Detectin' so you can go Fishin' Seminar. The Purgin ... · The...
Transcript of Detectin' so you can go Fishin' Seminar. The Purgin ... · The...
Slide 1Dial 1-904-779-4740 for e-Seminar Audio
OPTIMIZING FOR VOLATILES ANALYSIS:The Purgin'-Trappin'-Desorbin'-Resolvin'-Detectin' so you can go Fishin' Seminar.
Slide 2Dial 1-904-779-4740 for e-Seminar Audio
VOC ANALYSIS BY PURGE AND TRAP IS A COMPROMISE
Slide 3Dial 1-904-779-4740 for e-Seminar Audio
EPA METHOD 502/524 COMPOUNDS1. Dichlorodifluoromethane 21. 1,2-Dichloropropane 41. 1,1,2,2-Tetrachloroethane
2. Chloromethane 22. Dibromomethane 42. Brombenzene
3. Vinyl chloride 23. Bromodichloromethane 43. 1,2,3-Trichloropropane
4. Bromomethane 24. cis-1,3-Dichloropropene 44. n-Propylbenzene
5. Chloroethane 25. Toluene 45. 2-Chlorotoluene
6. Trichlorofluoromethane 26. trans-1,3-Dichloropropene 46. 1,3,5-Trimethylbenzene
7. 1,1-Dichloroethene 27. 1,1,2-Trichloroethane 47. 4-Chlorotoluene
8. Methylene chloride 28. Tetrachloroethene 48. tert-Butylbenzene
9. trans-1,2-Dichloroethene 29. 1,3-Dichloropropane 49. 1,2,4-Trimethylbenzene
10. 1,1-Dichloroethane 30. Dibromochloromethane 50. sec-Butylbenzene
11. cis-1,2-Dichloroethene 31. 1,2-Dibromomethane 51. 1,3-Dichlorobenzene
12. 2,2-Dichloropropane 32. Chlorobenzene 52. p-Isopropyltoluene
13. Bromochloromethane 33. 1,1,1,2-Tetrachloroethane 53. 1,4-Dichlorobenzene
14. Chloroform 34. Ethylebenzene 54. n-Butylbenzene
15. 1,1,1-Trichloroethane 35. m-Xylene 55. 1,2-Dichlorobenzene
16. Carbon tetrachloride 36. p-Xylene 56. 1,2-Dibromo-3-chloropropane
17. 1,1-Dichloropropene 37. o-Xylene 57. 1,2,4-Trichlorobenzene
18. Benzene 38. Styrene 58. Hexachlorobutadiene
19. 1,2-Dichloroethane 39. Bromoform 59. Naphthalene
20. Trichloroethene 40. Isopropylbenzene 60. 1,2,3-Trichlorobenzene
Slide 4Dial 1-904-779-4740 for e-Seminar Audio
ANALYTES ADDED TO EPA METHOD 524.2
61. Acetone 73. 1-Chlorbutane
62. Diethyl ether 74. Chloroacetonitrile
63. Methyl iodide 75. 2-Nitropropane
64. Acrylonitrile 76. Methylmethacrylate
65. Allyl Chloride 77. 1,1-Dichloro-2-propanone
66. Carbon disulfide 78. 4-Methyl-2-pentanone (MIBK)
67. Methyl-tert-butyl ether 79. Ethylmethacrylate
68. Propionitrile 80. 2-Hexanone
69. 2-Butanone 81. trans-1,4-Dichloropropanone
70. Methacrylonitrile 82. Pentachloroethane
71. Methyl acrylate 83. Hexachloroethane
72. Tetrahydrofuran 84. Nitrobenzene
Slide 5Dial 1-904-779-4740 for e-Seminar Audio
VOC ANALYSIS SYSTEM
TRAPPURGE
GAS
VENT
DETECTOR
GCCARRIER
GC
PURGE STATE
Slide 6Dial 1-904-779-4740 for e-Seminar Audio
VOC ANALYSIS SYSTEM
GC
DRAIN
TRAP
PURGEGAS
DETECTOR
GCCARRIER
DESORPTION & DRAIN
Slide 7Dial 1-904-779-4740 for e-Seminar Audio
PARAMETERS THAT AFFECT PURGE AND TRAP RESULTS
• Purge Flow
• Traps
• Desorb
• Bake
Slide 8Dial 1-904-779-4740 for e-Seminar Audio
PURGE FLOW
• High
• Low
• Time
• Sparger
