EMC Test Systems

download EMC Test Systems

of 24

Transcript of EMC Test Systems

  • 8/3/2019 EMC Test Systems

    1/24

    EMC TEST SYTEMS /

    INSTRUMENTS AND

    CONTROL SOFTWARE

    Compact Immunity Test System / BCI Test System - CIT-10acc. to IEC/EN 61000-4-6 / ISO 11452-4 / MIL-STD 461E

    Power Signal Generator - PSG-300acc. to EN 61000-4-16

    Compact Magnetic Field Test System /Low Frequency Test System - MTS-800acc. to automotive standards and MIL-STD 461E

    Immunity Test Systems - RIS 3000acc. to IEC/EN 61000-4-3

    Control SotwareRF-Lab, CD-Lab, BCI-Lab

  • 8/3/2019 EMC Test Systems

    2/24

    Content / Index

    EMC TEST SYSTEMS:

    Radiated Immunity Test Systemaccording to IEC/EN 61000-4-3, RIS 3000 3-5

    Conducted Immunity Test System / BCI Test System,according to IEC/EN 61000-4-6, ISO 11452-4, MIL-STD 461E, CIT-10 6-10

    Coupling / Decoupling Networks, CDNs 11-12Electromagnetic Coupling Clamp 12

    Compact Magnetic-Field Test System / Low Frequency Test Systemor Emission and Immunity Tests according to EN 55103-1/2,

    EN 61000-4-16, EN61000-4-8, SAE J1113-22, ISO 11452-8,MIL-STD-461E (CE101, RE101, CS101, CS109 and RS101),

    Automotive manuacturer standards, MTS-800 13-19Power-Signal-Generator acc. to IEC/EN 61000-4-16, PSG-300 20

    Control Sotware or Radiated Immunity Testsaccording to IEC/EN 61000-4-3, IEC/EN 61000-4-20,

    automotive- and military standards, RF-LAB 21Control Sotware or Conducted Immunity Tests

    according to IEC/EN 61000-4-6, CD-LAB 22Control Sotware or Immunity Tests with Bulk Current Injection

    according to ISO 11452-4, BCI-LAB 23

  • 8/3/2019 EMC Test Systems

    3/24

    Controlcomputerwith IEEEcard

    (NationalInstruments)

    Fieldmeasuringsystem

    Transmitting

    antenna

    Fieldprobe

    Testsite

    (anechoicchamber)

    Fiberopticcable

    RS232

    SignalgeneratorRFout

    RFout

    Reverse

    PowerForward

    Power

    Block diagram 1: Set-up o a test system according to IEC/EN 61000-4-3;Measurement o the uniorm area.

    RFin

    RF-Power-AmpliferFrankonia,

    type FLH/FLGGPIB

    Directionalcoupler

    RF-Power-MeterFrankonia,

    typePMS1084

    General

    For testing an EUT with respect to immunity to RF intererence in the

    requency range rom 80MHz to 6 GHz according to IEC/EN 61000-4-3, the EUT is exposed to a dened radiated electromagnetic eld.

    To avoid infuences on the equipment located in the environment and

    on the radio services, as well as or reasons o personal saety, this test

    should be perormed in RF shielded rooms (anechoic chambers).

    Set-up o Test Systems or testing in anechoic chambers.

    Block diagram 1 shows the set-up required or the measurement o the uniorm area.

    Control sotware

    RF-LAB is ully compliant to IEC/EN 61000-4-3, IEC/EN 61000-4-20,

    automotive- and MIL standards. The sotware RF-LAB controls the

    complete test system and prepares the test report. It perorms measure-ment o the uniorm area and generates reerence calibration data rom

    it. In alternative reerence data can be measured directly. Tests may beperormed manually and ully automatically. A ull automatic monitor-

    ing o the EUTs unction is possible whenever its compliance can be

    monitored with preset tolerance limits. Up to our values can be

    monitored and recorded or example by means o multimeters.

    Signal generator

    A commercial signal generator is used as signal source. It

    should cover at least the required requency range such

    as 80MHz to 6 GHz and allow amplitude modulation with

    a sine wave o 1kHz and 80%, as requested by the stan-dards. Besides, it should meet the requirements regarding

    requency step width (1% o the preceding value). I the

    equipment oers urther modulation depths and modes, aswell as a higher requency range and smaller steps, this may

    be advantageous or uture applications. Minimum require-

    ments are as ollows:

    Control computer

    The controller is a standard PC with operating system Microsot WIN-

    DOWS XP/VISTA. Depending on the system layout GPIB (IEEE 488) withNational Instruments interace card, serial bus RS 232, USB and other

    bus systems are supported.

    In the ollowing, the measuring and testing equipment is described in detail.

    Essential data o the sotware are:

    MicrosoftWindowsplatformWINVISTA,WINXP Simpleoperatorsguide Onlinehelpfunction Presentationoftheresultsinonlinegraphicsandreports Exportfunctionofthelesforfurtherprocessingunder

    Microsot Word, Microsot Excel, Measurementofhomogeneouseldincl.evaluation Calculationofreferencedata

    2dBsaturationtestonbaseofhomogeneouseldmeasurement compliant to standard Measurementofreferencedatawithxedtestlevel

    or prole o level vs. requency PermanentVSWRcontrolduringtestandoperator

    dened limitation as well as restriction o max.input level o amplier and max. allowed output power

    Automaticmultiplerepetitionoftest Manualtestmode Manualincrease/decreaseoftestlevel Automatictestmodeincl.monitoringoftheEUT HandshakefunctiontoEUTviaserialinterface Easyandfastgraphicaldeviceset-up,systemlayout

    can be printed FullycomplianttoIEC/EN61000-4-3,

    IEC/EN 61000-4-20, automotive- and MIL standards ControlofthetestsystembyGPIB,USBinterfaces, Customizedmodicationspossible

    Frequencyrange:9kHz3.0GHz(6.0GHz) RFoutput:-40dBmto0dBm Frequencyresolution:1Hz Levelresolution:0.1dB Amplitudemodulation:0to99.9% Furthermodulationtypes:frequencymodulation,

    phase modulation, pulse modulation Interface:GPIB(IEEE-488),RS232,USB)

    Immunity Test Systems according to IEC/EN 61000-4-3RIS 3000

    3//PAGE

  • 8/3/2019 EMC Test Systems

    4/24

    RF Power amplifer

    The sotware controls the level o the signal genera-

    tor output or each test requency. This signal level isamplied by the power amplier output in order to

    generate the required eld strength around the EUT. It

    depends on the testing set-up, the distance between

    EUT and antenna (1m,, 3m), and the test level / testeld strength (1V/m, 3Vm, 10V/m or special require-

    ments) whether an amplier output o 10W, 30W,

    200W,2kW, is required. Normally, a eld strength

    o 10V/m, with 1kHz / 80% AM, can be obtained with

    a 200W amplier in a testing distance o 3m.

