Corrections for physical effects in Positron Emission Mammography
description
Transcript of Corrections for physical effects in Positron Emission Mammography
Corrections for physical effects in
Positron Emission Mammography
Nuno C. FerreiraNuno C. FerreiraDep. Biofísica – IBILIDep. Biofísica – IBILI
Fac. Medicina – Univ. CoimbraFac. Medicina – Univ. Coimbra
PEM Workshop – Lisbon, 1-2 July 2002PEM Workshop – Lisbon, 1-2 July 2002
2002, 2002, IBILIIBILI, FMUC, FMUC
Why correct data for physical effects?
To improve To improve image qualityimage quality Image should represent the true activity distributionImage should represent the true activity distribution Corrections improve contrast and tumour detectability, Corrections improve contrast and tumour detectability,
remove artefactsremove artefacts
To preserve To preserve quantificationquantification Useful in comparative studies (e.g. follow-up to Useful in comparative studies (e.g. follow-up to
therapy)therapy)
ScatteredScatteredcoincidences coincidences componentcomponent
AttenuationAttenuationof radiationof radiation
Random Random coincidencescoincidencescomponentcomponent
Detector Detector efficiencyefficiency
effectseffects TrueTruecoincidencescoincidencescomponentcomponent
Data corrections are necessary... ...because the measured projections are not the same as ...because the measured projections are not the same as
the projections assumed during image reconstruction the projections assumed during image reconstruction
ObjectObject(uniform(uniformcylinder)cylinder)
projectionprojectionmeasuredmeasured
projectionprojectionassumedassumed
integral of the activity integral of the activity along the line or tube of along the line or tube of
responseresponse
2002, 2002, IBILIIBILI, FMUC, FMUC
True eventTrue event
Correct data for what physical effects?
ComptonCompton
RandomRandom
Event lost due to Event lost due to AttenuationAttenuation
Event lost dueEvent lost dueto sensitivityto sensitivity
Random coincidencesRandom coincidences
Compton ScatterCompton Scatter
AttenuationAttenuation
Sensitivity variationsSensitivity variations(Normalization correction)(Normalization correction)
2002, 2002, IBILIIBILI, FMUC, FMUC
Correct data for what physical effects?
Random coincidencesRandom coincidences
Compton ScatterCompton Scatter
AttenuationAttenuation
Sensitivity variations Sensitivity variations (Normalization correction)(Normalization correction)
RandomRandom
DELAYEDcoincidences = Estimate(RANDOMS)
Randoms Correction Using Delayed Channel
Detector 2 + delay
delay
time
Detector 1
Detector 2
= TRUES + RANDOMSPROMPT
coincidences
Estimate(TRUES) = (TRUES+RANDOMS) – Estimate(RANDOMS)
PROMPTS DELAYED“TRUES” –=
2002, 2002, IBILIIBILI, FMUC, FMUC
Random coincidences correction To improve the estimation of random To improve the estimation of random
coincidences:coincidences:
Option 1: Averaging of estimations obtained with various Option 1: Averaging of estimations obtained with various delaysdelays
Option 2: Measurement of the single events: Option 2: Measurement of the single events:
Detector 1Detector 1
Detector 2Detector 2
SS11
SS22
RR1212
Width of Width of coincidence coincidence
windowwindow
RR1212 = 2T . S = 2T . S11 . S . S22
RandomsRandomsraterate
Singles rate Singles rate in detectors in detectors
1 and 21 and 2
Random coincidences correction Randoms: PEM vs. PETRandoms: PEM vs. PET
PEM has less activity in the FOV PEM has less activity in the FOV less randoms than PET less randoms than PET In PEM most activity is outside the FOV and there are no In PEM most activity is outside the FOV and there are no
end-shields end-shields larger % of randoms from outside the FOV larger % of randoms from outside the FOV
ClearPEM: ClearPEM: LuAP is fast LuAP is fast very short coinc. window very short coinc. window less randoms less randoms
PETDetectors
End-shields
PEMDetectors
BrainBrain
MyocardiumMyocardium
BladderBladder
LiverLiver
TorsoTorso
Activity outside the FOV
Source: Raylman et al., IEEE Trans Nucl Sci (2001), 48(3):913-923 Source: Raylman et al., IEEE Trans Nucl Sci (2001), 48(3):913-923
MyocardiumMyocardium(9%)(9%)
TorsoTorso(82%)(82%)
LiverLiver(7%)(7%) BladderBladder
(1%)(1%)
PEMPEMDetectorsDetectors
Contributions of different sources to the total random coincidences from outside
the FOV BrainBrain(1%)(1%)
2002, 2002, IBILIIBILI, FMUC, FMUC
Correct data for what physical effects?
