ELEC 371 -Orthopaedics
Transcript of ELEC 371 -Orthopaedics
Bones• Structural supportJoints• MobilityLigaments• Connect bone to bone• Alignment & supportTendons• Connect muscle to bone• Energy transfer
1
ELEC 371 - Orthopaedics
PeriosteumContiguous exteriorAttachment site (tendons / ligaments)Absent on articular surfaces
Cortical BoneStructural rigiditySome Blood vessels
TrabecularCancellous or Spongy BoneFew blood vessels
Marrow CavityFlexible tissueMany blood vessels
2
ELEC 371 - Orthopaedics
Anatomic Axes
• actual anatomic bone path
• one for each bone
Mechanical Axes
• effective straight-line link between joint rotation axes
• one for each bone
Load Bearing Axis
• effective straight-line link between hip and ankle
• one for entire limb
• depends on limb configuration (eg. knee angle)
ELEC 371 - Orthopaedics
3
Hinge
Elbow, Knee
Knuckle
Pivot
Vertebral Column
Saddle
Thumb / Finger / Toe (base of digit)
Elipsoid
Wrist / Ankle
Hand / Foot
Gliding (planar)
Hand / Foot
Vertebral Column
Ball & Socket
Hip
Shoulder
ELEC 371 - Orthopaedics
4
Unlike the spine, there is no joint that provides axial rotation of any of the limbs or digits.
For example, axial rotation of the wrist involves coordinated motion of the ulna and radius.
ELEC 371 - Orthopaedics
5
6
Photo of a THA procedure taken just after:
• the incision into the lateral hip was made
• the skin and muscle was retracted
• the femur was dislocated from the acetabulum
Damage to the head of the femur is easily visible.
This damage is too severe to heal on its own.
The joint must be replaced to restore patient mobility.
ELEC 371 - Orthopaedics
This is not the same patent as the previous photo
Pre-surgical x-ray shows
• Substantial wear of the femoral head
• Substantial damage to the acetabular cup
Post-surgical x-ray shows
• Stem implant inserted into femur
• Spherical joint fitted to implant neck
ELEC 371 - Orthopaedics
7
Conventional THA surgical tools include
• hammers
• slide hammers
• broaches
• cup reamers
ELEC 371 - Orthopaedics
8
Stem implants
• many lengths, widths, curvatures
• rough proximal surface for bone in-growth
• tapered neck for spherical joint interface
• proximal locking mechanisms for revision procedures
Cup implants
• rough exterior surface for bone in-growth
• screw holes
• snap-in liner
ELEC 371 - Orthopaedics
9
Biocompatibility is mandatory
• rejection by autoimmune system requires immediate removal & replacement
Osteoinduction is mandatory
• if bone does not heal, mobility is never restored
Osteoconduction is desired
• implant stability is improved
• prevented by foreign substance such as bone cement
Osteointegration is desired
• eliminates need for revision surgery
ELEC 371 - Orthopaedics
10