-9

1 nF 1 ms 4 100 µs 5 10 µs 6/7 1 µs 7/8 100 ns 1-9 10 ns 1-9 1 ns 1-9

10 nF 10 ms 3 1 ms 4 100 µs 5 10 µs 6/7 1 µs 7/8 100 ns 1-9 10 ns 1-9

100 nF 100 ms 2 10 ms 3 1 ms 4 100 µs 5 10 µs 6/7 1 µs 7/8 100 ns 1-9

1 µF 1 s 1 100 ms 2 10 ms 3 1 ms 4 100 µs 5 10 µs 6/7 1 µs 7/8

10 µF 10 s 1 s 1 100 ms 2 10 ms 3 1 ms 4 100 µs 5 10 µs 6/7

100 µF 100 s 10 s 1 s 1 100 ms 2 10 ms 3 1 ms 4 100 µs 5

1 mF 1 ks 100 s 10 s 1 s 1 100 ms 2 10 ms 3 1 ms 4

10 mF 10 ks 1 ks 100 s 10 s 1 s 1 100 ms 2 10 ms 3

100 mF 100 ks 10 ks 1 ks 100 s 10 s 1 s 1 100 ms 2

1 F 1 Ms 100 ks 10 ks 1 ks 100 s 10 s 1 s 1

Example: for a cell with a capacity of 100 nF with a time constant of 1 ms, studied with a current range of 100 µA, the recommended bandwidth is the bandwidth number 4.

6.1.2 Bandwidth selection of a channel connected to ULC and/or one booster

The same kind of table can be proposed to help the user select the correct bandwidth when the channel board is connected with the booster. The selection of the bandwidth can be done as previously, considering the cell capacity and time constant of the cell.

Page 66: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

58

Table 7: Help for bandwidth selection for a standard board with a booster considering the capacity and the time constant (TC) of the studied cell, but also the experimental conditions. Considering these factors the most appropriate bandwidth (SF) is given.

1 µA/ 1 MΩ

10 µA/ 100 KΩ

100 µA/ 10 KΩ

1 mA/ 1 KΩ

10 mA/ 100 Ω

100 mA/ 10 Ω

1 A/ 1 Ω

TC SF TC SF TC SF TC SF TC SF TC SF TC SF

10 pF 10 µs 7/8 1 µs 1-9 100 ns 1-9 10 ns 1-9 1 ns 1-9 100 ps 1-9 10 ps 1-9

100 pF 100 µs 7 10 µs 7/8 1 µs 1-9 100 ns 1-9 10 ns 1-9 1 ns 1-9 100 ps 1-9

1 nF 1 ms 6 100 µs 7 10 µs 7/8 1 µs 1-9 100 ns 1-9 10 ns 1-9 1 ns 1-9

10 nF 10 ms 5 1 ms 6 100 µs 7 10 µs 7/8 1 µs 1-9 100 ns 1-9 10 ns 1-9

100 nF 100 ms 4 10 ms 5 1 ms 6 100 µs 7 10 µs 7/8 1 µs 1-9 100 ns 1-9

1 µF 1 s 3 100 ms 4 10 ms 5 1 ms 6 100 µs 7 10 µs 7/8 1 µs 1-9

10 µF 10 s 2 1 s 3 100 ms 4 10 ms 5 1 ms 6 100 µs 7 10 µs 7/8

100 µF 100 s 1 10 s 2 1 s 3 100 ms 4 10 ms 5 1 ms 6 100 µs 7

1 mF 1 ks 100 s 1 10 s 2 1 s 3 100 ms 4 10 ms 5 1 ms 6

10 mF 10 ks 1 ks 100 1 10 s 2 1 s 3 100 ms 4 10 ms 5

100 mF 100 ks 10 ks 1 ks 100 s 1 10 s 2 1 s 3 100 ms 4

1 F 1 Ms 100 ks 10 ks 1 ks 100 s 1 10 s 2 1 s 3

10 F 10 Ms 1 Ms 100 ks 10 ks 1 ks 100 s 1 10 s 2

100 F 100 Ms 10 Ms 1 Ms 100 ks 10 ks 1 ks 100 s 1

6.2 Floating mode

By default, the instrument is configured in grounded mode. Changing this configuration can be done in the Advanced Settings window of EC-Lab® software. Both the standard and floating mode can be used without modifying the cell’s connection.

Fig. 81: Floating/Grounded mode selection frame.

Note: In EC-Lab, when the channel board is in floating mode, the red LED is not bright as it is when the channel board is in Grounded mode.

6.3 Filtering

In some experimental conditions, it would be useful to use a filter, then three possibilities of current and potential filter are available with the instrument. The filter configuration can be done in the Advanced settings window of EC-Lab® software. Note the two gains are only available if the one of the filter is activated.

Fig. 82: filters selection frame.

Page 67: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

59

6.4 External device control and recording

6.4.1 General description

The EC-Lab software offers the user the ability to control external devices such as rotating electrodes and thermostatic baths and to record external analog signals through the auxiliary DB9 connector. The user has to configure the analog output to control an external device and configure the Analog In1 and Analog In2 inputs to record external signals. Our instruments can control and record analog signals from – 10 to + 10 V. Most external devices work in a 0 to + 5 V range. The figure below shows the external device window where the user sets parameters. Many instruments are already configured in the software to be controlled by our

potentiostat. This list continues to be expanded in new versions of EC-Lab software. To configure external devices select “External Device” frame.

Fig. 83: “External Devices” frame.

The user must define several parameters to configure the external device to either be controlled via the analog output (left column) or record/measure data via analog input 1 and 2 (right column). The procedure for the configuration of the auxiliary inputs/outputs is described is as follows:

1- Choose the channel to configure. Each channel can be configured for a specific device. One channel can control one device and the other one another device.

2- Select the Device Type in the list between None, Thermostat, RDE, QCM and other. According to the selected device type one or several device names are available.

3- Among the available devices some can be controlled by the analog output and some of them can only be used to record values with analog inputs 1 and 2. The user must tick the box to activate the input/output.

4- In the activated frame, the user must define the conversion between the input voltage and the variable to plot. This is a direct linear conversion in the range defined by the user between the min and the max value.

5- The user can also define the name and the unit of the variable they want to display. Click on “Custom Variables”. The figure below is displayed:

Page 68: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

60

Fig. 84: Custom Units window to define new variables.