    Power measurement

    During testing the eld strength probe is replaced by the EUT. A power measurement during the test

    runs assures that the EUT is actually exposed to the requested test conditions. For this purpose, adirectional coupler is connected at the amplier output. A power measuring equipment type PMS

    1084 determines the orward- and reverse power up to 6GHz. Both are stored and recorded by thecontrol sotware.

    Field strength measurement

    A eld strength measuring system consists o an isotropic probe, measuring unit and a bre-opticconverter which is connected to the PC. The system is used or the measurement o the eld homo-geneity o the test set-up.

    System Installation into 19 Rack: RIS 3000

    On request the test systems are supplied installed

    into 19-rack. Picture on the let shows an exampleo a radiated immunity test system acc. to IEC/EN

    61000-4-3, type RIS 3000.

    Transmitting antenna

    The broadband antenna Frankonia BTA-M covers the whole requency range rom 80MHz to 3000MHz.It may be loaded by up to 1000W. Above 1GHz horn antennas or small log-periodical antennas like

    our BTA-S save amplier power and costs. The antennas are also suitable or emission measurements.

    Main data:

    Type:FrankoniaFLH,FLG,

    Frequencyrange:1MHz-6GHz

    Nominaloutput:4W2000W

    Inputfornominaloutput:107dBmV (0dBm)

    Impedance:50Ohm

    Features:

    Moveablerack

    (on big rubber wheels) Instrumentsareinstalled

    into rack and cabled

    Mainswitch Emergencyswitch

    Easymountablecoverforfrontand rear side o the rack included

    Typical Dimensions (H x D x W):

    1,270 x 710 x 540mmWeight without instruments: 50kg

    Main data:

    Type: FrankoniaBTA-M FrankoniaBTA-S Frequencyrange: 30MHzto3GHz 0.7GHzto9GHz Impedance: 50Ohm 50Ohm Weight: 5kg 3.7kg

    Immunity Test Systems according to IEC/EN 61000-4-3RIS 3000

  • 8/3/2019 EMC Test Systems

    5/24

    Reerence data

    Reerence data can be determined with two dierent methods: (rom data o homogeneous eld) and .

    Calculate Reerence File

    Acc. to standard the reerence data o a desired test eld strength shall be calculated rom data measured in the eld homogeneity measurement.

    This can simply be done by use o this unction.

    Individual Reerence measurement (or individual test set-up)

    The sotware supports individual measurement o reerence iles. This may be necessary i an exactly deined test level at a deined

    EUT- position or a proile o test level versus requency is requested.

    Testing

    The testing set-up is described in block diagram2. On the basis othe reerence data (reerence calculation or reerence measure-

    ment) determined or the respective testing set-up, the requiredtest level and/or test ield strength is adjusted reproducibly in

    the test run by means o the control sotware or each requencystep. For this purpose, the signal is amplitude-modulated with

    a sine wave 1kHz / 80%. The veriication o the EUTs unction is

    perormed manually (optically) or, i possible, automatically. Inthe latter case the EUTs unction is tested by means o max. 4measuring values with respect to its compliance with preset tol-

    erance values. The determined data, the results, a descriptiono the test system as well as a comment regarding the measure-ment are summarized by the sotware in a measuring record.

    The output is realized by a printer connected to the PC. The datacan be exported or urther processing in other data processing

    programs, e.g. Microsot Word and Microsot Excel.

    Homogeneous / Uniorm Field

    To assure the reproducibility o the immunity test, the standard prescribes the homogeneity o the eld gener-

    ated. The anechoic chamber must guarantee a homogeneous eld within the size 1.5m x 1.5m in a distance

    o 1 to 3m rom the transmitting antenna, (e.g. Frankonia type BTA). I a smaller surace is sucient or expos-

    ing the EUT and its connection cables to radiation, the homogeneous eld can be reduced to 0.5m x 0.5m. The

    lowest part o the homogeneous eld surace is situated at 0.8m above the foor. To assure the correct displayo the eld probe the measurements are perormed without modulation in the empty anechoic chamber. The

    homogeneous eld has to be established in 1% steps, starting rom 80MHz up to 6GHz. The requested eld

    homogeneity or the respective requency is met, as soon as in 12 measuring points out o 16 (1.5m x 1.5m),or in 4 measuring points out o 4 (0.5m x 0.5m) the dierence between the highest and lowest eld strength

    value amounts to 0dB to +6dB. The high requirements regarding eld homogeneity cannot be met by normalRF-shielded cabins (without absorber lining). The sotware uses these data to check the 2dB saturation o the

    system as required by the standard.

    Block diagram 2: Set-up o a test system according to IEC/EN 61000-4-3

    in the test run.

    ControlcomputerwithIEEEcard

    (NationalInstruments)

    Transmitting

    antenna

    Testsite

    (anechoic chamber)

    Signalgenerator

    RFout

    RFout

    Reverse

    PowerForward

    Power

    RFin

    RF-Power-Amplifer Frankonia,

    typeFLH/FLGGPIB

    Directionalcoupler

    RF-Power-MeterFrankonia,

    typePMS1084

    Table

    EUT

    Position o the eld probe(equal distances)

    0.5m

    0.5m

    1.5m

    1.5m

    0.8m

    Uniorm area

    Ground

    Immunity Test Systems according to IEC/EN 61000-4-3RIS 3000

    5//PAGE

  • 8/3/2019 EMC Test Systems

    6/24

    Special Features:

