FreeStyle Optium Neo H Blood Glucose and Ketone Monitoring System
Healthcare professional portrayal.
Glucose Ketones
TheFreeStyleNeoHcanperformbothbloodglucoseandbloodKetonetests.Ensuretofamiliariseyourselfwithbothmodules.Clickonthemodulebelow.
Patient testing
Quality Control
Maintenance
Patient testing
Quality Control
Maintenance
AnIntroduc+ontoFreeStyleOp+umNeoH
TheFreeStyleOpAumNeoHSystemataGlance
High contrast display Displays meter home screen, test results
and other important information
Up button • Use to mark a venous blood test
• Adjust settings
Test strip port Where the test strip is inserted
Logbook
Down button • Used to mark a control solution test
• Adjust settings
Power button Press to:
• Turns meter on and off
• Returns to the home screen
• Saves settings
USB port Insert a data cable for uploading test results to a computer
TheFreeStyleOpAumNeoHBloodGlucoseTestStrip
TheFreeStyleOpAumNeoHmeterisdesignedtobeusedonlywithFreeStyleOpAumNeoHBloodGlucoseTestStrips.• 0.6μLsamplesize• 5-secondresultcountdown• Assayrange1.1-27.8mmol/L• Haematocritrange15-65%• ToporApsampleapplicaAon• VisualconfirmaAonofsufficientsample• Filltriggerensuressufficientsample• SampleTypes–Capillary,venous,arterial&neonatal• Eachpackcontains100individuallyfoilwrappedteststrips,calibraAonbarandteststripinsert
MetercalibraAon
• ThecalibraAonprocedureprogrammesthemeterwiththelotnumber,expirydateandteststriptechnology
• Alwayscalibratethemeterwitheachnewpackofteststrips–failuretocalibrateproperlywillcauseincorrectresults
• Withthelotnumberfacingtowardyou,insertthecontactbarsofthecalibratorintothemeter.ThemeterturnsonautomaAcally
• LOTandthelotnumberappearinthedisplaywindow• Checkthatthelotnumberonthemeterdisplaywindowmatches
withthenumberontheteststripcalibratorandthelastfivedigitsonboththeteststripfoilpacketandteststripinsert
• RetainthecalibraAonbarfortheduraAonofthatpack• CalibraAoniscomplete• Onlyusethecalibratorthatispackagedinthepackofteststrips
youareusing
PerformingaPa+entTestThe FreeStyle Optium Neo H system enables easy and accurate bedside testing of blood glucose levels
} 5 second test time
} Small 0.6 μL sample size
} Calibrator provides expiry check and
calibrates the meter
} Control solution mark
Blood glucose testing
PaAentTesAng1. Preparethesamplesite:
i. Beforeyouobtainablooddrop,makesurethesamplesiteisclean,dryandwarm
ii. Hangthearmdownbeforelancingthesamplesite,tohelpaidbloodflow
2. Opentheindividuallyfoilwrappedteststripatthenotchandteardown
3. InserttheteststripintothemeterunAlitstops.Thiswillturnthemeteron.
4. CheckthattheLOTnumberofteststripyouareusingmatchestheLOTnumbershownonthemeterdisplay
5. The blinks,indicaAngthemeterisreadyforyoutoapplyasampletotheteststripi. EnsureyoufolloworganisaAonalpolicyforsamplecollecAon
PaAenttesAng
6. IftesAngglucoseonvenousbloodpressthetomarkthetestunAlappearsindicaAngthatthemeterisnowreadyforyoutoapplyvenousbloodtotheteststrip7. Applyasamplewiththemeterinahorizontal(flat) posiAontoavoiditenteringtheteststripport8. Ifyouareapplyingthesamplefromasyringe,applyaway
fromthedirecAonoftheteststripport
PaAenttesAng
9. HoldsampletotheteststripunAl3shortlinesappearonthescreen
10. Theresultswillbedisplayedafera5-secondcountdown
PaAenttesAng
Whentheresultappearsonthemeterscreenthetestiscomplete.Theresultswillbestoredinthemeter’smemory.Pressandholdthebuhontoturnoffthemeter.