To create a new variable with its unit, click on “Add” and put the name and the unit of the new

variable in the frame. Then click on to validate. The new variable is displayed in the list in blue (as a custom variable) and can now be selected as the recorded variable for the analog inputs.

6- Finally click on “OK” to configure the selected channel to record the auxiliary input signal

The new selected variables for Analog In1 and Analog In2 are automatically displayed on the “Cell characteristics” window and activated for recording. In the “Selector” the created variables are displayed and can be plotted. These auxiliary variables can be used in several protocols as conditional limits of an experiment. Note: - The parameters set in Analog In1 and Analog In2 to define the linear slope can be

inverted to display an inverse relationship between the recorded value and the plotted value. - The configuration of external devices that can be controlled by the potentiostat (analog output) are described in detail in the corresponding sections of the manual.

- A manual control of external devices is also available on the left of the panel. - When a channel has been configured to control an external device, this device can be seen in the global view.

Page 69: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

61

6.4.2 Rotating electrodes control

The Bio-Logic instrument can control a rotating electrode such as a ALS-RRDE-3A Rotating Disk electrode model with the auxiliary input/output. A specific control panel has been designed to control the rotation speed. Note that no measurement of the actual rotation speed is available with the ALS-RRDE-3A model.

Fig. 85: ALS-RRDE-3A Rotating Disk electrode

6.4.2.1 Connections

Two cables are necessary for connection of a rotating disc electrode, the cell cable and a special cable for RRDE with a DB-9 connector. This cable has a DB9 connector on one end and three wires named Analog Out, Trigger Out and 1 Ground on the other end (PN: 092-22/11). The connection procedure is as follows:

Connect the DB9 cable to the front of the channel board (Auxiliary Input/Output).

Connect the “Analog Out” wire to the “IN” plug on the rear panel of the RRDE-3A.

Connect the “Ground” wire to the “GND” plug on the rear panel of the RRDE-3A.

To control the purge of the RRDE-3A, connect the “Trigger Out” to the “Purge” plug on the rear panel of the RRDE-3A.

Connect the REF1 and CA2 wires to the disk brush, REF3 and CA1 wires to the counter electrode, and REF2 wire to the reference electrode. For RRDE experiment, “CE to Ground” mode must be selected. In that case, use the appropriate connections for both the ring and the disc channels.

Page 70: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

62

Fig. 86: ALS-RRDE-3A Rotating Disk electrode connection to the cell cable.

Note: This menu is available only if channels designed to drive a RDE are currently connected to the RDE. If so, click on “External Devices” and the following window will be displayed:

Fig. 87: Rotating electrode control configuration.

Under “Device Name” one can select the RDE type. Three standard rotators are included in the list; ALS-RRDE-3A, PINE RRDE or RADIOMETER CTV101. For these devices, the

Page 71: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

63

calibration parameters are factory set. Other external systems can be used but are not pre-configured in EC-Lab. If you would like another device to be pre-configured, please contact Bio-Logic and it will be added onto the device list upon request. Note that the calibration parameters for the pre-configured devices cannot be modified by the user .If you would like to configure a device not in the list, select “Other” in the Device Name and configure the “Analog OUT” window to define the control parameters. Click on the “Apply” button to validate the settings. Note that this menu can be activated without any rotating electrode unit, but it will only have effects on electrochemical instruments equipped with a rotating system.

6.4.3 Temperature control

Temperature control is possible with the auxiliary output voltage of the instrument. Several thermostats have already been configured such as the Julabo series and the Haake Phoenix series.

Fig. 88: Haake Phoenix series thermostat control configuration.

The user can configure other thermostats to either record temperatures (Analog In) or to both control (Analog Out) and record (Analog In) temperature.

6.4.4 Electrochemical Quartz Crystal Microbalance coupling

The SEIKO EG&G QCM 922 quartz crystal microbalance can be coupled with our potentiostat / galvanostat to record both the frequency variation and the resistance variation. The configuration for the EQCM coupling is described in the following figure:

Page 72: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

64

Fig. 89: SEIKO EG&G QCM-922 configuration window.

One can see that both frequency and resistance variations are recorded on the potentiostat analog inputs. The user has to define both the frequency range and the resistance range. The results of this experiment are displayed to the right. A process is also available to calculate the amount of a species electro-disposed onto the QCM. To use this process, select the process data option in the Analysis menu.

Fig. 90: Frequency and resistance variations recorded from the analog inputs for an instrument coupled with a SEIKO EG&G QCM 922.

Page 73: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

65

6.5 Virtual potentiostat

The user can work with the EC-Lab software without being connected to an instrument. In that case, the software sees a virtual potentiostat that is not available for experiments but can be used as a user’s interface. The user can select his virtual instrument in the “Devices” frame,

click on the “Virtual potentiostat” button then, the corresponding software interface will be displayed.

Fig. 91: Virtual potentiostat button (in the red rectangle).

The available techniques and time base are different depending on the instrument selected.

Fig. 92: Virtual potentiostat window.

Page 74: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

66

7. Options & Accessories

The options and accessories of the VMP300-based instruments are described in this section.

7.1 Analog Ramp Generator (ARG)

This option is an electronic board which has to be plugged onto the potentiostat/galvanostat board. It allows users to apply a true linear voltage scan (analog and not digital) to the cell. The analog nature of the scan is critical in several applications such as catalyst study for fuel cell development or for voltammetry requiring fast scan rate (to detect intermediate species with short lifetime). As the linear scan generator allows fast scan investigation, hardware ohmic drop compensation is also implemented in the software. This correction is done continuously by the hardware which is much faster than the software compensation. For example, we can compare the standard and linear CV. Due to the feedback loop, the software compensation with the standard CV takes around 100 µs while the hardware

compensation is about 100 times faster. This option is compatible with ULC cable and booster boards.

Fig. 93: ARG option.

ARG specifications:

Scan ranges 1 MV/s, 10 kV/s, 100 V/s, 1 V/s.

Scan resolution 0.0015% FSR* (down to 15 μV/s).

Minimum voltage scan rate

0.1 mV/s

Sampling rate down to 1 µs

Buffer size for Fast mode

40 000 data points

Voltage range ± 10 V. The voltage range can be adjusted.

