Art: 714260-01E Rev. Date: 08/14/06 TECHNICAL BULLETIN i-STAT ® ANALYZER CODED MESSAGES From the time it powers up until the time it powers down, the i-STAT ® Analyzer performs numerous quality checks. The failure of any quality check causes the analyzer to halt the test cycle and display a “cause”, an “action” message, and a code. The Cause Message: This message describes the likely cause of the failed quality check. For example, when an overfilled cartridge is detected, the analyzer will display “Sample Positioned Beyond Fill Mark”. The Action Message: This message indicates the appropriate action. For example, if it is likely the quality check will fail again the next time the analyzer is used, the instruction “Use Electronic Simulator” will be displayed. If the problem is related to an operator or cartridge, the instruction “Use Another Cartridge” will be displayed. The Cause Code: This is a numeric code associated with the failed quality check. Since multiple codes can be associated with a single cause message, this is essential information when contacting i-STAT Technical Services or your local support organization for further assistance. The codes are stored in the analyzer’s memory along with other test records and are transmitted to the Central Data Station. The code list can be viewed and printed. Codes 1-15 and 95 usually indicate a condition related to the environment or the state of the analyzer. These conditions are usually benign and go away after the next cartridge or Electronic Simulator is inserted, or after the offending condition is corrected. Code Number Cause/Action Message on Display Explanation 1 Dead Batteries / Replace Batteries There is insufficient battery power to complete the testing cycle. Replace the disposable lithium batteries in the analyzer or recharge the rechargeable batteries. If you are experiencing this code frequently and use disposable batteries with the i-STAT 1 analyzer, you may want to consider the rechargeable battery system available with the i-STAT 1 Analyzer.

Transcript of i-STAT Technical BulleTin analyser coded...Since multiple codes can be associated ... 22, 25...

Art:714260-01E Rev.Date:08/14/06

Technical BulleTini-STAT®

ANALYZER CODED MESSAGES

Fromthetimeitpowersupuntilthetimeitpowersdown,thei-STAT®Analyzerperformsnumerousqualitychecks.Thefailureofanyqualitycheckcausestheanalyzertohaltthetestcycleanddisplaya“cause”,an“action”message,andacode.

The Cause Message:This message describes the likely cause of the failed quality check. For example, when an overfilledcartridgeisdetected,theanalyzerwilldisplay“SamplePositionedBeyondFillMark”.

The Action Message:Thismessageindicatestheappropriateaction.Forexample,ifitislikelythequalitycheckwillfailagainthenexttimetheanalyzerisused,theinstruction“UseElectronicSimulator”willbedisplayed.Iftheproblemisrelatedtoanoperatororcartridge,theinstruction“UseAnotherCartridge”willbedisplayed.

The Cause Code:This isanumericcodeassociatedwith the failedqualitycheck. Sincemultiplecodescanbeassociatedwithasinglecausemessage,thisisessentialinformationwhencontactingi-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.Thecodesarestoredintheanalyzer’smemoryalongwithothertestrecordsandaretransmittedtotheCentralDataStation.Thecodelistcanbeviewedandprinted.

Codes 1-15 and 95 usually indicate a condition related to the environment or the state of the analyzer. These conditions are usually benign and go away after the next cartridge or Electronic Simulator is inserted, or after the offending condition is corrected.

Code Number

Cause/Action Message on Display Explanation

1 Dead Batteries/ReplaceBatteries

Thereisinsufficientbatterypowertocompletethetestingcycle.Replacethedisposablelithiumbatteriesintheanalyzerorrechargetherechargeablebatteries.

Ifyouareexperiencingthiscodefrequentlyandusedisposablebatterieswiththei-STAT1analyzer,youmaywanttoconsidertherechargeablebatterysystemavailablewiththei-STAT1Analyzer.

Art:714260-01E Rev.Date:08/14/062

Code Number

Cause/Action Message on Display Explanation

2 Temperature Out of Range/CheckStatusPage

Theanalyzerisrecordingatemperatureoutsideitsoperatingrange.Movetheanalyzertoanareawithintheoperatingtemperatureofthetestbeingperformedandallowtheanalyzertocometothenewroomtemperature.Checktheanalyzer’stemperaturereadingontheStatusPage.