    ConductedRFimmunitytestsacc.toIEC/EN61000-4-6

    andBCItestsacc.toISO11452-4andMILSTD461E Signalgenerator,RF-power-amplier,RF-power-meter

    and directional coupler (optional) in one 19 case

    Stand-aloneoperationpossiblewithoptionalavailablenetbook

    Control-softwareincluded

    Mostimportantparametersareshownonaintegrateddisplay

    AutomaticEUT-monitoring

    OperationviaUSBportofaPCorNotebook

    CompleterangeofCDNsavailable

    General:

    The CIT-10 is a complete test system or conducted RF- im-

    munity tests according to IEC/EN 61000-4-6, ISO 11452-4,

    MIL STD 461E/F CS114, SAE-J1113-2, DC 10614 and similar

    standards. Its internal RF generator and RF-power-amplier

    produce output signals with max. up to 150W within a re-quency range rom 100 (10)kHz up to 400MHz. Generated

    signals are measured via one o the max. 3 internal RF-Volt-

    meters. Furthermore via an optional, internal directional

    coupler orward and refected power can be measured. The

    whole test system allows ull automatic tests or the speci-ed requency range. As a stand-alone test system the

    CIT-10 is convincing by its easy and comortable handling

    and the excellent cost-perormance ratio. Add-ons like

    coupling/decoupling devices are available as well.

    Applications:

    Immunity Testing:

    Testing according to IEC/EN 61000-4-6, ISO 11452-4, MIL STD 461E/F CS114, DC10614 can be perormed automatically.

    Generation, amplifcation and verifcation o RF-Signals:The internal amplier amplies any signal rom 100 (10)kHz up to 400MHz. Using the internal generator can also generate a desired nar-rowband signal. Signals up to 30dBm can be measured at the same time. I a directional coupler is installed, orward and refected power

    are measured as well.

    SignalGenerator

    DirectionalCoupler(o

    pt.)

    RF-power-amplifier

    RF-power-meter

    IN

    ONEUNIT:

    acc. to IEC/EN 61000-4-6 / ISO 11452-4 / MIL-STD 461E

    Compact Immunity Test System CIT-10, 10kHz - 400MHz

  • 8/3/2019 EMC Test Systems

    7/24

    Internal RF-Power-Amplifer

    Several ampliier modules are available. Highest output

    power can be 150 W over the speciied requency range.The ampliier input can be accessed via the back panel o

    the CIT-10, so that the ampliier can also be used with

    any external generator. 20W, 75W and 150W ampliiers

    are available as standard.

    Internal RF-Signal Generator

    As the internal generator generates its output signal without anyinternal mixing components, low harmonics and spurious re-

    quencies are assured.

    Amplitude Modulation

    Frequencies generated by the generator can also be modulatedwith a LF signal. Modulation requencies may vary rom 1 Hz up

    to 100kHz, modulation levels are available rom 0% to 100%.

    Set-up:

    The CIT-10 is a PC-controlled test equipment. It can be operated by any commercial IBM compatible PC (WIN/NT/2000/XP/VISTA) via USB port. All

    settings o the equipment, e.g. start requency, stop requency, step width, test voltage etc. are made by means o the control sotware whichis also included in the delivery. The three unctional units Signal generator, RF power amplier and RF voltmeter are set automatically by the

    sotware, depending on the pre-set test parameters. Each component, however, may also be called and operated as separate measuring and

    testing equipment. This means: using the CIT-10 as testing system, you have three ull, additional single units at your disposal, or which

    separate inputs and outputs are available as BNC connections. Due to the computer-aided control o the CIT-10, any modications which may

    become necessary, or example, due to the revision o standards, may be perormed without problems and without having to manipulate the

    hardware o the equipment.

    Functioning:

    The equipment is ready or operation immediately ater connec-

    tion o the USB port, installation o the drivers and the control

    sotware. Ater starting o the control sotware, the main menu

    oers the manual control o and . Further options in the menu are (, ) and (,

    ).

    Internal RF-Voltmeter

    Accurate measurements o RF signals rom -40 dBm up to +30

    dBm are done by the internal RF-voltmeter which can be accessed(or separate use) via a BNC connector. Two internal voltmeters

    measure the orward and reverse power on an optional available

    directional coupler or i no directional coupler is installed, the

    output voltage o the amplier is measured.

    BCI-Tests with additional Power-Meter

    For BCI-Tests the CIT-10 can be equipped with up to 3 piec-

    es internal power meters.

    User defned signals

    External signals (e.g. EUT Fail or external instruments) can be con-

    nected and monitored using the application sotware.

    Features:

    acc. to IEC/EN 61000-4-6 / ISO 11452-4 / MIL-STD 461E

    Compact Immunity Test System CIT-10, 10kHz - 400MHz

    7//PAGE

  • 8/3/2019 EMC Test Systems

    8/24

    CDN-Calibration:

    The CDNs (Coupling/Decoupling Networks) serve to inject the

    test voltage into the lines to be tested and/or to decouple anyconnected peripheral equipment rom the EUT. The charac-

    teristics o the CDNs as well as o the power amplier are not

    absolutely linear over the whole requency range, i.e. the

    amount o power required to generate a constant test voltageover the whole requency range varies slightly, depending on

    the requency. In the calibration run, the requency-depen-

    dent output level o the signal generator, which is necessary

    or a constant test voltage, will be determined and stored in

    the sotware, together with the dened requency range andthe desired test voltage. The data records thus created may

    then be stored and recalled or tests.

    Sel-Calibration:

    When selecting this menu option, the test equipment will per-

    orm a sel-calibration. In this case, the output o the signal

    generator must be connected to the input o the voltmeter.

    Test:

    The menu option oers the selection possibilities , and . The settings or a

    test, e.g. start and stop requency, step width and test voltageare made automatically via the calibration le o the selected

    coupling unit. It is now possible to decide whether the test is

    to be perormed exactly according to these pre-settings, i.e.exactly as in the calibration, or whether modications o the

    pre-settings shall be admissible. I the calibration run was per-

    ormed, or example, or a test voltage o 10V, and the test is to

    be perormed now with 3V without having to perorm a new

    calibration run or this purpose, this can be done by selectingmenu item .

    Is a suitable measuring instrument connected to the specied se-rial port o the CIT-10, EUT can be monitored automatically. Data

    are shown graphically. During all test routines the amplier out-put is monitored in a bar display. This guarantees correct tests.