Slide 9Dial 1-904-779-4740 for e-Seminar Audio
SPARGER
PURGEGAS
TO TRAP
PURGEGAS
TO TRAP
Slide 10Dial 1-904-779-4740 for e-Seminar Audio
GENERAL RECOVERY PATTERNS FOR VOCFor EPA Methods 502\524
3 6 9 1512 18PURGE TIME (min)
20
21
40
60
80
100
COMPOUNDS 8-38
COMPOUNDS 39-55
COMPOUNDS 56-60
COMPOUNDS 61-84
%RECOVERY
GASES 1-7
Slide 11Dial 1-904-779-4740 for e-Seminar Audio
TRAPSSpecified in EPA Methods
PURGEFLOW TENAX SILICA GEL CHARCOAL
OV1
FREON 12GASES
PURGEFLOW
PURGEFLOW
PURGEFLOW TENAX
VOLATILES &SEMIVOLATILES
TENAX SILICA GEL CHARCOAL
TENAX SILICA GEL
Slide 12Dial 1-904-779-4740 for e-Seminar Audio
TRAPSSpecified in EPA Methods
DESORBFLOW
DESORBFLOW
DESORBFLOW
PURGEFLOW
PURGEFLOW
PURGEFLOW
PURGEFLOW
DESORBFLOWTENAX SILICA GEL CHARCOAL
OV1
FREON 12GASES
TENAX
VOLATILES &SEMIVOLATILES
TENAX SILICA GEL CHARCOAL
TENAX SILICA GEL
Slide 13Dial 1-904-779-4740 for e-Seminar Audio
TRAPSSynthetic Charcoal Traps
PURGEFLOW
PURGEFLOW
PURGEFLOW
CARBOPAK C
CARBOPAK B
CARBOXEN 1000
CARBOXEN 1001
C B 1000 1001
B 1000 1001
C B
Slide 14Dial 1-904-779-4740 for e-Seminar Audio
TRAPS
• Substrates
• Age
• Temperature
Slide 15Dial 1-904-779-4740 for e-Seminar Audio
TRAPSAge
PURGEFLOW TENAX SILICA GEL CHARCOAL
PURGEFLOW
Slide 16Dial 1-904-779-4740 for e-Seminar Audio
TRAPSTemperature
SYNTHETIC CHARCOAL TRAPS
MAX 280°C
TENAX TRAPS
MAX 220°C
Slide 17Dial 1-904-779-4740 for e-Seminar Audio
DESORB
• Carrier Flow
• Water Management
• Time
• Temperature
Slide 18Dial 1-904-779-4740 for e-Seminar Audio
BAKE
• Time
• Temperature
• Purge Flow
Slide 19Dial 1-904-779-4740 for e-Seminar Audio
BAKETime Optimization
SAMPLECONCENTRATOR
CYCLE
TIME
Start
Start Start
Start
GC Analysis GC Stop Oven Cool Eqil
Desorb Bake TrapCooling/Stby
Purge
Ti
Tf
GC CYCLE
TEMPERATURE
Slide 20Dial 1-904-779-4740 for e-Seminar Audio
DETECTORPhotoionization Detector (PID)
• Flow
• Leak Check
• Window
Slide 21Dial 1-904-779-4740 for e-Seminar Audio
DETECTORElectrolytic Conductivity Detector (ELCD)
• Interface to PID
• Reaction Tube
• Reaction Temperature
• Transfer line to cell
Slide 22Dial 1-904-779-4740 for e-Seminar Audio
Rinsing the ELCD Transfer Line
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30
BEFORE
allenv,hp5.0126950a,4,1,0;0
Ti me( minutes)
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30
AFTER
Slide 23Dial 1-904-779-4740 for e-Seminar Audio
DETECTORMass Selective Detector (MSD)
What interfaces are being used?
Slide 24Dial 1-904-779-4740 for e-Seminar Audio
INTERFACES NEEDED FOR THE PURGE AND TRAP SYSTEM
PURGE AND TRAP
COLUMN
COLUMN
DETECTOR
Slide 25Dial 1-904-779-4740 for e-Seminar Audio
GAS FLOWSOptimal Helium Gas Flows
•Trap Desorb
•Megabore Columns
•Capillary Columns
30-40 mL/min
4-5 mL/min
2 mL/min
Slide 26Dial 1-904-779-4740 for e-Seminar Audio
MASS SPECTROMETER DETECTORSpecial Questions
• Is cryo being used?