3 New Software Installed/UseElectronicSimulator

ThismessageappearsonthePortableClinicalAnalyzerafternewsoftwarehasbeeninstalledor,insomecases,whenanewcustomizationprofileisreceived.

4, 8 Analyzer Interrupted/UseAnotherCartridge

Theanalyzerhasdetectedthatthelasttestcyclewasnotcompleted.Thiscanhappenifthebatterieswereremovedorweremakingpoorcontactwhileacartridgewasstillintheanalyzer.Batteriesthataretooshortwillnotmakepropercontact.Checkthatthebatteriesareinsertedproperlyandseatedwellintheanalyzer;checkthebatteryvoltageontheanalyzer’sStatusPageandreplacebatteriesiflow.NOTE:Patientresultsdisplayedbeforethiscodearevalid.

5, 6, 9 Analyzer Interrupted/ReadyforUse

ThePortableClinicalAnalyzerisunabletorefreshthedisplay.Thiscanhappenifpowerisinterruptedbeforetheanalyzerpowersitselfdown.Checkthatbatteriesareinsertedproperlyandseatedwellintheanalyzer.Batteriesthataretooshortwillnotmakepropercontact.CheckthebatteryvoltageontheStatusPage.

7 Batteries Changed/ReadyforUse

ThisisanormalresponseonthePortableClinicalAnalyzerwhenthebatteriesarechangedafteracode1hasoccurred.

10 Temperature In Range/ReadyforUse

Temperatureisbackinrangefollowingacode2onthePortableClinicalAnalyzer.

11 Date Invalid/CheckClockonStatusPage

Ifthedateintherealtimeclockprecedesthereleasedateprogrammedintotheapplicationsoftware,code11istriggered.Checkthedateontherealtimeclock.

Theaccuracyoftheclockischeckedatthebeginningofacoagulationtest.Iftheclockisinaccurate,Code11istriggered.

12 Invalid or Expired CLEW/SeeManual

TheCLEWstandardizationhasexpired.DownloadavalidCLEW.

ThedateontherealtimeclockexceedstheexpirationdateoftheCLEWsoftware.Checkthedateontherealtimeclock.

13 Invalid or Expired CLEW/SeeManual

TheCLEWiscorruptornotcompatiblewiththeapplicationsoftware(JAMS),orthereisnoCLEWintheanalyzer.DownloadavalidCLEW.IfthiscodeoccursafterasoftwareupgradeandthecustomizationapplicationisenabledintheCDS,changetheCLEWversionintheCustomizationProfiletothelatestversionandre-transmittheprofiletotheanalyzer.

14 Analyzer Error/SeeManual

Customizationprofileiscorrupted.Retransmitthecustomizationprofile.Ifcode14reoccurs,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.

15 Barcode Does Not Match Cartridge Type

Thebarcodescannedbytheuserdoesnotmatchtheimmunoassaycartridgetypeindicatedbytheidentificationchipinthecartridge.Theusershouldrunanothercartridge,beingcarefultoscanthebarcodefromtheportionpackofthespecificcartridgetypebeingrunontheanalyzer.

95 Test Cancelled by Operator

Thismessagewillappearinthestoredtestrecordsonthei-STAT1Analyzeriftheanalyzerpowersdownbeforemandatoryinformationwasentered.

� Art:714260-01E Rev.Date:08/14/06

The following codes are associated with the cartridge or fluid movement within a cartridge. These conditions can be operator or sample related. In most cases, a new cartridge must be used. If a condition persists, especially if isolated to one analyzer, there may an analyzer problem.

Code Number

Cause/Action Message on Display Explanation

19 No Clot Detetced/SeeManual

DuringthePT/INRcycle,noclotwasdetected.Runanothercartridge.Ifcode19reappears,runthesampleonanalternatemethodology.

22, 25 Cartridge Error/UseAnotherCartridge

Thesecodesoccuronlyforcoagulationcartridgesifthemixingofthesampleandreagentiscompromised.Thiscanbecausedbyaninsufficientorclottedsample,orbyairbubblesinthesample.