Accuracy < ±0.1% of range

Number of cycles 1 to 65535

7.2 Isolation System (IS1)

It is used to keep a potentiostat totally floating even when it is connected to a grounded external device via the auxiliary cable. The auxiliary cable is connected to the potentiostat board using the DB9 (or SubD 9) connector. Thanks to this cable, an external device can communicate with the potentiostat either by sending/receiving digital triggers or sending/receiving analog signals. When the auxiliary cable is connected to an external device, the potentiostat board becomes grounded to the internal ground of the external device. Consequently, if the channel was initially used in the floating mode, connecting it to the external device using the auxiliary cable makes it lose its floating character.

Page 75: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

67

With the IS1, even if the channel is connected to an external device, it keeps its floating character. The IS1 is powered by the potentiostat board using the LIO cable connected to the DB15 (or SubD 15) connector on the potentiostat or by an external 12V DC power supply. The auxiliary cable from the potentiostat board is connected to the DB9 connector on the IS1. The external device is connected to the IS1 thanks to a simple BNC cable. All the available I/O were separated in the same manner as the DB9 to BNC adaptor.

Fig. 94: IS1 option.

IS1 specifications:

Analog I/O Digital I/O

Input (IN1, IN2) : Input (Trigger In, Open In):

Input impedance: >1011Ω / 15 pF Input impedance 1 MΩ / 150 pF

Input voltage range: +/- 10 V Input voltage range 0 to 5 V

Bandwidth 50 kHz Input frequency max 400 kHz

Signal noise: 10 mVpp (peak to peak)

Low/high input threshold

[0; 0.8]V / [2 ; 5] V

Maximum propagation time/Minimum pulse width

1 µs

Output (Out , I monitor, E monitor) : Output (Trigger Out):

Output impedance 100 Ω Output impedance 100 Ω

Signal noise 10 mVpp Low/high input threshold

[0; 0.4] V / [4.6; 5] V

Output current max 10 mA Maximum output frequency:

400 kHz

Maximum propagation time/Minimum pulse width

1 µs

Page 76: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

68

7.3 Current Collector (CC5)

Up to five current boosters can be connected in parallel, increasing the maximum current that can be passed through the cell. In order to simplify the connections, a current collector was designed (Fig. 95). A maximum current of 60 A can flow through the CC5 (Current Collector for 5 boosters). This simple device collects the current coming from each booster in parallel and delivers it through two 60 cm power cables terminated by two sturdy crocodile copper clips, similar to automobile battery clips. The scheme below (Fig. 96) shows how to connect in parallel five boosters inserted in the VSP-300 chassis. All the boosters need to be connected to each other using the DB 15 connectors.

Fig. 95: CC5 option.

On each booster the power cables are connected to the CC5 as it is shown in Fig. 99 below. The crocodile clips of the CC5 are then clipped to the battery or the cell of interest. The total current will flow through these cables. The connection leads from the potentiostat board coming from the DB 25 connector, are connected directly to the battery.

Fig. 96: Connection of horizontal VSP-300 and CC5 to a battery

Page 77: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

69

Fig. 97: Connection of vertical VSP-300 and CC5 to a battery.

NOTE: If more than 5 boosters are needed, it is possible to use several CC5. For 2A and 10 A boosters, only 4 boosters can be used, in parallel or independently.

7.4 MP-MEA option

MP-MEA option is especially dedicated to applications using micro electrode array (MEA). Up to 256 electrodes can be handled with this multiplexer, the switching time is 8 ms between each electrode.

Fig. 98: MP-MEA option

Page 78: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

70

7.4.1 Description

At the top of the MP-MEA:

- The connector for the MEA (in the middle). To fit with most MEA, two types of adapters are available (connection from the top or from below)

- Three 2 mm-connectors (in the middle above the MEA connector). One to connect an external reference electrode, one to connect an external counter electrode and one to connect to ground of the instrument.

- Two LEDs. The LED “power” indicates if the MP-MEA is powered on and the “status” LED indicates experiments is running.

At the right side:

- Connector for the Ultra Low current probe. P1, S1, S2, S3, P2 and GND 2mm-connectors of the Ultra Low current probe have to be connected to the appropriate connectors.

The Ultra low current probe to cable and DB-15 connector to connect MP-MEA to the board are available at the rear panel of the MP-MEA.

Two cables are needed to connect MP-MEA to instrument board. The DB-15 connectors of the instrument and of the one of the MP-MEA module has to be connected with DB-15/DB-15 cable AND the ultra low current probe connected to the DB-25 of the board has to be connected to the dedicated plug.

7.5 Test boxes for calibration check and user training

These accessories can be used to check the instrument or to help user optimize their setup..

7.5.1 Dummy Cell 2 (DC2)

The Dummy Cell 2 is provided with each channel board. The circuitry is made of resistors (accuracy 1%) and capacitors.

On the left side there are two lugs, on the right

side two holes. This allows several DC-2s to be

bound together.

The dummy cells for boosters and DC2s can be

bound together as well.

Page 79: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

71

Fig. 99: Dummy Cell-2

7.5.2 Dummy cell for booster

The Dummy cell for booster is specifically dedicated to periodically check a booster. The dummy cell for booster is provided with each booster chassis.

Dummy cell for booster and DC2 can be bound together

as well.

Characteristics of the dummy cell are given in the following table:

Resistance

Tolerance Temp. coeff.

Current Max

Power Max

5 mΩ 1% +/- 50 ppm/°C

20 A 7 W

Fig. 100: Dummy Cell for Booster.

7.5.3 Test Box 2

The Test Box 2 is a tool specifically designed for checking the calibration of the standard channel boards of our instruments. This test box is made of one electrical circuit with high precision resistors for calibration check and two dummy cell circuits for user training. The high precision resistor circuit is made with 7 resistors, one to check each current range of the board.

Fig. 101: Test Box-2.

High precision resistor characteristics:

High precision resistor I range

10 Ohms +/- 0.04% 1 A

100 Ohms +/- 0.02% 100 mA

1 kOhms +/- 0.02% 10 mA

10 kOhms +/- 0.02% 1 mA

100 kOhms +/- 0.02% 100 µA

1 MOhms +/- 0.02% 10µA

10 MOhms +/- 0.04% 1 µA

Page 80: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

72

7.5.4 Test Box 3

Test Box 3 is a tool for teaching and practice on linear and non-linear electrochemical systems. This testing box is made of three electrical circuits simulating real electrochemical systems. Along with application notes #8 and #9 Test Box 3 can be used to highlight some commonly overlooked points about electrochemical impedance spectroscopy. Additionally, with Test Box 3 some general electrochemistry protocols like Cyclic Voltammetry and, corrosion techniques such as Linear Polarization or Generalized Corrosion can also be practiced.

