Knowledge Formulation for Supervised Evidential Classification
Evaluation of New Evidential Breath Alcohol Instrumentation€¦ · Evaluation of New Evidential...
Transcript of Evaluation of New Evidential Breath Alcohol Instrumentation€¦ · Evaluation of New Evidential...
Evaluation of New Evidential Breath Alcohol Instrumentation
Preliminary Results
R . G . Gullberg
Washington State Patrol
Breath Test Program
December 2007
Introduction
The purpose of this review was to evaluate the most recently
developed technology in forensic breath alcohol instrumentation .
The effort was largely investigational in order to learn what
features were now available and to determine what might be relevant
for the evidential breath test program in Washington State . Part of
the objective was also to consider revisions to Washington ' s breath
test program regarding both instrumentation and protocols. Since
Washington ' s current program was largely developed approximately
twenty years ago , many features offered by new instrumentation were
considered as possible useful additions to program development .
Four instruments were provided by vendors for evaluation .
These included : ( 1) Datamaster DMT - National Patent Analytical
Systems, Inc. , ( 2 Intoxilyzer 8000 - CMI Inc. , Intoximeter.
ECIR II - Intoximeters, Inc . , and ( 4 ) Alcotest 9510 - Drager Safety ,
Inc . All of the experimental work was performed at the Roanoke
office of the Breath Test Program in Seattle. None of the
instruments were re- calibrated in the Seattle Lab prior to beginning
the following analyses. The instruments were evaluated with the
calibration established by the manufacturer.
The following summary will present the principle results from
this preliminary evaluation .
Accuracy and Precision
Accuracy and precision were evaluated by performing n = 10
replicate simulator measurements on each of the four instruments at
eight concentrations. The solutions were obtained from the State
Toxicology Laboratory where reference values were determined by
headspace gas chromatography Figures 5 plot the bias
along with one coefficient of variation ( ) error bars at each of
the eight reference concentrations. Lines representing bias are
also shown for reference. Figures 6 through 9 are linearity plots
137
for each instrument . Each point represents the mean of n = 10
simulator replicates along with standard deviation ( SD ) error
bars . The linear regression equation is shown as well for each
instrument .
All instruments revealed very acceptable accuracy with bias
estimates being less than 4 . 7 % All instruments, except the
Alcotest 9510 , revealed acceptable precision with estimates being
less than 2 . 5 % . The Alcotest did not have heated simulator
tubing and showed some very large estimates at both the IR and EC
detectors. Figure 2 shows the bias plot for the Intoxilyzer 8000
when calibrated at the factor with the four point calibration
algorithm . Figure 3 shows the bias plots for the Intoxilyzer 8000
when calibrated with the one point calibration routine. The one
point calibration appears to yield very satisfactory results. The
plot in figure 5 shows the results on the Alcotest 9510 when the
simulator tubing was purged between runs of n = 10 . When doing so ,
the results were very acceptable. Heated simulator tubing was a
feature that Drager added at a later time.
Figure 10 plots the total error for each instrument determined
at each of the eight reference concentrations. Total error, as
shown by the equation in the plot , combines both bias and random
error ( CV) The Intoximeter ECIR performed the best overall by this
criteria .
Effect of Loose Simulator Jar
Figure shows the results of employing a tight and loose
simulator jar . set of n = 10 simulator measurements were performed
on the instrument initially with a tight simulator jar . Then the
jar was loosened slightly and another set of n = 10 measurements were
performed. instruments provided acceptable results with both
the tight and loose simulator jar except the Intoxilyzer 8000
shown in the figure, this instrument was very sensitive to the loose
jar condition . Both accuracy and precision suffered in this
condition . Very likely the vacuum sampling mechanism of the
Intoxilyzer 8000 was the cause of this . 11 also shows the
effect of shorter and simulator tubing in the Intoxilyzer
8000 . The effect of having a loose simulator jar would be an
important consideration where simulators are used as control
standards for every subject tested .
