Clinica Chimica Acta Volume 412 Issue 9-10 2011 [Doi 10.1016%2Fj.cca.2011.01.010] Paola Brunori;...

download Clinica Chimica Acta Volume 412 Issue 9-10 2011 [Doi 10.1016%2Fj.cca.2011.01.010] Paola Brunori; Piergiorgio Masi; Luigi Faggiani; Luciano Villani -- Evaluation of Bilirubin Concentration

of 4

Transcript of Clinica Chimica Acta Volume 412 Issue 9-10 2011 [Doi 10.1016%2Fj.cca.2011.01.010] Paola Brunori;...

  • 8/13/2019 Clinica Chimica Acta Volume 412 Issue 9-10 2011 [Doi 10.1016%2Fj.cca.2011.01.010] Paola Brunori; Piergiorgio M

    1/4

    Evaluation of bilirubin concentration in hemolysed samples, is it really impossible?

    The altitude-curve cartography approach to interfered assays

    Paola Brunori a,, Piergiorgio Masi a, Luigi Faggiani a, Luciano Villani a, Michele Tronchin b, Claudio Galli b,Clarissa Laube a, Antonella Leoni a, Maila Demi a, Antonio La Gioia a

    a Azienda USL 6 Livorno, Italyb Abbott Diagnostic Division, Roma, Italy

    a b s t r a c ta r t i c l e i n f o

    Article history:

    Received 18 November 2010

    Received in revised form 6 January 2011

    Accepted 6 January 2011

    Available online 14 January 2011

    Keywords:

    Bilirubin

    Icterus

    Hemolysis

    Interference

    Neonatal jaundice

    Background: Neonatal jaundice might lead to severe clinical consequences. Measurement of bilirubin in

    samples is interfered by hemolysis. Over a method-depending cut-off value of measured hemolysis, bilirubin

    value is not accepted and a new sample is required for evaluation although this is not always possible,

    especially with newborns and cachectic oncological patients. When usage of different methods, less prone to

    interferences, is not feasible an alternative recovery method for analytical signicance of rejected data might

    help clinicians to take appropriate decisions.

    Methods: We studied the effects of hemolysis over total bilirubin measurement, comparing hemolysis-

    interfered bilirubinmeasurement with the non-interfered value. Interference curves wereextrapolated overa

    wide range of bilirubin (030 mg/mL) and hemolysis (H Index 01100).

    Results: Interference altitude curves werecalculated and plotted.A bimodal acceptance table was calculated.

    Non-interfered bilirubin of given samples was calculated, by linear interpolation between the nearest lower

    and upper interference curves.

    Conclusions: Rejection of interference-sensitive data from hemolysed samples for every method should be

    based not upon the interferent concentration but upon a more complex algorithm based upon the

    concentration-dependent bimodal interaction between the interfered analyte and the measured interferent.

    The altitude-curve cartography approach to interfered assays may help laboratories to build up their ownmethod-dependent algorithm and to improve the trueness of their data by choosing a cut-off value different

    from the one (10% interference) proposed by manufacturers.

    When re-sampling or an alternative method is not available the altitude-curve cartography approach might

    also represent an alternative recovery method for analytical signicance of rejected data.

    2011 Elsevier B.V. All rights reserved.

    1. Introduction

    Hemolysis (H) is the most common cause of blood sample

    inadequacy and interferes in different ways with several assays [1].

    As far as bilirubin (B) measurement is concerned, the interference of

    hemolysis is caused by two kinds of events:

    1. Physical interference (light absorption of the heme group) and

    chemical interference (inhibition of diazotization).

    2. Physiological catabolism of plasma/serum heme groups into

    bilirubin.

    When a hemolysed sample arrives in the laboratory physiological

    generation of bilirubin from hemoglobin may be easily prevented or,

    at least, limited e.g. by immediate measurement of bilirubin, while

    physicalchemical interferences are already present and may only be

    evaluated in order to accept (or reject) the results from the analyzer.

