GCT PREMIER MASS SPECTROMETER
Do You Need:
• Wide dynamic range for characterization of complex mixtures?
• Rapid spectral acquisition rates to profile narrow chromatographic peaks?
• Highly specific exact mass measurement for structural elucidation?
• Selectivity of high resolution to provide low determination limits?
• High sensitivity full scan spectra to maximize sample information?
• A comprehensive suite of post-processing software to maximize productivity and information output?
Take GC/MS to the Highest Level
The Waters® Micromass GCT Premier™ Mass Spectrometer combines the proven elevated resolution and exact mass performance of the classic GCT™ with extended dynamic range. These unmatched capabilities (in addition to the high full scan sensitivity and rapid spectral acquisition rates) are ideal for full characterization and profiling of components in a whole range of sample types.
Use the GCT Premier for Characterization and Profiling in the following Application Areas:
• Metabolomics
• Metabonomics
• Synthetics confirmation and QC
• Natural products research
• Pesticide residues monitoring
• Unknown environmental contaminants ID
• Food research and QC
• Flavors and fragrances research and QC
• Material emissions monitoring and research
• Homeland security chemical monitoring
• Polymer research
• Petrochemicals
• Materials research and monitoring
Sales Offices:Austria 43 1 877 18 07
Australia 61 2 9933 1777
Belgium and Luxembourg 32 2 726 1000
Brazil 55 11 5543 7788
Canada 800 252 4752 X2205
Czech Republic 420 2 617 11384
Denmark 45 46 59 8080Denmark 45 46 59 8080Denmark
Finland 358 9 506 4140
France 33 1 3048 7200
Germany 49 6196 400600Germany 49 6196 400600Germany
Hong Kong 852 29 64 1800
Hungary 36 1 350 5086Hungary 36 1 350 5086Hungary
India 91 80 2837 1900
Ireland 353 1 448 1500
Italy 39 02 27 4211Italy 39 02 27 4211Italy
Japan 81 3 3471 7191
Korea 82 2 820 2700
Mexico 52 55 5524 7636
The Netherlands 31 76 508 7200
Norway 47 6 384 6050Norway 47 6 384 6050Norway
People’s Republic of China 86 10 8451 8918
Poland 48 22 833 4400
Puerto Rico 787 747 8445
Russia/CIS 7 095 931 9193
Singapore 65 6278 7997
Spain 34 93 600 9300
Sweden 46 8 555 11 500
Switzerland 41 62 889 2030
Taiwan 886 2 2543 1898
UK 44 208 238 6100UK 44 208 238 6100UK
US 800 252 4752
WATERS CORPORATION34 Maple St.Milford, MA 01757 U.S.A.T: 508 478 2000F: 508 872 1990www.waters.com
Waters, GCT, GCT Premier, ChromaLynx, MarkerLynx, TargetLynx, OpenLynx and Connections are trademarks of Waters Corporation.All other trademarks are property of their respective owners.©2005 Waters Corporation Printed in the U.S.A. June 2005 720001166EN AG&LW-CV
Waters Micromass GCT Premier Mass Spectrometer
Waters: For Complete Confidence
Waters Corporation is the world's leading provider of instruments, chemistries, software, and support services for chromatography, mass spectrometry, thermal analysis and laboratory informatics.
For more than 45 years, Waters has been supplying scientists with the analytical tools that allow them to better understand the secrets of chemistry and of life itself. Around the globe, Waters systems and solutions are used by pharmaceutical, life science, proteomics, food and beverage, environmental, clinical, industrial, and biotechnology researchers to push the boundaries of analytical science.
To learn more visit www.waters.com
Additional Techniques Provide Enhanced Versatility
The variety of ionization and probe techniques available on the GCT Premier Mass Spectrometermaximizes the amount of information that can be obtained across a wide range of sample types.
Figure 6. EI and CI are complementary techniques, providingstructural and molecular weight information.
Figure 7. FI is the softest ionization technique for generating molecular weight information on most compound types.