Day- to -Day Simulator Differences
Employing a new instrument in Washington State will likely
involve a change in the evidential test control standard
138
measurement. There will most likely be two simulator or gas
standards performed with each subject tested. In order to evaluate
the day - to - day variation, duplicate simulator tests were performed
on each instrument on different days . Figures 12 to 15 show these
results for each instrument The difference was determined by
subtracting the second result from the first Figures 12 and 14
( for the Datamaster DMT and Intoxilyzer 8000, respectively) a
tendency for the difference to be negative. This results from the
second result being slightly higher than the first This can result
in simulator tests where the tubing is not heated, as in the case of
the Intoxilyzer 8000 . The Datamaster DMT, on the other hand, had
heated simulator tubing of the differences along with the
actual results themselves, were very acceptable and would comply
with the current statutory requirements in Washington State.
Figure 16 shows another evaluation of day - to - day stability , in
this example for the Intoxilyzer 8000. Each point represents the
mean of n = 10 measurements of either the wet bath simulator or dry
gas standard . Also shown are error bars. Each run of
measurements were performed on different days . From the plot we see
that the wet bath simulator results tended to decrease over time at
a greater rate than the dry gas standard . This results from the
depletion effect which is more pronounced in use of the wet bath
simulator. We also see that the variability (magnitude of the 2SD
error bars) is generally greater with the simulator. This results
largely from the lack of heated simulator tubing in the Intoxilyzer
8000 . Figure 17 shows the same plot for the Datamaster DMT. Here
we see less depletion effect when compared to the Intoxilyzer 8000
in figure 16 . We also note that the variability is generally more
for the simulator than the dry gas standard. Figure 18 shows the
same results for the Intoximeter ECIR II. From this plot we see the
superior day - to - day precision compared to either the Intoxilyzer
8000 or the Datamaster DMT .
Carryover Effect
One important characteristic of an analytical system is that
each separate sample introduced for analysis must be considered
totally analytically independent. Each sample must be measured
without influence from prior samples introduced. This is not really
expected to be a problem with any of the instruments considered here
since they have purging routines along with zeroing requirements
before a subsequent sample can be introduced. In order the evaluate
the instruments for any carryover effect, alternatinghigh ( 0 .40
g /210L) and low ( 0 . 04 g /210L) were introduced. Some of the
instruments received the sample through the breath tube and some
through the simulator ports. Figures 19 to 22 reveal the results
for each instrument. Both the Datamaster DMT and the Drager 9510
produced very stable and reproducible results as seen in figures 19
and 20 . The Drager 9510 actually did not yield stable lower results
when the samples were introduced through the simulator ports . This
resulted from not having heated simulator tubing. Later , heated
simulator was installed which improved precision of the
lower results. The Intoxilyzer 8000 was very sensitive to the use
of non -heated simulator tubing. Figure 21 reveals the tendency for
the results to be increasing with sequential analysis until
measurement number when the tubing was cleared and the results
stabilized. The response of the Intoximeter ECIR in figure 22 was
very unusual Both the high and low sequential results tended
towards each other . These samples were also introduced through the
heated breath tube .
Wet Bath Simulators and Dry Gas Comparisons
One important feature of the instrument must be the ability to
employ either wet bath simulator solutions or dry gas as external
control standards . have such capability . Figure 23
shows the plot of means along with 2 SD error bars for both simulator
and dry gas standard results. Each point is the mean of n = 10
measurements . All sequential results were performed on different
days . From the figure we see that the dry gas standards yield
slightly better precision . This probably results from condensation
in the simulator headspace and tubing along with non - constant
temperature control. We also see a slight trend down for the
simulator results. Although the instrument employs recycle tubing
this represents some slight depletion over time . The total
measurements were well over 150 . Figure 24 shows the of means
with 2SD error bars for the Intoximeter ECIR . This plot also
reveals slightly better precision for the dry gas standards . We
also see a more significant depletion rate in the simulator results
compared to that of figure 23 for the Datamaster DMT. Figure 25
shows the results for the same analysis in the Intoxilyzer 8000 .