    Evaluation of hemolysis might be performed in a visual, qualitative

    way but also in a semiquantitative way using interference assays that

    are normally available in modern clinical chemistry analyzers.

    Interference assays are to be preferred for better reliability [2]. For

    every bilirubin methoda cut-off value for hemolysis(H) is established

    Clinica Chimica Acta 412 (2011) 774777

    Abbreviations:Bil, bilirubin (total)total bilirubin value; Bil, value of Bil in a non-

    hemolytic sample; BilHmeas, measured value of Bil in a hemolytic sample; Bil index,

    icterus index value; H, hemolysishemolysis index value; Hhigh, calculated value of H

    for the nearest higher interference curve; Hlow, calculated value of H for the nearest

    lower interference curve; Hmeas, measured value of H in a hemolytic sample; I%,

    interference; I%chosen, chosen interference level for altitude curves; I%low, nearest

    lower interference curve; I%high, nearest higher interference curve; l%mis, measured

    interference; POCT, point of care testing.

    Corresponding author. Laboratorio Ospedale Villamarina, via Forlanini 22, 57025

    Piombino (LI), Italy.

    E-mail address:[email protected](P. Brunori).

    0009-8981/$ see front matter 2011 Elsevier B.V. All rights reserved.

    doi:10.1016/j.cca.2011.01.010

    Contents lists available at ScienceDirect

    Clinica Chimica Acta

    j o u r n a l h o m e p a g e : w w w. e l s ev i e r. c o m / l o c a t e / c l i n c h i m

    http://dx.doi.org/10.1016/j.cca.2011.01.010http://dx.doi.org/10.1016/j.cca.2011.01.010http://dx.doi.org/10.1016/j.cca.2011.01.010mailto:[email protected]://dx.doi.org/10.1016/j.cca.2011.01.010http://www.sciencedirect.com/science/journal/00098981http://www.sciencedirect.com/science/journal/00098981http://dx.doi.org/10.1016/j.cca.2011.01.010mailto:[email protected]://dx.doi.org/10.1016/j.cca.2011.01.010
  • 8/13/2019 Clinica Chimica Acta Volume 412 Issue 9-10 2011 [Doi 10.1016%2Fj.cca.2011.01.010] Paola Brunori; Piergiorgio M

    2/4

    by the manufacturer's insert sheet [3] or laboratory experience.

    Therefore the Bil value of a given sample might be rejected if the H

    value exceeds this threshold; in this case a new sample without

    Hemolysis should be obtained for a correct measurement of

    bilirubin. In most cases this is not a great problem, but for newborns

    a new sample might not be easy to obtain since vein diameter in

    babies is very small, the newborn total blood volume may not allow a

    second sampling and, nally, newborn spontaneous hemolysis might

    be the causative agent of high bilirubin concentration[4].The measurement of bilirubin in neonatology is a critical point,

    since high concentration of bilirubinat birth maygenerate permanent

    and severe encephalopathy (kernicterus). Classical kernicterus is a

    well-described clinical tetrad of (i) abnormal motor control, move-

    ments and muscle tone, (ii) an auditory processing disturbance with

    or without hearing loss, (iii) oculomotor impairments, especially

    impairment of upward vertical gaze, and (iv) dysplasia of the enamel

    of deciduous teeth[5].

    Measurements must be accurate and with a short TAT since

    evolution of bilirubin in newborns may be very fast, and the clinical

    protocol must also take into account the patient's age (in hours)and B

    increment speed. The clinicians, therefore, are asking for a different

    solution in this context[6].

    A second source of problems is that, conventionally, the interfer-

    ence cutoffpoint is chosen when interference is equal to 10%, which is

    very close, but greater, than 9.8%, that represents the European

    inaccuracy goal for bilirubin [7] without considering any other

    possible source of inaccuracy.

    Therefore we decided to investigate in depth the inuence of

    hemolysis on bilirubin measurements in order to propose a different

    approach.