SummaryforbloodglucosetesAng
1. Insertsteststrip,blinks
2.Applyblood3.Sampledetected,5secondscountdownbegins
4.Resultisdisplayed
Outofrangeindicators
Highbloodglucoseresults• AresultfallingoutsidetherangesetbyyourorganisaAonmay
requiresomeintervenAonfromtheoperator,suchasrepeaAngthetest– Themeterwilldisplayasolidyellowarrow
• Forguidanceonoutofrangeresults,pleasefollowyourorganisaAonalpolicy
Lowbloodglucoseresults• AresultfallingoutsidetherangesetbyyourorganisaAonwill
requiresomeintervenAonfromtheoperator,suchasrepeaAngthetest– Themeterwilldisplayasolidredarrow
• Forguidanceonoutofrangeresults,pleasefollowyourorganisaAonalpolicy
DisplayMessages
Result is higher than or equal to 13.3 mmol/L
Glucose result is above 27.8 mmol/L
Result is less than 1.1 mmol/L
Results&Memory
Thetestiscompletewhentheresultappearsonthemeterscreen.• Theresultisstoredinthemeter’slogbook• Themeterlogbookcanstoreupto1,000events–includingbloodglucose,bloodketone,andcontrolsoluAonresults,andothermeterinformaAon• ThelogbookcanbeaccessedbygoingtotheHomeScreenandselecAng• TorecallpaAentresults:WiththeFreeStyleOpAumNeoHturnedon:
1. Pressthe toviewlogbookevents2. Press or toscrollthroughlogbookevents
While on the Home screen, press to view logbook events
Logbook:ViewLogbookEvents
Press to scroll through to view logbook events
Blood Glucose Result
Glucose Control Solution Result
Blood Ketone Result Venous Blood Result
Ketone Control Solution Result
Control solution test results will alternate between ‘CtL’ and result
Logbook:ViewBloodGlucoseAverages
Average
Number of tests In this average
1. While in the logbook, press any time to view the 7-day average
2. Press or to scroll through 7-, 14- and 30-day averages. Press to return to logbook events.
ErrorMessages
Message What it means
E1 Temperature is too hot or too cold for the meter to work correctly
E2/E9 Meter error
E3 i. Blood drop is too small
ii. Incorrect test procedure
iii. There may be a problem with the test strip
E4 i. The blood glucose level may be too high to be read by the system
ii. There may be a problem with the test strip
E5 i. Blood was applied to the test strip too soon
E6 i. Calibration error
ii. Test strip error
E7 i. Test strip may be damaged, used or the meter does not recognise it
PerformingaQualityControlTest
PerformingaQualityControlTest
• IfQCtesAngisNOTperformed,misleadingresultsmayleadtoinappropriatetreatment
• AlwaysfollowyourorganisaAonalpolicytoavoidpumngpaAentsatrisk• TheFreeStyleOpAumNeoHmetermaypromptwhenthisisdue• OnlyuseMediSenseGlucoseandKetoneControlSoluAonsandkeepthemeter
flattopreventexcesssoluAonrunningintothemeter• QCSoluAonisstablefor90daysfollowingopeningorunAltheexpirydate
printedonthesoluAonbohles,whicheverissooner.DiscardanddoNOTusecontrolsoluAonpastitsexpirydate
• CheckresultsagainsttheteststripinsertinsidethepackofteststripsbeingusedasQCrangeswillvarybetweenteststripLOTs
• ItistheresponsibilityofalluserstoensurethemeterhasbeentestedbeforepaAentuse
PerformingQualityControl(QC)Tests1. CheckthattheLOTnumberprintedontheteststripandinstrucAonsforuse
match2. Opentheindividualfoilteststripatthenotchandteardown3. Inserttheteststripintothemeterwhichwillturnthemeteronturnon4. WhentheblinkspressthebuhonunAlthisiconappears5. ThemeterisnowreadyforyoutoapplythelowlevelcontrolsoluAontothetest
stripi. InvertthebohletomixthesoluAonii. WipetheQCnozzlewithcleangauzebeforeandafereachtestiii. Replacethecapsecurelyonthebohleimmediatelyaferuse
7. ApplyadropofthecontrolsoluAonwiththemeterinahorizontalposiAontoavoiditenteringtheteststripport
PerformingQualityControl(QC)Tests8. ApplyadropofthecontrolsoluAontotheteststripunAl3shortlinesappearon
thescreen9. Theresultswillbedisplayedafera5-secondcountdown
i. ComparethecontrolsoluAonresultstotherangeprintedonthebloodglucoseteststripinsert; theresultsshouldfallwithinthisrangei. TheresultswillbestoredinmemoryasacontrolsoluAonresultii. RecordtheresultintheQCrecordbook
10. Ifresultsareout-of-rangei. Repeatthetestii. Iftherepeatedresultisout-of-rangefolloworganisaAonalpolicyforout-of-rangeresults