This reveals an even greater rate of depletion in the simulator
results . Again , the dry gas standard results reveal slightly better
precision than the simulator results. While figure 25 is a plot of
percent bias rather than actual mean concentrations , the trend and
relative precision are still observable . Figure 26 is another
illustration of similar analysis on the Intoxilyzer 8000 . The means
of n = 10 measurements for five sequential sets of runs are plotted
along with one coefficient of variation error bars . The three
configurations were simulator without clearing the tubing between
runs of n = 10, simulator with cleared tubing and dry gas standards .
The figure reveals that when the tubing is not cleared on the
instrument there is significant increase in variability . This
results from the simulator tubing not being heated as the instrument
140
was configured Clearing the tubing greatly improved precision ,
even better than the dry gas standard runs.
Interference Test Results
All of the instruments employed features designed to detect
the presence of an interfering substance . Figures 27 through 29
show the results of testing these features . Figure 27 shows the
results of first measuring the sample from a simulator containing on
ethanol. The simulator was arranged with four -way tubing so that
the same sample was received by all four instruments at the same
time. A common pressure source (portable compressor ) was employed
as the positive pressure sampling source . All four instruments
yielded very similar results for the ethanol only sample . Next, five
drops of acetone was added to the simulator to represent a low
concentration . The Drager 9510 and the Intoxilyzer 8000 yielded no
significant change in the result The Datamaster DMT and the
Datamaster CDM showed slightly increased measured results . Next , a
total of ten drops of acetone were in the simulator and finally a
total of fifteen drops were in the simulator . On these last two
occasions the Intoxilyzer 8000, Datamaster DMT and Datamaster CDM
all detected the presence of an interferents. In all three
conditions with added acetone , the Drager 9510 did not show any
significant change in the measured result . This results from the
use of 9 . 5 um in the IR region and fuel cell technology , both of
which do not respond to acetone . Figure 28 shows the results of
similar testing employing isopropanol. Again the four instruments
respond very comparably to the ethanol only simulator. Adding
200 of isopropanol slightly increased the result for the Drager
9510 instrument while it was detected as an interfering for the
other three instruments. Finally, adding a total of 400
isopropanol yielded an interfering result on the Drager 9510 . The
differing response of the Drager 9510 after adding 200
probably due to the use again of 9 . 5 um in the IR region along with
the fuel cell Finally, a separate set of tests were done on just
the Drager 9510 with the results shown in figure 29 . A set of ten
measurements were first performed using a simulator with ethanol
only . Then 200 of acetone was added and another measurement was
performed . additional acetone is added we see from the figure
than the IR result tends to increase while the fuel cell result
tends to decline. Eventually at 600 and 800 acetone, one of
the measurements for each run yielded interference on one of the
analytical methods. At 1000 of acetone both methods yielded
interference 200 of acetone, the vapor concentration is
approximately 926 ug/ is an extreme concentration only
likely to be present in uncontrolled diabetics .
141
Additional Analyses
There were a few additional analyses that were done with
different instruments. One evaluation done on the Intoxilyzer 8000
involved the calibration procedure. The instrument as initially
received was designed to be re - calibrated at four different
concentrations . This process would take about 30 minutes to
accomplish . This did not seem an acceptable process so the
manufacturer provided another instrument that was designed to employ
one-point calibration 30 and 31 show the results of this
analysis . Figure 30 plots the means and one coefficient of
variation error bars as percent bias for n = 10 measurements
throughout the concentration range where the instrument had received
the manufacturer four point calibration We see that all results
remained remain within 5 % of the reference value . 31 plots
the same analysis for the instrument calibrated with one point
calibration . Again we see that all results remain within 5 % of the
reference value. This supports the sufficiency of one -point
calibration in the Intoxilyzer 8000 which would be a more acceptable
procedure as part of an annual re- calibration program .
