i-STAT Technical BulleTin analyser coded...Since multiple codes can be associated ... 22, 25...
Transcript of i-STAT Technical BulleTin analyser coded...Since multiple codes can be associated ... 22, 25...
Art:714260-01E Rev.Date:08/14/06
Technical BulleTini-STAT®
ANALYZER CODED MESSAGES
Fromthetimeitpowersupuntilthetimeitpowersdown,thei-STAT®Analyzerperformsnumerousqualitychecks.Thefailureofanyqualitycheckcausestheanalyzertohaltthetestcycleanddisplaya“cause”,an“action”message,andacode.
The Cause Message:This message describes the likely cause of the failed quality check. For example, when an overfilledcartridgeisdetected,theanalyzerwilldisplay“SamplePositionedBeyondFillMark”.
The Action Message:Thismessageindicatestheappropriateaction.Forexample,ifitislikelythequalitycheckwillfailagainthenexttimetheanalyzerisused,theinstruction“UseElectronicSimulator”willbedisplayed.Iftheproblemisrelatedtoanoperatororcartridge,theinstruction“UseAnotherCartridge”willbedisplayed.
The Cause Code:This isanumericcodeassociatedwith the failedqualitycheck. Sincemultiplecodescanbeassociatedwithasinglecausemessage,thisisessentialinformationwhencontactingi-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.Thecodesarestoredintheanalyzer’smemoryalongwithothertestrecordsandaretransmittedtotheCentralDataStation.Thecodelistcanbeviewedandprinted.
Codes 1-15 and 95 usually indicate a condition related to the environment or the state of the analyzer. These conditions are usually benign and go away after the next cartridge or Electronic Simulator is inserted, or after the offending condition is corrected.
Code Number
Cause/Action Message on Display Explanation
1 Dead Batteries/ReplaceBatteries
Thereisinsufficientbatterypowertocompletethetestingcycle.Replacethedisposablelithiumbatteriesintheanalyzerorrechargetherechargeablebatteries.
Ifyouareexperiencingthiscodefrequentlyandusedisposablebatterieswiththei-STAT1analyzer,youmaywanttoconsidertherechargeablebatterysystemavailablewiththei-STAT1Analyzer.
Art:714260-01E Rev.Date:08/14/062
Code Number
Cause/Action Message on Display Explanation
2 Temperature Out of Range/CheckStatusPage
Theanalyzerisrecordingatemperatureoutsideitsoperatingrange.Movetheanalyzertoanareawithintheoperatingtemperatureofthetestbeingperformedandallowtheanalyzertocometothenewroomtemperature.Checktheanalyzer’stemperaturereadingontheStatusPage.
3 New Software Installed/UseElectronicSimulator
ThismessageappearsonthePortableClinicalAnalyzerafternewsoftwarehasbeeninstalledor,insomecases,whenanewcustomizationprofileisreceived.
4, 8 Analyzer Interrupted/UseAnotherCartridge
Theanalyzerhasdetectedthatthelasttestcyclewasnotcompleted.Thiscanhappenifthebatterieswereremovedorweremakingpoorcontactwhileacartridgewasstillintheanalyzer.Batteriesthataretooshortwillnotmakepropercontact.Checkthatthebatteriesareinsertedproperlyandseatedwellintheanalyzer;checkthebatteryvoltageontheanalyzer’sStatusPageandreplacebatteriesiflow.NOTE:Patientresultsdisplayedbeforethiscodearevalid.
5, 6, 9 Analyzer Interrupted/ReadyforUse
ThePortableClinicalAnalyzerisunabletorefreshthedisplay.Thiscanhappenifpowerisinterruptedbeforetheanalyzerpowersitselfdown.Checkthatbatteriesareinsertedproperlyandseatedwellintheanalyzer.Batteriesthataretooshortwillnotmakepropercontact.CheckthebatteryvoltageontheStatusPage.