11. Removetheteststrip,repeattheprocedureforhighlevelcontrolsoluAon12. RecordHIandLOQCtestresultintheQCrecordbook
QualityControlTesAngSummary
1. Insertteststrip,blinks.
2.PressandholdtheunAlappears.
3.ThemeterisnowreadyfortheoperatortoapplycontrolsoluAontotheteststrip.
4.ApplycontrolsoluAon.5.Sampledetected,countdown
begins.5.Resultisdisplayed.
Maintenance
Maintenanceofthemeter
1. GentlycleantheexterioroftheFreeStyleOpAumNeoHmeterwithadampclothi. AcceptablecleaningsoluAonsincludemilddetergentandwater,70%isopropylalcohol
oramixtureof1parthouseholdbleachand9partswaterii. Donotimmersethemeterinwaterorotherliquids.Avoidgemngdust,blood,control
soluAonandanyothersubstancesintheteststripport,USBportandbaherycompartment
iii. Allowthemetertodry
2. Donottrytocleantheteststripport.3. Donotpourliquidintotheteststripportorontothebuhons.4. Donotimmersethemeterinwateroranyotherliquid.
Maintenanceofthemeter1. Replacecoincellbaherieswhen‘LowBaheryPower’messageisdisplayed
1. Themeterrequires2replaceableCR2032coincellbaheries.2. Tochangethebaheriesturnthemeteroverandslidethebaherydoor
(locatedonthesideofthedevice)open,asshown.2. PerformQCtestasperorganisaAonalpolicy3. RefertotroubleshooAngguideinFreeStyleOpAumNeoHoperator’smanualfor
errormessages4. Whennotinuse,alwaysstoretheFreeStyleOpAumNeoHmeterasper
organisaAonalpolicy
Bloodβ-KetoneTes+ngwithFreeStyleOp+umNeoH
The FreeStyle Optium Neo H system enables easy and accurate bedside testing of blood ketone levels
} 10 second test time
} 1.5 μL sample size
} Purple calibrator calibrates the meter
} Control solution mark
Blood ketone testing
AnIntroduc+ontoFreeStyleOp+umNeoH
TheFreeStyleOpAumNeoHSystemataGlance
High contrast display Displays meter home screen, test results
and other important information
Up button • Use to mark a venous blood test
• Adjust settings
Test strip port Where the test strip is inserted
Down button • Used to mark a control solution test
• Adjust settings
Power button Press to:
• Turns meter on and off
• Returns to the home screen
• Saves settings
USB port Insert a data cable for uploading test results to a computer
FreeStyle Optium H Blood β-Ketone Test Strips
TheFreeStyleOpAumHBloodβ-KetoneTestStrip:• 1.5μLsamplesize• 10-secondresultcountdown• AssayRange0.0-8.0mmol/L• HaematocritRange30-60%• ToporApsampleapplicaAon• VisualconfirmaAonofsufficientsample• Filltriggerensuressufficientsample• SampleType–capillary(finger)andvenouswholebloodsamples• Eachpackcontains10individuallyfoilwrappedteststrips,calibraAonbarandteststripinsert
β-Ketone Benefits
Clinicalaccuracy• Measuresβ-hydroxybutyrate(themostsignificantketonebody)• AquanAtaAvevs.qualitaAvemeasure• ResultsunaffectedbymanycommonmedicaAons
Pa+entconvenience/compliance• Bloodvs.urinesampling• Foilwrappedteststrips• Testintegratedwithinbloodglucosemeters
MetercalibraAon
• ThecalibraAonprocedureprogrammesthemeterwiththelotnumber,expirydateandteststriptechnology
• Alwayscalibratethemeterwitheachnewpackofteststrips–failuretocalibrateproperlywillcauseincorrectresults
• Withthelotnumberfacingtowardyou,insertthecontactbarsofthecalibratorintothemeter.ThemeterturnsonautomaAcally
