Safety for electronic systems ESD Simulator Verification Greg Senko Business Manager - EMC Test...
-
Upload
adrian-callahan -
Category
Documents
-
view
220 -
download
2
Transcript of Safety for electronic systems ESD Simulator Verification Greg Senko Business Manager - EMC Test...
safety for electronic systems
ESD Simulator Verification
Greg SenkoBusiness Manager - EMC Test EquipmentSchaffner EMC
Ken WyattHardware Test Center ManagerAgilent Technologies, Colorado Springs
Copyright 2003 Schaffner EMC - All rights reserved
Van de Graaff generator, Boston Museum of Science (photo © 2003 by Kenneth Wyatt)
safety for electronic systems
Virtually every EMC laboratory has one or more ESD simulator.
Almost none are equipped to verify the ESD simulators’ performance.
We will cover:
• Verification techniques, including ISO, SAE, ANSI and IEC standards
• Proposed changes in the measurement setup
• Practical aspects of measurement setup and performance
• Live demonstration
Copyright 2003 Schaffner EMC - All rights reserved
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
3
safety for electronic systems
What parameters must be measured?
Tip voltage
Current waveform
• Peak
• Rise
• Current at 30ns
• Current at 60ns
• Time Constant (air discharge, auto manf)
• Current derivative - ANSI Draft (gives indication of smoothness)
Positive peak
Negative peak
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
4
safety for electronic systems
Measuring tip voltage
Measured at standard test levels:
±2kV, ±4kV, ±6kV, ±8kV, ±15kV and ±25kV
Measured using Electrometer or Giga-ohm meter
Most standards don’t specify requirements
ISO 10605 specifies 100 GOhm minimum input impedance
The simulator’s tip voltage not affected by the measurement
If a Giga-ohm meter is used, the simulator must continuously chargethe high-voltage capacitor
- Many older simulators provide an initial charge only,which can bleed off with time or with load
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
5
safety for electronic systems
Tip voltage measurement using Giga-ohm meter (Brandenburg Model 139D)
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
6
safety for electronic systems
Idealized ESD simulator waveform
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
7
safety for electronic systems
Actual waveform measurement (Tek 7104)
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
8
safety for electronic systems
How do we measure the current waveform?
A low impedance shunt (ESD target) is used to represent adischarge into a large metallic object
The shunt impedance is < 2.1 Ohms
Block diagram:
ATTENUATOR ATTENUATOR OSCILLOSCOPE
TARGET
GROUND PLANE
CABLE
Optional Attenuator for > 8 kV (20 dB)
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
9
safety for electronic systems
Typical ESD current measurement system
NOTE:
The reason a Faraday cage waswritten into the original standardwas that the analog phosphorstorage oscilloscopes were generallysusceptible to the high field energyproduced by simulators.
The digitizing oscilloscopes todayare much more immune and theFaraday cage is no longer a must. Youmust confirm your measurement systemis unaffected, however!
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
10
safety for electronic systems
Typical ESD current measurement system
ESD measurement systemat Schaffner, Switzerland.
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
11
safety for electronic systems
Typical ESD current measurement system(Agilent lab)
1.2m ground plane clampedto ESD table.
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
12
safety for electronic systems
Typical ESD current measurement system
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
13
safety for electronic systems
Performing a contact discharge into the older ESD target
Keytek MZ-15EC MiniZapSimulator.
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
14
safety for electronic systems
Target design history
IEC 801-2: 1991
• No longer referenced by any current ESD standard• No performance specifications• Poor design - lots of ringing
IEC 61000-4-2: 1995
• Referenced by virtually all current ESD standards• No performance specifications• Transfer function “zero” at 5-6 GHz
ANSI C63.16 Draft 9
• Proposed new design (uses sm resistors and tapered transitions)• Flat to 6GHz• “Driving” adapter to evaluate high frequency performance
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
15
safety for electronic systems
IEC 801-2 target “ball tip”
Old design is no longer specified
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
16
safety for electronic systems
IEC 61000-4-2 target
Presently specified in standards
Example: EMCO CTC-3,and others
The large flat disk tendsto build up a pre-coronadischarge, which slowsthe risetime and leads tovariable results for air-discharge measurements.