• Is jet separator being utilized?
• Is the column directly interfaced to the MSD?
• Would you like to use the split injector interface?
Slide 27Dial 1-904-779-4740 for e-Seminar Audio
SPLIT INJECTOR INTERFACEPurge State
Capillary Inlet
Low Dead VolumeUnion
Purge ControlValve
Column HeadPressure ControlTo Detector
1mL/min
SplitVent
Off
Septum PurgeControl
HeCarrier
Vent
Trap25°C
Sample ConcentratorValve Oven
Purge Gas
Total (mass)Flow Controller
Slide 28Dial 1-904-779-4740 for e-Seminar Audio
SPLIT INJECTOR INTERFACEDesorb State
HeCarrier
Trap180°C
Sample ConcentratorValve Oven
Low Dead VolumeUnion
Capillary Inlet
1 mL\min
To Detector
Purge ControlValve
Column HeadPressureControl
SplitVent
Off
Total (mass)Flow Controller
Septom PurgeControl
Slide 29Dial 1-904-779-4740 for e-Seminar Audio
PURGEABLES ANALYSIS WITH DB-VRX
1
23
4
5 6
7
8 910,11
12
13
14
15 1617,18
19
20
2122
23
24
26
2728
29,30
31
32
33
34
3536
37
3839
40,41
42
43,44
45
4647
48
49
50
51
52
53
54
55
56
57
58
59,60
61
62
63,64
65 66
67
68
69
70
71
7273
74
7576
77
78
79
80
81
82
83
84
85
86
87
88
89
25
0.00 5.00 10.00 15.00 20.00 25.00
OI Model 4560Purge: 11 min @ 35 ml/min (Helium)Desorb: 1.0 min @ 220° CTrap: (Tenax, Silica Gel, CMS)
Column: 60m x 0.25mm x 1.4 µm, DB-VRXHP 5890 GC Conditions
Oven: 40°C for 3min8°C/min to 90°C, hold 4 min6°C/min to 200°C hold 5 minCarrier: Helium total flow 30 ml/min,septum purge offcolumn 1.0 ml/min (@ 40°C )HP5971A Mass Selective DetectorScan: 35-300 amu (m/z 44 subtracted)Sample: 40 ppb each analyte in 5 ml water
Slide 30Dial 1-904-779-4740 for e-Seminar Audio
Obstacles To Further Advancement Toward Smaller I.D. Columns
• General dislike for using cryogenic cooling
• Sensitivity
• Sensitivity
• Sensitivity
Slide 31Dial 1-904-779-4740 for e-Seminar Audio
Sensitivity Limitations Using A Split Injector
Smaller diameter columns require higher split ratios in order to maintain efficient trap desorption
60:1 to 300:1 not uncommon
Higher split ratios give reduced sensitivity
Slide 32Dial 1-904-779-4740 for e-Seminar Audio
Sensitivity Limitations Using Split Injector High Split Ratio
Older MSD systems were simply not equipped to handle these high split ratios (i.e. - low sample-to-detector quantities)
The Agilent 5973 and 5973N have opened the door for further advancements toward using small diameter columns with Purge and Trap introduction
Slide 33Dial 1-904-779-4740 for e-Seminar Audio
Improved Sensitivity Of 5973 MSDCompared To 5970, 5971 And 5972
• Reduced source volume
• Improved electron multiplier design
• Optimized quadrapoles
Slide 34Dial 1-904-779-4740 for e-Seminar Audio
Considerations For Optimum Performance Using 5973 MSD For Volatiles
• The smallest I.D. column possible, within reason: - 20m x 0.18 mm I.D., 1.0 um film
• Use as high of a split as possible
• Use Standard Autotune, not BFB Autotune- Set Entrance Lens to zero- Lower the Repeller slightly
Slide 35Dial 1-904-779-4740 for e-Seminar Audio
EPA 8260B Using DB-VRXExcellent Resolution Of 114 VOCs In Under 8 Minutes
7.6 Minutes!
Slide 36Dial 1-904-779-4740 for e-Seminar Audio
Why Would Anyone Want A Volatiles Analysis That Is This Fast?
