From CCD to EMCCD Scientific imaging for today’s microscopy.

55
From CCD to EMCCD Scientific imaging for today’s microscopy

Transcript of From CCD to EMCCD Scientific imaging for today’s microscopy.

Page 1: From CCD to EMCCD Scientific imaging for today’s microscopy.

From CCD to EMCCDScientific imaging for today’s microscopy

Page 2: From CCD to EMCCD Scientific imaging for today’s microscopy.

From CCD to EMCCD

Is this a right topic?

Who should care what?

The CCD manufactories: Sony, Kodak, Texas Instruments, e2V ….

2. Cameras: Read noise, cooling, interface, speed, sensitivities …..

MAG !!

3. Imaging Systems:

1. CCD, CMOS, EMCCD, Interline, Color, Frame Transfer……

System Integrators, LIN Trading !!

Page 3: From CCD to EMCCD Scientific imaging for today’s microscopy.

From CCD to EMCCD

Is this a right topic?

What actually do you/researchers care about?

Beautiful image!!!

Publication and

Quantization !!!

Page 4: From CCD to EMCCD Scientific imaging for today’s microscopy.

From CCD to EMCCD

The right topic

What make a good image?

• better resolution

• higher Signal-Noise Ratio

• Good Contrast (Dynamic Range)

Page 5: From CCD to EMCCD Scientific imaging for today’s microscopy.

• Resolution (Digital Resolution, Spatial Resolution)

• Signal-Noise Ratio

• Contrast / Dynamic Range

From CCD to EMCCD

Revised topic: important characters of digital image

Page 6: From CCD to EMCCD Scientific imaging for today’s microscopy.

electronics

CCD / CMOS

image sensor

Camera (CCD?)

Page 7: From CCD to EMCCD Scientific imaging for today’s microscopy.

• Resolution (Digital Resolution, Spatial Resolution)

• Signal-Noise Ratio

• Contrast (Dynamic Range)

From CCD to EMCCD

Revised topic: important characters of digital image

Page 8: From CCD to EMCCD Scientific imaging for today’s microscopy.

From CCD to EMCCD

Digital Resolution

Page 9: From CCD to EMCCD Scientific imaging for today’s microscopy.

Approach:

1. Bigger Chip;

High cost for high grade chips

2. Smaller Pixels;

Lower sensitivity (Signal/Noise Ratio)

3. Micro Scanning

Good Balance of the above, slow speed.

From CCD to EMCCD

Digital Resolution

Page 10: From CCD to EMCCD Scientific imaging for today’s microscopy.

Useful resolution for microscopy

Camera Resolving Power > Optical Resolving Power

1. Specimen details resolved by the objectives, need to be acquired by the camera ;2. Avoid “empty resolution“, empty resolution only create unnecessary large files.

Tip to be remembered:Camera resolution should match optical resolution;Low magnifications normally require higher camera resolutions.

From CCD to EMCCD

Digital Resolution : The higher, the better ?

Page 11: From CCD to EMCCD Scientific imaging for today’s microscopy.

25 mm

2/3 “ Chip

Adaptation 1,0 x 18, 3% of view area 20

Adaptation 0,5 x , 58% of view area 20

sensor size 8,8 mm x 6,6 mm

8.8mm6

.6m

m

1.22 x N.A.Objective + N.A.Condenser

d0 =

Field of View = Field Number/mag.

Requested resolution = 2* FOV / d0

Requested Resolution

The higher, the better?

Page 12: From CCD to EMCCD Scientific imaging for today’s microscopy.

1435 x 108184714 x 53842 1,4100

921 x 69354459 x 34627 0,9100

2270 x 17091341139 x 85867 1,463

3320 x 24991951666 x 1254981,340

1906 x 1435112952 x 717560,7540

3255 x 24511911632 x 1229960,8025

3830 x 28862251921 x 14461130,7520

2548 x 19181501275 x 960 750,520

5097 x 38373002550 x 19201500,510

2548 x 19181501275 x 960750,2510

3063 x 23051801530 x 1152900,155

4905 x 36932882448 x 18431440,122,5

3210 x 24511921632 x 1229960,041,25

Necessary camera

resolution

Lines/mm(TV- 0,5 x)

Necessary camera

resolution

Lines/mm(TV-1.0 x)

N.A.Magnificati

on

Nyquist Theorem: Sampling frequency should be double the frequency of the signal.