Fig. 102: Test Box-3

#1

#2

#3

Fig. 103: stationary curve obtained with Test Box-3.

7.6 Temperature probe

To allow the user to measure temperature, temperature probe (PN: 092-22/13) can be connected to the auxiliary input/output of the potentiostat board. The temperature probe is fully controlled in EC-Lab® software.

Fig. 104: Temperature probe.

7.7 Battery holders

7.7.1 Battery Holder: BH-1

This accessory is designed to make battery testing easier. The holder is modular and compatible with a wide variety of battery types (coin cells, 18650 and 26650 cells). The BH-1 can accommodate up to 4 batteries in pairs of two. Perfect contact of the BH-1 connectors with the battery poles is ensured with spring-loaded connection points. Each of the batteries is secured in place by isolating separators

Fig. 105: Battery holder, BH-1.

Page 81: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

73

specifically designed for each battery type.

It is possible to test many batteries simultaneously by linking multiple holders in series. NOTE: Before starting a measurement check if the contact between the battery and the connector of the battery holder is efficient (stable voltage).

7.8 Labview VIs

The potentiostat can be controlled with the free EC-Lab® development package. This package is devoted to software developers who need to integrate the control of the Bio-Logic potentiostats/galvanostats/EIS with OEM software. The EC-Lab® development package includes a LLB LabVIEW® library where the functions of the DLL are implemented (Requirements: LabVIEW® V8.5 or higher). This development package includes a DLL with specific functions for:

- connection/disconnection to a selected instrument (Ethernet or USB), - initialization of the channels by loading the firmware, - loading protocols on the channels:

- all techniques available in EC-Lab® Express - starting/stopping the selected channel(s), - retrieving data. - Note that the time base with these LabVIEW® Vis depends on the selected protocol.

7.9 Electrochemistry accessories

Bio-Logic can provide accessories for many fields of electrochemistry and battery chemistry. Please consult the Bio-Logic website for more details: http://www.bio-logic.info/electrochemistry-ec-lab/accessories/

Page 82: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

74

8. Validation and calibration

8.1 Channel validation

The channel board is the potentiostat/galvanostat board alone (not connected to any Low Current option or current boosters) Actually, there are two procedures depending on the board option:

The first one can be done for every channel board,

The second one can only be done for boards with the EIS option For both procedures, the user needs a dummy cell (DC2) shown in Fig. 106. This is the dummy cell provided with each channel board. The cell cable has to be connected as shown in the Fig. 106.

Fig. 106: Front panel of a channel board of VMP300-based instruments.

Fig. 107: Connection of the cell cable to the Dummy cell (DC2).

Note: The two blue leads have to be stacked together and the two red leads have to be stacked together. CAUTION It is noteworthy that to avoid any external perturbation, the dummy cell has to be placed in a Faraday cage. This Faraday cage has to be connected to the chassis of the instrument (green plug on the rear panel).

Page 83: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

75

8.1.1 Channel board without EIS option

The procedure is the following: - Load the CV technique with the default settings on the channel of interest. The screenshot of the settings is shown in Fig. 108. NOTE: For longer cable the bandwidth parameters may be reduced [3].

- Start the measurement.

- Select the whole resulting curve - Fit the resulting data with the “Linear

Fit” tool available in the “Analysis” menu (or by clicking on “Shift+F2”).

Fig. 108: Default settings of the CV technique.

TIP: To get directly the value in Ohm ([Volt]/[Amperes]), you should shift the I vs. Ewe representation (default representation for CV) to Ewe vs. I representation. This shift can be easily done thanks to the representations tool available at the top of the Graphic frame (Fig. 109).

Fig. 109: Ewe vs. I representation.

Page 84: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

76

- Check the fitted values, the slope has to be close to the theoretical values which is 5 069 Ω ±1% for the Dummy Cell DC2 and the offset has to be around zero +/- 1 mV (Fig. 110).

Fig. 110: Result of the Linear fit.

8.1.2 Channel board with EIS option

The procedure is the following:

- Load the PEIS technique with the default settings on the channel of interest (Fig. 111).

Fig. 111: Default settings of PEIS technique.

NOTE: For longer cable the bandwidth parameters may be reduced.

- Start the measurement. - Select the whole resulting curve - Fit the resulting data with “ZFit” available in the “Analysis” menu (or by clicking on “F6”). - Click on the “Edit” button and select the R1 + C2/R2 + C3/R3 (Fig. 115).

Page 85: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

77

Fig. 112: Equivalent circuit selection.

- Click on the “Minimize” button (Fig. 113).

Fig. 113: Results of Z Fit.

- Check the fitted values. They have to be close to the theoretical values. The theoretical values for DC2 are summarized in Tab. 8.

-

Table 8: The theoretical values of the DC2.

Element Value

R1 499 Ω ±1% R2 or R3 1 k Ω ±1% NOTE:

The values R2/C2 and R3/C3 may be inverted, respectively.

Page 86: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

78

8.1.3 Validation of the ultra-low current module.

The procedure is exactly the same. The only difference is that the cell cable connected to the dummy cell will be the ultra-low current cell cable instead of the standard cell cable. Moreover, the user can check if the instrument goes through the current ranges of ULC (1A to 1pA) option. To check this, the user has to decrease the scan rate and check the current range indicated in the status bar. For the current range 10pA and 1pA an analog filter (50 kHz,1 kHz or 5 Hz) should be activated.

8.1.4 Validation of the booster board

For the booster board validation, the procedure is similar to the procedure described in II.1. Because of the involved current, the dummy cell, DC2, has to be exchanged by the dummy cell for booster (Fig.114).

Fig. 114: Dummy cell for boosters.

The voltage sweep, i.e. E1; E2 and Erange parameters of the CV technique, has to be set according to the specifications of the dummy cell and the booster. Some examples are given in Tab. 9. For your information, the specifications of the dummy cell for booster are:

5 mΩ (Accuracy: 1%)

Pmax = 7 W

Imax = 20 A.

Table 9: The E1, E2 and Erange values according to the booster type.