24 Cartridge Error/UseAnotherCartridge

Theelectricalresistanceofthecalibrantfluid(Rcal)usedtoverifytheelectrolyteconcentrationisoutofspecification.Thiscouldoccurifthecalibrantpackwasrupturedwellbeforethetestallowingevaporationtoresultinahigherelectrolyteconcentration.

Besidestheelectrolyteconcentration,theRcalisalsoaffectedbythetemperatureandtheheightandwidthofthefluidsegmentovertheconductometricsensor.Theanalyzeraccountsforthetemperature,buttheheightandwidthofthefluidsegmentcanvaryfromcartridgelottocartridgelot.Theanalyzerhasbeenprogrammedtocompensatefortheselot-to-lotdifferencesbymaintainingarunningaverageoftheRcalvaluesmeasuredfromthemostrecentcartridgeruns.Occasionally,thedifferencebetweentheRcalvaluesfortwocartridgelotsislargeenoughtocausetheintroductionofanewlottotriggercode24onthefirstfewcartridgeruns.TheCode24errorsshoulddisappearastherunningaverageadjusts.However,ifcode24persistsaftermorethan�cartridgerunsoneachanalyzer,contacti-STATTechnicalServicesoryourlocalsupportorganization.

26 Cartridge Error/UseAnotherCartridge

Thiscodeoccursiftherewasacoagulationspecificqualitycheckfailure:prematuresubstrateactivation,abnormallylowlevelsofsubstrate,orinvalidfluidmotion.

20, 27-29, 32, 33, 40, 41, 45, 87

Cartridge Error/UseAnotherCartridge

Thesecodesidentifyproblemswiththecartridgesuchas:calibrantfluidarrivingtoosoon,toolate,ornotatall,ornoiseinthecalibrantfluidsignals.Codes20,27,and41canbecausedbypoorcontactthatcansometimesbecorrectedbyconditioningthepinsintheanalyzerusingtheceramiccleaningcartridge.Thespecificconditioningprocedureisdescribedattheendofthisbulletin.

42, 43 Cartridge Error/UseAnotherCartridge

Thesecodesindicatethattheconductometricsensor(code42)ortheamperometricsensor(code4�)wasoutofspecification.Thiscouldbecausedbyapre-burstcalibrantpack,dirtycartridgecontactpads,oradirtyconnectorintheanalyzer.

79-81 Cartridge Error/UseAnotherCartridge

Badcontactbetweenthethermalprobesintheanalyzerandthemetalizationonthebackofthechipsinthecartridgetriggerthesecodes.Causesare:poormetalizationofthechips,dirtonthemetalization,orbentorbrokenthermalprobesintheanalyzer.

Art:714260-01E Rev.Date:08/14/064

Code Number

Cause/Action Message on Display Explanation

21 Cartridge Preburst/UseAnotherCartridge

Thiscodeindicatesthattheanalyzerdetectedfluidonthesensorsbeforeitshouldhave.Possiblecauses:mishandlingofcartridges(puttingpressureinthecenterofthecartridge),poorstorageconditionsofcartridges(frozen),orrerunningusedcartridges.

31, 34, 44 Unable to Position Sample/UseAnotherCartridge

Theanalyzerdidnotdetectmovementofsampleacrossthesensors.Thiscouldbeduetoaclotinthesample(especiallyinneonates),tonotclosingthesnapclosureonthecartridge,ortoanaberrantcartridge.

35, 36 Sample Positioned Short of Fill Mark/UseAnotherCartridge

Thecartridgewasunderfilled.Thesamplemustreachthefillmark.Tryanothercartridge.

30, 37 Sample Positioned Beyond Fill Mark/UseAnotherCartridge

Thecartridgewasoverfilled.Thesamplewaspastthefillmark.Tryanothercartridge.

38, 39 Insufficient Sample/UseAnotherCartridge

Thisismostlikelyduetoinsufficientsampleinthesamplewellofthecartridge,butcanalsobecausedbybubblesinthesample.Tryanothercartridgeandensuresufficientsampleisinthesamplewell.

46 Cartridge Error/UseAnotherCartridge

Theanalyzerdidnotdetectmovementofsampleacrossthesensors.Thiscouldbeduetoaclotinthesample(especiallyinneonates),tonotclosingthesnapclosureonthecartridge,ortoanaberrantcartridge.