7 Batteries Changed/ReadyforUse
ThisisanormalresponseonthePortableClinicalAnalyzerwhenthebatteriesarechangedafteracode1hasoccurred.
10 Temperature In Range/ReadyforUse
Temperatureisbackinrangefollowingacode2onthePortableClinicalAnalyzer.
11 Date Invalid/CheckClockonStatusPage
Ifthedateintherealtimeclockprecedesthereleasedateprogrammedintotheapplicationsoftware,code11istriggered.Checkthedateontherealtimeclock.
Theaccuracyoftheclockischeckedatthebeginningofacoagulationtest.Iftheclockisinaccurate,Code11istriggered.
12 Invalid or Expired CLEW/SeeManual
TheCLEWstandardizationhasexpired.DownloadavalidCLEW.
ThedateontherealtimeclockexceedstheexpirationdateoftheCLEWsoftware.Checkthedateontherealtimeclock.
13 Invalid or Expired CLEW/SeeManual
TheCLEWiscorruptornotcompatiblewiththeapplicationsoftware(JAMS),orthereisnoCLEWintheanalyzer.DownloadavalidCLEW.IfthiscodeoccursafterasoftwareupgradeandthecustomizationapplicationisenabledintheCDS,changetheCLEWversionintheCustomizationProfiletothelatestversionandre-transmittheprofiletotheanalyzer.
14 Analyzer Error/SeeManual
Customizationprofileiscorrupted.Retransmitthecustomizationprofile.Ifcode14reoccurs,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.
15 Barcode Does Not Match Cartridge Type
Thebarcodescannedbytheuserdoesnotmatchtheimmunoassaycartridgetypeindicatedbytheidentificationchipinthecartridge.Theusershouldrunanothercartridge,beingcarefultoscanthebarcodefromtheportionpackofthespecificcartridgetypebeingrunontheanalyzer.
95 Test Cancelled by Operator
Thismessagewillappearinthestoredtestrecordsonthei-STAT1Analyzeriftheanalyzerpowersdownbeforemandatoryinformationwasentered.
� Art:714260-01E Rev.Date:08/14/06
The following codes are associated with the cartridge or fluid movement within a cartridge. These conditions can be operator or sample related. In most cases, a new cartridge must be used. If a condition persists, especially if isolated to one analyzer, there may an analyzer problem.
Code Number
Cause/Action Message on Display Explanation
19 No Clot Detetced/SeeManual
DuringthePT/INRcycle,noclotwasdetected.Runanothercartridge.Ifcode19reappears,runthesampleonanalternatemethodology.
22, 25 Cartridge Error/UseAnotherCartridge
Thesecodesoccuronlyforcoagulationcartridgesifthemixingofthesampleandreagentiscompromised.Thiscanbecausedbyaninsufficientorclottedsample,orbyairbubblesinthesample.
24 Cartridge Error/UseAnotherCartridge
Theelectricalresistanceofthecalibrantfluid(Rcal)usedtoverifytheelectrolyteconcentrationisoutofspecification.Thiscouldoccurifthecalibrantpackwasrupturedwellbeforethetestallowingevaporationtoresultinahigherelectrolyteconcentration.
Besidestheelectrolyteconcentration,theRcalisalsoaffectedbythetemperatureandtheheightandwidthofthefluidsegmentovertheconductometricsensor.Theanalyzeraccountsforthetemperature,buttheheightandwidthofthefluidsegmentcanvaryfromcartridgelottocartridgelot.Theanalyzerhasbeenprogrammedtocompensatefortheselot-to-lotdifferencesbymaintainingarunningaverageoftheRcalvaluesmeasuredfromthemostrecentcartridgeruns.Occasionally,thedifferencebetweentheRcalvaluesfortwocartridgelotsislargeenoughtocausetheintroductionofanewlottotriggercode24onthefirstfewcartridgeruns.TheCode24errorsshoulddisappearastherunningaverageadjusts.However,ifcode24persistsaftermorethan�cartridgerunsoneachanalyzer,contacti-STATTechnicalServicesoryourlocalsupportorganization.