Booster board type

E1 = -E2 Erange

2A 10 mV +/-10 mV 4A 20 mV +/-20 mV 5A 25 mV +/- 25 mV 8A 40 mV +/- 40mV

10A 50 mV +/- 50 mV 20 A 100 mV +/- 100

mV

Moreover, the user can check if the instrument goes through the current ranges of the booster. To check this, the user has to check the current range indicated in the status bar.

Page 87: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

79

8.2 Channel calibration with EC-Lab® software

8.2.1 Potentiostat/galvanostat board calibration

To perform the calibration the user has to follow the procedure described in detail hereafter. The calibration procedure lasts a few minutes.

1) In EC-Lab software Select "Tools/channel calibration". You will receive the following message.

2) Answer "Yes" to the

warning message. Then the board calibration window appears.

3) Select the channel you want to calibrate with the drop-down box

4) Select the type of the calibration you want to run with the drop-down box: three different calibrations are offered i.e. only the “Channel”, the “Channel & Amplifier”, or only the

“Amplifier” . If you want to calibrate: - the channel board alone, select “Channel” - the booster and the channel board, select “Channel & Amplifier” - the booster alone, select “Amplifier”

5) Click on the “Calibrate” button and this warning message appears which indicates that

if you start the calibration all the settings of the channel selected will be reset.

Page 88: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

80

6) The calibration program (“ecal4.bin”) is loaded on the channel.

7) Specific connections have to be done between the calibration board and the channel and/or amplifier board(s) depending on the selected calibration. A wizard helps the user to use the appropriate connection. NOTE: Hereafter, the “Channel & Amplifier” is selected. If user selects “Channel” (steps #8 to #16), only the first part will be done and if user selects “Amplifier” (steps #17 to #22), only the second part will be done.

8) Click on the “Next”

button.

Page 89: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

81

9) Connect the cell cable of the channel board to the calibration board and click on the “Next” button.

10) Connect the two

DB15 connectors using DB15-BB15 cable and click on the “Next” button.

Page 90: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

82

11) Connect the two DB9 connectors using DB9-DB9 cable and click on the “Next” button.

If the DB15-DB15 is not correctly connected a warning message appears.

12) Then the calibration procedure starts.

Page 91: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

83

13) If all the items pass, the value of the calibration is saved in the memory of the board and/or in the ULC electrometer. Click on ”OK” button to complete the calibration process.

14) If one or more items failed, a warning

message indicating that the value of the calibration will not be saved and the calibration process is not completed. Before contacting your local representative, please check again if all the cables were properly connected.

15) If you selected “Channel” calibration the calibration procedure is finished. If not, the calibration procedure continues with the calibration of the amplifier board. As this part calibrates only the amplifier board, so the amplifier board has to be connected as well. If user select “Amplifier” calibration”, the calibration procedure starts from this step.

Page 92: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

84

8.2.2 Booster board calibration

16) Select the appropriate amplifier with the drop-down box and click on the “Next” button. Hereafter, only the calibration procedure of the 4A amplifier is described but the procedure is the same for the other boosters.

17) Connect the cell cable and the two DB15 connectors between amplifier and the calibration boards as described below and click on the “Next” button.

Page 93: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

85

18) Connect the two DB15 connectors of the channel and the amplifier boards as described below and click on the “Next” button.

19) Connect the P1/P2 connector coming from the amplifier board to the P1/P2 of the cell cable coming from the channel (two 4mm to 2mm adapters are provided with the amplifier board) and click on the “Next” button.

20) If all the items pass, the

value of the calibration is saved in the memory of the board and/or in the ULC electrometer. Click on the ”OK” button to complete the calibration process.

Page 94: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

86

21) If one or more items failed, a warning message indicating that the value of the

calibration will not be saved and the calibration process is not completed. Before contacting your local representative, please check again if all the cables were properly connected.

Page 95: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

87

22) Standard programs are reloaded to the channel to be ready to be used.

8.2.3 HCV-3048 booster

CAUTION The HCV-3048 (as well as all our instruments) is calibrated at the factory with high precision instruments. Both the voltage measurement and control and the current measurement and control are calibrated with a specific channel board. If the user performs the calibration using the calibration board, the accuracy that the instrument had when leaving the factory will be lost, because the calibration board does not have the highly precise components needed to calibrate over the whole 48 V range that is covered by the HCV-3048. Nevertheless, if the instrument has been used for years without calibration, then it can be useful to perform an user’s calibation. Of course, to have the factory accuracy, the best is to send it back to the factory. The other consequence is that if the user uses a different channel board than the one used when the calibration was performed, the factory calibration will also be lost. The factory and user accuracies are given in 10.2.7. For the user’s calibration please follow the following step:

Page 96: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

88

Note: the blue lead of the HV-48 must be connected to the white S2 connector on the calibration board and not on the blue S3 connector.

Page 97: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

89

Page 98: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

90

The next steps are similar to the in-built boosters.

Page 99: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

91

9. Maintenance

WARNING Before performing any maintenance power down the unit then disconnect the power cord and all test cables.

9.1 Calibration

9.1.1 SP-200/SP-240/SP-300/VSP-300/VMP-300

Our instruments do not require specific maintenance. Each channel board is calibrated at the factory before it is delivered to the customer. Due to the temperature differences between winter and summer, we recommend adjustment of the gains and offsets of the channel boards twice a year, especially if the instrument is not in an air-conditioned room. This adjustment is

performed using the EC-Lab software “Channel Calibration” in the “Tools” menu. We also recommend a full check-up of the instrument (at the factory) every two years.

9.1.2 HCV-3048 external booster

If the HCV-3048 is purchased with a potentiostat or with a channel board, it will be calibrated at the factory and retain the better accuracy values specified in 10.2.7. If the external booster is purchased separately from the controlling channel board then there are two options: . the user accepts to have the accuracy given by performing a calibration using the calibration board and specified in 10.2.7. . the user can send the controlling potentiostat board to the factory to allow a calibration with the HCV-3048, giving the factory accuracy. In both cases, we suggest a factory calibration of the HCV-3048 every year. In which case it has to be returned with the controlling potentiostat channel.

9.2 Ventilation

WARNING The user must carefully check that the ventilation grids are not obstructed on the right and left sides and under the chassis. An external cleaning can be made with a vacuum cleaner if necessary. Please see 2.1.3 for more detailed information.