47 Cartridge Not Inserted Properly/ReinsertCartridge

ThiscodeindicatesthecartridgeorElectronicSimulatormaynotbepushedinalltheway.ReinsertthecartridgeorElectronicSimulator.Iftheproblempersistsand/ortheuseriscertainthecartridgeorSimulatorisproperlyinserted,itmayindicateananalyzerproblem.Contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.

48 Analyzer Error/SeeManual

ThiscodeindicatesthecartridgeorElectronicSimulatormayhavebeen“cocked”wheninserted.PushthecartridgeorSimulatorstraightthroughthecartridgeport.Iftheproblempersists,andtheuseriscertainthecartridgeorSimulatorisproperlyinserted,itmayindicateananalyzerproblem.Contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.

49 Poor Contact Detected/SeeManual

Thesystemdetectedacontactproblemwithoneoftheconnectorpinswhilereadingtheidentificationchipintheimmunoassaycartridge.Thiscansometimesbecorrectedbyconditioningthepinsintheanalyzerusingtheceramicconditioningcartridge.Thespecificconditioningprocedureisdescribedattheendofthisbulletin.

Note:Ifyoudonothaveaceramicconditioningcartridge,pleasecontactyourSupportRepresentative.

� Art:714260-01E Rev.Date:08/14/06

The following conditions are related to electronic or mechanical failures in the analyzer.

Code Number

Cause/Action Message on Display Explanation

50 Analyzer Error/UseElectronicSimulator

Themotorhasmovedtoofar.Runningasimulatormaynotdetectthisproblem.Runthesimulatorandiftheanalyzerpasses,runacartridgetoseeifthecodereoccurs.Ifnot,continuetousetheanalyzer.Ifthecodereoccurs,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.

Iftestingimmunoassaycartridgesonani-STAT1Analyzer,thiscodecanberelatedtopoorelectricalconnectionbetweenthei-STAT1Analyzerandthecartridge.Thiscansometimesbecorrectedbyconditioningthepinsintheanalyzerusingtheceramicconditioningcartridge.Thespecificconditioningprocedureisdescribedattheendofthisbulletin.

Codes126and128aresometimesrelatedtoelectricalconnectionaswell.Ifyouexperiencemultipleoccurrencesofthese�codes(�0,126,and128)inashortperiodoftime,considerreturningtheanalyzerforservicingandreplacement

Thepresenceofsamplebubbleswhenrunningimmunoassaycartridgesmay,undersomecircumstances,alsoelicitthiscode.

51 Analyzer Error/UseElectronicSimulator

Themotormovedfortoolong.Runasimulator.IftheerroroccurredwhilerunninganACTcartridge,alsorunacartridge.Ifthecodedoesnotreoccur,continuetousetheanalyzer.Undersomeconditions,alowbatterywillcausethiserrorinsteadofcode1.Tryfreshbatteries.Ifthecodereoccurs,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.

52 Analyzer Error/UseElectronicSimulator

Themotorstalledwhilemoving.Runasimulator.IftheerroroccurredwhilerunninganACTcartridge,alsorunacartridge.Ifthecodedoesnotreoccur,continuetousetheanalyzer.Ifthecodereoccurs,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.

58-62 Analyzer Error/UseElectronicSimulator

Theanalyzerusuallyrecoversfromtheseerrorconditions.TheseerrorconditionscanbedetectedbytheElectronicSimulator.IftheanalyzerpassestheElectronicSimulatortest,continuetouseit.Ifnot,checkthebatteryvoltageandchecktheanalyzerwithanothersimulatortoruleoutasimulatorproblem.Ifthecodepersists,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.

Art:714260-01E Rev.Date:08/14/066

Code Number

Cause/Action Message on Display Explanation

23, 53, 55, 63, 65-68, 70, 72-74, 82, 85, 86, 89-94, 96, 97

Analyzer Error/SeeManual

Thesearemechanicalorelectronicfailuresfromwhichtheanalyzermaynotbeabletorecover.

Code2�maybecausedbypoorcontactbetweentheanalyzerpinsandthecartridgechip.Thiscansometimesbecorrectedbyconditioningthepinsintheanalyzerusingtheceramiccleaningcartridge.Thespecificconditioningprocedureisdescribedattheendofthisbulletin.