Random coincidencesRandom coincidences
Compton ScatterCompton Scatter
AttenuationAttenuation
Sensitivity variations Sensitivity variations (Normalization correction)(Normalization correction)
ComptonCompton
2002, 2002, IBILIIBILI, FMUC, FMUC
Effect of scatter Main effect:Main effect:
Reduces image contrast (target-to-Reduces image contrast (target-to-background activity ratio)background activity ratio)
Affects tumour detectabilityAffects tumour detectability
Other effects:Other effects: Reduces quantification accuracyReduces quantification accuracy Slight reduction of the total number of useful Slight reduction of the total number of useful
events detectedevents detected Slight degradation of spatial image resolutionSlight degradation of spatial image resolution
Scatter correction PEM vs. PETPEM vs. PET
Relative to PET, scatter is less important in PEM...Relative to PET, scatter is less important in PEM... ...but influence of activity from outside the FOV is more ...but influence of activity from outside the FOV is more
importantimportant Activity is mostly outside the FOV (heart, chest)Activity is mostly outside the FOV (heart, chest) No side shielding in PEMNo side shielding in PEM
Correction methods:Correction methods: Proposed by other groups for PEM: Proposed by other groups for PEM:
Monte-Carlo simulation Monte-Carlo simulation [Holdsworth et al., IEEE TNS, February 2002][Holdsworth et al., IEEE TNS, February 2002] Convolution-subtraction (Bergström)Convolution-subtraction (Bergström) [Raylman et al., IEEE TNS, [Raylman et al., IEEE TNS,
June 2001]June 2001] Proposed by our group for PET: Proposed by our group for PET:
Hybrid energy-based correction Hybrid energy-based correction [Ferreira et al., Phys Med Biol, May [Ferreira et al., Phys Med Biol, May 2002]2002]
Advantage: takes scatter from outside the FOV into accountAdvantage: takes scatter from outside the FOV into account
2002, 2002, IBILIIBILI, FMUC, FMUC
Correct data for what physical effects?
Random coincidencesRandom coincidences
Compton ScatterCompton Scatter
AttenuationAttenuation
Sensitivity variations Sensitivity variations (Normalization correction)(Normalization correction)
Event lost due to Event lost due to AttenuationAttenuation
Attenuation correction
Attenuation is more important in PET than in PEMAttenuation is more important in PET than in PEM Attenuation needs to be measured in PET, but probably not in PEM Attenuation needs to be measured in PET, but probably not in PEM
attenuation coefficient @ 511keV is ~ constant for the breastattenuation coefficient @ 511keV is ~ constant for the breast use of consistency conditions may be sufficiently accurateuse of consistency conditions may be sufficiently accurate
The amount of attenuation is mainly dictated by the dimensions of the The amount of attenuation is mainly dictated by the dimensions of the human body. human body.
When a radiation beam When a radiation beam penetrates a material with penetrates a material with thickness x and linear thickness x and linear attenuation coefficient attenuation coefficient , it is , it is attenuated by a factor:attenuated by a factor:ee--xxFor tissue, For tissue, =0.096 cm=0.096 cm-1-1 @ @ 511 keV511 keV
x=15 cm x=15 cm 24% 24%x= 5 cm x= 5 cm 62% 62%
x=50 cm x=50 cm 1% 1%
% of true % of true (unscattered) (unscattered) coincidences coincidences
detected:detected:
2002, 2002, IBILIIBILI, FMUC, FMUC
Attenuation correction
Source: Meikle, Bailey et al., Source: Meikle, Bailey et al., J NucJ Nuc MedMed (1995) (1995); 36:1680-1688; 36:1680-1688
Effect of Attenuation Correction (PET)Effect of Attenuation Correction (PET)
2002, 2002, IBILIIBILI, FMUC, FMUC
Correct data for what physical effects?