Another evaluation completed was the comparison of a different
simulator . A Guth model 590 simulator was compared to the Guth
model 2100 simulator. A total of n 20 measurements were performed
on each simulator using the same solution batch and the same
Intoxilyzer 8000 instrument Figure 32 shows the results While
both simulators yielded very similar precision estimates, the Guth
590 simulator results were slightly higher . A runs test on both
sets of data showed random results with no trend.
Desirable Instrument Features
Part of this evaluation was to learn of the existing technology
and identify desirable features for the Breath Test Program in
Washington State. The following is a partial list of desirable
features for the instrument:
1 . The ability employ either wet bath simulators or dry gas as
control standard devices.
2 . The ability to employ operator and technician card readers
using 2D bar code technology While all of the manufacturers
claimed to offer this feature, none actually fully demonstrated it
on their instrument.
3 . The ability of the instrument to provide a page Quality
Assurance Procedure printout of the final results following a QAP
procedure .
142
. Windows based operating system technology with touch screen
capability .
5 . The ability to store and produce when requested , the breath
alcohol concentration and breath flow curves .
6 . The ability to recall a complete breath test printout record
from a prior date when needed. This can be provided at the host
computer end.
7 . The ability to monitor and record the simulator temperature on
the actual printout document when a simulator control standard is
employed
8 . The ability during a QAP to monitor and record the accuracy ,
precision and test for outliers .
9 . The ability during a QAP to monitor and record critical voltage
settings in the instrument.
10 . Part of the contract should state that the State of Washington
will not have any ownership of or any access to the source code or
any copyrights thereof within the instrument .
Conclusions and Opinions
All of the instruments evaluated represent the technology ,
features and capabilities for state -of- the-art forensic breath test
instruments that are fully fit - for- purpose. After working with the
instruments it is my opinion that the Datamaster DMT or the Drager
9510 would be the instruments best suited to meet the needs for the
State of Washington these two instruments, I believe that the
Datamaster DMT is best suited for immediate use and implementation .
There are, however , some features still in need of revision on the
Datamaster DMT ( i . e . , the gas standard configuration ) . While the
Drager 9510 has many desirable features it is not yet ready for
implementation in my opinion . If time and program needs were to
allow it , I would like to continue the evaluation of these two
instruments for the next year . At that time both technologies would
be advanced and better suited for a more informed decision.
R G Gullberg
12 / 13/ 2007
143
Figure 1
PercentBias and One CoefficientofVariation Error Bars ForDatamaster DMT
PercentBias
8
- -
0 0 0 . 1 0 . 4 0 . 5
144
0 .2 0 . 3
Reference Value( g/ 210L )
Figure 2
Percent Bias and One Coefficient ofVariation Error Bars For Intoxilyzer 8000
PercentBias
- -
- 8
0 .0 0. 1 0 . 4 0 . 50.2 0.3
Reference Value (g /210L )145
Figure 3
PercentBias and One Coefficientof Variation Error Bars For Intoxilyzer 8000
Using One PointCalibration
PercentBias
87One PointCalibration
n = 10 at each point
6
E - - - - - -- -
- -
- 8
0. 1 0.2 0 . 40 3
Reference Value (g/210L )
Figure 4
Percent Bias and One Coefficient of Variation Error Bars For Intoximeter ECIR
PercentBias
- - - - - - - - - - - - - - - - - - - - - - -
- - - -
0 . 0 0. 10 . 1 0. 2 0.3 0.40 . 4 0.50 . 5
147
0 . 2 0 . 3
Reference Value ( g /210L )
Figure 5Percent Bias and One Coefficient of Variation Error Bars For Alcotest 9510
PercentBias
n = 10 at each point2 27 /2007
- - -
-- - - -
- 6
0 .00 0. 2 0.40. 3
Reference Value( g/210L)148
Figure 6
Linearity PlotFor Datamaster DMT
Measured Concentration (g / 210L )
0 . 5
0 . 4
0 . 3
Y = 0. 9873 X + 0 . 000048
0 .