    In the case o , a test is perormed over the

    complete selected requency range; in this case the test requency is in-

    creased by the control sotware according to the selected step width andthe entered dwell time. I there is a malunction o the EUT, the test may

    be stopped at any time. It is then possible to either increase or reduce

    the requency by any number o steps, as well as to switch on and o

    the modulation and test voltage. Besides, a description o the malunc-tion occurred may be entered in a comment line which is included in

    the test record.

    oers the possibility o testing the EUT at discrete requen-cies. This can be done either with a xed test voltage or, optionally, with

    a ramp unction. In case o the ramp unction, the start and stop voltage,

    the step width by which the test voltage is to be increased, as well as the

    dwell time between the individual steps may be preset by the tester.

    The standard consists o the head o the protocol and a dia-

    gram which shows the test results. In the head o the protocol the date

    and time are taken over rom the computer; in addition, details like

    temperature, air humidity, tester, as well as testing set-up and EUT,

    may be registered. The protocol may be printed directly. It is also pos-

    sible to edit the protocol individually.

    acc. to IEC/EN 61000-4-6 / ISO 11452-4 / MIL-STD 461E

    Compact Immunity Test System CIT-10, 10kHz - 400MHz

  • 8/3/2019 EMC Test Systems

    9/24

    acc. to IEC/EN 61000-4-6 / ISO 11452-4 / MIL-STD 461E

    Compact Immunity Test System CIT-10, 10kHz - 400MHz

    Technical data:

    RF Voltmeter (external in-/output)

    Frequency range 10 kHz to 400MHz

    Measuring range +30 dBm to - 40 dBm

    Accuracy 0.5 dB

    VSWR < 1.1 : 1

    Input BNC, 50 Ohm

    RF Generator

    Output BNC, 50 Ohm

    Frequency range 10 kHz to 400 MHz

    Frequency resolution 1 Hz

    Output level range 0 to - 60 dBm

    Output level resolution 0.1 dB

    Output level accuracy 0.5 dB ( 1 dB max)

    Accuracy (requency) 5 ppm (TCXO)

    Harmonics < -30 dBc

    Non harmonics < -45 dBc

    Amplitude modulation(internal)

    0 to 100%; resolution 0.5%(internal AF Generator)

    Amplitude modulation

    (external)

    BNC jack

    1 Hz to 100 kHz, 0 to 100%Input impedance > 100 kOhm

    Pulse modulation

    variable duty cycle 10 - 90%;resolution 1 % (internal AF Generator)

    VSWR < 1.5:1

    AF Generator

    Output jack BNC

    Frequency range 1 Hz to 100 kHz

    Frequency resolution 0.1 Hz

    Output voltage 0 to 1 V amplitude; resolution 5mV

    Accuracy (requency) 50 ppm

    Signal Sine wave / square wave / triangular

    RF Voltmeter (internal, 2 pcs.)

    Frequency range 10 kHz to 400 MHz

    Measuring range +53 dBm to - 0 dBm

    Accuracy 0.5 dB

    Directional coupler (optional)

    Frequency range 10 kHz to 400 MHz

    Power 200 W CWInsertion loss 0.5 dB max

    VSWR 1.25 : 1 max

    Directivity 20 dB min

    Technical data:

    Power amplifer

    Frequency range 100kHz (10kHz) to 400MHz (75W, 150W)

    100kHz to 230MHz (20W)

    Gain 51 dB 1.5 dB

    Output power 75W/150W (optional)

    20W (100kHz to 230MHz)

    Distortion

  • 8/3/2019 EMC Test Systems

    10/24

    Circuit diagram o a test system according to IEC/EN 61000-4-6:

    Circuit diagram o a BCI test system:

    acc. to IEC/EN 61000-4-6 / ISO 11452-4 / MIL-STD 461E

    Compact Immunity Test System CIT-10, 10kHz - 400MHz

    RS232

    GPIB

    RS232

    max.

    107 dBV

    max.

    100W

    Device 1

    Keithley 2000

    Device 2

    CAN-Bus Test

    Device 3

    HP34401A

    Device 1

    R&S CMD 55

    PC

    Power MeterChannel 1

    CIT-10 RS

    ReerenceDevice

    CIT-10 RS

    Generator

    CIT-10 RS

    Amplifer 2

    No amplier

    Amplifer 1

    CIT-10

    Power MeterChannel 2

    CIT-10 RS

    EUTAE

    ~

    ~ -

    Coupling Network CDN801 M3

    Relay 2Relay 1

    C5019

    DirectionalCoupler DC

    AE

    Current ProbeFCC F-52

    Injection ClampFCC F-120-8

    RS232

    GPIB

    RS232

    max.

    107 dBV

    max.

    100W

    Device 1

    Keithley 2000

    Device 2

    CAN-Bus Test

    Device 3

    HP34401A

    Device 1

    R&S CMD 55

    PC

    Power MeterChannel 1

    CIT-10 RS

    ReerenceDevice

    CIT-10 RS

    Generator

    CIT-10 RS

    Amplifer 2

    No amplier

    Amplifer 1

    CIT-10/75

    Power MeterChannel 2

    CIT-10 RS

    Relay 2

    Relay 3

    Relay 1

    C5019

    DirectionalCoupler DC

    EUT

  • 8/3/2019 EMC Test Systems

    11/24

    Compact Immunity Test System CIT-10Coupling / Decoupling Networks (CDNs)

    Description

    Immunity tests according to IEC/EN require coupling o RF disturbance voltages into any

    conducting cable o an EUT. Furthermore these disturbances should not be coupled intoany urther equipment so that a decoupling path to any auxiliary equipment is provided.

    We oer a wide range o CDNs or dierent types o interconnected lines which are ully

    calibrated or the requency range rom 150kHz to 230MHz. The ollowing CDNs are avail-

    able: M-, AF-, S-, T-, RJ, USB-types. Almost any network can be assembled on customersrequests. Guidance or selecting the appropriate CDN is given in the ollowing table:

    TheEUTshallbeplacedonanisolatingsupport,0.1mabovethegroundreferenceplane.For table-top equipment, the ground reerence plane may be placed on a table.

    Onallcablestobetested,couplinganddecouplingdevicesshallbeinserted.

