Hemodynamics Physics of Blood flow in the circulation.

55
Hemodynamics Physics of Blood flow in the circulation

Transcript of Hemodynamics Physics of Blood flow in the circulation.

Page 1: Hemodynamics Physics of Blood flow in the circulation.

Hemodynamics

Physics of Blood flow in the circulation

Page 2: Hemodynamics Physics of Blood flow in the circulation.

Circulatory System

• Heart:

Has 2 collecting chambers - (Left, Right

Atria)

Has 2 Pumping chambers - (Left, Right

Ventricles)

Page 3: Hemodynamics Physics of Blood flow in the circulation.
Page 4: Hemodynamics Physics of Blood flow in the circulation.

Circulation Schematic

Lungs Tissues

Left Side of Heart

Right Side of Heart

A V

V A

Pulmonary Vein

Pulmonary Artery

Aorta

Sup. & Inf. Vena Cava

Mitral Valve

Pulmonary Valve

Aortic Valve

Tricuspid Valve

Page 5: Hemodynamics Physics of Blood flow in the circulation.

Heart Valves

• Atrioventricular (A-V) valves - separate Atria from Ventricles

• Bicuspid (Mitral) - Left Side

• Tricuspid - Right Side

• Semi-Lunar Valves - separate ventricles from Arteries

Page 6: Hemodynamics Physics of Blood flow in the circulation.

Opening, Closing of Valves - Depends on Pressuredifferences between bloodin adjacent areas

Page 7: Hemodynamics Physics of Blood flow in the circulation.

Heart Sounds

• ‘Lubb’ (1st sound) - Closure of A-V valves• ‘Dupp’ (2nd sound) - Closure of S-L valves

Caused by Turbulence on closing.

Anything extra ’Murmur’ (swishing of blood)

Could be due to:• Stenosis of Valves (calcification)• Valves not closing properly

(Incompetence, Insufficiency)

Increases Pressure on heart

Page 8: Hemodynamics Physics of Blood flow in the circulation.

Heart Sounds and Phonocardiography

• Heart sounds are vibrations or sounds due to the acceleration or deceleration of blood during heart muscle contractions, whereas murmurs (a type of heart sounds) are considered vibrations or sounds due to blood turbulence.

• Phonocardiographyis the recording of heart sounds.

Page 9: Hemodynamics Physics of Blood flow in the circulation.

• Heart Sounds– The auscultation of the heart provides valuable

information to the clinician concerning the functional integrity of the heart.

Page 10: Hemodynamics Physics of Blood flow in the circulation.

Basic heart sounds

Page 11: Hemodynamics Physics of Blood flow in the circulation.

• The first heart sound is generated at the termination of the atrial contractions, just at the onset of ventricular contraction. This sounds is generally attributed to movement of blood into the ventricles, the artioventricular (AV) valves closing, and the sudden cessation of blood flow in the atria. Splitting of the first heart sound is defined as an asynchronous closure of the tricuspid and the mitral valves.

• The second heart sound is a low frequency vibration associated with the closing of the semilunar valves - the aortic and pulmonary valves. This sound is coincident with the completionof the T wave of the ECG.

Page 12: Hemodynamics Physics of Blood flow in the circulation.

• The third heart sound corresponds to the sudden cessation of the ventricular rapidfilling. This low-amplitude, low frequency vibration is audible in children and in some adults.

• The fourth heart sound occurs when the atria contracts and propel blood into the ventricles. This sound with very low amplitude and low frequency is not audible, but may be recorded by the phonocardiography (PCG).

Page 13: Hemodynamics Physics of Blood flow in the circulation.
Page 14: Hemodynamics Physics of Blood flow in the circulation.

• The sources of most murmurs, developed by turbulence in rapidly moving blood, are known. Murmurs are common in children during early systolic phase; they are normally heard in nearly all adults after exercise.

• Abnormal murmurs may caused by stenoses and insufficiencies (leaks) at the aortic, pulmonary, and mitral valves. They are detected by noting the time of their occurrence in the cardiac cycle and their location at the time of measurement.

Page 15: Hemodynamics Physics of Blood flow in the circulation.