Code70canoccuronthePortableClinicalAnalyzeriftheuserpressestheDISkeybeforetheElectronicSimulatorisrunafterasoftwareupdate.Ifthishappens,reseatthebatteriestoresettheanalyzer,andthenruntheElectronicSimulator.

Codes82and92typicallyindicateaproblemwiththepressuretransducersintheanalyzer.Ifthesecodespersist,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.

Forothercodes,runtheElectronicSimulatortwice,thenrunacartridgewithasample.Iftheanalyzerpassesthesimulatorcheckandaqualitycheckdoesnotoccurwiththesamplerun,continuetousetheanalyzer.Iftheanalyzerdoesnotpassthesimulatorcheckand/oraqualitycodeoccurswiththesamplerun,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.

69 Cartridge Type Not Recognized/UseAnotherCartridge

Thiscodecouldbeduetouseofacartridgetypethatisnotcompatiblewiththeversionofsoftwareintheanalyzer,ortheuseofexpiredcartridges.Checkthecartridgeexpirationdateonthecartridgeboxorpouch.Ifthecartridgeshavenotexpired,andifanewcartridgetypeisbeingrun,contacti-STATTechnicalServicesoryourlocalsupportorganizationforasoftwareupdate.

Whenrunningcoagulationcartridges,Code69maybecausedbypoorcontactbetweentheanalyzerpinsandthecartridgechip.Thiscansometimesbecorrectedbyconditioningthepinsintheanalyzerusingtheceramiccleaningcartridge.Thespecificconditioningprocedureisdescribedattheendofthisbulletin.

Duringimmunoassaycartridgeruns,thiscodewillbedisplayedifincorrectinformationisenteredinresponsetotheprompt“EnterorScanCartridgeLotNumber”.

Theinstrumentexpectsthebarcodeonthebackoftheindividualcartridgeportionpacktobescanned.Thecorrectbarcodelookslikethis:

Forimmunoassaycartridges,theinstrumentwillnotacceptkeypadentriesofthecartridgelotnumbernorascanofthebarcodeonthecartridgebox.

Thisconditionmaybeduetoanaberrantcartridge.However,iftheconditionoccursrepeatedlyononeanalyzer,theanalyzermayneedrepair.Contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.

7 Art:714260-01E Rev.Date:08/14/06

Codes 100-111 indicate a condtion with the PCx or PCx Plus Glucose Test Strip on the i-STAT 1 Analyzer

Code Number

Cause/Action Message on Display Explanation

100 Strip Error/UseAnotherStrip

Theusertriedtorunawetstrip.Removetheteststrip.Press1forTestOptions.Press1forSamePatientorLevel.Repeatthetest.

101 Strip Error/UseAnotherStrip

Theteststripwasremovedfromthestripportduringtesting.Press1forTestOptions.Press1forSamePatientorLevel.Repeatthetest.

103, 105, 106, 107, 111

Strip Error/UseAnotherStrip

Testunsuccessful.Anerrorwasdetectedduringtheanalysissequence.Removetheteststrip.Press1forTestOptions.Press1forSamePatientorLevel.Repeatthetest.Iftheproblempersists,recordthethreedigiterrorcodeandcontactAbbottDiabetesCaresupportservicesoryourlocalsupportorganization.

102, 104 Strip Error/UseAnotherStrip

Theteststripmalfunctionedorthebloodglucoselevelinthesampleisbeyondthemeasuringcapabilityoftheteststripandstripreader.Removetheteststrip.Press1forTestOptions.Press1forSamePatientorLevel.Repeatthetestwithanewteststrip.Iftheerroroccursagain,confirmtheresultbyperformingthetestonadifferentmethod.ContactAbbottDiabetesCaresupportservicesoryourlocalsupportorganization.

108, 109 Temperature Out of Range/CheckStatusPage

Duringthetest,roomtemperaturebecameunstableormovedoutsidethelimitswithinwhichtheteststripreadercanperformatest.Ensurethattheroomtemperatureiswithinthespecifiedlimits.Allowtheanalyzertostabilizetoaroomtemperatureof1�-40ºCor�9-104ºF.PresstheMenukeyuntiltheAdministrationMenuisdisplayed.Press1forAnalyzerStatuswheretheroomtemperaturereadingisdisplayed.