From CCD to EMCCD

Digital Resolution: the higher, the better?

Page 13: From CCD to EMCCD Scientific imaging for today’s microscopy.

1. GREEN resolution/QE ->50%;2. RED resolution/QE -> 25%; 3. BLUE resolution/QE -> 25%; 4. The color interpolation decreases camera resolution;

4.2 Pixels

From CCD to EMCCD

Digital Resolution: Color or Mono?

1 Pixels

Page 14: From CCD to EMCCD Scientific imaging for today’s microscopy.

Exposure Time: Color 3.34 ms VS. Mono 0.9 ms

From CCD to EMCCD

Digital Resolution: Color or Mono?

Page 15: From CCD to EMCCD Scientific imaging for today’s microscopy.

• Resolution (Digital Resolution, Spatial Resolution)

• Signal-Noise Ratio

• Contrast (Dynamic Range)

From CCD to EMCCD

Revised topic: important characters of digital image

Page 16: From CCD to EMCCD Scientific imaging for today’s microscopy.

Important Camera Specs affect Signal-Noise Ratio

1. Quantum Efficiency: higher signal

2. Noise: Photon noise, readout noise, dark current

3. Signal-Noise Ratio, Camera sensitivity

From CCD to EMCCD

Signal-Noise Ratio: Low light considerations

Page 17: From CCD to EMCCD Scientific imaging for today’s microscopy.

Quantum Efficiency : The Spectral Response / Photon to Electron converting efficiency

From CCD to EMCCD

Signal-Noise Ratio: Quantum Efficiency

Page 18: From CCD to EMCCD Scientific imaging for today’s microscopy.

Front vs Backside Illuminated CCD

From CCD to EMCCD

Signal-Noise Ratio: Quantum Efficiency

Page 19: From CCD to EMCCD Scientific imaging for today’s microscopy.

– Photon-induced shot noise– Readout noise– Dark current noise

• Total System Noise = all noise sources added in quadrature

Main Noise Sources in CCDs

From CCD to EMCCD

Signal-Noise Ratio: Noise

Page 20: From CCD to EMCCD Scientific imaging for today’s microscopy.

Photon Noise (Shot Noise)

- Law of physics- Square root relationship between signal and noise

noise = square root of number of electrons- Poisson distribution- When photon noise exceeds system noise, data is photon (shot) noise limited

- Law of physics- Square root relationship between signal and noise

Photon noise = √Signal electrons

- Poisson distribution- When photon noise exceeds system noise, image data is photon (shot) noise limited

From CCD to EMCCD

Signal-Noise Ratio: Noise

Page 21: From CCD to EMCCD Scientific imaging for today’s microscopy.

Serial Register

Preamplifier

Output NodeActive Array

ADC

From CCD to EMCCD

Signal-Noise Ratio: Noise

CCD Readout

Page 22: From CCD to EMCCD Scientific imaging for today’s microscopy.

Read Noise (preamplifier noise)

- Higher readout speed leads to higher Read Noise; example: Readout speed = 1 MHZ, Readout Noise = 3 e; -> 0.5 frame/second Readout speed = 20 MHZ, Readout Noise = 8 e; -> 10 frames/second

- Minimized by careful electronic design;

- Under low-light/low-signal conditions where read noise exceeds photon noise, data is read noise limited

- Read noise not as relevant in high-signal applications

From CCD to EMCCD

Signal-Noise Ratio: Noise

Page 23: From CCD to EMCCD Scientific imaging for today’s microscopy.

Dark current: • Electrons created by thermal emission;• Increases with time and temperature;• Cooling CCD reduces Dark Current;

Dark current is cut in half as the CCD temperature drops approximately every 6.7° C• Reduced by utilizing multi-pinned-phase (MPP) technology

Rule: 6~7 degree doubling

From CCD to EMCCD

Signal-Noise Ratio: Noise

Page 24: From CCD to EMCCD Scientific imaging for today’s microscopy.

