Chapter 7: Gradient Echo Imaging Methods

18
Chapter 7: Gradient Echo Imaging Methods Mark D. Herbst, MD, PhD

description

Chapter 7: Gradient Echo Imaging Methods. Mark D. Herbst, MD, PhD. Two Main Imaging Methods. Spin Echo – uses 180 degree refocusing RF pulse to refocus spins and form an echo that is captured and put into k-space Gradient Echo – uses gradient pulse to form the echo - PowerPoint PPT Presentation

Transcript of Chapter 7: Gradient Echo Imaging Methods

Page 1: Chapter 7: Gradient Echo Imaging Methods

Chapter 7: Gradient Echo Imaging Methods

Mark D. Herbst, MD, PhD

Page 2: Chapter 7: Gradient Echo Imaging Methods

Two Main Imaging Methods

• Spin Echo – uses 180 degree refocusing RF pulse to refocus spins and form an echo that is captured and put into k-space

• Gradient Echo – uses gradient pulse to form the echo

• Gradient Echo – abbreviated GE, or GRE for Gradient Recalled Echo, or FFE for Fast Field Echo, or SAGE (small angle GE)

Page 3: Chapter 7: Gradient Echo Imaging Methods

Advantages of GE Methods

• Faster than SE (spin echo) because of short TRs and short TEs

• Shows flowing blood as bright good for MRA• Good for functional MRI• Good for subtle hemorrhage (microhemorrhage

technique or magnetic susceptibility technique)• Can be combined with SE to produce very fast

images (GRASE=gradient and spin echo)

Page 4: Chapter 7: Gradient Echo Imaging Methods

Disadvantages of GE Methods

• Can have too much metal artifact• GE gives T2*-weighted images (“T-two-

star”) instead of T2WI• Need to be aware of fat and water being

in-phase or out-of-phase at interfaces between water and fat

Page 5: Chapter 7: Gradient Echo Imaging Methods

Spin Echo

Page 6: Chapter 7: Gradient Echo Imaging Methods

Gradient Echo

Page 7: Chapter 7: Gradient Echo Imaging Methods

Gradient Echo

Page 8: Chapter 7: Gradient Echo Imaging Methods

Spoiled Gradient Echo

• “Spoiling” refers to the destruction of any T2 contrast in the image, leaving you with a fast way of getting a T1WI using a gradient technique. The “spoiler pulse” completely dephases the spins in the xy plane, removing any evidence of T2 or T2* contrast.

Page 9: Chapter 7: Gradient Echo Imaging Methods

GRASS and Spoiled GRASS

Page 10: Chapter 7: Gradient Echo Imaging Methods

Spin Echo

Page 11: Chapter 7: Gradient Echo Imaging Methods

Effect of Flip Angle on Signal

Page 12: Chapter 7: Gradient Echo Imaging Methods

Effect of Flip Angle on Net Magnetization Vector

Page 13: Chapter 7: Gradient Echo Imaging Methods

Effect of Flip Angle on Longitudinal and Transverse Magnetization

Page 14: Chapter 7: Gradient Echo Imaging Methods

Effect of Flip Angle on Contrast

Page 15: Chapter 7: Gradient Echo Imaging Methods

Effect of TR on M

Page 16: Chapter 7: Gradient Echo Imaging Methods

GRASE

Page 17: Chapter 7: Gradient Echo Imaging Methods

Spin Echo Planar Imaging (EPI)

Page 18: Chapter 7: Gradient Echo Imaging Methods

Gradient Echo Planar Imaging (EPI)