IfthetemperaturereadingontheStatusPageiswithinthelimitsdescribedabove,yetthesecodespersist,theremaybeaproblemwithoneofthethermistorsintheanalyzer.Contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.

110 Strip Error/UseAnotherStrip

Atthestartoftheteststripcycle,theanalyzerpromptstheusertoapplybloodtothestripandwaitsfor20minutesfortheusertodoso.Thiserroroccurswhenthe20minuteshaveelapsedandtheanalyzerdidn’tdetectblood.Themostlikelycauseofthiserrorislackofuserinteraction.

Art:714260-01E Rev.Date:08/14/068

Codes in the range of 120 to 138 and 140 to 151 indicate a failure during an immuno cartridge cycle. In most cases, the cartridge is spent and another cartridge must be used. Only the i-STAT 1 Analyzer produces these codes, as the Portable Clinical Analyzer does not support immuno cycles.

Code Number

Cause/Action Message on Display Explanation

120-122, 124, 125, 133, 144, 148

Cartridge Error /UseAnotherCartridge

Thesecodesindicateaproblemwiththemovementoftheanalysisfluidduringthecartridgerun.Tryanothercartridge.

123 Cartridge Error /UseAnotherCartridge

Thequalitycontrolduringthecartridgerunfailedtoverifythepresenceofactiveimmunoreagents.Tryanothercartridge.

126 Cartridge Error /UseAnotherCartridge

Thiscodecanberelatedtopoorelectricalconnectionbetweenthei-STAT1Analyzerandthecartridge.Thiscansometimesbecorrectedbyconditioningthepinsintheanalyzerusingtheceramicconditioningcartridge.Thespecificconditioningprocedureisdescribedattheendofthisbulletin.

Codes�0and128aresometimesrelatedtoelectricalconnectionaswell.Ifyouexperiencemultipleoccurrencesofthese�codes(�0,126,and128)inashortperiodoftime,considerreturningtheanalyzerforreplacement.

127 Cartridge Error /UseAnotherCartridge

Awetsensorwasdetectedbeforetheinitialsamplemovement.Possibleoverfilledorusedcartridge.Tryanothercartridge.

128, 131, 132, 134, 135 - 138

Cartridge Error /UseAnotherCartridge

Thesecodesaremostoftenrelatedtopoorfillingofanimmunoassaycartridge,thepresenceofsamplebubbles,ortheabruptinsertionofacartridgeintotheanalyzer.

Guidelines for proper filling:

1. Discard(always)1dropfromdeliverydevicetoclearunseenbubbles.

2. Hangsingledropslightlylargerthanroundtargetwell.

�. Touchonedrop(only)toroundtargetwellallowingcartridgetodrawsamplein.

4. ConfirmsamplevolumelinesupwithtopofREDFILLLINEdiagram.

�. Closeslidecoverfromlefttoright.

Guidelines for cartridge insertion:

1. Afterclosingthecartridge,graspthecartridgeclosurebetweenyourfirstfingerandthumb.Thereisarecessforyourthumbintheclosure.

2. Guidethecartridgeintotheanalyzergently,untilasoftclickisheard.

129, 142, 143

Cartridge Error /UseAnotherCartridge

Theanalyzerdetectedanalysisfluidmixedwiththesample.Tryanothercartridge.

130 Cartridge Error /UseAnotherCartridge

Theanalyzerdetectedanairbubbleinthesamplesegment.Tryanothercartridge.

140 Lot Expired Theanalyzerdetectedanexpiredcartridgelot.Checktheexpirationdateandrepeatthetestusinganon-expiredcartridgelot.

9 Art:714260-01E Rev.Date:08/14/06

Code Number

Cause/Action Message on Display Explanation

141 Test Canceled by Operator

Thiscodewillbedisplayedifthecartridgebarcodeisnotscannedwithin60secondsofcartridgeinsertion.Thecorrectbarcodetoscanisthebarcodeonthecartridgeportionpack,nottheoneonthecartridgebox.Anexampleoftheportionpackbarcodeisfoundinthetablelistingforcode69above.