    Thecouplinganddecouplingdevicesshallbeplacedonthegroundreferenceplane,makingdirectcontactwithitatabout0.10.3mfromtheEUT.

    ThecablesbetweenthecouplinganddecouplingdevicesandtheEUTshallbeasshortaspossibleandshallneverbebundledorwrapped.

    Theheightabovethegroundreferenceplaneshallbebetween30and50mm(wherepossible). The6dBattenuatorshallbeplacedtothecouplinganddecouplingdeviceasnearaspossible.

    ThetestshallbeperformedwiththetestgeneratorconnectedtoeachoftheCDNsinturnwhiletheothernon-exitedRF-inputportsofthe

    CDNs are terminated by a 50 load resistor.

    The coupling and decoupling devices shall be placed on a ground reerence plane:

    *) The 150 loading (e.g. a 150 to

    50 adaptor terminated with a 50load) at the AE-port shall only be ap-plied to unscreened cables (screenedcables will have their screen con-nected to the ground reerence planeat the AE-side).

    Test procedure with Coupling/Decoupling (CDNs) Networks acc. to IEC/EN 61000-4-6:

    Calibration o Coupling Devices:

    Type Interconnected lines

    M1, M2, M3, M4, M5, M2+M3 Unscreened supply (mains)

    AF2, AF4, AF5, AF6, AF8 Unscreened nonbalanced lines

    S1, S2,S4, S8, S9, S15, S25, S36 Screened lines

    T2, T4, T8 Unscreened balanced lines

    RJ11, RJ45 Unscreened data lines

    RJ11/S, RJ45/S, USB Screened data lines

    11//PAGE

  • 8/3/2019 EMC Test Systems

    12/24

    1. The test generator (RF-out) shall be connected to the RF-input port o the coupling device via the 6dB-attenuator.2. The EUT port o the coupling device shall be connected in common-mode through the 150 to 50 adaptor to the RF-Voltmeters (calibration).

    3. The AE-port shall be loaded in common-mode with a 150 to 50 adaptor, terminated with 50.

    With direct injection to screened cable (CDN S types), the 150 load at the AE-port is not required as the screen will be connected to the

    ground reerence plane at the AE-port side.

    Although the 150 load at the AE-port is mandatory with CDN T, AF and M types calibration data are identical with the AE-port open or

    short. This is because a capacitor is connected to ground at the AE-port side, which leads to a RF-short-circuit similar to the CDN S types.This means that even with CDN M, AF and T types the 150 load at the AE-port is not required.

    Set-up or level setting at the EUT-port o coupling and decoupling devices:

    Electromagnetic Coupling Clamp acc. to IEC/EN 61000-4-6

    Direct Injection: Any shielded connection to an EUT can also be connected to the RF disturbance voltage via a direct injection adapter.

    Description

    The EM coupling clamp is used in all cases where CDNs are

    not advisable or not available. The lines to be tested are inthis case inserted in the clamp and the test voltage is inject-

    ed inductively. RF electromagnetic elds requently degrade

    electronic equipment by generating common mode currentson cables. The eect o these E and H elds on the equip-

    ment can be simulated by injecting common mode currents

    into the cables o the equipment being tested or RF im-

    munity. IEC/EN 61000-4-6 denes the methods or testing

    the immunity o electronic equipment to conducted modecurrents between 150kHz and 230MHz. The electromag-

    netic (EM) clamp is a high eciency broadband clamp-on

    injection device developed to test the immunity o electronicequipment when the standard IEC/EN 61000-4-6 CDN using

    the direct capacitive coupling technique is not possible norappropriate. The EM Clamp is oten used to test unshielded

    multiple conductor cables.

    To calibrate a CDN the ollowing items are required:

    adaptor fastening angle 50/150 adaptor

    Adapter with BNC jack, specic or each CDN,e.g. with 3 banana plugs or CDN M3

    Fasteningangle

    50 / 150 Ohm adpter withBNC connectors

    Fastening angle and 50/150 adaptor should be orderedor the rst CDN. For each ollowing CDN only the specic adap-

    tor has to be ordered.

    Nut

    Compact Immunity Test System CIT-10Coupling / Decoupling Networks (CDNs)

    Type F-203I-23MM

    Frequency range 10KHz to 1000MHz

    Max. allowed power:

    10KHz to 100 MHz: 100 watts CW or 15 minutes

    100 MHz to 230 MHz: 100 watts CW or 10 minutes

    230 MHZ to 1GHz: 50 watts CW or 10 minutes

    The test level o the immunity test with the specied valuescorresponds to a eld strength o 100V/m

    Directivity >10dB above 20 MHz

    Dimensions (L x W x H) 23 x 610 x 135 mm

    RF-disturbance connector N type

    Max. cable diameter 23mm

    Weight approx. 7.5 kg

    Options:

    Calibration set consisting o:- 2 pcs. calibration corners

    incl. 150/50 W adapter- 1 pcs. calibration cable

  • 8/3/2019 EMC Test Systems

    13/24

    General:

    The MTS-800 is a compact test system or broadband gener-

    ation and measurement o magnetic elds. Its internal com-

    ponents allow automatic EMC tests according to automotivestandards where high eld strength need to be generated

    or measured.

    In combination with our triaxial Helmholtz coils ull au-

    tomated susceptibility tests are possible at magnetic eldstrength up to 1000 A/m or requencies rom DC to 1 kHz.

    Lower eld strength can be generated or requencies up to

    250 kHz. Due to the triaxial set-up o our Helmholtz coil

    major improvement in device handling is achieved because

    there is no need to turn an EUT during tests.

    The MTS-800 complies to all magnetic eld requirements o

    relevant EMC and military standards.

    Tests and measurements are controlled by a program which

    will set most parameter automatically. For any relevant

    standard, which are ullled by the MTS-800, limit values

    are already included into the sotware package, although

    any dierent value can be dened by a user. Ater every testull reports will be created automatically. Report layout is

    pre-dened, though any user-dened layout is possible.

    High perormance is guaranteed by a sel-calibration processwhich utilizes an internal source as reerence.