26 Cartridge Error/UseAnotherCartridge
Thiscodeoccursiftherewasacoagulationspecificqualitycheckfailure:prematuresubstrateactivation,abnormallylowlevelsofsubstrate,orinvalidfluidmotion.
20, 27-29, 32, 33, 40, 41, 45, 87
Cartridge Error/UseAnotherCartridge
Thesecodesidentifyproblemswiththecartridgesuchas:calibrantfluidarrivingtoosoon,toolate,ornotatall,ornoiseinthecalibrantfluidsignals.Codes20,27,and41canbecausedbypoorcontactthatcansometimesbecorrectedbyconditioningthepinsintheanalyzerusingtheceramiccleaningcartridge.Thespecificconditioningprocedureisdescribedattheendofthisbulletin.
42, 43 Cartridge Error/UseAnotherCartridge
Thesecodesindicatethattheconductometricsensor(code42)ortheamperometricsensor(code4�)wasoutofspecification.Thiscouldbecausedbyapre-burstcalibrantpack,dirtycartridgecontactpads,oradirtyconnectorintheanalyzer.
79-81 Cartridge Error/UseAnotherCartridge
Badcontactbetweenthethermalprobesintheanalyzerandthemetalizationonthebackofthechipsinthecartridgetriggerthesecodes.Causesare:poormetalizationofthechips,dirtonthemetalization,orbentorbrokenthermalprobesintheanalyzer.
Art:714260-01E Rev.Date:08/14/064
Code Number
Cause/Action Message on Display Explanation
21 Cartridge Preburst/UseAnotherCartridge
Thiscodeindicatesthattheanalyzerdetectedfluidonthesensorsbeforeitshouldhave.Possiblecauses:mishandlingofcartridges(puttingpressureinthecenterofthecartridge),poorstorageconditionsofcartridges(frozen),orrerunningusedcartridges.
31, 34, 44 Unable to Position Sample/UseAnotherCartridge
Theanalyzerdidnotdetectmovementofsampleacrossthesensors.Thiscouldbeduetoaclotinthesample(especiallyinneonates),tonotclosingthesnapclosureonthecartridge,ortoanaberrantcartridge.
35, 36 Sample Positioned Short of Fill Mark/UseAnotherCartridge
Thecartridgewasunderfilled.Thesamplemustreachthefillmark.Tryanothercartridge.
30, 37 Sample Positioned Beyond Fill Mark/UseAnotherCartridge
Thecartridgewasoverfilled.Thesamplewaspastthefillmark.Tryanothercartridge.
38, 39 Insufficient Sample/UseAnotherCartridge
Thisismostlikelyduetoinsufficientsampleinthesamplewellofthecartridge,butcanalsobecausedbybubblesinthesample.Tryanothercartridgeandensuresufficientsampleisinthesamplewell.
46 Cartridge Error/UseAnotherCartridge
Theanalyzerdidnotdetectmovementofsampleacrossthesensors.Thiscouldbeduetoaclotinthesample(especiallyinneonates),tonotclosingthesnapclosureonthecartridge,ortoanaberrantcartridge.
47 Cartridge Not Inserted Properly/ReinsertCartridge
ThiscodeindicatesthecartridgeorElectronicSimulatormaynotbepushedinalltheway.ReinsertthecartridgeorElectronicSimulator.Iftheproblempersistsand/ortheuseriscertainthecartridgeorSimulatorisproperlyinserted,itmayindicateananalyzerproblem.Contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.
48 Analyzer Error/SeeManual
ThiscodeindicatesthecartridgeorElectronicSimulatormayhavebeen“cocked”wheninserted.PushthecartridgeorSimulatorstraightthroughthecartridgeport.Iftheproblempersists,andtheuseriscertainthecartridgeorSimulatorisproperlyinserted,itmayindicateananalyzerproblem.Contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.
