IEEE EMBC 2006, 30 August – 3 September, 2006, New York City - We hypothesize that intracardiac...
-
Upload
cora-gibbs -
Category
Documents
-
view
216 -
download
1
Transcript of IEEE EMBC 2006, 30 August – 3 September, 2006, New York City - We hypothesize that intracardiac...
IEEE EMBC 2006, 30 August – 3 September, 2006, New York City
- We hypothesize that intracardiac impedance can provide real-time hemo- dynamic information to automatically control the upper and lower pacing rate limits of a rate adaptive pacemaker.
- A number of technical challenges must be overcome to incorporate bioimpedance based automatic control of rate within the constraints of an implantable device.
- We will introduce some basic results of our collaborative study in this subject.
Intracardiac Electrical BioimpedanceIntracardiac Electrical Bioimpedanceas a Basis for Controlling of Pacing Rate Limitsas a Basis for Controlling of Pacing Rate Limits
Andres KinkAndres Kink, Rodney W. SaloRodney W. Salo, Mart Min Mart Min , Toomas ParveToomas Parve, and Indrek RätsepIndrek Rätsep
Guidant Corporation (USA), Smartimplant Ltd, and Tallinn University of Technology (ESTONIA)
IEEE EMBC 2006 30 Aug – 3 Sept 2006, New York City 2 A.Kink, R.W.Salo, M.Min, T.Parve, I.Rätsep. Intracardiac Electrical Bioimpedance as a Basis for Controlling of Pacing Rate Limits
1. Over pacing and Under pacing
Fig. 1. A rate-adaptive pacing system
The pacing rate PR should follow the body workload Wbody, which correlates with minute ventilation MV = RR x TV estimated, e.g., through a transthoracic bioimpedance (Fig. 1).
Problem: Inadequate estimation can lead to dangerous, too high or too low PR
Automatic control of upper and lower PR limits prevents both, over and under pacing.
Cardiac output CO = HR SV
Veins
P
LLuunnggss
O2
Arteries
Pacing Rate PR
Skeletal Muscles and other tissues
Activity
Acceleration Minute Volume MV
Myocardium
P R
Heart Rate
Stroke Volume
Respiration Rate Tidal Volume
Minute Volume MV = RR TV
Over pacing takes place when- the cardiac cycle is too short to permit adequate filling, thus limiting cardiac output, or - cardiac output is maintained, but at the cost of excessive oxygen consumption.
Under pacing occurs when the sensed demand is low and the paced rate is either - insufficient to meet first cardiac or overall metabolic demand, or - the cardiac output is maintained only by an excessive increases in preload and high filling pressures.
IEEE EMBC 2006 30 Aug – 3 Sept 2006, New York City 3 A.Kink, R.W.Salo, M.Min, T.Parve, I.Rätsep. Intracardiac Electrical Bioimpedance as a Basis for Controlling of Pacing Rate Limits
2. Why the paced heart rate is critical?
Fig. 2. Ventricular pressure-volume loop (a),
and variation of arterial pressure (b):
1. HR = Hrest 2. HR = 2·Hrest
Metabolic demand can exceed the capabilities of a damaged heart (W > E).
Problem: artificial pacing may drive the heart into failure.
A “positive” balance E ≥ W must be maintained !
Pacing rate limits are needed, but incorrect, fixed limits can exacerbate problems such as: - Postural hypotension, caused by shifts in blood volume to lower extremities, which can lead to syncope, and - Neurogenic syncope, which is even more insidious problem because a sudden drop in blood pressure may occur minutes after the precipitating event.Dynamically adaptive limits are needed!
Sdem < Ssup
Wext ≈ ΔP · SV E ≈ ΔP · tdiast
external work energy supply
IEEE EMBC 2006 30 Aug – 3 Sept 2006, New York City 4 A.Kink, R.W.Salo, M.Min, T.Parve, I.Rätsep. Intracardiac Electrical Bioimpedance as a Basis for Controlling of Pacing Rate Limits
3. Myocardial Energy ?
Fig. 3. P-V diagram with energy consumption areas Sdem and Spot , where ESPVR is the end-systolic P-V relationship line.
