FUNDUS MODULE 300Quick guide – Image capture step-by-step
Image capture step-by-stepPreparing slit lamp
■ Move microscope and illumination tower to the right until stop position (for BM 900 set to 60°)
Installing Fundus Module 300 ■ Lift Fundus Module 300 from cradle (if Fundus Module 300 has been turned off, press ON button for starting)
■ Latch Fundus Module 300 to base plate on microscope
Preparing Fundus Module 300 for capture
■ Move Fundus Module 300 forward until stop position ■ Select ‘manual mode’ and enter patient’s refractive error
Start capture ■ Start examination AFTER Windows recognizes Fundus Module 300 in EyeSuite (wait for Windows connection sound or pop-up)
■ Move Fundus Module 300 towards patients eye ■ Centre retina and move in until the picture detail of IR image shows section of interest
Capturing ■ Press release bar on control panel or capture button on Fundus Module 300
■ Captured image is shown briefly then transmitted to EyeSuite
How to avoid image artefacts Disturbing re-flexes in lower part
Reason: Camera is not close enough
Action: Move camera slightly closer to patient’s eye
Image is out of focus
Reason: Camera focus was not on retina
Action: Enter patient’s refractive error to bring focus on retina
Disturbing re-flexes in upper part
Reason: Camera is too close
Action: Move camera slightly fur-ther from patient’s eye
Picture detail incomplete
Reason: Patient’s pupil was too small
Action: Darken room further or if necessary dilate patient’s eye
Image too dark
Reason: Flash intensity is too low
Action: Increase flash intensity
Poor detail despite sharp image
Reason: Image was captured through hazy media
Action: Best im-ages can only be achieved through relatively clear media
Image too bright
Reason: Flash intensity is too high
Action: Reduce flash intensity
Ideal image
HS-Art.No. 1504.7020040.02020
©HAAG-STREIT AG, 3098 Koeniz, Switzerland2. Edition / 2019 – 01
HAAG-STREIT Holding AGwww.haag-streit-holding.com
HAAG-STREIT AG, Diagnosticswww.haag-streit.com
HAAG-STREIT AG, Verkauf Schweizwww.haag-streit.ch
HAAG-STREIT Deutschland GmbHwww.haag-streit.de
HAAG-STREIT Far Eastwww.haag-streit-fareast.com
HAAG-STREIT Medtech AGwww.haag-streit-medtech.com
HAAG-STREIT Surgical GmbHwww.haag-streit-surgical.com
HAAG-STREIT UKwww.haag-streit-uk.com
HAAG-STREIT USAwww.haag-streit-usa.com
CLEMENT CLARKE Ltd.www.clement-clarke.com
HS DOMS GmbHwww.hs-doms.com
IPRO GmbHwww.ipro.de
John Weiss Ltd.www.johnweiss.com
Möller-Wedel GmbH & Co KGwww.haag-streit-surgical.com
Möller-Wedel Optical GmbHwww.moeller-wedel-optical.com
OptoMedical Technologies GmbHwww.haag-streit-surgical.com
Reliance Medical Inc.www.haag-streit-usa.com
SPECTROS AGwww.spectros.ch
Members of HAAG-STREIT Group
HAAG-STREIT AGGartenstadtstrasse 103098 KoenizSwitzerlandPhone +41 31 978 01 11Fax +41 31 978 02 [email protected]
Distributed by:
Manufactured by:Optomed OyYrttipellontie 190230 Oulu, Finland
0598