145 Cartridge Error /UseAnotherCartridge

Theanalyzerfailedtodetectfluidarrivalupontheinitialsamplepush.Thismaybecausedbya(n):

• cartridgeleak,

• failuretoclosethecartridgecompletely.Ensurethattheslidecoverisfullyengagedbeforeinsertingthecartridgeintotheanalyzer

• underfilledcartridge.Onceasingledropofsampleistouchedtothetargetwell,immunoassaycartridgeswillfillautomaticallybywickingthesampleatafixedspeed.Tryingtoinjectthesampleintothecartridgeoraddingmoresampletothetargetwellwillnotmakethecartridgefillfaster.Waitforthesampletoreachthe“fillto”mark,andthenclosethecartridge.

146 Cartridge Error /UseAnotherCartridge

Overfilledcartridge.Repeatthetest.

147 Analyzer Error /SeeManual

Inordertorunanimmunoassaycartridge,thei-STAT1Analyzermust:

• bearthe symboland

• becustomizedfor

a) CartridgeBarcodeRequired,or

b) CartridgeInformationFirstRequiredandCartridgeLotNumberRequired.

Ifeitherofthesetwoconditionsabovearenotmet,theanalyzerwilldisplaythiscode.

149 - 151 Cartridge Error/UseAnotherCartridge

Theanalyzerdetectedanatypicaldatastreamfromtheanalyzer.Tryanothercartridge.

Art:714260-01E Rev.Date:08/14/0610

The following conditions are related to the Electronic Simulator

Code Explanation How to Respond

Numerical Code

SeeunderAnalyzerCodedMessages. SeeunderAnalyzerCodedMessages.

L Potentiometricchanneloutoflimits.Canoccurifmoisturecollectsonthecontactpinsinsidetheanalyzerwhentheanalyzerissubjectedtoambienttemperaturechange.

Allowanalyzertoequilibrateinnewenvironmentfor�0minutesandrepeattest.Ifcodereoccurs,returnanlayzer.

G Amperometricchanneloutoflimits.Canoccurifexternalsimulatornotinsertedstraight.

Reinsertthesimulatorstraight.Ifcodereoccurs,returnanalyzer.

R, r Resistancereadingonconductometricchanneloutoflimits.

Returnanalyzer.

t Thermalprobefailure. Returnanalyzer.

B Potentiometricchanneloutoflimits. Returnanalyzer.

NOTE: Anytimerepetitivecodesoccurwhichcannotbeaddressedorcorrectedthroughtraining,contacti-STATTechnicalServicesoryourlocalsupportorganizationforfurtherassistance.

11 Art:714260-01E Rev.Date:08/14/06

PROCEDURE FOR USING AN i-STAT CERAMIC CONDITIONING CARTRIDGE (CCC) FOR ANALyzER PIN CONDITIONING

Step Number Explanation

1.RunanexternalElectronicSimulator. IftheanalyzerisconfiguredwiththeinternalElectronicSimulatorenabled,runanexternalElectronicSimulator.RunningtheexternalElectronicSimulatorensurestheinternalSimulatorcyclewillnotexecuteduringthepinconditioningprocess,whichcouldleadtotheprematureterminationoftheprocess.

2.RuntheCCCtwotimes. InitiatetheCCCcycleasyouwouldinitiateanexternalElectronicSimulatorcycle.TheinstrumentwillidentifytheCCCasanexternalElectronicSimulatoranddisplayaSimulatorFailureCode(i.e.rRGL)whenthecycleiscomplete.Disregardthecode,asthisisexpectedbehavior.

�.UpdatetheCCCUsageLog Thelogislocatedonpage�oftheTechnicalBulletinentitled“InstructionsforRestoringAnalyzersThatProduce***forHematocritandQualityCheckCode2�”,whichisshippedwiththeCCC.UpdatingthelogallowstheusertokeeptrackofthenumberofpinconditioningcyclesperformedwiththecurrentceramicstripintheCCC.Ifnecessary,replaceorrotatetheceramicstripsotheCCCisreadyforfutureuse.

4.Returntheanalyzertoservice.

Art:714260-01E Rev.Date:08/14/0612