Auscultation and Stethoscopes• Heart sounds travel through the body from the

heart and major blood vessels to the body surface.• The physician can hear those sounds with a

stethoscope. • Basic heart sounds occur mostly in the frequency

range of 20 to 200 Hz. • Certain heart murmurs produce sounds in the

1000-Hz region, and some frequency components exist down to 4 or 5 Hz.

• Some researchers even reported that heart sounds and murmurs have small amplitudes with frequencies as low as 0.1 Hz and as high as 2000Hz.

Page 16: Hemodynamics Physics of Blood flow in the circulation.

Stethoscopes: Historical and Current

• Has been used for almost 200 years, and still being used nowadays for screening and diagnosis in primary health care.

Page 17: Hemodynamics Physics of Blood flow in the circulation.

The typical frequency-response curve for a stethoscope

Page 18: Hemodynamics Physics of Blood flow in the circulation.

• Many types of electronic stethoscopes have been proposed by engineers. These devices have selectable frequency-response characteristics ranging from the "ideal" flat-response case and selected bandpass to typical mechanical-stethoscope responses.

• Physicians, however, have not generally accepted these electronic stethoscopes, mainly because they are unfamiliar with the sounds heard with them. Their size, portability, convenience, and resemblance to the mechanical one are other important considerations.

Page 19: Hemodynamics Physics of Blood flow in the circulation.

Phonocardiography• Phonocardiography is an mechano-electronic

recording technique of heart sounds and murmurs. • It is valuable in that it not only eliminates the

subjective interpretation of these sounds, but also makes possible an evaluation of the heart sounds and murmurs with respect to the electrical (such as ECG) and mechanical (carotid pulse recorded in the midneck region) events in the cardiac cycle.

• It is also valuable in locating the sources of various heart sounds.

Page 20: Hemodynamics Physics of Blood flow in the circulation.

• A PCG machine is usually consist of four main parts: – A microphone or PCG transducer, – filtering (mechanical and electrical), – processing unit, and – display.

Page 21: Hemodynamics Physics of Blood flow in the circulation.

• There are optimal recording sites for the various heart sounds or PCG signals. Because of the acoustical properties of the transmission path, heart sound waves are attenuated but not reflected. Figure shows four basic chest locations at which the intensity of sound from the four valves is maximized.

• Auscultatory areas on the chest A, aortic; P, pulmonary; T, tricuspid; and M, mitral areas.

Page 22: Hemodynamics Physics of Blood flow in the circulation.

Blood Vessels

• Arteries• Capillaries• Veins

Systemic Pathway:Left Ventricle Aorta Arteries Arterioles

of Heart Capillaries

Venules Veins Right Atrium of Heart

Page 23: Hemodynamics Physics of Blood flow in the circulation.

Blood

• Composition:– Approx 45% by Vol. Solid Components

» Red Blood Cells (12m x 2 m)

» White Cells

» Platelets

– Approx 55% Liquid (plasma)» 91.5% of which is water

» 7% plasma proteins

» 1.5% other solutes

Page 24: Hemodynamics Physics of Blood flow in the circulation.

Blood Functions

• Transportation

of blood gases, nutrients, wastes

• Homeostasis (regulation)

of Ph, Body Temp, water content

• Protection

Page 25: Hemodynamics Physics of Blood flow in the circulation.

As a Result …….

• Blood behaves as a simple Newtonian Fluid when flowing in blood vessels

i.e. Viscous stresses Viscosity, strain rate

dy

du y u(y)

No slip at wall

Page 26: Hemodynamics Physics of Blood flow in the circulation.

• Viscosity of Blood = 3 3.5 times of water

• Blood acts as a non-newtonian fluid in smaller vessels (including capillaries)

Page 27: Hemodynamics Physics of Blood flow in the circulation.

Cardiac Output• Flow of blood is usually measured in l/min• Total amount of blood flowing through the

circulation = Cardiac Output (CO)

Cardiac Ouput = Stroke Vol. x Heart Rate

= 5 l/min

Influenced by Blood Pressure & Resistance

Force of bloodagainst vessel wall

•Blood viscosity•Vessel Length•Vessel Elasticity•Vasconstriction / Vasodilation

with water retention with dehydration, hemorrage

Page 28: Hemodynamics Physics of Blood flow in the circulation.