    2. Methods

    Pools of sera with different values of bilirubin(Bil) were generated

    from blood donors' samples without relevant hemolysis (inclusion

    criteria was a free hemoglobin concentration b20 mg/dL) and were

    divided into quotes and frozen until usage. Immediately before

    testing, red blood cells from donors without increased Bil wereseparated from plasma, manually hemolysed and centrifuged. The

    hemolysed supernatant was used to generate interfering samples by

    dilution with a phosphate-buffered saline solution. Samples for

    measurement were generated adding 1 part of one interfering sample

    and 9 part of one icteric sample. Final free hemoglobin concentration

    in the above-mentioned samples ranged from 35 to 1050 mg/dL. Non-

    interfered samples (nal free hemoglobin concentration ranged from

    2 to 12 mg/dL) were generated by adding 1 part of saline solution

    instead of the interfering sample. For each sample we measured the

    interference index (Hmeas) and total bilirubin (BilHmeas) with the

    reagent (total bilirubin LN 8G6220, Abbott Diagnostics, Chicago, IL)

    and instruments (ARCHITECT ci8200, Abbott Diagnostics, Chicago, IL)

    routinely used in our laboratory. Each point was measured at least

    twice with two different instruments. Subsequent data processingwas performed using a standard worksheet (MS Excel) program. Data

    were compared with non-interfered bilirubin (Bil) to calculate

    interference (I%) as

    I% = BilHmeas BilB

    = BilB 100:

    The theoretical values of I%=5%, 10%, 15%, 20%, 25%,

    30%,35%,40%,45%,50%,55%,60%,65%, 70%, 75%, 80%,

    85%, 90% and95%were chosen(I%chosen) forcalculatingthe valuesof

    experimental interference altitude curves. Data for I%chosen were

    calculate from H (HI%) and Bil (BilI%) for any chosen interference

    altitudecurve between two experimental points (BilHmeas1; Hmeas1)

    and (BilHmeas2; Hmeas2) whose I% were respectively lower (I%meas1)

    and greater (I%meas2) than the chosen I% (I%chosen), using classical

    interpolation formulas

    HI% = Hmeas1 + I%chosen I%meas1 = I%meas2 I%meas1 Hmeas2 Hmeas1

    BilI% = BilHmeas1 + I%chosen I%meas1 = I%meas2 I%meas1

    BilHmeas2 BilHmeas1 :

    Later, second order regression curves with the formula

    H = A Bil2

    + B Bil + C

    were calculated over (HI%; BilI%) points of the chosen interference

    altitudecurves.

    After completing this phase a worksheet was build to plot into

    altitude curves any experimental (BilHmeas; Hmeas). In this worksheet

    H values in the interference curves are calculated for the given value

    of (BilHmeas) and compared to (Hmeas). The H value of the nearest

    lower altitude curve of (BilHmeas; Hmeas) is dened as Hlowand the H

    value of the nearest higher altitude curve of (BilHmeas; Hmeas) is

    dened as Hhigh. This procedure allows to calculate the position of

    Hmeas in the line (Hlow: Hhigh) and therefore to evaluate the

    interference according to the formula

    I% = I%low + Hmeas Hlow = Hhigh Hlow

    I%high I%low

    :

    Therefore Bil is calculated with the formula

    BilBcalc = Bilmeas= 1 + I% :

    Four fresh samples (verication samples), with signicant Bil

    values were treated like the former ones to test the worksheet and

    compared their measured Bil (1 part of saline instead of interference

    sample) with the calculated Bil.

    3. Results

    Fig. 1A shows average bulk decay curves of B versus H. Visual

    analysis of bulk decay curves was used to evaluate the coverage of the

    measured area and eventually to check the single experimental points

    in order to replicate eventual non-tting points before proceeding

    with calculations.

    Fig. 1.Bulk bilirubin measurements (A: mg/dL) and interferences (B: %) versus index

    measurements (H).