Figure 8. DCI is a softer alternative to solids probe CI for thermally labile compounds. The molecular ion can be seen at m/z 380.
Enhance Your Lab’s GC/MS Capabilities with the GCT Premier Technology Platform
The GCT Premier is a powerful orthogonal acceleration time-of-flight (oa-TOF) GC/MS that offers fast acquisition of full spectrum data with high sensitivity, extended dynamic range and elevated resolution. The GCT Premier’s elevated resolution minimizes the likelihood of mass interference. Exact mass capability, which leads to elemental composition determination, offers another dimension over traditional GC/MS.
With the GCT Premier, your lab will benefit from:
Compact Size: Powerful oa-TOF with reflectron technology minimizing linear bench space without sacrificing performance
Elevated Resolution: 7000 FWHM
Exact Mass Measurement: Sub-5 ppm RMS with internal lock mass correction
Sensitivity: High duty cycle compared to scanning instruments providing low- and sub-picogram sensitivity with full spectra
Dynamic Range: Enhanced up to four orders for both exact measurement and quantitative analysis
High Spectral Sampling Rate: Ion beam sampled up to 25,000 times per second eliminating spectral skew associated with traditional scanning instruments
High Spectral Acquisition Rate: Up to 20 spectra per second can be acquired to disk, enabling narrow chromatographic peaks to be accommodated and enhancing the performance of spectral deconvolution routines
Versatile Range of Sources and Probes: A full range of source and probe options are available, including Electron Impact (EI), Chemical Ionization (CI), Desorption Chemical Ionization (DCI), Field Ionization (FI), Field Desorption (FD) and Direct Insertion Probe (DIP)
GC and AutoSampler Control: Integrated control of Agilent®
6890N GC and Agilent 7683, and CTC-GCPal™ and CTC-CombiPal™ autosamplers
ChromaLynx™ Application Manager: Facilitates the rapid detection, identification and semi-quantitative determination of components in complex mixtures while enabling the comparison of sample and control extracts
MarkerLynx™ db Application Manager: Dynamic statistical processing and visualization of complex data sets for metabolomic and metabonomic applications
TargetLynx™ Application Manager: Dedicated target compound quantitation software allowing easy method setup and a full range of regulatory QC checks
OpenLynx™ Application Manager: Automated exact mass processing for batch confirmation of synthetics
Figure 1. Proven oa-TOF technology in a compact bench top GC/MS system.
Figure 2. Exact mass MS measurement enables compounds to be more easily identified. Confirming a measured m/z to within 5 ppm significantly increases confidence in assigning elemental composition.
Maximize Detection Capabilities with Dynamic Range Enhancement (DRE)
The advantage of the full mass range spectra produced by the GCT Premier enables a wide range of components, both known and unknown, to be detected with high sensitivity. Full mass range spectra are beneficial when trying to characterize samples in terms of identity and quantity.
The extended dynamic range of the GCT Premier delivers up to four orders of magnitude for exact mass measurement and quantitative studies allowing you to maximize information produced from a single analysis. Figure 3. This example of hexachlorobenzene (500 fg
to 8 ng on column) demonstrates >4 orders of linear dynamic range with DRE enabled.
Compound 1 thujopsene
Compound 3 cyclamen aldehydeCompound 2 fruitate
Figure 4. Example of spectral deconvolution of overlapping peaks in a complex chromatogram. This processing was performed using ChromaLynx Application Manager.
Increase Productivity with Rapid Spectral Sampling and Acquisition Rate
The GCT Premier samples the source ion beam at 25,000 times per second. This eliminates the spectral skew associated with scanning instruments, which means that accurate spectra and quantitative data can be obtained with fewer scans over the chromatographic peak. In addition, the high spectral acquisition rate (up to 20 spectra per second) enables narrow chromatographic peaks to be processed. In combination with Waters deconvolution algorithm, pure spectra can be obtained from partially co-eluting components in complex chromatograms.
Figure 5. Illustration of the absence of spectral skew from the GCT Premier Mass Spectrometer as chromatographic intensity changes.