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
17
safety for electronic systems
ANSI C63.16 target
Proposed design
Example: Schaffner MD-102,Amplifier Research CTR-2,and others
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
18
safety for electronic systems
Old versus new ESD targets
EMCO CTC-3 (left)Schaffner MD-102 (right).
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
19
safety for electronic systems
New target with “driving” adapter to measure transfer characteristics
Schaffner MD-102
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
20
safety for electronic systems
ANSI C63.16 target specifications
Reflection coefficient of target and adapter < 0.1
• Equivalent to VSWR < 1.22
Insertion loss < 0.3dB up to 4 GHz
Variation of attenuation of the target
-attenuator-cable chain
< ±0.3dB from DC to 1GHz (< ±3.51%)
< ±0.8dB from 1GHz to 4GHz (< ±9.65%)
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
21
safety for electronic systems
Waveforms of IEC 801-2 target vs. ANSI target
IEC 801-2Target
ANSI Target -less HF ringingand shows truepeak shape
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
22
safety for electronic systems
Actual waveform measurement (Agilent 54855A, 1.5 GHz BW)
Old target New Target
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
23
safety for electronic systems
Actual waveform measurement (Agilent 54855A, 6 GHz BW)
Old target New Target
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
24
safety for electronic systems
Choosing attenuators
• Target transfer function is ~1V/A when loaded by 50 Ohms
• Contact mode peak current at 8kV is ~30A
• Input range of most oscilloscopes is < 10V in 50 Ohm mode
• Therefore, an attenuator is needed to reduce the signal level
• 20dB is typically chosen for 10:1 ratio
• Contact mode to 25kV may require additional attenuation
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
25
safety for electronic systems
Choosing attenuators
• Low power attenuators may damaged by the short term peak power
• Attenuators are available with 1kW peak power ratings
• Use an 18GHz attenuator with low SWR, < 1.25 to 8GHz
• The attenuator accuracy requires that the entire chain be calibrated
Accuracy variation dB Percentage0.1 1.16%0.3 3.51%0.5 5.93%0.7 8.39%0.9 10.92%
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
26
safety for electronic systems
Choosing cables
• A low loss cable is required
• Cable length < 1m is required by most standards
• Double shielding is required by most standards
• The ANSI standard recommends RG 400
• RG 214 is twice the dia, 1/2 the loss and is commonly available
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
27
safety for electronic systems
Oscilloscopes - Bandwidth
All standards require at least 1GHz bandwidth
The BW/risetime of the oscilloscope is the single most limiting factor to accurately measure the pulse risetime
The true risetime is related to the observed risetime as follows:
The above correction is proposed in the ANSI draft standard and
assumes a Gaussian rolloff in frequency response. However most
digitizers use a sharper cutoff filter, 20dB/decade or higher.
2_
2_ chaintmeasuremenobservedpulseesd risetimerisetimerisetime
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
28
safety for electronic systems
Oscilloscopes - Bandwidth
How does bandwidth affect observed risetime?
Let‘s assume a Gaussian rolloff
Pulse Risetime
(ps)
Scope BW (GHz)
Scope Risetime
(ps)
Observed Risetime
(ps) Difference % Error
700 1 350 783 82.62379 11.8%
700 1.5 233 738 37.86479 5.4%
700 2 175 722 21.54348 3.1%
700 3 117 710 9.655629 1.4%
700 4 88 705 5.447553 0.8%
700 6 58 702 2.42635 0.3%
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
29
safety for electronic systems
Oscilloscopes - Sampling rate
Single-shot sampling rate is the key
• A fast-edge triangular peak requires fast sample rate
• Risetime of 800 ps from 10%-90% is 80% of waveform
• 10 Gs/s = 100 ps/sample
• 8 samples in 800 ps or 10%/sample!
• Since peak is symmetrical and somewhat rounded actual error is < 5% (assumes a triangle shape)
Effective sampling rate increased by capturing multiple shots
• Must have stable waveform
• Useful for contact mode only - never for air discharge
• Shot to shot variation is low for most simulators
• Should be used for verification - not for calibration
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
30
safety for electronic systems
Shot-to-shot variation - 20 shots
33.3 A peakStd dev .425 ±0.64% ofpeak
898 ps RiseStd dev 11.9 ±0.66% ofrisetime
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
31
safety for electronic systems
Oscilloscopes - Sampling rate
SAE and ISO recommend 4Gs/s minimum
2 Gs/s - 27.87A-16.0%
5 Gs/s - 31.92A-3.8%
20 Gs/s - 33.18A
10 Gs/s - 32.23A-2.9%
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
32
safety for electronic systems
Oscilloscopes – Interpolation - sin(x)/x
ON or OFF?