The area
Sdem ≈ ΔP · SVin Fig. 3 represents the myocardium’s
external work Wext.
The area
Spot ≈ ΔP·(Ves ‑ V0 ) ∕ 2 is proportional to the myocardium’s
internal work Wint ,
Total energy consumption W of myocardium can be expressed as
Scons = Sdem + Spot = = ΔP·(2·SV + Ves ‑ V0) ∕ 2
IEEE EMBC 2006 30 Aug – 3 Sept 2006, New York City 5 A.Kink, R.W.Salo, M.Min, T.Parve, I.Rätsep. Intracardiac Electrical Bioimpedance as a Basis for Controlling of Pacing Rate Limits
4. Myocardial Energy and Impedance
The myocardial energy balance can be expressed as
Ssup = Sdem ∙ kpot , where kpot expresses the role of internal work and
depends almost linearly on relative stroke volume SV/Ves ,
but only very slightly on relative pressures Pas/Pad (Fig.4).
It can be approximated as
kpot = (1.7-1.8) SV/Ves
Rough estimation of the overall energy balance
E = Wext + Wint does not require pressure measurements.
We can extract the needed information (time intervals and relative ventricular volumes) solely from the measured intracardiac impedance waveform !
Fig. 4 Relationship of the relative energy demand Sdem/Spot
versus relative stroke volume SV/Ves
for some values of Pas /Pad .
IEEE EMBC 2006 30 Aug – 3 Sept 2006, New York City 6 A.Kink, R.W.Salo, M.Min, T.Parve, I.Rätsep. Intracardiac Electrical Bioimpedance as a Basis for Controlling of Pacing Rate Limits
5. Experiments and Modeling
Fig. 5. A comparison between right ventricular stroke volume measured by aortic flow, and right ventricular peak-to-peak resistance, ΔR, in a dog during an infusion of dobutamine.
1) the actual stroke volume in ml measured by aortic flowmeter in an intact canine model,
and
2) the peak-to-peak resistance change, ΔR (in ohms), during a cardiac cycle measured with a four electrode right ventricular catheter.
In Fig. 5. is shown an experimental comparison between
IEEE EMBC 2006 30 Aug – 3 Sept 2006, New York City 7 A.Kink, R.W.Salo, M.Min, T.Parve, I.Rätsep. Intracardiac Electrical Bioimpedance as a Basis for Controlling of Pacing Rate Limits
Fig. 6. A comparison between the stroke volumes computed from resistances (vertical axis) measured across the left ventricle (from right ventricular apex to left ventricular free wall)
and the actual left ventricular volumes (horizontal axis).
Fig. 6 shows the relationship between actual volumes, and the volumes that were calculated from impedances (vertical axis).
A finite-difference model of heart conductance containing a standard implantable electrode system was used to generate these data.
The near-to-linear relationship indicates the feasibility of following changes in relative stroke volume using a standard electrode system.
IEEE EMBC 2006 30 Aug – 3 Sept 2006, New York City 8 A.Kink, R.W.Salo, M.Min, T.Parve, I.Rätsep. Intracardiac Electrical Bioimpedance as a Basis for Controlling of Pacing Rate Limits
Fig. 7. Dependence of the stroke volume SV and diastolic time tdiast on pacing rate PR.
Stabilizing the Stroke Volume
The energy supply diminishes with the higher rate because the dia-stolic time tdiast
shortens (Fig.7).
When stabilizing the stroke volume at the predetermined reference value
SV = SVref , (see Fig.8) the myocardium’s
energy consumption will remain constant.
The energy balance can be expressed directly through the pacing rate.