Enhance Your Lab’s GC/MS Capabilities with the GCT Premier Technology Platform
The GCT Premier is a powerful orthogonal acceleration time-of-flight (oa-TOF) GC/MS that offers fast acquisition of full spectrum data with high sensitivity, extended dynamic range and elevated resolution. The GCT Premier’s elevated resolution minimizes the likelihood of mass interference. Exact mass capability, which leads to elemental composition determination, offers another dimension over traditional GC/MS.
With the GCT Premier, your lab will benefit from:
Compact Size: Powerful oa-TOF with reflectron technology minimizing linear bench space without sacrificing performance
Elevated Resolution: 7000 FWHM
Exact Mass Measurement: Sub-5 ppm RMS with internal lock mass correction
Sensitivity: High duty cycle compared to scanning instruments providing low- and sub-picogram sensitivity with full spectra
Dynamic Range: Enhanced up to four orders for both exact measurement and quantitative analysis
High Spectral Sampling Rate: Ion beam sampled up to 25,000 times per second eliminating spectral skew associated with traditional scanning instruments
High Spectral Acquisition Rate: Up to 20 spectra per second can be acquired to disk, enabling narrow chromatographic peaks to be accommodated and enhancing the performance of spectral deconvolution routines
Versatile Range of Sources and Probes: A full range of source and probe options are available, including Electron Impact (EI), Chemical Ionization (CI), Desorption Chemical Ionization (DCI), Field Ionization (FI), Field Desorption (FD) and Direct Insertion Probe (DIP)
GC and AutoSampler Control: Integrated control of Agilent®
6890N GC and Agilent 7683, and CTC-GCPal™ and CTC-CombiPal™ autosamplers
ChromaLynx™ Application Manager: Facilitates the rapid detection, identification and semi-quantitative determination of components in complex mixtures while enabling the comparison of sample and control extracts
MarkerLynx™ db Application Manager: Dynamic statistical processing and visualization of complex data sets for metabolomic and metabonomic applications
TargetLynx™ Application Manager: Dedicated target compound quantitation software allowing easy method setup and a full range of regulatory QC checks
OpenLynx™ Application Manager: Automated exact mass processing for batch confirmation of synthetics
Figure 1. Proven oa-TOF technology in a compact bench top GC/MS system.
Figure 2. Exact mass MS measurement enables compounds to be more easily identified. Confirming a measured m/z to within 5 ppm significantly increases confidence in assigning elemental composition.
Maximize Detection Capabilities with Dynamic Range Enhancement (DRE)
The advantage of the full mass range spectra produced by the GCT Premier enables a wide range of components, both known and unknown, to be detected with high sensitivity. Full mass range spectra are beneficial when trying to characterize samples in terms of identity and quantity.
The extended dynamic range of the GCT Premier delivers up to four orders of magnitude for exact mass measurement and quantitative studies allowing you to maximize information produced from a single analysis. Figure 3. This example of hexachlorobenzene (500 fg
to 8 ng on column) demonstrates >4 orders of linear dynamic range with DRE enabled.
Compound 1 thujopsene
Compound 3 cyclamen aldehydeCompound 2 fruitate
Figure 4. Example of spectral deconvolution of overlapping peaks in a complex chromatogram. This processing was performed using ChromaLynx Application Manager.
Increase Productivity with Rapid Spectral Sampling and Acquisition Rate
The GCT Premier samples the source ion beam at 25,000 times per second. This eliminates the spectral skew associated with scanning instruments, which means that accurate spectra and quantitative data can be obtained with fewer scans over the chromatographic peak. In addition, the high spectral acquisition rate (up to 20 spectra per second) enables narrow chromatographic peaks to be processed. In combination with Waters deconvolution algorithm, pure spectra can be obtained from partially co-eluting components in complex chromatograms.
Figure 5. Illustration of the absence of spectral skew from the GCT Premier Mass Spectrometer as chromatographic intensity changes.