Overall

• Greater Pressure Greater Blood

Differences Flow

• Greater Resistance Lesser Blood Flow

Page 29: Hemodynamics Physics of Blood flow in the circulation.

Driving force for blood flow is pressure created by ventricular contraction

Elastic arterial walls expand and recoil

continuous blood flow

Blood PressureBlood Pressure

Page 30: Hemodynamics Physics of Blood flow in the circulation.

Blood pressure is highest in the arteries and falls continuously . . .

Systolic pressure in Aorta: 120 mm Hg

Diastolic pressure in Aorta: 80 mm Hg

Page 31: Hemodynamics Physics of Blood flow in the circulation.

Typical values of circulatory pressures SP is the systolic pressure, DP the diastolic pressure, and MP the mean pressure.

Page 32: Hemodynamics Physics of Blood flow in the circulation.

Ventricular pressure difficult to measure

arterial blood pressure assumed to indicate driving pressure for blood flow

Arterial pressure is pulsatile

useful to have single value for driving pressure: Mean Arterial Pressure

MAP = diastolic P + 1/3 pulse pressure

Page 33: Hemodynamics Physics of Blood flow in the circulation.

Pulse Pressure = systolic pressure - ??

= measure of amplitude of blood pressure wave

Page 34: Hemodynamics Physics of Blood flow in the circulation.

MAP influenced by

• Cardiac output

• Peripheral resistance

MAP CO x Rarterioles

• Blood volume– fairly constant due to homeostatic mechanisms

(kidneys!!)

Page 35: Hemodynamics Physics of Blood flow in the circulation.

BP too low:BP too low:

• Driving force for blood flow unable to overcome gravity

O2 supply to brain

Symptoms?

Page 36: Hemodynamics Physics of Blood flow in the circulation.

BP too high:BP too high:

• Weakening of arterial walls - AneurysmWeakening of arterial walls - Aneurysm

Risk of rupture & hemorrhageRisk of rupture & hemorrhage

Cerebral hemorrhage: ?Cerebral hemorrhage: ?

Rupture of major artery: Rupture of major artery:

Page 37: Hemodynamics Physics of Blood flow in the circulation.

Auscultation of brachial artery with stethoscope

Laminar flow vs. turbulent flow

BP estimated by SphygmomanometryBP estimated by Sphygmomanometry

Page 38: Hemodynamics Physics of Blood flow in the circulation.

Typical indirect blood-pressure measurement system The sphygmomanometer cuff is inflated by a hand bulb to pressures above the systolic level. Pressure is then slowly released, and blood flow under the

cuff is monitored by a microphone or stethoscope placed over a downstream artery. The first Korotkoff sound detected indicates systolic

pressure, whereas the transition from muffling to silence brackets diastolic pressure.

Page 39: Hemodynamics Physics of Blood flow in the circulation.

Principles ofPrinciples of SphygmomanometrySphygmomanometry

Cuff inflated until brachial artery compressed and blood flow stopped what kind of sound?what kind of sound?

Page 40: Hemodynamics Physics of Blood flow in the circulation.

turbulent flow

Slowly release pressure in cuff:

Page 41: Hemodynamics Physics of Blood flow in the circulation.

Pressure at which . . .

. . . sound (= blood flow) first heard:

. . . sound disappeared:

Page 42: Hemodynamics Physics of Blood flow in the circulation.

Ultrasonic determination of blood pressure A compression cuff is placed over the transmitting (8 MHz) and receiving (8 MHz ±ƒ) crystals. The opening and closing of the blood vessel are detected as the applied cuff

pressure is varied.

Page 43: Hemodynamics Physics of Blood flow in the circulation.

• Pressure can be stated in terms of column of fluid.

Pressure Units

mm Hg cm H2O PSI ATM

50 68 0.9 0.065

100 136 1.9 0.13

200 272 3.8 0.26

300 408 5.7 0.39

400 544 7.6 0.52

Page 44: Hemodynamics Physics of Blood flow in the circulation.