The Next Advancement In Purge And Trap Analysis• Dual Purge and Trap injection into a single GC/MS instrument
- PT2 from EST- Duet Dual Sampling System from Tekmar
• Allows laboratories to double sample throughput
Slide 37Dial 1-904-779-4740 for e-Seminar Audio
Break Number 1
For Questions and Answers
Press *1 on Your Phone to
Ask a Question
Slide 38Dial 1-904-779-4740 for e-Seminar Audio
COLUMN SELECTION
What are my options when choosing the best column for purge and trap analysis?
Slide 39Dial 1-904-779-4740 for e-Seminar Audio
3 PRIMARY COLUMN CONSIDERATIONS
• Stationary Phase
• Film Thickness
• Length
Slide 40Dial 1-904-779-4740 for e-Seminar Audio
WHAT ARE WE CHROMATOGRAPHING?
• Polars
• Halogenated
• Gases
• Solids
Slide 41Dial 1-904-779-4740 for e-Seminar Audio
CHROMATOGRAPHIC RESOLUTION
= 1/4 -1 k
k +1 n
= relative retention
n = theoretical platesk = partition ratio
αααααααα
αααα
Slide 42Dial 1-904-779-4740 for e-Seminar Audio
RELATIVE RETENTION
Highly selective phase requiredto resolve 60+ analytes
How far apart 2 peaks are
T, phase=ƒα
Slide 43Dial 1-904-779-4740 for e-Seminar Audio
PARTITION RATIO
Thick films needed to obtain adequateretention of most volatile compounds
Decreasing column diameter increasesefficiency but limits capacity andflow rate
d f , r Tc ,= ƒ
Slide 44Dial 1-904-779-4740 for e-Seminar Audio
THEORETICAL PLATES
Column length provides high columnhead pressures for dense desorb band
Column length increases run time
, rcn L= ƒ
Slide 45Dial 1-904-779-4740 for e-Seminar Audio
COLUMN LENGTH
= 1/4 -1 k
k +1Lh
LSince n =
To double resolution you must increase length by 4
h
αα
Slide 46Dial 1-904-779-4740 for e-Seminar Audio
GC ANALYSIS OF EPA METHOD 502.2 ANALYTES
1. Dichlorodifluromethane2. Chloromethane3. Vinyle Chloride4. Bromomethane5. Chloroethane6. Trichlorofluoromethane7. 1,1-Dichloroethene8. Methylene chloride9. trans-1,2-Dichloroethane
10. 1,1-Dichloroethane11. cis-1,2-Dichloropropane12. 2,2-Dichloropropane13. Bromochloromethane14. Chloroform15. 1,1,1,-Trichloroethane16. Carbon tetrachloride17. 1,1-Dichloropropene
35. m-Xylene36. p-Xylene37. o-Xylene38. Styrene39. Bromoform40. Isopropylbenzene41. 1,1,2,2-Tetrachloroethane42. Bromobenzene43. 1,2,3-Trichloropropane44. n-Propylbenzene45. 2-Chlorotoluene46. 1,3,5-Trimethylbenzene47. 4-Chlorotoluene48. tert-Butylbenzene49. 1,2,4-Trimethylbenzene50. sec-Butylbenzene51. 1,3-Dichlorobenzene
52. p-Isopropyltoluene53. 1,4-Dichlorobenzene54. n-Butylbenzene55. 1,2-Dichlorobenzene56. 1,2-Dibromo-3-chloropropane57. 1,2,4-Trichlorobenzene58. Hexachlorobutadiene59. Naphthalene60. 1,2,3-Trichlorobenzene
18. Benzene19. 1,2-Dichloroethane20. Trichloroethene21. 1,2-Dichloropropane22. Dibromomethane23. Bromodichloromethane24. cis-1,3-Dichloropropene25. Toluene26. trans-1,3-Dichloropropane27. 1,1,2-Trichloroethane28. Tetrachloroethene29. 1,3-Dichloropropane30. Dibromoethane31. 1,2-Dibromethane32. Chlorobenzene33. 1,1,1,2-Tetrachloroethane34. Ethylbenzene
Slide 47Dial 1-904-779-4740 for e-Seminar Audio