Tip: Readout Noise is the major equipmental noise contributor for a cooled camera!

Total equipment noise=√readout noise2+dark noise2

We use a typical readout noise = 8e, Dark noise =√total dark current =√dark current x exposure time

camera A cooled 25°C lower than ambient, dark current = 0.15e/p/scamera B cooled to -25°C, dark current = 0.015e/p/s

With exposure 30s, Total noise of camera A = 8.27e Total noise of camera B = 8.02e

With exposure 1mins, Total noise of camera A = 8.54e Total noise of camera A = 8.06e

From CCD to EMCCD

Signal-Noise Ratio: the cooler, the better?

Page 25: From CCD to EMCCD Scientific imaging for today’s microscopy.

Ultimately, a High-Performance CCD camera is limited only by Readout Noise and Photon Noise.

– Photon Noise - A law of physics!

– Readout Noise - Reduced by careful electronics design

– Dark Current Noise - Reduced by cooling and MPP

Noise Reduction in CCD

From CCD to EMCCD

Signal-Noise Ratio

Page 26: From CCD to EMCCD Scientific imaging for today’s microscopy.

From CCD to EMCCD

Signal-Noise Ratio: The final Equation

Signal-to-Noise Ratio of an Image = Total Photon collected / Noise1. Total Photon Collected where P=total incident photons,

DQE = QE at specific wavelength

2. Shot Noise

Page 27: From CCD to EMCCD Scientific imaging for today’s microscopy.

From CCD to EMCCD

Signal-Noise Ratio: The final Equation

Sensitivity of a camera: the lowest signal can be differentiated from background noise by the camera

Read Noise limited region

Photon Noise limited region

Page 28: From CCD to EMCCD Scientific imaging for today’s microscopy.

• Resolution (Digital Resolution, Spatial Resolution)

• Signal-Noise Ratio

• Dynamic Range (Contrast)

From CCD to EMCCD

Revised topic: important characters of digital image

Page 29: From CCD to EMCCD Scientific imaging for today’s microscopy.

Well capacity (Well depth):Number of electrons can be hold by a pixel before saturation

Well Capacity will be higher when pixel size is bigger:

• Same resolution, larger chip size;

• Same chip size, lower resolution;

• Binning

Note: If the charge capacity is exceeded,the excess charge will overflow into

adjacent pixels and produce artifacts known as blooming and smear.

NoiseOverflowing and Blooming

Charging

From CCD to EMCCD

Revised topic: important characters of digital image

Page 30: From CCD to EMCCD Scientific imaging for today’s microscopy.

Dynamic Range

Dynamic Range = Well capacity / Read noiseDynamic Range (dB) = 20 x Log10 (Well capacity /Read noise)

Tip: if your sample contains both very dark and very bright signals, a higher dynamic range camera is needed to imaging them in one shot!

From CCD to EMCCD

Revised topic: Dynamic Range

Dynamic Range of CCD should be matched to A/D Converter. 12, 14, 16 bit

Page 31: From CCD to EMCCD Scientific imaging for today’s microscopy.

Binning

- Higher Dynamic Range- Higher Signal-to-Noise Ratio- Faster Readout- Dynamically Change Pixel Size/Aspect Ratio

From CCD to EMCCD

Revised topic: Dynamic Range

Page 32: From CCD to EMCCD Scientific imaging for today’s microscopy.

Slower readout -> Lower noise

From CCD to EMCCD Low light – slower readout or longer exposure

Longer exposure -> Stronger

Signal

same exposure same readout speed

Page 33: From CCD to EMCCD Scientific imaging for today’s microscopy.

short exposure less photon collected

high readout high noise

From CCD to EMCCD

Low light & High Speed -> Short exposure + High Readout ??!!

Signal-to-Noise Ratio (SNR) = Total Photon collected / Noise

Page 34: From CCD to EMCCD Scientific imaging for today’s microscopy.