Pressure = Height x Density

or P = gh

If Right Atrial pressure = 1 cm H2O in an open column of

blood

Pressure in feet = 140 cm H2O

Rupture

Venous Valves

Density of blood= 1.035 that of water

Incompetent venous valves Varicosities

Actual Pressure in foot= 4-5 cm H2O

Page 45: Hemodynamics Physics of Blood flow in the circulation.

Pressures in the circulation

• Pressures in the arteries, veins and heart chambers are the result of the pumping action of the heart

• The right and left ventricles have similar waveforms but different pressures

• The right and left atria also have similar waveforms with pressures that are similar but not identical

Page 46: Hemodynamics Physics of Blood flow in the circulation.

1. The LV pressure begins to rise after the QRS wave of the ECG

2. Pressure rises until the LV pressure exceeds the aortic pressure

The blood begins to move from the ventricle to the aorta

3. As blood enters the aorta, the aortic pressure begins to rise to form the systolic pulse

4. As the LV pressure falls in late systole the aortic pressure falls until the LV pressure is below the aortic diastolic press.

5. Then the aortic valve closes and LV pressure falls to LA pressure

Page 47: Hemodynamics Physics of Blood flow in the circulation.

•The first wave of atrial pressure (the A wave) is due to atrial contraction•The second wave of atrial pressure (the V wave) is due to ventricular contraction

Page 48: Hemodynamics Physics of Blood flow in the circulation.

Normal Pressures

• RV and pulmonary systolic pressure are 12-15 mm Hg

• Pulmonary diastolic pressure is 6-10 mm Hg• LA pressure is difficult to measure because access

to the LA is not direct

Page 49: Hemodynamics Physics of Blood flow in the circulation.

• The severity of AS is determined by the pressure drop across the aortic valve or by the aortic valve area

• The high velocity of blood flow through the narrowed valve causes turbulence and a characteristic murmur AS can be diagnosed with a stethoscope

AS produces a pressure gradient between the aorta and LV

i.e. For blood to move rapidly through a narrowed aortic valve orifice, the pressure must be higher in the ventricle

Page 50: Hemodynamics Physics of Blood flow in the circulation.

(a) Systolic pressure gradient (left ventricular-aortic pressure) across a stenotic aortic valve. (b) Marked decrease in systolic pressure gradient

with insertion of an aortic ball valve.

Page 51: Hemodynamics Physics of Blood flow in the circulation.
Page 52: Hemodynamics Physics of Blood flow in the circulation.

Pressure Measurement

• Accurate pressure measurements are essential to understanding the status of the circulation

• In 1733 Steven Hales connected a long glass tube directly to the left femoral artery of a horse and measured the height of a column of blood (8 feet, 3 inches) to determine mean BP

• Direct pressure measurements are made frequently in the cardiac catheterization laboratory, the ICU and the OR

Page 53: Hemodynamics Physics of Blood flow in the circulation.

• A tube is inserted into an artery and connected to an electrical strain gauge that converts pressure into force that is sensed electrically

• The output of the transducer is an electrical signal that is amplified and recorded on a strip chart

• For correct pressure measurements the cannula and transducer must be free of air, the cannula should be stiff and short

Page 54: Hemodynamics Physics of Blood flow in the circulation.

Extravascular pressure-sensor system A catheter couples a flush solution (heparinized saline) through a disposable pressure sensor with an integral flush device to the sensing port. The three-way stopcock is used to take

blood samples and zero the pressure sensor.

Sensing port

Roller clamp

Flush solution under pressure

Sample and transducerzero stopcock

Electrical connector

Disposable pressure transducer with an integral flush device

Page 55: Hemodynamics Physics of Blood flow in the circulation.

(a) Physical model of a catheter-sensor system. (b) Analogous electric system for this catheter-sensor system. Each segment of the catheter has its own resistance Rc, inertance Lc, and compliance Cc. In addition, the

sensor has resistor Rs, inertance, Ls, and compliance Cs. The compliance of the diaphragm is Cd.

Diaphragm

Sensor

LsLc RsRcLcRc Lc

P

Rc

Cc Cc Cc Cd =Cs

V

V

P

Liquid Catheter

(a)

(b)

Incrementallength