49 Poor Contact Detected/SeeManual
Thesystemdetectedacontactproblemwithoneoftheconnectorpinswhilereadingtheidentificationchipintheimmunoassaycartridge.Thiscansometimesbecorrectedbyconditioningthepinsintheanalyzerusingtheceramicconditioningcartridge.Thespecificconditioningprocedureisdescribedattheendofthisbulletin.
Note:Ifyoudonothaveaceramicconditioningcartridge,pleasecontactyourSupportRepresentative.
� Art:714260-01E Rev.Date:08/14/06
The following conditions are related to electronic or mechanical failures in the analyzer.
Code Number
Cause/Action Message on Display Explanation
50 Analyzer Error/UseElectronicSimulator
Themotorhasmovedtoofar.Runningasimulatormaynotdetectthisproblem.Runthesimulatorandiftheanalyzerpasses,runacartridgetoseeifthecodereoccurs.Ifnot,continuetousetheanalyzer.Ifthecodereoccurs,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.
Iftestingimmunoassaycartridgesonani-STAT1Analyzer,thiscodecanberelatedtopoorelectricalconnectionbetweenthei-STAT1Analyzerandthecartridge.Thiscansometimesbecorrectedbyconditioningthepinsintheanalyzerusingtheceramicconditioningcartridge.Thespecificconditioningprocedureisdescribedattheendofthisbulletin.
Codes126and128aresometimesrelatedtoelectricalconnectionaswell.Ifyouexperiencemultipleoccurrencesofthese�codes(�0,126,and128)inashortperiodoftime,considerreturningtheanalyzerforservicingandreplacement
Thepresenceofsamplebubbleswhenrunningimmunoassaycartridgesmay,undersomecircumstances,alsoelicitthiscode.
51 Analyzer Error/UseElectronicSimulator
Themotormovedfortoolong.Runasimulator.IftheerroroccurredwhilerunninganACTcartridge,alsorunacartridge.Ifthecodedoesnotreoccur,continuetousetheanalyzer.Undersomeconditions,alowbatterywillcausethiserrorinsteadofcode1.Tryfreshbatteries.Ifthecodereoccurs,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.
52 Analyzer Error/UseElectronicSimulator
Themotorstalledwhilemoving.Runasimulator.IftheerroroccurredwhilerunninganACTcartridge,alsorunacartridge.Ifthecodedoesnotreoccur,continuetousetheanalyzer.Ifthecodereoccurs,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.
58-62 Analyzer Error/UseElectronicSimulator
Theanalyzerusuallyrecoversfromtheseerrorconditions.TheseerrorconditionscanbedetectedbytheElectronicSimulator.IftheanalyzerpassestheElectronicSimulatortest,continuetouseit.Ifnot,checkthebatteryvoltageandchecktheanalyzerwithanothersimulatortoruleoutasimulatorproblem.Ifthecodepersists,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.
Art:714260-01E Rev.Date:08/14/066
Code Number
Cause/Action Message on Display Explanation
23, 53, 55, 63, 65-68, 70, 72-74, 82, 85, 86, 89-94, 96, 97
Analyzer Error/SeeManual
Thesearemechanicalorelectronicfailuresfromwhichtheanalyzermaynotbeabletorecover.
Code2�maybecausedbypoorcontactbetweentheanalyzerpinsandthecartridgechip.Thiscansometimesbecorrectedbyconditioningthepinsintheanalyzerusingtheceramiccleaningcartridge.Thespecificconditioningprocedureisdescribedattheendofthisbulletin.
Code70canoccuronthePortableClinicalAnalyzeriftheuserpressestheDISkeybeforetheElectronicSimulatorisrunafterasoftwareupdate.Ifthishappens,reseatthebatteriestoresettheanalyzer,andthenruntheElectronicSimulator.
Codes82and92typicallyindicateaproblemwiththepressuretransducersintheanalyzer.Ifthesecodespersist,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.