• LOTandthelotnumberappearinthedisplaywindow• Checkthatthelotnumberonthemeterdisplaywindowmatches
withthenumberontheteststripcalibratorandthelastfivedigitsonboththeteststripfoilpacketandteststripinsert
• RetainthecalibraAonbarfortheduraAonofthatpack• CalibraAoniscomplete• Onlyusethecalibratorthatispackagedinthepackofteststrips
youareusing
PerformingaPa+entTest
PaAentTesAng1. Preparethesamplesite:
i. Beforeyouobtainablooddrop,makesurethesamplesiteisclean,dryandwarm
ii. Hangthearmdownbeforelancingthesamplesite,tohelpaidbloodflow
2. Opentheindividuallyfoilwrappedteststripatthenotchandteardown
3. InserttheteststripintothemeterunAlitstops.Thiswillturnthemeteron.
4. CheckthattheLOTnumberofteststripyouareusingmatchestheLOTnumbershownonthemeterdisplay
5. The blinks,indicaAngthemeterisreadyforyoutoapplyasampletotheteststripi. EnsureyoufolloworganisaAonalpolicyforsamplecollecAon
PaAenttesAng
6. IftesAngketonesonvenousbloodpressthetomarkthetestunAlappearsindicaAngthatthemeterisnowreadyforyoutoapplyvenousbloodtotheteststrip7. Applyasamplewiththemeterinahorizontal(flat) posiAontoavoiditenteringtheteststripport8. Ifyouareapplyingthesamplefromasyringe,applyaway
fromthedirecAonoftheteststripport
PaAenttesAng
9. HoldsampletotheteststripunAl3shortlinesappearonthescreen
10. Theresultswillbedisplayedafera10-secondcountdown
PaAenttesAng
Whentheresultappearsonthemeterscreenthetestiscomplete.Theresultswillbestoredinthemeter’smemory.Pressandholdthebuhontoturnoffthemeter.
DisplayMessages
Message Explanation
Result above 8.0 mmol/L
Thisappearswhenaresultishigherthan8.0mmol/LThissuggestsveryhighbloodketonelevelsorthattheremaybeanerrorwiththeteststripRepeatthetestwithanewteststripThisrequiresimmediateacAonasdefinedbyyourfacility’spoliciesandprocedures
1. Insert ketone test strip, blinks and KET appears
2. Apply blood
3. Sample detected, 10 seconds countdown begins
4. Result is displayed
Blood Ketone Testing summary
ErrorMessages
Message What it means
E1 Temperature is too hot or too cold for the meter to work correctly
E2/E9 Meter error
E3 i. Blood drop is too small
ii. Incorrect test procedure
iii. There may be a problem with the test strip
E4 i. The blood glucose level may be too high to be read by the system
ii. There may be a problem with the test strip
E5 i. Blood was applied to the test strip too soon
E6 i. Calibration error
ii. Test strip error
E7 i. Test strip may be damaged, used or the meter does not recognise it
PerformingaQualityControlTest
PerformingaQualityControlTest
• IfQCtesAngisNOTperformed,misleadingresultsmayleadtoinappropriatetreatment
• AlwaysfollowyourorganisaAonalpolicytoavoidpumngpaAentsatrisk• TheFreeStyleOpAumNeoHmetermaypromptwhenthisisdue• OnlyuseMediSenseGlucoseandKetoneControlSoluAonsandkeepthemeter
flattopreventexcesssoluAonrunningintothemeter• QCSoluAonisstablefor90daysfollowingopeningorunAltheexpirydate
printedonthesoluAonbohles,whicheverissooner.DiscardanddoNOTusecontrolsoluAonpastitsexpirydate
• CheckresultsagainsttheteststripinsertinsidethepackofteststripsbeingusedasQCrangeswillvarybetweenteststripLOTs
• ItistheresponsibilityofalluserstoensurethemeterhasbeentestedbeforepaAentuse