0 . 0
0 . 0 0 : 1 0 . 2 0 . 3 0 . 4 0 . 5
149 Reference Value ( g /210L)
Figure 7Linearity PlotForIntoxilyzer 8000
MeasuredConcentration ( g /210L )
0 . 5
0 . 4
Y = 0 . 9806 X + 0.00089
0 . 0 +0 . 1 0 . 2 0 . 3 0 . 4 05
150
Reference Value ( g/210L )
Figure 8
Linearity Plot For Intoximeter ECIR
Measured Concentration ( g/ 210L )
0 .5
0 . 4
0 . 3
Y = 0 . 9810 X + 0. 00428
0 .0
0 . 0 0 . 1 0. 2 0 . 3 0 . 4 0 . 5
ReferenceValue ( g/210L )
Figure 9PlotFor Alcotest 9510
Measured Concentration ( g / 210L )
0 . 5
0 . 4
0 . 3
0 .
Y = 0 . 992 X + 0 . 00026
0 . 1
0 . 0
0.00 . 0 0 1 0 . 21 0 . 3 0. 4 0 . 5
152
Reference Value (g/210L )
Figure 10
Plots of TotalError For Each InstrumentDetermined atEightConcentrations
Percent TotalError
Datamaster DMT
.... .... . Intoxilyzer 8000
- - - Intoximeter ECIR
Datamaster CDM
= bias + 2 CV-
4
---
0 . 0 0153
0 . 2 0. 3
Reference Concentration (g/210L)
Figure 11
Percent Bias and One CV Error Bars Comparing Tight and Loose Simulator
Percent Bias
2 .5 inch tubing
- 20Loose SimulatorTightSimulator
Seven inch tubing
-30
-40
Intoxilyzer 8000 DatamasterDMT Intoximeter ECIR Alcotest9510
Instrumentand SimulatorArrangement154
Figure 12PlotofDifferences Between Duplicate Simulator Tests Performed on Different
Days For the Datamaster DMT
Difference Sim 1-Sim ( g /210L)
0 . 010DatamasterDMT
Day- to -day precision: 0 .0015 g/ 210L
0.005
0 . 000
- . 005
- 0 .010
3020
Test Sequence155
Figure 13
Plot of Differences Between Duplicate Simulator Tests Performed on Different
Days For the Intoximeter ECIR
Difference ( g / 210L )
0.Intoximeter ECIR
Day to -day precision : 0 .0008 g/ 210L
0 .005
0.000
- 0 . 005
-0 . 010
10 20 30156
Sequence
Figure 14Plot of Differences Between Duplicate Simulator Tests Performed on Different
Days Forthe Intoxilyzer8000
Difference( g/ 210L)
0. Intoxilyzer 8000
Day- to -day precision : 0 .0028 g / 210L
0 .005
0 .000
- 0 . 005
- 0 .010 +
10 30157
20
TestSequence
Figure 15Plot of Differences Between Duplicate Simulator Tests Performed on Different
For the Drager 9510 IR Detector
Difference( g/210L)
0 . 010
Drager 9510 IR result
Day- to -day precision : 0.0008 g/210L
0 . 005
0 .000
- 0 .005
-0.01010
158
Test Sequence
Figure 16PlotofMean Along With 2SD Error Bars For Simulator and Dry Gas Results on
Intoxilyzer 8000
Percent Bias
10n = 10 at each pointGas Standard Reference : 0 . 100 g /210L
Simulator Reference : 0 .0827 g / 210L
5 + - - - - - - - -
Simulator results
- 5 -!