Forothercodes,runtheElectronicSimulatortwice,thenrunacartridgewithasample.Iftheanalyzerpassesthesimulatorcheckandaqualitycheckdoesnotoccurwiththesamplerun,continuetousetheanalyzer.Iftheanalyzerdoesnotpassthesimulatorcheckand/oraqualitycodeoccurswiththesamplerun,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.
69 Cartridge Type Not Recognized/UseAnotherCartridge
Thiscodecouldbeduetouseofacartridgetypethatisnotcompatiblewiththeversionofsoftwareintheanalyzer,ortheuseofexpiredcartridges.Checkthecartridgeexpirationdateonthecartridgeboxorpouch.Ifthecartridgeshavenotexpired,andifanewcartridgetypeisbeingrun,contacti-STATTechnicalServicesoryourlocalsupportorganizationforasoftwareupdate.
Whenrunningcoagulationcartridges,Code69maybecausedbypoorcontactbetweentheanalyzerpinsandthecartridgechip.Thiscansometimesbecorrectedbyconditioningthepinsintheanalyzerusingtheceramiccleaningcartridge.Thespecificconditioningprocedureisdescribedattheendofthisbulletin.
Duringimmunoassaycartridgeruns,thiscodewillbedisplayedifincorrectinformationisenteredinresponsetotheprompt“EnterorScanCartridgeLotNumber”.
Theinstrumentexpectsthebarcodeonthebackoftheindividualcartridgeportionpacktobescanned.Thecorrectbarcodelookslikethis:
Forimmunoassaycartridges,theinstrumentwillnotacceptkeypadentriesofthecartridgelotnumbernorascanofthebarcodeonthecartridgebox.
Thisconditionmaybeduetoanaberrantcartridge.However,iftheconditionoccursrepeatedlyononeanalyzer,theanalyzermayneedrepair.Contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.
7 Art:714260-01E Rev.Date:08/14/06
Codes 100-111 indicate a condtion with the PCx or PCx Plus Glucose Test Strip on the i-STAT 1 Analyzer
Code Number
Cause/Action Message on Display Explanation
100 Strip Error/UseAnotherStrip
Theusertriedtorunawetstrip.Removetheteststrip.Press1forTestOptions.Press1forSamePatientorLevel.Repeatthetest.
101 Strip Error/UseAnotherStrip
Theteststripwasremovedfromthestripportduringtesting.Press1forTestOptions.Press1forSamePatientorLevel.Repeatthetest.
103, 105, 106, 107, 111
Strip Error/UseAnotherStrip
Testunsuccessful.Anerrorwasdetectedduringtheanalysissequence.Removetheteststrip.Press1forTestOptions.Press1forSamePatientorLevel.Repeatthetest.Iftheproblempersists,recordthethreedigiterrorcodeandcontactAbbottDiabetesCaresupportservicesoryourlocalsupportorganization.
102, 104 Strip Error/UseAnotherStrip
Theteststripmalfunctionedorthebloodglucoselevelinthesampleisbeyondthemeasuringcapabilityoftheteststripandstripreader.Removetheteststrip.Press1forTestOptions.Press1forSamePatientorLevel.Repeatthetestwithanewteststrip.Iftheerroroccursagain,confirmtheresultbyperformingthetestonadifferentmethod.ContactAbbottDiabetesCaresupportservicesoryourlocalsupportorganization.
108, 109 Temperature Out of Range/CheckStatusPage
Duringthetest,roomtemperaturebecameunstableormovedoutsidethelimitswithinwhichtheteststripreadercanperformatest.Ensurethattheroomtemperatureiswithinthespecifiedlimits.Allowtheanalyzertostabilizetoaroomtemperatureof1�-40ºCor�9-104ºF.PresstheMenukeyuntiltheAdministrationMenuisdisplayed.Press1forAnalyzerStatuswheretheroomtemperaturereadingisdisplayed.
IfthetemperaturereadingontheStatusPageiswithinthelimitsdescribedabove,yetthesecodespersist,theremaybeaproblemwithoneofthethermistorsintheanalyzer.Contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.