Random coincidencesRandom coincidences
Compton ScatterCompton Scatter
AttenuationAttenuation
Sensitivity variations Sensitivity variations (Normalization correction)(Normalization correction)
Includes various Includes various corrections, mainly corrections, mainly dependent of the dependent of the scanner:scanner:
- geometrical effectsgeometrical effects
- intrinsic efficiencies intrinsic efficiencies of individual of individual detectorsdetectors
- interferences interferences between individual between individual detectorsdetectors
- deadtime deadtime
Results from bibliographySource: Raylman et al., IEEE Trans Nucl Sci (2001), 48(3):913-923 Source: Raylman et al., IEEE Trans Nucl Sci (2001), 48(3):913-923
FDGFDG
Water (Cold)Water (Cold)
NormalizedNormalizedRaw imageRaw image(no corrections)(no corrections)
NormalizedNormalized+ Random correction+ Random correction
NormalizedNormalized+ Random correction+ Random correction+ Scatter correction+ Scatter correction
Scatter Scatter phantophantom m
uncorrecteduncorrectedcorrectedcorrected
Results from bibliography (2)Source: Raylman et al., IEEE Trans Nucl Sci (2001), 48(3):913-923 Source: Raylman et al., IEEE Trans Nucl Sci (2001), 48(3):913-923
(?)(?)
BreastBreastphantophantom m
Raw imageRaw image(no corrections)(no corrections) NormalizedNormalized NormalizedNormalized
+ Random correction+ Random correctionNormalizedNormalized
+ Random correction+ Random correction+ Scatter correction+ Scatter correction
2002, 2002, IBILIIBILI, FMUC, FMUC
Summary Specificities of PEM Specificities of PEM
Corrections:Corrections: Randoms correctionRandoms correction
Less randoms than PETLess randoms than PET Larger % of randoms from Larger % of randoms from
outside the FOV than PEToutside the FOV than PET Same methods as in PET Same methods as in PET
Scatter correctionScatter correction Less scatter than PETLess scatter than PET More scatter from outside the More scatter from outside the
FOV than PETFOV than PET Hybrid energy-based methodsHybrid energy-based methods
Attenuation correctionAttenuation correction Less attenuation than PETLess attenuation than PET Attenuation in breast is more Attenuation in breast is more
homogeneous than in PEThomogeneous than in PET No transmission measurement No transmission measurement
(but attenuation correction (but attenuation correction seems feasible)seems feasible)
Normalization correctionNormalization correction Correction is scanner Correction is scanner
dependent, as in PET dependent, as in PET
Specificities Specificities of PEM:of PEM:
Large influence of Large influence of activity outside the activity outside the FOV FOV
Little activity in Little activity in the FOVthe FOV
Most activity is Most activity is outside the FOVoutside the FOV
Side-shielding is Side-shielding is not possiblenot possible
Small FOV, objectSmall FOV, object Less attenuationLess attenuation
High sensitivity High sensitivity Large solid angleLarge solid angle Less attenuationLess attenuation
High resolutionHigh resolution
Indications for Indications for PEM camera PEM camera design:design:
Randoms correctionRandoms correction Smallest possible Smallest possible
coinc. window (LuAP is coinc. window (LuAP is OK)OK)
Capability to count Capability to count singles/channel...singles/channel...
...or count randoms ...or count randoms with multiple delayswith multiple delays
Scatter correctionScatter correction Good energy Good energy
information (~256 information (~256 energy channels)energy channels)
List-mode (with energy List-mode (with energy information)information)
Normalization correctionNormalization correction Information about Information about
single events (for single events (for deadtime correction)deadtime correction)