    Special Features:

    Frequencyrangeforemissionandimmunitymeasurements:

    DC250kHz 800Wprecisionpoweramplier,signalgeneratorand

    spectrum analyzer in one compact unit

    Allinstrumentsmayaswellbeusedasstand-alonedevices

    Powerfulbuteasytooperatesoftware,fullyexpandable

    or uture standards modications

    Standardsoftwareallowseasyoperation,reportgeneration

    and integration o external measuring instrument or EUT

    monitoring

    PreparedforconnectionofexternalmultimeterforEUTcontrol

    FullyautomatedtestswithtriaxialHelmholtzcoil.Software

    controlled generation o magnetic eld in x-, y- and

    z- direction; no need to turn the EUT!

    Largevarietyofextensiveaccessoriesavailable

    800Wprecision

    poweramplifier,

    SpectrumAnalyzer,

    SignalGenerator

    IN

    ONEUNIT:

    According to

    EN 55103-1/2, EN 61000-4-16, EN 61000-4-8,SAE J1113-22, ISO 11452-8,

    MIL-STD-461E (CE101, RE101, CS101, CS109 and RS101),

    Automotive manuacturer standards

    Magnetic-Field Test System / Low-Frequency Test Systemor Emission and Immunity Tests / MTS-800

    13//PAGE

  • 8/3/2019 EMC Test Systems

    14/24

    Magnetic Field Generation

    MTS-800 enables a user to generate strong magnetic ields

    up to 1000 A/m. Even alternating ields up to 250 kHz canbe generated by the magnetic test system.

    Automotive Testing

    Intensive testing is required or new products which should

    be used in any automotive application. The MTS-800 al-lows ast and easy testing according to many automotive

    standards as described beore.

    Components

    MTS-800 consists o three independent module: a signal genera-tor(DC250kHz),apoweramplifier(800Woutputmaximum,

    DC1MHzbandwidth)andspectrumanalyzer(16Bit,1MSPS

    sampling rate). All modules can be used as stand-alone units.

    Automatic Testing Capabilities

    Full compliance with several immunity test asISO 11452-8, MIL-STD-461E/F RS101, CS101, CS109,

    IEC/EN 55103-2, IEC/EN 61000-4-8, SAE J1113-2,

    SAE J1113-22, Ford ES-XW7T-1A278-AC, GM W3097,

    PSA B21 7110, Renault 36-00-808, DC-11224,

    DC 10614 and similar standards.

    Furthermore the MTS-800 allows emission measurements ac-

    cording to MIL-STD-461E/F RE101, CE101 and IEC/EN 55103-1.

    Sotware

    Any unction is controlled via an application sotware which

    also guide the user through any test or measurement. Ad-

    aptation o signal strength or measurement graphs are pos-

    sible at any stage. User dened signals complement the

    usage or ast and reliable tests. The application sotwareis written in LabVIEW which guarantees stable and ast per-

    ormance on any Microsot Windows platorm.

    Accessories

    Frankonia provides also many dierent coils and loop sensor which

    are ideally suited or the described tests. Any additional equipment

    is ready to use without a need or recalibration. Not only our own

    equipment can be used with the MTS-800, but also user denedcoils. A calibration mode is included in the sotware to complement

    the magnetic test system with any urther equipment.

    Sel-calibration

    Using an ultra-stable voltage source sel-calibration correction

    values are stored in an internal EEPROM. Any voltage signal or

    voltage measurement device is calibrated at a sel-calibration

    process automatically in about a minute.

    Low Frequency emission and immunity tests

    acc. to MIL-STD 461E, CE 101, RE 101, CS 101, CS 109 and RS 101.

    Individual sotware modules and hardware accessories are avail-able or each o these tests.

    Applications:

    Features:

    Magnetic-Field Test System / Low-Frequency Test Systemor Emission and Immunity Tests / MTS-800

  • 8/3/2019 EMC Test Systems

    15/24

    The sotware starts with the generator/amplier control panel.This window allows basic settings o generator and amplier.

    Control Panel

    Open the Magnetic eld measurement window or spectrum

    analyzer measurements. Perorm a single or continuous mea-surement. Perorm test according to predened standards.

    Measurement results

    Edit a predened standard or create a new one.Load, save and print data.

    Example standard le or MIL-STD-461E / RS101_Navy

    Open the Magnetic eld generation window or suscep-tibility tests according to predened standards.

    Standard generation window

    Open the continuous generation window or long term

    magnetic eld test.

    Magnetic eld continuous generation window

    Magnetic-Field Test System / Low-Frequency Test Systemor Emission and Immunity Tests / MTS-800

    Further eatures and possibilities:

    Susceptibility tests with xed requencies and test levels or

    use the ramp unction to sweep rom start to stop level. Veriythe generated eld o any radiating coil with loop sensor.

    Short term generation window or short term magnetic eldtests (optional).

    Scope mode window.

    Determine the coil actor o an unknown coil

    Sel calibration o the MTS-800

    15//PAGE

  • 8/3/2019 EMC Test Systems

    16/24

    Magnetic-Field Test System / Low-Frequency Test Systemor Emission and Immunity Tests / MTS-800

    Type MTS-800

    Voltage input (Analyzer)

    Frequency range DC - 250 kHz

    Input impedance 1MOhm / 50 Ohm switchable

    Connector XLR, unbalanced

    Max. input voltage 100V continuous (attenuator autoset at overvoltage); 10 V at 50 Ohm

    Gain -20/0/20 dB Preamplier, 0/20/40 dB ADC Amplier; Sel-calibration with ultra stable on-board reerence

    Current input

    Frequency range DC - 250 kHz

    Shunts 10mOhm / 1 Ohm / 100 Ohm

    Max. input current 20 A continuous (overload protection); 1 Ohm and 100 Ohm shunt are protected by an additional 1.5 A useConnector 4 mm saety jack (+, -) measurement via insulation amplier or input jacks

    Measurement range 20A, 10A, 1A, 100mA, 10mA, 1mA automatic oset and gain;Sel-calibration with ultra stable on-board reerence

    AD converter

    Resolution 16 Bit

    Sampling rate 1.25 MSPS

    Aliasinglter 0.01dB Tschebysche lter, g = 260 kHz; lter may be switched o

    Generator

    Frequency range DC - 250 kHzOutput impedance 50 Ohm

    Connector BNC, unbalanced

    Signal Sine wave / triangular /square wave/ DC

    Amplitude 0 - 10V AC, - 10V - + 10V DC

    Resolution 12 Bit (2.5mV), Switchable - 20dB Attenuator; Sel-calibration with ultra stable on-board reerence

    Amplifer

    Frequency range DC - 1 MHz

    Connector 4mm saety jacks (output); BNC, unbalanced (input)

    Current 16Arms

    Voltage 50Vrms / 75 VDC

    Distortion (DC-100kHz, load 4Ohm) < 0.10%

    General data

    EUT control / Connector 9-pin Sub-D; RS-232

    Connection to Computer USB

    Temperature range 0 to 40C

    Warm-up time 15 min.