110 Strip Error/UseAnotherStrip
Atthestartoftheteststripcycle,theanalyzerpromptstheusertoapplybloodtothestripandwaitsfor20minutesfortheusertodoso.Thiserroroccurswhenthe20minuteshaveelapsedandtheanalyzerdidn’tdetectblood.Themostlikelycauseofthiserrorislackofuserinteraction.
Art:714260-01E Rev.Date:08/14/068
Codes in the range of 120 to 138 and 140 to 151 indicate a failure during an immuno cartridge cycle. In most cases, the cartridge is spent and another cartridge must be used. Only the i-STAT 1 Analyzer produces these codes, as the Portable Clinical Analyzer does not support immuno cycles.
Code Number
Cause/Action Message on Display Explanation
120-122, 124, 125, 133, 144, 148
Cartridge Error /UseAnotherCartridge
Thesecodesindicateaproblemwiththemovementoftheanalysisfluidduringthecartridgerun.Tryanothercartridge.
123 Cartridge Error /UseAnotherCartridge
Thequalitycontrolduringthecartridgerunfailedtoverifythepresenceofactiveimmunoreagents.Tryanothercartridge.
126 Cartridge Error /UseAnotherCartridge
Thiscodecanberelatedtopoorelectricalconnectionbetweenthei-STAT1Analyzerandthecartridge.Thiscansometimesbecorrectedbyconditioningthepinsintheanalyzerusingtheceramicconditioningcartridge.Thespecificconditioningprocedureisdescribedattheendofthisbulletin.
Codes�0and128aresometimesrelatedtoelectricalconnectionaswell.Ifyouexperiencemultipleoccurrencesofthese�codes(�0,126,and128)inashortperiodoftime,considerreturningtheanalyzerforreplacement.
127 Cartridge Error /UseAnotherCartridge
Awetsensorwasdetectedbeforetheinitialsamplemovement.Possibleoverfilledorusedcartridge.Tryanothercartridge.
128, 131, 132, 134, 135 - 138
Cartridge Error /UseAnotherCartridge
Thesecodesaremostoftenrelatedtopoorfillingofanimmunoassaycartridge,thepresenceofsamplebubbles,ortheabruptinsertionofacartridgeintotheanalyzer.
Guidelines for proper filling:
1. Discard(always)1dropfromdeliverydevicetoclearunseenbubbles.
2. Hangsingledropslightlylargerthanroundtargetwell.
�. Touchonedrop(only)toroundtargetwellallowingcartridgetodrawsamplein.
4. ConfirmsamplevolumelinesupwithtopofREDFILLLINEdiagram.
�. Closeslidecoverfromlefttoright.
Guidelines for cartridge insertion:
1. Afterclosingthecartridge,graspthecartridgeclosurebetweenyourfirstfingerandthumb.Thereisarecessforyourthumbintheclosure.
2. Guidethecartridgeintotheanalyzergently,untilasoftclickisheard.
129, 142, 143
Cartridge Error /UseAnotherCartridge
Theanalyzerdetectedanalysisfluidmixedwiththesample.Tryanothercartridge.
130 Cartridge Error /UseAnotherCartridge
Theanalyzerdetectedanairbubbleinthesamplesegment.Tryanothercartridge.
140 Lot Expired Theanalyzerdetectedanexpiredcartridgelot.Checktheexpirationdateandrepeatthetestusinganon-expiredcartridgelot.
9 Art:714260-01E Rev.Date:08/14/06
Code Number
Cause/Action Message on Display Explanation
141 Test Canceled by Operator
Thiscodewillbedisplayedifthecartridgebarcodeisnotscannedwithin60secondsofcartridgeinsertion.Thecorrectbarcodetoscanisthebarcodeonthecartridgeportionpack,nottheoneonthecartridgebox.Anexampleoftheportionpackbarcodeisfoundinthetablelistingforcode69above.