9.3 Cleaning

9.3.1 Ventilation grids

External cleaning can be made with a vacuum cleaner if necessary. Use a damp cloth or mild, water-based cleaner to clean the instrument. Clean the exterior of the box only, never the circuit board. WARNING Do not apply cleaner directly to the box or allow liquids to enter or spill on the box.

9.3.2 Air filters cleaning

Filters are placed underneath the instrument to filter the air and retain the dust that may come with the inlet air flow that is cooling the power supply (and not the heat sinks). It is recommended to clean periodically the filter. This filter is located on the underside of the instrument and is outlined in red (Fig. 115)

Page 100: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

92

Fig. 115: Bottom side of the VMP-300 with the air filter outlined in red.

Fig. 116: Bottom side of the vertical VSP-300 with the air filter outlined in red.

Fig. 117: Bottom side of the HCV-3048 with the air filter outlined in red.

9.3.3 Fuses:

WARNING To maintain protection from electric shock and fire, replace fuses, with the same rating and type (described in the Technical Specifications section just below).

Page 101: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

93

On the HCV-3048, the fuses are located on the rear of the instrument on the left of the power socket.

Fuse location on the VMP-300 family instruments.

Page 102: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

94

10. Technical Specifications

10.1 Equipment Ratings

10.1.1 Electrical & Mechanical specifications

Table 10: Instrument electrical ratings.

Instrument type

Input Output Mechanical

Voltage Range/

Vac

Frequency

range/Hz

Power max/

W Fusesa Voltage Range/Current

Size (HxWxD)/mm

Weight without cable/kg

SP-200 [90;264] [50; 60] 350 2 x 4 AF 10 Vdc / 500 mA 221x170x384 7.2

SP-240 [90;264] [50; 60] 350 2 x 4 AF 10 Vdc / 500 mA or 14 Vdc / 4A (with 4A/14V booster)

225x205x410 7.5

SP-300 [90;264] [50; 60] 350 2 x 4 AF 10 Vdc / 500 mA or 48 Vdc/1A (with 1A/48V booster) or 30 Vdc/2A (with 2A/30V booster) or 14 Vdc/4A (with 4A/14V booster) or 5 Vdc/10A (with 10A/5V booster)

225x205x410 7.5

VSP-300 [90;264] [50; 60] 650 2 x 5 AF 260x530x510 22

VSP-300 (Vertical)

[90;264] [50; 60] 650 2 x 5 AF 337x254x517 20

VMP-300 [90;264] [50; 60] 1500 2 x

10 AF 533x315x563 30

HCV-3048 [85-264] [47-440] [1700-2000]*

2 x 16 AF [0;48] Vdc / ±30 A 135x400x430 26

* See power contour plot of HCV-3048

10.1.2 Environmental

- Indoor use - Operating Temperature: 10°C to +40°C Indoor use - Storage Temperature: 0°C to +50°C - Pollution degree: 1 (no pollution or only dry) - Altitude: <2000 m above sea level - Humidity: 10% to 80% non-condensing - Case protection: IP20 - Warm-up: 1 hour to rated accuracy - Cooling: Internal DC Fans - Vibration: not specified - Choke: not specified - Noise (audible) : <60 dBa at full power

Safety complies with EN61010-1. EMC complies with EN61326. RoHS complies with 2002/95/EC

10.2 Channel specifications

10.2.1 Channel board

General Functions

Potentiostat Yes

Galvanostat Yes

Impedance analyzer Yes (option)

Coulometer Yes

Linear scan generator Yes (option)

Page 103: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

95

Floating mode Yes

IR compensation Yes

Analog filtering Yes

External input/outputs Yes

Cell connection 2, 3, 4 or 5 terminal leads (+ ground)

Control amplifier

Compliance ± 12 V

Maximum current ± 500 mA continuous

Gain-Bandwidth compensation 9 programmable stability factors

Highest unity gain bandwidth 1.4 MHz

Slew rate (no load) > 20 V/µs

Rise/Fall time (no load) < 500 ns

Voltage control

Ranges Adjustable from ± 10 V down to ± 30 mV (1 mV resolution)

DC level shift ± 10 V, 300 µV resolution. For example: For an Erange of [0;20]V, the resolution is 333.33 µV. For [0;19.66]V, the resolution is 300 µV. For [0;13.1]V, the resolution is 200 µV. For [0; 6.55]V, the resolution is 100 µV. For [0; 3.27]V, the resolution is 50 µV. For [0; 1.31]V, the resolution is 20 µV. For [0; 0.65]V, the resolution is 10 µV. For [0; 0.32]V, the resolution is 5 µV. For [0; 0.13]V, the resolution is 2 µV. For [0; 0.06]V, the resolution is 1 µV.

Accuracy < ± 1 mV ± 0.03% of setting

Lowest resolution 1 µV

Current control

Ranges ± 1 A, ± 100 mA, ± 10 mA, ± 1 mA, ± 100 µA, ± 10 µA, ± 1 µA (7 ranges + Autorange)

Additional ranges ± 100 nA, 10 nA with gain

Accuracy < ± 0.1 % of range ± 0.03 % of setting

Resolution 0.0033 % of range

Voltage measurement

Ranges ± 10 V, ± 5 V, ± 2.5 V, ± 250 mV, ± 25 mV

DC level shift ± 10 V, 300 µV resolution

Accuracy (DC) < ± 1 mV, ± 0.03 % of reading

Maximum resolution < 0.0033 % of range

Bandwidth (-3 dB) 8 MHz

Filtering low-pass 4 poles Sallen-Key filters

50 kHz, 1 kHz and 5 Hz

Acquisition speed 1,000,000 samples/s Current measurement

Ranges ± 1 A, ± 100 mA, ± 10 mA, ± 1 mA, ± 100 µA, ± 10 µA, ± 1 µA

Additional ranges ± 100 nA, ± 10 nA

Accuracy (DC) < ± 0.1 % of range ± 0.03 % of reading

Maximum resolution 0.0033 % of range

Bandwidth (-3 dB) 8 MHz

Filtering low-pass 4 poles Sallen-Key filters

50 kHz, 1 kHz and 5 Hz

Acquisition speed 1,000,000 samples/s Electrometer

Page 104: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

96

Input impedance 1 TΩ ΙΙ 25 pF typical

Input Bias current < 10 pA

Bandwidth (-3 dB) 8 MHz

Common mode rejection ratio > 60 dB at 100 kHz

Ground to chassis impedance

Floating mode 1 MΩ ΙΙ 25 pF typical

Grounded mode < 100 Ω

IR compensation

Resistance determination EIS

Compensation mode Hardware or software positive feedback

Compensation range Programmable from 0 to 100 % of the current range resistor

Auxiliary Inputs/outputs

External Input Can be used to apply an external waveform directly to the control amplifier