Enhance Your Lab’s GC/MS Capabilities with the GCT Premier Technology Platform
The GCT Premier is a powerful orthogonal acceleration time-of-flight (oa-TOF) GC/MS that offers fast acquisition of full spectrum data with high sensitivity, extended dynamic range and elevated resolution. The GCT Premier’s elevated resolution minimizes the likelihood of mass interference. Exact mass capability, which leads to elemental composition determination, offers another dimension over traditional GC/MS.
With the GCT Premier, your lab will benefit from:
Compact Size: Powerful oa-TOF with reflectron technology minimizing linear bench space without sacrificing performance
Elevated Resolution: 7000 FWHM
Exact Mass Measurement: Sub-5 ppm RMS with internal lock mass correction
Sensitivity: High duty cycle compared to scanning instruments providing low- and sub-picogram sensitivity with full spectra
Dynamic Range: Enhanced up to four orders for both exact measurement and quantitative analysis
High Spectral Sampling Rate: Ion beam sampled up to 25,000 times per second eliminating spectral skew associated with traditional scanning instruments
High Spectral Acquisition Rate: Up to 20 spectra per second can be acquired to disk, enabling narrow chromatographic peaks to be accommodated and enhancing the performance of spectral deconvolution routines
Versatile Range of Sources and Probes: A full range of source and probe options are available, including Electron Impact (EI), Chemical Ionization (CI), Desorption Chemical Ionization (DCI), Field Ionization (FI), Field Desorption (FD) and Direct Insertion Probe (DIP)
GC and AutoSampler Control: Integrated control of Agilent®
6890N GC and Agilent 7683, and CTC-GCPal™ and CTC-CombiPal™ autosamplers
ChromaLynx™ Application Manager: Facilitates the rapid detection, identification and semi-quantitative determination of components in complex mixtures while enabling the comparison of sample and control extracts
MarkerLynx™ db Application Manager: Dynamic statistical processing and visualization of complex data sets for metabolomic and metabonomic applications
TargetLynx™ Application Manager: Dedicated target compound quantitation software allowing easy method setup and a full range of regulatory QC checks
OpenLynx™ Application Manager: Automated exact mass processing for batch confirmation of synthetics
Figure 1. Proven oa-TOF technology in a compact bench top GC/MS system.
Figure 2. Exact mass MS measurement enables compounds to be more easily identified. Confirming a measured m/z to within 5 ppm significantly increases confidence in assigning elemental composition.
Maximize Detection Capabilities with Dynamic Range Enhancement (DRE)
The advantage of the full mass range spectra produced by the GCT Premier enables a wide range of components, both known and unknown, to be detected with high sensitivity. Full mass range spectra are beneficial when trying to characterize samples in terms of identity and quantity.
The extended dynamic range of the GCT Premier delivers up to four orders of magnitude for exact mass measurement and quantitative studies allowing you to maximize information produced from a single analysis. Figure 3. This example of hexachlorobenzene (500 fg
to 8 ng on column) demonstrates >4 orders of linear dynamic range with DRE enabled.
Compound 1 thujopsene
Compound 3 cyclamen aldehydeCompound 2 fruitate
Figure 4. Example of spectral deconvolution of overlapping peaks in a complex chromatogram. This processing was performed using ChromaLynx Application Manager.
Increase Productivity with Rapid Spectral Sampling and Acquisition Rate
The GCT Premier samples the source ion beam at 25,000 times per second. This eliminates the spectral skew associated with scanning instruments, which means that accurate spectra and quantitative data can be obtained with fewer scans over the chromatographic peak. In addition, the high spectral acquisition rate (up to 20 spectra per second) enables narrow chromatographic peaks to be processed. In combination with Waters deconvolution algorithm, pure spectra can be obtained from partially co-eluting components in complex chromatograms.
Figure 5. Illustration of the absence of spectral skew from the GCT Premier Mass Spectrometer as chromatographic intensity changes.
GCT PREMIER MASS SPECTROMETER
Do You Need:
• Wide dynamic range for characterization of complex mixtures?
• Rapid spectral acquisition rates to profile narrow chromatographic peaks?