145 Cartridge Error /UseAnotherCartridge
Theanalyzerfailedtodetectfluidarrivalupontheinitialsamplepush.Thismaybecausedbya(n):
• cartridgeleak,
• failuretoclosethecartridgecompletely.Ensurethattheslidecoverisfullyengagedbeforeinsertingthecartridgeintotheanalyzer
• underfilledcartridge.Onceasingledropofsampleistouchedtothetargetwell,immunoassaycartridgeswillfillautomaticallybywickingthesampleatafixedspeed.Tryingtoinjectthesampleintothecartridgeoraddingmoresampletothetargetwellwillnotmakethecartridgefillfaster.Waitforthesampletoreachthe“fillto”mark,andthenclosethecartridge.
146 Cartridge Error /UseAnotherCartridge
Overfilledcartridge.Repeatthetest.
147 Analyzer Error /SeeManual
Inordertorunanimmunoassaycartridge,thei-STAT1Analyzermust:
• bearthe symboland
• becustomizedfor
a) CartridgeBarcodeRequired,or
b) CartridgeInformationFirstRequiredandCartridgeLotNumberRequired.
Ifeitherofthesetwoconditionsabovearenotmet,theanalyzerwilldisplaythiscode.
149 - 151 Cartridge Error/UseAnotherCartridge
Theanalyzerdetectedanatypicaldatastreamfromtheanalyzer.Tryanothercartridge.
Art:714260-01E Rev.Date:08/14/0610
The following conditions are related to the Electronic Simulator
Code Explanation How to Respond
Numerical Code
SeeunderAnalyzerCodedMessages. SeeunderAnalyzerCodedMessages.
L Potentiometricchanneloutoflimits.Canoccurifmoisturecollectsonthecontactpinsinsidetheanalyzerwhentheanalyzerissubjectedtoambienttemperaturechange.
Allowanalyzertoequilibrateinnewenvironmentfor�0minutesandrepeattest.Ifcodereoccurs,returnanlayzer.
G Amperometricchanneloutoflimits.Canoccurifexternalsimulatornotinsertedstraight.
Reinsertthesimulatorstraight.Ifcodereoccurs,returnanalyzer.
R, r Resistancereadingonconductometricchanneloutoflimits.
Returnanalyzer.
t Thermalprobefailure. Returnanalyzer.
B Potentiometricchanneloutoflimits. Returnanalyzer.
NOTE: Anytimerepetitivecodesoccurwhichcannotbeaddressedorcorrectedthroughtraining,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.
11 Art:714260-01E Rev.Date:08/14/06
PROCEDURE FOR USING AN i-STAT CERAMIC CONDITIONING CARTRIDGE (CCC) FOR ANALyzER PIN CONDITIONING
Step Number Explanation
1.RunanexternalElectronicSimulator. IftheanalyzerisconfiguredwiththeinternalElectronicSimulatorenabled,runanexternalElectronicSimulator.RunningtheexternalElectronicSimulatorensurestheinternalSimulatorcyclewillnotexecuteduringthepinconditioningprocess,whichcouldleadtotheprematureterminationoftheprocess.
2.RuntheCCCtwotimes. InitiatetheCCCcycleasyouwouldinitiateanexternalElectronicSimulatorcycle.TheinstrumentwillidentifytheCCCasanexternalElectronicSimulatoranddisplayaSimulatorFailureCode(i.e.rRGL)whenthecycleiscomplete.Disregardthecode,asthisisexpectedbehavior.
�.UpdatetheCCCUsageLog Thelogislocatedonpage�oftheTechnicalBulletinentitled“InstructionsforRestoringAnalyzersThatProduce***forHematocritandQualityCheckCode2�”,whichisshippedwiththeCCC.UpdatingthelogallowstheusertokeeptrackofthenumberofpinconditioningcyclesperformedwiththecurrentceramicstripintheCCC.Ifnecessary,replaceorrotatetheceramicstripsotheCCCisreadyforfutureuse.
4.Returntheanalyzertoservice.