    775P. Brunori et al. / Clinica Chimica Acta 412 (2011) 774777

    http://www.westgard.com/europe.htmhttp://www.westgard.com/europe.htm
  • 8/13/2019 Clinica Chimica Acta Volume 412 Issue 9-10 2011 [Doi 10.1016%2Fj.cca.2011.01.010] Paola Brunori; Piergiorgio M

    3/4

    Fig. 1B shows the interaction of I% and H for the same data. The

    manufacturer indication for interference cutoff point (absolute value

    of interference 10% which correspond to I%10%) indicates an H

    value of 31. The decay trends of Bil=0.99 mg/dL and Bil=1.73 mg/

    dL when H=31 show an absolute value of interference greater than

    10% (17% and 13% respectively) while two other ones show absolute

    values of interference between 5% and 10% (8.5%, and 5% for

    Bil= 4.46 mg/dL, Bil= 5.94 mg/dL, respectively), and negligible

    (b

    5%) for the other ones (Bil=8.43 mg/dL, Bil=12.62 mg/dL andBil= 29.98 mg/dL).

    I% =10% was found at H values of 19, 24, 35, 49, 65, 84, and 109

    f or Bi l = 0.99 mg/dL, Bi l = 1.73 mg/dL, Bi l = 4.46 mg/dL,

    Bil =5.94 mg/dL, Bil =8.43 mg/dL, Bil=12.62 mg/dL, and

    Bil=29.98 mg/dL, respectively.

    Fig. 2 shows calculated altitude data points obtained with

    previous data interpolation as described above, with their second-

    order tting polynomial curves. Calculated points for curves 80%,

    85%, 90% and 95% were less than 4 for each curve, therefore the

    second-ordertting polynomial curve for this data was forced to be a

    rst-order polynomial curve.

    Square correlation coefcient (R2) ranged from 0.977 for

    I%chosen=5%, to 0.999 for I%chosen=80%. Fit parameters were

    used to generate data pairs (BilTmisversus Hmis) inTable 1for three

    chosen altitudecurves: 2.5%, 5%, 7.5% and 10% respectively.

    Fit parameters from 5%, 10%, 15%, 20%, 25%, 30%,

    35%, 40%, 45%, 50%, 55%, 60%, 65%, 70%, 75%, 80%, 85%,

    90%and95%curves respectivelywere used to generate an altitude

    plot, as shown inFig. 3, in a worksheet le.Table 2shows data of four

    verication samples that were treated as described above, and were

    used to test the worksheet le. Baseline H (samples with saline)

    showed free hemoglobin b20 mg/dL for all samples except #4 (free

    hemoglobin=66 mg/dL). InTable 2the icterus index value (Bilindex)

    is also shown as a comparison.

    Data calculation was not possible for sample 1C (Bilmisb0.1 mg/dL).

    Bilcalc of all other samples were in agreement with the measured Bil. It

    should be considered that, a not-null value of index is also present in

    samples with saline. Therefore the correction formula was used also

    with Bilmeasto generate a corrected Bilmeas. Comparison of all Bilcalcversus corrected Bilmeas showed a better agreement than the

    comparison versus non-corrected Bilmeas. Average bias of Bilcalc versus

    corrected Bilmeas was 37% (50% versus non-corrected Bilmeas) for

    sample #1.Average bias of Bilcalc versus corrected Bilmeas was 9% (11%

    versus non-corrected Bilmeas) for sample #2. Average bias of Bilcalcversus corrected Bilmeas was 2% (4% versus non-corrected Bilmeas)

    for sample #3. Average bias of Bilcalc versus corrected Bilmeas was1%

    (8% versus non-corrected Bilmeas) for sample #4.

    Fig. 4 shows a bias and precision plot from the data of the

    verication samples.

    4. Discussion

    Bilirubin measurement must identify the infants at risk for

    permanent brain damages (kernicterus) and other pathological

    conditions [6]. It is of concern that many bilirubin automated kit

    methods suffer from hemolysis interference. Since hemolysis may be

    one of the causes of high bilirubin concentration, many methods often

    fail to measure bilirubin.