Substituting tdiast = (60/PR) – tsyst in kpot·(SVref) · ΔP ≤ (tdiast) · ΔP
we obtain that PR ≤ 60 SVref ⁄ (1 + kpot · SVref · tsyst).
t cycle , s
0 60 120 180 PR, bpm
1 s
Rest 100W 400 Body work, W
0.5
tsyst
SV, ml
1.5
50 ml SVrest
SV – stroke volume
tdiast
60 1 tcycle = ------- = ------- = tsyst + tdiast PR fbeat
IEEE EMBC 2006 30 Aug – 3 Sept 2006, New York City 9 A.Kink, R.W.Salo, M.Min, T.Parve, I.Rätsep. Intracardiac Electrical Bioimpedance as a Basis for Controlling of Pacing Rate Limits
Fig. 8. A closed-loop system to control pacing rate
by maintaining a fixed stroke volume.
A closed loop stabilizer of Stroke Volume
Pacing rate is controlled automatically as a tool to stabilize the stroke volume
SVactual = SVref
using a closed-loop or feedback system.
IEEE EMBC 2006 30 Aug – 3 Sept 2006, New York City 10 A.Kink, R.W.Salo, M.Min, T.Parve, I.Rätsep. Intracardiac Electrical Bioimpedance as a Basis for Controlling of Pacing Rate Limits
Modeling of the Heart and Intracardiac Impedance
Fig. 9. 3D digital model of the heart with 16 measurement electrodes.
Fig. 10. Experimentation on a virtual heart reflecting impedance variations during a cardiac cycle
measured between different pairs of electrodes:
a) 1 and 4, b) 2 and 4, c) 3 and 4,
applying excitation current to electrodes 1 and 4.
IEEE EMBC 2006 30 Aug – 3 Sept 2006, New York City 11 A.Kink, R.W.Salo, M.Min, T.Parve, I.Rätsep. Intracardiac Electrical Bioimpedance as a Basis for Controlling of Pacing Rate Limits
Experimentation with an isolated pig’s heart
A picture of an experimental setup with an isolated pig heart is shown in Fig. 11.
The resuscitated heart was continuously perfused by an artificial blood circulation system permitting active experimentation for several hours.
An original impedance measurement and data acquisition system was developed for performing in vivo experiments.
PCT Patent Application WO2005/122889 A1 (29.12.2005).
Fig. 11. Experimental setup with an isolated pig’s heart.
IEEE EMBC 2006 30 Aug – 3 Sept 2006, New York City 12 A.Kink, R.W.Salo, M.Min, T.Parve, I.Rätsep. Intracardiac Electrical Bioimpedance as a Basis for Controlling of Pacing Rate Limits
6. Discussion
As a result of our study, we assume that:
A. it becomes possible to use 1) hemodynamic information derived from bioimpedance measurements alone for feed-forward control of pacing rate, while2) simultaneously monitoring myocardial energy balance to preclude potentially damaging rates using feedback corrections.
B. the pacing rate is only an intermediate parameter, available as a control tool to 1) satisfy the patient’s metabolic demands, and2) fulfilling the myocardium’s energy supply needs.
C. a potential benefit of stabilizing the stroke volume we would maintain a relatively constant preload and myocyte “stretch”, minimizing hypertrophic signaling and subsequent cardiac remodeling.
IEEE EMBC 2006 30 Aug – 3 Sept 2006, New York City 13 A.Kink, R.W.Salo, M.Min, T.Parve, I.Rätsep. Intracardiac Electrical Bioimpedance as a Basis for Controlling of Pacing Rate Limits
Conclusions
Our experimental studies and theoretical speculations confirm that:
- Increased concern over cardiac efficiency and maintenance of energy balance within the heart may be addressed by novel pacing control algorithms that require only relative stroke volume information, derivable from bioimpedance measurements applied to a feedback control system.
- New impedance measurement methods can permit more reliable results to make such feedback systems feasible for rate control.
- Model based design appears to be a fruitful tool for the synthesis of complicated and nonlinear closed loop systems for pacing rate control.
Thank you for your attention !