• Highly specific exact mass measurement for structural elucidation?
• Selectivity of high resolution to provide low determination limits?
• High sensitivity full scan spectra to maximize sample information?
• A comprehensive suite of post-processing software to maximize productivity and information output?
Take GC/MS to the Highest Level
The Waters® Micromass GCT Premier™ Mass Spectrometer combines the proven elevated resolution and exact mass performance of the classic GCT™ with extended dynamic range. These unmatched capabilities (in addition to the high full scan sensitivity and rapid spectral acquisition rates) are ideal for full characterization and profiling of components in a whole range of sample types.
Use the GCT Premier for Characterization and Profiling in the following Application Areas:
• Metabolomics
• Metabonomics
• Synthetics confirmation and QC
• Natural products research
• Pesticide residues monitoring
• Unknown environmental contaminants ID
• Food research and QC
• Flavors and fragrances research and QC
• Material emissions monitoring and research
• Homeland security chemical monitoring
• Polymer research
• Petrochemicals
• Materials research and monitoring
Sales Offices:Austria 43 1 877 18 07
Australia 61 2 9933 1777
Belgium and Luxembourg 32 2 726 1000
Brazil 55 11 5543 7788
Canada 800 252 4752 X2205
Czech Republic 420 2 617 11384
Denmark 45 46 59 8080Denmark 45 46 59 8080Denmark
Finland 358 9 506 4140
France 33 1 3048 7200
Germany 49 6196 400600Germany 49 6196 400600Germany
Hong Kong 852 29 64 1800
Hungary 36 1 350 5086Hungary 36 1 350 5086Hungary
India 91 80 2837 1900
Ireland 353 1 448 1500
Italy 39 02 27 4211Italy 39 02 27 4211Italy
Japan 81 3 3471 7191
Korea 82 2 820 2700
Mexico 52 55 5524 7636
The Netherlands 31 76 508 7200
Norway 47 6 384 6050Norway 47 6 384 6050Norway
People’s Republic of China 86 10 8451 8918
Poland 48 22 833 4400
Puerto Rico 787 747 8445
Russia/CIS 7 095 931 9193
Singapore 65 6278 7997
Spain 34 93 600 9300
Sweden 46 8 555 11 500
Switzerland 41 62 889 2030
Taiwan 886 2 2543 1898
UK 44 208 238 6100UK 44 208 238 6100UK
US 800 252 4752
WATERS CORPORATION34 Maple St.Milford, MA 01757 U.S.A.T: 508 478 2000F: 508 872 1990www.waters.com
Waters, GCT, GCT Premier, ChromaLynx, MarkerLynx, TargetLynx, OpenLynx and Connections are trademarks of Waters Corporation.All other trademarks are property of their respective owners.©2005 Waters Corporation Printed in the U.S.A. June 2005 720001166EN AG&LW-CV
Waters Micromass GCT Premier Mass Spectrometer
Waters: For Complete Confidence
Waters Corporation is the world's leading provider of instruments, chemistries, software, and support services for chromatography, mass spectrometry, thermal analysis and laboratory informatics.
For more than 45 years, Waters has been supplying scientists with the analytical tools that allow them to better understand the secrets of chemistry and of life itself. Around the globe, Waters systems and solutions are used by pharmaceutical, life science, proteomics, food and beverage, environmental, clinical, industrial, and biotechnology researchers to push the boundaries of analytical science.
To learn more visit www.waters.com
Additional Techniques Provide Enhanced Versatility
The variety of ionization and probe techniques available on the GCT Premier Mass Spectrometermaximizes the amount of information that can be obtained across a wide range of sample types.
Figure 6. EI and CI are complementary techniques, providingstructural and molecular weight information.
Figure 7. FI is the softest ionization technique for generating molecular weight information on most compound types.
Figure 8. DCI is a softer alternative to solids probe CI for thermally labile compounds. The molecular ion can be seen at m/z 380.
GCT PREMIER MASS SPECTROMETER
Do You Need:
• Wide dynamic range for characterization of complex mixtures?