Gas standard results
1015
Day159
Figure 17
Plot of Mean Along With 2SD Error Bars For Simulator and Dry Gas Results onDatamaster DMT
Standard Results (g/ 210L )
0. 11Gas standard results
n = 10 at each point
Gas Standard Reference: 0 100 210L
Simulator Reference: 0 .0820 g /210L
0 . 100 g /210L
. 10
0 . 10
. 09
Ref. 0. 0820 g /210L
0.08
Simulator results
0 5 10 1515160
Day
Figure 18PlotofMean Along With 2SD Error Bars For Simulator and Dry Gas Results on
Intoximeter ECIR
StandardResults ( g/ 210L )
0. 100n = 10 at each point
Simulator Reference: 0. 0814 g/ 210L
Gas Standard Reference: 0 .0820 g /210L
0 . 095
Simulator results
0. 090
0.085Dry Gas Reference = 0. 0820 g/210L
HOH
0.080Simulator Reference = 0 .0814 g /210L
Gas standard results
0 .075
0 .070
10 15
Day161
Figure 19Sequential Plotof Alternating High and Low Alcohol Concentrations
Alcohol Concentration ( g / 210L)
0 .5 DatamasterDMT
Samples through simulatorports
mean = 0 .3940 g /210LSD = 0 . 0039 g /210L
mean = 0 .0442 g /210L
SD = 0 .0017 g/210L
8 10 126SequentialTest
162
Figure 20
Sequential Plot of Alternating High and Low AlcoholConcentrations
Alcohol Concentration (g /210L )
0 .57
Alcotest 9510 IR Resultthrough breath tube
0. 4
mean = 0 .4339 g /210L
SD = 0 .0021 g/ 210L
mean = 0 . 0383 g /210L
SD = 0 .0007g/ 210L
0 . 1
163Sequential Test
Figure 21
Sequential Plotof Alternating High and Low Alcohol Concentrations
Alcohol Concentration ( g /210L ) Intoxilyzer 8000
Samples through simulator ports0.5
0 . 4
mean = 0 .4086 g /210L
SD = 0 .0062 g / 210L
Cleared tubingmean = 0 .0537 g /210L
SD = 0 . 0088g/210L
0 . 1
0. 0 +
164 Sequential Test
Figure 22Sequential Plot of Alternating High and Low Alcohol Concentrations
AlcoholConcentration (g/210L) Intoximeter ECIR
samples through breath tube
mean = 0 .3561 g / 210LSD = 0 .024 g /210LCV = 6 . 7 %
mean = 0 .0774 g/210LSD = 0 .024g
CV = 31 %
0 . 0
4 6 10 12165
Sequential Test
Figure 23
PlotofMeans and ErrorBars ForBoth Simulator and Dry Gas Results
Standard Results ( g/210L )0 . 117
Gas standard results
Datamaster DMT
n = 10 at each point
Gas Standard Reference: 0 . 100 g /210LSimulator Reference : 0 .0820 g /210L
I I Ref. = 0. 100 g/210L
0. 10
0 .09
Ref. = 0 .0820 g /210L
0 . 08
Simulator results
0. 0710 15
Day
Figure 24
PlotofMeans and Error Bars For Both Simulator and Dry Gas Results
Standard Results ( g /210L )0 . 100
Intoximeter ECIRn = 10 ateach point
Simulator Reference : 0 .0814 g / 210L
Gas Standard Reference: 0. 0820 g / 210L0 .095
Simulator results0. 090
0.085Dry GasReference = 0 . 0820 g /210L
0 .080Simulator Reference = 0.0814 g / 210L
Gas standard results
0.075
0. 070 +
10 15
167 Day
Figure 25
PlotofMeans and 2SD Error Bars For Both Simulator and Dry Gas Results
PercentBias
Intoxilyzer 8000
n = 10 at each point
Gas Standard Reference: 0. 100 g /210L
Simulator Reference: 0 .0827 g/210L
5
Simulator results
101
:
-5 + - - - - -
Gas standard results
1010 20 2515
Day168
Figure 26Percent Bias and One Coefficient ofVariation Error Bars Comparing
Simulator Tubing Configurations and Dry Gas Standards on the Intoxilyzer 8000
Percent Bias
- -
n = 10 ateach point
3 /19/ 2007
- 2
Simulator Without Cleared TubingDry Gas Standard
Simulator With Cleared TuabingA
169Testing Sequence
Figure 27Single MeasurementResults Where Increasing Amounts of Acetone
Are Added to a Simulator Initially Containing Only Ethanol
Result ( g /210L ).