PURGEABLES ANALYSIS WITH DB-502.2Column: 105m x 0.53mm x 3.0 µm, DB-502.2
GC ConditionsOven: 35°C for 10 min,35-200°C at 4°C/min200°C for 5 minCarrier: Helium at 10 ml/minInjector: Purge and trap (O.I.A 4560)Trap: Tenax/Silica gel/CMSDesorb: 200°C for 0.6 minDetector B: ELCD (O.I.A 4420)
ELCD
1
2 3
4
56
7
8
9 10
11,12
14
1315171619
20
2123
22
2426
27
28
29
30
31
32
33
39
41
43
42
4547
5153
55
56
57
58
60
Slide 48Dial 1-904-779-4740 for e-Seminar Audio
PURGEABLES ANALYSIS WITH DB-624
Column: 75m x 0.53mm x 3.0 µm, DB-624GC Conditions
Oven: 35°C for 8 min35-200°C at 10°C/minCarrier: Helium at 10 ml/minInjector: Purge and trap (O. I. A. 4560)20 ppb per compound in 5 ml waterTrap: Tenax/Silica gel/CMSDesorb: 200°C for 2 minDetector B: ELCD (O.I.A 4420)
ELCD
1
2
3
45
6
7
89 10
11, 12
13
14
15
16, 17
19
20
21
22
23
24
26
27
28, 29
30
3132
33
39
4142, 43
4547
515355
56
57
58
60
Slide 49Dial 1-904-779-4740 for e-Seminar Audio
PURGEABLES ANALYSIS WITH DB-VRXColumn: 75m x 0.45mm x 2.55 µm, DB-VRXGC Conditions
Oven: 35°C for 12 min, 35-60°C at 5°C/min60°C for 1 min, 60-200° at 17°C/min200°C for 3 minCarrier: Helium at 10 ml/min,measured at 35°CInjector: Purge and trap (LSC 2000)20 ppb per compound in 5 ml water
Trap: VOCARB 4000Preheat: 240°CDesorb: 245°CHelium makeup gas at 30 ml/minDetector B: ELCD (O.I.A 4420)
ELCD1
2 3
4
5
67
8
910
11
13
14
121915
17
16
22
21
20
23
24
26
27
29
30
31
2833
32
39
41
43
42
46
47
51
53
55
56
57
58
60
Dial 1-904-779-4740 for e-Seminar Audio
COLUMNS OF CHOICE FOR VOC METHODS
• VOC Methods using 5970,71 & 72 MSD detectors and split injector interface
DB-VRX, 60m x 0.25mm ID x 1.4um film
• VOC Methods using 5973 MSD and split injector interface
DB-624, 20m x 0.18mm ID x 1.0um film
DB-VRX, 20m x 0.18mm ID x 1.0um film
• VOC Methods using standard detectors
DB-VRX, 75m x 0.45mm ID x 2.55um film
DB-624, 75m x 0.45mm ID x 2.55um film
• Contact Technical Support for complete column recommendation
Slide 51Dial 1-904-779-4740 for e-Seminar Audio
BASICS OF PURGE AND TRAP TROUBLESHOOTING
Divide and conquer by elimination
Take nothing for granted-CHECK EVERYTHING
Slide 52Dial 1-904-779-4740 for e-Seminar Audio
BASICS OF PURGE AND TRAP TROUBLESHOOTING
•Begin with "high" concentration standard
•If two detectors are being used, compare same run results
•Discern between chemical and electromechanical problems by turning things off
Slide 53Dial 1-904-779-4740 for e-Seminar Audio
CONCLUSIONSPURGE AND TRAP CONCENTRATOR
• Purge Flow
• Desorb
• Bake
• Trap
Slide 54Dial 1-904-779-4740 for e-Seminar Audio
CONCLUSIONSCOLUMN SELECTION
•Stationary Phase Selectivity
•Film Thickness
•Column Dimensions
Slide 55Dial 1-904-779-4740 for e-Seminar Audio
J&W Scientific Technical Support
800-227-9770 (phone: US & Canada)*
302-993-5304 (phone)*
* Select option 4, then option 2.
916-608-1964 (fax)
www.agilent.com/chem
Slide 56Dial 1-904-779-4740 for e-Seminar Audio
Wrap-up E-Seminar Questions
Thank you for attending Agilent e-Seminars. Our e-Seminar schedule is expanding every
week. Please check our website frequently at: www.agilent.com