2 Analog Inputs Automatic ± 2.5 V, ± 5 V, ± 10 V ranges - 16 bits resolution

1 Analog Output 16 bits resolution on the ± 10 V range

2 Digital Inputs TTL level Trigger input and Open input

1 Digital Output TTL level Trigger output

2 Monitor Outputs Cell current and compensated working electrode potential

Impedance specifications (EIS option only)

Frequency range 10 µHz to 7 MHz

Frequency resolution < 10 ppm of the setting

Sinus amplitude 0.5 mV to 5 V with 76 µV resolution 0.1 % to 100 % of the current range with resolution of 0.004 % of the range

Accuracy See contour plot

Mode Single sine, Multisine, FFT analysis

Contour plot

10.2.2 Ultra Low Current

Cell Control Maximum Current ± 500 mA (± 1 A continuous with 1 A/48 V)

Page 105: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

97

Maximum Current resolution

0.004 % of the range (down to 76 aA on the ± 1 pA)

Applied Current Accuracy

< ±0.1% of range ±0.03% of setting for ±500 mA to ±100 nA ranges < ±0.1% of range ±1% of setting for ±10 nA range to ±1 nA ranges < ±0.2% of range ±2% of setting for ±100 pA range

Current Measurement Ranges ± 100 pA, ± 1 nA, ± 10 nA, ± 100 nA

Additional Ranges with gain

± 1 pA, ± 10 pA

Maximum Resolution 0.004 % of the range

Accuracy < ±0.1% of range ±0.03% of setting for ±500 mA to ±100 nA ranges < ±0.1% of range ±1% of setting for ±10 nA range to ±1 nA ranges < ±0.2% of range ±2% of setting for ±100 pA range < ±1% of range ±2% of setting for ±10 pA range < ±10% of range ±2% of setting for ±1 pA range

Impedance specifications (EIS option only)

Accuracy

10.2.3 1A/48V booster

Control amplifier

Compliance voltage ± 49V

Compliance current ± 1 A

Control voltage ± 48V

Bandwidth (-3dB) >2MHz

Page 106: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

98

Operating area

Slew rate (no load) >15V/µs

Rise/Fall time (no load) <250ns

Floating mode Yes

Parallel ability No

Current accuracy 0.1% range

Impedance specifications (if booster board connected to board with EIS option)

Accuracy Same as the potentiostat/galvanostat board but limited at 2 MHz

10.2.4 2A/30 V booster

Control amplifier

Compliance voltage ± 30V

Compliance current ± 2 A

Control voltage ± 30V

Bandwidth (-3dB) >3MHz

Operating area Operating area was specified in the worst case

-60

-40

-20

0

20

40

60

-1,5 -1 -0,5 0 0,5 1 1,5

E(V)

I(A)

-40

-30

-20

-10

0

10

20

30

40

-3 -2 -1 0 1 2 3

E(V

)

I(A)

Page 107: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

99

Impedance specifications (if booster board connected to board with EIS option)

Accuracy

10.2.5 4A/14V booster

Control amplifier

Compliance voltage - 3 V ; +14 V

Compliance current ± 4 A

Control voltage - 3 V ; +10 V

Operating area For VSP-300 chassis there is derating for the slot #5 and #6 (red curve). For all the other case, consider the blue curve.

Bandwidth (-3 dB) >4 MHz

Slew Rate (no load) 50 V/µs

Rise/Fall time (no load) <200 nA

Floating mode yes

Parallel ability yes

Current accuracy 0.1% range

Slew rate (no load) >50 V/µs

Rise/Fall time (no load) <200 ns

Floating mode yes

Parallel ability Yes

Current accuracy 0.1% range

Page 108: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

100

Impedance specifications (if booster board connected to board with EIS option)

Accuracy

10.2.6 10 A/ 5V booster

Control amplifier

Compliance voltage -1V; +6V

Compliance current ± 10 A

Control voltage -1V, 6V

Bandwidth (-3dB) >8MHz

Operating area

Slew rate (no load) >50V/µs

Rise/Fall time (no load) <200ns

Floating mode yes

Parallel ability Yes

Current accuracy 0.1% range

Impedance specifications (if booster board connected to board with EIS option)

-4

-2

0

2

4

6

8

-12 -8 -4 0 4 8 12

E(V)

I(A)

Page 109: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

101

Accuracy

10.2.7 HCV-3048 used with HV-48 electrometer

Control amplifier

Compliance voltage [-1,50] V

Compliance current ± 30 A

Current ranges 3 (30 A, 3 A, 300 mA)

Control voltage [0,48] V

Accuracy

User (using calibration board, when HCV-3048 purchased separately from pstat board)

Voltage control 0.1 % + 0.1 % (range + reading)

Current control 0.3 % + 0.5 % (range + reading)

Voltage measurement 0.1 % + 0.1 % (range + reading)

Current measurement 0.3 % + 0.5 % (range + reading)

Factory (when HCV-3048 purchased with pstat board, or pstat board returned to factory for calibration with HCV-3048)

Voltage control 0.03 % + 0.03 % (range + reading)

Current control 0.1 % + 0.3 % (range + reading)

Voltage measurement 0.03 % + 0.03 % (range + reading)

Current measurement 0.1 % + 0.3 % (range + reading)

Bandwidth (-3dB) 120 kHz

Operating area Control/measurement : [-30,30] A @ [0, 48] V (with 2.5 m cables)

Slew rate (no load) 20 V/µs

Rise/Fall time (no load) 3 µs

Floating mode Yes (350 kΩ isolation)

Parallel ability Yes up to 4 (120 A)

Page 110: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

102

EIS contour plot (if HCV-3048 connected to board with EIS option)

Power contour plot

WARNING When used with 110 Vac, if the HCV-3048 delivers [+25;+30] A, the fuses will blow.