    Interference is not only a matter of hemoglobin concentration but

    depends also upon bilirubin concentration since the same value of H

    produces different values of I% upon different bilirubin concentrationsand therefore it is a double-factor event. The interaction is

    reproducible, therefore a mathematical approach might expand

    measuring possibilities beyond their traditional limits. It is clear that

    evaluation of bilirubin from the icterus index is a possible approach,

    but is not recommended[3] since it may generate untrue, although

    precise, measurement. A different approach might be non-invasive

    patient-side (POCT) measurement of transcutaneous bilirubin with

    multiwavelength spectral reectance analysis[8].

    Multiwavelength spectral reectance analysis, is methodologically

    similar to the above-mentioned index method, without the need of

    blood sampling, and therefore without needle-induced hemolysis.

    Fig. 2.

    Altitude-curve

    points with regression curves.

    Table 1

    Bimodal acceptance table for data rejection.

    Bil Tmeas(mg/dL) H index Hb (mg/dL) values for interference curves

    2.5% 5.0% 7.5% 10%

    0 1 3 3 3

    1 3 7 9 12

    2 5 10 16 21

    3 7 14 21 29

    4 8 17 27 37

    5 10 20 32 45

    6 12 23 37 52

    7 13 26 42 58

    8 14 29 47 65

    9 16 32 51 70

    11 17 34 55 76

    12 18 37 59 81

    13 19 39 62 86

    14 21 41 66 90

    15 22 43 69 94

    16 23 45 71 98

    17 23 47 74 101

    18 24 49 76 103

    19 25 50 78 106

    20 26 51 80 108

    21 27 54 82 111

    22 27 55 83 111

    23 28 56 84 11224 28 57 84 112

    25 29 57 84 111

    Fig. 3. Usageof altitudecurves: a patient () withBilmeas=15.0 mg/dLand Hmeas=600

    is evaluated to be affected by a 39% interference therefore Bilcalc=24.5 mg/dL.

    776 P. Brunori et al. / Clinica Chimica Acta 412 (2011) 774777

    http://localhost/var/www/apps/conversion/tmp/scratch_2/image%20of%20Fig.%E0%B3%80http://localhost/var/www/apps/conversion/tmp/scratch_2/image%20of%20Fig.%E0%B2%80
  • 8/13/2019 Clinica Chimica Acta Volume 412 Issue 9-10 2011 [Doi 10.1016%2Fj.cca.2011.01.010] Paola Brunori; Piergiorgio M

    4/4

    Nevertheless, transcutaneous bilirubin measurement might overesti-

    mate blood bilirubin[10], when signicant skin bilirubin is present.

    The altitude curve approach produces a reliable estimation of

    interference and Bil. This approach may improve traditional data

    rejectiondening thehemolysis cutoffpoint no more as a singlepoint,

    but asa function ofBilHmeas (Table 1). Indications fordata rejectionare

    traditionally indicated by manufacturers when the absolute value of

    interference is equal to 10%, which is very close, but greater, than 9.8%,

    that represents the European inaccuracy goal for bilirubin[7]without

    considering any other possible source of inaccuracy. With the

    altitude curve approach more restrictive cutoffs are available,

    therefore in Table 1 not only the traditional minimal (10%)

    curve is indicated but also the 2.5% (desirable as of the

    European total inaccuracy goal), the 5% (acceptable as of the

    European total inaccuracy goal) and 7.5% (permissiveas of the

    European total inaccuracy goal) curves respectively.

    Interference plots like inFigs. 2and3or a commercial worksheetlike inTable 2may help evaluate Bil from BilHmeasand Hmeas if the

    laboratory is more condent with data calculation.