• Rapid spectral acquisition rates to profile narrow chromatographic peaks?
• Highly specific exact mass measurement for structural elucidation?
• Selectivity of high resolution to provide low determination limits?
• High sensitivity full scan spectra to maximize sample information?
• A comprehensive suite of post-processing software to maximize productivity and information output?
Take GC/MS to the Highest Level
The Waters® Micromass GCT Premier™ Mass Spectrometer combines the proven elevated resolution and exact mass performance of the classic GCT™ with extended dynamic range. These unmatched capabilities (in addition to the high full scan sensitivity and rapid spectral acquisition rates) are ideal for full characterization and profiling of components in a whole range of sample types.
Use the GCT Premier for Characterization and Profiling in the following Application Areas:
• Metabolomics
• Metabonomics
• Synthetics confirmation and QC
• Natural products research
• Pesticide residues monitoring
• Unknown environmental contaminants ID
• Food research and QC
• Flavors and fragrances research and QC
• Material emissions monitoring and research
• Homeland security chemical monitoring
• Polymer research
• Petrochemicals
• Materials research and monitoring
Sales Offices:Austria 43 1 877 18 07
Australia 61 2 9933 1777
Belgium and Luxembourg 32 2 726 1000
Brazil 55 11 5543 7788
Canada 800 252 4752 X2205
Czech Republic 420 2 617 11384
Denmark 45 46 59 8080Denmark 45 46 59 8080Denmark
Finland 358 9 506 4140
France 33 1 3048 7200
Germany 49 6196 400600Germany 49 6196 400600Germany
Hong Kong 852 29 64 1800
Hungary 36 1 350 5086Hungary 36 1 350 5086Hungary
India 91 80 2837 1900
Ireland 353 1 448 1500
Italy 39 02 27 4211Italy 39 02 27 4211Italy
Japan 81 3 3471 7191
Korea 82 2 820 2700
Mexico 52 55 5524 7636
The Netherlands 31 76 508 7200
Norway 47 6 384 6050Norway 47 6 384 6050Norway
People’s Republic of China 86 10 8451 8918
Poland 48 22 833 4400
Puerto Rico 787 747 8445
Russia/CIS 7 095 931 9193
Singapore 65 6278 7997
Spain 34 93 600 9300
Sweden 46 8 555 11 500
Switzerland 41 62 889 2030
Taiwan 886 2 2543 1898
UK 44 208 238 6100UK 44 208 238 6100UK
US 800 252 4752
WATERS CORPORATION34 Maple St.Milford, MA 01757 U.S.A.T: 508 478 2000F: 508 872 1990www.waters.com
Waters, GCT, GCT Premier, ChromaLynx, MarkerLynx, TargetLynx, OpenLynx and Connections are trademarks of Waters Corporation.All other trademarks are property of their respective owners.©2005 Waters Corporation Printed in the U.S.A. June 2005 720001166EN AG&LW-CV
Waters Micromass GCT Premier Mass Spectrometer
Waters: For Complete Confidence
Waters Corporation is the world's leading provider of instruments, chemistries, software, and support services for chromatography, mass spectrometry, thermal analysis and laboratory informatics.
For more than 45 years, Waters has been supplying scientists with the analytical tools that allow them to better understand the secrets of chemistry and of life itself. Around the globe, Waters systems and solutions are used by pharmaceutical, life science, proteomics, food and beverage, environmental, clinical, industrial, and biotechnology researchers to push the boundaries of analytical science.
To learn more visit www.waters.com
Additional Techniques Provide Enhanced Versatility
The variety of ionization and probe techniques available on the GCT Premier Mass Spectrometermaximizes the amount of information that can be obtained across a wide range of sample types.
Figure 6. EI and CI are complementary techniques, providingstructural and molecular weight information.
Figure 7. FI is the softest ionization technique for generating molecular weight information on most compound types.
Figure 8. DCI is a softer alternative to solids probe CI for thermally labile compounds. The molecular ion can be seen at m/z 380.