PerformingQualityControl(QC)Tests1. CheckthattheLOTnumberprintedontheteststripandinstrucAonsforusematch2. Opentheindividualfoilteststripatthenotchandteardown3. Inserttheteststripintothemeterwhichwillturnthemeteronturnon4. WhentheblinkspressthebuhonunAlthisiconappears5. ThemeterisnowreadyforyoutoapplythelowlevelcontrolsoluAontothetest
stripi. InvertthebohletomixthesoluAonii. WipetheQCnozzlewithcleangauzebeforeandafereachtestiii. Replacethecapsecurelyonthebohleimmediatelyaferuse
7. ApplyadropofthecontrolsoluAonwiththemeterinahorizontalposiAontoavoiditenteringtheteststripport*KETwillappearonthescreenasapurpleketoneteststriphasbeeninserted
PerformingQualityControl(QC)Tests8. ApplyadropofthecontrolsoluAontotheteststripunAl3shortlinesappearon
thescreen9. Theresultswillbedisplayedafera10-secondcountdown
i. ComparethecontrolsoluAonresultstotherangeprintedonthebloodglucoseteststripinsert; theresultsshouldfallwithinthisrangei. TheresultswillbestoredinmemoryasacontrolsoluAonresultii. RecordtheresultintheQCrecordbook
10. Ifresultsareout-of-rangei. Repeatthetestii. Iftherepeatedresultisout-of-rangefolloworganisaAonalpolicyforout-of-rangeresults
11. Removetheteststrip,repeattheprocedureforhighlevelcontrolsoluAon12. RecordHIandLOQCtestresultintheQCrecordbook
QualityControlTesAngSummary
1. Insertteststrip,blinks.
2.PressandholdtheunAlappears.
3.ThemeterisnowreadyfortheoperatortoapplycontrolsoluAontotheteststrip.
4.ApplycontrolsoluAon.5.Sampledetected,countdown
begins.5.Resultisdisplayed.
Maintenance
Maintenanceofthemeter
1. GentlycleantheexterioroftheFreeStyleOpAumNeoHmeterwithadampclothi. AcceptablecleaningsoluAonsincludemilddetergentandwater,70%isopropylalcohol
oramixtureof1parthouseholdbleachand9partswaterii. Donotimmersethemeterinwaterorotherliquids.Avoidgemngdust,blood,control
soluAonandanyothersubstancesintheteststripport,USBportandbaherycompartment
iii. Allowthemetertodry
2. Donottrytocleantheteststripport.3. Donotpourliquidintotheteststripportorontothebuhons.4. Donotimmersethemeterinwateroranyotherliquid.
Maintenanceofthemeter1. Replacecoincellbaherieswhen‘LowBaheryPower’messageisdisplayed
1. Themeterrequires2replaceableCR2032coincellbaheries.2. Tochangethebaheriesturnthemeteroverandslidethebaherydoor
(locatedonthesideofthedevice)open,asshown.2. PerformQCtestasperorganisaAonalpolicy3. RefertotroubleshooAngguideinFreeStyleOpAumNeoHoperator’smanualfor
errormessages4. Whennotinuse,alwaysstoretheFreeStyleOpAumNeoHmeterasper
organisaAonalpolicy
CongratulaAonsoncompleAngthismodule
FreeStyle blood glucose meters assist in the monitoring of blood glucose (and ketone) levels for people diagnosed with diabetes mellitus. People diagnosed with diabetes with elevated blood ketone levels should seek medical advice. Information contained herein for distribution outside of the U.S. Only. FreeStyle and related brand marks are trademarks of Abbott Diabetes Care Inc. in various jurisdictions. All other trademarks are the property of their respective owners. Abbott Diabetes Care Australia, 666 Doncaster Road, Doncaster VIC 3180. For further information call Customer Service on 1800 801 478 or visit www.myfreestyle.com.au MSE160420044321