    Housing 19 Subrack or desktop case

    Width / heigth / depth 449 mm / 177 mm / 580 mm

    Weight (shipping) approx. 40 kg (net 34kg)

    Gain 10 0.1% (0.01 % /C)

  • 8/3/2019 EMC Test Systems

    17/24

    CommonmodetestadapterforbalancedsignalandcontrolconnectionsaccordingtoEN55103-3

    CalibrationnetworkforcommonmodetestadapteraccordingtoEN55103-2

    CurrenttransducerforbalancedvideoconnectionsaccordingtoEN55103-2

    EnclosedvariabletransformerforshorttermeldaccordingtoEN61000-4-8;prim.230V,sec.0to230V,max.current20A;incl.supplycable

    Accessories:

    Helmholtz Coils:Several Helmholtz coils are available or susceptibility tests. We also oer tri-axial Helmholtz coils which are suitable or MTS-800.To achieve 1000 A/m at 1 kHz, it is absolutely necessary to use our Helmholtz coils and an optional available compensation board.

    Helmholtz coil HCS_50/28_TAP with Loop Sesor RLS_133

    Triaxial Helmholtz coil HCST_50/28_TAP

    Magnetic-Field Test System / Low-Frequency Test Systemor Emission and Immunity Tests / MTS-800

    Coil-Type Specification

    HCST_50/28_TAP Tapped triaxial Helmholtz coil or immunity tests.HCS_50/28TAP Tapped single axis Helmholtz coil or immunity tests.

    Designed or the generation o magnetic elds with eldstrength > 1000A/m.

    HCS_125/75_TAP Tapped single axis Helmholtz coil or immunity testsaccording to EN 55103

    17//PAGE

  • 8/3/2019 EMC Test Systems

    18/24

    Loop Sensors / Radiating Loops:

    For immunity tests radiating loops are necessary to generate magnetic ields.

    Suitable loops are available. Measuring emissions require loop sensors whichcan also be ordered rom us.

    Loop sensor LS_040

    Radiating Loop / Loop Sensor RLS_133

    Radiating loop RL_120

    Magnetic-Field Test System / Low-Frequency Test Systemor Emission and Immunity Tests / MTS-800

    Coupling Transormer:

    MIL-STD-461E CS 101 requires a coupling transormer or conducted suscep-

    tibility tests. We developed a coupling transormer which meets all require-

    ments. Due to direct coupling to voltage mains, the coupling transormer

    has an additional dierential ampliier or common mode rejection o the AC

    mains. Using the coupling transormer without this ampliier can destroy anymeasurement instrument due to overvoltage.

    Coupling transormer CT2.5/50AC with dierential ampliier

    Coil-Type Specification

    RL_120 120mm radiating loop according to MIL-STD-461E

    LS_040 Electrostatically shielded loop sensor according to MIL-STD-461E

    LS_133 Electrostatically shielded loop sensor according to MIL-STD-461E

    RLS_133 Electrostatically shielded loop sensor according toEN 55103-1/2. Useas Radiating loop and loop sensor

  • 8/3/2019 EMC Test Systems

    19/24

    Accessories selecting table:

    Magnetic-Field Test System / Low-Frequency Test Systemor Emission and Immunity Tests / MTS-800

    Current transducer incl. correction network Calibration network Common mode test adapter

    Testing equipment acc. to EN 55103-2:

    EN 55103-2 requires certain immunity tests or requencies rom 50 Hz to 10 kHz. The ollowing test equipment ulils all requirements

    according to EN 55103-2, annex B.

    Technical speciications or all accessories are available on request.

    Test equipment MIL-STD 461E Recommended Model CE101 CS101 CS109 RE101 RS101

    Measurement receiver MTS-800

    Current probe Pearson Model 3525

    Signal generator MTS-800

    Power amplier MTS-800

    Data recording device MTS-800

    Oscilloscope MTS-800

    Coupling transormer CT_2.5/50AC

    Isolation transormer Any commercially available model

    LISNs Any commercially available model

    Radiating loop 12cm RL_120

    Loop sensor 4cm LS_040

    Loop sensor 13.3cm LS_133

    Ohmmeter Any commercially available model

    Standards

    CE101 Conducted Emission, Power Leads, 30Hz to 10kHz

    CS101 Conducted Susceptibility, Power Leads, 30Hz to 150kHz

    CS109 Conducted Susceptibility, Structure Current, 60Hz to 100kHz

    RE101 Radiated Emission, Magnetic Field, 30Hz to 100kHz

    RS101 Radiated Susceptibility, Magnetic Field, 30Hz to 100kHz

    19//PAGE

  • 8/3/2019 EMC Test Systems

    20/24

    Features:

    - Short circuit and overload protection- Completely linear and low noise design

    - Outstanding DC stability

    - Over temperature switch o

    - Protection / Ready LED

    The number one choice or allapplications with the need or astand powerul signals, e.g.:

    - Simulation o DC/AC supply lines- Generation o magnetic elds with

    Helmholtz or similar coils

    - Control o piezo actors

    - Immunity testing according to

    EN 61000-4-16

    - Calibration devices etc.