    The altitude curveapproach has its own limitations, rst of all it

    may estimate chemicalphysical interference but not the physiolog-

    ical after-sampling production of bilirubin from hemolised samples

    [1]. Furthermore, results of this work are limited to the method in use

    in our laboratory, but other laboratories that employ different

    methods or instruments may use the same approach to build their

    own altitude curves. While the choice of different reagents less

    prone to interference by hemolysis will be of more help in specialized

    settings[9], this method will be useful in most general labs that serve

    both in- and outpatients, as it is common in Italy and generally in

    Europe; in these settings it is difcult to envision theusage of different

    reagents according to the population assayed. If reagents more

    sensitive to hemolysis are in use, our method may be useful, and

    though it is validated only for the Architectreagent, it may be adapted

    also to different ones. The altitude-curve cartography approach may

    give new sights and solutions to the problem of interferences,

    especially where a new sampling is not possible or convenient, and

    the interfered analyte should be quickly determined to drive

    important therapeutic decisions, like bilirubin in the neonatology

    eld.

    References

    [1] BradyJ, O'Leary N. Interference due to hemolysis inroutine photometric analysisa survey. Ann Clin Biochem 1998;35:12834.

    [2] Glick MR, Ryder KW, Glick SJ, et al. Unreliable visual estimation of the incidenceand amount of turbidity, hemolysis, and icterus in serum from hospitalizedpatients. Clin Chem 1989;35:8379.

    [3] Manufacturer's insert sheetAbbott Laboratories 8G62.[4] Cohen RS, Wong RJ, Stevenson DK. Understanding neonatal jaundice a perspective

    on causation. Pediatr neonatol 2010;51:1438.[5] Shapiro SM. Chronic bilirubin encephalopathy: diagnosis and outcome. Semin

    Fetal Neonatal Med 2010;15:15763.[6] Kirk JM. Neonatal jaundice: a critical review of the role and practice of bilirubin

    analysis. Ann Clin Biochem 2008;45:45262.[7] http://www.westgard.com/europe.htm.[8] Bhutani VK, Gourley GR, Adler S, Kreamer B, Dalin C, Johnson LH. Noninvasive

    measurement of total serum bilirubin in a multiracial predischarge newbornpopulation to assess the risk of severe hyperbilirubinemia. Pediatrics 2000;106:e17.

    [9] Algeciras-Schimnich A, Cook WJ, Milz TC, Saenger AK, Karon BS. Evaluation ofhemoglobin interference in capillary heel-stick samples collected for determina-tion of neonatal bilirubin. Clin Biochem 2007;40:13116.

    [10] Ahmed M, Mostafa S, Fisher G, Reynolds TM. Comparison between transcutaneousbilirubinometry and total serum bilirubin measurements in preterm infantsb35 weeks gestation. Ann Clin Biochem 2010;47:727.

    Table 2

    Effect of addition of saline or hemolysed sera (A, B, C, D, E, F and G) over bilirubin

    measurement. Uncorrected Bil is Bilmeasof saline-added samples, and corrected Bil is

    Bilcalcof saline-added samples.

    Sample Addition

    (10%) of

    Bilmeas(mg/dL)

    Hmeas

    (mg/dL of Hb)

    Calculated

    interference

    Bilcalc(mg/dL)

    Bilindex(mg/dL)

    #1 Saline 0.9 8 9% 1.0 1.5

    A 0.6 47 50% 1.2 1.5

    B 0.3 85 80% 1.5 1.5

    C b0.1 170 N. a. N. a. N. a.#2 Saline 2.9 3 1% 2.9 2.6

    A 2.4 42 20% 3.0 2.7

    B 1.9 88 44% 3.4 2.6

    C 1.1 168 66% 3.2 2.6

    #3 Saline 6.7 12 3% 6.9 5.8

    B 5.4 94 21% 6.9 5.8

    C 4.4 173 39% 7.2 5.8

    E 2.7 311 61% 6.9 5.7

    #4 Saline 15.7 66 7% 16.9 15.7

    D 13.5 223 23% 17.5 15.6

    F 10.8 420 38% 17.6 15.4

    G 7.7 699 51% 15.7 15.3

    Fig. 4.Bias and precision plot from the data of the verication samples.

    777P. Brunori et al. / Clinica Chimica Acta 412 (2011) 774777

    http://localhost/var/www/apps/conversion/tmp/scratch_2/image%20of%20Fig.%E0%B4%80