Ethanolonly5 dropsacetone added
10 dropsacetone added
15 drops acetoneadded Interference Detected
0 . 12
0. 10
0 .08
0. 06Drager 9510 Intoxilyzer 8000 Datamaster Datamaster CDM
170Instrument
Figure 28
Single MeasurementResults Where Increasing Amounts of
Are Added to a Simulator Initially Containing Only Ethanol
Result ( g/ 210L )
0 . 14
Ethanol only
Isopropanol
400ul Isopropanol
Interference Detected0.
0 . 10
0 .08
0 . 06
Drager 9510 Intoxilyzer8000 DatamasterDMT DatamasterCDM
Instrument
171
Figure 29Results Where Increasing Amounts of Acetone Are Added to a Simulator InitiallyContaining Only EthanolFor the Drager 9510 to Compare the Response of the
IR and EC
Result (g / 210L )All samples introduced through breath tube
200ul acetoneyields approx. 926 ug/L vapor concentration0. 117
Interference DetectedIR Result
EC Result
0. 10
These two series had one
interferentand one quantified result0 . 09
meanof ten results
alongwith bars
0 .08
0.07
0. 06EthanolOnly
Acetone Acetone Acetone Acetone
1000uL
Acetone
172
Simulator Solution
Figure 30
PlotofBias and One Coefficientof Variation Error Bars ForFour-Point
Factory Calibration on the Intoxilyzer 9000
PercentBias
FactoryCalibrationn = 10 at each point
- - - - - - -
-
0.0 0 1 0 . 4 0 . 50. 2
Reference Value (g /210L)173
Figure 31Plotof Bias and One Coefficient ofVariation Error Bars ForFour-Point
Factory Calibration on the Intoxilyzer 9000
Percent BiasOne PointCalibration
n = 10 at each point
-
-
0 . 0 0. 1 0. 2 0 . 4 0 . 50 3
Reference Value ( g/210L )174
Figure 32SequentialResults From Two DifferentSimulators Tested on the Intoxilyzer8000
Concentration ( g/ 210L )
0.090
0 . 088Guth 590 Simulatormean 0843 g /210LSD = 0 0005 g /210L
Runs test: Z 0 .49 p = 0 .62
0. 086
0 . 084 bo boooo
0 . 082 + - - - - - - - -
Reference Value 0.0820 g /210L
0 .080Guth 2100 Simulator
mean = 0 .0830 g /210LSD = 0.0003 g /
Runstest: Z = 0 p = 1. 0
0 .078
10 15 20175
SequentialTest
Proposed Protocol For New Instrument Evaluation
Washington State Patrol
Breath Test Program
November 2006
A . Introduction
1. Purposea to evaluate new breath alcohol instrumentation and
technology for establishing type approval and possibleimplementation within Washington State
b . to develop a forensically sound breath alcohol test
program in Washington State centered around an
instrument that has been found to be fit - for -purpose
to develop documented evidence having a high degree of
assurance that a particular instrument type with perform
according to pre - determined specifications and
analyticalattributes
to provide documented for the validity and
subsequent selection of a particular instrument type
that has been shown to fit - for - purpose with a high
degree of confidence
2 . Manufacturers will be asked to come to Seattle and present
an overview of their instrument and leave one for evaluation
Manufacturers will be provided with a copy of this proposed
evaluation protocolThe evaluation results will be available to all interested
individuals5 . The evaluation is expected to take approximately six months
to completeThe instruments will be tested together as much as possible
in order to minimize any differences due to experimental
design , procedure or the use of other equipment .
A Datamaster CDM will also be included in the testing
protocol.
3 .
7 .
Desirable Instrumental Features
1 .
3.
Infrared and / or Electrochemical analytical method
Data collectionRe - cycle tubing for simulator analysis
Ability to continue using Guth 2100 digital simulators
a minimize length of simulator tubing
b ability to heat simulator tubing
c future purchase of new instruments must include a Guth
176