    General:

    The PSG-300 contains a linear precision power amplier with a wide bandwidth (DC-300kHz), suitable or all applications concerningast alternating signals at high output power. The built in generator provides sine, square and triangle waves. Communication between

    PSG-300 and PC is via USB connection. The application soware is suited or general power generator applications and or immunity testsaccording to EN 61000-4-16 as well as to IEC/EN 61543. Short term tests are enabled by phase controlled switching o an external power

    source. The PSG-300 is equipped with a silent, temperature-controlled an. Internal saeguards protect the amplier rom overheating and

    high power dissipation. They also assure protection against short-circuits and overload.

    accordingto:EN61000-4-16,DC300kHzandIEC/EN61543,1kHz150kHz

    Power Signal Generator PSG-300

    Type PSG-300 / 260W PSG-300 / 600W

    Amplifer:

    Frequency range DC - 1 MHz (small signal -3 dB)

    Power bandwidth DC200kHz

    Slew rate 100 V/s

    Oset 1 mV (0.1 mV/C)

    Gain 10 0.1 % (0.01 %/C)

    Output voltage 50 Ve / 75 Vpeak

    Output current 5 Ae / 7.5 Apeak 16 Ae / 23 Apeak

    Distortion(DC100kHz,load 4 ) < 0.10%

    Input impedance 100 k

    Max. input voltage 80 V (cont.), 100 V (< 1 min)

    Noise(10Hz1MHz,input:50) 0.5 mVe

    Power dissipation (each side) 260 W (100 ms) 600 W (100 ms)

    Output connector 4mm MCOutput connector 50 Ohm BNC

    Generator:

    Frequency range DC, 0.05Hz - 300 kHz

    Frequency resolution 0.05 Hz

    Frequency accuracy +/-20ppm

    Waveorm Sine, square, triangle

    External generator input BNC

    Input connector or phase controlled switching o external

    switching o external power source power source

    General data:Remote control USB connector

    Dimension (B x H x T) 84 TE x 3 HE x 410 mm 84 TE x 4 HE x 580 mm

    Weight approx. 14 kg approx. 40 kg

  • 8/3/2019 EMC Test Systems

    21/24

    Special Features:

    Online-helpfunction Graphicaldevicespecication Inputofcalibrationdataforalldevices Testfunctionforsystemcheck Generationofcalibrationlefortestingbycalculation

    rom homogeneous eld data or measurement Saturationtestoftestsystemasrequestedbystandard Testingbymanualopticalmonitoring Allows detailled examinationby manual increase /

    decrease o the applied test level during test Testingbyautomaticalmonitoring/evaluationof

    EUT with up to 4 measuring instruments Measurementandevaluationofelduniformity Presentationofresultsinon-linegraphics

    Professionalreportingsystem ExportfunctionforlesforfurtherprocessinginMicrosot Word, Microsot Excel, ...

    German/Englishlanguageswitchable Customizedmodicationspossible

    Technical Data:

    Microsot Windows-platorm: WIN XPRequires standard PCControl o test system byIEEE-488 (GPIB)-BUS, RS232-bus and USBRecommended GPIB-interace card: National Instruments

    Short Description:

    The sotware RF-LAB controls radiated immunity test systems. The sys-

    tem conguration is done in a graphical set-up. The calibration les or

    the testing can be calculated rom the uniormity measurement or mea-

    sured in calibration runs. These data are stored. Tests can be perormed

    manually with optical monitoring o the EUT or ully-automaticallyby up to 4 measuring instruments. Besides the sotware supports the

    measurement and evaluation o eld uniormity as described in IEC /

    EN 61000-4-3. Calibration data, test results, and eld uniormity are

    presented in proessional reports, which contain all necessary data.

    acc. to common immunity test standard IEC / EN 61000-4-3, IEC/ EN 61000-4-20, automotive- and military standards

    RF-LAB / Control Sotware or Radiated Immunity Tests

    21//PAGE

  • 8/3/2019 EMC Test Systems

    22/24

    acc. to common immunity test standard IEC / EN 61000-4-6

    CD-LAB / Control Sotware or Conducted Immunity Tests

    Special Features:

    Online-helpfunction

    Graphicaldevicespecication Inputofcalibrationdataforalldevices Testfunctionforsystemcheck Testingbymanualopticalmonitoring

    Allows detailled examination by manual increase /decrease o the applied test level during test

    Saturationtestoftestsystemasrequestedbystandard Testingbyautomaticalmonitoring/evaluationof

    EUT by up to 4 measuring instruments Supportsmeasurementoflterattenuation Presentationofresultsinon-linegraphics Professionalreportingsystem Exportfunctionforlesforfurtherprocessingin

    Microsot Word, Microsot Excel, German/Englishlanguageswitchable Customizedmodicationspossible

    Technical Data:

    Microsot Windows-platorm: WIN XPRequires standard PCControl o test system byIEEE-488 (GPIB)-BUS, RS232-bus and USBRecommended GPIB-interace card: National Instruments

    Short Description:

    The sotware CD-LAB controls conducted immunity test systems. The

    system conguration is done in a graphical set-up. The sotware per-orms calibration runs and stores the data thereo. Tests can be perormed

    manually with optical monitoring o the EUT or ully-automatical by upto 4 measuring instruments. Calibration data and test results are pre-

    sented in proessional reports, which contain all necessary data.

  • 8/3/2019 EMC Test Systems

    23/24

    acc. to common immunity test standard ISO 11452-4

    BCI-LAB / Control Sotware or Immunity Tests with Bulk Current Injection

    Specail Features:

    Online-helpfunction

    Graphicaldevicespecication Inputofcalibrationdataforthedevices Testingbymanualopticalmonitoring Testingbyautomaticalmonitoring/evaluationof

    EUT by up to 4 measuring instruments Saturationtestoftestsystemasrequestedbystandard Presentationofresultsinon-linegraphics Professionalreportingsystem Exportfunctionforlesforfurtherprocessingin

    Microsot Word, Microsot Excel, ... Testfunctionforsystemcheck German/Englishlanguageswitchable Customizedmodicationspossible

    Technical Data:

    Microsot Windows-platorm: WIN XPRequires standard PCControl o test system byIEEE-488 (GPIB)-BUS, RS232-bus and USBRecommended GPIB-interace card: National Instruments

    Short Description:

    The sotware BCI-LAB controls BCI-immunity test systems. The sys-

    tem conguration is done in a graphical set-up. The sotware supportsboth, online control method and calibrated clamp method. Tests can

    be perormed manually with optical monitoring o the EUT or auto-matically by up to 4 measuring instruments. Test results are presented

    in proessional reports, which contain all necessary data.

    23//PAGE

  • 8/3/2019 EMC Test Systems

    24/24

    Tel : +49 (0) 91 91 / 73 666 0 Web http://www rankonia emv com

    Frankonia EMC Test-Systems GmbH

    Daimlerstrae 17