• EMCCD: Electron Multiplying Charge Coupled Device

• Operates by applying high voltage during readout before the preamp stage of the CCD. Occurs through a probabilistic phenomenon where the gain is determined by:

Gain = (1 + g)N

where g is the probability of creating a second electron (typically in the vicinity of 0.01 – 0.016) and N is the number of elements (usually 500+)

From CCD to EMCCD Low light & High Speed -> The EMCCD Technology !

Signal-to-Noise Ratio (SNR) = Total Photon collected / Noise

Page 35: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer CCD

Serial Register

Preamplifier

Output NodeActive Array

Page 36: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer CCD

Serial Register

Preamplifier

Output NodeActive Array

Page 37: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer CCD

Serial Register

Preamplifier

Output NodeActive Array

Page 38: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer CCD

Serial Register

Preamplifier

Output NodeActive Array

Page 39: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer CCD

Serial Register

Preamplifier

Output NodeActive Array

ADC

ReadoutSignal = 1

Note: if read noise is 1 then S/N = 1/1

Page 40: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer EMCCD

Active Array

Output Node

Page 41: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer EMCCD

Serial Register

Active Array

Preamplifier

Output Node

EM Register

Page 42: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer EMCCD

Serial Register

Active Array

Preamplifier

Output Node

EM Register

Page 43: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer EMCCD

Serial Register

Active Array

Preamplifier

Output Node

EM Register

Page 44: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer EMCCD

Serial Register

Active Array

Preamplifier

Output Node

EM Register

Page 45: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer EMCCD

Serial Register

Active Array

Preamplifier

Output Node

EM Register

Page 46: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer EMCCD

Serial Register

Active Array

Preamplifier

Output Node

EM Register

Page 47: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer EMCCD

Serial Register Preamplifier

Output Node

Active Array

EM Register

Page 48: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer EMCCD

Serial Register

Active Array

Preamplifier

Output Node

EM Register

Page 49: From CCD to EMCCD Scientific imaging for today’s microscopy.

Frame Transfer EMCCD

Serial Register Preamplifier

Output Node

Active Array

ADC

ReadoutSignal = 5

Note: if read noise is 1 then S/N = 5/1! Vast improvement

EM Register

Page 50: From CCD to EMCCD Scientific imaging for today’s microscopy.

On-Chip Multiplication Gain CCD SNR:

SNR=[S*QE]÷√[S*QE*F2 + D*F2 +(σR/G)2]

Note: F is the excess noise factor.

From CCD to EMCCD SNR: The new equation

Page 51: From CCD to EMCCD Scientific imaging for today’s microscopy.

From CCD to EMCCD Types of Noise in EM Cameras

• Dark Current– Dependent on exposure time– Increases when gain is increased -> cooling important

• Read Noise– Changes with readout speed

• Spurious Noise (aka clock induced charge)– Not dependent on exposure time– Lower cooling increases chance of spurious charge– Occurs during high pulse clocking of CCD and generates a

secondary electron, even though no primary is present– Usually combined with the overall dark charge

• Excess Noise Factor– Based on deviation or uncertainty in on-chip multiplication gain

Page 52: From CCD to EMCCD Scientific imaging for today’s microscopy.

EM camera Applications

• Total internal reflection fluorescence (TIRF) microscopy

• Spinning-disk confocal microscopy

• Dynamic ratio imaging (e.g., pH and low-concentration flux)

• Fluorescence recovery after photo bleaching (FRAP)

• Live-cell fluorescent protein imaging

Very high sensitivity up to single molecular detection!

Page 53: From CCD to EMCCD Scientific imaging for today’s microscopy.

From CCD to EMCCD When to use EMCCD?

Signal-to-Noise ratio curve

Page 54: From CCD to EMCCD Scientific imaging for today’s microscopy.

From CCD to EMCCD

Dual Amplifier EMCCD: Traditional Amplifier for Wide-dynamic range operation

Page 55: From CCD to EMCCD Scientific imaging for today’s microscopy.

From CCD to EMCCDScientific imaging for today’s microscopy

Thank you for your attention!