2 Gs/s - 29.98A-9.6%
5 Gs/s - 32.32A-2.6%
2 Gs/s - 27.87A-16.0%
5 Gs/s - 31.92A-3.8%
Interpolation ON Interpolation OFF
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
33
safety for electronic systems
Calibrating the target-attenuator-scope chain
It is recommended that the DC transfer function of the entire chainbe measured as follows:
• Inject a known current• Measure the resulting voltage at the oscilloscope• The attenuation factor = Injected current / observed voltage• Attenuation factor is used to correct waveform amplitude
ATTENUATOR ATTENUATOR OSCILLOSCOPE
TARGET
GROUND PLANE
CABLECURRENT SOURCE
Optional Attenuator for > 8 kV (20 dB)
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
34
safety for electronic systems
Other factors - Do’s and don’ts
Shielding
• Do we need it?
Position of ground cable
• Will it affect waveform?
Orientation of simulator
• Will it affect waveform?
Automatic Measurements
• Must use Min and Max values to calculate 10% and 90% points
Other cables
• Keep them well separated
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
35
safety for electronic systems
Oscilloscope shielding - Do we need it?
Standards say yes, but probably not necessary - use distance test
Scope insideFaraday cage
Scopenext tosimulator
Scopeat cornerof plane
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
36
safety for electronic systems
Ground cable position
Does affect results - peak, rise and duration
Natural loop
Loop closerto plane
20 Gs/s - 33.6A,891ps
20 Gs/s - 36.9A,926ps
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
37
safety for electronic systems
Simulator orientation to target
Does affect results - peak, rise and duration
Simulator on axis
Tip down 30º
Simulator tip down 10º
20 Gs/s - 33.6A,891ps
20 Gs/s - 33.6A,913ps
20 Gs/s - 34.5A,945ps
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
38
safety for electronic systems
Air discharge - What risetime/peak do you want?
Approach speedand environmentalfactors will greatlyaffect results - notRepeatable!
Obtaining a passingwaveform is amatter of patience!
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
39
safety for electronic systems
Measurement uncertainty
The estimated bounds of the deviation of a measured quantity from its true value
• List all the possible error sources and compute the uncertainty
• Uncertainty budget for each measured parameter
• Statement of confidence that can be placed in the value of uncertainty
• Does measured result truly fall within acceptable limits?
National Association for Measurement and Sampling publication NIS81, The Treatment of Uncertainty in EMC Measurements
Link to CE-Mag site
http://www.ce-mag.com/ARG/Senko.html
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
40
safety for electronic systems
Target plane size
ANSI - 1.2m x 1.2m, IEC 1.5m x 1.5m, ISO - N/A, SAE - N/A
Mini TargetPlane
1.2m2
Target Plane
20 Gs/s - 31.85A
20 Gs/s - 33.18A
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
41
safety for electronic systems
Demonstration equipment
Simulator: Schaffner NSG 435 / Keytek Minizap MZ-15EC
New Target: Schaffner MD 102 (designed to new ANSI stnd)
Old Target: Emco CTC-3 (designed to meet IEC 61000-4-2 stnd)
Target Plane: Small sized plane for demo purposes
Attenuator: Weinschel Model 2-20, 20dB, 5W, 1000W peak
Cable: RG-214 1m
Oscilloscope: Agilent Infiniium 54855A 6GHz BW, 20Gs/s scope
ESD Monitor: Credence Technologies CTC034-3 (counts andbeeps for each ESD event) www.credencetech.com
Copyright 2003 Schaffner EMC Inc. - All Rights Reserved
42
safety for electronic systems
Thank you for your attention
Your feedback is welcome
Greg SenkoBusiness Manager - EMC Test EquipmentSchaffner EMC
(603) [email protected]
Ken WyattSr. EMC EngrHardware Test MgrAgilent Technologies
(719) [email protected]