Page 111: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

103

Meaning of the front panel LED

S = Solid (stays on) The colour of the letter is the colour of the LED. Status n°1: Normal operation => Permanent or solid green LED Status n°2: Static error => Permanent or solid red LED Status n°3: with the CLIO cable connected, the LED turns from red to green while the firmware is loading and the PC is finally connected to the device

thanks to the CLIO cable. Status n°4: if the CLIO cable is disconnected no firmware loading is possible => solid red LED Status n°5: if CLIO cable is unplugged during normal operation => solid green LED turns into solid red LED (Status n°4)

10.2.8 MP-MEA

General

Number of channel 256

Switching time 8 ms

Maximum Current ± 500 mA

Dimension 307 * 210 * 50 mm

Electrometer Impedance <1010 Ohm//<3pF

Bias Current <10 pA

Bandwidth 1 MHz

EIS accuracy with ULC

10.3 PC requirements

Recommended: - Pentium 2 GHz

Status n° Description LED

1 Status OK S

2Static alarm (T°C/Power

supply/Open In/Emergency Stop)S

3Plugged CLIO: Off -> On (Firmware

loaded)S -> S

4Unplugged CLIO:Off -> On

(Firmware loaded)S

5Unplugging CLIO cable during

experimentS -> S

Page 112: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

104

- 512 Mo RAM - 80 GB Hard Drive - Screen resolution 1280*1024 - Ethernet board with 10/100 base T or USB port - Windows XP© (SP2), Seven, Eight (32 or 64 bits), 10

© Microsoft Corporation. All rights reserved.

Page 113: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

105

11. Troubleshooting

11.1 Data saving

Problem: Data cannot be saved from a given channel (this channel appears in yellow in EC-

Lab, and the program displays an error message while attempting to save data): Solution(s): - ensure that the file being saved to has not been moved or destroyed, or opened by another

application, - if the saved file is on a network drive, ensure that you have the right to write data into the

same directory (create and destroy a text file). Otherwise see your network authorizations…,

- in EC-Lab, select File, Repair... Then select the saved file and click on the Repair button, - ensure that the computer’s IP Address has not been modified since the beginning of the

experiment, - if the problem persists, please contact Bio-Logic.

11.2 PC Disconnection

Problem: The PC is disconnected from the instrument ("Disconnected" is displayed in red on

the EC-Lab status bar): Solution(s): - check the PC – instrument connection,

- direct connection: verify that the Ethernet cable is plugged in at both ends, - network connection: verify that the yellow LED is blinking on the instrument front panel

and that you have access to your network directories from the PC, - check that the green led is blinking (this ensures that the multichannel potentiostat is

running properly), - in the Tera Term Pro window type "r" or "R": this will restart the Ethernet connection

program that is a part of the instrument firmware, WARNING: this operation is not a simple task, proceed with this only in extreme circumstances.

- if the problem persists, please contact Bio-Logic

11.3 Effects of computer save options on data recording

Electrochemical experiments often have a long duration (hours to days or longer). During the experiment, the user must ensure that the computer is always ready and able to record the data. If the power saving options for the hard disk are activated, there is a risk of the computer not being able to continuously record data from the instrument. In order to avoid this, we advise the user to turn off all of the power save option for the computer in the settings panel.

Page 114: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

106

12. EC declaration of conformity

12.1 SP-200

Page 115: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

107

12.2 SP-240/SP-300

Page 116: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

108

12.3 VSP-300

12.3.1 Horizontal VSP-300

Page 117: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

109

12.3.2 Vertical VSP-300

Page 118: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

110

12.4 VMP-300

Page 119: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

111

12.5 HCV-3048

Page 120: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

112

13. Glossary

This glossary is made for the user to understand most of the terms used in this instrument installation and configuration manual. The terms are classified in the alphabetical order. Amplifier: current power booster that can be added to each channel individually. Bandwidth: It depends on the electrochemical cell impedance. The bandwidths values are going from 1 to 9. Bandwidth selection is important for the stability of the potentiostat during measurement. Calibration: operation that has to be done for each channel in order to reduce the difference between a controlled value (for example Ectrl) and the corresponding measured value (for example Ewe). Cell connection: connection of the instrument channel board to the electrochemical cell with five leads. Mixed DB-25: connector with 25 pins on the instrument front panel where the cable connecting the channel board and the electrochemical cell is set. DB-9: connector with 9 pins on the instrument front panel used as auxiliary input/output. DB-15: connector with 15 pins on the instrument front panel used to connect the channel board to the booster or to the calibration board. Default settings: settings defined and saved as default by the user and automatically opened when the corresponding protocol is selected. Firmware upgrading: the firmware is the operating system of the instrument. With new improvements on the instrument, it is necessary to upgrade the firmware and the software to benefit of the most recent version. Gateway: IP address allowing the connection of computers from different networks onto an instrument. Impedance: defined by the ratio E/I. Low current: option providing a sub-aA resolution that can be added to each channel by the mean of a cable with an electrometer. This option extends the current range down to 1 pA with gain. This option can be added both to standard or Z option channel boards. Network: group of computers connected together in which the multichannel potentiostat can be added. Several users with different computers can lead experiments in one or more channels of the potentiostat. Specifications: Characteristics of the instrument such as cell control or current and potential measurement. Subnet mask: IP number used when the instrument is not in the same network as the computer.

Page 121: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

113

TCP/IP: Transfer Control Protocol/Internet Protocol using IP addresses to identify hosts on a network. Triggers: option that allows the instrument to set a trigger out (TTL signal) at experiment start/stop or to wait for an external trigger in to start or stop the run.

Page 122: Installation and Configuration Manual for VMP-300 …public.bnl.gov/sites/rep/Equipment Media/ELE-BR-001 VMP-300... · Instruments Pvt. Ltd Contact: Samuel Varghese ... Class I review

Manual for VMP300-based instruments

114

14. Index

Battery Holder ......................................................................................................................61 Channel

Board installation ..............................................................................................................62 Calibration ........................................................................................................................67

Connection IP Address change ...........................................................................................................30

Connections cell ....................................................................................................................................38 Instrument ........................................................................................................................23

Dummy cell for booster .........................................................................................................59 EQCM ..................................................................................................................................51 External device

QCM-922 ..........................................................................................................................51 File

Repair ...............................................................................................................................86 Maintenance .........................................................................................................................76 Network ................................................................................................................................16 QCM-922..............................................................................................................................51 Trigger ..................................................................................................................................44 Ultra Low Current option .......................................................................................................37

Windows Vista ....................................................................................................................15

Manual Rev_I (April 2016)